TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong

Similar documents
PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

MEASUREMENT UNCERTAINTY OF ROTATING TORQUE TRANSDUCERS WHEN USED IN PARTIAL LOAD RANGES

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements


THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

INVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE

Multi-Capacity Load Cell Concept

Unit 3 Transducers. Lecture_3.1 Introduction to Transducers

T20WN. Data Sheet. Torque transducers. Special features. Installation example with bellows couplings. B en

T e c h n i c a l L e x i c o n

Force and Displacement Measurement

Sensors and Transducers. mywbut.com

Module 2 Mechanics of Machining. Version 2 ME IIT, Kharagpur

(Refer Slide Time 03:12)

1 Force Sensing. Lecture Notes. 1.1 Load Cell. 1.2 Stress and Strain

Part 2. Sensor and Transducer Instrument Selection Criteria (3 Hour)

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Final Report on the Torque Key Comparison CCM.T-K2 Measurand Torque: 0 kn m, 10 kn m, 20 kn m Dirk Röske 1), Koji Ogushi 2)

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Application example. Measuring Force Sensors Rigid. Six series Nano, Mini, Gamma, Delta, Theta, Omega. Range of measurement, force ± 36 N..

Uncertainty restrictions of transfer click torque wrenches

Glossary Innovative Measurement Solutions

COMPLETION AND MEASUREMENT UNCERTAINTY BUDGET OF THE MULTI-COMPONENT MEASURING DEVISE FOR FORCE UP TO 1 MN AND TORQUE UP TO 2 KN M

Applying the Wheatstone Bridge Circuit

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

INSTRUMENTATION ECE Fourth Semester. Presented By:- Sumit Grover Lect., Deptt. of ECE

Investigation of Piezoelectric Force Measuring Devices in Force Calibration and Force. Standard Machines

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING

3-D Finite Element Modelling of a High Pressure Strain Gauge Pressure Transducer. M. H. Orhan*, Ç. Doğan*, H. Kocabaş**, G.

INSTITUTE OF AERONAUTICAL ENGINEERING Dundigal, Hyderabad DEPARTMENT OF ECE QUESTION BANK. : G.Lakshminarayana, Asst.

Comparison between the CENAM (Mexico) 150 kn and the INRiM (Italy) 1 MN force standard machines

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

Chapter 3. Lecture 3 Chapter 3 Basic Principles of Transducers. Chapter 3 - Definitions. Chapter 3. Chapter 3 7/28/2010. Chapter 3 - Definitions.

Theory and Design for Mechanical Measurements

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

Data Logger V2. Instruction Manual

2. (a) Differentiate between rare metal thermocouples and base metal thermocouples.

MET 301 EXPERIMENT # 2 APPLICATION OF BONDED STRAIN GAGES

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

DYNAMIC ROTARY TORQUE MEASUREMENT USING SURFACE ACOUSTIC WAVES

MEASUREMENT OF MAGNETIC MATERIAL

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING UNIT I - INTRODUCTION SYLLABUS

Loss analysis of a 1 MW class HTS synchronous motor

AE60 INSTRUMENTATION & MEASUREMENTS DEC 2013

Mechatronics II Laboratory EXPERIMENT #1: FORCE AND TORQUE SENSORS DC Motor Characteristics Dynamometer, Part I

Fitting PS P - B - R2

Piezoelectric Resonators ME 2082

MANUFACTURE OF STRAIN GAUGES & TRANSDUCERS

APPLICATIONS OF VIBRATION TRANSDUCERS

EE 3324 Electromagnetics Laboratory

Strain Measurement. Prof. Yu Qiao. Department of Structural Engineering, UCSD. Strain Measurement

Modelling of a dynamic torque calibration device and determination of model parameters

TORQUE CAPACITY ENHANCEMENT OF A MAGNETORHEOLOGICAL FLUID CLUTCH USING THE SQUEEZE-STRENGTHEN EFFECT

Type UB1 Load Cell. Option Stainless steel cable gland

Conventional Paper-I-2011 PART-A

The secondary winding have equal no. of turns. The secondary windings are placed identically on either side of the primary winding.

