Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv)

Similar documents
Measurements of protein electrophoretic mobility using the Zetasizer Nano ZSP

The ultimate in desktop particle characterization

The ultimate in desktop particle characterization

The Diffusion Barrier Technique, Practical Aspects and Data interpretation

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

ZS90. The ultimate in desktop particle characterization. Particle size. Zeta potential. Molecular weight

10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS

Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Electrophoretic Light Scattering Overview

Use of DLS/Raman to study the thermal unfolding process of lysozyme

Introduction to the calculators in the Zetasizer software

Studying the Effect of Steric Layer Thickness on Emulsion Stability

ZETASIZER NANO Series PERFORMANCE, SIMPLICITY, VERSATILITY

REVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID.

HIGH PERFORMANCE PARTICLE SIZER THE FIRST SYSTEM TO COMBINE HIGH SENSITIVITY WITH A HIGH CONCENTRATION CAPABILITY

Using the Zetasizer Nano to optimize formulation stability

Nanobubble Applications and Characterization by Nanoparticle Tracking Analysis

Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications

DEVELOPING A BIOFORMULATION STABILITY PROFILE

Component specific particle characterisation of the active components in a pharmaceutical topical formulations

Litesizer Sheet 1. [Home]

Differentiation of polymer branching and composition using the Mark Houwink plot

next generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

The power of multi-detector SEC in the analysis of antibodies

GPC/SEC Analysis of Polymer Degradation

Overcoming and quantifying Wall Slip in measurements made on a rotational rheometer

Zeta potential - An introduction in 30 minutes

Understanding the conformational stability of protein therapeutics using Raman spectroscopy

Executive Summary. Introduction WHITEPAPER

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy

Amended Specifications as per the Pre Bid meeting held on in Chapter 4

Page 1 of 5. Is it alright to estimate dñ/dc in SLS measurements?

Characterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy

How to use GPC/SEC for compositional analysis

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

Introduction to Dynamic Light Scattering for Particle Size Determination

Particle Characterization Laboratories, Inc.

Measuring sub nanometre sizes using dynamic light scattering

Inform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years

How switching mobile phases can improve your OMNISEC results

Scattering intensity fluctuations

Determining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer

Surface Zeta Potential Cell (ZEN1020)

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement :

ENV/JM/MONO(2015)17/PART1/ANN2

Measuring NP Aggregation Propensities

The Coulter Principle for Outlier Detection in Highly Concentrated Solutions

Zeta Potential Analysis using Z-NTA

Particle Concentration Particle Size Zeta Potential

Using runt Disposable Cuvettes

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector

K + -Selective Optical Microsensors Based On Surface Modified Polystyrene Microspheres

South pacific Journal of Technology and Science

Accessory Publication

Nanoparticle Analyzer

Using rheology to study the hardness and spreadability of butter

Size and Zeta Potential of Colloidal Gold Particles

CHAPTER TWO: EXPERIMENTAL AND INSTRUMENTATION TECHNIQUES

Fabrication Facility. 1. Turn on the. the power. For. - polystyrene. - Do. -Quartz quivette. quivette -Low

Bohlin. Rheological instruments backed with rheological experience. Rheological properties

Supplementary Information

Particle Analysis at the Touch of a Button. Litesizer series

Particle Size Analysis with Differential Centrifugal Sedimentation. Marc Steinmetz Sales & Support Manager CPS Instruments Europe

200 to +200mV. Particle Diameter Measurement Range 0.3nm to 8µm. Zeta Potential Measurement. Molecular weight to g/mol

Method validation for laser diffraction measurements

YOUR PARTNER IN PARTICLE CHARACTERIZATION

ELECTRONIC SUPPLMENTARY INFORMATION (ESI) Doubly crosslinked microgel-colloidosomes: a versatile

Particle Analysis at the Touch of a Button. Litesizer series

PARTICLE SIZE CHARACTERISATION OF AGRICULTURAL SPRAYS USING LASER DIFFRACTION

Evaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications

An introduction to particle size characterisation by DCS:

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison

Supporting Information for

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

Supracolloidal Polymer Chains of Diblock Copolymer Micelles

An Introductions to Advanced GPC Solutions

How DLS Works: Interference of Light

Electronic Supplementary Information. Selective detection of Al 3+ and citric acid with a fluorescent amphiphile

Meet Stunner: The one-shot protein concentration and sizing combo

Advances in particle concentration measurements

Supporting Information

CP Chapter 15/16 Solutions What Are Solutions?

