GEMMA RAURET, J S MESTRES and J F GARCIA

Similar documents
Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE

SHERIDAN BOWMAN British Museum Research Laboratory London WC1B 3DG, England

THE POTENTIAL OF THE LONDON UNDERGROUND FOR LIQUID SCINTILLATION COUNTING

Jerde, Peterson and Stein (1967) indicated that high-energy radiation

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

SIGURDUR A EINARSSON and PALL THEODORSSON Science Institute, University of Iceland Dunhaga 3, IS-107 Reykjavik, Iceland

DETERMINATION OF RADON BY LIQUID SCINTILLATION a/f PARTICLE SPECTROMETRY: Towards the Resolution of a 14C Dating Problem

PERFORMANCE AND DESIGN OF 03-ML TO 10-ML SYNTHETIC SILICA LIQUID SCINTILLATION VIALS FOR LOW-LEVEL 14C DETERMINATION

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

Time-Resolved Liquid Scintillation Counting

Basics of a/b Discrimination for Liquid Scintillation Counting

Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE.

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

CIEMAT/NIST STANDARDIZATION METHOD EXTENDED TO ANODE OUTPUTS FOR BETA AND ELECTRON-CAPTURE NUCLIDES J. F. ORTIZ

Application Note. Abstract. Introduction. Experimental

Hidex 300 SL Automatic TDCR Liquid Scintillation Counter

RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

RADIOCARBON DATING REPRODUCIBILITY AT THE MUSEO DE LA PLATA RADIOCARBON LABORATORY

GAS PROPORTIONAL VERSUS LIQUID SCINTILLATION COUNTING, RADIOMETRIC VERSUS AMS DATING PALL THEODORSSON

Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Unit 2. Instrumentation. Experts Teaching from Practical Experience

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

LS6500 Scintillation Counting Systems

BUREAU INTERNATIONAL DES POIDS ET MESURES

IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2

COMBINED LSC-BASED METHOD FOR RADON IN AIR MEASUREMENTS

[RADIOCARBON, VOL 31, No. 3, 1989, P ]

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa.

THE STATISTICS OF LOW-LEVEL COUNTING USING THE NEW GENERATION OF PACKARD LIQUID SCINTILLATION COUNTERS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

arxiv: v1 [physics.ins-det] 3 Feb 2011

EXTERNAL STANDARD METHOD OF QUENCH CORRECTION: ADVANCED TECHNIQUES. B. H. Laney Searle Analytic, Inc. Des Plaines, Illinois ABSTRACT

TECHNETIUM-99 IN WATER

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Measurement of the carbon 14 activity at natural level in air samples

Simultaneous measurement of 89 Sr and 90 Sr in aqueous samples by liquid scintillation counting using the spectrum unfolding method

Radiation Detection. 15 th Annual OSC Readiness Training Program.

TECHNETIUM-99 IN SOIL

Primary Standardisation of I-125

Liquid Scintillation Counter

2100TR Liquid Scintillation Counter

Advances in radiocarbon measurement of water samples

Fundamentals of Radionuclide Metrology

1 of :32

TECHNETIUM-99 IN WATER

HIGH-PRECISION 14C MEASUREMENT OF GERMAN AND IRISH OAKS TO SHOW THE NATURAL 14C VARIATIONS FROM 7890 TO 5000 BC

Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd.

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Hidex Liquid Scintillation Counters

Status of the AMoRE experiment searching for neutrinoless double beta decay of 100 Mo

Teflon Vials For Precise C-14 in Benzene Measurements by LSC Technique

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Development of an active scintillating target for fission studies. J. Aupiais, G. Bélier CEA, DAM, DIF, F Arpajon France

Čerenkov counting and liquid scintillation counting of 36 Cl

MEASUREMENT AND DETECTION OF RADIATION

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner

Scintillation Detector

Hidex LSCs. LSC 2013 Barcelona

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Radio-HPLC Flow Cells

COMBINED LSC BASED METHOD FOR RADON IN AIR MEASUREMENTS

corresponding l Nuclear Physics Laboratory and Quaternary Isotope Laboratory University of Washington, Seattle, Washington INTRODUCTION

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2

Determination of 210 Pb and 210 Po in Water Samples

Quality Assurance. Purity control. Polycrystalline Ingots

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC)

Detection of 222 Rn and 226 Ra in environmental samples by scintillation method

The LSC Approach to Radon Counting in Air and Water

On the Calibration of Alpha Sciences Proportional Counters

Beta Screening Options

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY

Gamma Analyst Performance Characteristics (MDAs)

RESULTS YOU CAN COUNT ON

Measurement of 39 Ar in Underground Argon for Dark Matter Experiments

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures

Radiochemical Sensor for Continuous and Remote Liquid Effluents Monitoring

WM2018 Conference, March 18-22, 2018, Phoenix, Arizona, USA. PVT and LaBr3(Ce)-based Radon Express Analyzers 18164

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Radiometric Measurements of Environmental Radioactivity

LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS. The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine.

