A simplified method of VNA uncertainty estimation

Similar documents
VNA Tools II S-parameter uncertainty calculation

Guidelines on the Evaluation of Vector Network Analysers (VNA) EURAMET Calibration Guide No. 12 Version 3.0 (03/2018)

Traceability of on-machine measurements under a wide range of working conditions

Analyzing Uncertainty Matrices Associated with Multiple S-parameter Measurements

Calibration Uncertainty Estimation. at the Wafer Level

Accuracy Estimations of S-Parameter Measurements at the Wafer-Level

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Ripple Method for Evaluating Residual Errors

Develop and characterize a flow metering standard down to 1 nl/min

Temperature Coefficients for DC Resistance of Match and Reference Attenuators

Numerical modeling of the primary source in a hemi-anechoic room

Advances in Radio Science

VNA error models: Comments on EURAMET/cg-12/v.01

Influence of directivity and spectral shape on the measured sound power level

New Methods for Measuring the Effective Source Reflection Coefficient of Microwave Generators. Jens Leinhos, Thomas Reichel and Jochen Simon

Nanoholes for leak metrology

Analysis of shock force measurements for the model based dynamic calibration

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

Estimation of Complex and Linear Uncertainities in S-Parameter Measurements for Metrology Applications

Guidelines on the Calibration of Static Torque Measuring Devices

SI Broader Scope - Projects

Communicating uncertainties in sea surface temperature

European Project Metrology for Radioactive Waste Management

Importance and estimation of mismatch uncertainty for RF parameters in calibration laboratories

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Lecture 13. Vector Network Analyzers and Signal Flow Graphs

Contents. ! Transmission Lines! The Smith Chart! Vector Network Analyser (VNA) ! Measurements. ! structure! calibration! operation

Science of Stability 2017

Accuracy Calibration Certificate

The in-flight calibration system for the airborne imager GLORIA

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Advancements in mm-wave On-Wafer Measurements: A Commercial Multi-Line TRL Calibration Author: Leonard Hayden Presenter: Gavin Fisher

Measurement of S-Parameters. Transfer of the Reference Plane. Power Waves. Graphic Representation of Waves in Circuits

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements

Waves on Lines. Contents. ! Transmission Lines! The Smith Chart! Vector Network Analyser (VNA) ! Measurements

Open Access Repository eprint

Experiment 06 - Extraction of Transmission Line Parameters

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Measurement and Calibration of a High-Sensitivity Microwave Power Sensor with an Attenuator

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Contents. Transmission Lines The Smith Chart Vector Network Analyser (VNA) ü structure ü calibration ü operation. Measurements

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Calibration of Temperature Block Calibrators

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Comparison protocol EURAMET project

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3

UNCERTAINTY EVALUATION OF SINUSOIDAL FORCE MEASUREMENT

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Traceable UV-Vis Reference Materials

TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 2017

Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note

MEASUREMENT UNCERTAINTY EVALUATION OF TORQUE MEASUREMENTS IN NACELLE TEST BENCHES

Contents. Transmission Lines The Smith Chart Vector Network Analyser (VNA) ü structure ü calibration ü operation. Measurements

Certificate of Analysis

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

This section reviews the basic theory of accuracy enhancement for one-port networks.

Uncertainty Propagation for Force Calibration Systems

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

An Introduction to the Differences Between the Two Most Recognized Force Standards

Document No: TR 12 Issue No: 1

Vocabulary of Metrology

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Unknown Thru Calibration Algorithm

ANNEXE 8. Technical protocols for the interlaboratory comparisons

Investigations for the model based dynamic calibration of force transducers by using shock excitation

DEVELOPMENT OF MEASUREMENT STANDARD FOR DYNAMIC PRESSURE AT MIKES

Practical Considerations and Solutions for Temperature-Dependent S-Parameter Measurement for Accurate Parameter Extraction of

