Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Similar documents
MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Introduction to the Q Trap LC/MS/MS System

TANDEM MASS SPECTROSCOPY

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Lecture 8: Mass Spectrometry

Mass Spectrometry in MCAL

Types of Analyzers: Quadrupole: mass filter -part1

Lecture 8: Mass Spectrometry

Mass Spectrometry. What is Mass Spectrometry?

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

Harris: Quantitative Chemical Analysis, Eight Edition

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Chemistry 311: Topic 3 - Mass Spectrometry

CHAPTER D4 ORTHOGONAL TIME OF FLIGHT OPTICS

Week 5: Fourier Tranform-based Mass Analyzers: FT-ICR and Orbitrap

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS)

MS Goals and Applications. MS Goals and Applications

MS/MS .LQGVRI0606([SHULPHQWV

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Analysis of Polar Metabolites using Mass Spectrometry

ICPMS Doherty Lecture 1

Extrel Application Note

Mass Spectrometry. Fundamental LC-MS. Mass Analysers

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Suzanne Bell Second Edition

1) Introduction. 2) Types of Mass Analyzers - Overview. Assigned Reading: Nier mass spectrometry paper Extra Reading:

The Aerosol Ion Trap Mass Spectrometer (AIMS): Instrument development and first experimental results

Introduction to GC/MS

CHAPTER D3 TOF ION OPTICS

MS Goals and Applications. MS Goals and Applications

Secondary Ion Mass Spectroscopy (SIMS)

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

Mass Selective Ejection by Axial Resonant Excitation from a Linear Ion Trap

Introduction to LC-MS

This is the total charge on an ion divided by the elementary charge (e).

Design considerations for linear Paul trap mass spectrometer under development

CHM 424 EXAM 4 CRIB - COVER PAGE FALL

Ion traps. Quadrupole (3D) traps. Linear traps

Mass Analysers for LC MS

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Photoelectron Spectroscopy using High Order Harmonic Generation

1. The range of frequencies that a measurement is sensitive to is called the frequency

PARTICLE ACCELERATORS

- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

Quadrupole Storage Mass Spectrometry

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Molecular Mass Spectrometry

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Section 4 : Accelerators

Computations on Gabor lens having two different field distributions

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Interface (backside) & Extraction Lens

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

Space Charge Effects in Linear. Quadrupole Ion Traps

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Introduction to Elemental Analysis

Translational Biomarker Core

Quattro Micro - How does it work?

Simulation of External Ion Injection, Cooling and Extraction Processes with SIMION 6.0 for the Ion Trap/Reflectron Time-of-flight Mass Spectrometer

Penning Traps. Contents. Plasma Physics Penning Traps AJW August 16, Introduction. Clasical picture. Radiation Damping.

Theory English (Official)

Principles of Quadrupole and Ion Trap Mass Analyzers

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER

Technical Note # TN-24 Extending the Capacity in Modern Spherical High Capacity Ion Traps (HCT)

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Mass spectrometry.

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Molecular Mass Spectrometry

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Mass Spectrometry. Ionizer Mass Analyzer Detector

Appendix A2. Particle Accelerators and Detectors The Large Hadron Collider (LHC) in Geneva, Switzerland on the Border of France.

MASS SPECTROMETRY. Topics

Mass Spectrometry. A truly interdisciplinary and versatile analytical method

Time-of-Flight Mass Analyzers

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

Introduction Introduction

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

Welcome to Organic Chemistry II

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

Particle physics experiments

M M e M M H M M H. Ion Sources

Asymmetrical features of mass spectral peaks produced by quadrupole mass filters

Ion Trajectories in the Injection Quadrupole for SIFT-MS

NEW ION TRAP FOR FREQUENCY STANDARD APPLICATIONS*

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Using a Microwave Interferometer to Measure Plasma Density Mentor: Prof. W. Gekelman. P. Pribyl (UCLA)

Characteristics and classification of plasmas

Force Due to Magnetic Field You will use

Secondary ion mass spectrometry (SIMS)

