DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS

Similar documents
DSC AS PROBLEM-SOLVING TOOL: BETTER INTERPRETATION OF Tg USING CYCLIC DSC

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14

High Pressure DSC Differential Scanning Calorimeter

APPLICATIONS OF THERMAL ANALYSIS IN POLYMER AND COMPOSITES CHARACTERIZATION. Wei Xie TA Instruments

Thermal Methods of Analysis

Differential Scanning Calorimetry

Modulated DSC Paper #8 Use Of Quasi-isothermal Mode for Improved Understanding of Structure Change

Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC)

Apparent Melting: A New Approach to Detecting Drug-Excipient Incompatibility

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

TGA Thermal Gravimetric Analysis

How to Use Kinetic Modeling Analysis To Predict Profile Part Failure

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

Chapter 31. Thermal Methods

DETERMINATION OF THE DECOMPOSITION PRODUCTS OF CALCIUM OXALATE USING THERMAL GRAVIMETRY AND INFRARED SPECTROMETRY

TECHNICAL UPDATE. Ricon Resins Peroxide Curing Data and Use as a Reactive Plasticizer in Polyphenylene Ether Based CCL and PWB

HYPHENATED TECHNOLOGY GUIDE

TOPEM the new advanced multi-frequency TMDSC technique

Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis

APPLICATION NOTE. Characterization and Classification of Recycled Polyamides by Means of Identify. Dr. Ekkehard Füglein

HYPHENATED TECHNOLOGY GUIDE

Nutshells of Thermal Analysis. Heat it up! Burn it! Thermal Analysis

Characterization of Solid State Drugs by Calorimetry

Thermal methods. P A Sathe Ramnarain Ruia College, Mumbai

Thermal and Mechanical Properties of EPR and XLPE Cable Compounds

Analyzing & Testing Business Unit. Tau-R Mode for Advanced DSC Analysis Applications Newsletter 1/ /09, Dr. Stefan Schmölzer

DSC Methods to Quantify Physical Aging and Mobility in Amorphous Systems: Assessing Molecular Mobility

Evolved gas analysis by simultaneous thermogravimetric differential thermal analysis-fourier transformation infrared spectroscopy (TG-DTA-FTIR)

It is the purpose of this paper to demonstrate that the detection limit is directly related to the sample size, for both TG and TG/FTIR.

ISO INTERNATIONAL STANDARD. Plastics Differential scanning calorimetry (DSC) Part 4: Determination of specific heat capacity

THERMO- GRAVIMETRIC ANALYZER TGA 1000

UNIT 11 DIFFERENTIAL THERMAL ANALYSIS, SCANNING CALORIMETRY AND THERMOMETRIC TITRATIONS

TAWN tests for quantitatively measuring the resolution and sensitivity of DSCs (version 2.1)

The THT micro reaction calorimeter RC

A Thermal Analysis Technique That Combines Differential Scanning Calorimetry and Light Microscopy

Thermal Analysis Premium

Thermal analysis unlocks the secrets of elastomers

DSC PT 10. Applications

MORPHOLOGY AND THERMAL STUDIES OF NICKEL CARBONATE NANOPARTICLES

Complete Materials Deformulation Using TGA-IR

Total analysis with DSC, TMA and TGA-EGA

1) INTRODUCTION 2) ROLE IN PREFORMULATION 3) CLASSIFICATION OF THERMAL ANALYSIS. 4) DIFFERENT METHODS OF THERMAL ANALYSIS.

Automated multi-vapor gravimetric sorption analyzer for advanced research applications

Scheme 1: Reaction scheme for the synthesis of p(an-co-mma) copolymer

Common Definition of Thermal Analysis

The THT micro reaction calorimeter μrc

Thermochimica Acta 531 (2012) Contents lists available at SciVerse ScienceDirect. Thermochimica Acta

Thermal Analysis Excellence

Chemically recyclable alternating copolymers with low polydispersity from

Studies on thermal degradation behavior of nano silica loaded cotton and polyester fabrics

SILMALAB is a laboratory of analysis for plastics industry that takes advantage of the experience gained by our engineers in 30 years of work and

Techniques in Thermal Analysis: Hyphenated Techniques, Thermal Analysis of the Surface, and Fast Rate Analysis

VOCs Emissions and Structural Changes of Polypropylene During Multiple Melt Processing

Heat Transfer in Polymers

Stabilizing vitamin D 3 by conformationally selective co-crystallization

Perseverance. Experimentation. Knowledge.

