STATISTIQUE DE PERFORMANCE

Similar documents
STATUS OF THE WIGOS DEMONSTRATION PROJECTS

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

World Meteorological Organization

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

CALIBRATION CERTIFICATE # 503

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

Introduction of RIC Tsukuba (Japan, RAII)

CALIBRATION REPORT FOR THERMOHYDROMETER

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Metrology and Environment

CERTIFICATE OF CALIBRATION

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A

OverviewofRIC Tsukuba (Japan,RAII )

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

STATUS OF INSTRUMENTS TRACEABILITY IN NMHSs

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Test report CVN Project number Page 2 of 2

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

Measurement Uncertainty in Mechanical Testing

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Vocabulary of Metrology

SIM AUV.A-K1.prev Microphone Intercomparison

Analysis of a national low dc resistance inter-laboratories comparison

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Meteorological Service

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Calibration and verification: Two procedures having comparable objectives and results

Interlaboratory Comparison & Proficiency Testing Program

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

The Fundamentals of Moisture Calibration

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

A FIRST REVIEW OF CALIBRATION DEVICES ACCEPTABLE FOR METROLOGY LABORATORY

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

EURAMET Project no Inter-comparison of a 1000 L proving tank

CERTIFICATE OF CALIBRATION

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

SUB-REGIONAL PYRANOMETER INTERCOMPARISON FOR THE SOUTH-EASTERN EUROPE ABSTRACT

High pressure comparison among seven European national laboratories

Document No: TR 12 Issue No: 1

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

CERTIFICATE OF CALIBRATION

Survey on alternatives for dangerous and obsolete instruments WMO - IOM 117

Final Report August 2010

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

Interlaboratory Comparison & Proficiency Testing Program

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Science of Stability 2017

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

8001 E. BLOOMINGTON FREEWAY BLOOMINGTON, MN Service:

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Accuracy Calibration Certificate

Calibration of Paroscientific Model 205 Pressure Sensor for use at Heard Island.

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Method Validation and Accreditation

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12

MASS AND VOLUME COMPARISONS AT MIKES

Test Report EN (2008)

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

R SANAS Page 1 of 7

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

Calibration Traceability Guidelines

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

R & D AKUSTIKA * SIA. Test Report of laboratory Sound Insulation measurements Nr. 569 / 2011 AL 8.4. CUSTOMER: AEROC AS (Estonia)

Uncertainty of Calibration. Results in Force Measurements

Physikalisch- Technische Bundesanstalt

ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY

Appendix B1. Reports of SMU

ANNEXE 8. Technical protocols for the interlaboratory comparisons

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

The AWS based operational urban network in Milano: achievements and open questions.

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Commissioning of the Beta Secondary Standard (BSS2)

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Certificate of Accreditation

Transcription:

INTERCOMPARISON OF STANDARD CALIBRATION, METHODS AND MEASUREMENTS FOR PRESSURE, TEMPERATURE AND HUMIDITY BETWEEN RIC CASABLANCA, MOROCCO AND RIC Ljubljana SLOVENIA 3(10) azizmounir@gmail.com The operators who participated in the calibration relating to this comparison are: - Mr. Mounir AZIZ: RIC Responsible, Casablanca Morocco - Mr. Ales Razpotnic: Pression Field, Slovenian Environment Agency - -Mr. Drago Groselj, RIC Responsible, Slovenian Environment Agency, drago.groselj@gov.si ABSTRACT A CIL (Comparison interlaboratory) of standard calibration instruments and methods for pressure, humidity and temperature was organised in 2015 between the laboratories of metrology of Meteo Morocco and the Slovenian Environment Agency. The two RIC laboratories are equipped and work identically in terms of equipments and methods. Both laboratories use the ISO 17025:2005 standard and have an unbroken chain of traceability of their standards. The results show and prove an agreement between the measurements from the laboratories for both pressure, temperature and humidity calibration.