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

ORTEC AN34 Experiment 10 Compton Scattering

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

SODIUM PERTECHNETATE ( 99m Tc) INJECTION (FISSION): Revised Final text for addition to The International Pharmacopoeia (January September 2009)

Transcription:

[RADIOCARBON, VOL31, No. 3, 1989, P 380-3861 OPTIMIZATION OF LIQUID SCINTILLATION COUNTING CONDITIONS WITH TWO KINDS OF VIALS AND DETECTOR SHIELDS FOR LOW-ACTIVITY RADIOCARBON MEASUREMENTS GEMMA RAURET, J S MESTRES and J F GARCIA Radiocarbon Laboratory, Department of Analytical Chemistry University of Barcelona - Institute of Catalan Studies, Barcelona, Spain ABSTRACT. Low-level radiocarbon measurements were made using two liquid scintillation (LS) counters, both with passive lead shields of approximately the same dimensions. The first one had a conventional background reduction with low-activity lead and "high-low" coincidence bias selection, and the second one had a spectrum analysis background reduction based on the different three-dimensional pulse-height signal of background and beta pulses. The performance of commercial teflon-copper and glass vials were compared with both LS counters. Optimum counting conditions for the two detectors studied were also established. From results obtained, quality parameters and characteristics of quench curves were studied. Performances of each counter in the working conditions established were also compared. INTRODUCTION Liquid scintillation (LS) counting of low-level 14C activity ensures the lowest detection limits as well as high-precision measurement. In our laboratory, we have tested two counters which have different background reduction design concepts. An LKB 1217 Rackbeta Kangaroo with a conventional passive low-activity lead shield surrounding the counting chamber, "high-low" coincidence bias selection, and the optimum use of teflon-copper (TC) vials for background reduction (Polach et al, 1983b; Calf & Polach, 1974; Kuc & Rozanski, 1978; Polach et al, 1983a; Rauret, Mestres & Garcia, 1988). The other counter tested is a Packard TriCarb 2000 CA/LL with spectrum analysis which recognizes the different threedimensional pulse-height distribution of background pulses and 14C beta pulses, thus reducing the background counts at the expense of efficiency (Operation manual TriCarb 2000/LL). This paper compares the performance of the two LS counters for low-level 14C measurements under optimum conditions. Quality parameters and characteristics of quench curves were also studied. One of the quality parameters calculated was the oldest attainable age that corresponds to the lowest measurable activity determined according to a defined criterion. The sample count rate was considered greater than background when the difference between both was larger than twice the standard deviation of the background (Stuiver & Polach, 1977). The determination of 14C activity by LS counting of pure benzene has the advantage of working under constant quench conditions. In practice, however, we realized that even under these conditions the spectral end-point of 14C is not constant but presents some small variations (Rauret, Mestres & Garcia, 1988), so that each counting vial must be individually calibrated based on external standard ratio determinations of counting efficiency. 380

