Electron Microprobe Analysis and Scanning Electron Microscopy

Similar documents
Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Gaetano L Episcopo. Scanning Electron Microscopy Focus Ion Beam and. Pulsed Plasma Deposition

Modern Optical Spectroscopy

h p λ = mν Back to de Broglie and the electron as a wave you will learn more about this Equation in CHEM* 2060

Generation of X-Rays in the SEM specimen

X Rays & Crystals. Characterizing Mineral Chemistry & Structure. J.D. Price

EDS User School. Principles of Electron Beam Microanalysis

CHARACTERIZATION of NANOMATERIALS KHP

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Contrasted strengths and weakness of EDS, WDS and AES for determining the composition of samples

Chapter Six: X-Rays. 6.1 Discovery of X-rays

Why microscopy?

Transmission Electron Microscopy

Chapter 9. Electron mean free path Microscopy principles of SEM, TEM, LEEM

ABC s of Electrochemistry series Materials Characterization techniques: SEM and EDS Ana María Valenzuela-Muñiz November 3, 2011

Practical course in scanning electron microscopy

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Praktikum zur. Materialanalytik

Fundamentals of Nanoscale Film Analysis

object objective lens eyepiece lens

AP5301/ Name the major parts of an optical microscope and state their functions.

EE 527 MICROFABRICATION. Lecture 5 Tai-Chang Chen University of Washington

Chemical Analysis in TEM: XEDS, EELS and EFTEM. HRTEM PhD course Lecture 5

Weak-Beam Dark-Field Technique

Basic structure of SEM

EPMA IMAGES. Figure 9. Energy-dispersive spectra of spot mineral analyses in sample 89GGR-33A for locations 1-5 in Figure 8.

6. Analytical Electron Microscopy

X-RAY SPECTRA. Theory:

Scanning Electron Microscopy & Ancillary Techniques

Understanding X-rays: The electromagnetic spectrum

KMÜ 396 MATERIALS SCIENCE AND TECH. I PRESENTATION ELECTRON ENERGY LOSS SPECTROSCOPY (EELS) TUĞÇE SEZGİN

Massachusetts Institute of Technology. Dr. Nilanjan Chatterjee

X-rays. X-ray Radiography - absorption is a function of Z and density. X-ray crystallography. X-ray spectrometry

An Introduction to Diffraction and Scattering. School of Chemistry The University of Sydney

Characterisation of Catalysts Using Secondary and Backscattered Electron In-lens Detectors

Understanding X-rays: The electromagnetic spectrum

Bragg reflection :determining the lattice constants of monocrystals

April 10th-12th, 2017

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

CBE Science of Engineering Materials. Scanning Electron Microscopy (SEM)

Interactions with Matter

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Lecture 23 X-Ray & UV Techniques

EEE4106Z Radiation Interactions & Detection

SOLID STATE PHYSICS PHY F341. Dr. Manjuladevi.V Associate Professor Department of Physics BITS Pilani

X-ray practical: Crystallography

HOW TO APPROACH SCANNING ELECTRON MICROSCOPY AND ENERGY DISPERSIVE SPECTROSCOPY ANALYSIS. SCSAM Short Course Amir Avishai

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Electron Microscopy I

Overview of scattering, diffraction & imaging in the TEM

MS482 Materials Characterization ( 재료분석 ) Lecture Note 5: RBS

Analytical Methods for Materials

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

4. Inelastic Scattering

SCANNING ELECTRON MICROSCOPE

Writing: Text Types and purposes 1. Write arguments to support claims in an analysis of substantive topics or texts, using valid reasoning and

Unit title: Atomic and Nuclear Physics for Spectroscopic Applications

Emphasis on what happens to emitted particle (if no nuclear reaction and MEDIUM (i.e., atomic effects)

X-ray Energy Spectroscopy (XES).

X-Ray Emission and Absorption

Part II: Thin Film Characterization

X-Ray Photoelectron Spectroscopy (XPS)

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

Preamble: Emphasis: Material = Device? MTSE 719 PHYSICAL PRINCIPLES OF CHARACTERIZATION OF SOLIDS

Atomic Physics. Chapter 6 X ray. Jinniu Hu 24/12/ /20/13

Chapter 31 Atomic Physics

Everhart-Thornley detector

Interaction X-rays - Matter

MANIPAL INSTITUTE OF TECHNOLOGY

Electron Probe Microanalysis (EPMA)

MEMS Metrology. Prof. Tianhong Cui ME 8254

Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture

Scanning Electron Microscopy

SEM stands for Scanning Electron Microscopy. The earliest known work describing

MSE 321 Structural Characterization

Invited Lecture. "Different Aspects of Electron Microscopy. Sardar Vallabhbhai National Institute of Technology, Surat. Deepak Rajput & S.K.

