Proton radiography with a range telescope and its use in proton therapy

Similar documents
Towards Proton Computed Tomography

Initial Studies in Proton Computed Tomography

Reconstruction for Proton Computed Tomography: A Monte Carlo Study

a) Dipartimento di Fisiopatologia Clinica, Università degli Studi di Firenze, v.le Morgagni 85, I-

Loma Linda University Medical Center, Loma Linda, CA 92354

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

The Most Likely Path of an Energetic Charged Particle Through a Uniform Medium

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Calibration of the Modular Neutron Array (MoNA)

R. P. Redwine. Bates Linear Accelerator Center Laboratory for Nuclear Science Department of Physics Massachusetts Institute of Technology

Energy Dependence of Biological Systems Under Radiation Exposure

Physics of particles. H. Paganetti PhD Massachusetts General Hospital & Harvard Medical School

Development of beam delivery systems for proton (ion) therapy

Fast and compact proton radiography imaging system for proton radiotherapy

GEANT4 simulation of the testbeam set-up for the ALFA detector

Monte Carlo Simulation concerning Particle Therapy

Measuring Cosmic Ray Muon Decay Constant and Flux

Beam Optics for a Scanned Proton Beam at Loma Linda University Medical Center

Radiation protection issues in proton therapy

8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania

First Results and Realization Status of a Proton Computed Radiography Device

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

THE mono-energetic hadron beam such as heavy-ions or

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.

Tracking properties of the ATLAS Transition Radiation Tracker (TRT)

Estimating proton beam energy spread using Bragg peak measurement

PROTON radiation therapy is one of the most precise forms

IAC-08-A MONTE CARLO SIMULATIONS OF ENERGY LOSSES BY SPACE PROTONS IN THE CRATER DETECTOR

Neutron pulse height analysis (R405n)

CLASSIFICATION AND IMAGING OF SPENT NUCLEAR FUEL DRY CASKS USING COSMIC RAY MUONS

Queen s University PHYS 352

EDUCATION University of Surrey, Guildford, Surrey, UK PhD in Physics/ Medical Physics 2009

Ideas for an Interferometric Thermometer

Hadron Therapy Medical Applications

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY

Outline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

Prototype Tracking Studies for Proton CT

The EDM Polarimeter Development at COSY-Jülich

STUDY ON IONIZATION EFFECTS PRODUCED BY NEUTRON INTERACTION PRODUCTS IN BNCT FIELD *

High Energy Electron Radiation Exposure Facility at PSI

PoS(EPS-HEP2015)232. Performance of a 1 m 2 Micromegas Detector Using Argon and Neon based Drift Gases

arxiv: v1 [physics.ins-det] 1 Sep 2009

MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY *

Characterizations and Diagnostics of Compton Light Source

Applications of MCBEND

Examples for experiments that can be done at the T9 beam line

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko

Using Prompt gamma ray emission to address uncertainties in proton therapy

Appendix A2. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France.

Detectors in Nuclear Physics: Monte Carlo Methods. Dr. Andrea Mairani. Lectures I-II

CRaTER Science Requirements

Medical Neutron Science

The KASCADE-Grande Experiment

Information about the T9 beam line and experimental facilities

Shielded Scintillator for Neutron Characterization

arxiv: v1 [physics.ins-det] 26 Sep 2018

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

STATUS OF ATLAS TILE CALORIMETER AND STUDY OF MUON INTERACTIONS. 1 Brief Description of the ATLAS Tile Calorimeter

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

Neutron Transport Calculations Using Monte-Carlo Methods. Sean Lourette Fairport High School Advisor: Christian Stoeckl

arxiv:nucl-ex/ v1 21 Dec 2004

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

A measurement of the air fluorescence yield

Tracking Study of 200 MeV Protons within a PMMA. Phantom

The NArCos project (Neutron ARray for Correlation Studies)

CRaTER Pre-Ship Review (PSR) Instrument Calibration Science Requirements Compliance

Progress in measuring GMR in unstable nuclei: Decay detector calibration and inverse reaction experiment. J. Button, Y.-W. Lui, and D.H.

