SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM

Similar documents
S is referred to as sour. The sulfide loading lends an unpleasant odor and makes the water corrosive. Additionally, the presence of H 2

Reprinted from February Hydrocarbon

Physicochemical Processes

SULFUR RECOVERY ANALYZERS; UV Diode-Array Fiber- Optics Technology

Cyanide Analysis of Wastewater Samples from FCC and Hydrocracking Operations

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2

MEASURING H 2 S IN CRUDE OIL FOR QUALITY CONTROL & TRANSPORTATION SAFETY

A New Hyperspectral Spherical-Cavity Absorption Meter

ISA QATAR : 90 th Seminar

Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

Method for Nitrite determination on Low-Range Samples

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Determining the Concentration of a Solution: Beer s Law

SERVOTOUGH SpectraScan

Mass Transfer Operations I Prof. Bishnupada Mandal Department of Chemical Engineering Indian Institute of Technology, Guwahati

Final Exam. Physical Constants and Conversion Factors. Equations

Operating Instructions for Ammonium ISE Specifications

MASS TRANSFER AND GAS ABSORPTION EQUIPMENT

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

2B Technologies, Inc. An InDevR Company

Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons. August 9, 2011

Chromatography & instrumentation in Organic Chemistry

Lab 5: Calculating an equilibrium constant

Determining the Concentration of a Solution: Beer s Law

Carbon dioxide removal processes by alkanolamines in aqueous organic solvents Hamborg, Espen Steinseth

Physical Separations and Chromatography

WEEK 2 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY

Distillation of Liquids: Separation of 2-Propanol from Water by Fractional Distillation

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes

ph Probe ReallyEasyData com

UNIT 4 Part 2 Measurements of Environmental Air Pollution Parameters j

Sampling. Information is helpful in implementing control measures for reducing pollutant concentration to acceptable levels

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Rule 2. Rule 1. Rule 4. Rule 3. Rule 5. Rule 6. Rule 7. Rule 8

Physics 197 Lab 11: Spectrometer

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

8 Enthalpy of Reaction

(Refer Slide Time: 00:00:43 min) Welcome back in the last few lectures we discussed compression refrigeration systems.

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

3 - Atomic Absorption Spectroscopy

Advanced Pharmaceutical Analysis

Measurement Technique and its Application to Trace Components of Atmospheric Gas

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

Brooklyn College Department of Chemistry

Corresponding Author: Jason Wight,

Determination of an Equilibrium Constant

STATE OF COLORADO DESIGN CRITERIA FOR POTABLE WATER SYSTEMS WATER QUALITY CONTROL DIVISION. Price: $5.00. Revised March 31, 1997

GENERAL INSTRUCTIONS GENERAL PREPARATION

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

Experiment 7A ANALYSIS OF BRASS

Sodium Chloride - Analytical Standard

A Study of Beer s Law Prelab

Validating Slope Spectroscopy Methods: A Formula for Robust Measurements

Distillation is a method of separating mixtures based

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

UV Quality Assurance. Dark Current Correction

Optical In-line Control of Web Coating Processes with UV-VIS-NIR Spectroscopy. AIMCAL Europe Web Coating Conference Peter Lamparter

SEPARATION TECHNIQUES

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

FEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada.

FLOW-THROUGH CAPACITOR ASSEMBLY FOR THE TREATMENT OF A. Field of application

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Figure 4-1: Pretreatment schematic

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Benefits of Heat Stable Salts in Tail Gas Treaters

Aspirin Synthesis H 3 PO 4

Design of A chemical Absorption System for The Separation of Propane/Propylene Mixture

Physics P202, Lab #12. Rydberg s Constant

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

S reacts with the catalyst s metallic surface to substitute sulfur atoms for oxygen atoms.

Lecture (9) Reactor Sizing. Figure (1). Information needed to predict what a reactor can do.

