QUALIFICATION OF FLOW CONDTIONING DEVICES ACCORDING TO THE NEW API 14.3 PART 2 PROCEDURE. Klaus J. Zanker and Dale Goodson. Daniel Industries, Inc.

Similar documents
NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

The Expansibility Factor Equations in ISO and ISO : Do They Deliver What They Promise?

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Discharge coefficient behaviour in presence of four perforated plates flow conditioners - Numerical investigation Part 1-

Discharge Coefficient Behaviour in Presence of Four Perforated Plates Flow Conditioners

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Laboratory work No 2: Calibration of Orifice Flow Meter

FINAL DRAFT: October 25, 2010 API TECHNICAL REPORT FUEL GAS MEASUREMENT. FINAL DRAFT: October 25, 2010 Page 0

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

Lecture 13 Flow Measurement in Pipes. I. Introduction

ME332 FLUID MECHANICS LABORATORY (PART I)

Smart DP Flow Metering A summary of Established Technology and Latest Innovation. Gary Fish

Major and Minor Losses

Conditioning of swirling and stratified pipe-flow: Analysis with laser-doppler velocimetry

Experiment (4): Flow measurement

Water Circuit Lab. The pressure drop along a straight pipe segment can be calculated using the following set of equations:

V-Cone Flowmeter DATASHEET

Flow Measurement in Pipes and Ducts COURSE CONTENT


Universal type Port size AS13 1F-M3. M3 x 0.5 M5 x 0.8 AS23 1F-01 AS23 1F-02 AS33 1F-02 AS33 1F-03 AS43 1F-04 R 1/8 R 1/4 R 1/4 R 3/8 R 1/2

Flowmeter Discharge Coefficient Estimation

ALASTAIR MCLACHLAN Applications Cameron

Large Rocket Engine Environments

ME 4600:483 Lab Notes Revised 11/16/2015. Flow Measurement

A REVIEW OF THE V-CONE METER FOR NATURAL GAS FLOWS

SELECCIÓN DE CAUDALIMETROS. Tablas de Fabricantes

Installation effects on an ultrasonic flow meter with implications for self diagnostics

An Essential Requirement in CV Based Industrial Appliances.

HEAT TRANSFER ENHANCEMENT IN HEAT EXCHANGER USING TANGENTIAL INJECTOR TYPE SWIRL GENERATOR

Sudden Expansion Exercise

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

507F. Multi-Port Sanitary Ball Valve. Full Port ISO 5211 Mounting Pad. For Automated & Manual Service. Pharmaceutical. Bio-Technology.

EVALUATION OF A CLOSE COUPLED SLOTTED ORIFICE, ELECTRICAL IMPEDANCE, AND SWIRL FLOW METERS FOR MULTIPHASE FLOW. A Thesis MUHAMMET CEVIK

) where the inlet pressure ( P 1

Measurement of fluid flow in closed conduits

Experiment No.4: Flow through Venturi meter. Background and Theory

ABB Flowmeters Review of industrial processing flowmeters

Flow rate and mass flow rate

PROCEEDINGS American School of Gas Measurement Technology 2003 PROCEEDINGS. PAGE xxv AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

Prediction of Performance Characteristics of Orifice Plate Assembly for Non-Standard Conditions Using CFD

Flow Measurement. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/22 Siemens FI

VENTURIMETER EXPERIMENT

Rapid Averaging Probe (RAP)

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz

Measurement of fluid flow in closed conduits

Calculation of Pipe Friction Loss

ISO 5136 INTERNATIONAL STANDARD. Acoustics Determination of sound power radiated into a duct by fans and other air-moving devices In-duct method

TransCanada / Daniel Using Advances Diagnostics to Identify Measurement Uncertainties

Roll No. :... Invigilator s Signature :.. CS/B.Tech (EIE)/SEM-5/EI-501/ INDUSTRIAL INSTRUMENTATION II

ISO 9906 INTERNATIONAL STANDARD. Rotodynamic pumps Hydraulic performance acceptance tests Grades 1 and 2

White Paper FINAL REPORT AN EVALUATION OF THE HYDRODYNAMICS MECHANISMS WHICH DRIVE THE PERFORMANCE OF THE WESTFALL STATIC MIXER.

ME 309 Fluid Mechanics Fall 2010 Exam 2 1A. 1B.