X-CELL Analytical. Petroleum Fluids Analysis. St. Francis Xavier University Analytical Services Lab. Equipment Specification Brochure

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Meet Our Uncle: 12 Stability Applications on One Platform

Interaction of Gold Nanoparticle with Proteins

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

Measuring particle aggregation rates by light scattering

Particle Tracking on. Exosomes. ZetaView. Multiparameter NTA sizing, counting and zeta potential. C. Helmbrecht and H. Wachernig

of Polystyrene 4-arm Stars Synthesized by RAFT- Mediated Miniemulsions.

Holographic Characterization of Protein Aggregates

Supporting Information

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

UV-vis Analysis of the Effect of Sodium Citrate on the Size and the Surface Plasmon Resonance of Au NPs. Eman Mousa Alhajji

Measuring Zeta Potential

Suspension Stability; Why Particle Size, Zeta Potential and Rheology are Important

Introduction to Zeta Potential Measurement with the SZ-100

Transcription:

Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv) Contents Size using Non-Invasive back-scatter (NIBS) optics Zetasizer Nano S/ZS...2 Minimum size specification (d.nm), Peak mode 0.6nm, Minimum size 0.3nm...2 Maximum size specification (d.nm), Peak mode 8.9 microns, Maximum size 10.0 microns...2 Size using 90 degree optics Zetasizer Nano S90/ZS90...3 Minimum size specification (d.nm), Peak mode 0.6nm, Minimum size 0.3nm...3 Maximum size specification (d.nm), Peak mode 3.0 microns, Maximum size 5.0 microns...3 Zetasizer APS...4 Minimum size specification (r.nm), Peak mode 0.3nm, Minimum size 0.15nm...4 Maximum size specification (r.nm), Peak mode 0.5 microns, Maximum size 1.0 microns...4 Zetasizer µv...5 Minimum size specification (r.nm), Peak mode 0.3nm, Minimum size 0.15nm...5 Maximum size specification (r.nm), Peak mode 0.5 microns, Maximum size 1.0 microns...5 Temperature performance...6 o Temperature control (+/- 0.1 C)...6 Concentration performance...7 Minimum concentration specification for size, 0.1ppm...7 Minimum concentration specification for size, 0.1mg/mL lysozyme...7 Maximum concentration range for size 40wt%...8 Zeta potential Zetasizer Nano Z/ZS/ZS90...9 Minimum particle size for zeta potential measurement (d.nm) 3.8nm...9 Maximum particle size for zeta potential measurement (d.nm) 100.0 microns...9 Maximum mobility range: > +/- 20µ.cm/V.s (no effective maximum)...10 Maximum zeta potential range: > +/- 500mV (no effective limitations)...10 Maximum sample conductivity for zeta potential measurement 200mS/cm...10 Maximum sample concentration zeta potential measurement, Zetasizer Nano Z/ZS 40%w/v...10 Molecular Weight Zetasizer Nano S/ZS...11 Minimum absolute molecular weight 980Da...11 Molecular Weight Zetasizer Nano S90/ZS90...12 Minimum absolute molecular weight 9,800Da...12 Molecular Weight Zetasizer μv...13 Minimum absolute molecular weight 980Da...13 7 Maximum absolute molecular weight 2 x 10 Da Zetasizer series...14 Introduction This document summarises sample measurements performed on the Zetasizer Nano series, the Zetasizer APS and the Zetasizer µv. These measurements were designed to test the performance of the Zetasizer systems using careful sample preparation and measurement design. 1 Zetasizer technical note MRK1515-01

Size using Non-Invasive back-scatter (NIBS) optics Zetasizer Nano S/ZS All Zetasizer units provide a Z-average figure which is the intensity weighted mean hydrodynamic size of the ensemble collection of particles measured by dynamic light scattering (DLS). Many samples are polydisperse a range of material sizes and as such to achieve distribution measurements it is important that the system can monitor sizes beyond the range of the peak mode. Minimum size specification (d.nm), Peak mode 0.6nm, Minimum size 0.3nm Figure 1 shows an overplot of 3 measurements of sucrose, at a concentration of 30%w/v. The peak had a mode of 0.6nm. To characterise this size, the system has a capability of measuring the size of the distribution down to 0.3nm Figure 1: Verifying the minimum size on the Zetasizer Nano S/ZS Maximum size specification (d.nm), Peak mode 8.9 microns, Maximum size 10.0 microns. Figure 2 shows an overplot of three measurements of an 8.9 micron polystyrene latex, dispersed in 13%w/v sucrose. This dispersant is chosen to density match the latex to prevent sedimentation. This demonstrates the capability of the Zetasizer Nano S and ZS to measure particles with a size mode of 8.9 microns. To characterise this size, the system has a capability of measuring the size of the distribution up to 10.0 microns Figure 2: Verifying the maximum size on the Zetasizer Nano S/ZS 2 Zetasizer technical note MRK1515-01