Comparison of V and 10 V DC Voltage References

Progress of total ozone data retrieval from Phaeton - REG(AUTH)

Technique for the electric and magnetic parameter measurement of powdered materials

Accurate Measurement of Transmittance and Reflectance for Engineering Applications

Development of procedures for calibration of meteorological sensors. Case study: calibration of a tipping-bucket rain gauge and data-logger set

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Uncertainty of Calibration. Results in Force Measurements

Measurement standards and the general problem of reference points in

THE NEW WATT BALANCE EXPERIMENT AT METAS

Infrared Thermometer Calibration A Complete Guide

Metrological Characterization of a Primary Vickers Hardness Standard Machine - NIS Egypt

Prior information, but no MCMC:

A CALIBRATION PROCEDURE FOR TWO-PORT VNA WITH THREE MEASUREMENT CHANNELS BASED ON T-MATRIX

Final Report CCEM.RF-K8.CL COMPARISON CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Quantifying and Verifying trace stability of vector network analyzers at frequencies below 1GHz

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

WE PRESENT an iterative algorithm, which we first

The Role of Proficiency Tests in the Estimation of Measurement Uncertainty of PCDD/PCDF and PCB Determination by Isotope Dilution Methods

Broad-band space conservative on wafer network analyzer calibrations with more complex SOLT definitions

MVS Uncertainty of Measurement

Liaison with the World Meteorological Organization (WMO) Isabelle Rüedi

Product Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541

Transcription:

A simplified method of VNA uncertainty estimation Markus Zeier, Johannes Hoffmann, Jurg Ruefenacht, Michael Wollensack METAS ANAMET Workshop, PTB Braunschweig, 17.12.2014 ANAMET 1 METAS

Outline Introduction Basic Idea Method Summary ANAMET 2 METAS

Problem Rigorous uncertainty propagation through VNA measurement model requires specification of basic uncertainty influences and use of software. Residual error method with ripple technique requires handling of beadless airlines and makes questionable assumptions in its current form. Is it possible to estimate VNA uncertainty in a simpler way? ANAMET 3 METAS

Possible Solution Black Box approach that combines verification process with uncertainty estimation. Credits: Based on a method developed by Frederic Pythoud (EMC lab METAS). The method is promoted by the Swiss Accreditation System for use in EMC labs. Method is being refined and is at this point still preliminary. ANAMET 4 METAS

Principle Previously characterized verification standards are measured The deviation needs to be within previously defined tolerance intervals The quoted uncertainty is based on the tolerance intervals and other contributions based on additional measurements. ANAMET 5 METAS

Simple example: Measurement of matched load Ingredients: Declared tolerance interval T low reflect Characterized OSL kit for VNA calibration Characterized verification load: S11 load ± u ( S11 load ) Steps: VNA is calibrated with OSL Verification load is measured: M11 load Tolerance check: M11 load S11 load < T low reflect DUT load measured: S11 dut ( ) ± u rep S dut 11 u ( S11 dut ) = (u (S11 load )) 2 + ( Tlow reflect 2 ) 2 + (urep (S dut 11 )) 2 ANAMET 6 METAS

Uncertainty influences that need to be addressed: Noise Drift Linearity Cable Connector Calibration standards ANAMET 7 METAS

Ingredients A set of traceably characterized calibration standards Open, Short, Load A set of traceably characterized stable verification standards High reflects: Open, Short Load Something with S 11 0.5, e.g. T-checker Attenuators Transparent device: Beaded airline or adapter ANAMET 8 METAS

Verification process Characterized verification standards: S ver xx, u (S ver xx ) Previously defined Tolerance intervals/regions (ev. f dependent): T ver Verification measurements: M ver xx Verify that M ver xx Assign basic uncertainties: S ver xx < T ver ( ) Low reflect: u lr = (u (S11 load )) 2 2 + Tlr 2 High reflect: u hr = (max (u (S11 open ), u (S11 short ))) 2 + Linearity (S 11 0.5): u l = (u (S T check 11 )) 2 + ( ) 2 Tlin 2 ( ) 2 Thr 2 ANAMET 9 METAS

