XPS: Issues with Data Acquisition and Data Processing

Similar documents
The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

Quantification of JEOL XPS Spectra from SpecSurf

CasaXPS Manual Rev 1.2. Copyright 2009 Casa Software Ltd

for XPS surface analysis

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka

Why Did We Spend 10 Years to Make the Digital SpecMaster XPS Spectral Data-Base System?

Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer 1

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

Estimation of IMFP except for a constant factor using only XPS background-optimized peak intensities and cross sections

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

raw materials C V Mn Mg S Al Ca Ti Cr Si G H Nb Na Zn Ni K Co A B C D E F

AES Features in CasaXPS

Application of Surface Analysis for Root Cause Failure Analysis

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

Surface and Interface Characterization of Polymer Films

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance

Auger Electron Spectroscopy (AES)

A DIVISION OF ULVAC-PHI

ISO INTERNATIONAL STANDARD. Surface chemical analysis X-ray photoelectron spectroscopy Repeatability and constancy of intensity scale

XPS/UPS and EFM. Brent Gila. XPS/UPS Ryan Davies EFM Andy Gerger

Auger Electron Spectroscopy *

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Auger Electron Spectroscopy

X-Ray Photoelectron Spectroscopy (XPS)

Repeatability of Spectral Intensity Using an Auger Electron Spectroscopy Instrument Equipped with a Cylindrical Mirror Analyzer

Photoemission Spectroscopy

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

X-ray Photoelectron Spectroscopy (XPS)

RESEARCH REPOSITORY.

INORGANIC ELEMENT FUNCTIONAL GROUP DATABASE ON PULVERIZED COAL SURFACE BASED ON XPS METHOD

Nanoscale IR spectroscopy of organic contaminants

Applications of XPS, AES, and TOF-SIMS

Surface Science Spectra

An Introduction to Auger Electron Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy

The active background method in XPS data peak-fitting

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

Fast, Effective XPS Point Analysis of Metal Components

Practical Surface Analysis

A DIVISION OF ULVAC-PHI. Quantera II. Scanning XPS Microprobe

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

The Standard Auger Electron Spectra in The AIST DIO-DB(111); Ge(111)

A Review of XPS Data-Banks

X-Ray Photoelectron Spectroscopy (XPS)-2

Methods of surface analysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Auger Electron Spectroscopy Overview

Developments & Limitations in GSR Analysis

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

Scanning Auger Microprobe

Student Projects for

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

X-Ray Photoelectron Spectroscopy (XPS)

Optical Properties of Thin Semiconductor Films

Supporting Online Material for

Spa & Wellness for Pipettes

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Photon Interaction. Spectroscopy

Measuring Colocalization within Fluorescence Microscopy Images

STUDY AND ANALYSIS OF THE FAILED USB DEVICES USED IN COPPER MINE USING ENVIRONMENTAL SCANNING ELECTRON MICROSCOPY AND X-RAY PHOTOELECTRON SPECTROSCOPY

Agilent All Ions MS/MS

Spa & Wellness for Pipettes

Inorganic Element Functional Group Database on Pulverized Coal Surface Based on XPS Method

Signal to Noise Comparison of IS-Instruments Ltd High Throughput OEM Spectrometer and a Czerny Turner Instrument. July 2013

Ma5: Auger- and Electron Energy Loss Spectroscopy

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

PHI. Scanning XPS Microprobe

Theta Probe: A tool for characterizing ultra thin films and self assembled monolayers using parallel angle resolved XPS (ARXPS)

1 of :32

Analyst Software. Automatic Optimization Tutorial

Trends in X-ray Microanalysis: Enabling Rapid Discovery One Pixel at a Time

Pulsar. Delivering NMR to your benchtop

Introduction to Fourier Transform Infrared Spectroscopy

NUCL 3000/5030 Laboratory 2 Fall 2013

High Resolution Optical Spectroscopy

Photoelectron Spectroscopy. Xiaozhe Zhang 10/03/2014

NMR Data workup using NUTS

Electronic Supplementary Information

Table S1. Structural parameters of shell-by-shell fitting of the EXAFS spectrum for reduced and oxidized samples at room temperature (RT)

IFM Chemistry Computational Chemistry 2010, 7.5 hp LAB2. Computer laboratory exercise 1 (LAB2): Quantum chemical calculations

Scintillation Detector

CIM PACA Characterisation Lab

Multi-residue analysis of pesticides by GC-HRMS

ToF-SIMS or XPS? Xinqi Chen Keck-II

LAB 6 - GRAVITATIONAL AND PASSIVE FORCES

WHAT DO RADIAL VELOCITY MEASUREMENTS TELL ABOUT RV TAURI STARS?