TRACERCO Diagnostics FCCU Study

MEASUREMENT: PART II

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Systems Engineering Spring Group Project #1: Process Flowsheeting Calculations for Acetic Anhydride Plant. Date: 2/25/00 Due: 3/3/00

THE APPLICATION OF PROCESS MASS SPECTROMETRY TO FUMED SILICA PRODUCTION

1901 Application of Spectrophotometry

UV3000. Accurate, precise, and portable ambient gas point analyzer

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

CHAPTER 3 MODELLING AND ANALYSIS OF THE PACKED COLUMN

Experiment 9. Emission Spectra. measure the emission spectrum of a source of light using the digital spectrometer.

The Synthesis and Analysis of Aspirin

SYNTHESIS OF 1-BROMOBUTANE Experimental procedure at macroscale (adapted from Williamson, Minard & Masters 1 )

2 Copyright Pearson Education, Inc., or its affiliates. All Rights Reserved.

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW

OPTICAL GAIN AND LASERS

Advancements in Tracer Flow Testing; Development of Real-Time Technology for Flow and Enthalpy Measurements

Lab Investigation 4 - How could you make more of this dye?

25 Instruments for Optical Spectrometry

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water

Transcription:

SOUR WATER ANALYSIS USING A DUAL- COLUMN HEADSPACE SAMPLING SYSTEM Andy Burgess, David Greaves, Daniel Murphy and Sam Rolley Applied Analytics, Inc. 29 Domino Drive, Concord, MA 01742 KEYWORDS Sour Water, Headspace, H 2 S, NH 3, Spectroscopy ABSTRACT In the petroleum refining process one of the main byproducts is wastewater rich in hydrogen sulfide and ammonia, commonly referred to as sour water. These chemicals must be removed to prevent contamination and corrosion. A process called Sour Water Stripping accomplishes this removal. Accurate measurement of hydrogen sulfide and ammonia concentrations at the outlet of the stripper provides critical process validation information and allows for corrective action in the event of stripper failure. To carry out this measurement in real time, a dual-column headspace sampling system is utilized in conjunction with a UV-diode array based spectrometer. This system allows for the simultaneous measurement of hydrogen sulfide and ammonia with a single device. A detailed discussion of sour water measurement benefits, the unique features of headspace sampling, and experimental results are presented. INTRODUCTION Sour water is defined as any refinery process water that contains hydrogen sulfide (H 2 S). With such a broad definition, sources of sour water vary significantly. Atmospheric crude columns and vacuum crude towers produce sour water from condensed stripping steam removed by overhead condensing systems. Steam cracking and fluid catalytic cracking units produce sour water as condensates from steam used in injection stripping and aeration [1]. Refineries create sour water as a byproduct in many of their processes. Unfortunately, sour water contains ammonium salts such as ammonium bisulfide and ammonium Session 7.3: Page 1

carbonate. These ammonium salts can cause several problems. In an extensive study of corrosion rates in sour water, a number of variables were considered including ammonium bisulfide concentration, velocity, H 2 S partial pressure, temperature, chloride concentration, hydrocarbon content, and chemical treatments. It was confirmed that the concentration of ammonium bisulfide is the most influential variable contributing to corrosion [5]. Ammonium salts can also settle on orifices causing plugging and increase maintenance concerns. As a result, refineries need to remove the hydrogen sulfide and ammonia (NH 3 ) from the water to avoid the formation of these ammonium salts when the water is reused for other plant processes. To remove the H 2 S and NH 3, a sour water stripper is often utilized. The plant operators need to know how efficient the stripping process is and if the water is suitable for reintroduction. TECHNOLOGY To address the need brought to light in the introduction, Applied Analytics developed a system utilizing the OMA-300 for the measurement of H 2 S and NH 3 in sour water. It utilizes an ultraviolet-visible full-spectrum spectrometer and a dual column headspace sampling approach. The combined system offers many benefits. It allows a large dynamic range in the measurement of the analytes due to flexibility in wavelength selection. Cross interference from moisture is avoided, because the measurement is being done in the UV- VIS region where water doesn t absorb light. Finally, consumables and/or reagents are not necessary for operation. The technology is broken down into two sections for ease of explanation: an analytical section and a sample conditioning section. ANALYTICAL The analytical configuration consists of the following equipment: a xenon flash lamp, an ultraviolet-visible full-spectrum spectrometer, collimating lenses, fiber optic cables, and a flow through cell. This layout can be seen in Figure 1. Absorbance is defined using the following equation: A = log 10 (I 0 /I) (1) Where: A = Absorbance I 0 = Light In I = Light Out Session 7.3: Page 2