Optimization of LS Counting Conditions 381 EXPERIMENTAL We used two counters, an LKB-Wallac Rackbeta 1217 Kangaroo and a Packard TriCarb 2000 CALL. The characteristics of the first counter are: high voltage stabilization circuit, "high-low" coincidence bias selection, linear amplification and logarithmic 256 channel spectra (1-2800 key range) display, four independent channels with variable windows for data accumulation, automatic external standardization with a Ra-226 source, and refrigeration (10 C). The characteristics of the second counter are: linear amplification, MCA of 4096 channels (1-2000 key range) a variable range background discriminator based on the three-dimensional spectrum analyses SPEC- TRALYZERTM, three independent regions with variable limits for data accumulation, external standard source of Ba-133, spectrum storage possibility and refrigeration. Two kinds of vials were used: teflon-copper (TC) 7m1(LKB-Wallac), modified (Rauret, Mestres & Garcia, 1988) and the standard lowpotassium-content borosilicate 20m1 glass vials (Packard). We also used analytical reagents grade (AR), benzene and toluene (Carlo Erba, RPE) and 14C-labeled toluene (Packard). We prepared active toluene in benzene solutions of 2428.2 dpm/ml (A) and 3697.7 dpm/gr (B) from the previous solution. As quencher, solutions of CC14 in benzene: 0.5, 1.0, 2.0 and 4.0% (vlv) were used. The four scintillators used were: commercial cocktail Instagel, butyl-pbd (2.5% w/v and 5% w/v in benzene), PPO (0.4% wlv) in toluene and a mixture PPO (0.4% wlv) dimethyl- POPOP (0.01% w/v) in toluene (all Packard). The standard procedure employed 3ml of total counting solution obtained with 2.4m1 sample benzene and 0.6m1 of scintillator solution. In all cases, including commercial cocktail Instagel, we maintained the same 4:1 ratio between sample benzene and scintillator solutions. To carry out performance comparison of glass and teflon vials, we used solutions with PPOdimethyl-POPOP as scintillator and 3m1 of total counting volume for the LKB counter and butylpbd and 4m1 for the Packard counter. The blank was solutions with the same composition and conditions as the samples. In the quenching efficiency correlation study, the standards were prepared as follows: for the LKB counter, 3.Oml of A solution, 0.4m1 of butyl- PBD (5%), variable volumes of CCl4 benzene solution and inactive benzene up to a total volume of 4.Oml were added to each TC vial. Twelve activity standard solutions with increasing CC14 concentrations from 0 to 0.40% (v/ v) were obtained. For the Packard counter, 1.Oml of B solution was weighted in the glass vial previously tared; then the same procedures as before were followed. With the LKB counter, the spectra were plotted manually 10 channels at a time. Overall counting time for the standard 14C was 6 hr and for the background 12-20 hr. With the Packard, the spectra were obtained with the low-level correction option and with a counting time of 2-4 hr for the standard 14C and 6-12 hr for the background. For the study of the correlation between quenching and efficiency, the two sets of activity standards were counted for 10 periods of 10 min each, in

382 G Rauret, JS Mestres and JF Garcia the corresponding counters. The quenching parameter was provided by the external standard. For the LKB counter, an external standard channel ratio was used; the external standard window partition used is that which furnishes the maximum variation of the parameter between the less-quenched and the more-quenched solutions. The t-sie parameter was used in the Packard counter. RESULTS AND DISCUSSION The following results were obtained from the integration of experimentally obtained spectra. In each case, the integration thresholds chosen as the optimum window were those which gave the best figure of merit. Performance of Glass and Teflon Vials The characteristics of TC and low-potassium borosilicate glass vials were compared in both counters. From the results obtained for background and efficiency, the figure of merit for each case was calculated. Table 1 TABLE 1 Background, efficiency and figure of merit of glass and teflon vials LKB B(cpm) E(%) FM Type of vigl Glass 2.36 33.8 Teflon-copper 1.41 Scintillator Instagel 1.54 66.7 Butyl-PBD 1.78 72.5 PPO 1.43 PPO+dimethyl- POPOP 1.33 Volume 3m1 1.78 72.5 5m1 2.47 73.7 7ml 3.39 73.8 PACKARD shows the results. For the LKB counter, the figure of merit is much higher for TC vials than for glass because of the higher background count rate at low energies in the glass vials due to Cerenkov radiation. Figure 1 shows the spectra of the background for the glass and TC vials. For the Packard counter, on the contrary; the glass vials yield a much higher figure of merit than the one obtained with TC. Scintillator Effect Once the best class of vial for each counter was chosen, the effect of different scintillators on performance was tested. Table 1 shows results relating to efficiency (E) and background (B) and figure of merit for each counter and four scintillators.

Optimization of LS Counting Conditions 383 PM-B 7 6 5 4 3 2 1 50 6 70 80 90 100 110 120 130 140 150 160 170 CHANNELS Fig 1. Radiocarbon (S) and background (B) spectra for glass and teflon vials Based on the figure of merit, it can be concluded that for the LKB, as described earlier (Rauret, Mestres & Garcia, 1988), butyl-pbd is the best scintillator, followed by Instagel, PPO and PPO + dimethyl-popop. These results agree with those found by Polach et al (1983). On the other hand, for Packard, the best scintillator is Instagel and the descending order of figure of merit is: Instagel > butyl-pbd > PPO > PPO + dimethyl- POPOP. Although the 14C beta particle to photon energy transfer is better using butyl-pbd as scintillator, the shape of the 14C spectrum in this counter shifted for Instagel to lower energy values, allowing us to work with a window for which the background is smaller for Instagel than for butyl :PBD with a similar efficiency. Under these conditions, the figure of merit for Instagel is better. Volume Effect In order to compare the volume effect, we chose butyl-pbd as scintillator for both counters. It has been established previously (Rauret, Mestres & Garcia, 1988) that, for the LKB counter, background increases linearly with volume whereas efficiency remains approximately constant with volume, so that the figure of merit increases when volume decreases. Similar behavior has been observed with the Packard counter. The values of efficiency (E), background (B) and figure of merit for both counters are also given in Table 1. Considering that loss of sensivity occurs with small volumes of solution, we propose to standardize on a total volume of 4.Om1(3.2m1 sample + 0.8m1 scintillator solution).