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Transcription:

Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted from the ion source and sent to a device called the Mass Analyzer (Figure 1a). Gas phase ions in motion can be characterized by their momentum, kinetic energy, velocity, and response to electric or magnetic field. Mass analyzers make use of one or more of these properties to distinguish between ions. There are many types of mass analyzers and they can be classified in a variety of ways. One classification (based on their general principle of operation) sorts them into two groups: Mass filters are devices in which only ions of interest make it through the mass analyzer to the detector (Figure 1b). All other ions are neutralized. Conditions on the mass analyzer are then reset so that other ions are guided to the detector. Beam analyzers are devices in which the ions are separated according to their m/z such that each ion travels through the mass analyzer and arrives at the detector at q different time or different location on q focal plane (Figure 1c). Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Figure 1a: Ions produced in the ion source (blue and red dots) are extracted (from cloud of neutral compounds and ions) and accelerated by the ion optics into the mass analyzer.

Figure 1b: General principle of operation of a mass filter. With mass filters a mixture of ions that are continuously extracted by the ion optics are introduced into the mass analyzer. Under a set of conditions imposed on the mass analyzer, only one group of ions (with a unique m/z) makes it through to the detector (the red ones in the diagram). Other ions are neutralized and removed from the mass analyzer (the blue ions produced in the ion course in Figure 1a). Figure 1c: General principle of operation of beam analyzers. In an ideal situation, all m/z ions successfully make it through the mass analyzer, but they arrive at the detector at different times or at different points in space (i.e. the ions are spatially resolved). In the diagram, the blue and red ions will make it to the detector at different times.

1. Magnetic Sector Mass Analyzers Ion Extraction optics Magnet Ion flight tube Ion Source Ion Detector Figure 2: Schematic diagram of a magnetic sector instrument. These are first generation mass analyzers that date back to the days of J.J. Thompson experiment on analysis of positive rays. As the name indicates, they are comprised of a magnet that typically is in a semi-circular shape. An ion flight tube runs through the magnet (see Figure 2). The interaction of ions with the magnetic field can be modeled by the following expression: m z 2 2 B r e Where m is the mass of the ion (in kg), z is the number of charges on the ion, e is the electronic charge (1.602 x 10-19 C), r is the radius of the ion path through the flight tube, V is the acceleration potential of the ions and B is the strength of the magnetic field (in Tesla, T). 2V From this expression, it can be predicted that at a given value of B and V, ions with smaller m/z will travel in a path of smaller radius compared to an ions with a bigger m/z ratio. However, only ions that travel in a path that is similar in radius to the flight tube can successfully make it through to the detector. Other ions travel in a path that leads them to collide with the wall of the flight tube. They become neutralized and are pumped out of the tube under the prevailing low pressure (~10-6 Torr) condition. Varying the magnetic field while keeping the acceleration potential (V) constant will vary which ions (of select m/z) successfully make it through the flight tube to the detector. Monitoring the signal that is generated by the detector for each group of ions and plotting that signal versus the m/z of each group results in the mass spectrum

2. Quadrupole Mass Analyzers Figure 3: Schematic diagram of a quadrupole mass analyzer. The quadruple mass analyzer is also called a mass filter (Figure 3). It consists of four rods (conducting metals). Separation of gas phase ions is achieved by creating an electric field that makes the ions oscillate within the space between the rods. At a particular set of conditions, only one m/z ion has the correct oscillation to make it through the quadrupole to the ion detector. Other ions oscillate with so much amplitude that they collide with one of the rods and are neutralized and pumped out of the system. To accomplish the oscillation, a positive dc voltage is applied to one pair of the rods and a negative dc potential is applied to the other pair. Additionally, an alternating radiofrequency voltage is superimposed on the rods. At an instant of time, the ions are attracted to one pair of the rods and repelled by the other pair.