A Beginners Guide to XPS

Atomic Level Analysis of SiC Devices Using Numerical Simulation

The Use of Synchrotron Radiation in Modern Research

QUESTIONS AND ANSWERS

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

Growth and Trends in Auger Electron Spectroscopy and X-ray Photoelectron Spectroscopy

Analyst Software. Peptide and Protein Quantitation Tutorial

Introduction to Computer Tools and Uncertainties

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 7: Instrumentation for IR spectroscopy

Transcription:

Invited Paper XPS: Issues with Data Acquisition and Data Processing J.T. Grant Research Institute University of Dayton 3 College Park, Dayton OH 45469-51, USA j.grant@ieee.org (Received: November 26, 27; Accepted: December 21, 27) Methods for both data acquisition and data processing are very important for analyzing samples in x-ray photoelectron spectroscopy (XPS/ESCA). I find that many users do not understand the operation of their instruments and therefore are not taking data using optimal conditions. This can be as simple as using the wrong combinations of analyzer apertures and/or pass energies, energy intervals, and even the x-ray source when more than one source is available. Sometimes, limitations for data acquisition are imposed by the acquisition software. In data processing, the analyst is often limited to the processing software provided by the instrument manufacturer. There are however, some software packages that can convert data from several (proprietary) formats into ASCII. Also, software developers are increasingly providing output of their data into the ISO format for surface analysis. This, in principle, allows one to select from a wider variety of data processing packages. Many software packages today are very powerful and provide quite a variety of processing options. This paper discusses and illustrates some basic issues with data acquisition and data processing in XPS/ESCA. 1. Introduction During the past year, I have seen many examples of problems with the analysis of surfaces using x-ray photoelectron spectroscopy (XPS/ESCA). In many cases, instrument operators do not understand how the instruments acquire spectra, and therefore are not taking data using optimal conditions. Such conditions can be as basic as using the wrong combinations of analyzer apertures and/or pass energies, and using too many energy steps in spectra, particularly for survey scans. Even when more than one x-ray source is available, operators often use just one source for all their analyses. Sometimes, limitations in data acquisition are imposed by the acquisition software. Examples of this are (a) selecting from a fixed set of energy step intervals, (b) selecting from a fixed set of acquisition times (dwell times) per energy step, or (c) having to take data at only one possible dwell time per energy step. In data processing, the analyst is often limited to the processing software provided by the instrument manufacturer, sometimes because of difficulty in outputting the data into a non-proprietary format for use with other software. Examples of this are (a) using rather poor peak identification routines where peaks are often misidentified, (b) the inability to combine chemistry information from peak fitting higher resolution spectra into a quantification table obtained from a survey scan, or (c) the inability to fix peak intensities between chemical components of different elements according to their known chemistries. There are, however, some software packages that can convert the original data from several proprietary formats into the ISO14975 and 14976 formats. Also, instrument manufacturers themselves are increasingly providing output of their data into the ISO format for surface analysis. This, in principle, allows one to select from a wider variety of data processing packages should the need arise. Many software packages today are very powerful and provide quite a variety of processing options. This paper discusses and illustrates several basic issues with data acquisition and data processing in XPS/ESCA, using several case studies. These case studies are (a) selection of apertures and pass energies with a double-pass cylindrical mirror analyzer in order Copyright (c) 28 by The Surface Analysis Society of Japan -398-

