PHOSPHOR THERMOMETRY TECHNIQUES FOR THE REALIZATION OF THERMAL STANDARDS *

Similar documents
International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

How do you really know what the temperature is? Michael de Podesta

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

HT TRANSIENT THERMAL MEASUREMENTS USING THERMOGRAPHIC PHOSPHORS FOR TEMPERATURE RATE ESTIMATES

The International Temperature Scale of 1990

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Fundamental Temperature Measurement: Re-defining the Boltzmann Constant How do you really know what the temperature is? Michael de Podesta

Study on a thermal diffusivity standard for the laser flash method measurements 1

Calibration of temperature sensors within Length Standard Section of NMIJ

Thermometry at Low and Ultra-low Temperatures

EFFECTS OF PROTON IRRADIATION ON TRIBOLUMINESCENT MATERIALS SUCH AS ZNS:MN

Introduction to Blackbody Sources

Total radiation measurements of thermodynamic temperature

INC 331 Industrial Process Measurement. Instrument Characteristics

Abstract. 1. Introduction

Goodbye, Mr. kilogram!

Photonics applications II. Ion-doped ChGs

Academic and Research Staff. Prof. R. Weiss. Graduate Students. G. D. Blum T. R. Brown S. Ezekiel

Lecture 32. Lidar Error and Sensitivity Analysis

Chapter 7: Optical Properties of Solids. Interaction of light with atoms. Insert Fig Allowed and forbidden electronic transitions

Chemistry Instrumental Analysis Lecture 5. Chem 4631

1 THE CONCEPT OF TEMPERATURE

Nanometrology and its role in the development of nanotechnology

A NEW REALIZATION OF TERRESTRIAL TIME

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

IR spectral characterization of customer blackbody sources: first calibration results

Unit-2 LASER. Syllabus: Properties of lasers, types of lasers, derivation of Einstein A & B Coefficients, Working He-Ne and Ruby lasers.

Temperature. 3

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Calibrating the Thermal Camera

The Redefinition of the Kelvin

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

European Association of National Metrology Institutes

Transient measurements using thermographic phosphors

How do you really know what the temperature is? Michael de Podesta

Section 7. Temperature Measurement

by S. Solve +, R. Chayramy +, M. Stock +, D. Vlad*

The SI in 2019: Traceability Questions for SIM A Presentation to Encourage Discussion and Understanding Alan Steele SIM General Assembly Montevideo,

Chapter 1 Overview of Semiconductor Materials and Physics

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Good morning everyone, and welcome again to MSLs World Metrology Day celebrations.

A NEW REALIZATION OF TERRESTRIAL TIME

Ozone cross-sections (Hartley band) Progress towards new accurate values

MISE EN PRATIQUE OF THE REALIZATION OF THE KELVIN

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

Computational Study of Amplitude-to-Phase Conversion in a Modified Unitraveling Carrier Photodetector

SUMMARY OF THE BIPM QUANTUM METROLOGY WORKSHOP. CARL WILLIAMS PML DEPUTY DIRECTOR October 19, 2017

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

Thermographic Phosphors Temperature measurements using Laser Induced Phosphorescence (LIP)

Mass determination of 1 kg silicon spheres for Avogadro project

THERMOGRAPHIC PHOSPHORS AS A MEANS OF ESTIMATING HEATING RATE TO SOLVE THE INVERSE HEAT CONDUCTION PROBLEM. Paul R. Crim. Thesis

BUREAU INTERNATIONAL DES POIDS ET MESURES

The Kilogram Redefined:

CESIUM ATOMIC FOUNTAIN CLOCKS AT NMIJ

Excitation-Wavelength Dependent and Time-Resolved Photoluminescence Studies of Europium Doped GaN Grown by Interrupted Growth Epitaxy (IGE)

Lecture PowerPoint. Chapter 28 Physics: Principles with Applications, 6 th edition Giancoli

