Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Similar documents
Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

The New Agilent 7700 Series ICP-MS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

The 30-Minute Guide to ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Trace Analyses in Metal Matrices Using the ELAN DRC II

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Atomic Spectroscopy AA/ICP/ICPMS:

Agilent Technologies at TIAFT 2013

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Agilent 8900 Triple Quadrupole ICP-MS

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Partial Energy Level Diagrams

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Fergus Keenan Shona McSheehy Julian Wills

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

OES - Optical Emission Spectrometer 2000

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

ICP-MS. Inductively Coupled Plasma Mass Spectrometry. A Primer

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Samples compliant with the WATER

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Lecture 7: Atomic Spectroscopy

Introduction to the Q Trap LC/MS/MS System

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Extrel Application Note

Agilent 7700 Series. The Smallest, Most Powerful ICP MS Ever Made

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

Trace elemental analysis solutions for your application. June 6, 2018

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Spectrometric Methods of Analysis. OCN 633 Fall 2013

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Mass Spectrometry in MCAL

ICPMS Doherty Lecture 1

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Agilent ICP-MS Journal September 2009 Issue 39

Sources of Errors in Trace Element and Speciation Analysis

ICP-OES Application Note Number 35

Sample Analysis Design PART II

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Analysis Repeatability of Trace and Major Elements in a Water Sample

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Interference-free determination of ultra-trace levels of Arsenic and Selenium using methyl fluoride as reaction gas in ICP-MS/MS

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Sample Analysis Design PART III

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Optical Atomic Spectroscopy

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

CHAPTER 4: ANALYTICAL INSTRUMENTATION

Transcription:

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry

Outline Some background and fundamentals of ICPMS Proposed ICPMS method for ASTM Preliminary Results for D7111-11 Preliminary Results for D5185-09 Advantages of ICP-MS Summary

Inductively Coupled Plasma Mass Spectrometry Key applications are: Environmental Foods Semiconductor Clinical Chemical/Petrochemical Pharmaceutical Consumer Goods Forensic Geological Nuclear Academic/Research ICP-MS market growing fast as technology improves and encroaches upon traditional GFAA and ICP-OES markets Page 3 October 21, 2013

Comparison of Elemental Analysis Techniques for Liquid Samples

Firstly, What is ICP-MS? An elemental analysis technique ICP - Inductively Coupled Plasma MS - Mass Spectrometer high temperature ion source decomposes, atomizes and ionizes the sample Uses quadrupole mass analyzer Mass range from 7 to 250 amu (Li to U...) separates all elements in rapid sequential scan isotopic information available Measures ions, using dual mode detector Sub ppt to ppm levels ICP-MS combines the detection limits of Graphite Furnace AA (or better) and the sample throughput of ICP-OES Page 5 October 21, 2013

Simplified Schematic the functional components Sample Introduction Plasma Ion Source Ion lenses Collision/ Reaction Cell Quadrupole MS Detector Converts liquid sample to aerosol Decomposes sample matrix and forms ions Focuses ions and removes photons and neutrals Removes spectral interferences (polyatomic ions) Separates ions by m/z unit mass resolution Detects ions and transfers counts to data system Nearly all modern ICP-MS instruments employ some form of Collision/Reaction Cell to remove polyatomic (molecular) interferences Page 6 October 21, 2013

General ICP-MS Components/Technology 1. The liquid sample is mixed with argon gas by the nebulizer to form an aerosol.. 3. The sample is desolvated and ionized in the plasma 5. Ion lenses focus and collimate the ions. The Omega lens bends the ion beam offaxis to prevent photons from striking the detector 7. The quadrupole mass spectrometer separates ions based on their mass to charge ratio. The selected ions continue on to the detector 2. The smallest droplets pass through the spray chamber and into the ion source - the plasma 4. Ions are extracted from the plasma by extraction lenses in the interface region 6. Interference Ions are removed from the ion path by reaction/collision/ked effects in the cell 8. Ions are measured using a discrete dynode detector providing 9 orders of linear dynamic range

Processes in ICP-MS Plasma Mostly the same as ICP-OES

The ICP-MS Torch Torch and plasma nearly identical to ICP-OES

ICP-MS Plasma/Vacuum Interface Plasma gases and sample ions pass through sample cone into the interface, or expansion stage First stage of 3 stage vacuum system Gases expand into a vacuum The ions are sampled by the skimmer cone and pass into the intermediate stage Cone geometry is critical for efficient transport of ions Extraction lenses create a voltage field behind skimmer. Voltage field begins the process of focusing the ion beam. Ext 1 Ext. 2