to collect appropriate data in minimum time, (b) selection of energy step intervals for acquiring spectra with multi-channel detector systems and its impact on quantitative analysis using peak area, (c) selection of pass energies and energy step intervals in a hemispherical analyzer (HSA), again regarding the collection of appropriate data in minimum time, (d) acquisition times for appropriate signal-to-noise and detection limits, (e) a possible analyzer problem involving analyzer scattering, and (f) issues regarding the identification of peaks in spectra. studies of uniform samples, one must use the large apertures and change the pass energy for energy resolution in order to collect appropriate data in the shortest time. 2. Case studies The issues with data acquisition and data processing discussed in this paper are presented as several case studies. 2.1 Pass energies and apertures in a DPCMA: collecting appropriate data in minimum time A double-pass cylindrical mirror analyzer (DPCMA) was manufactured and sold by Physical Electronics for XPS (and AES) analysis [1]. The analyzer has two sets of apertures, one being in the central region separating the two cylindrical analyzers that are in series, the other being at the exit of the analyzer. There are three possible apertures in each set, and these are referred to as small, medium and large. Larger apertures give more luminosity, but degrade the energy resolution. I had used such an analyzer for several years and was familiar with its operating design. One steel research laboratory that I visited was using the same model DPCMA for their XPS work, but I was surprised with the rather poor signal-to-noise of their spectra. In checking the instrument operation with the laboratory technician, I found that he always took his XPS data using the two small aperture settings, and selected the pass energy to determine the spectral resolution. His samples were relatively large and the coatings that he was examining were uniform laterally, so this was the worse possible setting for his XPS analysis. As mentioned earlier, larger apertures give more luminosity, but degrade the energy resolution. However, luminosity increases as the square of the aperture diameter whereas energy resolution degrades linearly with aperture diameter. Therefore in XPS Fig. 1. An example showing a three-times improvement in signal-to-noise, as would be obtained using large-large apertures (top) rather than small-small apertures (bottom) in a DPCMA. The spectra have been offset for clarity. By using the large apertures, the technician was able to acquire similar data ten times more quickly than before, or obtain a three-fold improvement in signal-to-noise (and detection limit) for the same acquisition time that he used for many years beforehand. An example of such an improvement is shown in Fig. 1. These spectra are displayed after converting the raw data into ISO format, as are many of the spectra presented later. 2.2 Energy step intervals with multi-channel detector systems: impact on quantitative analysis using peak area Before acquiring spectra, it is important to consider what is needed from the analysis. If one is interested in obtaining peak areas for quantitative analysis, the acquisition parameters used should be quite different than if one is interested in peak lineshapes for studying chemical effects. Most XPS instruments sold today have multiple-channel energy detection [2], allowing one to use rather large energy step intervals for measuring peak areas. Some operators use small energy step intervals, so the peaks look the right shape and feel that if the peaks look distorted, the peak areas will be quite incorrect. Nothing could be further from the truth, -399-