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Design and Development of Impact Load Sensor for Dynamic Testing Purposes

Pressure transmitters for industrial applications MBS 32 and MBS 33

Bending Load & Calibration Module

Unconventional Double-Current Circuit for Deflection and Temperature Simultaneous Measurement

AC Circuits Homework Set

ME 515 Mechatronics. Overview of Computer based Control System

Guidelines on the Calibration of Static Torque Measuring Devices

CIVL222 STRENGTH OF MATERIALS. Chapter 6. Torsion

D a t a s h e e t. T o r q u e s t a n d a r d. S e r i e s D m - T N. ( 1 N m N m ) Benefits/Application. Options/Accessories

AOE 3054 Measuring static displacement and deformation in structures

Experimentation. Third Edition. Wheeler. Ahmad R. Ganji. School of Engineering. With Third Edition contributions by. V. V. Krishnan. Brian S.

Exercise 1: Thermocouple Characteristics

ENCAPSULATION AND UNCERTAINTIES OF STRAIN-GAUGE SENSORS FOR STRESS-MONITORING OF CONSTRUCTIONS

Strain Gauges and Accessories

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

Enhancement of buckling load of thin plates using Piezoelectric actuators

Z7A. Load cells. Special features. Data sheet. Wiring code

An Introduction to Electrical Machines. P. Di Barba, University of Pavia, Italy

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

MAS.836 PROBLEM SET THREE

SECTION 3 BASIC AUTOMATIC CONTROLS UNIT 12 BASIC ELECTRICITY AND MAGNETISM

Exercise 2: Bending Beam Load Cell

TF 202 / TF 202-Ex. Field mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels

1. Mark the correct statement(s)

Gurley Model A58 Absolute Encoder

Engineering Science. 1 Be able to determine the behavioural characteristics of elements of static engineering systems

MEASUREMENT OF THE MECHANICAL STATE OF SUBSURFACE LAYERS BY EDDY-CURRENT METHOD

Strain Measurements. Isaac Choutapalli

CHAPTER 5. BRIDGES AND THEIR APPLICATION Resistance Measurements. Dr. Wael Salah

coil of the circuit. [8+8]

DESIGN OF A HIGH-EFFICIENCY MAGNETORHEOLOGICAL VALVE

PIEZOELECTRIC TECHNOLOGY PRIMER

INSTITUTE OF AERONAUTICAL ENGINEERING (Autonomous) Dundigal, Hyderabad

Alternating Current Circuits

Single-phase Transistor Lab Report. Module: EEE 108

Transcription:

IMEKO 22 nd TC3, 15 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong Torque Laboratory, Mechanical Department, National Institute of Metrology (Thailand), Pathumthani, Thailand 12120 Tel: +66 2577 5100, Fax: +66 2577 5095 Abstract: Effect of temperature and humidity on torque measuring devices were studied.with the aim to estimate measurement uncertainties due to environmental varying conditions, torque transfer wrenches and transducers were carried out by the measurement in an environmental controlled chamber under conditions of range 16ºC to 31ºC and 35 %RH to 65 %RH. The measurement responses from two kind of amplifiers, alternating current and direct current excitation amplifier were taken into account as well. The temperature sensitivity was significantly minimized causing by temperature compensation circuit. Moreover, the humidity sensitivities from several kinds of torque device structure consisted of directly contact to ambient, covering protection, waterproof protection and hermetically sealed seem to be similar and cannot be distinguished. On the other hand, amount of humidity sensitivity obtained by using direct current excitation amplifier are smaller than using alternating current excitation amplifier. Keywords: Torque measuring devices, temperature sensitivity and humidity sensitivity 1. INTRODUCTION National Institute of Metrology (Thailand) was requested to review uncertainty evaluation of torque measurement that one of quantities in ISO/IEC 17025 accreditation scope. Sources of measurement uncertainties are concerned including to the uncertainty due to environmental conditions such as temperature and humidity effect. Thus the temperature sensitivity and humidity sensitivity are necessary to known clearly. So these effects were studied. The former research, A Study of Torque Transfer Wrench Stability [1] shows that an influence of temperature significantly affects the accuracy of torque transfer wrenches in case of on-site calibration, but using temperature compensation circuit can reduce these effect. However for humidity, the experimental results show that there is no significant difference on humidity sensitivity among different humidity protections of torque transfer wrench. Thus, the level of moisture absorption in strain gauge is not a dominant influence of instability of the transducer due to humidity. For this reason, how humidity affects the transducer should be deliberately studied and understood. Principally, a bridge circuit of torque transducer is the main part to transform mechanical deformation to electrical signal. The circuit consists of electrical components, which are connector, cable, lead wire and strain gage. The researcher is interested in studying the humidity effect on electrical characteristic of the devices, in particularly the impedance of bridge circuit. To prove this hypothesis, torque measuring devices are sampled and measured repeatedly in order to compare temperature and humidity sensitivity from using two types of amplifier: alternating current (AC) excitation amplifier and direct current (DC) excitation amplifier. In case different measurement results for different devices with humidity sensitivity, it is affirmed that the impedance characteristic of circuit causes the humidity effect. To ensure that high accuracy torque transducers, one of torque measuring devices, have been tested under the same condition as torque transfer wrenches, high accuracy torque transducers are also under experiment as well. 2. IMPEDANCE IN BRIDGE CIRCUIT Normally strain gauge bridge circuit is a simple resistance circuit as shown in Fig. 1. The excitation voltage is stabilized by feedback control system; although there is voltage across the resistances pertained to cables. 1 2 Vex 4 3 Fig. 1, Strain gauge bridge circuit The output signal of bridge circuit equals to R / R, when the supply voltage is DC. On the other hand when the supply voltage is AC, impedance in strain gauge bridge circuit should be included in calculation. The impedance consists of resistance, inductance and capacitance. Inductance equation is described in equation (1), A L r N 2 0. (1) l Where L denotes the inductance, 0 denotes the magnetic constant (4π 10 7 H/m), r denotes the relative permeability, N denotes the number of turns, A denotes the cross-section area and l denotes length of the coil. Capacitance equation is described in Equation 2; V