CHARGED PARTICLE INTERACTIONS

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

Exercise 1 Atomic line spectra 1/9

ELECTRON CURRENT IMAGE DIFFRACTION FROM CRYSTAL SURFACES AT LOW ENERGIES

Methods of surface analysis

The Basic of Transmission Electron Microscope. Text book: Transmission electron microscopy by David B Williams & C. Barry Carter.

Surface Sensitivity & Surface Specificity

Overview of X-Ray Fluorescence Analysis

Applications for PIXE and other Ion Beam Analysis (IBA)

Interaction of charged particles and photons with matter

III. Energy Deposition in the Detector and Spectrum Formation

Electron-Matter Interactions

Chapter 28. Atomic Physics

Properties of Electrons, their Interactions with Matter and Applications in Electron Microscopy

INTERACTIONS OF RADIATION WITH MATTER

Any first year text, sections on atomic structure, spectral lines and spectrometers

Scanning electron microscopy

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

X-RAY PRODUCTION. Prepared by:- EN KAMARUL AMIN BIN ABDULLAH

Lab Manual: Determination of Planck s constant with x-rays

Technical University of Denmark. Center for Electron Nanoscopy. Advanced TEM (16 September 2010) Microanalysis in the electron microscope

CHEM*3440. X-Ray Energies. Bremsstrahlung Radiation. X-ray Line Spectra. Chemical Instrumentation. X-Ray Spectroscopy. Topic 13

Transcription:

Electron Microprobe Analysis and Scanning Electron Microscopy Electron microprobe analysis (EMPA) Analytical technique in which a beam of electrons is focused on a sample surface, producing X-rays from the material in the sample. Used for chemical mapping of a sample surface, and also for quantitative chemical analysis.

What is EMPA used for? -Quantitative or qualitative chemical analysis of a very small focussed spot (<1um) on a polished sample surface. A nondestructive technique. Useful for: -Mineral identification -Descriptive petrology -Geothermometry and geobarometry -Experimental petrology -Cosmochemistry -Zoning in minerals -Diffusion studies -Particle analysis

What is EMPA used for? (con t) -Chemical mapping of a polished sample surface. Accomplished by sweeping (rastering) the beam over the sample surface. Used for many of the same types of studies described above. Particularly useful for: -Location and distribution of phases -Quick location and identification of trace particles -Subtle zoning patterns in mineral samples -Examination of the interfaces between phases

Secondary electron imaging (SEI, also commonly SEM) Secondary electron imaging (SEI, also commonly SEM) A related technique in which a very finely-focussed beam of electrons is swept across a sample surface, producing secondary electrons which are then detected by a scintillation detector. This technique is used for highmagnification and resolution imaging of a 3-D sample surface.

What is secondary electron imaging used for? -Morphological imaging of a rough sample surface. Also accomplished by rastering the beam over the sample surface, but does only yields morphological, not chemical, information. Useful for: -Determining the morphology and morphological relationship between mineral phases that can be difficult to detect with the petrographic microscope -Determining size of very small mineral phases

The instrumentation which allows these two types of analyses is closely related, in that the electron optics system is very similar. The CAMECA SX-100 instrument is primarily an electron microprobe designed for quantitative chemical analysis, but also can also be used for secondary electron imaging.

History of Electron Probe Microanalysis X-RAYS 1895- Roentgen discovers X-rays X-rays are high-frequency electromagnetic radiation with energy intermediate between far- UV and gamma ray regions. This leads to development in numerous fields, particularly medical. 1911- Barkla recognized characteristic radiation This scientist recognizes that x-rays emitted from different elements produce x-rays of different energy and wavelength.

