CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

Similar documents
SIZE-SELECTED AEROSOL FILTER ANALYSIS USING ELECTROSTATIC CLASSIFICATION AND LIBS

MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

TSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE

Introduction to Fourier Transform Infrared Spectroscopy

Observation of Extreme Phase Transition Temperatures of Water Confined Inside Isolated Carbon Nanotubes

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

Introduction to Fourier Transform Infrared Spectroscopy

Heat Flow Calculations Made Using the Models 8386 and 8386A VELOCICALC Plus Air Velocity Meters Application Note ITI-124

Quantitative Analysis of Carbon Content in Bituminous Coal by Laser-Induced Breakdown Spectroscopy Using UV Laser Radiation

Estimation of Mass with the Model 3321 APS Spectrometer

Imaging Carbon materials with correlative Raman-SEM microscopy. Introduction. Raman, SEM and FIB within one chamber. Diamond.

MEASURING NANOPARTICLE EXPOSURE

Liberating the Nanotube: From Lab to Large Scale Production

Obsolete Product(s) - Obsolete Product(s)

HCF4555B DUAL BINARY TO 1 OF 4 DECODER/DEMULTIPLEXER OUTPUT HIGH ON SELECT

RamanStation 400: a Versatile Platform for SERS Analysis

Supporting Information s for

TECHNICAL SPECIFICATION

BAR42FILM BAR43FILM SMALL SIGNAL SCHOTTKY DIODE. Table 1: Main Product Characteristics I F(AV) 0.1 A V RRM. 30 V T j 150 C 0.33 and 0.

Workshop II Nanomaterials Surfaces and Layers Commercialising Carbon Nanotubes

Supplementary Materials for

International Standardization for Measurement and Characterization of Nanomaterials

HCF4532B 8-BIT PRIORITY ENCODER

Obsolete Product(s) - Obsolete Product(s)

HCF4035B 4 STAGE PARALLEL IN/PARALLEL OUT SHIFT REGISTER

Nanocrystalline Cellulose:

M74HC20TTR DUAL 4-INPUT NAND GATE

TECHNICAL SPECIFICATION

Trapping Lithium into Hollow Silica Microspheres. with a Carbon Nanotube Core for Dendrite-Free

Obsolete Product(s) - Obsolete Product(s)

ICP-MS. High Resolution ICP-MS.

M74HCT688TTR 8 BIT EQUALITY COMPARATOR

2N3904 SMALL SIGNAL NPN TRANSISTOR

Obsolete Product(s) - Obsolete Product(s)

National Technical University of Athens School of Applied Mathematical and Physical sciences

Obsolete Product(s) - Obsolete Product(s)

ab83360 Ammonia Assay Kit

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

M74HC147TTR 10 TO 4 LINE PRIORITY ENCODER

TEA6422 BUS-CONTROLLED AUDIO MATRIX

Supporting Information

ab83360 Ammonia Assay Kit

Carbon Nanotube Reference Materials and Characterization Jeffrey A. Fagan

Mercury(II) detection by SERS based on a single gold microshell

Supplementary information for:

Obsolete Product(s) - Obsolete Product(s)

HCF4089B BINARY RATE MULTIPLIER

The Future of Spectroscopy

Executive Summary. Introduction WHITEPAPER

Carbon Nanotubes for Interconnect Applications Franz Kreupl, Andrew P. Graham, Maik Liebau, Georg S. Duesberg, Robert Seidel, Eugen Unger

IN QUALITATIVE ANALYSIS,

HYPHENATED TECHNOLOGY GUIDE

Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy

Metallic/semiconducting ratio of carbon nanotubes in a bundle prepared using CVD technique

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

ELECTROSTATIC CLASSIFIER MODEL 3082

Fundamental investigation on CO 2 laser-produced Sn plasma for an EUVL source

M74HCT138TTR 3 TO 8 LINE DECODER (INVERTING)

The NanoSpectralyzer. Models: NS1 NS2 NS3 NS MiniTracer. for Nanomaterial and Carbon Nanotube Characterization

BAT60J SMALL SIGNAL SCHOTTKY DIODE FEATURES AND BENEFITS VERY SMALL CONDUCTION LOSSES NEGLIGIBLE SWITCHING LOSSES LOW FORWARD VOLTAGE DROP

STPS1L40M LOW DROP POWER SCHOTTKY RECTIFIER. Table 1: Main Product Characteristics I F(AV) V RRM. 1 A 40 V T j (max) 150 C V F (max) 0.