384 G Rauret, JS Mestres and JF Garcia Quench Correlation Curves If TC vials are used in the LKB counter, the quenching parameter ranges from 474 to 534. For glass vials with the Packard counter, this range is narrower, ranging from 769 to 813. This has compelled us to study the correlation between efficiency and quenching for each counter in order to 744 Fig 2. Quenching correlation curve for the Parckard counter x IN 475 525 575 Fig 3. Quenching correlation curve for the LKB counter x

Optimization of LS Counting Conditions 385 correct the small variations of efficiency in the working window. Figures 2 and 3 show the quenching correlation curves obtained for each counter. Efficiency drops quickly when the quenching increases for the LKB counter and it remains nearly constant in the quenching range studied for the Packard counter. In both cases, the working window used was that which gives the best figure of merit corrected for the possible interference of tritium. With the LKB counter, the possibility of using a window, the efficiency of which was not sensitive to small variations of quench, was rejected because of the drop of figure of merit obtained in these conditions, which was considered less advantageous for these kinds of measurements. Optimum Working Conditions for Both Counters From the results obtained in the previous sections, we have established the optimum working conditions for LKB and Packard counters to measure low-level activity 14C samples. Table 2 gives these conditions. TABLE 2 Optimum working conditions for the counters studied Counter Vial Scintillator Volume LKB Teflon-copper Butyl-PBD 3m! PACKARD Glass Instagel 3m! Oldest Attainable Age Let us assume that the smallest measurable activity is 2S(B) Am = axe where a is the weight of benzene measured and E is the efficiency of counting. Then, the greatest attainable age is given by: _ 5568 tm 1n2 In AoN 2S(B) According to this equation, the oldest attainable age may be calculated in two ways: with the theoretical standard deviation corresponding to background counting, and with the experimental standard deviation of the background. The oldest attainable ages for each counter in optimum working conditions with theoretical and experimental background standard deviations are nearly the same value for the LKB counter, 44,300 yr, and for the Packard counter, 45,000 yr, for theoretical (Poisson) and 41,100 for experimental. The experimental standard deviation gives, with the Packard counter, smaller values because the experimental standard deviation of the background is higher than the theoretical. ae

386 G Rauret, JS Mestres and JF Garcia CONCLUSIONS For measurements of low 14C levels, both counters perform well. Nevertheless, it is necessary to optimize conditions to attain lower detection limits. The main differences observed under optimal conditions are related to the vial and the scintillator. If we compare the performances of each counter with optimal conditions, we can conclude that the LKB has a more stable background but its efficiency is more dependent on the degree of quenching. Contrasting behavior is observed for the Packard counter. REFERENCES Calf, G E and Polach, H, 1974, Teflon vials for liquid scintillation counting of carbon-14, in Liquid scintillation counting: Recent developments: New York, Academic Press. Kuc, T and Rozanski, K, 1978, A small volume teflon-copper vial for 14C low level liquid scintillation counting: Internatl Jour Applied Radiation Isotopes, v 30, p 452. Operation manual TriCarb 2000 CALL: United Technologies, Packard Instruments, Zurich. Polach, H, Gower, J, Kojola, H and Heinonen, A, 1983a, An ideal vial and cocktail for low level scintillation counting, in Advances in liquid scintillation counting: Alberta, Univ Alberta Press, p 508-526. Polach, H, Nurmi, J, Kojola, H and Soini, E, 1983b, Electronic optimization of scintillator counter for detection of low-level H-3 and C-14, in Advances in liquid scintillation counting: Alberta, Univ Alberta Press, p 420-442. Rauret, G, Mestres, J S and Garcia, J F, 1988, Effect of counting vials and scintillators on the detection of low-level 14C by liquid scintillation spectrometry: Nuclear Instruments & Methods, v B34, p 373-376. Stuiver, M and Polach, H A, 1977, Discussion: Reporting of 14C data: Radiocarbon, v 19, no. 3, p 355-363.