ScienceDirect. Gas flow visualization using laser-induced fluorescence

Investigation of Water Fragments

SCINTILLATORS AND WAVELENGTH SHIFTERS FOR THE DETECTION OF IONIZING RADIATION *

arxiv: v1 [physics.optics] 6 Jul 2009

Atomic clocks. Clocks

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

Quantum Technologies CCEM Workshop March 23 rd, 2017

MIT Department of Nuclear Science & Engineering

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

Phys 2310 Fri. Dec. 12, 2014 Today s Topics. Begin Chapter 13: Lasers Reading for Next Time

Multi-cycle THz pulse generation in poled lithium niobate crystals

Part Number Range Accuracy Printing WP * Probe Special User LDC with Page Logging included Feature Cal Backlight

Figure 1 Relaxation processes within an excited state or the ground state.

Traceability and Quality Control in a Radiation Thermometry Laboratory

Chemistry 11. Unit 2: Introduction to Chemistry. Measurement tools Graphing Scientific notation Unit conversions Density Significant figures

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

Lecture 1. Chapter 1 (Halliday/Resnick/Walker, Fundamentals of Physics 8 th edition)

Bureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya

Temperature-dependent spectroscopic analysis of F 2 + ** and F 2 + **-like color centers in LiF

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

Chapter 14 Temperature and Heat

Bureau International des Poids et Mesures

THE WATT BALANCE ROUTE TOWARDS A NEW DEFINITION OF THE KILOGRAM. A. Eichenberger, H. Baumann, and B. Jeckelmann.

Final Report August 2010

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

Spectral Response of GaAs(Cs, NF 3 ) Photocathodes

Introduction to metrology

Temperature and Synchrotron Radiation

Sub-percent precision Møller polarimetry in experimental Hall C

Le système SI et son évolution; le rôle de la seconde dans le future SI Terry Quinn

Spectroscopic investigations of Rb- and Cs- rare gas systems

RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

Chemistry 11. Measuring and Recording Scientific Data. Measurement tools Graphing Scientific notation Unit conversions Density Significant figures

New analysis for determination of hemispherical total emissivity by feedback-controlled pulse-heating technique

NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT

F.G. Major. The Quantum Beat. The Physical Principles of Atomic Clocks. With 230 Illustrations. Springer

In a metal, how does the probability distribution of an electron look like at absolute zero?

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Magnetic Field measurements with Csmagnetometers

Combined method for establishment and dissemination of the international temperature scale

Transcription:

PHOSPHOR THERMOMETRY TECHNIQUES FOR THE REALIZATION OF THERMAL STANDARDS * S. W. Allison and M. R. Cates G. T. Gillies Sr. Development Staff Members Research Professor Oak Ridge National Laboratory University of Virginia Oak Ridge, TN 37831-8058 Charlottesville, VA 22904-4746 KEYWORDS Thermographic, phosphors, temperature, SI units, thermal standards, calibration, secondary standards ABSTRACT The possibility of using single-crystals of thermographic phosphors in a new realization of the degree kelvin has been explored. The experimental technique would involve the laser-induced fluorescence of a rare-earth-doped phosphor, the optical decay lifetime of which would be measured relative to an atomic clock, in order to obtain a primary standard of temperature based only on atomic quantities. In principle the technique could provide the first realization of the kelvin over continuous ranges of temperature, without the need for interpolation between fixed thermodynamic transition points. Possible experimental arrangements for carrying out such a realization of the degree kelvin will be discussed. INTRODUCTION Thermometer calibrations at all standards laboratories are carried out within the context of the International Temperature Scale of 1990 (ITS-90). This temperature scale is maintained between the values of 13 K and 1234 K in terms the well-defined thermodynamic transition points (melting, freezing, vapor pressure and triple points) of several pure materials, including the gases hydrogen, neon, oxygen and argon, and the metals mercury, gallium, indium, tin, zinc, aluminum and silver, as well as with water (1). The base unit of the SI system associated with the measurement of thermodyanamic temperature, the degree kelvin, is defined as 1/273.16 of the thermodynamic temperature of the triple point of water (2). It is realized via interpolation over sub-ranges of the ITS-90 span, with calibration errors on the order of a few tenths of a millikelvin propagating from it. The structure, maintenance and improvement of the ITS-90 is made possible by a massive effort on the part of many physicists, chemists

