UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT

Similar documents
THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

Force/Torque measuring facility for friction coefficient and multicomponent

COMPLETION AND MEASUREMENT UNCERTAINTY BUDGET OF THE MULTI-COMPONENT MEASURING DEVISE FOR FORCE UP TO 1 MN AND TORQUE UP TO 2 KN M

INVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

MEASUREMENT UNCERTAINTY OF ROTATING TORQUE TRANSDUCERS WHEN USED IN PARTIAL LOAD RANGES

Modelling of a dynamic torque calibration device and determination of model parameters

Concept and setup of a multi-component facility for force and torque in the range of 1 MN and 2 kn m

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Evaluation of Force Standard Machines by using Build-up System

Investigation of Piezoelectric Force Measuring Devices in Force Calibration and Force. Standard Machines

THE UNCERTAINTY OF MEASUREMENT IN CALIBRATION USING A COMPARISON FORCE STANDARD MACHINE

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

PRELIMINARY DRAFT INDIAN STANDARD METHOD FOR CALIBRATION AND CLASSIFICATION OF TORQUE MEASURING DEVICES INCLUDING TORQUE WRENCH TESTER

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

Uncertainty of Calibration. Results in Force Measurements

FEEDBACK CONTROLLED DEADWEIGHT MACHINE

EVALUATION OF STATIC AND DYNAMIC PARASSITIC COMPONENTS ON THE INRIM 1 MN PRIMARY FORCE STANDARD MACHINE BY MEANS THE 500 kn SIX-COMPONENT DYNAMOMETER

Investigation of basic elements loading and tension of heavy hydraulic presses for metallurgical production

Guidelines on the Calibration of Static Torque Measuring Devices

Final Report on the Torque Key Comparison CCM.T-K2 Measurand Torque: 0 kn m, 10 kn m, 20 kn m Dirk Röske 1), Koji Ogushi 2)

50kNm TORQUE STANDARD MACHINE

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Comparison between the CENAM (Mexico) 150 kn and the INRiM (Italy) 1 MN force standard machines

Reliable Test Results

Analysis of shock force measurements for the model based dynamic calibration

MEASUREMENT UNCERTAINTY EVALUATION OF TORQUE MEASUREMENTS IN NACELLE TEST BENCHES

Foundations of Ultraprecision Mechanism Design

NATIONAL CERTIFICATE (VOCATIONAL) APPLIED ENGINEERING TECHNOLOGY NQF LEVEL 4 NOVEMBER 2009

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Optical Method for Micro Force Measurement. Yusaku FUJII Gunma University

L13 Structural Engineering Laboratory

Glossary Innovative Measurement Solutions

The calibration system of force measurement devices - conceptions and principles

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Torque Measuring Devices

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

MODEL-BASED ANALYSIS OF THE DYNAMIC BEHAVIOUR OF A 250 KN SHOCK FORCE CALIBRATION DEVICE

T e c h n i c a l L e x i c o n

3 Hours/100 Marks Seat No.

INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3

TE 75R RESEARCH RUBBER FRICTION TEST MACHINE

When a rigid body is in equilibrium, both the resultant force and the resultant couple must be zero.

T20WN. Data Sheet. Torque transducers. Special features. Installation example with bellows couplings. B en

Chapter 8. Centripetal Force and The Law of Gravity

ENGINEERING MECHANICS: STATICS AND DYNAMICS

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

INVESTIGATION OF IMPACT HAMMER CALIBRATIONS

Key words: Polymeric Composite Bearing, Clearance, FEM

Project PAJ2 Dynamic Performance of Adhesively Bonded Joints. Report No. 3 August Proposed Draft for the Revision of ISO

Rotating Bending with Constant Torsion and Rotated Bending with Constant or Variable Torsion

Messunsicherheit des Drehmomentes in multi-mw Windenergieanlagen Prüfständen. Torque measurement uncertainty in multi-mw nacelle test benches

NR ROTARY RING TABLE: FLEXIBLE IN EVERY RESPECT

Appendix. Vectors, Systems of Equations

Calibration Traceability Guidelines

Model parameter identification from measurement data as a prerequisite for dynamic torque calibration Measurement results and validation