and all one has to do is take some data in such instruments with different step intervals to see the effects. Au 4f spectra taken in an instrument with approximately 1 channels of detection and a pass energy of 16 ev are shown in Fig. 2. Five energy step intervals were used, namely, 2. (bottom spectrum), 1.5, 1.,.5, and.25 (top spectrum) ev. Area measurements on these five spectra are all within 2% of the average value obtained for the area. The spectrum at the top took eight times longer to acquire than the bottom spectrum, and even though the peak shape looks nice, it does not improve the accuracy of the area measure ment. In such an instrument, it is generally not necessary to take survey scan data with step intervals of.5 or 1 ev for peak area measurements. Minimizing analysis time is important, where possible, as commercial analytical laboratories typically charge hundreds of US$ per hour of analysis time. Of course, where peak overlap problems might occur, a separate analysis over a small energy range with more data points per ev, or a better energy resolution, might be necessary. Fig. 3 shows similar spectra to Fig. 2 except that the pass energy was reduced to 2 ev and an additional energy step of.125 ev was included in the data set. With a lower pass energy of 2 ev, the energy spread Table 1. Peak area measurements for the Au 4f doublet taken from the spectra displayed in Figs. 2 and 3 for pass energies (PE) of 16 and 2 ev. Peak areas obtained for pass energies of 8 and 4 ev are also included in the table..energy Step (ev) Peak Area for 16 ev PE Peak Area for 8 ev PE Peak Area for 4 ev PE 2. 229 k 232 k 216 k 212 k 1.5 229 k 241 k 222 k 219 k 1. 229 k 237 k 222 k 218 k.5 225 k 235 k 222 k 216 k.25 223 k 232 k 22 k 216 k.125 216 k Peak Area for 2 ev PE Fig. 2. Au 4f doublet taken with a monochromatic Al x-ray source in a Kratos Ultra XPS with approximately 1 channels of detection, using a pass energy of 16 ev. The energy step intervals were (bottom to top) 2., 1.5, 1.,.5, and.25 ev. The spectra have been offset for clarity. of the analyzer output across the 1 channels of detection is proportionally smaller and is now about 2 ev, so this would be the smallest pass energy where 2 ev steps would give an accurate area measurement. Of course, in practice one would never be using a 2 ev step interval at 2 ev pass energy as such a pass energy would be used Fig. 3. Au 4f doublet taken with a monochromatic Al x-ray source in a Kratos Ultra XPS with approximately 1 channels of detection, using a pass energy of 2 ev. The energy step intervals were (bottom to top) 2., 1.5, 1.,.5,.25, and.125 ev. The spectra have been offset for clarity. for lineshape analysis and would require many more channels per ev. However, there are still sufficient detection channels to accurately measure peak areas with large energy step intervals, and again the peak areas are all within 2% of the average peak area. The peak areas measured from the data shown in Figs. -4-

2 and 3 are listed in Table 1, showing their close agreement. Note that the peak area measurements were automatically corrected for step interval and for the different analyzer transmissions at the two pass energies shown. The analyzer transmission is saved with the spectral data and used in the data processing software for measuring peak areas [3]. The peak areas calculated for the 2 ev pass energy are consistently a few percent below those calculated for a pass energy of 16 ev, indicating that the transmission correction provided by the instrument calibration procedure was not perfect. 2.3 Pass energies and energy step intervals in a HSA: collecting appropriate data in minimum time Recently I was asked to help on a semiconductor project because the instrument that had been used for this project (at another university) was not available at the time. I offered to help, and was given part of an earlier report on an analysis. The instrument that I used for the analysis was similar (PHI 57), but I noticed that the other university took two hours to acquire their survey spectra. This is much longer than the accepted norm. I also noticed that the pass energy used for the survey scan was 58.7 ev, lower than what I use (188 ev). The energy step interval was not printed in the header with the spectra, but an attached text report stated that a step size of 1 ev was used for a survey scan and a step size of.2 ev was used for multiplex data (over the elements of interest). However, the pass energy used was the same (58.7 ev) for both acquisitions! The survey scan from the other university is shown in Fig. 4. Note the acquisition time of 125 min. The text report provided (from a different sample) is shown in Table 2. I requested the raw data file from the other university to see what step interval was used for the survey scan and it was.4 ev, which did not agree with either of the step intervals given in the associated table. Further, the concentration table (shown in Table 2), lists values for peak heights and not peak areas, so it was not clear if the quantitative analysis was made using the peak heights listed, or actual peak areas. Such discrepancies in reports can be of great concern for the experienced user, but are often overlooked by less experienced personnel requesting an analysis BM_31.SPE: Company Name 27 Aug 8 Al mono 35. W. μ. 58.7 ev 6.472e+4 max 125.4 min SUR/Area1/1 x 1 4 8 BM_31.SPE 7 -O1s 6 -N1s 5 -Ga2p3 c/s 4 3 2 1 -Ga2p1 -In MNN -OKLL -OKLL -In3p1 -F1s -In3p3 -In3d3 -In3d5 -C1s -Cl2s -Cl2p -Al2s -Al2p -Ga3d -In4d 12 1 8 6 Binding Energy (ev) Fig. 4. XPS survey scan taken from a semiconductor sample with a pass energy of 58.7 ev, a step interval of.4 ev, and a total acquisition time of 125 min. 4 2-41-