Correlations in Prompt Neutrons and Gamma Rays from Fission

Basic Radiation Physics of Protons

Secondary Particles Produced by Hadron Therapy

Recent results at the -meson region from the CMD-3 detector at the VEPP-2000 collider

OPPORTUNITY TO JOIN IEEE AND NPSS

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT

The relevance of XENON10 constraints in this low-mass region has been questioned [15] C.E. Aalseth et al. arxiv: v1

Measurement of the transverse diffusion coefficient of charge in liquid xenon

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS AND WENDI-2 MEASUREMENTS

Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy

M d e i di l ca A pplilli t ca i ttions o f P arti ttic ti l P e h Physics Saverio Braccini INSEL

Polarimetry Options for E160

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

AEPHY: Nuclear Physics Practise Test

Fitting the Bragg peak for accurate proton range determination

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY

Investigation of nuclear Equation of State (EoS) is an attractive subject not only for nuclear

spherical-shield-neutron-dose

SLAC National Accelerator Laboratory FACET & TEST BEAM FACILITIES PROPOSAL

The Cosmic Ray Telescope for the Effects of Radiation (CRaTER) Investigation for the Lunar Reconnaissance Orbiter

Technical University of Denmark

Simulations in Radiation Therapy

Michigan State University, East Lansing MI48824, USA INTRODUCTION

THE FORWARD DETECTOR OF THE ANKE SPECTROMETER. SCINTILLATION AND CHERENKOV HODOSCOPES

Compton Camera. Compton Camera

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Measurement of induced radioactivity in air and water for medical accelerators

Transcription:

Proton radiography with a range telescope and its use in proton therapy Juan L. Romero Department of Physics, University of California, Davis, CA Invited talk at Massachusetts General Hospital, Boston, Ma, May 31, 1995. Because of the advent of relatively high energy (245 MeV) proton accelerators for proton therapy, we were lead to reexplore the old idea of proton radiography, as first used by A. M. Koehler in 1968, and by Hanson et al. in the 70's, now in the light of present (high-energy) nuclear physics techniques. The basic idea is to make use of the proton beam per se to help position the patient/tumor relative to the beam. We describe results of recent tests done at proton energies of 70 MeV to measure density variation (like tumors) in an relatively homogeneous material like water or plastic (biological tissue). These tests were specifically designed to deal with simple, but realistic, shapes and geometries, so that we could compare the experimental results with Monte Carlo simulations. Extensions of the method at 245 MeV are discussed. 1) On our involvement (circa 1993) -- UC Davis, together with LBL, are (were?) joining

efforts to build a proton cancer treatment facility in Sacramento. NCI was considering at the time building two similar facilities, the other one located at Massachusetts General Hospital. -- We (Crocker Nuclear Laboratory and Physics at UCD) were approached by Del Kubo (UCDMC) to see if there is a simple way, other than the usual x-ray method, to position accurately (to within perhaps 5 mm) the proton beam in the tumor of the patient. -- "Why not use a sample of the beam itself and make use of energy loss sensitivity to surface density of the medium?", we asked ourselves. -- Preliminary Monte Carlo simulations showed us that there was a good possibility. Also, we soon found out about the pioneering work by others on proton radiography... (see below). -- At this point, we established a collaboration group to explore these ideas.

Collaborators J. L. Romero, J. H. Osborne, F. P. Brady, W. Caskey, D. A. Cebra, M. D. Partlan Department of Physics, University of California, Davis B. Kusko, R. S. King, I. Mirshad Crocker Nuclear Laboratory, University of California, Davis H. Kubo, I. Daftari Department of Radiation Oncology, University of California Davis Medical Center W. Chu Lawrence Berkeley Laboratory, University of California, Berkeley The results presented here have been published in: Nucl. Instr. and Methods A 356 (1995) 558-565 2) Earlier work on proton radiography The value of using beams of high-energy charged particles, in particular proton beams, as sensitive probes to detect small density variations has been recognized for many years. In 1968, Koehler used 160 MeV protons from the Harvard University cyclotron to obtain radiography of fabricated phantoms [7]. [7] A. M. Koehler, Proton radiography, Science 160 (1968) 303. Koehler showed that density differences as small as

0.05% could be detected. Working with Steward in the early 1970s, they obtained proton radiography of real tumors of the breast [1] and brain [2]. [1] V.W. Steward and A.M. Koehler, Surg. Neurology,2 (1974) 283-284. [2] V.W. Steward and A.M. Koehler, Radiol. 110 (1974) 217-221. In the late 1970s, Hanson et al. used a 240 MeV proton beam from the Los Alamos Meson Production Factory (LAMPF) to obtain computed tomography of the highest quality and sensitivity to density variation of the order of 0.013 g/cm 2 [3,4] [3] K. M. Hanson, IEEE Trans. on Nucl. Sci. N3-26 (1979) 1635. [4] K. M. Hanson et al., IEEE Trans. on Nucl. Sci. N3-25 (1978) 657. They also compared proton radiography with conventional x-ray radiography and showed that for similar quality images, protons resulted in a much lower dose to the patient than did x-rays, and that treatment doses could be located more accurately and away from sensitive areas. A general discussion on the physics of imaging with ionizing radiations is found in Kouris, Spyrou and Jackson [5] [5] Kypros Kouris, Nicholas M. Spyrou and Daphne. F. Jackson, "Imaging with ionizing radiations", Surrey University Press, 1982. 3) Description of method Principle of proton radiography For a given incident proton energy and a given density variation, the residual (detected) energy changes much