1. Introduction 3 2. REFERENCE DOCUMENTS 3 3. Instrument used 3 4. Program and pressure calibration procedures: 3 5. Diagram measurement benches mounting pressure: 3 6. RESULALTS: 4 6.1. Uncertainties: 5 6.2. Drift values obtained from testing: 5 7. STATISTIQUE DE PERFORMANCE 6 8. ERRORS DUE TO TRANSPORTATION : 7 9. Conclusion: 7

1. INTRODUCTION This paper presents the results of the comparison inter laboratory for calibration of barometers, between the laboratories of Moroccan Meteorology Department and Slovenian Environment Agency, performed on the occasion of Expert Team on Operational Metrology (ET OM) meeting in Ljubljana, Slovenia, from 1st 4th December 2015, hosted by Regional Instrument Center of the Slovenian Environment Agency. The first calibrations were made in Morocco on November 30, 2015, and return on March 1, 2016, the calibration by Slovenian Environment Agenc, took place on 3 and 4 December 2015. Interlaboratory comparison (ILC) serves as a tool for comparison of measurement results carried out by both calibration laboratories in the field of Pressure, temperature and humidity measurement. ILC represents the best means to demonstrate technical competence of the participant and also serves as a part of technical bases for accreditation, it is the most important element for monitoring of quality of measurement results as required by ISO/IEC 17025:2005 standard for laboratories in part 5.9. 2. REFERENCE DOCUMENTS The certificates produced or used in this comparison are: The calibration certificate No. 5122015 Labo. Slovenia and the certificate No. CEP15170 Labo. Morocco. The labo. Morocco joined his traceability certificate (Cal Cert ID: 121,595 in the accredited Mensor laboratory Cert2066.01 ZL. 3. INSTRUMENT USED The instrument calibrated by both laboratories is: In Calibration pressure: Display PTB 330 barometer (VAISALA), Serial C1040029 4. PROGRAM AND PRESSURE CALIBRATION PROCEDURES: The laboratory of Slovenian Environment Agency calibrated on its own calibration bench with a standard traceable to the international system, the calibration was done in 11 points between 850 hpa and 1100 hpa. Three down climb cycles were applied for 2 Slovénen and Moroccan laboratories, with bearings 2500 Pa. The Morocco Laboratory calibrate the barometer compared to its generator and reference standard "MENSOR CPC 6000" in 11 points from 85000 Pascal to 110000 pascal. 5. DIAGRAM MEASUREMENT BENCHES MOUNTING PRESSURE: Installation diagrams of two laboratories are fairly similar, but with two different brands.

calibration bench assembly diagram of Moroccan laboratory, with integrated standard automatic generator CPC6000 POMPE PC Baromètre réf : pression PPC1 Générateur A Z O T E Liaison série Circuit pneumatique Alimentation secteur Distributeur de pression calibration bench assembly diagram of Slovenian laboratory Baromètre à étalonner 6. RESULALTS:

The results of calibrations used for intercomparison, are averages of three cycles of correction term measures, accompanied by the uncertainty calculated for each calibration laboratory and according to the means used. The ambient conditions in laboratory of Slovenia is 22 C to 25.5 C with a relative humidity between 25% and 32%; and in the laboratory of Morocco is 23 C to 23.6 C. 6.1. Uncertainties: Both laboratories express calibration uncertainty with a coverage factor k = 2 And take into account four important components: Uncertainty calibration means Uncertainty due to the repeatability of the sensor at 3 cycles, The maximum hysteresis observed for all of the bearings. The uncertainty due to the resolution of the instrument under test Laboratory of Slovenian Environment Agency declared uncertainties are set at the maximum pressure used = 1100 hpa. This method simplifie the use of uncertainty whereas the range of maximum pressure measurements (850 to 1100 hpa) does not justify the method of polynomial calculation and the discrepancies between the low pressure and considered high pressure, are small compared to the overall uncertainty. Details uncertainties determined by the two laboratories: Laboratory Slovenia ( Pa) Laboratory Morocco ( Pa) u -moyens / 2 u -moyens / 2 U= 1,5 Pa + 3,3 *10-5 * P u- répétabilité u- répétabilité 3 u_ hyster ptb u_ hyster ptb 0.005/ 3=0,289 u_ resolution ptb u_ resolution ptb 0.005/ 3=0,289 U (k=2) = 8 Pa U(k=2) = 6 Pa + 2*10-3 P soit au max. = 8,2 Pa Slovénie : 2* Root (u -moyens 2 +u -répétabilité2 +u _hyster_ptb 2 +u _resolution_ptb2 ) U = 8 Pa Maroc : 2*Root (u -moyens 2 +u -répétabilité 2 +u _hyster_ptb 2 +u _resolution_ptb 2 ) U = (6 Pa +2*10-3 P) = 8,2 Pa The error means the maximum difference observed between the standard and the test sensor. Tolerance or compliance of customer, declared by the Moroccan Laboratory is 0,5 mbar or 50 Pa The decision criterion is "the greater error (E Accuracy, Fidelity E, E hysteresis) increased the uncertainty (U) must be less than the specified tolerance. Ei + U <(T / 2), T = customer tolerance Similarly, the adjustment criterion in test barometers established by the formula Error + U <( C / 2), C = Customer Compliance = T 6.2. Drift values obtained from testing:

All the points of steps executed by Laboratory Slovenia and Laboratory Morocco is reported in this table, where measurements are made using the same bearings are analyzed. SETPOINTS (Pa) UNCERTAINTY U (Pa) Morocco Slovenia U Maroc U Slovénie Réf = Pe Pe-Pi Réf = Pe Pe-Pi 110000 110000-34 110000-33 7,7 8 107500 107500-34 107500-33 7,7 8 105000 105000-33,5 105000-33 7,8 8 102500 102500-31 102500-32 7,8 8 100000 100000-31,5 100000-32 7,9 8 97500 97500-30,5 97500-32 7,9 8 95000 95000-32 95000-32 8,0 8 92500 92500-33,5 92500-32 8,0 8 90000 90000-35 90000-32 8,1 8 87500 87500-34,5 87500-32 8,1 8 85000 85000-36 85000-33 8,2 8 Comparative curves in Pascal Note that the sensor used, presents a gap between 32 and 33Pa for the values obtained by Laboratory Slovenia and 30.5 and 36 Pa to the values obtained by Laboratory Morocco. The uncertainty bars are representative of U declared 8 Pa Laboratory Slovenia and between 7.7 and 8.2 for Laboratory Morocco, these error bars show the overlay measurements of this sensor by 2 calibration methods. To this end, it seems necessary to calculate the standard deviation of all measurements, to check whether the inter comparison pressure is considered satisfactory. 7. STATISTIQUE DE PERFORMANCE Performance statistic used in the context of this intercomparison is the standard deviation "EN"; described according to DIN EN ISO / IEC 17043: 2010. It is obtained as follows: En M S / (U M U S )

X M and X S are the results, respectively, error indications or corrections of participating laboratory from morocco and slovenia. U Labi et U Labp are respectively the expanded uncertainties of participating laboratory from morocco and slovenia. ERRORS MOROCCO SLOVENIA Ecarts = M U Ecarts = S U EN Maroc Slovenie 34 7,7 33 8 0,09 34 7,7 33 8 0,09 33,5 7,8 33 8 0,04 31 7,8 32 8 0,09 31,5 7,9 32 8 0,04 30,5 7,9 32 8 0,13 32 8 32 8 0,00 33,5 8 32 8 0,13 35 8,1 32 8 0,26 34,5 8,1 32 8 0,22 36 8,2 33 8 0,26 8. ERRORS DUE TO TRANSPORTATION : SETPOINTS DERIVE /REF DERIVE/REF BEFORE AFTER DRIFT UP DOWN UP DOWN UP DOWN 110000 2 3 6 4 5 3 107500 2 1 4 3 2 2 105000 0 1 3 3 3 2 102500 1-1 2 2 1 3 100000 1 0 4 4 3 4 97500 0-1 3 3 3 4 95000 0 0 2 3 2 3 92500 1 2 2 3 1 1 90000 2 1 4 6 2 5 87500 2 2 5 4 3 2 85000 2 1 6 4 4 3 For the travel of the sensor, calibration report shows excellent retention of the equipment after transportation, the maximum deviation is 5 Pascal between calibrations values before and after transport. 9. CONCLUSION:

RESULTATS DE L INTER COMPARAISON PRESSION EN-S-11-09-20 Version: A Date: 14/08/2013 In this case comparing two laboratories and in particular for the pressure calibration method, it is found that the values obtained by the two laboratories have a superposition inducing a good agreement. The standard deviation is well below the limit value of 1. The comparison is satisfactory. Both laboratories have unbroken traceability chains for humidity and pressure and their calibration results are in good agreement. Full results will be available at TECO2016.