M.J. CIESLAK, K.A.A. GAMAGE Department of Engineering, Lancaster University Lancaster, LA1 4YW United Kingdom ABSTRACT

Similar documents
Digital two-parametric processing of the output data from radiation detectors

SCINTILLATION DETECTORS AND PM TUBES

arxiv: v1 [physics.ins-det] 29 Jun 2011

EEE4106Z Radiation Interactions & Detection

Nuclear Physics and Astrophysics

Chinese Physics C. Study of n-γ discrimination in low energy range (above 40. kevee) by charge comparison method with a BC501A liquid

An investigation into a suitable scintillator for localising neutron capture within a detector

Louis Baum University of Michigan, REU Student Summer UM Nuclear Sciences Group

Radiation Detection and Measurement

Neutron Detection. n interactions with matter n detection High/low energy n detectors

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT

Radiation Detection and Measurement

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Mass of the electron m 0

national security, nonproliferation, defense, border security, and homeland security applications.

DETECTORS. I. Charged Particle Detectors

Chapter 4 Scintillation Detectors

RADIATION DETECTION AND MEASUREMENT

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Realizing Real-Time Capabilities of an Embedded Control System for Fast-Neutron Scintillation Detectors

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Timing Characterization of Helium-4 Fast Neutron Detector with EJ-309 Organic Liquid Scintillator

Measurements of liquid xenon s response to low-energy particle interactions

Characterization of the new scintillator Cs 2 LiYCl 6 :Ce 3+ *

Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

Development of a Radioxenon Detection System Using CdZnTe, Array of SiPMs, and a Plastic Scintillator

BUREAU INTERNATIONAL DES POIDS ET MESURES

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Nuclear Instruments and Methods in

Pulsed Neutron Interrogation Test Assembly - PUNITA

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Calibration of EJ-301, a Liquid Scintillator

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

Quality Assurance. Purity control. Polycrystalline Ingots

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

A Comparison between Channel Selections in Heavy Ion Reactions

Neutron Spectroscopy in Proton Therapy

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Cf-252 spontaneous fission prompt neutron and photon correlations

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Dedicated Arrays: MEDEA GDR studies (E γ = MeV) Highly excited CN E*~ MeV, 4 T 8 MeV

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Scintillation detectors

Scintillation efficiency measurement of Na recoils in NaI(Tl) below the DAMA/LIBRA energy threshold

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Scintillators General Characteristics

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007

arxiv:physics/ v1 3 Aug 2006

1 Geant4 to simulate Photoelectric, Compton, and Pair production Events

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

XMASS: a large single-phase liquid-xenon detector

Investigation of Ce-doped Gd2Si2O7 as a scintillator. Author(s) Kurashige, Kazuhisa; Ishibashi, Hiroyuki; Furusaka, Instructions for use

Scintillation Detector

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner

PHYS 3650L - Modern Physics Laboratory

MCNP6 simulation validation of fast neutron coincidence detection system for nuclear security and safeguards applications

X ray Imaging Android [XIA]

LAAPD Performance Measurements in Liquid Xenon

High Energy Neutron Scattering Benchmark of Monte Carlo Computations

MEASUREMENT AND DETECTION OF RADIATION

Development of a High Precision Axial 3-D PET for Brain Imaging

Light ion recoil detector

Measurement of Muon Lifetime

DELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS

Chapter Seven (Nuclear Detectors)

Radiation (Particle) Detection and Measurement

Test bench for measurements of NOvA scintillator properties at JINR D.S. Velikanova, 1 A.I. Antoshkin, 1 N.V. Anfimov, 1 O.B.

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Ashley Richard Jones. MEng. Nuclear Engineering (Hons) This thesis is submitted in accordance with the requirements of the.

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Shielded Scintillator for Neutron Characterization

THE COMPTON EFFECT Last Revised: January 5, 2007

arxiv: v2 [physics.ins-det] 28 Jul 2010

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

This is a repository copy of Characteristics of alpha, gamma and nuclear recoil pulses from NaI(Tl) at kev relevant to dark matter searches.