APPLICATION OF A SCANNING VIBROMETER FOR THE PERIODIC CALIBRATION OF FORCE TRANSDUCERS

The University of Melbourne Engineering Mechanics

MECHANICAL CHARACTERISTICS OF STARCH BASED ELECTRORHEOLOGICAL FLUIDS

Flywheels-Function need and Operation

Investigations for the model based dynamic calibration of force transducers by using shock excitation

SAMCEF For ROTORS. Chapter 1 : Physical Aspects of rotor dynamics. This document is the property of SAMTECH S.A. MEF A, Page 1

1.103 CIVIL ENGINEERING MATERIALS LABORATORY (1-2-3) Dr. J.T. Germaine Spring 2004 LABORATORY ASSIGNMENT NUMBER 6

M.S Comprehensive Examination Analysis

HELICAL BUCKLING OF DRILL-STRINGS

Final Exam April 30, 2013

Torque. Physics 6A. Prepared by Vince Zaccone For Campus Learning Assistance Services at UCSB

An Analysis Technique for Vibration Reduction of Motor Pump

General Physics I Lab. M1 The Atwood Machine

Study of Rupture Directivity in a Foam Rubber Physical Model

CWR track vibration characteristics varying with the change of supporting condition

Contents. Dynamics and control of mechanical systems. Focus on

D : SOLID MECHANICS. Q. 1 Q. 9 carry one mark each. Q.1 Find the force (in kn) in the member BH of the truss shown.

Shape Optimization of Oldham Coupling in Scroll Compressor

Backlash-free Torque Limiting Clutches

CHAPTER 6 FRICTION AND WEAR ANALYSIS FOR BUSHING

Torque. Objectives. Assessment. Assessment. Equations. Physics terms 6/2/14

Levers. 558 A Textbook of Machine Design

Mechanical System Elements

Module 4 : Deflection of Structures Lecture 4 : Strain Energy Method

Selection Calculations For Linear & Rotary Actuators

Stress Analysis Lecture 4 ME 276 Spring Dr./ Ahmed Mohamed Nagib Elmekawy

TUTORIAL SHEET 1. magnitude of P and the values of ø and θ. Ans: ø =74 0 and θ= 53 0

two structural analysis (statics & mechanics) APPLIED ACHITECTURAL STRUCTURES: DR. ANNE NICHOLS SPRING 2017 lecture STRUCTURAL ANALYSIS AND SYSTEMS

MECHANICS OF MATERIALS. Prepared by Engr. John Paul Timola

ME 202 STRENGTH OF MATERIALS SPRING 2014 HOMEWORK 4 SOLUTIONS

Coupling System for Ultra Precision Machining

Question Bank Chapter -4- Part -2-

SOLUTION (17.3) Known: A simply supported steel shaft is connected to an electric motor with a flexible coupling.

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Development of Distributed Optical Torque Sensors for Realization of Local Impedance Control of the Robot Arm

DESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE

Metrology Prof. Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere

ANALYSIS AND IDENTIFICATION IN ROTOR-BEARING SYSTEMS

Strain Measurement Techniques for Composite Coupon Testing

Measurement Uncertainty in Mechanical Testing

Random Eigenvalue Problems in Structural Dynamics: An Experimental Investigation

PLEASURE VESSEL VIBRATION AND NOISE FINITE ELEMENT ANALYSIS

Transcription:

UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT Dieter Kenzler 1, Jörg Oblasser 2, Andreas Subaric-Leitis 2 and Christian Ullner 2 1 ) SCHATZ AG, 42897 Remscheid, Germany 2 ) Federal Institute for Materials Research and Testing (BAM), 12000 Berlin, Germany Abstract Calibration machines up to a torque range of 20 kn m need extensive design if the force is performed by dead weights. A machine based on two force reference standards and a vertical torque axis may serve to avoid these specific problems. The precision force transducers work at a lever arm of precision length. The test piece is mounted in vertical arrangement between a hydraulically operating rotary-actuator and the torque measuring system. Such a calibration machine allows additionally performing easily continuous calibration. Regarding an overall relative uncertainty not greater than 2 10-4 there are some significant problems. The estimation of the uncertainty has been proved by comparative tests with the national torque standard by the PTB according to the guidelines EA-10/14 and EA-2/03. Boundary conditions The best measuring capability is given by a system with a lever-arm (beam) with direct effect of mass and the influence of gravity and earth-acceleration (Figure 1). To compare this type of calibration machine with the introduced type we will discuss the most important points: Figure 1: Principially Torque Primary Standard