Size using 90 degree optics Zetasizer Nano S90/ZS90 All Zetasizer units provide a Z-average figure which is the intensity weighted mean hydrodynamic size of the ensemble collection of particles measured by dynamic light scattering (DLS). Many samples are polydisperse a range of material sizes and as such to achieve distribution measurements it is important that the system can monitor sizes beyond the range of the peak mode. Minimum size specification (d.nm), Peak mode 0.6nm, Minimum size 0.3nm Figure 3 shows an overplot of 3 measurements of sucrose, at a concentration of 20%w/v. The peak had a mode of 0.6nm. To characterise this size, the system has a capability of measuring the size of the distribution down to 0.3nm. Figure 3: Verifying the minimum size on the Zetasizer Nano S90/ZS90 Maximum size specification (d.nm), Peak mode 3.0 microns, Maximum size 5.0 microns Figure 4 shows an overplot of three measurements of a 3.0 micron polystyrene latex, dispersed in 13%w/v sucrose. This dispersant is chosen to density match the latex to prevent sedimentation. This demonstrates the capability of the Zetasizer Nano S and ZS to measure particles with a size mode of 3.0 microns. To characterise this size, the system has a capability of measuring the size of the distribution up to 5.0 microns Figure 4: Verifying the maximum size on the Zetasizer Nano S90/ZS90 3 Zetasizer technical note MRK1515-01

Zetasizer APS All Zetasizer units provide a Z-average figure which is the intensity weighted mean hydrodynamic size of the ensemble collection of particles measured by dynamic light scattering (DLS). Many samples are polydisperse a range of material sizes and as such to achieve distribution measurements it is important that the system can monitor sizes beyond the range of the peak mode. Minimum size specification (r.nm), Peak mode 0.3nm, Minimum size 0.15nm Figure 5 shows an overplot of 3 repeat measurements of sucrose, at a concentration of 20%w/v. The peak had a mode of 0.3nm. To characterise this size, the system has a capability of measuring the size of the distribution down to 0.15nm Figure 5: Verifying the minimum size on the Zetasizer APS Maximum size specification (r.nm), Peak mode 0.5 microns, Maximum size 1.0 microns Figure 6 shows the measurement of a 0.5 micron radius polystyrene latex, dispersed in 13% w/v sucrose. This dispersant was chosen to density match the latex to prevent sedimentation. This demonstrates the capability of the Zetasizer APS to measure particles with a size mode of 0.5 microns. To characterise this size, the system has a capability of measuring the size of the distribution up to 1.0 micron. Figure 6: Verifying the maximum size on the Zetasizer APS 4 Zetasizer technical note MRK1515-01

Zetasizer µv All Zetasizer units provide a Z-average figure which is the intensity weighted mean hydrodynamic size of the ensemble collection of particles measured by dynamic light scattering (DLS). Many samples are polydisperse a range of material sizes and as such to achieve distribution measurements it is important that the system can monitor sizes beyond the range of the peak mode. Minimum size specification (r.nm), Peak mode 0.3nm, Minimum size 0.15nm Figure 7 shows an overplot of 3 measurements of sucrose, at a concentration of 15%w/v. The peak had a mode of 0.3nm. To characterise this size, the system has a capability of measuring the size of the distribution down to 0.15nm. Figure 7: Verifying the minimum size on the Zetasizer μv Maximum size specification (r.nm), Peak mode 0.5 microns, Maximum size 1.0 microns Figure 8 shows an overplot of three measurements of a 0.5 micron radius polystyrene latex, dispersed in 13%w/v sucrose. This dispersant was chosen to density match the latex to prevent sedimentation. This demonstrates the capability of the Zetasizer µv to measure particles with a size mode of 0.5 microns. To characterise this size, the system has a capability of measuring the size of the distribution up to 1.0 micron. Figure 8: Verifying the maximum size on the Zetasizer μv 5 Zetasizer technical note MRK1515-01