Principle of for removing polyatomic interferences He Mode and KED (Kinetic Energy Discrimination) Polyatomic ions Analyte ions Energy distribution of analyte and interfering polyatomic ions with the same mass Polyatomic ions Bias voltage rejects low energy (polyatomic) ions Analyte ions Energy Energy At cell entrance, analyte and polyatomic ion energies overlap. Energy spread of both groups of ions is narrow, due to ShieldTorch System Cell Entrance Energy loss from each collision with a He atom is the same for analyte and polyatomic ion, but polyatomics are bigger and so collide more often Cell Exit By cell exit, ion energies no longer overlap; polyatomics are rejected using a bias voltage step. Analyte ions have enough residual energy to get over step; polyatomics don t (energy discrimination) Page 11 October 21, 2013

Energy Discrimination Ionic Radii m / z 75 m / z 51 Any polyatomic species will have a larger-cross section than single ions The larger polyatomics will collide with the cell gas a greater number of times than the smaller analyte ions and loose energy Low energy ions cannot enter the QP Approx Ionic Radii (pm) 300 250 200 150 100 50 0 As+ ArCl+ Simple Species V+ ClO+

Energy Discrimination Ionic Radii Energy Discrimination is NOT a reaction process does NOT rely on reaction pathways that are species dependant Simply acts as a physical filter for the larger polyatomic ions Is applicable for ANY polyatomic Approx Ionic Radii (pm) 400 350 300 250 200 150 100 50 0 Anything larger than 150pm will be filtered out Cu S 2 m / z 65 SO 2 NOCl Complex Species ArMg CaOH

ORS 3 Reaction Mode Another tool for removing interferences 1. On-Mass Measurement: Unreactive analyte does not react with chosen cell gas, remains at original m/z and so can be separated from reactive interferences Reaction gas Reaction product ion On-mass interference Analyte Analyte Analyte and interfering ions enter reaction cell Interference M + MR + Interference reacts to form product ion Quad set to original analyte mass rejects interference product ion(s) Reactive interferences are converted to product ions at a new mass can be rejected by analyzer quad, which is set to original analyte mass

ORS 3 Reaction Mode Another tool for removing interferences 2. Mass-Shift Measurement: Reactive analyte reacts with chosen cell gas, is moved to a new product ion mass and can be separated from unreactive interferences Reaction gas Original interfering ion On-mass interference Analyte Analyte and interfering ions enter reaction cell Analyte M + MR + Analyte reacts to form product ion Analyte product ion Quad set to analyte product ion mass rejects original interfering ions Reactive analyte is converted to product ions at a new mass interferences remain at original mass and are rejected by analyzer quad

Quadrupole Mass Analyzer Also referred to as the quad, or QP 4 hyperbolic rods are electronically paired. RF and DC voltages are applied and ramped - x and y rods are out of phase with each other. QP behaves as a filter -- all masses may enter, however only one mass exits the QP at any point in time - All ions except the target ion are unstable they exit the QP out to the side, and are pumped away by the vacuum system. - QP continuously scans the mass range during acquisition. - Can acquire each mass between 0-260 amu in approx. 0.1s. Single mass at QP exit

Electron Multiplier Detector Ions exiting the QP strike the first dynode of the detector. Creating a cascade of electrons which is amplified and counted One ion results in one pulse of electrons.

Key Hardware and Software Requirements For Organic Analysis O 2 Option Gas MFC Added to Aux gas at torch Allows Organics to be analyzed directly by ICPMS Auto setup of conditions New MassHunter Software and Torch Design Modified firmware and software startup settings ensure reliable ignition with organic solvents New torch design provides improved tolerance to volatile organic solvents ORS 3 Octopole Reaction System 3 rd Generation Octopole Allows analyses of difficult elements - Si, P, S & Se

Preliminary Results D7111-11 Determination of Trace Elements in Middle Distillate Fuels (Modified for ICP-MS)

Preliminary Results D7111-11 Modified for ICP-MS Instrument Acquisition mode(s) Sample introduction devices Sample uptake Interface Autosampler Agilent 7700x ICP-MS Hydrogen and High Energy Helium (HEHe) cell modes Glass nebulizer and 700 mm length capillary tube Quartz torch with 1.5 mm bore injector Natural aspiration Platinum sampling cone and skimmer cone I-AS Reagents Standard solutions of organic metals were CONOSTAN oil standards. Kerosene, as the diluent solvent, was purchased from Kanto Chemicals. No further purification was applied to kerosene. Calibration 5ppm (w/w in kerosene) of mixed standard was prepared from CONOSTAN standards. 0, 2, 10 and 50ppb (w/w) solutions were obtained by serial dilution of this stock standard solution. Yttrium internal standard was added into each working standard solution.