faster near the range of the incident proton as anywhere else (about 10 times for 245 MeV protons). Range-straggling Fig. 1 presents the well-known relationship between the proton energy and the range in water and also the corresponding range straggling. It is seen that for water, straggling is typically 1% of the range. For a specimen of 40 cm of water, straggling is about 4 mm. Multiple (Coulomb) scattering... of the proton beam with the atoms of the material affects the lateral precision. However, a determination of the trajectory of the proton, measured before and after traversing the phantom, allows us to correct for this effect. Typically, one expects to correct the lateral component of the proton trajectory position to within 1 mm. Conceptual design Considering the above, a conceptual design of a detector array to obtain a radiography of a patient tumor or phantom structure, by measuring the range of a high energy proton beam is sketched in Fig. 2. A (very) low intensity proton beam

passes an X-Y position-determining device such as a multi-wire proportional chamber, impinges on a patient, traverses a density anomaly such as a tumor or bone structure, emerges through a second X-Y device and finally is stopped in a total energy detector, such as the range telescope. In this case, if one uses 1 mm scintillating detectors, the typical average error in determining the range (or depth sensitivity ) is 0.05 g/cm 2. 4) Experimental setup Geometry and detectors Fig. 3 shows the experimental set-up we used at CNL. Use was made of existing DE detectors and available wire chambers to assemble a fast range telescope. The low intensity proton beam impinges on a phantom and onto the detecting telescope, where it stops. The fast range telescope consists of 5 thin DE detectors (DE1, DE2,..., DE5), with thicknesses ranging from 0.8 mm to 2 mm. An adjustable energy degrader and a trigger DE detector are located before the phantom. Wire chambers WC1 and WC2, located before and after the phantom, respectively, are used to determine more precisely the individual proton trajectory. The multiwire chambers give position (one standard deviation) close to s= ±1 mm, and also allow corrections for the effects of multiple Coulomb scattering. Electronics

The anode signal from each DE detector is fed to a fast discriminator. The threshold for the latter is set in a separate calibration run using full energy protons (67.5 MeV), guaranteeing in this way the detection of low energy protons when the phantom is used. The output from the fast discriminator is fed into a CAMAC coincidence register, which establishes how many of the five DE detectors are in coincidence with the trigger detector. The latter initiates the sequence of electronic coincidences for each event. The coincidence register and the X and Y signals from the wire chambers are stored event-by-event onto magnetic tape for off-line analysis. In this manner, each individual proton is registered as an electronic coincidence between the trigger detector and at least one of the five DE detectors. For example, an event that has a coincidence register set to detectors 1, 2 and 3 is accepted in the analysis. A coincidence register set to detectors 1, 2, 3 and 5 is rejected. In the next section, we present results from events that have consecutive hits starting from detector 1. 5) Results of experiment and simulations at 70 MeV Results to test general idea of the method Fig. 4 (right) presents the results corresponding to an acrylic phantom shaped as in Fig. 4 (left). The phantom has, as shown, a raised pattern 1.6 mm in thickness on top of an acrylic cube 30 mm in thickness. The very low intensity proton beam (about 160 protons/sec with an energy at the phantom of 66.0 MeV) impinged normal to the phantom along the Z direction. Each panel in Fig. 4 shows the distribution of events stopped in the corresponding DE detector. The X, Y values

correspond to the hits in WC1 placed upstream of the phantom, as in the setup in Fig. 3. The sequence of scatter plots of positions at the entrance wire chamber for the cases where the proton stopped in the first, second,..., and fifth detector is an excellent illustration of both the principle and sensitivity of the method. This radiography of the phantom was obtained with about 3x10 4 protons and it took 3 minutes. It is seen that most of the protons incident on the thin raised pattern are stopped in DE2, while most of the protons incident outside are stopped in DE4. Straggling, beam energy width (s 0.3 MeV), wire chambers resolution (s 1.3 mm) and multiple scattering produce diffuseness in depth (as measured by the detector where the proton stopped) and in the X,Y position (as measured by the hit in WC1). The rather good depth sensitivity allows us, in the present configuration, to distinguish thickness differences of at least 0.05 g/cm 2 (see below). Results to test simulations Another set of measurements were made using a variety of simple phantoms. The purpose here was to explore the capability of our simulation code to reproduce the experimentally observed results. Here we discuss the results using the phantom and geometry shown in Fig. 5. Simulations Briefly, the Monte Carlo simulations are done as follows: Individual protons, each with an energy of 66.0 MeV, are sent along the beam direction, the Z-axis, at random X,Y points, so they span the phantom. The incident energy is smeared according to a Gaussian distribution corresponding to a beam