Temperature performance Temperature control (+/- 0.1 o C) It is difficult to measure the temperature of a sample without affecting the results. In design and production, specialist equipment is used to measure the temperature at the extremes of performance, and also the rate of change of temperature. An alternative way of assessing temperature control and stability is to monitor the size of a known standard with which temperature deviation will be picked up by size variation. To test the Zetasizer a 60nm latex standard was measured three times in 10mM NaCl solution at a temperature of 20 o C and the result recorded. This initial measurement was 63.43nm diameter with a SD of 0.14nm which is within the accepted tolerance limits. The temperature was then reduced in steps of 5 o C down to 0 o C with three measurements taken at each step. If the sample did not achieve the set temperature then you would expect to see variations in the measured size. The sizes measured at the different temperatures are shown in figure 9 and they indicate that the Zetasizer is capable of achieving set temperatures within +/-0.1 o C. Note: This relies on the viscosity being set correctly, and in the Zetasizer, for water, this is set automatically 69 67 Z-Ave size Vs Temperature º C Figure 9: Verifying the temperature stability of the Zetasizer platforms Z-Ave sizw (d.nm) 65 63 61 59 57 55 0 5 10 15 20 25 Degrees Celcius 6 Zetasizer technical note MRK1515-01

Concentration performance Minimum concentration specification for size, 0.1ppm The minimum concentration is specified using a 60nm latex. Figure 10 shows an overplot of 5 measurements of a 60nm latex, diluted in 10mM NaCl. The average size recorded was 63.7nm, with a standard deviation of 0.19nm. Figure 10: Verifying the minimum concentration range of 0.1ppm for the Zetasizer Nano S/ZS Minimum concentration specification for size, 0.1mg/mL lysozyme Figure 11 shows a plot of the size of peak 1 of measurements of lyzozyme as a function of concentration. Each point shows an error bar which is one standard deviation of a set of 5 repeat measurements. At 0.1mg/mL, the error in the measurement of size is larger than at higher concentrations but the mean value is in the expected size range for lysozyme. Figure 11: Verifying the minimum concentration range in of 0.1mg/mL lysozyme for the Zetasizer Nano S/ZS 7 Zetasizer technical note MRK1515-01

Maximum concentration range for size 40wt% Figure 12 shows an overplot of 3 measurements of a nominally 100nm silica at a concentration of 40%w/v, demonstrating the repeatability of measurement at this concentration. Figure 12: Verifying the maximum concentration range at 40%w/v 8 Zetasizer technical note MRK1515-01

Zeta potential Zetasizer Nano Z/ZS/ZS90 Minimum particle size for zeta potential measurement (d.nm) 3.8nm The measurement capability of the Zetasizer Nano was tested using a solution of lysozyme. This molecule has a hydrodynamic diameter of 3.8nm when dissolved in a salt solution. Figure 13 shows the mobility measurement of 3 aliquots of sample, each measurement repeated 7 times. The lysozyme was prepared at two concentrations, 5mg/mL and 20mg/mL dispersed in 100mM KCl with 10mM BisTris. The average result of each of the aliquots is 0.7 +/- 0.1 mobility units, or +9mV in terms of zeta potential. This demonstrates the repeatability of the result at this size, and is in agreement with the referenced published literature value plotted as the black line. Figure 13: Verifying the minimum particle size for zeta potential measurement Maximum particle size for zeta potential measurement (d.nm) 100.0 microns To measure the maximum particle size for zeta potential, a 97 micron polystyrene latex sample was used (figure 14). The sample was dispersed in 13%w/v sucrose to density match the latex. This was done to in order to do a number of measurements before the particles sedimented from the measurement volume, to show measurement repeatability. Figure 14: Verifying the maximum particle size for zeta potential measurement 9 Zetasizer technical note MRK1515-01