NUMERICAL EVALUATION OF DISTORTED VELOCITY PROFILE EFFECTS ON INSERTION FLOW METERS

COURSE CODE : 3072 COURSE CATEGORY : B PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE

GRANULAR PRESSURE MEASUREMENT IN FLUIDIZED BEDS

Discharge Coefficient Prediction for Multi hole Orifice Plate in a Turbulent Flow through Pipe: Experimental and Numerical Investigation

Flow Measurement in Pipes and Ducts COURSE CONTENT

Bounding the Uncertainties of the Meter Factors for LEFM CheckPlus Flow Elements in Feedwater Flow Measurements

MODEL: WB009GMFI19HLD (120V) Wall Mount DC Inverter Fan Coil Unit 9,000 BTUH

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER

EXPERIMENTAL INVESTIGATION OF THE INSTALLATION EFFECTS ON THE VENTURI FLOWMETER PERFORMANCE

Air Flow through Woven Stainless Steel Mesh

Lab Section Date. ME4751 Air Flow Rate Measurement

VELOCITY PROFILE EFFECTS IN CORIOLIS MASS FLOWMETERS: RECENT FINDINGS AND OPEN QUESTIONS

LECTURE 6- ENERGY LOSSES IN HYDRAULIC SYSTEMS SELF EVALUATION QUESTIONS AND ANSWERS

Fundamentals of Flow Metering

FLOW MEASUREMENT IN PIPES EXPERIMENT

NEW VSL Conditioning plate!

Addendum to ARI Standard , Air Terminals September 2002

Direct-operated 2/2-way compact solenoid valves Type EV210A

P303 Pressure Regulator

Calibration of Orifice Flow Meter and Venturi Flow Meter

Measurements using Bernoulli s equation

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

Compressible Gas Flow

Numerical and Experimental Study on the Effect of Guide Vane Insertion on the Flow Characteristics in a 90º Rectangular Elbow

Decay of Rotational Airflow with Flow Conditioner in Larger Diameter Ducts for Dust Concentration Measurement using Isokinetic Sampling

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere

9.1 Introduction. 9.2 Fluid Flow. CHAPTER 9 Flow

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

ABSTRACT I. INTRODUCTION

GAS MASS FLOW METERING AN ALTERNATIVE APPROACH Eric Sanford, Vortek Instruments Koichi Igarashi, Azbil North America Kim Lewis, DP Diagnostics

Fluid Flow Analysis Penn State Chemical Engineering

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

VD1 CLC. Small Valves DIN-range with a higher k v -value

INTERNATIONAL OIML D 25 DOCUMENT. Vortex meters used in measuring systems for fluids ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Orifice and Venturi Pipe Flow Meters

MEDIDAS EN FLUIDOS. INEL 5205 Instrumentación. Monday, November 19, 12

A MULTIPATH ULTRASONIC METER WITH REDUCING NOZZLE FOR IMPROVED PERFORMANCE IN THE LAMINAR/TURBULENT TRANSITION REGION

Differential Pressure Producing Flow Elements

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

Physics 3 Summer 1990 Lab 7 - Hydrodynamics

Flow Measurement in Pipes and Ducts

CALIBRATION CERTIFICATE # 503

Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

DESIGN AND CONSTRUCTION OF A WATER TUNNEL. Stephen C. Ko

"Experimental investigation of heat transfer enhancement by u-shaped Turbulator"

Transcription:

QUALIFICATION OF FLOW CONDTIONING DEVICES ACCORDING TO THE NEW API 14.3 PART 2 PROCEDURE. Klaus J. Zanker and Dale Goodson Daniel Industries, Inc.

Old Tube Bundle A D a 1. a/d < 1/4 2. L/a > 1 3. A < 1/16(π D 2 /4) 4. Tubes any size - symmetrical 5. Tubes any shape 6.Tubes mounted into end rings 7. Tubes welded together 8. Tubes thin walled 9. All tubes mounted axially

New Tube Bundle Tube bundle OD Items 1, 2, 3, 4, 5, & 6 no longer apply, only 19 tube uniform concentric tube bundle allowed All tubes must touch one another Tubes must be welded together OD >.95 D t < 2.5 % of D Tube wall thickness = t

API Type Approval Requirements Baseline Calibration, evaluating performance of test facility To prove that the mechanical baseline configuration is valid, the baseline CD values should lie within the 95% confidence interval for the RG equation. Standard does not demand an accredited facility

Disturbance Tests Good Flow Conditions - test evaluating impact of flow conditioner on fully developed velocity profile. Two 9 Elbows in Perpendicular Planes - testing of flow conditioner performance in handling combination of a modest swirl and a non-symmetrical velocity profile. Gate Valve 5% Closed - test evaluating flow conditioner performance in strongly non-symmetrical velocity profile. High Swirl - test assessing flow conditioner performance in flows with high swirl angle (over 25 ).