SUPPLEMENTARY INFORMATION

Regression and Nonlinear Axes

ab Caspase 3, Caspase 8 and Caspase 9 Multiplex Activity Assay Kit (Fluorometric)

Introduction to Spectroscopy: Analysis of Copper Ore

Remote Raman & Fluorescence Capabilities for Chemical Detection at University of Hawaii

Obsolete Product(s) - Obsolete Product(s)

ab Carrez Clarification Reagent Kit

Carcino embryonic Antigen Human ELISA Kit

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX

HCF4543B BCD-TO-7-SEGMENT LATCH/DECODER/LCD DRIVER

Differential Spectral Imaging of the CN Violet Band in Laser-Induced Plasmas on TNT Simulant Molecules

Electronic Supplementary Information

Performance and Control of the Agilent Nano Indenter DCM

STTH6004W. Ultrafast high voltage rectifier. Table 1: Main product characteristics I F(AV) 60 A V RRM. 400 V T j (max) 175 C V F (typ)

Supporting Information

Instantaneous reduction of graphene oxide at room temperature

STPS2030CT/CG/CR LOW DROP POWER SCHOTTKY RECTIFIER MAJOR PRODUCTS CHARACTERISTICS. 2x10A 30 V Tj (max) 150 C

A Fluorometric Analysis of Quinine in Tonic Water

SAXS COURSE Richard Koschuch. HECUS M.Braun Graz X-Ray Systems. and. Institute of Biophysics and X-Ray Structure Research,

specified quantity of a solvent at a given temperature. To deconvolute the value from the

Highly efficient SERS test strips

M74HC4543TTR BCD TO 7 SEGMENT LATCH/DECODER/LCD DRIVER

Analytical Performance & Method. Validation

HYPHENATED TECHNOLOGY GUIDE

Supporting information S1. Procedures Materials and methods Preparation of the reagents Preparation of the calibration standard solutions

Snowy Range Instruments

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

Obsolete Product(s) - Obsolete Product(s)

CapaciTorr B MK5

FSA TM Multi-bit Digital Processors

Optimization of laser-produced plasma light sources for EUV lithography

Monitoring of Airborne Nanoparticles in Research Laboratories at the Center for High-Rate Nanomanufacturing (CHN)

1N581x LOW DROP POWER SCHOTTKY RECTIFIER MAJOR PRODUCTS CHARACTERISTICS. 1 A 40 V Tj 150 C

ab Alkaline Phosphatase Assay Kit (Luminometric)

Processing and Applications of Carbon Nanotubes, Graphene, and Beyond

Transcription:

CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER APPLICATION NOTE LIBS-019 Carbon Nanotubes (CNTs) are used in an ever increasing number of applications, from engineering applications (lighter, stronger, more flexible materials) to medical applications in cancer research. Laser-Induced Breakdown Spectroscopy (LIBS) is very well suited to help with the analysis of metallic residues in the CNTs, stemming from catalysts, substrates and impurities. Samples Two different samples of CNTs were analyzed. The physical and chemical characteristics are shown in Table 1 below: Table 1: Physical and Chemical Characteristics of CNT Samples Purity Ash content OD (nm) Length (µm) C Ni Co SWCNT >60% <1.5% 1-2 5-30 96.3 2.91 Fe MWCNT >90% <1.5% 50-80 10-20 97.3 1.88 0.56 In order to get a wider range of compositions, a number of mixtures of the two CNT samples from Table 1 were prepared. Table 2 below shows the relative amounts (by mass, in %) of the two samples in the mixtures: Table 2: Mixing ratios for CNT samples MWCNT Mix 5 Mix 2 Mix 3 Mix 1 Mix 4 Mix 6 SWCNT % SWCNT 0 7 29 53 64 96 99.2 100 %MWCNT 100 93 71 47 36 4 0.8 0 Sample Preparation Mixtures were prepared by weighing amounts of the two samples in Table 1 and mixing them in a plastic bag. For all samples, the powder (mixture or pure) was pressed into a pellet of 7 mm diameter and 2 to 3 mm thick in a fume hood. Each pellet had a mass of ~40 mg. Pellets were placed in a holder to keep the pellet in place during analysis and facilitate transport and storage. TSI and TSI logo are registered trademarks of TSI Incorporated.