Maximum mobility range: > +/- 20µ.cm/V.s (no effective maximum) The optical and electronic design of the Zetasizer Nano series limits the mobility of particles that can be measured to 3,460 mobility units (µm.cm/v.s). In practice, the highest mobility measured of a particle in an aqueous dispersion is about 10 mobility units. This capability then counts as no effective maximum. For specification purposes a value is declared of > +/- 20 µm.cm/v.s Maximum zeta potential range: > +/- 500mV (no effective limitations) The range of mobilities that can be measured by the Zetasizer, for samples in aqueous systems is equivalent to a zeta potential of +/- 44,300mV, and even higher for samples in non-aqueous dispersants. For samples with a mobility of 10µm.cm/V.s in an aqueous dispersant at 25ºC, the zeta potential calculated using the Smoluchowski limit is 128mV. This capability counts as no effective maximum. For specification purposes a value is declared of > +/- 500mV Maximum sample conductivity for zeta potential measurement 200mS/cm Figure 15 shows 10 consecutive measurements of 30nm Ludox, ph 10 adjusted to a conductivity of 209mS/cm using sodium chloride. A 60 second pause was used between each measurement. Figure 15: Verifying the maximum conductivity of 200mS/cm for zeta potential measurement Maximum sample concentration zeta potential measurement, Zetasizer Nano Z/ZS 40%w/v Figure 16 shows a plot of zeta potential as a function of concentration for an aqueous polyurethane dispersion with the addition of 5% triethylamine, demonstrating the ability to measure samples up to 40%w/v Figure 16: Verifying the maximum concentration of 40%w/v for zeta potential measurement 10 Zetasizer technical note MRK1515-01

Molecular Weight Zetasizer Nano S/ZS The absolute molecular weight is measured using a Debye plot. This requires the measurement of the intensity of a scatterer with a known Rayleigh ratio, usually toluene, followed by the measurement of the scattering intensity of a number of known concentrations of a solution of the sample. This requires that the system is sensitive enough to detect the excess scattering from the sample over the dispersant, and that the count rate is exceptionally stable. Minimum absolute molecular weight 980Da Figure 17 shows a Debye plot using three concentrations of a 980 Dalton polystyrene standard, showing the capability of the Zetasizer Nano S and ZS to measure the molecular weight. Figure 17: Verifying the minimum molecular weight for the Zetasizer Nano S/ZS 11 Zetasizer technical note MRK1515-01

Molecular Weight Zetasizer Nano S90/ZS90 The absolute molecular weight is measured using a Debye plot. This requires the measurement of the intensity of a scatterer with a known Rayleigh ratio, usually toluene, followed by the measurement of the scattering intensity of a number of known concentrations of a solution of the sample. This requires that the system is sensitive enough to detect the excess scattering from the sample over the dispersant, and that the count rate is exceptionally stable. Minimum absolute molecular weight 9,800Da Figure 18 shows a Debye plot using three concentrations of a 9,800 Dalton polystyrene standard, showing the capability of the Zetasizer Nano S90 and ZS90 to measure the molecular weight. Figure 18: Verifying the minimum molecular weight for the Zetasizer Nano S90/ZS90 12 Zetasizer technical note MRK1515-01

Molecular Weight Zetasizer μv The absolute molecular weight is measured using a Debye plot. This requires the measurement of the intensity of a scatterer with a known Rayleigh ratio, usually toluene, followed by the measurement of the scattering intensity of a number of known concentrations of a solution of the sample. This requires that the system is sensitive enough to detect the excess scattering from the sample over the dispersant, and that the count rate is exceptionally stable. Minimum absolute molecular weight 980Da Figure 19 shows a Debye plot using three concentrations of a 980 Dalton polystyrene standard, showing the capability of the Zetasizer µv to measure the molecular weight. Figure 19: Verifying the minimum molecular weight for the Zetasizer µv 13 Zetasizer technical note MRK1515-01

Maximum absolute molecular weight 2 x 10 7 Da Zetasizer series The molecular weight of protein solutions can be measured on the Zetasizer Nano and the Zetasizer µv using a Debye plot. This involves measurement of a number of known concentrations of a solution of molecules at the scattering angle of the system. The error in using a single angle for measurement for a globular protein, rather than using a multi angle system, can be calculated theoretically. The specification is then set at an acceptable level of error. For the Zetasizer Nano S/ZS there is a maximum 5% error for molecules with an Rg of 27nm (figure 20). From known relationships this is equivalent to a globular protein with a molecular weight of 20 million Daltons. For the Zetasizer Nano S90/ZS90 and Zetasizer µv, the errors are much lower, hence the upper specification could be higher, however the specification is left at 20 million Daltons as this is sufficient to cover all practical samples. Figure 20: Verifying the maximum molecular weight of the Zetasizer series Malvern Instruments Ltd Enigma Business Park Grovewood Road Malvern Worcestershire UK WR14 1XZ Tel: +44 (0)1684 892456 Fax: +44 (0)1684 892789 Malvern Instruments Worldwide Sales and service centers in over 50 countries for details visit www.malvern.com/contact Malvern Instruments Ltd 2010 more information at www.malvern.com 14 Zetasizer technical note MRK1515-01