A Fixed Surface Potential Probe with the Swing Capacitive Electrometer Compared to the Vibrating Kelvin Probe

Similar documents
Trek electrostatic voltmeters Setup, environment, working conditions

Core Technology Group Application Note 3 AN-3

Institute for Electron Microscopy and Nanoanalysis Graz Centre for Electron Microscopy

Measurement of hardness, surface potential, and charge distribution with dynamic contact mode electrostatic force microscope

Carbonized Electrospun Nanofiber Sheets for Thermophones

Integrating MEMS Electro-Static Driven Micro-Probe and Laser Doppler Vibrometer for Non-Contact Vibration Mode SPM System Design

Active elastomer components based on dielectric elastomers

CHAPTER 5 FIXED GUIDED BEAM ANALYSIS

(a) (b) Supplementary Figure 1. (a) (b) (a) Supplementary Figure 2. (a) (b) (c) (d) (e)

HIGH-PERFORMANCE PERFLUORINATED POLYMER ELECTRET FILM FOR MICRO POWER GENERATION

Experimental Validation of Damping Model for a MEMS Bistable Electrostatic Energy Harvester

Vacuum Kelvin Force Probe Research Richard Williams August 1st 2008

DYNAMIC ROTARY TORQUE MEASUREMENT USING SURFACE ACOUSTIC WAVES

Piezoelectric Resonators ME 2082

Capacitors. Charging a Capacitor. Charge and Capacitance. L05: Capacitors and Inductors

LOW-COST CAPACITIVE PERSONNEL DETECTOR

Supplementary Figures

Electromagnetics in COMSOL Multiphysics is extended by add-on Modules

Chapt ha e pt r e r 9 Capacitors

MEMS INERTIAL POWER GENERATORS FOR BIOMEDICAL APPLICATIONS

Electrostatic Microgenerators

EIS of Organic Coatings and Paints

Reduced Order Modeling Enables System Level Simulation of a MEMS Piezoelectric Energy Harvester with a Self-Supplied SSHI-Scheme

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

ERROR SOURCE IDENTIFICATION AND STABILITY TEST OF A PRECISION CAPACITANCE MEASUREMENT SYSTEM

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2009 PROBLEM SET #7. Due (at 7 p.m.): Thursday, Dec. 10, 2009, in the EE C245 HW box in 240 Cory.

Development and Characterization of High Frequency Bulk Mode Resonators

CST EM : Examples. Chang-Kyun PARK (Ph. D. St.) Thin Films & Devices (TFD) Lab.

Supplementary Materials to Addressable Photo-Charging of Single Quantum Dots Assisted with Atomic Force Microscopy Probe

MOS CAPACITOR AND MOSFET

Reactor Characteristic Evaluation and Analysis Technologies of JFE Steel

Inductive Versus Capacitive Position Sensors

Nonstationary electrical charge distribution on the fused silica bifilar pendulum and its effect on the mechanical Q-factor

Charge spectrometry with a strongly coupled superconducting single-electron transistor

The interfacial study on the Cu 2 O/Ga 2 O 3 /AZO/TiO 2 photocathode for water splitting fabricated by pulsed laser deposition

CIRCUIT ELEMENT: CAPACITOR

Electrochemistry of Semiconductors

ME 237: Mechanics of Microsystems : Lecture. Modeling Squeeze Film Effects in MEMS

MEMS design and fabrication of an electrostatic vibration-to-electricity energy converter

MOS Transistors. Prof. Krishna Saraswat. Department of Electrical Engineering Stanford University Stanford, CA

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2011 C. Nguyen PROBLEM SET #7. Table 1: Gyroscope Modeling Parameters

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY

Lecture 23: Negative Resistance Osc, Differential Osc, and VCOs

MAGNETIC FIELDS & UNIFORM PLANE WAVES

Analysis of Geometrical Aspects of a Kelvin Probe

DAMPING CONTROL OF A PZT MULTILAYER VIBRATION USING NEGATIVE IMPEDANCE CIRCUIT

NorthStar Battery Company DCN: SES DCR: 1038-S07 Date: Page 1 of 12

Light, Electricity and Liquids. By Vitaliy Zamsha and Vladimir Shevtsov

THERMOCOUPLE CHARACTERISTICS TRAINER

Supplementary Figure 1: SAW transducer equivalent circuit

Biosensors and Instrumentation: Tutorial 2

DESIGN AND FABRICATION OF THE MICRO- ACCELEROMETER USING PIEZOELECTRIC THIN FILMS

Modelling of Different MEMS Pressure Sensors using COMSOL Multiphysics

Large, light-induced capacitance enhancement in semiconductor junctions simulated by capacitor-resistor nets