Calibration Standards: One Port PRELIMINARY: DUT S 11 : Uncertainty contribution to S 11 due to one port standards ( ) u 1p S dut = u lr + S dut (u lr + u hr ) 11 for u r = max (u lr, u hr ): ( ) ( u 1p S dut = 1 + 2 11 11 S11 dut ) u r ANAMET 10 METAS

Calibration Standards: Two Port Reflection PRELIMINARY: SOLT: DUT S 11 : Uncertainty contribution to S 11 due to one port standards: ( ) u 1p S dut = u lr + S dut (u lr + u hr ) + S dut u lr 11 for u r = max (u lr, u hr ): ( ) ( u 1p S dut = 1 + 2 11 11 S11 dut Unknown Thru: Not yet done. + S21 dut ) u r 21 ANAMET 11 METAS

Calibration Standards: Two Port Transmission PRELIMINARY: SOLT: DUT S 21 : Uncertainty contribution to S 21 due to one port standards: ( ) u 1p S dut = 2 S dut S dut u 2 lr + u lru hr 21 11 for u r = max (u lr, u hr ): ( ) u 1p S dut = 2 2 21 21 S11 dut S21 dut Unknown Thru: Not yet done. u r ANAMET 12 METAS

Drift Repeat verification procedure after DUT measurement again and recheck that deviations are still within tolerances. ANAMET 13 METAS

Noise and Connector Measurements of DUT are repeated at n (n 4) different connector orientations: u rep (S xx ) = max {s (Re [S xx ]), s (Im [S xx ])} Standard deviation: n s (x) = 2 i=1 (x i ˆx) 2 n 1 ANAMET 14 METAS

Cable One port measurements: Keep cable fixed Two port measurements: If possible use Unknown Thru calibration to avoid cable movements For any cable movements: 1. Calibrate VNA. 2. Connect DUT. 3. Repeat movement of cable during 1. or 2. with DUT connected and record change in S-parameters. 4. Take maximum difference in S-parameters as an additional uncertainty contribution: u cable (S xx ) = max ( S xx ) ANAMET 15 METAS

Linearity Reflection: u lin ( S11 dut ) = 2u l Transmission: in work ANAMET 16 METAS

Combined uncertainty Reflection: ( ( )) u S dut 2 11 = ( ( )) u 1p S dut 2 ( ( )) 11 + ulin S dut 2 ( ( )) 11 + ucable S dut 2 ( ( )) + urep S dut 2 Transmission: in work 11 11 ANAMET 17 METAS

Advantages Simple cookbook type procedure, clear pass/fail criteria. No handling of beadless airlines No software for uncertainty propagation needed. No previous characterization of VNA needed. The connector effect is taken care of. Not limited in frequency. The method combines verification and uncertainty evaluation. ANAMET 18 METAS

Disadvantages The method is not strictly GUM. Uncertainties are not propagated through a measurement model. The method does not calculate uncertainty intervals with a defined coverage (e.g. 95%). The coverage is > 95%. No detailed uncertainty budget and therefore no deeper understanding where the uncertainties are coming from. ANAMET 19 METAS

Conclusion The method is correct in the sense that VNA uncertainties are not underestimated. The method is suitable for labs with low accuracy requirements. The method is suitable for labs where S-parameters are just secondary quantities. The efforts of implementation are manageable. An assessor can relatively easy verify if a lab fullfills the requirements. We recommend to put it in cg-12! ANAMET 20 METAS

Acknowledgement This work is funded through the European Metrology Research Programme (EMRP) Project SIB62 Metrology for New Electrical Measurement Quantities in High-frequency Circuits. The EMRP is jointly funded by the EMRP participating countries within EURAMET and the European Union. ANAMET 21 METAS