MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

Similar documents
The power of multi-detector SEC in the analysis of antibodies

Measurements of protein electrophoretic mobility using the Zetasizer Nano ZSP

Differentiation of polymer branching and composition using the Mark Houwink plot

GPC/SEC Analysis of Polymer Degradation

Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv)

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

The Diffusion Barrier Technique, Practical Aspects and Data interpretation

Introduction to the calculators in the Zetasizer software

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

How to use GPC/SEC for compositional analysis

Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Use of SEC-MALS. (Size Exclusion Chromatography - Multi Angle. Light Scattering) for protein quality and characterization

How switching mobile phases can improve your OMNISEC results

Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

Malvern Macromolecular Solutions. Innovative Solutions in Material Characterization

DEVELOPING A BIOFORMULATION STABILITY PROFILE

Use of DLS/Raman to study the thermal unfolding process of lysozyme

10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS

Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector

Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments

The ultimate in desktop particle characterization

Electrophoretic Light Scattering Overview

REVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID.

The ultimate in desktop particle characterization

Component specific particle characterisation of the active components in a pharmaceutical topical formulations

Determining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer

Nanobubble Applications and Characterization by Nanoparticle Tracking Analysis

Studying the Effect of Steric Layer Thickness on Emulsion Stability

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

Sample preparation and characterization around SAXS

Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications

Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution

Overcoming and quantifying Wall Slip in measurements made on a rotational rheometer

The Illumination of Structure using Light Scattering. Michael Caves Product Technical Specialist for Biophysical Characterisation

ZS90. The ultimate in desktop particle characterization. Particle size. Zeta potential. Molecular weight

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Tips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

An Introductions to Advanced GPC Solutions

Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation

GPC / SEC Theory and Understanding

Understanding the conformational stability of protein therapeutics using Raman spectroscopy

Static and dynamic light scattering. Cy Jeffries EMBL Hamburg

Size Exclusion Chromatography in the Presence of an Anionic Surfactant

Progress toward reliable NC molecular mass distribution by GPC

Tips & Tricks GPC/SEC: Inter-Detector Delay

Evaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications

Protein Aggregate Assessment of Ligand Binding Assay (LBA) Reagents Using SEC-MALS

Executive Summary. Introduction WHITEPAPER

Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy

Characterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy

BSA pegylation PG Journal of chromatographic A, 1147 (2007)

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Using runt Disposable Cuvettes

Interaction of Gold Nanoparticle with Proteins

Clemens Stilling Characterisation, Analytical Sciences for Biologicals, UCB Pharma S.A. Alun, living with Parkinson s disease

Using rheology to study the hardness and spreadability of butter

OVERVIEW INTRODUCTION

How Molecular Weight and Branching of Polymers Influences Laser Sintering Techniques

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Inform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

Using the Zetasizer Nano to optimize formulation stability

Highly automated protein formulation development: a case study

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Sample characterization: Quality control and sample handling prior to data collection

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison

Introduction to Biological Small Angle Scattering

Impurities In Biomolecules Institute for International Research. Monitoring & Predicting Biomolecular Aggregation Using Light Scattering

Biophysics Service at the MPIB Biochemistry Core Facility Stephan Uebel, Biochemistry Core Facility

Setting the Standard for GPC. Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies. The Right Instrument for Your Application

Sem /2007. Fisika Polimer Ariadne L. Juwono

Sepax SRT -C, Zenix -C SEC Phases

Strategies to compare far-uv Circular Dichroism spectra of similar proteins using Chirascan -plus ACD

Gel Permeation Chromatography

GPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC

High Performance Liquid Chromatography

Supporting Information

Meet Our Uncle: 12 Stability Applications on One Platform

Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection

No.106. Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series. Contents. Page

Particle Characterization Laboratories, Inc.