Section 12: Intro to Devices

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

3/24/11. Introduction! Electrogenic cell

CONTENTS. by J.S.S. Rao by Dr. M.L. Kothari. xv xvii. xxiii 1

ECE2262 Electric Circuits. Chapter 6: Capacitance and Inductance

Metallized Polypropylene Film Capacitor Related Document: IEC

Switching Current Study: Hysteresis Measurement of Ferroelectric Capacitors using Current-Voltage Measurement Method

THICK-FILM THERMALLY EXCITED RESONATOR FOR MASS FLOW MEASUREMENT. D. Crescini, D. Marioli and A. Taroni

Electrochemical methods : Fundamentals and Applications

Micromechanical Instruments for Ferromagnetic Measurements

Objective: Competitive Low-Cost Thin-Film Varactor Technology. Integrated Monolithic Capacitors using Sputtered/MOCVD material on low-cost substrates

Technology Document Version

A Novel Tunable Dual-Band Bandstop Filter (DBBSF) Using BST Capacitors and Tuning Diode

Properties of Materials

Bridge Measurement 2.1 INTRODUCTION Advantages of Bridge Circuit

Analytical Optimization of High Performance and High Quality Factor MEMS Spiral Inductor

Shielding Tutorial Gentex EME Lab

Supplementary Information

Keywords. 1=magnetron sputtering, 2= rotatable cathodes, 3=substrate temperature, 4=anode. Abstract

Elizabethtown College Department of Physics and Engineering PHY104

Supplementary Figure 1. Theoretical calculation results to optimize the FEP layer thickness

EE 4345 Chapter 6. Derek Johnson Michael Hasni Rex Reeves Michael Custer

OR Explain thermal breakdown in solid dielectrics. How this mechanism is

Fitting PS P - B - R2

LOW FREQUENCY NOISE OF TANTALUM CAPACITORS*

DESIGN AND FABRICATION OF A MICRO ELECTROSTATIC VIBRATION-TO-ELECTRICITY ENERGY CONVERTER

Supporting Information: Poly(dimethylsiloxane) Stamp Coated with a. Low-Surface-Energy, Diffusion-Blocking,

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Applicability of Self-Powered Synchronized Electric Charge Extraction (SECE) Circuit for Piezoelectric Energy Harvesting

Final Exam: 11 May 2001, Friday

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

Performance Evaluation of MEMS Based Capacitive Pressure Sensor for Hearing Aid Application

Design and fabrication of multi-dimensional RF MEMS variable capacitors

Novel Devices and Circuits for Computing

Electrical Characterization of 3D Through-Silicon-Vias

Low Frequency Noise in MoS 2 Negative Capacitance Field-effect Transistor

Electromechanical Sensors and Actuators Fall Term

Sensor Measurements For Diagnostic Equipment

Lecture 4 Scanning Probe Microscopy (SPM)

Microwave Loss Reduction in Cryogenically Cooled Conductors R. Finger and A. R. Kerr

AC Circuits Homework Set

The Basic Capacitor. Dielectric. Conductors

Microstructure cantilever beam for current measurement

Today s SPM in Nanotechnology

Capacitive Generation and Detection of Third Sound Resonances in Saturated Superfluid 4He Films

Tunable MEMS Capacitor for RF Applications

Transcription:

Proc. 2017 Annual Meeting of the Electrostatics of America 1 A Fixed Surface Potential Probe with the Swing Capacitive Electrometer Compared to the Vibrating Kelvin Probe Michael Reznikov Dept. of Integrated Systems Physical Optics Corp., Torrance, CA phone: (1) 310-320-3088 e-mail: mreznikov@poc.com Abstract The novel fixed (non-vibrating) Volta potential probe was evaluated in comparison with the standard, vibrating (400 Hz) commercial Kelvin probe. While both tested probes demonstrated the similar behavior, there were significant practical advantages of the novel probe. Specifically, the novel probe allows for the elevated distance to the surface and provides more sensitivity to charge under the coating layer. These differences are explained by the analysis of probe circuitry together with the electrostatic approach to the dynamics of the Volta potential measurement. I. INTRODUCTION The capacitive probe method for the measurement of charge induced in the probe electrode was proposed by Lord Kelvin (W. Thomson) in 1898 as a capacitive quadrant electrometer switching between the immovable probe and discharge shunt while the biasing voltage is applied to the material specimen [1] (see Fig. 1(a)). The using of biasing voltage and thus the "zero signal" measurement were implemented because the capacitance of the quadrant electrometer is varied during the measurement and thus the measured value depends on the capacitance between the probe electrode and the surface of the specimen. Due to the manual switching and adjustment of biasing voltage, the using of original Kelvin probe was very time-consuming.. In 1932 the vibrating probe method was developed by W.A. Zisman [2] that allowed the use of an electronic amplifier and became widely known as a Kelvin probe (for example, [3-5]) illustrated in Fig. 1(b). Either in the capacitive quadrant electrometer or in the vibrating probe, the capacitance between the probe and specimen is varied and the induced charge is redistributed between this capacitance and parasitic capacitances between the specimen and probe relatively to the ground. The significant nonlinearity of the measured signal is mitigated by the application of compensating bias to the sample (or sample substrate) to reduce the measured signal to zero. Such a "zero point" measurement introduces the challenge of precious adjustment of biasing voltage at the very small output signal at the realistic electromagnetic background noise. Because the measured signal is, in fact, the derivative of varied electric charge, the very small gap between the probe and the specimen is required to keep the corresponding capacitance larger than parasitic capacitances.

Proc. 2017 Annual Meeting of the Electrostatics of America 2 Fig. 1 Transformation of Kelvin Probe: (a) Original capacitive quadrant electrometer of W. Thomson; (b) Modern scanning Kelvin Probe (top - SKP5050 from KP Technology, UK, bottom - scheme of KP 6500 from McAllister Technical Services, USA); (c) Micromechanical electrometer [6]; (d) Vibrating capacitor XL7900, Philips; (e) Varactor based electric field sensor [8]. In some application [6], probe electrode is not vibrating and the variable capacitor is installed inside of the probe in parallel to the fixed capacitor (typically the parasitic capacitance to the ground) as shown in Fig. 1(c). In fact, it is the micromechanical implementation of vibrating capacitor vacuum device XL7900 from Philips [7] (see Fig. 1(d)). The pure solid state (no mechanical parts) electrometer was proposed by M. A. Noras [8] utilizing the electric field modulated semiconductor structure (a varactor or metalinsulator-semiconductor structure) as shown in Fig. 1(e). While allowing for the immobilized probe electrode, all these devices (but original Lord Kelvin's probe) are producing the alternating current through the specimen due to the redistribution of charge between the varied probe capacitor(s) and fixed capacitance between the specimen and probe, which are connected in serials. While this is not an issue with the conductive object (very short relaxation time), the problem arises if the specimen is coated with the material, which dielectric constant depends on the frequency thus the measured signal depends on the frequency of vibrations. The investigation of this concern is the subject of presented work. II. CONCEPT OF MEASUREMENT Independently of the nature of surface potential (either the electrochemical, Fermi potential of the sample material or the localized space charge on the surface), the electrical potential over the surface is defined by the electric field of the space charge. This electric field induces the charge in the probe electrode as illustrated in Fig. 2, where the substrate of the specimen may be grounded or linked to the ground through the capacitance Csg = C specimen-ground and the surface of the specimen may be coated or uncoated that affects the capacitance Cps = C probe - specimen.