A C r 0. (2) d Where C denotes the capacitance, A denotes the area of overlap of the two plates, r denotes the relative static permittivity, 0 denotes the electric constant (ε 0 8.854 10 12 F m 1 ) and d denotes the separation between the plates. Considering the phase relationship, the origin of the different signs for capacitive and inductive reactance is the phase factor in the impedance; 1 Z C j C (3) and j L. (4) Where Z L ZC denotes the capacitive reactance, Z L denotes the inductive reactance, j 1 and denotes the angular frequency. The simple impedance bridge circuit is shown in Fig. 2(a) and equivalent circuit is shown in Fig. 2(b). From the equation (1) and (2), the impedance of circuit depends on humidity, but for DC supply voltage the inductance and capacitance can be negligible. This difference of the circuits may cause the different output signals. (a) (b) Z1 Z4 1 2 4 3 1 2 4 3 Z2 Z3 Fig. 2, a) the simple impedance bridge circuit and b) the equivalent impedance bridge circuit. Z Z Z Z V Vex V Vex 3. EXPERIMENTAL SETUP In the experiment, torque measuring devices and amplifiers are set up and left in environmental controlled chamber in order to ensure identical of temperature and humidity absorption of the two devices and the environment is controlled as same situation as real application. The environmental controlled chamber produces by Airplus Apply Co., Ltd. It has a capacity of 8 m 3. The stability of temperature control is ± 0.3 C and humidity control is ± 1 %RH. Furthermore, to ensure that the devices and amplifiers are saturated by temperature and humidity, they have been left in controlled conditions at least 4 hours before measurement process begins The environmental conditions are; 1) To study on temperature stability, temperature is controlled between 16 C 28 C at 50 %RH constant humidity. 2) To study on humidity stability, humidity is controlled between 35 %RH - 65 %RH at 22 C constant temperature. The measurement points are approximately 0%, 50% and of torque capacity. Torque is applied by direct weight loading on a constant distance of arm length of torque transfer wrench itself. For torque transducer, torque is applied from weight loading to lever arm fabricated as shown in Fig. 3. 4. TORQUE MEASURING DEVICES For this research, torque transfer wrenches (TTW) and high accuracy torque transducers (TTS) have various structures, but can be classification by; Bridge circuit types: with temperature compensation circuit and without temperature compensation circuit. Humidity protection: hermetically sealed, water proof protection, open housing with protective covering on strain gauge and open housing with strain gauge directly contacted to an ambient humidity. 4.1 Commercial TTW Model: TTS, capacity: 50 N m S/N: 014640002, manufactured by HBM, Germany. Where sensing body is open housing type with radial shear design. Temperature compensation and protective coverings of full bridge circuit strain gauges are employed. 4.2 Commercial TTW Model: DmTS, capacity: 20 N m S/N: 48039, manufactured by GTM, Germany. Where sensing body is open housing type with bending design. Temperature compensation and protective coverings of full bridge circuit strain gauges are employed. 4.3 Commercial TTW Model: TTW 10, capacity: 10 N m S/N: TTW 101, manufactured by CEH, Germany. Where sensing body is open housing type with bending design. Protective coverings of full bridge circuit strain gauges were employed but no temperature compensation is furnished. 4.4 Fabricated TTW Model: Chock 03, capacity: 50 N m S/N:03.04.2551, fabricated by NIMT. Where sensing body is open housing type with bending design, without

