General Accreditation Guidance. Liquid-in-glass thermometers selection, use and calibration

Similar documents
technical guide Working Thermometers Calibration Procedures

Specific Accreditation Guidance

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing

CHEM 334 Quantitative Analysis Laboratory

Thermometer Line. today s laboratory and field

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

WACEL AGGREGATE LABORATORY TESTING TECHNICIAN

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

[Unless otherwise stated all tolerances are ±5%] SAMPLES

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

MIXTURES, COMPOUNDS, & SOLUTIONS

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

PURE PHYSICS THERMAL PHYSICS (PART I)

By D.B. Radtke, J.K. Kurklin, and F.D. Wilde. Temperature... T Equipment and supplies A Maintenance, cleaning, and storage...

Density of an Unknown

Unit 11: Temperature and heat

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

A MODIFIED HALDANE GAS ANALYZER FOR ANALYSIS OF MIXTURES WITH ONE HUNDRED PER CENT ABSORBABLE GAS

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level

Thermometry And Thermal Expansion

High temperature He is hot

Station 1: Adiabatic Decompression Materials: Can of compressed air Infrared thermometer Graph paper

Dr.Salwa Alsaleh fac.ksu.edu.sa/salwams

MATERIAL AND EQUIPMENT STANDARD FOR. INHIBITOR FOR HYDROCHLORIC ACID AND HCl+HF DESCALING AND PICKLING SOLUTION FOR OIL AND GAS WELLS ORIGINAL EDITION

* Defining Temperature * Temperature is proportional to the kinetic energy of atoms and molecules. * Temperature * Internal energy

MS20 Laboratory Seawater Salinity and Density

Experiment 8 Introduction to Volumetric Techniques I. Objectives

Core practical 13: Determine a value for the specific latent heat of ice

PREPARATIVE TASK GRAND PRIX CHIMIQUE PETNICA SCIENCE CENTER, VALJEVO, SERBIA 9 TH -14 TH OCTOBER 2017

Lecture 2: Zero law of thermodynamics

GENERAL INSTRUCTIONS GENERAL PREPARATION

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Calibration of Temperature Block Calibrators

States of Matter: Solid, Liquid, and Gas

ICY HOT LAB. In at least one sentence, explain your reasoning behind your prediction of curve:

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

Sodium Chloride - Analytical Standard

Distillation of Liquids: Separation of 2-Propanol from Water by Fractional Distillation

Q11: WHAT IS A MEANT GOOD ELECTRICAL CONNECTION?

Unit 6: Thermal Physics

Cambridge International Examinations Cambridge Ordinary Level

Put sufficient ice cubes into water (1 M) and wait for equilibrium (both exist) (1 M)

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

SOUTH AFRICAN NATIONAL STANDARD

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education CHEMISTRY


Core practical 5: Investigate the oxidation of ethanol

6. Common Laboratory Techniques

Pearson Education Limited Edinburgh Gate Harlow Essex CM20 2JE England and Associated Companies throughout the world

States of Matter: Solid, Liquid, and Gas

Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Introduction. Measurement of temperature is generally considered to be one of the simplest and most accurate measurements performed in engineering.

PART I: MEASURING MASS

Introduction to Small Scale Chemistry

A STUDY OF REACTION RATES

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education CHEMISTRY

Determination of an Equilibrium Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16

Experimental Procedure. Lab 406

Dr. L. I. N. de Silva. Student Name Registration Number: Assessed By: Lecturers Remarks

MEASUREMENT: PART II

Cambridge International Examinations Cambridge International General Certifi cate of Secondary Education

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

GENERAL INSTRUCTIONS GENERAL PREPARATION

STANDARD OPERATING PROCEDURE

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Ordinary Level

Cambridge International Examinations Cambridge International General Certificate of Secondary Education

5072 CHEMISTRY (NEW PAPERS WITH SPA) TOPIC 1: EXPERIMENTAL CHEMISTRY 5067 CHEMISTRY (NEW PAPERS WITH PRACTICAL EXAM) TOPIC 1: EXPERIMENTAL CHEMISTRY

Chapter 17 Temperature and heat

Al-Saudia Virtual Academy Online tuiton Pakistan Online Tutor Pakistan. Heat

1. How much heat was needed to raise the bullet to its final temperature?

Basic Equipment. 2D.1 Equipment for Measuring Mass

#09 Investigating the Relationship between the Mass of a Liquid and its Volume Ken Lyle, St. John s School, Houston, TX