Proc. 2017 Annual Meeting of the Electrostatics of America 3 Fig. 2 The concept of measurement. In the lump element presentation depicted in Fig. 2, all 3 capacitors (if present), C sg, C ps (either including the capacitance of coating layer or not), and the capacitance of probe transducer, C p, are connected in serials. As a result, the electric field of the space charge, which would be equal to both directions in case of the ideal charged surface in the infinite space, redistributed between capacitor Csg and capacitors Cps and Cp connected in serials thus the electric potential of specimen's substrate relatively to the ground is equal for both branches. For the localized charge of trapped carriers, there is always energy barrier (and correspondingly the electric potential difference) between the charge and the bulk material of the substrate. The similar barrier is formed even in the ideal conductor or semiconductor where any localized surface states are absent as illustrated in Fig. 3. Fig. 3. Illustration of space charge formed at the measurement of work function in metal (on left) and semiconductor (on right) Therefore, the simulation model should be based on the conservation of charge while accounting for the redistribution of the electric field (and correspondingly the induced charge) in the measurement circuit. Our concept of the device is presented in Fig. 4 where the probe transducer contains the compound capacitor built from two capacitors, which are varied in the opposite mode (swing capacitor). The main idea behind this concept is to keep the constant total capacitance of transducer thus there is no charge oscillation outside of the transducer and probe is measuring the steady state induced charge similarly to the original device of Lord Kelvin..

Proc. 2017 Annual Meeting of the Electrostatics of America 4 Fig. 4.The evaluated concept of the device. We have used the acoustic modulation to minimize the parasitic electromagnetic influence. Correspondingly, condenser microphones were used to build the swing capacitor. Because the modulated parameter is the distance between electrodes, i.e. the capacitance non-linearly depends on the excitation, the simulation of charge redistribution between the probe and specimen under the evaluation was performed with the Mathcad software. The capacitance of the non-exited microphone was set to 50 pf ( a typical value for good quality condenser microphone and the distance between electrodes was modulated from 13% to 87%. The capacitance between the probe and object was set to 1 pf that corresponds either to the very small tip of the probe (high resolution) or relatively far distance to the object. The modulation frequency was set to 12 khz. Results of this simulation are presented in Fig. 5. Fig. 5. Simulation of processes in the swing transducer: (a) Variation of total capacitance between the object and ground; (b) Oscillation of charge in branches of swing capacitor, Q1 and Q2, and the total charge in the transducer; (c) Currents between branches of swing capacitor and the measurable differential current (red curve). As Fig. 5 demonstrates, the variation of total capacitance does not exceed 1%, the total charge in the transducer is constant, and the current follows the modulation.

Proc. 2017 Annual Meeting of the Electrostatics of America 5 III. EXPERIMENTAL RESULTS. The Fixed Electrode Kelvin Probe (FEKP), which is schematically presented in Fig. 4, was tested in comparison with the commercial Kelvin probe, CKP (KP 6500 from McAllister Technical Services). The FEKP transducer is show in Fig. 6 while Fig. 7 presents the experimental setup. Fig. 6. FEKP transducer: (a) Sampler with brass electrodes (distance to the surface of sample - 1 mm); (b) Transducer in the sound-shielded case (open for the photo); (c) Disassembled transducer (left, acoustic part is normally placed on the top of right, "swing capacitor" part). Fig. 7. Experimental setup for the linear scan with CKP or FEKP. To evaluate and compare FEKP and CKP, we have used improvised, in-house fabricated samples of Aluminum alloy with coatings by epoxy primer and polyurethane paint. Some substrates were treated by electrolytic oxidation to generate the corrosion layer. Fig. 6 shows results of measurement for the resulting Volta potential with the evaluated FEKP and the CKP for bare substrate, the corroded (1 hour) substrate, substrate coated with the primer only, and fully coated substrate (primer and paint) without the "corrosion" and "corroded" during 1 hour and 5 hours. To mitigate differences in the material of probe tip (stainless steel in CKP and brass in FEKP), and the gain of electronics, measured data were normalized to match data for the bare metal substrate., full OML layer and OML layers deposited on the substrates corroded during 1 hour and 5 hours., was tested with to reveal the magnitude and sign of corrosion produced space charge, with the results presented in Fig. 8.