temperature compensation and protective coverings of full bridge circuit strain gauges. 4.5 Fabricated TTW Model: Chock 04, capacity: 50 N m S/N: 290910, fabricated by National Institute of Metrology (Thailand), NIMT. Where sensing body was design based on the principle of load cell application in torque [2]. Temperature compensation and protective coverings of full bridge circuit strain gauges are employed. 4.6 Fabricated TTW Model: Chock 05, capacity: 200 N m S/N: 020910, fabricated by NIMT. Where sensing body is hermetically sealed type hollow shaft design, without temperature compensation and protective coverings of full bridge circuit strain gauges [3]. 4.7 Commercial TTS Model: DmTN, capacity: 20 N m S/N: 57340, manufactured by GTM, Germany. Where sensing body is open housing type with solid shaft design. Temperature compensation and waterproof protection of full bridge circuit strain gauges are employed. 4.8 Commercial TTS Model: TT1/10 N m, capacity: 10 N m S/N: 36838-04, manufactured by Raute Precision Oy. Where sensing body is open housing type with solid shaft design. Temperature compensation and protective coverings of full bridge circuit strain gauges are employed. 4.9 Commercial TTS Model: TB2/1000 N m, capacity: 1,000 N m S/N: #120730044, manufactured by HBMr, Germany. Where sensing body is hermetically sealed type axial shear principle design, Temperature compensation is employed. 1 with DC excitation voltage 10 VDC as shown in Fig. 4(c). (a) (b) 2.0 volt/div., 2.0 ms/div. 2.0 volt/div., 2.0 ms/div. (c) (a) Chok 03 (b) DmTS (c) TTW10 (d) TB2 (e) DmTN (f) Chok 04 (g) Chok 05 (h) TT1 (i) TTS Fig. 3, Experiment setup of each torque measuring devices. 5. AMPLIFIERS The experiment was set up with 2 amplifiers as followings. 5.1 Alternating current excitation amplifier, model: MGC Plus with measurement card model: ML38 and ML 30, manufactured by HBM, S/N: 801058930. AC excitation voltage 5 V rms, frequency 225.05 Hz and 600 Hz for ML38 and ML30 respectively are shown in Fig. 4(a) and 4(b). 5.2 Direct current excitation amplifier, manufacture by GTM, model: VN, S/N: CKDB3J015. Accuracy of class 5.0 volt/div., 5.0 µs/div. Fig. 4, Show the supply voltage signal of a) and b) AC excitation amplifier with measurement card ML38 and ML30 respectively, and c) DC excitation amplifier. All of torque measuring devices are measured by AC excitation amplifier and some of torque measuring devices (TTS model: TB2, TTS model: TT1, TTS model: DmTN, TTW model: Chock 03 and TTW model: TTW10) are also re-measured by DC excitation amplifier. 6. TEMPERATURE EFFECT Temperature stability is used to calculate temperature sensitivity and standard deviation in the form of absolute value and relative value described in Table 1. From the results, there is consistency of the relative value of sensitivity at 50% increasing, 50% decreasing and of capacity as shown in Fig. 5. Therefore, temperature sensitivity should be described in term of relative value.

