high temp ( K) Chapter 20: Atomic Spectroscopy

Similar documents
CH. 21 Atomic Spectroscopy

very high temperature for excitation not necessary generally no plasma/arc/spark AAS

Optical Atomic Spectroscopy

Emission spectrum of H

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

9/13/10. Each spectral line is characteristic of an individual energy transition

2101 Atomic Spectroscopy

Chemistry Instrumental Analysis Lecture 18. Chem 4631

Partial Energy Level Diagrams

Ch. 8 Introduction to Optical Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy

Atomic Absorption & Atomic Fluorescence Spectrometry

Atomization. In Flame Emission

Chapter 8: An Introduction to Optical Atomic Spectrometry

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

3 - Atomic Absorption Spectroscopy

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

a. An emission line as close as possible to the analyte resonance line

ATOMIC SPECROSCOPY (AS)

Atomic Absorption Spectroscopy (AAS)

INTRODUCTION TO OPTICAL ATOMIC SPECTROSCOPY (Chapter 8)

PRINCIPLES OF AAS atomization flame furnace atomization absorption

Chemistry Instrumental Analysis Lecture 17. Chem 4631

CHAPTER 8 Introduction to Optical Atomic Spectrometry

Atomic Emission Spectroscopy

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Spectroscopy Problem Set February 22, 2018

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

II. Spectrophotometry (Chapters 17, 19, 20)

Atomic Spectroscopy AA/ICP/ICPMS:

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Chapter 8: Introduction to Atomic Spectrometry

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS

1. Cyclic voltammetry involves the measurement of a diffusion controlled at an electrode in which the is controlled. (4 points)

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Ionization Techniques Part IV

Chem 434 -Instrumental Analysis Hour Exam 1

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Questions on Instrumental Methods of Analysis

Chem 155 Quiz 3 Review Topics: Quiz 3 outline

Spectral Resolution. Spectral resolution is a measure of the ability to separate nearby features in wavelength space.

Electron temperature is the temperature that describes, through Maxwell's law, the kinetic energy distribution of the free electrons.

Sample Analysis Design Polyatomic Interferences

Atomic Absorption Spectroscopy

Chem 434 Instrumental Analysis Test 1

PRINCIPLE OF ICP- AES

Ch 313 FINAL EXAM OUTLINE Spring 2010

FLAME PHOTOMETRY AIM INTRODUCTION

A Spectrophotometric Analysis of Calcium in Cereal

Chemistry Instrumental Analysis Lecture 34. Chem 4631

25 Instruments for Optical Spectrometry

Final Exam. Physical Constants and Conversion Factors. Equations

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Lecture 10. Lidar Effective Cross-Section vs. Convolution

Characterization of Catalysts and Surfaces. Elemental Analysis (ICP, AAS etc.) Fall Semester 2016 Bodo Hattendorf HCI G105

OES - Optical Emission Spectrometer 2000

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Chapter 5 Light and Matter

Spectrometric Methods of Analysis. OCN 633 Fall 2013

Atomic absorption spectroscopy

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Chem Homework Set Answers

Chapter 4. Spectroscopy. Dr. Tariq Al-Abdullah

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Molecular spectroscopy

Sources of Errors in Trace Element and Speciation Analysis

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Lecture No. 4. i 0 TETA I RY:

Fundamentals of Spectroscopy for Optical Remote Sensing. Course Outline 2009

Introduction to Elemental Analysis

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Extrel Application Note

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Ultraviolet-Visible and Infrared Spectrophotometry

Prof. Dr. Biljana Škrbić, Jelena Živančev

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Example: model a star using a two layer model: Radiation starts from the inner layer as blackbody radiation at temperature T in. T out.

Spectral Line Shapes. Line Contributions

Trace elemental analysis solutions for your application. June 6, 2018

A study of the predissociation of NaK molecules in the 6 1 state by optical optical double resonance spectroscopy

Sample Copyright. Academic Group ATOMIC STRUCTURE 1. Topics covered in this chapter:

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

HEAVY METALS IN LAKE NAKURU ATOMIC SPECTROSCOPY

Comments to Atkins: Physical chemistry, 7th edition.

