Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Similar documents
CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Chemistry Instrumental Analysis Lecture 35. Chem 4631

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Types of Analyzers: Quadrupole: mass filter -part1

MS Goals and Applications. MS Goals and Applications

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

MS Goals and Applications. MS Goals and Applications

Mass Spectrometry in MCAL

Lecture 8: Mass Spectrometry

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

Lecture 8: Mass Spectrometry

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

TANDEM MASS SPECTROSCOPY

Introduction to the Q Trap LC/MS/MS System

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry of proteins, peptides and other analytes: principles and principal methods. Matt Renfrow January 11, 2008

Harris: Quantitative Chemical Analysis, Eight Edition

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

1. The range of frequencies that a measurement is sensitive to is called the frequency

This is the total charge on an ion divided by the elementary charge (e).

Chemistry 311: Topic 3 - Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Ion traps. Quadrupole (3D) traps. Linear traps

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Translational Biomarker Core

Molecular Mass Spectrometry

MS/MS .LQGVRI0606([SHULPHQWV

Molecular Mass Spectrometry

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

Time-of-Flight Mass Analyzers

MASS SPECTROMETRY. Topics

CHAPTER D3 TOF ION OPTICS

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Introduction to GC/MS

ICPMS Doherty Lecture 1

1 Mass Analysis. Gene Hart-Smith and Stephen J. Blanksby

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Finnigan LCQ Advantage MAX

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

Mass spectrometry.

Skoog/Holler/Crouch Chapter 19 Principles of Instrumental Analysis, 6th ed. CHAPTER 19

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Secondary Ion Mass Spectroscopy (SIMS)

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Mass Analysers for LC MS

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Thermo Finnigan LTQ. Specifications

Quattro Micro - How does it work?

LC-MS Based Metabolomics

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Mini review for final

Ch. 22 Mass Spectrometry (MS)

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Choosing the metabolomics platform

FT-ICR Mass Spectrometer QUANTRA. Technical Description September quantra

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Space Charge Effects in Linear. Quadrupole Ion Traps

vacuum analysis plasma diagnostics surface science gas analysis

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

Basics of Mass Spectrometry

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Accelerator Physics, BAU, First Semester, (Saed Dababneh).

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

First described by W. Paul and H. Steinwedel in

Interface (backside) & Extraction Lens

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Screening of pesticides residues by the time of flight analyzer (ToF) : Myth or reality?

Ever since the mass spectrometer was introduced,

VACUUM PUMPING METHODS

Accelerators and radiation spectra

AQA Chemistry A-Level : Atomic Structure

Accelerators Ideal Case

Hypervelocity Impacts of Macroscopic Bodies

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced

Accelerators. W. Udo Schröder, 2004

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen

Selecting and planning for the right mass spectrometer

Collision Cross Section: Ideal elastic hard sphere collision:

Transcription:

Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS)

Mass Analyzers Resolution R = m / Δm Accuracy/Precision mass measurement accuracy/reproducibility Transmission % of ions allowed through the analyzer Mass Range Highest m/z that can be analyzed Scan Speed How many spectra per unit of time

Double-Focusing Magnetic Sector

Double-Focusing Magnetic Sector Magnetic Sector m q = r 2 B 2 2V B = magnetic field strength r = radius of curvature in magnetic field V = accelerating voltage m = ion mass q = ion charge All ions of the same m/z will have the same radius Only if the Ion kinetic energy is constant

Double-Focusing Magnetic Sector Electric Sector r = 2Ek qe Ek = ion kinetic energy r = radius of curvature in electric field E = magnitude of electric field q = ion charge All ions exiting the electric sector have the same kinetic energy

Double-Focusing Magnetic Sector

Magnetic Sector Typically a voltage of 5-10kV is used to accelerate ions To obtain a full spectrum, magnetic field is scanned To obtain a HR scan, voltage is scanned at constant magnetic field To gain maximum sensitivity at one mass SIM scan is done B and E are constant for one or more masses

Double-Focusing Magnetic Sector Advantages Very High Resolution (60,000) High Accuracy (<5 ppm) 10,000 Mass Range Disadvantages Very Expensive Requires Skilled Operator Difficult to Interface to ESI Low resolution MS/MS without multiple analyzers

http://www.asms.org Quadrupole Mass Filter

Quadrupole E = U - Vcos(2πνt) E = potential applied to the rods U = DC potential V = RF amplitude ν = RF frequency t = time Quadrupole is scanned at a constant U/V

http://www.asms.org Quadrupole Mass Filter

Quadrupole Typically U varies from 500-2000 V V varies from 0-3000V (-3000 to +3000) Scanning U/V at a fixed ratio gives a full scan Higher values of U/V give higher resolution RF only (U=0) transmits all ions Higher sensitivity through SIM scan Jumping to specific points on the U/V line