SD. SD. Parameter Torque Measuring Devices Table 1: The temperature sensitivity obtained by using AC excitation amplifier. (mv/v)/k Temperature effect %/K 1.00 0.80 0.60 0.40 0.20 (a) TTS 1) DmTS 2) TTW10 2) Chok 03 2) Chok 04 2) Chok 05 2) DmTN 2) TT1 2) TB2 2) Noload 4.9E-04 4.5E-05 6.2E-02 5.7E-03-1.9E-05 2.8E-06-4.9E-03 7.2E-04-4.0E-05 4.9E-06-5.1E-03 6.3E-04-1.8E-05 3.1E-06-4.6E-03 7.8E-04 f0 6.0E-07 5.5E-07 7.6E-05 7.0E-05 h 1.5E-06 1.1E-06 3.8E-04 2.8E-04 Noload 9.7E-05 4.4E-06 5.9E-03 2.7E-04 6.8E-05 3.0E-06 8.2E-03 3.6E-04 1.4E-04 5.8E-06 8.2E-03 3.5E-04 6.2E-05 2.8E-06 7.6E-03 3.3E-04 f0-2.2e-06 2.4E-06-1.3E-04 1.4E-04 h -5.4E-06 1.5E-06-6.5E-04 1.8E-04 Noload 6.5E-05 4.1E-06 9.7E-03 6.1E-04 2.1E-04 2.0E-06 6.4E-02 5.9E-04 4.3E-04 4.4E-06 6.3E-02 6.5E-04 2.1E-04 2.0E-06 6.4E-02 6.0E-04 f0 7.0E-08 1.8E-07 1.0E-05 2.7E-05 h -4.2E-07 4.2E-07 9.9E-01 9.8E-01 Noload 6.7E-04 9.2E-06 4.9E-02 6.7E-04 2.6E-04 2.9E-06 3.9E-02 4.3E-04 5.2E-04 5.9E-06 3.8E-02 4.3E-04 2.6E-04 2.0E-06 3.8E-02 3.0E-04 f0-5.1e-07 7.5E-07-3.7E-05 5.5E-05 h -3.9E-06 1.5E-06-1.5E+00 5.6E-01 Noload -2.7E-05 6.3E-06-2.8E-03 6.5E-04-5.0E-06 5.0E-06-1.0E-03 1.0E-03-1.3E-05 9.6E-06-1.3E-03 9.9E-04-3.1E-06 4.7E-06-6.5E-04 9.7E-04 f0 5.8E-07 7.5E-07 6.1E-05 7.7E-05 h 1.7E-06 2.8E-06 3.5E-04 5.8E-04 Noload -2.5E-05 2.4E-06-4.0E-03 3.9E-04 1.0E-04 2.0E-06 3.3E-02 6.3E-04 2.0E-04 2.6E-06 3.3E-02 4.2E-04 1.0E-04 1.8E-06 3.2E-02 5.6E-04 f0-7.7e-07 3.5E-07-1.2E-04 5.5E-05 h -1.5E-06 1.2E-06-4.7E-04 3.8E-04 Noload 1.7E-05 1.0E-06-2.2E-03 1.3E-04 2.4E-05 5.8E-07-6.2E-03 1.5E-04 5.0E-05 1.1E-06-6.3E-03 1.3E-04 2.4E-05 8.0E-07-6.2E-03 2.0E-04 f0 3.4E-07 1.8E-07-4.3E-05 2.3E-05 h 1.5E-07 2.9E-07-7.3E-02 1.4E-01 Noload 4.1E-05 3.8E-06 3.8E-03 3.6E-04 2.7E-05 1.7E-06 5.0E-03 3.2E-04 4.7E-05 3.0E-06 4.4E-03 2.8E-04 2.4E-05 1.9E-06 4.4E-03 3.5E-04 f0 2.4E-08 1.7E-07 2.3E-06 1.6E-05 h -3.0E-06 1.0E-06-2.7E+00 9.3E-01 Noload 3.9E-06 3.2E-07 7.8E-04 6.4E-05 1.8E-05 3.0E-07 3.6E-03 6.0E-05 f0-4.8e-08 8.6E-08-9.6E-06 1.7E-05 Note: This measurement results are without compensation of coefficient of linear thermal expansion (CTE) of lever arm; 1) Made of stainless steel, CTE is 1.73 10 3 %/K, 2) Made of aluminum, CTE is 2.38 10 3 %/K, % -0.20 1.00 0.80 0.60 0.40 0.20-0.20 1.00 0.80 0.60 0.40 0.20-0.20 14 16 18 20 22 24 26 28 30 32 34 Temperature in C 14 16 18 20 22 24 26 28 30 32 34 Temperature, C 14 16 18 20 22 24 26 28 30 32 34 Temperature in C Fig. 5, Temperature stability on a), b) 50% of applied torque, and c) no load signal represent no temperature compensation represent temperature compensation. Stability of temperature for torque measurement have been classified to torque measuring devices without temperature compensation and with temperature compensation as shown in Fig. 6. /K 7.0E-02 6.0E-02 5.0E-02 4.0E-02 3.0E-02 2.0E-02 1.0E-02 0.0E+00-1.0E-02 (b) (c) Chok05 TTW 10 Chok 03 TTS DmTS Chok04 DmTN TT1 TB2 No temp. compensation Temp. compensation Fig. 6, Temperature stability of torque measuring devices classified by temperature compensation.