Secondary Ion Mass Spectrometry (SIMS)

Laser Dissociation of Protonated PAHs

Introduction to Plasma

Exam 3 Chem 454 April 25, 2018, Name

Chemistry 311: Topic 2 - Atomic Spectroscopy Topic 2: Spectroscopy:

The Fundamentals of Spectroscopy: Theory BUILDING BETTER SCIENCE AGILENT AND YOU

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Chem 155 Midterm Exam Page 1 of 10 Spring 2010 Terrill

Transcription:

high temp (2000-6000K) Chapter 20: Atomic Spectroscopy

20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS)

Atomic Absorption experiment Flame in Atomic Spec. Cuvette in Mol. Spec. (Path length in Flame: 10 cm)

Atomic Fluorescence experiment Atomic fluorescence from Pb at 405.8 nm Atomization: 6 ns pulse of 1064 nm laser radiation

Atomic Emission experiment Bandwidth of radiation that is absorbed or emitted: Molecular spectroscopy: ~100 nm Atomic spectroscopy: ~0.001 nm (very sharp line spectra little overlap between the spectra of different elements can measure multi-elements simultaneously; ~70 elements) Fig.21-3. A portion of emission spectrum

20-2. Atomization: Flames, Furnaces, and Plasmas Atomizer :converts a sample to an atomic vapor Atomization: Flames (AAS, AES; liquid sample) Furnaces (AAS, AES; liquid & solid sample) Plasma (AES, ICP-MS; liquid sample) The precision and accuracy of AS are critically dependent upon atomization step & sample introduction method

Flames Premix burner: fuel, oxidant, and sample are premixed. Nebulization: formation of a small droplets Aerosol: a fine suspension of liquid (solid) particles in a gas Nebulizer: create an aerosol from the liquid sample aerosol reaching the flame contains only about 5% of initial sample Pneumatic nebulizer

Flames most common For refractory elements Refractory elements: high-boiling elements

Furnaces An electrically heated graphite furnace - offers greater sensitivity than that provided by flames - requires less sample (1-100 L for furnace, 1-2 ml for flames)

Furnaces

Inductively Coupled Plasma -Exclusively for emission (not absorption) or mass spec. -6000-10000K in ICP (higher and more stable temp than flame) -Telsa coil ionize Ar gas Free electron is accelerated and oscillated in rf induction coil -Accelerated free electrons collide with Ar atom and transfer energy -> Joule Heating (ir 2 )

20-3. How Temperature Affects Atomic Spectroscopy Temp is a critical factor in determining - the degree to which a given sample breaks down to atoms (atomization efficiency) - the extent to which a given atom is found in its ground state, excited state or ionized states Boltzman Distribution -describes the relative populations of different states at thermal equilibrium -the relative population of any two states is N*/N o = (g*/g o ) exp(-δe/kt) E* Eo ΔE g* = 3, excited states g o = 2, ground states g: degeneracy (the number of states available at each energy level) T; temperature, k; Boltzman constant(1.31 X 10-23 J/K)

Effect of Temp on Excited-State Population Sodium atom at 2600K (in acetylene-air flame) N*/N o = (g*/g o ) exp(- ΔE/kT) ΔE = 3.371 x 10-19 J/atom g o = 1, g* = 2 Thus, N*/No = (2/1)exp[(-3.371 x 10-19 )/(1.381 x 10-23 J/K)(2600K)] = 1.67 x 10-4 0.0167% of the atoms are in the excited state (99.9833% in ground state) If the temperature is 2610K (from 2600K) N*/No = (2/1)exp[(-3.371 x 10-19 )/(1.381 x 10-23 J/K)(2610K)] = 1.74 x 10-4 0.0174% of the atoms are in the excited state (99.9826% in ground state) [(1.74-1.67) x 100%]/1.67 = 4% increase in the population of the excited state [99.9833-99.9826) x 100%]/99.9833 = 0.07% decrease in the population of the ground state