and metrologists at national standards laboratories around the world, all of whom contribute to the quest for lower overall uncertainty in the realization of the degree kelvin via this scale. The ITS-90 is disseminated to industry by a variety of primary, secondary and transfer-standard thermometers. The overall realization and use of this temperature scale is thus intricately dependent on a wide variety of material artifacts, devices and interpolative algorithms, with no direct tie to atomic quantities that might be independent of such dependencies. The other base unit of the SI system for which a similar situation holds is the kilogram. At present the international prototype for it is a platinum-iridium mass preserved at the International Bureau of Weights and Measures in France. To realize the kilogram and disseminate its value, many similar such masses are intercompared with each other (and occasionally with the international prototype) during the course of several complex series of precision weighings. The mass standards of individual laboratories can then be calibrated against the kilogram as maintained at each of the various national standards laboratories. See, for instance, reference (3) for further details of that process. Ideally, all of the base units of the SI system would be realized in terms of atomic quantities that are independent of artifacts. The units could then be established with essentially equal precision at any laboratory anywhere. Indeed, great effort is being expended on research into the realization of the kilogram on the basis of building it from single crystals of pure silicon, the atomic planes of which are regular enough in form and structure to permit single-atom counting and summation. Our purpose here is to propose something similar in terms of the realization of the degree kelvin. By measuring the laserinduced fluorescence lifetimes of very regular single crystals of rare-earth-doped ceramic oxides and/or oxysulfides in terms of the ticks of an atomic clock, one could in principle establish a new kind of temperature scale tied directly to atomic constants from which the degree kelvin could be realized and disseminated, possibly with improved precision. In what follows we describe the critical features of the proposed technique and discuss some preliminary attempts at achieving demonstration of a prototype. PROPOSED TECHNIQUE Remote thermometry via thermographic phosphors has been explored in depth over the past 20 or so years, and a large number of applications for this methodology have been described in the literature (4). The existence of charge-transfer states in materials such as Y 2 O 3 :Eu and La 2 O 2 S:Eu provides for a temperature-dependence in the amplitudes and exponential decay lifetimes of laser-induced fluorescence in such materials, with the excitation energy being shared between radiative (photon) and nonradiative (phonon) de-excitation processes. See (5) for a reference to a thorough listing of several candidate materials that exhibit this type of behavior. The physics underlying the temperature-dependence of the ratio between these de-excitation mechanisms is reasonably well understood and useful models exist for the quantum mechanics of the decay rates (4). If satisfactory predictions of temperature vs. decay rate can be made for an appropriate material, and if the subsequent experimentally measured decay rates could be determined using an apparatus that has a cesium beam atomic clock as its time base, then the essential features are in place for realizing a temperature scale in terms of atomic quantities. The general components of an experimental arrangement for accomplishing this type of measurement are shown in Figure 1. A tunable dye laser or some other suitable excitation source is used to optically pump the phosphor sample which is located inside an (ideally) gradient free furnace. The fluorescence signal is detected by a fast, low-noise photodetector and a waveform-processing oscilloscope and/or off-