SD. SD. Parameter Torque Measuring Devices The stability of torque measuring devices without temperature compensation circuit are obviously affected by temperature with the sensitivity on temperature between 0.033 %/K to 0.066 %/K. These sensitivity is related to temperature expansion coefficient on gauge factor of strain gauge and Young s modulus of sensing body material. While for torque measuring devices with temperature compensation, the sensitivity is between 12 %/K to 35 %/K, which is one tenth of the previous groups. However, temperature compensation technique of commercial torque transducers is inadequate for other activities requiring greater temperature stability such as measurement comparison. 7. HUMIDITY EFFECT Humidity sensitivity and standard deviation are described in Table 2 and Fig. 7, the results of AC excitation amplifier. 0.06 0.04 0.02-0.02-0.04-0.06 0.06 0.04 0.02-0.02-0.04-0.06 0.40 0.30 0.20 0.10-0.10-0.20 (a) 75 (b) 75 (c) 75 Fig. 7, Humidity stability on a), b) 50% of applied torque, and c) no load signal represent no protective covering represent protective covering represent waterproof protection represent hermetically sealed Table 2: The humidity sensitivity obtained by using AC excitation amplifier. TTS i) DmTS ii) TTW10 ii) Chok 03 ii) Chok 04 ii) Chok 05 ii) DmTN ii) TT1 i) TB2 i) (mv/v)/%rh Humidity effect %/%RH Noload -2.3E-05 5.9E-06-3.0E-03 7.5E-04-1.9E-06 6.2E-07-4.9E-04 1.6E-04-4.4E-06 1.3E-06-5.6E-04 1.7E-04-3.2E-06 6.7E-07-8.0E-04 1.7E-04 f0-3.9e-07 1.5E-07-5.0E-05 1.9E-05 h -1.2E-06 2.5E-07-3.0E-04 6.4E-05 Noload 1.6E-07 1.1E-06 9.8E-06 6.5E-05 2.2E-06 4.1E-07 2.6E-04 4.9E-05 6.0E-06 9.5E-07 3.7E-04 5.8E-05 1.9E-06 6.2E-07 2.3E-04 7.6E-05 f0-5.9e-07 3.7E-07-3.5E-05 2.3E-05 h -6.2E-07 5.2E-07-7.6E-05 6.3E-05 Noload -1.1E-06 8.0E-07-1.6E-04 1.2E-04-1.8E-06 7.5E-07-5.4E-04 2.3E-04-2.2E-06 1.0E-06-3.2E-04 1.6E-04-9.9E-07 7.5E-07-3.0E-04 2.2E-04 f0 4.3E-08 5.3E-08 6.4E-06 7.9E-06 h 3.2E-07 2.0E-07-5.0E+00 3.0E+00 Noload 1.4E-04 7.0E-06 1.0E-02 5.1E-04 5.1E-06 9.3E-07 7.5E-04 1.4E-04 1.1E-05 1.2E-06 8.0E-04 8.7E-05 4.3E-06 8.9E-07 6.4E-04 1.3E-04 f0 3.6E-07 1.4E-07 2.6E-05 1.0E-05 h -3.9E-07 8.8E-07-1.6E-01 3.6E-01 Noload -2.7E-05 2.0E-06-2.8E-03 2.1E-04-1.2E-06 1.1E-06-2.6E-04 2.3E-04 9.3E-07 1.9E-06 9.7E-05 2.0E-04 3.7E-07 1.0E-06 7.7E-05 2.2E-04 f0 1.7E-07 2.7E-07 1.7E-05 2.8E-05 h 1.6E-06 5.1E-07 3.4E-04 1.1E-04 Noload -2.2E-06 1.6E-06-3.5E-04 2.5E-04-2.3E-06 1.2E-06-7.5E-04 3.9E-04-1.2E-06 2.2E-06-2.0E-04 3.4E-04-1.6E-06 1.2E-06-5.2E-04 3.7E-04 f0 2.6E-07 2.7E-07 4.1E-05 4.2E-05 h -1.3E-07 3.0E-07-4.2E-05 9.5E-05 Noload 4.9E-07 3.5E-07-6.1E-05 4.4E-05-8.0E-07 1.0E-07 2.0E-04 2.6E-05-2.4E-06 2.3E-07 3.0E-04 2.9E-05-1.9E-06 1.5E-07 4.8E-04 3.8E-05 f0-2.8e-08 6.8E-08 3.5E-06 8.5E-06 h -1.2E-06 1.5E-07 3.9E-01 4.9E-02 Noload -3.6E-06 4.0E-07-3.4E-04 3.8E-05 4.3E-06 1.3E-06 8.1E-04 2.5E-04 2.1E-06 1.3E-06 1.9E-04 1.2E-04 6.5E-07 7.8E-07 1.2E-04 1.5E-04 f0-2.3e-07 8.0E-08-2.1E-05 7.4E-06 h -3.7E-06 8.7E-07-6.2E+00 1.5E+00 Noload 1.2E-07 1.0E-07 2.3E-05 2.0E-05-1.3E-06 1.7E-07-2.6E-04 3.5E-05 f0-3.2e-08 3.7E-08-6.3E-06 7.3E-06 Note: The measurement results are obtained by using AC excitation amplifier with 2 types of frequency; i) represent frequency is 225.05 Hz and ii) represent frequency is 600 Hz