Morphology and Thermal studies of Copper Carbonate Nanoparticles

CuSO 4.5H 2 O Analysis with a Hyphenated TGA 8000 and Hiden Analytical HPR-20 EGA Mass Spectrometer

K. Ishikiriyamaa and B. Wunderlich*

POLYAMIDE-6,9 WITH CARBAZOLE

high performance OPTIC 3 and applications a total analytical solution

THEORY AND APPLICATIONS OF MODULATED TEMPERATURE PROGRAMMING TO THERMOMECHANICAL TECHNIQUES

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group

DATA ANALYSIS WITHOUT FOURIER TRANSFORMATION FOR SAWTOOTH-TYPE TEMPERATURE-MODULATED DSC

SIMULTANEOUS THERMAL ANALYSIS STA PT 1000 STA PT 1600 STA HP

7. Thermal Properties

CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter

Supplementary Information

INTERNATIONAL STANDARD

APPLICATION NOTE. Investigation of Alkali Salts with the STA 449 F5 Jupiter. Dr. Alexander Schindler and Dr. Michael Schöneich.

Calorimetric Principles and TAM III

Chem Page IX - 1 LAB MANUAL Differential Scanning Calorimetry 09_dsc131.docx EXPERIMENT IX

Thermal Analysis. Short Courses POLYCHAR 25 Kuala Lumpur. Copyright 2017 by Jean-Marc Saiter

Title: Bulk Thermal Stability Characterization via the SBAT Apparatus

DEVELOPMENT OF IMPROVED METHODS FOR CHARACTERISING THE CURE OF COMPOSITE MATERIALS

Buy1Get1 FREE. TA Instruments. The World Leader in Thermal Analysis and Rheology Systems

Differential Scanning Calorimeter Analysis of Hydrino-Producing Solid Fuel

Analyzing & Testing. Thermogravimetric Analysis TGA. Technique, Instrument, Applications TG 209 F1. Leading Thermal Analysis

Thermal and Mechanical Properties of EPR Cable Compound. Steven Boggs ICC Educational Session 7 November 2005 Phoenix, Arizona

TOPEM, a new temperature modulated DSC technique: A critical review

Differential Scanning Calorimetry

NP-180R. U.L. file number E98983

Anhydrous Proton Conductivities of Squaric Acid Derivatives

Physico-chemical characterization and comparison of fluorinated commercial Ski-Waxes.

5.2 Thermal analysis Thermal analysis

Thermal Analysis Excellence

Thermal Analysis measurements

Comprehensive Handbook of Calorimetry and Thermal Analysis

Thermal stability of azurite and malachite in relation to the formation of mediaeval glass and glazes

_ TA221 Brochure: Thermal Analysis & Rheology Systems Product Overview. _ TA006 Product Bulletin: Mechanical Cooling Accessory for TMA 2940

Thermal Analysis. Fundamentals and Applications to Polymer Science. Second Edition

Technical Data Sheet: Thermokinetics 3

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Simultaneous DSC-synchronous XRD for understanding solid-state phase transformations

Measurement techniques

Nucleation Behavior of Polypropylene with a Nan-O-Sil Additive Dr. Previn Shah, Rheoplast Associates

The solid-state stability of aspartame hemihydrate (APM) sweetener is important information for the food industry.

Thermal Analysis as an Important Research Tool for Colleges and Universities

Christin T. Choma TA Instruments, 109 Lukens Drive, New Castle, DE 19720, USA

The solid-state stability of aspartame hemihydrate (APM) sweetener is important information for the food industry.

Transcription:

DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS Problem A scientist working for a major pharmaceutical R&D center is having difficulties in interpreting the DSC results obtained on an excipient material. The results obtained on the compound are displayed in Figure 1 and the scientist is uncertain as to the assignment of the observed transitions. Solution To order to enhance the characterization of pharmaceutical materials, as well as many other types of samples, cyclic DSC and simultaneous TG/DTA provide two very powerful approaches. Differential scanning calorimetry (DSC) measures heat flow into or out of a sample as the material is either heated, cooled or held isothermally, or a combination of these. Cyclic DSC is performed by heating a sample, cooling it back to the desired starting temperature, and then reheating. A direct comparison can be made between the data obtained during the first and second heating segments to establish the nature of the transitions taking place within a sample as it is heated. Using cyclic DSC, events can be better identified as to whether they are reversible, such as Tg, or irreversible, such as the loss of water or crystallization. Cyclic DSC experiments can be performed relatively quickly, by heating at 20 C/min, cooling back at 40 C/min, and then reheating at 20 C/min. Advantages offered by cyclic DSC over temperature modulated DSC (TMDSC) are the time savings and ease of use. The maximum heating rate that can be used with TMDSC is 5 C/min. If a sample is heated from -40 to 250 C, the total time of the TMDSC experiment would be 58 minutes, at 5 C/min. Using cyclic DSC, as described above, the total time would only be 45 minutes. The Seiko EXSTAR DSC6000 offers the following major advantages for the study of pharmaceutical compounds: High sensitivity (the Seiko EXSTAR DSC6100 offers the highest sensitivity of any DSC instrument on the market) Very stable baseline performance for highly reproducible results Exceptional subambient performance for the study of Tg's near or below 0 C State-of-the-art add-on robotic accessory for unattended operation 1 of 6