Figure 4: Representation of the pair of rods on the y-z plane of a quadrupole mass analyzer. Assuming that the ions are positively charged, when they oscillate in the y-z plane (that is they are vibrating in the y direction while travelling in the z direction), they experience repulsion from the positively charged rod (Figure 4). This repulsion focuses them onto the z-axis. This is strongly the case during the positive cycle of the superimposed rf potential. However, during the negative cycle of the rf potential, the ions are slightly defocused away from the z-axis. Relatively high m/z ions are less affected by this defocussing. Relatively low m/z ions are more affected and can oscillate high enough that they collide with the rods. Upon collision with the rods, they are neutralized and are removed from the analyzer chamber (i.e. they are filtered out). With a certain combination of dc (U) and ac (Vsin t) potentials, a group of ions with similar m/z will successfully travel through the analyzer to the detector. When the dc-ac conditions are changed, another group of ions of another m/z will successfully navigate their way to the detector. Repeating this process such that a range of ions (i.e. a range of m/z) are eventually guided to the detector sequentially is called scanning. Two dimensionless parameters, a and q (shown below) define the conditions for successful oscillation. 8zeU a u a x a y m r, 4zeV qu qx qy m 2 r 2 Re-arranging we get: 2 2 o o

m 8eU m 4eV and 2 2 z a r z q r u 2 2 o u o Where r o is the inscribed radius (or ½ the distance between opposite pair of rods, in m ) = 2f, and f is the radio frequency m is the mass of a single ion in kg e is electronic charge (1.602 x 10-19 C), U is the dc potential (in V) V is the rf potential (in V) During scanning, U and V are varied linearly to keep their ratio constant ( a/q = 2U/V ). At a particular combination of U and V, values of a and q are set and this determines which m/z ion navigates successfully though to the detector. 3. Time-of-flight Mass Analyzer In a time-of-flight (TOF) mass analyzer, ions from the ion source are directed into a region where they are repelled and accelerated into a field-free flight tube. The repulsion and acceleration of an ion depends on its m/z ratio. Therefore, ions in the flight tube travel at different speeds and arrive at the detector at different times (Figure 5).

Figure 5: Schematic diagram of a time-of-flight mass analyzer. The time of arrival of an ion is given by the expression: L m t L v 2qV Or L m t L v 2zeV Where t is the time, v is the velocity, and L is the distance (in m) between the extraction optics and the detector. L can be approximated with the length of the flight tube. Note that ions with smaller m/z will travel faster and reach the detector sooner. A TOF mass analyzer is calibrated with compounds of known molecular weight and fragments so that the time can be related to several m/z values. A generally accepted calibration equation is: m at 2 b q A TOF mass analyzer is an example of a beam analyzer because all the ions in a packet are transmitted to the detector.

4. Ion Trap Mass Analyzers Ion traps are mass analyzers that can perform dual roles of ion storage and mass analysis. A few things to note about ion traps are that: The ions are typically stored for only a limited period of time, There is a limit to the amount of ions that can be stored in a given device otherwise development of space charge may lead to resolution problems. Ion traps can also function as molecule-fragmentation and mass a selective device There are several kinds of ion traps which can be broadly classified as indicated in Figure 1 below: Figure 1: A Classification of ion trap devices. Note: LIT means linear ion trap. QIT means Quadrupole ion trap The 3D QIT will be discussed here as an example (representative) of ion traps. Threedimensional quadrupole ion traps (3D, QITs ) are commonly employed in the GC-MS systems that are frequently found in undergraduate laboratories. These devices confer multiple stage mass spectrometric (MS n ) capability, small size and low cost nature to the bench-top GC-MS systems.