SD. SD. Parameter Torque Measuring Devices Fig. 8 shows humidity sensitivity of each torque measuring device group by classification of humidity protection of strain gauge as following; Group A, without protective covering Group B, protective covering on strain gauge Group C, waterproof protection Group D, hermetically sealed /%RH 1.0E-03 8.0E-04 6.0E-04 4.0E-04 2.0E-04 0.0E+00-2.0E-04-4.0E-04-6.0E-04-8.0E-04-1.0E-03 TTW10 Chok 03 DmTN TT1 TB2 Table 3: The humidity sensitivity obtained by using DC excitation amplifier. (mv/v)/%rh Humidity effect %/%RH Chok 03 TTS DmTS Chok04 TT1 TTW 10 DmTN Chok05 TB2 No protective Protective covering Waterproof protection Hermetically sealed Fig. 8, Humidity stability of torque measuring devices classified by humidity protection. From the experimental results, both relative and absolute humidity sensitivity has no correlation with applied torque. Thus humidity sensitivity value claimed should be derived from the experiment with at least 2-3 steps of torque applied. Considering the sensitivity of each group, there is no relationship between sensitivity quantity and humidity protection quality. It shows that humidity protection is unable to reduce humidity effect as of its purpose. In other word, humidity absorption of strain gauge is not the main parameter affecting the sensitivity. To substantiate a hypothesis that an impedance of bridge strain gauge circuit for alternating current supply voltage is the cause of stability due to humidity, some of torque transducer was measured to study on humidity sensitivity by using DC excitation amplifier. The measurement results as shown in Table 3. The objective is to provide measurement results obtained from a bridge strain gauge circuit with only resistance and compare them with the experiment results obtained from a bridge strain gauge circuit with elements of impedance. The results show that the stability of the DC excitation amplifier is significantly less affected by humidity than using AC excitation amplifier as shown in Fig. 9-13, graph of measurement comparison results between DC excitation amplifier and AC excitation amplifier. Therefore, it is concluded that the humidity stability does not mainly come from moisture absorption of stain gauge directly but mainly comes from electrical properties of the bridge strain gauge circuit, especially the impedance. 0.015 5-5 0.015 5 Noload -1.9E-05 1.5E-06-2.8E-03 2.3E-04-1.2E-06 9.5E-07-3.7E-04 2.9E-04-2.9E-06 1.2E-06-4.4E-04 1.8E-04-1.3E-06 1.2E-06-3.8E-04 3.6E-04 f0-1.9e-05 1.5E-06-2.8E-03 2.3E-04 h 6.5E-07 4.5E-07 5.4E-01 3.7E-01 Noload 1.3E-04 5.4E-06 9.7E-03 4.0E-04-2.7E-07 1.1E-06-4.0E-05 1.6E-04 1.3E-06 1.7E-06 9.4E-05 1.3E-04 1.2E-06 1.4E-06 1.7E-04 2.1E-04 f0-8.2e-08 3.0E-07-6.0E-06 2.2E-05 h 1.5E-06 1.8E-06 4.9E-01 6.1E-01 Noload -1.6E-07 2.8E-06 2.1E-05 3.6E-04-2.4E-07 1.6E-07 6.2E-05 4.0E-05-4.9E-07 5.9E-07 6.2E-05 7.5E-05-4.5E-07 5.9E-07 1.2E-04 1.5E-04 f0 5.6E-08 8.6E-08-7.1E-06 1.1E-05 h -2.4E-07 4.5E-07 9.1E-02 1.7E-01 Noload -5.0E-06 3.2E-07-4.6E-04 2.9E-05 3.4E-07 1.1E-06 6.3E-05 2.0E-04-1.1E-06 1.5E-06-1.0E-04 1.4E-04 2.8E-07 1.2E-06 5.3E-05 2.2E-04 f0 1.9E-07 2.4E-07 1.8E-05 2.2E-05 h -9.0E-08 9.2E-07-1.7E-05 1.7E-04 Noload 8.4E-08 1.6E-07 1.7E-05 3.2E-05-1.0E-06 6.5E-07-2.0E-04 1.3E-04 f0-1.7e-08 1.2E-07-3.4E-06 2.4E-05 DmTN_ of applied torque. DmTN_50% of applied torque. -5 Fig. 9, The comparison results of torque measuring devices model DmTN.