FIGURE 1. ANALYTICAL HARDWARE The measurement algorithm used for analysis is a classical least squares regression of Beer s Law [2]: A l = e l bc (2) Where: A = Absorbance l = Wavelength e = Molar Absorptivity b = Path Length c = Concentration The spectrometer measures I 0 during its zero calibration and I when the analyte is in the flow through cell of path length b. e l is found by introducing a known concentration of the analyte into the flow-through cell and is saved for use in subsequent readings to determine unknown concentrations of the analyte. This would be adequate if the absorbance at a given wavelength could be attributed to only one component. However, this isn t the case with H 2 S and NH 3 in the ultraviolet region. At each wavelength in the UV where NH 3 absorbs, so does H 2 S. This is illustrated in the following spectra (Figures 2 & 3). Session 7.3: Page 3

H2S (gas) Absorbance Curve 1.2 1 ) U (A e c n a rb o s b A 0.8 0.6 0.4 0.2 0 215 220 225 230 235 240 245 250-0.2 Wavelength (nm) FIGURE 2. H 2 S ABSORBANCE CURVE NH3 (gas) Absorbance Curve 1 0.9 0.8 ) U 0.7 (A e c 0.6 n a rb 0.5 o s b 0.4 A 0.3 0.2 0.1 0 215 220 225 230 235 240 245 250 Wavelength (nm) FIGURE 3. NH 3 ABSORBANCE CURVE Session 7.3: Page 4

Therefore, in this case the equation must be extended to the following: A lx = e 1lx * b * c 1 + e 2lx * b * c 2 (3) A ly = e 1ly * b * c 1 + e 2ly * b * c 2 (4) Where lx could be 217nm and ly could be 220nm, for example (the over determination is omitted for simplicity). The final concentration readings are determined by solving for c 1 and c 2 using the previous equations. Accordingly, before the gas can be analyzed it must be coerced into the vapor phase. This is accomplished by the headspace sample conditioning system. SAMPLE CONDITIONING The heart of the headspace system is the stripping column (Figure 4). Its dimensions are 24 in height by 2 in diameter. The column is sealed with gaskets that are held together by sanitary clamps. The column is filled with Pall rings for efficient mass transfer between the liquid sample and the carrier gas. The column is heated with three band heaters and controlled with a thermocouple and a temperature controller. The flow rates of the carrier gas and the sample liquid are controlled by variable area flow meters with integrated flow control valves. A backpressure regulator regulates the pressure in the stripping column and the flow through cell. During operation, the liquid sample enters through a fitting on the top of the column and exits through a fitting on the bottom of the column. A carrier gas is introduced through another fitting on the bottom of the column and exits through a fourth fitting on the top of the column. This establishes a countercurrent flow between the liquid sample and the carrier gas. The Pall rings allow for efficient mass transfer between the liquid sample and the carrier gas by greatly increasing the total internal surface area of the column. After the carrier gas leaves the column it flows to the flow through cell for measurement. The headspace system is based on the principle of Henry s Law: C = k*p gas (5) Where: P = Partial Pressure k = Henry s Law Constant c = Concentration Henry s Law states that the amount of a gas dissolved in a solution at a given temperature is directly proportional to the partial pressure of the gas above the solution [3]. Session 7.3: Page 5

FIGURE 4. HEADSPACE SAMPLE CONDITIONING SYSTEM Therefore, given constant conditions such as temperature, pressure, carrier gas flow rate and liquid sample flow rate the headspace gas being measured can be correlated to the concentration of the liquid sample. Although the headspace sampling method can be used in a variety of applications, it is most advantageous in measurements of components with high volatility in opaque samples that have interfering components with low volatility. Sour water is a good example of such a stream, since it can often contain phenols that were formed from reactions between steam and cyclic hydrocarbons. Phenols have a very strong absorbance in the low UV region (1500 L/(mol*cm) @ 228nm), which would make measurement of the H 2 S or NH 3 in the liquid phase very difficult. Fortunately, phenols also have a much higher boiling point (see Table I) than H 2 S or NH 3. As a result, only very low phenol concentrations can exist in the headspace gas, thus making the measurement easier by minimizing the interfering component. Session 7.3: Page 6