Measuring Cosmic Ray Muon Decay Constant and Flux

Overview of Current Radioxenon Research

PHYS 3446 Lecture #12

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

a monte carlo simulation of photomultiplier resolution

Neutron Time-Of-Flight Spectrometer Based on HIRFL for Studies of Spallation Reactions Related to ADS Project

Detection and measurement of gamma-radiation by gammaspectroscopy

Radioactivity. Lecture 6 Detectors and Instrumentation

Chemical Engineering 412

arxiv: v1 [physics.ins-det] 16 May 2017

III. Energy Deposition in the Detector and Spectrum Formation

Lecture 3 - Compton Scattering

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

SILVER NEUTRON ACTIVATION DETECTOR FOR MEASURING BURSTS OF 14 MEV NEUTRONS

Dead-Time Measurement for Radiation Counters by Variance-to-Mean

OPPORTUNITY TO JOIN IEEE AND NPSS

HIGH-PRESSURE HELIUM-3 SCINTILLATION POSITION-SENSITIVE DETECTOR OF THERMAL NEUTRONS

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Transcription:

PULSE SHAPE DISCRIMINATION CHARACTERISTICS OF STILBENE CRYSTAL, EJ-299-33 AND 6 Li LOADED PLASTIC SCINTILLATORS FOR A HIGH RESOLUTION CODED-APERTURE NEUTRON IMAGER M.J. CIESLAK, K.A.A. GAMAGE Department of Engineering, Lancaster University Lancaster, LA1 4YW United Kingdom ABSTRACT Pulse shape discrimination performances of single stilbene crystal, EJ-299-33 plastic and 6 Li loaded plastic scintillators have been compared. Pulse Gradient Analysis pulse shape discrimination algorithm has been tested for each scintillator sample, assessing their neutron/gamma separation. In this study each scintillator sample was irradiated with a 252 Cf neutron source and, a real-time fast digitiser was used to collect the data. The figure-of-merit was utilised to compare the discrimination quality of the tested scintillator samples. 1. Introduction Pulse shape discrimination (PSD) is a well-established method of separating fast neutron and gamma-ray interactions within organic scintillation medium. The method is based on the difference in the decay time of fluorescence emitted within an organic scintillator as a result of an interaction between the ionising particle and the scintillant. The fluorescence decay time observed for heavy ionising particles, such as protons, is longer when compared to electrons [1]. Fast neutrons and gamma-ray photons interact with an organic scintillant predominantly through elastic scattering with a proton and Compton scattering, respectively. Consequently, the fluorescence decay rate exhibited by recoil protons and recoil electrons (Compton scattering) can be compared to infer the origin of the interaction [2]. The difference in the fluorescence decay rate formed a basis for neutron/gamma PSD techniques in organic scintillators. In this paper, PSD performance of three different scintillator types is analysed. Pulse Gradient Analysis (PGA) [3] method is applied aiming to identify the most suitable scintillator material for a real-time scintillation based coded-aperture neutron detector. In order to evaluate neutron/gamma separation performance of the scintillator, the figure-of-merit (FOM) values were estimated for each scintillator sample. 2. Experimental Method Three different organic solid scintillator types were in turn irradiated with a 252 Cf source located at Lancaster University, Lancaster, UK. Single stilbene crystal (20 mm diameter, 20 mm thick) - was provided by Inradoptics [4], whereas pure PSD plastic scintillator sample (25 mm diameter, 25 mm thick sample number 5706) and 6 Li loaded PSD plastic scintillator sample (40 mm diameter, 25 mm thick sample number 9023) were both provided by the Lawrence Livermore National Laboratory (LLNL), USA.