Effect of Temp on Absorption and Emission Sodium atom at 2600K 0.0167% of the atoms are in the excited state (99.9833% in ground state) If the temperature is 2610K (from 2600K) 0.0174% of the atoms are in the excited state (99.9826% in ground state) 10-K rise in temperature: Almost no change (0.07% decrease) in the ground-state population: No noticeable change in an atomic absorption 4% increase in the excited-state population: Emission intensity rise by 4% In atomic emission spectroscopy, the flame temperature should be kept very stable (ICP: high temp high sensitivity, stable plasma temp reproducible data)

20-4. Instrumentation Atomic Linewidths In atomic spec, absorption and emission lines are very sharp Three sources of line broadening: (1) Uncertainty effect: ~ 10-4 nm (2) Doppler effect: ~ 10-2 nm 10-3 nm (3) Pressure effect (collision broadening): ~ 10-2 10-3 nm Uncertainty broadening (natural line width) Heisenberg uncertainty principle: the shorter the lifetime of the excited state, the more uncertain in its energy relative to the ground state δe δt h/2π δe; the uncertainty in the energy difference between the ground and excited states δt; life time of the excited state before it decays to the ground sate (~ ns in AS) h; Plank s constant

Atomic Linewidths Doppler broadening Atoms in hot flame: atomic motions in every direction causes Doppler broadening Doppler linewidth ; δ = (7 x10-7 ) (T/M) 1/2 T; temperature, M; atomic mass A rapidly moving atom in flames If the motion is toward a photon detector, the wavelength of radiation emitted or absorbed: decreases (high frequency) If the atom is receding from the photon detector, the wavelength of radiation emitted or absorbed: increase (low frequency)

Atomic Linewidths Doppler broadening For an emission near = 300 nm from Fe (M = 56) at 2500K Doppler linewidth ; δ = (7 x10-7 ) (T/M) 1/2 = (300 nm) (7 x10-7 ) (2500K/56) 1/2 = 0.0014 nm Pressure broadening An order of magnitude greater than the natural linewidth Collisions of the absorbing or emitting atoms with other atoms or ions in heating medium shorten the lifetime of the excited state or causes the change in the energy level of the ground and excited states. line broadening High pressure higher collisional frequency greater line broadening

Hollow-Cathode Lamp in AAS Monochromators cannot isolate lines narrower than 10-3 10-2 nm. To get narrow lines of the correct frequency, Use of hollow cathode lamp containing the same element that being analyzed Filled with Ne or Ar at a pressure of 130 ~ 700 Pa High voltage (~300V) is applied between the anode and cathode Filler gas is ionized and positive ions are accelerated toward the cathode Accelerated positive ions strike the cathode with enough energy to sputter metal atoms from the cathode into the gas phase Free atoms are excited by collisions with high-energy electrons: photon emission Atomic radiation has the same frequency as that absorbed by the analyte atoms

Hollow-Cathode Lamp in AAS - The linewidth in hollow cathode lamp is sufficiently narrow Atoms in lamp are cooler than atoms in a flame lamp emission is narrower than the absorption bandwidth in the flame - A different lamp is required for each element - Some lamps are made with more than one element in the cathode - The purpose of a monochromator in atomic spectroscopy: 1) to select one line from hollow cathode lamp (usually filter) 2) to reject as much emission from the flame or furnace as possible

1 2 3 1,2,3: removed by filter or monochromator

Multielement Detection in ICP-AES ~ 70 elements can be detected simultaneously

Echelle Monochromator Two dispersion elements arranged in series : higher dispersion and resolution than an echellette of the same size or grating (low dispersion) Used in atomic emission spectroscopy