TABLE I. BOILING POINT COMPARISON Phenol H 2 S NH 3 Boiling Point 181.7-60.28-33.34 EXPERIMENT The objective of the experiment is to quantitatively demonstrate the typical accuracy and stability that the headspace sampling system offers when correlated with a liquid sample of H 2 S and NH 3 in water. PARAMETERS A number of parameters need be determined before starting the experiment. The proper liquid sample flow rate shall be constant. The liquid sample flow rate was experimentally determined by filling the column with water and draining it with just gravity as a motive force. The column was packed with Pall rings, as it would be in normal operation, during this test. The time in which it takes the water to fully evacuate the column was recorded. The volume of water as it leaves the column was captured in a graduated beaker, thus allowing the determination of the volume of the column with Pall rings installed. That information was then used to find the volumetric flow rate (V/t). Since Henry s Law is dependent upon temperature, it was controlled and kept constant at a temperature that promotes low solubility of the gas in the water. The dependence of the solubility on temperature is described by the following equations [4]: k = k t * exp [ (-enthalpy/r) * ((1/T) (1/T t )) ] (6) (-d*ln(k))/(d*(1/t)) = enthalpy/r (7) Where k = Henry s constant of a specific gas (M/atm) k t = Henry s constant of a specific gas at a temperature of 298.15 K (M/atm) R = universal gas constant (8.314 (J/mol*k)) T = System temperature (K) T t = 298.15 K (-d*ln(k))/(d*(1/t)) = Temperature Dependence (K) TABLE II. HENRY S LAW CONSTANT & TEMPERATURE DEPENDENCE H 2 S NH 3 Henry's Law Constant (M/atm) 0.1 61 Temperature Dependence (K) 2200 4200 Session 7.3: Page 7

Table II gives a value for temperature dependence, which according to Equation 7 is equal to enthalpy/r. That value was substituted into Equation 6 to find the Henry s Law Constant at different temperatures. Then Equation 5 was used to predict the concentration of the gas in the water. TABLE III. HENRY S LAW CONSTANT FOR H 2 S AT DIFFERENT TEMPERATURES k (M/atm) k t (M/atm) E/R (K) T (K) T t (K) 0.0787 0.1 2200 308.15 298.15 0.0628 0.1 2200 318.15 298.15 0.0509 0.1 2200 328.15 298.15 0.0417 0.1 2200 338.15 298.15 0.0346 0.1 2200 348.15 298.15 0.0290 0.1 2200 358.15 298.15 0.0245 0.1 2200 368.15 298.15 TABLE IV. HENRY S LAW CONSTANT FOR NH 3 AT DIFFERENT TEMPERATURES k (M/atm) k t (M/atm) E/R (K) T (K) T t (K) 38.6184 61 4200 308.15 298.15 25.1617 61 4200 318.15 298.15 16.8277 61 4200 328.15 298.15 11.5251 61 4200 338.15 298.15 8.06690 61 4200 348.15 298.15 5.75998 61 4200 358.15 298.15 4.18873 61 4200 368.15 298.15 Tables III and IV show that as temperature increases, the Henry s Law constant and the concentration of the dissolved gas in the water decreases. Therefore, it is beneficial to choose a high temperature for the column set point because it forces more of the dissolved gas from the liquid phase into the vapor phase. A larger concentration in the vapor phase results in more absorbance in the path length of the flow cell. Furthermore, the accuracy of the analyzer increases as absorbance increases. Keeping this in mind, along with practical limitations of the setup, 50 degrees C was chosen for the set point of the column. The carrier gas flow rate was also kept constant. The ideal carrier flow rate was set based on the highest flow rate that was achieved without decreasing the concentration of the analyte in the stripping carrier gas. This was determined experimentally by analyzing the carrier gas at several carrier gas flow rates and recording the results. Session 7.3: Page 8