Long term stability 7 hours continuous analysis 10ppb spiked diesel oil 23 Na [ H2 ] 1.9% 1.2 24 Mg [ H2 ] 1.8% Normalized Concentration 1 0.8 0.6 0.4 0.2 0 23 Na [ H2 ] 24 Mg [ H2 ] 31 P [ HEHe ] 40 Ca [ H2 ] 48 Ti [ HEHe ] 51 V [ HEHe ] 52 Cr [ HEHe ] 55 Mn [ H2 ] 56 Fe [ H2 ] 59 Co [ HEHe ] 60 Ni [ HEHe ] 63 Cu [ HEHe ] 66 Zn [ HEHe ] 75 As [ HEHe ] 78 Se [ H2 ] 95 Mo [ HEHe ] 107 Ag [ HEHe ] 114 Cd [ HEHe ] 118 Sn [ HEHe ] 138 Ba [ H2 ] 208 Pb [ HEHe ] 0 50 100 150 200 250 300 350 400 450 Elapsed time, min 31 P [ HEHe ] 0.5% 40 Ca [ H2 ] 1.5% 48 Ti [ HEHe ] 1.8% 51 V [ HEHe ] 1.3% 52 Cr [ HEHe ] 1.1% 55 Mn [ H2 ] 0.9% 56 Fe [ H2 ] 1.0% 59 Co [ HEHe ] 1.1% 60 Ni [ HEHe ] 1.3% 63 Cu [ HEHe ] 0.9% 66 Zn [ HEHe ] 2.0% 75 As [ HEHe ] 2.7% 78 Se [ H2 ] 1.5% 95 Mo [ HEHe ] 1.4% 107 Ag [ HEHe ] 0.9% 114 Cd [ HEHe ] 1.3% 118 Sn [ HEHe ] 1.4% 138 Ba [ H2 ] 1.1% 208 Pb [ HEHe ] 1.4%

Preliminary Results D7111-11 Metallic impurities in fuel oils (ug/kg), ppb kerosene alcohol white gas A white gas B diesel oil regular gas high-octane gas 10 B H 2 n. d. 920 16 25 1.5 5 n. d. 23 Na H 2 n. d. 19 n. d. 25 4.1 n. d. n. d. 24 Mg H 2 n. d. 5.2 3.6 37 6.6 2.6 n. d. 27 Al HEHe 0.4 8 n. d. 31 8.3 1.4 n. d. 40 Ca H 2 0.3 6.4 3 42 10 11 0.3 48 Ti HEHe 0.1 0.9 0.2 33 8.1 1.8 0.2 51 V HEHe 0.5 0.5 0.4 31 7.7 1.4 0.9 52 Cr HEHe 0.1 1.8 1.3 37 8 1.9 0.3 55 Mn H 2 0.3 2.6 1.3 36 8 2.2 0.4 56 Fe H 2 1.2 41 0.6 39 8.9 10 0.6 59 Co HEHe 0.2 1.1 0.1 0.9 0.3 0.2 0.4 60 Ni HEHe 0.4 2 0.9 30 6.4 1.9 0.1 63 Cu HEHe 2.1 14 4.8 41 7.4 2.8 0.5 66 Zn HEHe 58 78 8.2 35 52 250 3.6 95 Mo HEHe 0.1 0.6 0.2 31 6.9 1.5 0.3 107 Ag HEHe 0.1 4.7 0.1 27 6.8 1.5 0.2 114 Cd HEHe 0.1 12 0.5 30 6.4 1.5 0.3 118 Sn HEHe n. d. 610 3.8 36 7.4 2 0.2 138 Ba H 2 0.9 0.2 0.1 30 8.4 17 0.4 208 Pb HEHe 0.3 0.9 0.4 27 6.6 1.6 0.4