7817.spe: NE 31, new area Sur1/Full/1 15 x 1 4 7817.spe 1.4616e+5 max AFRL/MLBT 4.75 min 1 c/s 5 12 1 8 6 Binding Energy (ev) Fig. 5. XPS survey scan taken from a semiconductor sample with a pass energy of 188 ev, a step interval of.8 ev, and a total acquisition time of 5 min. 4 2 Table 2. Data given in a report related to the spectrum shown in Fig. 4. AlGaN/GaN/SiC (post reactor clean) NE31 XPS on 8-8-7. Survey of binding energy from -14eV. (DM_96). Pass energy = 58.7 Step size = 1eV Area of scan =.8mm Multiplex of NE1 (DM_961) Pass energy = 58.7 Step size =.2 ev Scan area =.8mm material Atomic Sensitivity Peak Peak height Atomic % orbital factor position N2 1s.499 1122 49.7 Al 2p.256 8242 37.2 In 3d5 4.53 6247 2.8 Ga 3d.438 2283 1.3 A survey spectrum that I took from a similar sample is shown in Fig. 5. This spectrum took only 5 min to acquire (a factor of 25 times faster) and is of similar quality as that shown in Fig. 4. This spectrum was taken with a pass energy of 188 ev and a step interval of.8 ev. Data at 188 ev pass energy could have been taken at a step interval of 1.6 ev for area measurements as this instrument has a 16 channel detector, and that would have resulted in a further reduction in time. This is more typical of survey spectra acquisition times needed in modern instruments. 2.4 Acquisition times for appropriate signal-to-noise and detection limits A scientist was studying an adhesive bond failure (involving a Cr-based primer) and was concerned that the failure might have occurred at the primer interface. He had been looking for Cr in the XPS spectrum, and as I walked past him, he asked do you see anything else in the spectrum besides O, Mg, C and Si? ; there isn t any Cr. In looking at the spectrum, the signal-to-noise was quite poor so I asked what the detection limit for Cr was in the spectrum and he just said there wasn t any. He was not interested in taking any more data, since he was sure that the failure was not at the primer interface, as he did not detect any Cr. The spectrum he took is shown in Fig. 6. From this spectrum, I later calculated the Cr detection limit to be.5 at.% based on the noise level in the Cr 2p region of the spectrum. Based on my experience, Cr levels detected -42-

Cr 2p.3 at% Fig. 6. XPS survey scan taken from a sample in an adhesive failure study involving a Cr-based primer. The detection limit for Cr is only.5 at.%. Fig. 8. XPS data from 52-62 ev binding energy showing the Cr 2p doublet detected from one side of the failure, indicating that the failure occurred at the primer interface. The main peak is O 1s. Cr? clearly in Fig.8, where spectra from the two failed surfaces show that one surface has.3 at.% Cr and the other would be <<.1 at.%, indicating that failure occurred at the primer interface. Acquisition times are very important as they must be sufficiently long to achieve the desired signal-to-noise for the required detection limit. It is not necessary to scan the entire survey region (typically -1 ev) to get the required signal-to-noise for a certain element. It is more time efficient to scan only the energy region of interest for that element for the necessary time. Fig. 7. XPS survey scans taken from opposite faces of an adhesive failure. The detection limit for Cr is <<.1 at.%. The spectra are offset for clarity. at surfaces of Cr-based primers are often less than.5 at.%. I have also studied such adhesive failures, and I feel that it is also necessary to look at both surfaces in a failure. XPS data that I had taken from two such surfaces are shown in Fig. 7. This study also involved a Cr-based primer. The most obvious difference between the spectrum in Fig. 6 and the two spectra shown in Fig. 7 is the higher signal-to-noise for the data in Fig. 7. To the casual observer, it might appear that there is no Cr peak detected from either surface in this case also. However, there is a significant difference between the spectra in the Cr 2p region of 57-59 ev, just beyond the loss structure from the O peak near 53 ev. This can be seen more 2.5 Possible analyzer problem involving analyzer scattering Another interesting example concerned an instrument that was awaiting repair under a service contract with the manufacturer. I had wanted to use the instrument, but was told by the responsible scientist (who had over 2 years of experience with such equipment), that I could not use it as it was not working properly and that the service engineer had been called to repair it. I enquired as to the problem, and was told that there was a high background in part of the spectrum. The lower part of Fig. 9 shows a spectrum similar to the one that I was shown, and was told, the background was too high in the middle and was actually sloping upwards (to the left of some peaks). I said that it probably was not a problem with the instrument at all, and that the increasing background indicated a layered structure as has been discussed many times by Tougaard [4]. I suggested -43-