The analyzer (UV absorbance) was calibrated on gas standards of H 2 S in N 2 and NH 3 in N 2. These spectral standards were used to establish an extinction coefficient for each gas at each wavelength in the range of interest. This calibration was done prior to water testing. It is also important to note that ph plays an important role in sour water stripping. At low ph levels near 5.5, NH 3 tends to be held in solution and H 2 S tends to be more volatile. However, at higher ph values near 9.0, NH 3 is more volatile and H 2 S tends to stay in solution [1]. For the purposes of these experiments a constant ph was assumed. PROCEDURE A sample of water without H 2 S or NH 3 was introduced into the column. The temperature of the column, the carrier gas flow rate and the liquid sample flow rate were held constant at their optimal values. This sample was used as a zero reference by the analyzer. A sample of water with a known concentration of H 2 S was prepared in a reservoir large enough to hold the volume required to run a 1-hour stability test. The temperature of the column, the carrier gas flow rate and the liquid sample flow rate were held constant at their optimal values. The concentration log function on the analyzer was activated and the system was left to run for 1 hour. The start time and the end time were recorded. At the end of the 1-hour run, a reservoir of clean water was connected to wash the system. Second, third, and fourth samples of water with different known concentrations of H 2 S were prepared and introduced to the system in the same manner as the first. A 1-hour run was not required as these samples were for the purpose of testing accuracy, not stability. After the system was washed with clean water again, a sample of water with a known concentration of NH 3 was prepared in a reservoir large enough to hold the volume required to run a 1-hour stability test. The concentration log function on the analyzer was activated and the system was left to run for 1 hour. The start time and the end time were recorded. At the end of the 1-hour run, a reservoir of clean water was connected to wash the system. Second, third, and fourth samples of water with different known concentrations of NH 3 were prepared and introduced into the system in the same manner as the first. RESULTS The data retrieved from the concentration log file was used to show the stability and accuracy of the headspace system using H 2 S and NH 3 concentrations in water. Session 7.3: Page 9

TABLE V. H 2 S STABILITY TABLE Test Duration (hours) 1 Reading Stored (1/sec) 60 Average Value (PPM) 980.54 Maximum Value (PPM) 987.75 Minimum Value (PPM) 968.87 Standard Deviation (PPM) 6.54 Slope (PPM/reading).36 Slope * 60 (PPM/hour) 2.35 H 2 S Stability 2000.00 Concentration (ppm) 1500.00 1000.00 500.00 0.00 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 Time (seconds) FIGURE 5. H 2 S STABILITY GRAPH The first set of data is the stability of H 2 S in water over 1 hour. The average value over the 1-hour run was determined to be 980.54, with the maximum reading at 987.75 and the minimum reading at 968.87. The standard deviation of the data set is 6.54 and it has a slope of.36. TABLE VI. NH 3 STABILITY Test Duration (hours) 1 Reading Stored (1/sec) 60 Average Value (PPM) 999.99 Maximum Value (PPM) 1009.82 Minimum Value (PPM) 990.73 Standard Deviation (PPM) 4.09 Slope (PPM/reading) 0.02 Slope * 60 (PPM/hour) 1.10 Session 7.3: Page 10

NH 3 Stability 2000.00 ) 1800.00 m p1600.00 (p 1400.00 n1200.00 tio 1000.00 tra 800.00 n 600.00 e c 400.00 n o 200.00 C 0.00 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 55 58 61 64 Time (seconds) FIGURE 6. NH 3 STABILITY GRAPH The second set of data is the stability of NH 3 in water over 1 hour. The average value over the 1-hour run was determined to be 999.99, with the maximum reading at 1009.82 and the minimum reading at 990.73. The standard deviation of the data set is 4.09 and it has a slope of 0.02. TABLE VII. H 2 S ACCURACY TABLE Liquid Concentration Sample 5 (% of Cal Sample) 43.5 Liquid Concentration Sample 6 (% of Cal Sample) 61.25 Liquid Concentration Sample 7 (% of Cal Sample) 91.25 Liquid Concentration Sample 8 (% of Cal Sample) 100 Gas Concentration Sample 5 (PPM) 1300 Gas Concentration Sample 6 (PPM) 2600 Gas Concentration Sample 7 (PPM) 3800 Gas Concentration Sample 8 (PPM) 4200 Linearity R 2 Value.9996 The third set of data is the accuracy of H 2 S in water. The gas concentration reading for Sample 1 was 1300ppm. The gas concentration reading for Sample 2 was 2600ppm. The gas concentration reading for Sample 3 was 3800ppm. The gas concentration reading for Sample 4 was 4200ppm. The linearity (R 2 value) of the gas concentration readings was.9996. Session 7.3: Page 11