line computer is used to analyze the signal and extract the exponential decay-time constant. The time base for the whole system would be a cesium beam atomic clock or, possibly, a suitable stable signal derived from one (such as the U.S. NIST WWV radio signal). Given a satisfactory theoretical understanding of the relationship between the temperature of the single-crystal phosphor sample and the associated fluorescence decay time, the resulting measurements will yield the thermodynamic temperature of the sample over the entire response range of the particular phosphor being used. A thermometer in thermal equilibrium with the sample, eg., a platinum resistance thermometer as suggested in Figure 1, can then be calibrated directly against this new type of primary standard. Realization of the degree kelvin over continuous spans several hundred degrees would then be possible, since some phosphors have very slow variations of decay time with temperature (i.e., they do not quench rapidly). PROTOTYPE EXPERIMENTAL ARRANGEMENTS We have carried out a series of experiments aimed at establishing a thermal measurement chain of the type shown in Figure 1. The details of the apparatus are described in several of our early papers on the subject (6,7,8) and in the thesis work of L. J. Dowell (9,10). The original purpose of the studies was to use the system in the inverse sense, and calibrate the response of the phosphor thermometer in terms of a NIST-traceable reference thermometery. This was done, and most of the sources of potential experimental uncertainty in such an experimental arrangement were identified (6). One of the principle findings of that work was that one could obtain very pure fluorescence decays from materials such as Y 2 O 3 :Eu, following the initial photo-optical charging of the phosphor. See Figure 2 for an example of the measured exponential decay of the fluorescence amplitude. With the availability of samples of single-crystal phosphors suitable for experimental evaluation (LaPO 4 :Eu and Y 2 O 3 :Eu), we then carried out sets of studies aimed at demonstrating the feasibility of realizing the degree kelvin with this approach. The initial results with LaPO 4 :Eu are described in (8), and those for Y 2 O 3 :Eu are described in (10). A sample of the temperature-dependent decay times obtained with the LaPO 4 :Eu crystals and phosphor is shown in Figure 3. Also shown there are three of the ITS-90 reference points. Using this particular phosphor, continuous measurements (rather than interpolations) between the freezing points of Zn and Al would be possible. A phosphor with a broader thermal roll-off or lower quenching temperature would be needed to reach lower temperatures. An example is YVO 4 :Dy phosphor which is also shown in Figure 3. A simple model based on a publication by Fonger and Struck (11) appears to match the data very well. The equation is = [(1/ 0 ) + b e -(E/kT) ] -1 where 0 is the the low temperature decay time, b is a ratio of energy transfer rates between energy levels, k is the Boltzmann factor, T is temperature and E is an activation energy. E determines at which temperature quenching begins to occur, b controls the slope of temperature versus decay time. 0 is determined by measurement and the curve is produced by adjusting the parameters E and b to match the data.

0 ( s) E (cm -1 ) b YVO 4 :Dy 154 11,800 5.5*10 15 LaPO 4 :Eu 2760 13,500 5*10 13 Optical fibers are used to convey the excitation and fluorescence signal to/from the phosphor under test. This is done because heat leaks associated with clear-air paths into and out of the calibration furnace can then be avoided, thus reducing the size of the thermal gradient in the furnace interior. We mapped the gradient in one candidate furnace used in these studies (Lindberg model 51442), and found that at 1273 K the maximum measured gradient was < 730 mk mm -1 (12). Details of the analysis algorithms used to extract the exponential decay lifetimes from the raw signals are available elsewhere (13, 14, 15). PROSPECTS FOR IMPLEMENTATION The realization of a new primary standard of thermodynamic temperature would be a very ambitious undertaking, and it would ultimately require the resolution of many important metrological issues. The conversion of the temperature measurement into one of obtaining and analyzing a time interval, however, moves the concept in a direction that is generally aligned with modern progress in metrology, since time and frequency are among the most precisely measured of all variables. For instance, Dowell (10) found that by using a Frequency and Time Systems model FTS 4050 cesium beam clock as the time base for his measurements, he was able to achieve a precision in his decay time-based temperature measurements of 0.6% relative to existing thermal standards at temperatures on the order of 1000 C. He also identified several of the principal components of the error budget in such a measurement. (In fact, the time base was not the limiting factor, but by using the atomic clock for that purpose, the feasibility of making an absolute time-based thermal measurement was demonstrated.) To fully realize the possibilities suggested by this approach, an accurate quantum mechanical model of the temperature dependence of the phosphor s decay time will be needed. Efforts aimed at developing such models have been reported (16, 17), and further work on this topic is underway. To make such a system readily available to industrial as well as standards laboratories, inexpensive alternatives to laserbased excitation of the phosphor should be used. We have demonstrated that blue LED excitation of rare-earth-doped phosphors is possible (18), thus providing one such alternative excitation source. Lastly, phosphors with a temperature dependent fluorescence decay time that covers as wide a temperature span as possible should be developed and used. We are studying both single-doped (19) and double-doped (20) phosphors that show promise of satisfactory performance. Eventually, long term experiments will be required to establish the durability of the fluorescent material for extended operation at high temperature and over many high/low temperature cycles. In work to be described in more detail elsewhere, an effort has been made to determine the ultimate precision capability in the vicinity of 800 C using the Y 2 O 3 :Eu phosphor (21,22). A large thermal mass was placed in a high temperature oven so that when the oven s thermostat was switched off, the temperature decreased rather slowly and approximately linearly in time over the region of interest. The results indicate that a relative drop in temperature of 8.5 mk could be discerned. This matches fairly well investigations by Zhang et al (23) who worked close to ambient temperature with a fiber tipped with a Cr:LiSAF crystal.