Sample Analysis Samples were analyzed in a TSI LIBS Desktop Analyzer. The TSI LIBS Desktop Analyzer used a 50 mj 1064 nm Nd:YAG laser and a 4 channel broadband spectrometer. Instrument settings were as follows: Laser output: 60% Spot size: 100 µm Sample shots: 5 (shots per spot) Layout: 4 x 8 grid (32 spots total) Spacing: 0.5 mm x 0.5 mm Acquisition delay: 1 µs Data Analysis The spectra of the 32 spots per pellet were averaged to increase the signal to noise ratio. The averaged spectrum of Mix 3 is shown below. Figure 1: LIBS spectrum (average of 32 shots) of CNT Mix #3 Peaks were chosen for the elements analyzed (C, Co, Ni, Fe) and the peak area was integrated for these peaks for all analyzed mixes. Figure 2 below shows the different regions chosen for the four peaks, again for the spectrum of Mix 3. As the C concentration does not vary very much between samples (96.3 to 97.3%), the C peak at 247.8 nm is a good internal standard. Therefore, all peak areas were divided by the C 247 peak for creating calibration curves. Page 2 of 5

Figure 2: Integrated peak areas for elements of interest, shown for mixture #3. Figure 3 below shows the linear calibration curves for the three metals analyzed (Ni, Co, Fe). The dotted lines indicate the linear regression curves for the metal calibrations. Error bars indicate 1 standard deviation where several measurements were made. Figure 3: Calibration curves for the 3 metals of interest. Error bars indicate ± 1 standard deviation for samples with several measurements, and dotted lines depict the linear regression curves. The regression results (slope, intercept, regression coefficient R 2 ) are listed in Table 3 below. Table 3: Regression values of merit for metals analyzed Metal Co Ni Fe Slope 9.39 4.32 20.47 Intercept -0.024 0.009 0.033 R 2 0.8877 0.9295 0.9378 Page 3 of 5

Concentration Predictions To test the calibration, the following procedure was used: For each element tested, one datapoint was removed and the calibration curve was calculated with the remaining datapoints (The pure SWCNT and MWCNT were not removed, in order not to have to extrapolate the data). The removed datapoint was then predicted using the new calibration curve. Figure 4: Known versus predicted metal concentration graphs. Known concentrations are plotted on the x-axis, predicted values on the y-axis. Dashed lines indicate 1-1 relationship. Each graph shows the known metal concentration on the x-axis and the predicted value on the y- axis. The dotted lines indicate the 1-to-1 ratio. With few exceptions, the predicted values are very close to the measured ones, indicating that this method is very good for metal concentration measurements in Carbon Nanotubes. Results can be improved by including more mixtures, several measurements for each mixture, and improving the homogeneity of the mixtures. It is important to create a calibration with matrix-matched materials so CNT impurity analyses should be conducted with CNT standards. Pellets should also be made in a consistent matter, always using the same pressure and dwell time when pressing them, to ensure reproducible sample density. The LIBS method is an elemental analysis technique, and while it can measure the concentration of carbon (not shown here), it cannot differentiate between pure SWCNT and pure MWCNT. With Raman measurements on the same samples, which could give additional information (SW vs. MW CNT, tube diameter, etc.), it is believed a comprehensive analysis of CNT materials may be possible. Page 4 of 5

Conclusions LIBS analysis of several mixes of Carbon Nanotubes with different metal concentrations showed good ability to predict metal impurities and contamination in CNTs. Minimal sample preparation was required to create pellets for analysis. TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 LIBS-019 Rev C 2013 TSI Incorporated Printed in U.S.A.