20 point temperature calibration for unsurpassed accuracy 10 point enthalpic calibration for highly accurate heat capacities and heats of transition Thermogravimetric analysis (TGA) measures weight loss (or gain) as a sample is either heated or held under isothermal conditions. The data obtained via a TGA instrument can be further enhanced through the ability to simultaneously measure both the TGA weight loss and the heat flow response. The Seiko TG/DTA provides this powerful ability which greatly aids in the characterization of the properties of a sample. The Seiko EXSTAR TG/DTA6000 offers the following advantages: Simultaneous measurement of TGA and DTA signals Direct temperature calibration since thermocouple is in direct contact with sample pan True horizontal purging for optimal performance with coupled techniques such as TGA-FTIR or TGA-MS Ability to perform measurements in auto stepwise isothermal mode for unsurpassed resolution of severely overlapping weight loss events High sensitivity Stable baseline performance 20 point temperature calibration for the greatest accuracy across the entire temperature range (up to 1500 C) Direct conversion of DTA signal to DSC units (mw) via 10 point DSC enthalpic calibration Figure 1 2 of 6

Displayed in Figure 1 are the DSC results obtained on a pharmaceutical compound. Without further information about the sample, it is difficult to interpret the results. What is the cause of the large endothermic peak at 108 C? What is the nature of the small event at 180 C? Both cyclic DSC and TG/DTA can help in the assignments of the transitions and make data interpretation much easier. Figure 2 Displayed in Figure 2 are the complete cyclic DSC results obtained for the pharmaceutical compound. The plot shows the DSC heat flow and sample temperature as a function of time. In the two heating ramps, the sample is heated at a rate of 20 C/min while, during the cooling segment, the sample is cooled at a rate of 40 C/min. The temperature-time profile demonstrates the outstanding temperature control offered by the Seiko DSC using 22 bit, ultra high resolution A/D converter technology (US Patent 4734678). Three separate data files can be obtained from the cyclic DSC experiment: the 'asreceived' or 1st heat results; the 2nd heat data, which reflects the reversible aspects of the sample; and the subtracted file obtained by simply subtracting the 2nd heat data from the 1st heat results. The subtracted file represents the irreversible characteristics of the sample. 3 of 6

Figure 3 Displayed in Figure 3 are the three separate data sets for the pharmaceutical compound obtained using cyclic DSC. The lower most curve represents the DSC results for the 'as received' sample. The upper most trace is that of the 2nd heating segment and this represents the reversible aspects. The center curve is the subtracted data set and reflects the irreversible aspects of the material. The cyclic DSC results reveal that the endothermic event at 108 C is irreversible in nature and is most likely due to the evolution of moisture from the sample. This can be easily confirmed using TGA. The 2nd heat results show that the small transition observed near 180 C is actually a glass transition event. A classic stepwise change in heat flow is clearly observed in the 2nd heat data set confirming that the transition at 180.5 C is a Tg with Figure 4 4 of 6

a change in heat capacity, Cp, of 0.249 mj/mg deg. The conclusions made using cyclic DSC were tested for accuracy using the Seiko TG/DTA and these results are displayed in Figure 4. This figure shows the TGA weight loss (%) and the simultaneous DTA data which was automatically converted to DSC heat flow units (mw). The TGA/DSC results confirm that the DSC endothermic event observed in the vicinity of 108 C is indeed due to the loss of volatiles, most likely water given the temperature. The amount of water is found to be 13.8%, based on the TGA weight loss data. The simultaneous TGA/DSC results show that the TGA weight loss corresponds exactly to the DSC endothermic peak. The main decomposition of the pharmaceutical compound is observed as a large mass loss event in the TGA signal with an onset temperature of 419.8 C. The total mass loss is 82.6% due to the thermal degradation of the sample. Figure 5 The simultaneous DTA-DSC signal associated with the Seiko TG/DTA has enough sensitivity to detect the small glass transition event associated with the pharmaceutical compound. Displayed in Figure 5 is an enlarged view of the DSC and TGA signals in the region near Tg. Since the sample does not lose mass in this region, the TGA data is flat while the simultaneous DSC signal does exhibit the small change in heat flow associated with the sample's glass transition event. The Tg as obtained using the Seiko TG/DTA agrees very well with that obtained via the dedicated Seiko DSC instrument (180.7 C from TG/DTA and 180.5 C from DSC). 5 of 6

Summary Cyclic DSC experiments can be performed on materials to better aid in the identification of transitions observed while heating. Cyclic DSC experiments can be conducted relatively quickly and three data sets are obtained: 1st heat or the as received results; the 2nd heat data, reflecting reversible aspects and making it easier to see glass transition events; and the subtracted data set which reflects irreversible aspects of the sample. TG/DTA further aids in the analysis of a material by helping in the assignment of irreversible events, such as the loss of volatiles. The simultaneous measurement of TGA and DTA signals makes the assignment of the transitions much less ambiguous. The data is further enhanced using the automatic conversion of the DTA signal to user friendly DSC units. 6 of 6