Freezing point depression (Item No.: P )

Distillation. Boiling

Sodium Chloride - Analytical Standard

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

Heat and Temperature

The Synthesis and Analysis of Aspirin

Basic Analytical Techniques, Calorimeter, and Conductivity Meter

SPECIFIC HEAT OF WATER LAB 11-2

Cambridge International Examinations Cambridge International General Certificate of Secondary Education

Titration of citric acid in juice

[Unless otherwise stated all tolerances are ±5%] SAMPLES

Experiment 12: Grignard Synthesis of Triphenylmethanol

ASEN 2002 Experimental Laboratory 1: Temperature Measurement and an Blow Dryer Test

ESSENTIAL EXPERIMENTS CHEMISTRY

ST. STEPHEN S GIRLS COLLEGE Mid Year Examination PHYSICS Time Allowed: 1 hour 30 minutes NAME: F.3 ( ) MARKS:

Po Kok Secondary School S.1 Integrated Science Chapter 1.1 Classwork What is Science? Class: S.1 ( ) Date: A. What is Science? P.

Latent Heat of Fusion of Ice

Properties of Gases. The perfect gas. States of gases Gas laws Kinetic model of gases (Ch th ed, th ed.) Real gases

Ideal Gas Law. To demonstrate the dependence of pressure on temperature for a fixed volume of real gas.

3. EFFECTS OF HEAT. Thus, heat can be defined as a form of energy that gives the sensation of hotness or coldness

PHYS102 Previous Exam Problems. Temperature, Heat & The First Law of Thermodynamics

Lesson 1 Solids, Liquids, and Gases

European Association of National Metrology Institutes

SPECIFIC HEAT CAPACITY AND HEAT OF FUSION

Ordinary Level Physics Long Questions: TEMPERATURE AND HEAT

METEOROLOGY 110 Lab 4 Name. Meteorological Measurements

Transcription:

General Accreditation Guidance Liquid-in-glass thermometers selection, use and calibration January 2018

Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright under the Commonwealth of Australia Copyright Act 1968. NATA s accredited facilities or facilities seeking accreditation may use or copy this publication or print or email this publication internally for accreditation purposes. Individuals may store a copy of this publication for private non-commercial use or copy a reasonable portion of this publication in accordance with the fair dealing provisions in Part III Division 3 of the Copyright Act 1968. You must include this copyright notice in its complete form if you make a copy of this publication. Apart from these permitted uses, you must not modify, copy, reproduce, republish, frame, upload to a third party, store in a retrieval system, post, transmit or distribute this content in any way or any form or by any means without express written authority from NATA.

Liquid-in-glass thermometers selection, use and calibration 1. Introduction This guidance document describes procedures for selecting liquid-in-glass thermometers, their use, the rectification of faults, and the comparison of working thermometers against calibrated reference thermometers via ice-point checks and single point checks at the temperature of use. 2. Construction and behaviour of thermometers A liquid-in-glass thermometer consists of a bulb, a fine uniform bore capillary and one or more expansion chambers at the top of the column. There may also be an expansion chamber between the bulb and the zero mark or at the beginning of the graduated scale. The liquid filling may be mercury or a suitable liquid such as ethanol or toluene coloured with a non-staining light-fast dye. The glass in a thermometer, when first made, may be in a highly stressed state. This stress slowly relaxes with time, causing the bulb to contract, rapidly at first but decreasing with time. This continuous, irreversible change is called the secular change and should be monitored by checking the reading of a working thermometer at the ice-point or at one point in working range, every six months. Since the secular change is due to contraction of the bulb, its effect is to raise the reading of the thermometer and a correction must be applied to the thermometer readings to determine the true temperatures. This correction will apply equally to readings at all points of the scale and for reference thermometers it is only necessary to repeat the entire scale calibration at ten-yearly intervals. For working thermometers it is necessary to repeat an entire scale calibration at five-yearly intervals. If a wide range mercury-in-glass thermometer is heated to its maximum temperature and then reduced to ambient temperature, a temporary depression of the ice-point will be caused through bulb expansion which may take several days to recover. Mercury-in-glass thermometers can be used over the range 38 C to 450 C. Mercury has a large coefficient of thermal expansion which is uniform over this range and it forms a well-shaped meniscus in a fine bore glass tube. Special thermometers containing a thallium/mercury alloy permit measurements to be made down to 50 C. Temperatures below 38 C can be measured using alcohol or toluene thermometers, which cover the ranges 80 C to 75 C and 80 C to 110 C respectively. Since the liquids wet glass, such thermometers should not be warmed too quickly as this may cause the formation of gas bubbles. Cooling too quickly can lead to drainage problems. 3. Thermometer calibration schedule NATA criteria for the calibration and checking of liquid-in-glass thermometers are specified in the NATA Accreditation Criteria. 4. Inspection and repair Before carrying out calibration checks, each thermometer should be carefully examined for defects such as bubbles of gas trapped in the bulb, breaks in the liquid column, uneven graduations and faults in markings. Gas inclusions may be removed and broken liquid columns rejoined by carefully cooling the bulb of the thermometer in a mixture of ice and methanol, dry ice and methanol, or liquid nitrogen, depending upon the temperature range of the thermometer. This will draw all the liquid into the bulb and allow the gas inclusion to be expelled, or the broken thread to be rejoined. Cooling too rapidly may freeze the liquid and break the bulb, and should be avoided. Gently tap the thermometer if the broken thread sticks in the bore. Spirit left in the expansion chamber or near the top of the thermometer can be distilled off by gently heating the stem. January 2018 Page 3 of 7