Quadrupole Mass Filter Advantages Inexpensive Easily Interfaced to Many Ionization Methods Disadvantages Low Resolution (<4000) Low Accuracy (>100ppm) MS/MS requires multiple analyzers Low Mass Range (<4000) Slow Scanning

Quadrupole Ion Trap HPLC Inlet Octopoles Nebulizer Gas lnlet Nebulizer Needle Capillary Skimmer Lenses Main RF Auxillary RF (1/3 Frequency of Main RF) Conversion Dynode Endcap Electron Multiplier Drying Gas Optimized Asymptote Angle Vacuum Partition Ring Electrode

Quadrupole Ion Trap

Quadrupole Ion Trap Ions are injected into the trap and all ions are trapped RF and DC are scanned to sequentially eject ions for detection Specific ions can be trapped while others are ejected Ion velocity can be increased to induced fragmentation

Quadrupole Ion Trap Advantages Inexpensive Easily Interfaced to Many Ionization Methods MS/MS in one analyzer Disadvantages Low Resolution (<4000) Low Accuracy (>100ppm) Space Charging Causes Mass Shifts Low Mass Range (<4000) Slow Scanning

Time-of-Flight (TOF) mv 2 2 = zvs = Ek d t = v t 2 = m z d 2 2Vs B = magnetic field strength v = ion velocity Vs = accelerating voltage m = ion mass q = ion charge All ions of the same m/z will have the same flight time Only if the Ion kinetic energy is constant

Linear Time-of-Flight (TOF) Advantages Extremely High Mass Range (>1 MDa) Fast Scanning Disadvantages Low Resolution (4000) Low Accuracy (>200ppm) MS/MS not possible

TOF Ions are accelerated with 5-35 kv Space focusing of source ions is accomplished by delayed extraction An electrostatic analyzer (reflectron) is used correct for kinetic energy spread

Reflectron Time-of-Flight (MALDI-TOF) IS 1 IS 1 / IS 2 G round potential PCIS / Deflection Drift region P 2 + P 1 + P 1 + P 1 + Linear detector T arget Pare nts and Ref lec tor Ref lec tor plate fragments detector

Reflectron Time-of-Flight (ESI-TOF) Reflectron Drift Region Orthogonal Interface: Accelerator Detector Flight Tube RF Ion Guide ESI Source Spray Chamber Courtesy Bruker Datonics BioTOF user s Manual

Reflectron Time-of-Flight (TOF)

Reflectron Time-of-Flight (TOF) Advantages High Resolution (>20,000 in some models) High Accuracy (<3ppm) 10,000 Mass Range Fast Scanning >100 Hz Disadvantages Low Resolution for MS/ MS (PSD)

FT-ICR-MS

qvb = mv2 r Centripital Force FT-ICR-MS qb 2πm = v 2πr B = magnetic field strength v = ion velocity f = orbital frequency m = ion mass q = ion charge r= orbital radius = r and v drop out f v = 2πf r Circular Path At constant B, orbital frequency is inversely related to m/z Frequency is independent of kinetic energy

FT-ICR-MS Ions are all trapped radially by a magnetic field (typically 3-15 T) Axial trapping by DC potential Ion radius is increased by RF pulse also brings orbits into phase Orbiting ions induce RF current in receiver plates Image current is a composite of all frequences in time domain FFT gives frequency (mass) spectrum

FT-ICR-MS Actively shielded magnet ESI source source chamber transfer stage analyzer stage capillary ion guide cell ESI needle (atmosphere) 5 L/sec rotary vane 10-1 mbar 250 L/sec turbo-drag pump 10-4 mbar 500 L/sec turbo pump 10-6 mbar 70 L/sec turbo pump 10-8 mbar 500 L/sec turbo pump 10-10 mbar

FT-ICR-MS

FT-ICR-MS

FT-ICR-MS

FT-ICR-MS

FT-ICR-MS

FT-ICR-MS

FT-ICR-MS Electrospray: Broadband Spectrum of Bovine Serum Albumin (66kDa) 7.0T Actively Shielded Magnet 52+ Δm = 0.01933 a.u. 1/Δm = 51.7308 a.u. mass = 64428 a.u. 52+ 1278.3 1278.8 m/z 60+ 36+ 1200 1400 1600 1800 m/z

FT-ICR-MS Electrospray: Deconvoluted Spectrum of Bovine Serum Albumin (66kDa) 7.0T Actively Shielded Magnet Δm = 1.004 a.u. 66410 66430 66450 m/z

FT-ICR-MS Advantages Extremely High Resolution (>500,000) Very Good Accuracy (<1 ppm) MS/MS in one analyzer Disadvantages Expensive Requires Superconducting Magnet Slow MS/MS