Proc. 2017 Annual Meeting of the Electrostatics of America 6 Fig. 8. Normalized results of measurement with FEKP (Fixed Kelvin Probe) and CKP (Vibrating Kelvin Probe). Fig. 8 clearly illustrate the advantage of a fixed, nonvibrating OPS probe over the standard, vibrating (400 Hz) Kelvin probe. The crucial difference is in the periodic redistribution of the electric field between the space charge and substrate and the varied capacitance between the space charge and Kelvin probe. Specifically, the measurement for the corroded only substrate (no coating) shows a comparable Volta potential for both probes (~30% difference) because the space charge in the corrosion layer is located on the surface of the sample, and the polarization time constant is small. The induced charge in the Kelvin probe almost follows the variation of the probe-to-surface capacitance. When the corrosion layer is covered by the primer and paint structure, the polarization time constant increases, which results in the only partial induction of a charge in the vibrating Kelvin probe. To evaluate this explanation, we have performed the electrochemical impedance spectrometry (dry test) of the same polyurethane paint and epoxy primer- coated coupons. Results indicated that the coated material demonstrate a multipeak dielectric loss spectrum as shown in Fig. 9, where the imaginary part of capacitance (dielectric loss) is presented as the real part of the complex impedance. Fig. 9. The real part of the complex impedance for the pain-on-primer coated samples:(a) Effect of corrosion under the coating; (b) Deconvolution of spectral components for the non-corroded sample.

Proc. 2017 Annual Meeting of the Electrostatics of America 7 Spectra in Fig. 9(a) show that the peak near ~ 700 Hz is revealed for all samples, including the sample coated without the preliminary oxidation while the oxidation notably reduces dielectric loss at lower frequencies. Deconvolution of spectrum for the nonoxidated samples in Fig. 9(b) reveals at least 3 main components with peaks at 6 Hz, 50 Hz, and 850 Hz. Comparing Fig. 9(a) and 9(b), the conclusion is made that the peak at 850 Hz should be attributed to the Aluminum Oxide film, which always exists on the surface of the Aluminum substrate. The electrochemically induced growing of this film doesn t change the position of the peak that indicates the polarization of bulk oxide in contrast to the interfacial polarization. In contrast, peaks at lower frequencies (6 Hz and 50 Hz) are associated with the polarization of the external coating. This explanation supports data obtained with the FEKP and CKP shown in Fig. 8: the low-frequency polarization of coating does not follow the commercial Kelvin probe vibration frequency (in our experiments, 420 Hz). In contrast, the fixed electrode probe allows for the full static induction of charge, which is measured by the internal swing electrometer. This explains why the FEKP always shows the higher Volta potential (i.e., the higher density of induced charge) if the original charge, to be detected, is located under the dielectric layer. IV. CONCLUSIONS The implementation of compound, "swing" capacitor in the transducer of fixed electrode Kelvin probe allows to minimize the redistribution of charge in the measurement circuit and thus to measure the established, steady state value of the induced charge. The preliminary experimentation with the fixed electrode Kelvin probe demonstrated that the topcoating notably reduces the potential reading by the commercial, vibrating Kelvin probe, while the fixed electrode probe s reading is much less affected. REFERENCES [1] L. Kelvin, Contact Electricity of Metals, Philos. Mag., vol. 46, pp. 82 120, 1898. [2] W. A. Zisman, A New Method of Measuring Contact Potential Differences in Metals, Rev. Sci. Instrum., vol. 3, no. 4, pp. 367 368, 1932. [3] Karl Heinz Besocke, " Electrically driven oscillating capacitor device," U.S. Patent 4,100,442, 1978. [4] James R. Champion, Clarke A. Yeager, "Electrostatic probe," U.S. Patent 4625176 A, 1983. [5] Robert B. Greegor, Richard H. Bossi, "MEMS based Kelvin probe for material state characterization," U.S. Patent 8,618,824, 2013. [6] P.S. Riehl, K.L. Scott, R.S. Muller, R.T. Howe, J.A. Yasaitis, " Electrostatic Charge and Field Sensors Based on Micromechanical Resonators," J. Microelectromechanical Systems, Vol. 12, No. 5, pp. 5770589, 2003. [7] Aart G. Van Nie, Jean H. J. Lorteije, Johannes J. Z. Van Zelst, "Vibrating Capacitor", CA patent 685685 A to Philips, 1964. [8] M.A. Noras, " Electric Field Sensor Based On A Varactor Diode/MIS/MOS Structure," Industry Applications Society Annual Meeting (IAS), 2010 IEEE.