Additional Tests Orifice b ratio: Check swirl test at β=.4 and β=7 Reynolds Number Sensitivity: Test at two ranges 1 4 < Re <5 x 1 5 and Re > 1 6 with approximately a 7:1 ratio. Scaling: Test at two sizes selected from D < 4 and D > 8

Zanker Plate Design Zanker straightener 1959 at BHRA, NEL conference in 196, was included in ISO 5167 in 198 Mitsubishi in 1973 US patent 38451 Elizabeth Laws in 1992 Modified Zanker: Thick plate, no honeycomb

Flow Straightener

Zanker Plate

.67b : Good Flow Conditions.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 2 4 6 8 1 12 14 16 18 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock

67b : Two 9 Elbows in Perpendicular Planes 17 D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 2 4 6 8 1 12 14 16 18 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : Two 9 Elbows in Perpendicular Planes 29D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : Two 9 Elbows in Perpendicular Planes 45D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 4 45 5 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : 5% Closed Gate Valve 17D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 2 4 6 8 1 12 14 16 18 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : 5% Closed Gate Valve 29D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : 5% Closed Gate Valve 45D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 4 45 5 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

Swirl Producing Vane

Swirl produced 35 3 Swirl angle (degrees) 25 2 15 1 5.5 1 1.5 2 Distance from pipe wall (in)

.67b : 17 D Meter Tube : 32 swirl.8 % Deviation of Actual C D from Baseline CD.4.2 -.2 -.4 - -.8 2 4 6 8 1 12 14 16 18 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : 32º Swirl : 29D Meter Tube % Deviation of Actual CD from Baseline CD.8.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67b : 32º Swirl : 45D Meter Tube % Deviation of Actual CD from Baseline CD.8.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 4 45 Zanker Plate Distance in Pipe Diameters from the Orifice Plate 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.4 b : 17 D Meter Tube : 32 swirl.8 % Deviation of Actual CD from Baseline C D.4.2 -.2 -.4 - -.8 2 4 6 8 1 12 14 16 18 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.4 b : 29 D Meter Tube : 32 swirl % Deviation of Actual CD from Baseline CD.8.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.4 b : 45 D Meter Tube : 32 swirl.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 12 O'Clock 9 O'Clock Error Band

.67 b : 4" OFU: High Reynolds Number: Good Flow Condition.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 9 O'Clock Error Band

.67 b : 8" OFU : Good Flow Condition.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 3 O'Clock 9 O'Clock Error Band

4' OFU :.67b :Ceesi & Daniel Baselines 14 12 1 CD 8 6 4 2.E+ 5.E+5 1.E+6 1.5E+6 2.E+6 2.5E+6 3.E+6 3.5E+6 Pipe Reynolds Number CEESI Daniel RG Equa. +Uncert. -Uncert

Good Flow Conditions : 4" Hi R D & Low R D : 8 inch Deviation of Actula C D from Baseline C D.8.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate from the Orifice Plate in Pipe Diameters Lo Rd Hi Rd +.23% -.23% 8 inch

.67 b : 8" OFU : 5% Closed Gate Valve : 17D Meter Tube.8 % Deviation of Actual C D from Baseline C D.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 12 O'Clock 6 O'Clock Error Band

.67 b : 8" OFU : 5% Closed Gate Valve : 29D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 12 O'Clock 6 O'Clock Error Band

.67 b : 8" OFU : 5% Closed Gate Valve : 45D Meter Tube.8 % Deviation of Actual CD from Baseline CD.4.2 -.2 -.4 - -.8 5 1 15 2 25 3 35 Distance of Zanker Plate in Pipe Diameters from the Orifice 12 O'Clock 6 O'Clock Error Band

Conclusions The three pairs of taps are not always all in tolerance The high swirl test requires better definition The Zanker Flow Conditioner has passed all the type approval tests This makes a reasonably priced nonproprietary flow conditioner available to the industry

Conclusions Recommendation: - use 17D meter tube with conditioner 7.5D from the orifice plate, as this arrangement passed all tests The conditioner can be used as follows: Beta up to 7 with any disturbance in any pipe size with no upper limit on Reynolds Number