looking at a different sample, and it turned out that there was no unusual background with it. The scientist quickly called the company to cancel the service call, and I was allowed to use the instrument. 2.6 Identification of peaks in spectra An interesting situation arises in identifying the weaker peaks in the lower spectrum of Fig. 7, specifically in the region -3 ev binding energy. The Si peaks near 1 and 15 ev are rather obvious, as are the Cl peaks near 2 and 27 ev. The peak near 13 ev might be thought to be the 2p peak of P, but it cannot be P as the 2s peak of P is absent in the spectrum. This can be seen clearly in Fig. 1(a) where the P 2s peak (which would be near 19 ev) is absent. Careful examination of the Cl peak intensities shows that the Cl 2s peak near 27 ev is also too intense relative to the Cl 2p peak near 2 ev. Therefore a peak from some other element is overlapping completing a qualitative analysis. Some data processing software packages claim to do this, but I find that this is often not the case. Another example of not checking relative peak intensities can be seen in Fig. 4, where the software identified the peak near 4 ev as N 1s and neglected the fact that some of this intensity is from an Auger peak from Ga, even though Ga was identified by the software to be present in this sample. Cl Cl P P x 1 4 7821.spe 12 1 (a) 8 c/s 6 4 2 backgrounds Sr Cl + Sr Cl Sr 12 1 8 6 Binding Energy (ev) 4 2 Fig. 9. XPS survey scans taken from a defective instrument (lower spectrum) since it had an increasing background near the center of the spectrum. A spectrum taken from a different sample (upper spectrum) did not show such a background, indicating there was no problem with the instrument. The spectra are offset for clarity. the Cl 2s peak. This other element is Sr, with its 3d peak near 13 ev. The overlapping peak near 27 ev is from Sr 3p 3/2 and the Sr 3p 1/2 peak can be seen as a shoulder on the rapidly rising C 1s peak; see Fig. 1(b). Sr could be further confirmed by taking a higher energy resolution scan of the Sr 3d region showing that this peak is a doublet with the appropriate energy separation. This is shown in Fig. 11. It is important to check relative peak intensities before (b) Fig. 1. Binding energy range of -3 ev from Fig. 7, showing (a) the initial labeling of peaks as Cl and P (the Si peaks near 1 and 15 ev are not labeled), and (b) the final labeling of the peaks as Cl and Sr. -44-

Sr 3d Fig. 11. The Sr 3d region of Fig. 1 (upper spectrum) taken at a higher energy resolution (lower spectrum) to observe the Sr 3d doublet and confirm the identification of Sr. 3. Summary The examples shown in this paper illustrate some basic issues that can occur in the acquisition and analysis of XPS data. Care should be taken in acquiring data to minimize acquisition times for the required analysis, and also in processing data to obtain accurate and meaningful results. 4. References [1] P.W. Palmberg, J. Electron Spectrosc. Relat. Phenom. 5, 691 (1974). [2] I.W. Drummond, XPS: Instrumentation and Performance, in Surface Analysis by Auger and X-ray Photoelectron Spectroscopy, ed. by D. Briggs and J.T. Grant, Chapt. 5, pp. 117-144, IM Publications and SurfaceSpectra (23). [3] CasaXPS, by Casa Software Ltd, Bay House, 5 Grosvenor Terrace, Teignmouth, TQ14 8NE, United Kingdom. [4] S. Tougaard, Quantification of Nano- structures by Electron Spectroscopy, in Surface Analysis by Auger and X-ray Photoelectron Spectroscopy, ed. by D. Briggs and J.T. Grant, Chapt. 12, pp. 295-343, IM Publications and SurfaceSpectra (23). -45-