energy resolution with s= 0.3 MeV. Each proton is then propagated through the phantom and, incrementally (in the beam direction) the new energy (decreased due to energy loss) and new position due to multiple scattering are calculated. The latter is done in 3-dimensions, so at each increment in the Z-axis a new position (X,Y,Z) and a new energy is obtained. This process is continued until the proton energy is zero (normally it stops in one of the DE detectors). At the point of stopping, the final position is determined by smearing the stopping position according to the range-straggling curve, using a Gaussian distribution with a standard deviation obtained from Fig. 1. Finally, the X,Y hits at the location of the wire chambers are also smeared according to the expected resolution of the wire chambers (s= 1.3 mm). The process is repeated for the desired number of protons. 6) Predictions at 245 MeV Bone structure (Fig. 9) The good agreement observed between the data and the simulations for 66 MeV protons gives us confidence in our Monte Carlo simulations and allows us to make predictions on the quality of the method for higher energy protons. An example of such predictions is presented in Fig. 9. The setup corresponds to that of Fig. 2. The phantom, shown on the left on Fig. 9, is (for simplicity) a cylinder, 2 cm in diameter and 3 cm in depth, and positioned at the center of a surrounding

media consisting of H20 (35 cm in depth, r = 1.0 g/cm 3 ). We assume the cylinder to have a composition similar to that of bone, i.e., CaO2 (r = 1.5 g/cm 3 ). The proton beam, with an energy of 245 MeV, impinges perpendicular to the main axis of the cylinder (Fig. 9). We send 3x10 5 protons over an area of 7 cm (horizontal) by 3 cm (vertical). The X-Y positions correspond to that of WC1 (s=1.3 mm). The data is binned in X-Y areas of 1 mm 2, and every 100 hits are averaged in each bin, to take advantage of statistical sampling. Increased resolution with beam intensity (fig. 10) The top part of Fig. 9 (right) shows the average (100 hits per bin) X-Y distribution at WC1 corresponding to protons stopping in detectors DE10 through DE16. The cylindrical bone was positioned with its center at X= 6 cm. The bottom panel of Fig. 9 (right) plots the detector where each proton stops (average over 100) versus the X position, and provides another way of displaying the results. (This is possible because of the symmetry of this phantom along the Y-axis.) The bone structure is discerned quite well even with much less statistics, as seen in Fig. 10, where we compare the results with the case corresponding to averaging only every ten hits (i.e., 10 times less integrated beam intensity, or 3x10 4 protons). As mentioned above, a more involved tomographic reconstruction has not been pursued, since we have dealt with rather simple phantoms in our tests and simulations. Future work should involve more complex phantoms and possibly tomographic reconstruction.

7) Recent results by the group at PSI The group at the Paul Scherrer Institute, in Switzerland have also pursued proton radiography as a tool in proton therapy. Their ideas are very similar to what we have explored, illustrating again that when technology is ready, ideas/methods pop up simultaneously in many places. Uwe Schneider and Eros Pedroni, " Proton radiography as a tool for quality control in proton therapy", Med. Phys. 22 (4), April 1995. Their studies differ from ours in several respects. Here are the most important ones: -- use of a 219 MeV proton beam -- use of a range shifter -- use of a NaI detector to measure residual energy (75 mm in diameter, 100 mm thick) -- they took proton radiography of the Alderson phantom -- counting rate limited to only 1000 protons per second, which implied long running times, 2 h. -- they used the spatial information of the two wire chambers to reconstruct the intercepts of the most likely proton trajectory of every event at every point between the two chambers. Then they used the intercept coordinates of the calculated proton trajectory and the plane to be reconstructed to form images at different depths in the phantom. This is an important work. They also have done detailed studies of multiple scattering and range uncertainties. They are setting up a system like ours, with a range telescope, to speed up the data rate at least 2 order of magnitude. They are also starting a Monte Carlo simulation program. 8) Final remarks

We have described a method to locate density variations in phantoms that makes use of high energy proton beams. The method is thought to be applicable to the positioning of a patient in a proton therapy facility. The method makes use of a proton range telescope for density variation and X-Y position sensitive detectors for planar positioning. We have tested the principle of the method with measurements at a proton energy of 66 MeV. Good visual quality is seen in the tests. We have compared the measurements with detailed Monte Carlo simulations and achieved good agreement. Further simulations with high energy proton beams (245 MeV) show that the method should provide good visual quality and sensitivity for positioning. In the case of a proton therapy facility, the latter could be achieved by using the proton therapy beam per se in a (very) low intensity mode.