Test certificate Page 1 of 4 Issued by Hugo de Grootplein 1 3314 EG Dordrecht The Netherlands Notified Body Number 0122 In accordance with Applicant In respect of Paragraph 8.1 of the European Standard on Metrological aspects of non-automatic weighing instruments EN 45501:1992/AC:1993 and by application of the OIML International Recommendation R60 (Edition 1991). The applied error fraction p, meant in the paragraph 3.5.4. of the standard is 0.7. Precia Molen B.V. Franse Akker 1 4824 AL Breda The Netherlands A compression load cell, with strain gauges, tested as a part of a weighing instrument. Manufacturer : Molen or Precia Molen Type : CSP-M- t-..-c..-ss/eex(d)/eex(i) Characteristics Maximum capacity (E max ) Accuracy class Maximum number of load cell verification intervals (n max ) Ratio of minimum LC verification interval Y = E max / V min 5, 10, 25, 40, 60 or 100 t C 1000 2000 3000 3000* 3500 3500* 5000 10000 10000 24000 11666 24000 * = special type with additional type description MB In the description number TC2194 revision 3 further characteristics are described. Nederlands Meetinstituut Hugo de Grootplein 1 3314 EG Dordrecht Telephone +31 78 6332332 Telefax +31 78 6332309 NMi B.V. (Chamber of Commerce no.27.228.701) Subsidiary companies: NMi Van Swinden Laboratorium B.V. (27228703) (27.233.418) Verispect B.V. (27.228.700) This document is issued under the provision that NMi. B.V. nor its subsidiary companies accept any liability. Reproduction of the complete document is allowed. Parts of the document may only be reproduced after written permission.
Test certificate Page 2 of 4 Description and The load cell is described in the description number TC2194 revision 3 and documentation documented in the documentation folder TC2194-3, appertaining to this test certificate. Remarks Summary of the test involved: see Appendix number TC2194 revision 3. This revision test certificate replaces the earlier version(s), including its documentation folder. Dordrecht, 30 June 2008 Ing. C. Oosterman Manager Product Certification
Description Page 3 of 4 1 General information about the load cell All properties of the load cell, whether mentioned or not, may not be in conflict with the standard mentioned in the test certificate. 1.1 Essential parts Description Drawing number Rev. Remarks Load cell assembly CSP-M-40-EEx-C 7.882.005.0.635 0 Mechanical Outline Dimensions Special CSP-M 10t / 100 t 799523 - Mechanical Load cell Assy. type CSP 7.882.005.0.638 0 Electrical Cable: - The load cell is provided with a 4-wire system; - The cable length has to be approximately 10 meters for capacity of 5 t and 10 t; - The cable length has to be approximately 20 meters for capacity of 25, 40, 60 and 100 t; - The cable should be a shielded cable, the shield is connected to the load cell. 1.2 Essential characteristics Minimum dead load : 0 kg Safe overload : 150 % of E max Rated Output : 2 mv/v Input impedance : 450 Ω ± 4.5 Ω Output impedance : 480 Ω ± 4.8 Ω Recommended excitation : 5-20 V DC/AC Excitation maximum : 25 V DC/AC See also drawing Specifications load cell CSP-M-..t-., drawing number HV-930217.079. 1.3 Essential shapes The load cell is built according to the drawings: - Load cell assembly CSP-M-40-EEx-C, drawing number 7.882.005.0.635; - Outline Dimensions Special CSP-M 10t / 100 t, drawing number 799523. The data plate is secured against removal by sealing or will be destroyed when removed. The data plate mentions at least the information and markings as described in the OIML R60 document. In the countries where it is mandatory the load cell should bear this test certificate number: TC2194. Securing: The connecting cable of the load cell or the junction box is provided with possibility to seal.
Appendix Page 4 of 4 Tests performed for this test certificate: Test Institute type, version, remarks Temperature test and repeatability (20, 40, -10 and 20 C) Temperature effect on minimum dead load output (20, 40, -10 and 20 C) CSP-M-25t-SC-C3.5-MB CSP-M-25t-SC-C3.5-MB Creep (20, 40 and 10 C) CSP-M-25t-SC-C3.5-MB Minimum dead load output return (20, 40 and 10 C) CSP-M-25t-SC-C3.5-MB Barometric pressure effects at room temperature CSP-M-25t-SC-C3.5-MB Humidity test CSP-M-25t-SC-C3.5-MB
Documentation folder Number TC2194-3 Page 1 of 1 Description Drawing number Rev. Remarks Load cell assembly CSP-M-40-EEx-C 7.882.005.0.635 0 Mechanical Outline Dimensions Special CSP-M 10t / 100 t 799523 - Mechanical Load cell Assy. type CSP 7.882.005.0.638 0 Electrical Specifications load cell CSP-M-..t-. HV-930217.079 -