The back of each scintillator sample was covered with EJ-510 reflective coating and each sample was then coupled to an ET Enterprises 9107B PMT with EJ-550 silicone grease. The PMT module supplied by ET Enterprises was enclosed in a light-proof tube. Depending on the scintillator type the positive high voltage supply, connected to the cathode of the PMT was varied between 850 V and 900 V. The PMT anode was connected to a real-time Mixed Field Analyser (MFA) developed by Hybrid Instruments Ltd. [5]. Experimental setup is shown in Fig. 1. Fig 1. Experimental set-up. 252 Cf source is located in the centre of a water filled, metal tank. During the experiments the source is moved to the edge of the tank pointed by the arrow on the left. Discrimination data were collected from the MFA digitiser for each scintillator sample. MFA is already programmed to discriminate neutrons and gamma-rays based on PGA algorithm and, the results corresponding to each scintillator sample, were then compared. 3. Results Number of events were plotted against the discrimination index, which was calculated as a ratio of first integral over second integral. It was used to assess the separation of neutron and gamma-ray events. The FOM as defined in Eq. (1) was calculated for each scintillator to compare their performance in separating neutrons from gamma-rays. Results are presented in Table 1. FOM = Peak separation (1) FWHM γ + FWHM n Scintillator Exposure time FOM 6 Li Loaded plastic 30 min - PSD plastic 30 min 0.551 Stilbene crystal 30 min 0.941

Tab 1: FOM calculations for each scintillator

a) b) c) Fig 2. Scatter plots using PGA - first integral against second integral for each scintillator sample: a) 6 Li loaded PSD plastic b) pure PSD plastic c) single stilbene crystal (Note: all the plots presented in the paper were produced using Matplotlib package.) Discussion Separation quality between neutrons and gamma-rays has been assessed by comparison of FOM values for each scintillator sample. There is a clear indication based on FOM values, as well as scatter plots in Fig. 2 that single stilbene crystal presents superior separation quality. Although PGA algorithm failed to separate neutrons from gamma-ray photons within 6 Li loaded plastic scintillator (see Fig.2. a), FOM value for the pure PSD plastic scintillator suggests relatively good discrimination performance. This difference in separation quality between two plastic scintillators is related to the scintillator loading, which enables neutrons thermalised within the detector to be captured by high capture cross-section 6 Li. As such, 6 Li loaded plastic scintillator can be beneficial for certain application but neutron capture events are difficult to separate from gamma-ray photons as presented in this and previous studies [6]. Acknowledgments The authors would like to acknowledge the funding support from EPSRC via Faculty of Science and Technology, Lancaster University, UK and Sellafield Ltd., UK.

References [1] F.D. Brooks, A scintillation counter with neutron and gamma-ray discriminators, Nucl. Instruments Methods. 4 (1959) 151 163. doi:10.1016/0029-554x(59)90067-9. [2] G.F. Knoll, Radiation Detection and Measurement, 4th ed., John Wiley & Sons, Hoboken, 2010. [3] B. D Mellow, M.D. Aspinall, R.O. Mackin, M.J. Joyce, A.J. Peyton, Digital discrimination of neutrons and -rays in liquid scintillators using pulse gradient analysis, Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip. 578 (2007) 191 197. doi:http://dx.doi.org/10.1016/j.nima.2007.04.174. [4] Inradoptics, Stilbene Single Crystals - Data Sheet, (n.d.). http://inradoptics.com/pdfs/datasheets/inradoptics_datasheet_stilbene_final.pdf. [5] M.J. Joyce, M.D. Aspinall, F.D. Cave, A. Lavietes, A 16-channel real-time digital processor for pulse-shape discrimination in multiplicity assay, IEEE Trans. Nucl. Sci. 61 (2014) 2222 2227. doi:10.1109/tns.2014.2322574. [6] M.J.I. Balmer, K.A.A. Gamage, G.C. Taylor, Comparative analysis of pulse shape discrimination methods in a 6Li loaded plastic scintillator, Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip. 788 (2015) 146 153. doi:10.1016/j.nima.2015.03.089.