PARTICLE SIZE CHARACTERISATION OF AGRICULTURAL SPRAYS USING LASER DIFFRACTION

Gel Permeation Chromatography

MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)

next generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size

Size Exclusion Chromatography

EXCLUSION HIGH PERFORMANCE LIQUID CHROMATOGRAPHY ANALYSIS OF A CELLULOSE ACETATE-AMMONIUM NITRATE GAS GENERATOR PROPELLANT

FFF-MALS method development and measurements of size and molecular weight

COURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download)

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Vrije Universiteit Amsterdam

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography

Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns

The ph-responsive behaviour of aqueous solutions of poly(acrylic acid) is dependent on molar mass

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Transcription:

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20 Introduction Protein aggregation is recognised as a major issue in the biopharmaceutical industry. Proteins have a tendency to aggregate over time and the risk for biologic drugs is that the presence of aggregates will stimulate an immune response. Size-exclusion chromatography (SEC) is a powerful tool that is commonly used to look at the aggregation of proteins. SEC separates proteins by size. It is commonly used to measure their molecular weight and characterise aggregation. While most SEC systems use a single concentration detector such as ultraviolet (UV), the addition of light scattering allows the measurement of the molecular weight of the protein independent of its retention volume. The molecular weight of the protein monomer, oligomers and aggregates in a sample can then be measured to understand its behaviour. Multi angle light scattering (MALS) can also be used to measure the radius of gyration (Rg) of molecules that scatter light anisotropically (unevenly with respect to angle). The Viscotek SEC-MALS 20 system (figure 1) is a 20 angle light scattering device capable of measuring the molecular weight of protein and protein aggregates independent of elution volume. The addition of further detectors such as concentration detectors (RI, UV), intrinsic viscosity (IV) and dynamic light scattering (DLS) can greatly increase the amount of information available from a single SEC measurement. In this application note, a selection of proteins are separated using SEC. The molecular weights of their oligomers and aggregates are measured by SEC-MALS and the results discussed. Figure1: The Viscotek SEC-MALS 20 system 2013 Malvern Instruments Limited MRK1927-01 1

Materials and methods The SEC-MALS 20 system was connected to a Viscotek TDAmax system using the TDA RI detector to measure concentration. Samples were separated using 2 x Viscotek protein columns. The mobile phase was phosphate-buffered saline and all the samples were dissolved in the mobile phase. The SEC-MALS system was calibrated using bovine serum albumin, a protein with a well characterised molecular weight. The detectors and columns were all held at 30 C to ensure a good separation and maximise baseline stability of the detectors. Results Samples of BSA are typically a mixture of monomer, oligomers and larger aggregates. Here, monomer, dimer and trimer peaks are clearly visible and their molecular weights have accurately been measured by MALS. The sample also contains some more polydisperse higher order aggregates that are not individually resolved. The BSA chromatogram can be seen in figure 2 and the results from the BSA measurement can be seen in table 1. Figure 1: Chromatogram of BSA showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. Table 1: Molecular weights of the different components of the BSA sample. Aggregates Trimer Dimer Monomer Peak RV (ml) 13.84 14.31 15.37 17.05 Mn (kda) 337.3 199.4 133.6 66.1 Mw (kda) 378.3 201.2 133.9 66.3 Mw / Mn 1.122 1.009 1.003 1.002 Rg(w) (nm) N/C N/C N/C N/C Wt Fr (Peak) 0.04 0.057 0.423 0.745 The molecular weights across the monomer, dimer and trimer peaks are clearly seen to be very stable. This is expected with proteins which individually have very tightly controlled molecular weights and are considered to be monodisperse. While these molecular weights are stable, indicating that these are oligomers, the molecular weight trace is more variable in the earliest eluting material that is made up from a more polydisperse population. This is more typical of denatured aggregates which are likely to no longer be functional. The fraction of the total sample in each peak is identified in the Wt Fr row. SEC-MALS 20 application note - MRK1927-01 2

The second protein studied was thyroglobulin and its chromatogram can be seen in figure 2. The MALS signals from the different angles can be seen in the MALS plot in figure 3 and the calculated results can be seen in table 2. Figure 2: Chromatogram of thyroglobulin showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. Figure 3: Chromatogram of SEC-MALS detector signals for thyroglobulin. SEC-MALS 20 application note - MRK1927-01 3