The turning point for the lever-arm (beam) is realized with an air-bearing to reduce the friction to a minimum value. The test measuring results carried out in the past 10 years gives a friction value of < 1 10-7 (0.00001%). This type of calibration machine is called supported beam. The measuring axis is horizontal. Every of the six measuring ranges are divided into ten disk-weights manipulated step by step like a chain. The uncertainty of the disk-weights is given by the German Gauge Office (Eichamt) within a tolerance of ± 0.0005 %. The point of coupling according to the length is corrected by the factor of gravity and earthcentrifugal force, the lift component of air and an ambient temperature of 21 ± 0.5 C with a length of 1019.384 mm of both sides and an uncertainty of < 5 µm. For the force application to the lever a thin metal belt (foil) with a thickness of 20 µm is used. The gravity with the influence of the earth-centrifugal force is measured traceable to the German gravity network with a value of 9.8112401 m s -2 with a tolerance of < 5 10-8 m s -2. The time-variation of the gravity as a result of influence of the gravity of sun and moon is about 3 10-7 m s -2. In the measuring range between 1 N m and 1000 N m, all significant torque values can be realized with the torque values being constantly increased or decreased. This is achieved by selecting different mass echelons for clockwise torque and counterclockwise torque. When both loading systems are used at the same time, the respective actual torque results from the difference between clockwise and counterclockwise torque. To build such a type of calibration device up to a torque of 20 kn m need an extensive and very expensive design concerning the air-bearing and the dead weights. The other idea is to build a machine based on two force reference standards and a vertical torque axis may serve to avoid these specific problems (Figure 2). 1 Precision tension force transducers 2 Lever-arm of precision length 3 Test piece (for instance torque reference transducer) 4 Adaptation 5 Hydraulically operated rotary-actuator 6 Oil-hydraulic pump 7 Directional setting valve 1 Figure 2: Simplified function-diagram of the 20 kn m-calibration device A pair of precision force transducers is fixed to the frame and work at both ends of a lever-arm of precision length. For the force application to the lever a thin metal belt is used. The test piece is

mounted in vertical arrangement between a hydraulically operating rotary-actuator and the leverarm. This rotary-actuator generates torque up to 20 kn m with an oil pressure up to 21 MPa and a rotating angle up to ±45. The transducer under test with the lever-arm and the coupled force transducer give a virtual point of torque axis. That causes no bearing is necessary. This system is called unsupported beam. Such a calibration machine allows additionally performing easily continuous calibration for shorter and simplified testing times. Possibilities and influence analysis of errors The most important detail is the stability of the frame to minimize twist or torsion. To reduce bending moments and side forces an exactly adjustment of the pair of force is required for a symmetrical load to the frame and to produce a precise virtually axis of torque. Misalignment by concrete behaviour of the transducer is connected with missing gradation of freedom. Geometrical uncertainty in the setting of structural precision e.g. the introduction of the force is not parallel and not vertical to the torque axis. The thin metal belt for the introduction of force is not close fitting to the end of the lever. That causes an uncertain tangential introduction of force to the lever. The axial introduction of force to the force standard transducers may be not ideal. An additional uncertainty contribution arises from axially moving the actuator. With increasing torque the results can be influenced by the superposed bending. Reading errors caused by polling problems of the measuring amplifier, in the case of a 2- channel measuring system, for example DMP40, the measuring of the third channel is possible only by multiplexing the channels. This causes errors in case of fast pass through of measurement, especially by using continuous calibration. Results of comparison measurements In the last 2 years there have been measurements within international comparison. The results have been compared with the measurements performed in the calibration laboratory of SCHATZ. Especially within the test measurements the possible faults mentioned in the section possibilities and influence analysis of errors have been erased consciously to determine their influence on the measuring results. The results showed relatively high fault shares especially in the lower sections of the corresponding measuring ranges. Especially high non-systematic faults appeared during fast changing of measuring values as common at the continuous calibration.

rel. interpolation deviation of reading rel. deviation in % 0,12 0,10 0,08 0,06 0,04 0,02 0,00 Linear Square Cubic -0,02 0 5 10 15 20 25 kn m Figure 3: Typically calibration result of a 20 kn m torque transfer standard transducer with an expanded uncertainty of about 0.01%. The calibration procedure is according to the guideline EA-10/14. PTB - BAM 01/00 comparison 20 kn m 0,40 0,35 rel. deviation in % 0,30 0,25 0,20 0,15 0,10 0,05 0,00 0 5 10 15 20 25 kn m dev up dev down Figure 4: Result of the comparison to the national torque standard at the PTB.