In future papers, we hope to describe a fully integrated primary standard based on this approach, and present a comprehensive error budget for it. ACKNOWLEDGMENTS The authors thank former U.Va. and U.T. students L. J. Dowell, W. N. Lutz, and A. R. Bugos for their efforts on these projects over the years. We also thank our colleagues D. Beshears, B. Noel and many staff members at the U.S. NIST and the BIPM for many useful and stimulating conversations. Oak Ridge National Laboratory is operated for the U.S. Department of Energy by Oak Ridge National Laboratory, managed by UT-Battelle, LLC, for the U.S. Department of Energy under contract DE- AC05-00OR22725. REFERENCES 1. U.S. National Institute of Standards and Technology, http://www.cstl.nist.gov/div836/836.05, (August 10, 2001). 2. The Comité International des Poids et Mesures, The International System of Units, 7 th Edition, Bureau International des Poids et Mesures, Sèvres, France, 1998, pp. 83-152. 3. U.K. National Physical Laboratory, http://www.npl.co.uk/npl/reference/mass.html, (2001). 4. S. W. Allison and G. T. Gillies, Remote Thermometry with Thermographic Phosphors: Instrumentation and Applications, Rev. Sci. Instrum., Vol. 68(7), pp. 2615-2650, 1997. 5. L. J. Dowell, G. T. Gillies, S. W. Allison, and M. R. Cates, A Research Bibliography on the Temperature Dependencies of Thermographic Phosphors, J. Lumines., Vol. 36, pp. 375-376, 1987. 6. L. J. Dowell, W. N. Lutz, G. T. Gillies, R. C. Ritter, S. W. Allison, and M. R. Cates, Calibration of a Laser Pumped, Thermographic Phosphor Remote Thermometry System, Proc. Laser Inst. Am., Vol. 57, pp. 135-141, 1987. 7. G. T. Gillies, L. J. Dowell, W. N. Lutz, S. W. Allison, M. R. Cates, B. W. Noel, L. A. Franks, and H. M. Borella, Noncontact Thermometry via laser Pumped, Thermographic Phosphors: Characterization of Systematic Errors and Industrial Applications, Proc. Laser Inst. Am., Vol. 63, pp. 15-21, 1988. 8. L. J. Dowell, G. T. Gillies, A. R. Bugos, and S. W. Allison, High Precision Single Crystal Fluorescence thermometry: A Possible Temperature Transfer Standard, Proc. Laser Inst. Am., Vol. 66, pp. 122-130, 1989.