5. Types of thermometer Types of thermometers Temperature range ( C) Smallest division ( C) Expected accuracy ( C) Laboratory Precision* 40 to 100 25 to 100 5 to 50 5 to 105 35 to 30 0.2 0.2 0.2 0.05 0.02 General Purpose 80 to 30 20 to 60 10 to 50 10 to 110 10 to 250 10 to 360 Industrial 38 to 27 10 to 250 90 to 260 0.3 Special Types: Maximum reading Max/Min reading Soil Gas Main Dairy 0 to 200 30 to 50 0 to 115 0 to 60 10 to 110 *Those listed as laboratory precision thermometers are suitable for use as reference thermometers for calibrating working thermometers. 6. Selection of thermometers for precision and general use The choice of a reference thermometer is determined by the range of temperature measurements to be made and by the required uncertainty of measurement. In general, the uncertainty of calibration of a reference thermometer should be 1/5th of the uncertainty of calibration required of the working thermometer. The reference thermometer should have scale divisions equal to, or less than, the uncertainty of measurement required of the working thermometer. For most purposes, a thermometer with a scale marked in divisions of or 0.2 is suitable as a reference thermometer. When selecting any thermometer for use, it should be checked for compliance with the relevant Australian Standard or other standard, for example: AS 1006 1995: Solid-stem general purpose thermometers It should also be examined to ensure that the glass is free of flaws which could distort the reading and that the bore and expansion chambers are free from oxidised deposits. 6.1 Selection of a total or partial immersion thermometer The choice of a total or partial immersion thermometer is dependent upon the working environment in which the temperature is to be measured. For example, if it is required to measure the temperature of ambient air in a room or a heated enclosure with a glass window, or of a liquid in a glass container, a total immersion thermometer should be used. It should be fully immersed or placed so that only 1 or 2 divisions of the scale are outside the medium to be measured. A partial immersion thermometer, immersed to the depth marked on the stem, should be used where the temperature is to be measured within a heated enclosure such as an oven, chamber or metal tank, or in a thermometer pocket, or in any situation where the depth of immersion is limited or where a thermometer cannot be moved. January 2018 Page 4 of 7