Recovery of Certified Elements in NIST-1634C Residual fuel oil Element Reference mg/kg Result mg/kg Recovery Na 37 37.7 102% V 28.19 + 0.40 28.0 99% Co 0.1510 + 0.0051 0.147 97% Ni 17.54 + 0.21 17.2 98% As 0.1426 + 0.0064 0.148 104% Se 0.1020 + 0.0038 0.102 100% Ba 1.8 1.87 104%

Preliminary Results D5185-09 Determination of Additive Elements, Wear Metals, and Contaminants in Used Lubricating Oils and Determination of Selected Elements in Base Oils (Modified for ICP-MS)

Preliminary results D5185-09 Modified for ICP-MS Essentially same methodology, instrument conditions, calibration, and sample prep were used as for D7111-11 Samples were diluted into either Kerosene or O-Xylene Lubricating oils were purchased from various sources on common market

Calibrations prepared from Conostan 21 Dilution in kerosene Unit CALBLK CALSTD1 CALSTD2 CALSTD3 CALSTD4 Na ppm 0 0.005 0.02 0.1 0.5 Mg ppm 0 0.5 2 10 50 P ppm 0 0.5 2 10 50 Ca ppm 0 0.5 2 10 50 Zn ppm 0 0.5 2 10 50 Mo ppm 0 0.1 0.4 2 10 Al ppb 0 0.2 0.8 4 20 Fe ppb 0 0.2 0.8 4 20 Cd ppb 0 0.2 0.8 4 20 Ti ppb 0 0.2 0.8 4 20 V ppb 0 0.2 0.8 4 20 Cr ppb 0 0.2 0.8 4 20 Mn ppb 0 0.2 0.8 4 20 Ni ppb 0 0.2 0.8 4 20 Cu ppb 0 0.2 0.8 4 20 Ag ppb 0 0.2 0.8 4 20 Sn ppb 0 0.2 0.8 4 20 Ba ppb 0 0.2 0.8 4 20 Pb ppb 0 0.2 0.8 4 20

Method Performance Detection limits Analyte m/z mode DL (ppb) BEC(ppb) B 10 H 2 0.85 2.1 Na 23 H 2 0.79 1.9 Mg 24 H 2 0.18 2.5 Al 27 He 0.03 0.11 P 31 He 28 33 Ca 40 H 2 0.14 0.13 Ca 44 H 2 2.2 1.2 Ti 49 He 0.14 0.01 V 51 He 0.005 0.01 Cr 52 He 0.05 0.26 Fe 54 H 2 0.06 0.02 Mn 55 He 0.01 0.005 Ni 60 He 0.02 0.02 Cu 63 He 0.05 0.26 Zn 66 He 0.21 0.33 Zn 67 He 0.33 0.29 Mo 95 He 0.005 0.005 Ag 107 He 0.01 0.01 Cd 114 He 0.005 0.01 Sn 118 He 0.005 0.01 Ba 137 H 2 0.01 0.01 Pb 208 He 0.005 0.01 Concentration Measured (ppb) Recovery Spiked unspiked spiked (%) B 50 4.3 47 85 Na 50 4.3 69 119 Mg 50 2.9 57 108 Al 10 19 28 90 Ca 50 145 198 106 Ti 50 2.3 59 113 V 50 0.15 56 112 Cr 10 1.3 13 117 Mn 50 0.77 59 116 Fe 50 5.8 34 62 Ni 10 2.8 12 112 Cu 10 5.1 17 119 Zn 50 2 62 120 Mo 50 0.83 55 108 Ag 10 0.37 11 106 Cd 10 1.1 12 109 Sn 10 0.45 12 116 Ba 50 5.7 60 109 Pb 50 8.2 54 86 Kerosene Certified Value % (uncertainty) Calcium 0.36 0.359 (0.011) Magnesium 0.78 0.821 (0.038) Phosphorus 0.78 0.788 (0.028) Zinc 0.87 0.873 (0.022) Recoveries - NIST 1848 Lubricating Oil Additive Package

Summary ICP-MS methods can be easily adapted from existing ICP-OES methods with only minor changes. ICP-MS offers 2 or more orders increased sensitivity ICP-MS suffers fewer interferences easily managed using CRC technology ICP-MS can provide isotopic information useful for confirmation

Page 29