TT1_ of applied torque. Chok03_ of applied torque. TT1_50% of applied torque. Chok03_50% of applied torque. Fig. 10, The comparison results of torque measuring devices model TT1. - - - - y = -x + R² = 0.263 TTW10_ of applied torque. y = -x + 0.016 R² = 0.369 TTW10_50% of applied torque. Fig. 11, The comparison results of torque measuring devices model TTW10. - Fig. 12, The comparison results of torque measuring devices model Chok 03. TB2_ of applied torque. Fig. 13, The comparison results of torque measuring devices model TB2. In order to clearer understand the effect of AC excitation with humidity sensitivity, thus the sensitivity comparison of two different types of excitation amplifier is shown in Fig. 14. /K 1.0E-03 8.0E-04 6.0E-04 4.0E-04 2.0E-04 0.0E+00-2.0E-04-4.0E-04-6.0E-04-8.0E-04 TTW 10 Chok 03 DmTN TT1 TB2 DC excitation AC excitation Fig. 14, The humidity sensitivity from 2 different types of excitation amplifier.

8. CONCLUSION Major influence on the temperature sensitivity is temperature coefficient on gauge factor of strain gauge and Young's modulus of sensing body material. Although the temperature compensation circuit can reduce this impact approximately 10 times but the result is still not proper to be used in the secondary standard or inter laboratory comparison activities. Thus the study to find the effect of temperature must be done. For torque measuring devices, degree of humidity protection has insignificant impact in reducing of humidity sensitivity while an electrical impedance property is the main factor affecting to sensitivity. The conditions of bridge strain gauge circuit with only resistance or bridge strain gauge circuit with direct current excitation are the best condition in reducing sensitivity. Impedance quantity is varied according to frequency of alternating current power supply. This results change in of humidity sensitivity, as it is a result from frequency changing. Nowadays there are various amplifier frequencies so the impact of diverse frequency to humidity sensitivity should be studied. [6] DAkkS-DKD R 3-7, Guideline of static calibration of reference torque wrenches, Edition 1/2010. [7] DIN 51309, Guideline of material testing machines calibration of static torque measuring devices, Edition 2005. 9. ACKNOWLEDGMENT We would like to express our sincere gratitude to our colleague, thanks to Force laboratory, National Institute of Metrology (Thailand) for Direct current excitation amplifier lending. Thanks to Thailand Institute of Scientific and Technological Research (TISTR), Calibration Engineering Hohmann (CEH) and Gassmann Testing Metrology GmbH (GTM GmbH) for torque measuring devices lending. Thanks to Airplus Apply Co., Ltd for their courtesies of temperature and humidity control chamber lending supports for the experiment. Many thanks to Miss Apichaya Meesaplak for her kind helps and advices in composing this paper. I would particularly thanks to Mr. Bunjob Suktat for the invaluable support and guidance. 10. REFERENCE [1] N. Arksonnarong, T. Sanponpute, A Study of Torque Transfer Wrench Stability, The 3rd TSME International Conference on Mechanical Engineering, Chiang Rai, October 2012. [2] T. Sanponpute, C. Watthong, PROTOTYPE OF LOAD CELL APPLICATION IN TORQE MEASUREMENT, IMEKO 2010 TC3, TC5 and TC22 Conferences Metrology in Modern Context, November 22-25, 2010. [3] T. Sanponpute, C. Wattong, N. Arksornarong, A Design on Hermetically Sealed Torque Transfer Wrench, The Asia- Pacific Symposium on Measurement of Mass, Force, Torque, Xi an, 2011. [4] William H. Hayt, Jr., ENGINEERING ELECTROMAGNETICS, McGRAW-HILL INTERNATIONAL EDITIONS, ISBN 0071004882. 5 th edition [5] Final Report on the Torque Key Comparison CCM.T-K1, BIPM key comparison database, Appendix C, URL:http://kcdb.bipm.org/appendixb/appbresults/ccm.tk1/ccm.t-k1_final_report.pdf, access on 24/02/2011