Color Art Insert Color Plate 23 Polychromator for ICP-Emission with one detector for each element

Color Art Insert Color Plate 24 Polychromator for ICP-Emission with one detector for all element

Background Correction Background signals arises from absorption, emission, or scattering from matrix as well as flame, plasma or furnace Graphite furnace absorption spectrum of bronze dissolved in HNO3 analyte singnal (1.0) background signal (0.3) Background correction is necessary

Background Correction Beam chopping method (or source modulation): eliminate flame emission Atomic absorption + Flame emission (signal a) Only flame emission (signal b) Signal a signal b = analyte signal Source modulation: the power supply of the source is designed for ac operation so that the source is switched on and off at the desired constant frequency

- The detection limits for furnaces is two orders of magnitude lower (thus, better) than that observed with a flame. - The detection limits for ICP are intermediate between flame and furnace.

20-5. Interference in AS Interference is any effect that changes the signal while analyte concentration remains unchanged. Types of interference: (1) Spectral interference: overlap of analytical signal with unwanted signals due to other elements in the sample/ or due to flames -Cd line causes spectral interference with the As line in most spectrometers. -Two peaks are separated with 1-m Czerny-Turner monochromator with a resolution of 0.005 nm from 160-320 nm and 0.01 nm from 320-800 nm

(1) Spectral interference: Problem solving: Overlap between lines of different elements in the sample: choose another wavelength Or Use of high-resolution spectrometers Elements that form very stable diatomic oxides are incompletely atomized at the temp of the flame and furnace: much broader and more complex spectrum than that of an atom

Types of interference: (2) Chemical interference: chemical reactions decreasing the extent of atomization of analytes Example: SO 4 2- and PO 4 3- hinders the atomization of Ca 2+ by forming nonvolatile salts. Solution: Add releasing agent (protective agent) - EDTA, 8-hydroxyquinoline: forms stable but volatile species with Ca 2+ -La 3+ : preferentially reacts with sulfate or phosphate and frees Ca 2+

(3) Ionization interference: - ionization of analyte atoms decreases the concentration of neutral atoms - energy levels of ionized atoms are different from those of neutral atoms <gas phase ionization reaction> M(g) M + (g) + e - (g) - a problem in the analysis of alkali metals even at relatively low flame temperature (alkali metal: low ionization potential) at 2450K and 0.1 Pa, Na: 5% ionized, K: 33% ionized Solution: add an ionization suppressor to decrease the extent of ionization of analyte Example: for the analysis of Sr, add K (more easily ionized) K K + + e- Sr Sr + + e- --- I.P (K) = 4.3 ev --- I.P (Sr) = 5.7 ev providing high concentration of electrons to the flame (from K + + e - ) suppress the ionization of the analyte (Sr)

Virtues of Inductively Coupled Plasma 1. ICP is twice as hot as a conventional flame 2. The residence time of analyte in the plasma is twice as long Atomization is more complete Signal is enhanced Formation of oxides and hydroxides are negligible 3. ICP is free of background radiation 4. The temperature in ICP is more uniform almost no self-absorption

20-6. ICP-Mass Spectrometry (negative: attract positive ions) (water cooling) Vacuum to avoid collisions between ions and gas molecules that divert the ions from their trajectory in a field Isobaric interference: Interference by ions of similar mass e.g. 40 Ar 16 O + = 56 Fe +, 40 Ar 2+ = 80 Se + Use of collision cell: dissociate ArO +, Ar 2+, ArCl +, -- Or high resolution mass

ICP-MS - Most important techniques for elemental analysis - Mass range: 3-300 (recently up to 4000 m/z) - Ability to resolve ions differing in m/z by 1-90% elements in periodic table can be determined - Detection limits; 0.1 10 ppb - RSD: 2 ~ 4% - Dynamic range: 6 orders - Fast analysis: 10 s for 1 elements

Laser Ablation-ICP-MS

Laser Ablation-ICP-MS