Table 2: Measured molecular weights of the different peaks of the thyroglobulin sample. Aggregates Monomer Peak RV (ml) 11.60 12.50 Mn (kda) 1601.0 681.5 Mw (kda) 2588.0 686.0 Mw / Mn 1.617 1.007 Rg(w) (nm) N/C N/C Wt Fr (Peak) 0.167 0.833 The thyroglobulin sample has been separated into 2 peaks according to the RI detector but 3 according to the SEC-MALS detector. The SEC-MALS detector is identifying a very small amount of very large aggregates that are only just visible to the RI detector. The molecular weight of the monomer has been measured at 686 kda. This is within 3% of the 669 kda molecular weight of thyroglobulin. This is stable across the peak indicating this population is monodisperse. The aggregate peaks, which have been analysed as a single component, show a higher and more variable (polydisperse) molecular weight clearly identifying them as some disordered aggregates. It is unlikely that any of the proteins in this peak will have any activity and if it were a drug sample, these aggregates would be very likely to be immunogenic. The aggregates account for approximately 17% of the sample. It is worth noting that despite the high molecular weight of this protein, the SEC-MALS display clearly shows that the scattering from the monomer and the aggregates is isotropic. As such, it is not possible to measure the Rg of any of the peaks in this sample. The final sample measured was pepsin and its chromatogram can be seen in figure 4. The SEC-MALS signals from the different angles can be seen in the MALS plot in figure 5 and the calculated results can be seen in table 3. Figure 4: Chromatogram of pepsin showing the refractive index (red) and SEC-MALS (90 ) (orange) detector signals. The molecular weight measured by SEC-MALS is overlaid in black. SEC-MALS 20 application note - MRK1927-01 4

Figure 5: Chromatogram of SEC-MALS detector signals for pepsin. Table 3: Measured molecular weights of the different peaks of the pepsin sample. Aggregates Monomer Degradation products Peak RV (ml) 10.84 18.33 20.92 Mn (kda) 3892.0 34.4 4.7 Mw (kda) 4431.0 34.7 6.4 Mw / Mn 1.138 1.008 1.364 Rg(w) (nm) 69.9 N/C N/C Wt Fr (Peak) 0.008 0.569 0.423 The pepsin sample has been separated into 2 main peaks. The molecular weight of the second peak is measured at 34.7 kda which is very close to known molecular weight of pepsin of 35 kda. The larger peak has a much higher and more variable molecular weight clearly identifying it as some disordered aggregates which are unlikely to be active. Despite being such a large light scattering peak, the aggregates account for less than 1% of the total material in the sample. In this case, the MALS plot shows that these aggregates are strongly anisotropic scatterers. This means that their Rg can actually be measured, and is shown in the table to be nearly 70 nm. Pepsin has a nonglobular structure which means that it is larger than generally expected for a protein of that molecular weight. The size of the aggregate shows that the structure actually persists in the aggregates, which then become very large. This compares with the thyroglobulin aggregates which were much smaller (<15 nm) despite being over half the molecular weight of the pepsin aggregates. The pepsin monomer is too small for the measurement of Rg. There is a third peak in the sample which is quite poorly defined but it is possible to measure its molecular weight as approximately 6 kda. As a digestive enzyme, it is quite likely that this pepsin sample is digesting itself and that this represents some of the digestion products. Discussion This application note has shown the successful measurement of the molecular weight and aggregate content of a number of proteins using the Viscotek SEC-MALS 20 system. This instrument can measure protein molecular weight independent of a protein s elution volume or structure. The molecular weight of BSA oligomers and thyroglobulin and pepsin aggregates were successfully measured and their amounts quantified. Where the aggregates are large enough, their size (Rg) can also be measured. SEC-MALS 20 application note - MRK1927-01 5

Malvern Instruments Ltd Enigma Business Park Grovewood Road Malvern Worcestershire UK WR14 1XZ Tel: +44 (0)1684 892456 Fax: +44 (0)1684 892789 Malvern Instruments Worldwide Sales and service centers in over 50 countries for details visit www.malvern.com/contact Malvern Instruments Ltd 2013 more information at www.malvern.com SEC-MALS 20 application note - MRK1927-01 6