Gas Concentration (PPM) 4500 4000 3500 3000 2500 2000 1500 1000 500 0 H2S Accuracy R² = 0.9996 0 20 40 60 80 100 120 Liquid Concentration (% of Cal Sample) FIGURE 7. H 2 S ACCURACY GRAPH TABLE VIII. NH 3 ACCURACY TABLE Liquid Concentration Sample 5 (% of Cal Sample) 25 Liquid Concentration Sample 6 (% of Cal Sample) 50 Liquid Concentration Sample 7 (% of Cal Sample) 75 Liquid Concentration Sample 8 (% of Cal Sample) 100 Gas Concentration Sample 5 (PPM) 115.80 Gas Concentration Sample 6 (PPM) 184.15 Gas Concentration Sample 7 (PPM) 245.10 Gas Concentration Sample 8 (PPM) 314.06 Linearity R 2 Value 0.9994 The fourth set of data is the accuracy of NH 3 in water. The gas concentration reading for Sample 5 was 115.80ppm. The gas concentration reading for Sample 6 was 184.15ppm. The gas concentration reading for Sample 7 was 245.10ppm. The gas concentration reading for Sample 8 was 314.06. The linearity (R 2 value) of the gas concentration readings was.9994. Session 7.3: Page 12

Gas Concentration (PPM) 350 300 250 200 150 100 50 0 NH3 Accuracy R² = 0.9994 0 20 40 60 80 100 120 Liquid Concentration (% of Cal Sample) FIGURE 8. NH 3 ACCURACY GRAPH CONCLUSIONS Refineries create sour water as a byproduct in many of their processes. Sour water has been proven to cause corrosion and plugging due to the presence of ammonium salts, such ammonium bisulfide. Therefore, it is beneficial for refineries to remove the hydrogen sulfide and ammonia so the water can be reused or discarded. This is often done through a stripping process. Measurement of the H 2 S and NH 3 in sour water occurs afterwards to ensure the effectiveness of the stripper before the water is sent for reuse. The measurement is not particularly easy without advanced sampling techniques. Therefore, a sampling technique was developed that used some of the same principles of stripping that the refineries themselves use to remove H 2 S and NH 3 from the sour water stream. After the concentrations of each of these chemicals are measured in the gas phase, they are correlated back to a liquid concentration. In conclusion, the results from the stability test demonstrate a drift of less than 0.5% full scale. Also, the headspace system yields a linearity of 0.9994 or better when comparing carrier gas to sample stream analyte concentrations. Coupled with the benefits mentioned in the introduction, these analytical results make the UV-VIS full-spectrum spectrometer Session 7.3: Page 13

with the headspace sample conditioning system an ideal solution for the measurement of H 2 S and NH 3 in sour water. REFERENCES 1. Armstrong, Tim, Scott, Bruce, Taylor, Kin, and Gardner, Art, Sour Water Stripping, Today s Refinery, June, 1996. 2. Mark, Howard, Workman, Jerome, Classical Least Squares, Part 1: Mathematical Theory, Spectroscopy, May, 2010. 3. Zumdahl, Steven S., Zumdahl, Susan A., Chemistry 6 th Edition Houghton Mifflin, Boston, Massachusetts, 2006. 4. Sander, Rolf, Compilation of Henry s Law Constants for Inorganic and Organic Species of Potential Importance in Environmental Chemistry, Max Planck Institute of Chemistry, April 1999. 5. Horvath, Richard J., Lagand, Vishal V., Srinivasan, Sridhar, Kane, Russel D., Prediction and Assessment of Ammonium Bisulfide Corrosion Under Refinery Sour Water Service Conditions Part 2., NACE International, 2010. Session 7.3: Page 14