The mathematical formulas to the calculation of compares the uncertainty are according to the paper Uncertainty Scope of the Force Calibration Machines, A. Sawla [4]: M TCM = M RefTra * (1 - Drift_RefTra ) * (1 - c ϑ ) * (1 - Connect ) (1) = M RefTra * (1 - Drift_RefTra ) * (1 - c ϑ ) * ( 2 - [M mid,tcm * (1 - RelDev ) * (1 - HysTCM ) / M mid,tcm ] - Drift_TranTra - Realization (2) Reduction of parasitic influences for lower uncertainty As before reported the estimated uncertainty in calibration of torque sensors is given. The sources of the uncertainty shall be discussed with regard to the method of reduction and to improve the design to reduce bending and bending moments and the reading errors. The reduction of bending moments is possible by using coupling elements which are highly flexible for bending but very stiff for torsion e.g. flexible coupling in exchange for the adaptation (Fig. 2 index 4). This solution is comprehensive discussed in [3]. To supervise the centre of the torque axis a measuring system shall be used. Additional the forces must be applied in such a way that the high precision of the lever length is preserved during the complete test. The close fitting of the thin metal belt shall be guarantee by using a pair of clamping springs. To reduce the reading errors caused by polling problems of the measuring amplifier DMP40 an exchange to a system with three measuring channels with absolute synchronically data acquisition e.g. MGCplus with three ML38 is necessary. This system allows in combination with qualified software a continuous data acquisition including online-interpolation. On the other hand the design of the calibration machine makes it possible to investigate the effect of superposed bending on the torque sensor using a dead weight on the lever-arm. The test results can be used for correcting the calibration results. General estimation Experiments with the introduced calibration device have shown, that in case of using appropriate force transducers, flexible coupling with hydraulic clampings and a three-channel synchronically working electronic precision data acquisition system, a calibration device with a measuring range from 500 N m up to 20 kn m only has to be equipped with 6 pairs of force reference transducers to reach an expanded relative uncertainty of U < 5 10-4. If you compare prices for calibration devices you will see that prices for calibration systems working with the "continuous comparison method" are much lower than for systems working with "absolute calibration". If this calibration method is chosen one should consider, that it is much more appropriate for most torque measuring instruments as the step-by-step method regarding the practical application.

Supported by special software - including interfaces for suitable data recording systems - it is also possible to include calibrations of measuring chains. In some special cases it might be impossible to embed the transducer to be tested into the automated data determination process. Then it is also possible to do a step-by-step calibration if the control of the drive system is prepared correspondingly. Conclusions The consideration to do a calibration following the comparison method at this machine makes possible an optimized procedure regarding time together with a most favorable load procedure concerning the application. This procedure also offers a load process which can be varied in wide scopes. Thus even exotic measuring ranges of transducers can be calibrated without problems because discrete steps are not necessary for the calibration. The ranges of steps can be edited easily via software. The described procedure should be used and offers not only high cost savings but also precise measuring results together with a fully automated procedure. References [1] Kenzler, D.: Application of the Continuous Calibration Procedure for Industrial Standard Calibration of Torque- and Force-Transducers ; IMEKO TC3 1998, Taejon, Korea [2] Oblasser, J.: BAM entwickelt Drehmoment-Kalibrierstand, Materialprüfung 32 (1990) 3,47 [3] Brüge, A., Peschel, D., Röske, D.: The influence of misalignment on torque transducers ; IMEKO 2001, Vol. 3, Paper 25 [4] Sawla, A.: Uncertainty Scope of the Force Calibration Machines, IMEKO 2000 Contact Point Dieter Kenzler, SCHATZ AG, Calibration Laboratory Koelner Str. 71, D-42897 Remscheid, Germany Phone: +49 2191 698 ext 287 Fax: +49 2191 60023 E-mail: dkenzler@t-online.de