An error of several degrees can result if a total immersion thermometer is used at partial immersion or if a partial immersion thermometer is used at full immersion. 7. Ice-point checks Working thermometers should be checked every six months at the ice-point (0 C) for changes in calibration due to relaxation of the glass. (Alternatively, working thermometers can be checked at one point in working range against a reference temperature measuring device). The reference thermometer should be properly conditioned and checked at ice point to ascertain any secular change, before being used. When a thermometer is heated and then cooled to its initial temperature, the bulb does not contract immediately to its original volume. It is therefore important that the thermometer be conditioned by the same method before any determination of the secular change. When the complete calibration of a thermometer is to be relied upon to within an uncertainty of 0.02 C, the reference temperature should be read as precisely as possible. The temperature of a carefully prepared icepoint is reproducible to better than 0.001 C and, by following the method given below, it is possible under optimum conditions to determine the thermometer reading at the ice point to within 0.003 C. 7.1 Ice-point check to an uncertainty of less than 0.02 C The ice-point apparatus consists of a thermally-insulated container such as a Dewar flask or other container about 25 cm deep and provided with a means of supporting the thermometers. The thermometer scales should be observed through a 5X magnification viewer which can be adjusted to avoid parallax error. 7.1.1. Wash the thermometer with distilled water and store at 0 C for several hours before taking readings at the ice point. 7.1.2. Prepare a mixture of distilled water and finely crushed ice made from distilled water. Mix it until it no longer appears opaque. Place it in the insulated container, stirring thoroughly and draining if necessary until there is no free water. 7.1.3. Fully immerse the thermometer in the ice mixture to above the mercury column. Before taking a reading, lift the thermometer until the mercury is just visible through the viewer. It should then be raised a few centimetres and quickly lowered to the reading position. 7.1.4. Repeat this operation two or three times, or until there is no change in the reading. Wipe the thermometer stem with absorbent paper and take a reading. Note: The purpose of this procedure is to avoid the tendency for the mercury to stick in the capillary bore. The magnitude of this effect is unpredictable and may amount to as much as a whole scale division. 7.1.5. Keep a record of all ice-point readings obtained during the life of the thermometer so that any anomalies may be detected. 8. Comparison of working thermometers with a reference thermometer 8.1 General procedures The reference and working total immersion thermometers should be supported side-by-side in a temperature-controlled bath in such a manner that they may be easily raised and lowered. The bath must be allowed to stabilise at the temperature at which the readings of the thermometers are to be compared. If the thermometers are to be compared at a particular temperature, then two sets of readings should be taken, on either side of the required temperature, and the two sets of readings averaged to find the correction to be applied to the readings of the working thermometer at the specified temperature. An example of this procedure is given below, including sample readings and the steps involved in calculating the corrections and readability. January 2018 Page 5 of 7

8.2 Correction of partial immersion thermometers When a thermometer is used at partial immersion, the temperature of the emergent liquid in glass column may be different from that at the time of calibration. For this reason, the mean temperature of the emergent column must be measured and a correction applied to adjust the reading of the thermometer to the conditions at the time of calibration. The procedure for calculating the emergent stem correction is detailed in NMI Monograph 9 Liquid-in-Glass Thermometry, by Corinna Horrigan and Robin E. Bentley. Example To check a thermometer at the 40.0 C mark. Firstly, check the calibration at 39.9 C. Thermometer Reference (ºC) Working (ºC) Reading 1 39.88 39.8 Reading 2 39.92 40.0 Averages 39.90 39.9 Reference correction from calibration certificate +0.02 Corrected reference temperature (1) 39.92 Subtract average bath temperature from average working thermometer reading to obtain error in working thermometer Correction (1) to working thermometer (change sign) ( 39.92) 0.02 +0.02 Note: If the working thermometer is only being checked at one point then only one set of readings is required. If the thermometer is being subjected to an initial check before being put into service then readings should be taken at approximately ten degree intervals along its scale. Secondly, check the calibration at 4 C. Thermometer Reference ( C) Working ( C) Reading 1 40.08 4 Reading 2 46 40.2 Averages 42 45 Reference correction from calibration certificate +0.02 Corrected reference temperature (2) 44 Subtract average temperature from average working thermometer reading to obtain error in working thermometer Correction (2) to working thermometer (change sign) Average of correction (1) and (2) to working thermometer at 40 C Round to C, the readability of the thermometer ( 44) +0.01 0.01 +0.005 0.0 Note: Working thermometers may alternatively be checked against a calibrated platinum resistance thermometer. January 2018 Page 6 of 7

9. References AS 1006 1995: Solid-stem general purpose thermometers. NMI Monograph 9 Liquid-in-Glass Thermometry, by Corinna Horrigan and Robin E. Bentley AMENDMENT TABLE The table below provides a summary of changes made to the document with this issue. Section Amendment Entire document This document replaces the former Technical Note 19. The document has been renamed to reflect the new accreditation criteria documentation structure. January 2018 Page 7 of 7