Figure 2: Schematic diagram showing the electrodes of a quadrupole ion trap (QIT). Three-dimensional QIT s comprise of three electrodes: two end-caps and a ring electrode (see Figure 2). Both end cap electrodes have orifices for ion injection and ejection. An ion source is placed in front of the entrance end-cap electrode and a transducer is positioned behind the exit end-cap electrode. Operation of the QIT can be broken into following steps: ion injection, ion trapping, ion fragmentation (if necessary), and sequential ion ejection and detection (i.e. mass analysis). Each of the steps is further described below: Ion Injection Electron and chemical ionization devices are used for generating ions from thermally stable and volatile molecules, while electrospray ionization device is used to generate ions from polar and non-volatile molecules. These externally generated ions are extracted with appropriate lenses into the trap. The amount of ions injected/extracted into the trap is controlled to avoid space charge which can subsequently disrupt mass resolution. Ions are injected into the trap in batches. The ions are trapped, fragmented (if necessary), analyzed and detected before the next batch of ions is introduced trapping chamber. Ion Trapping Since the kinetic energies of externally generated ions are typically too high to permit efficient trapping, a bath gas (e.g. helium) is allowed into the chamber to cause their collisional cooling (i.e. the reduction of their kinetic energy). This cooling also enables the confining of ions to (or near) the center of the trapping chamber. This is important because ions that travel closer to the electrodes are easily lost due to the steeper gradient of the electric field near the chamber. To sustain the ions in stable trajectories within the chamber, a high ac voltage called the fundamental RF is applied to the ring electrode while the end caps are kept grounded. This

makes each ion to oscillate axially while travelling in a circular trajectory which is similar to the Lissajous shape. Frequency of oscillation of the ions under this condition is called the secular frequency. Ions of same m/z have a unique secular frequency of oscillation. Ion Fragmentation To facilitate fragmentation of the trapped ions, a small auxiliary RF voltage is applied to the endcaps. When the frequency of this auxiliary voltage matches the secular frequency of an ion, the amplitude of its axial oscillation is increased and it collides with the bath gas and fragments. Such an ion is described as tickled and the voltage that is used in effecting the excitation is called the tickle voltage. Magnitude of the tickle voltage (or auxiliary voltage) is about 100mV. Ion Ejection/Mass Analysis A few methods are utilized for ion ejection from the trap but only the resonance ejection approach will be described here for simplicity. To initiate resonance ejection of ions with a particular m/z, first the corresponding fundamental RF is set, and the amplitude (magnitude) of the auxiliary RF is increased until its frequency matches the secular frequency of the desired ion (i.e. m/z). The trajectory of these ions is destabilized, as their axial oscillation increases and they escape through the orifice of the exit end cap to the detector. The fundamental and auxiliary rf voltages are then changed the next ion is then ejected. Advantages of ITMS a. ITMS generally have high sensitivity. This is facilitated by their ion storage/accumulation ability, coupled with the fact that ions that are not of interest can be rejected. Combination of these two factors leads to high signal to-noise ratio and consequently high sensitivities. b. ITMS can be used for Multistage mass spectrometry (MS n ) studies in which parent/precursor ion can be fragmented, and the fragment can be further fragmented to obtain structural information of molecules. c. QIT in particular can be coupled with ions sources that are operated at modest (e.g. CI) to high pressures (ESI). d. The small size of these mass analyzers enables the design of bench-top mass spectrometers (e.g. GC-MS). Disadvantages a. ITMS are not generally good for quantitation b. They have limited dynamic range c. Collision energies are not well defined when operated in the collision induced dissociation (fragmentation) mode, thereby leading to irreproducibility of ions. d. There are too many parameters to control to obtain quality mass spectrum

Describing Performance of Mass Analyzers Several important parameters describe the performances of a mass analyzer The mass range is the highest m/z that can be analyzed on a particular mass analyzer. Resolution is a measure of the mass analyzer s ability to separate and distinguish ions of similar masses. Mathematically, it is calculated using the expression: m 2 R m Where m 2 is the larger of the two masses and m is their difference. Mass Accuracy is defined as the error in measuring the mass in m/z of a particular ion divided by its calculated exact mass. mass error Mass Accuracy 10 exact mass 6 ppm Mass Error = Exact Mass Measured mass (Note: measured mass is the experimentally measured mass while Exact Mass is the calculated mass based on the sum of exact masses of each atom in the molecule). Tables of exact masses of atoms are commonly available (e.g. http://www.nist.gov/pml/data/comp.cfm )