9. L. J. Dowell, A Comprehensive Error Analysis of a Phosphor Thermography System, M.Sc. Thesis, University of Virginia, May 1987. 10. L. J. Dowell, Investigation and Development of Phosphor Thermometry, Ph.D. Dissertation, University of Virginia, May 1989. 11. W. H. Fonger and C. W. Struck, Eu +3 5 D Resonance Quenching to the Charge-Transfer States in Y 2 O 2 S, La 2 O 2 S, and LaOCl, J. Chem. Phys. Vol. 52(12), pp. 6364-6372, 1970. 12. W. N. Lutz, G. T. Gillies, and S. W. Allison, Computer-Aided Data-Acquisition System for thermal Gradient mapping and calibrations of High-Temperature Cavities, Rev. Sci. Instrum., Vol. 60(4), pp. 673-682, 1989. 13. L. J. Dowell, G. T. Gillies, M. R. Cates, and S. W. Allison, "Precision Limits of Waveform Recovery and Analysis in a Signal Processing Oscilloscope," Rev. Sci. Instrum., Vol. 58(7), pp. 1245-1250, 1987. 14. L. J. Dowell and G. T. Gillies, Precision Limits of Lifetime Estimation Algorithms as Determined by Monte Carlo Simulation: A Comparison of Theory and Experiment, Rev. Sci. Instrum., Vol. 59(8), pp. 1310-1315, 1988. 15. L. J. Dowell and G. T. Gillies, "Errors Caused by Baseline Offset and Noise in the Estimation of Exponential Lifetimes," Rev. Sci. Instrum., Vol. 62, pp. 242-243, 1991. 16. Z. Y. Zhang, K. T. V. Grattan, A. W. Palmer, V. Fernicola, and L. Crovini, Temperature- Dependence of the YAG:Cr 3+ Fluorescence Lifetime over the Range 77 to 900 K, Phys. Rev. B, Vol. 51(5), pp. 2656-2660, 1995. 17. Z. Y. Zhang and K. T. V. Grattan, Temperature-Dependence of YAG:Cr 3+ Fluorescence Lifetime up to 550-K, J. Lumines., Vol. 62(6), pp. 263-269, 1994. 18. S. W. Allison, M. R. Cates, and G. T. Gillies, Excitation of Thermographic Phosphors with a Blue LED, Proceedings of the 45 th International Instrumentation Symposium, Instrument Society of America, Research Triangle Park, North Carolina, 1999, pp. 235-244. 19. S. W. Allison, D. L. Beshears, M. R. Cates, and G. T. Gillies, A Wide-Range Phosphor Thermometry Technique, Proceedings of the 44 th International Instrumentation Symposium, Instrument Society of America, Research Triangle Park, North Carolina, 1998, pp. 604-612. 20. S. W. Allison, L. A. Boatner, and G. T. Gillies, Characterization of High-Temperature Thermographic Phosphors - Spectral Properties of LuPO 4 :Dy(1%),Eu(2%), Appl. Opt., Vol. 34(25), pp. 5624-5627, 1995. 21. S. W. Allison, W. S. Key, M. R. Cates, D. L. Beshears, A. J. Sanders, R. J. Newby, J. W. Campbell, and G. R. Shunk, Thermal Metrology for a Space-Based Gravity Experiment, in the

Proc. SPIE Vol. 3116, p. 97-104, Small Spacecraft, Space Environments, and Instrumentation Technologies, Oct. 1997. 22. M. R. Cates, A. J. Sanders, J. Newby, Ultra High Precision Phosphor Thermometry Near 1100 K, manuscript in preparation. 23. Z. Y. Zhang, K. T. V. Grattan, and A. W. Palmer, Sensitive fiber optic thermometer using Cr:LiSAF fluorescence for bio-medical sensing applications, in the Proceedings of the 8 th Optical Fiber Sensors Conference, Monterey, pp. 93-96, Jan. 1992.

Figure 1. Experimental arrangement for a phosphor thermography measurement system. Figure 2. Fluorescence decay of Y 2 O 3 :Eu (lower curve) vs. a pure exponential (upper curve).

Figure 3. Fluorescence lifetime for LaPO 4 :Eu vs. temperature. For comparison, the ITS-90 reference points for Sn, Zn and Al (freezing points) are shown at A, B and C, respectively.