1913-1914- Moseley quantifies the relationship between atomic number and x-ray wavelength. Moseley photographed and studied characteristic x-ray radiation. He determined that the atomic number of an atom is more important than the atomic weight in determining the characteristics of x-rays generated from that element. Formulated Moseley's law: λα 1/Z 2 where λ = x-ray wavelength Z = atomic number 1912- Von Laua. Diffraction Von Laua and coworkers speculated that wavelength of x- rays and crystal lattice parameters are similar, so x-rays could be diffracted by crystals.

1914- Bragg. Developed Bragg's Law Bragg's law defines a diffraction relationship between the wavelength of an incoming ray and the d-spacing of a diffracting crystal. Bragg's Law: n λ = 2d sinθ where n = the order of reflection λ = wavelength of incident ray d = interplanar spacing of the crystal θ = angle of incidence and reflection of incident ray This is the principle under which the wavelength-dispersive detectors operate.

1951 R. Castaing. Developed the first electron microprobe Castaing was a doctoral student at the University of Paris. His doctoral thesis was titled: "Application of electron beams to a method of local chemical and crystallographic analysis". He is the originator of the electron microprobe as we know it today.

Fundamental Beam Interactions

-Characteristic X-rays Produced by electronic transitions within inner electron shells. Can be explained by examining the Rutherford-Bohr model of the atom, in which electrons orbit in a number of shells around a nucleus. -Backscattered electrons Primary electrons from the beam that undergo violent collisions with atoms in the sample and are scattered back from the sample surface. The proportion of backscattered electrons is directly proportional to the mean atomic number of the sample. Backscatter from C is 8%, from U is 50%. Modern BSE detectors can detect mean Z differences of between 0.1 and 0.3 Z. -Secondary electrons Secondary electrons are produced by ionization of atoms in the sample by interaction with the primary beam. Secondary electrons have low energies, and cannot escape from deep within the sample. Therefore, this electron signal produces the highest-resolution morphological images of a sample surface.

-Cathodoluminescence Visible light emitted by minerals bombarded by electrons. Mechanism depends on the mineral in question. This property of certain minerals, particularly diamond and some types of feldspar, allow the beam to be viewed optically which is useful for focusing. -Continuous X-rays Most interactions between the primary beam do not produce the characteristic x-rays discussed earlier. Many electrons from the incident beam are simply decelerated by the interactions with outer orbital shell electrons from atoms within the sample. The energy released during these decelerations produces a continuous (or bremsstrahlung) x-ray spectrum. These x-rays are noise, and are undesirable, but unavoidable.

-Auger Electrons When an atom is ionized in an inner shell, energy may be released by ejecting another bound electron, instead of producing an x-ray. These electrons can have characteristic energy. Most effective for low atomic number elements (<10).

-Characteristic X-rays Produced by electronic transitions within inner electron shells. Can be explained by examining the Rutherford-Bohr model of the atom, in which electrons orbit in a number of shells around a nucleus. From Reed, 1996

From Reed, 1996

From Reed, 1996

From Reed, 1996

-Backscattered electrons Primary electrons from the beam that undergo violent collisions with atoms in the sample and are scattered back from the sample surface. The proportion of backscattered electrons is directly proportional to the mean atomic number of the sample. Backscatter from C is 8%, from U is 50%. Modern BSE detectors can detect mean Z differences of between 0.1 and 0.3 Z.

-Secondary electrons Secondary electrons are produced by ionization of atoms in the sample by interaction with the primary beam. Secondary electrons have low energies, and cannot escape from deep within the sample. Therefore, this electron signal produces the highest-resolution morphological images of a sample surface.

Electron interaction volumes Effect of beam interaction (damage) in plastic (polymethylmethacrylate), from Everhart et al., 1972. All specimens received same beam dosage, but were etched for progressively longer times, showing in (a) strongest electron energies, to (g) the region of least energetic electrons. Note teardrop shape in (g) From J. Fournelle http://www.geology.wisc.edu/~johnf/g777/

Major components of an electron microprobe 1. Electron optic system. -Electron gun -focussing lenses and apertures -associated electronics 2. Optical microscope and sample stage -optical/video apparatus for viewing sample -high-precision stage that holds samples and standards 3. X-ray optic system -spectrometers -secondary electron detector -backscatter electron detector -flow-proportional gas counters

Major components of an electron microprobe (cont ) 4. Vacuum system -roughing pumps -diffusion pump -ion pump 5. Amplification/electronics -amplification of signals 6. Computer Controller -interface between instrument and operator -image collection and processing systems -quantitative data reduction programs