Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori

Similar documents
The NMR Inverse Imaging Problem

MRI Physics I: Spins, Excitation, Relaxation

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Basic MRI physics and Functional MRI

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

The Basics of Magnetic Resonance Imaging

Contrast Mechanisms in MRI. Michael Jay Schillaci

Nuclear Magnetic Resonance Imaging

Introduction to Biomedical Imaging

Biomedical Imaging Magnetic Resonance Imaging

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging

Principles of Magnetic Resonance Imaging

K-space. Spin-Warp Pulse Sequence. At each point in time, the received signal is the Fourier transform of the object s(t) = M( k x

Fundamental MRI Principles Module Two

Apodization. Gibbs Artifact. Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2013 MRI Lecture 5. rect(k x )

Nuclear Magnetic Resonance Imaging

Basis of MRI Contrast

Chapter 14:Physics of Magnetic Resonance

Field trip: Tuesday, Feb 5th

Introduction to MRI Acquisition

Magnetic Resonance Imaging in a Nutshell

A Hands on Introduction to NMR Lecture #1 Nuclear Spin and Magnetic Resonance

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES

Physical fundamentals of magnetic resonance imaging

Magnetic Resonance Imaging (MRI)

Physics of MR Image Acquisition

Introductory MRI Physics

Basic p rinciples COPYRIGHTED MATERIAL. Introduction. Atomic s tructure

MR Fundamentals. 26 October Mitglied der Helmholtz-Gemeinschaft

The Theory of Nuclear Magnetic Resonance Behind Magnetic Resonance Imaging. Catherine Wasko Physics 304 Physics of the Human Body May 3, 2005

NMR and MRI : an introduction

ELECTRON SPIN RESONANCE & MAGNETIC RESONANCE TOMOGRAPHY

Magnetic resonance imaging MRI

NMR/MRI examination (8N080 / 3F240)

Principles of MRI. Vinyl Record. Last time: Today: Homework Due tonight! EE225E / BIO265. Transforms a temporal signal to a spatial signal

Chemistry 431. Lecture 23

The Physical Basis of Nuclear Magnetic Resonance Part I ESMRMB. Jürgen R. Reichenbach

MRI Physics II: Gradients, Imaging. Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor

Lecture 12 February 11, 2016

NMR, the vector model and the relaxation

Topics. Spin. The concept of spin Precession of magnetic spin Relaxation Bloch Equation

Measuring Spin-Lattice Relaxation Time

Advanced Topics and Diffusion MRI

Tissue Characteristics Module Three

Spectral Broadening Mechanisms

BME I5000: Biomedical Imaging

Technical University of Denmark

BMB 601 MRI. Ari Borthakur, PhD. Assistant Professor, Department of Radiology Associate Director, Center for Magnetic Resonance & Optical Imaging

Sketch of the MRI Device

Introduction to the Physics of NMR, MRI, BOLD fmri

Exam 8N080 - Introduction to MRI

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1

Outlines: (June 11, 1996) Instructor:

MRI in Review: Simple Steps to Cutting Edge Part I

Velocity Images. Phase Contrast Technique. G. Reiter 1,2, U. Reiter 1, R. Rienmüller 1

10.4 Continuous Wave NMR Instrumentation

Chapter 7. Nuclear Magnetic Resonance Spectroscopy

NMR Imaging in porous media

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR

Basic principles COPYRIGHTED MATERIAL. Introduction. Introduction 1 Precession and precessional

Relaxation times in nuclear magnetic resonance

Topics. The History of Spin. Spin. The concept of spin Precession of magnetic spin Relaxation

Spin. Nuclear Spin Rules

Part III: Sequences and Contrast

Spectral Broadening Mechanisms. Broadening mechanisms. Lineshape functions. Spectral lifetime broadening

Bioengineering 278" Magnetic Resonance Imaging" Winter 2010" Lecture 1! Topics:! Review of NMR basics! Hardware Overview! Quadrature Detection!

June 16, Signal generation and gradient fields in MRI. Maximilian Oehm. Summary of physical fundamentals. Motivation. Complex representation

Me myself and MRI: adventures in not understanding nuclear physics.

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

The physics US and MRI. Prof. Peter Bogner

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

1 Magnetism, Curie s Law and the Bloch Equations

Part II: Magnetic Resonance Imaging (MRI)

FREQUENCY SELECTIVE EXCITATION

Functional Magnetic Resonance Imaging (FMRI) is an imaging technique for

Low Field MRI of Laser Polarized Noble Gases. Yuan Zheng, 4 th year seminar, Feb, 2013

Nuclei, Excitation, Relaxation

BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD

MR Advance Techniques. Flow Phenomena. Class I

Lecture #6 NMR in Hilbert Space

NMR Spectroscopy Laboratory Experiment Introduction. 2. Theory

Chem 325 NMR Intro. The Electromagnetic Spectrum. Physical properties, chemical properties, formulas Shedding real light on molecular structure:

Spin. Nuclear Spin Rules

Magnetic Resonance Imaging

On Signal to Noise Ratio Tradeoffs in fmri

Lecture 21. Nuclear magnetic resonance

V27: RF Spectroscopy

EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 9 Solutions. Due April 29th, 2013

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

How does this work? How does this method differ from ordinary MRI?

Medical Imaging Physics Spring Quarter Week 9-1

MRS: IN VIVO SPECTROSCOPIC IMAGING MAIN POINTS

Physical Background Of Nuclear Magnetic Resonance Spectroscopy

Tissue Parametric Mapping:

Magnetic Resonance Imaging in Medicine

NMR BMB 173 Lecture 16, February

Transcription:

Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori Enseignant-chercheur Equipe IMAGES - Télécom ParisTech pietro.gori@telecom-paristech.fr September 20, 2017 P. Gori BIOMED 20/09/2017 1 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 2 / 76

Magnetic Resonance Imaging It is a medical imaging technique and diagnostic tool in radiology based on the Nuclear Magnetic Resonance (NMR). Main pros non-invasive and 3D no ionizing radiations very good spatial resolution (1mm isotropic for anatomy) anatomy, functional and physiological state of the internal organs both healthy and pathological tissues Main cons expensive long scan time (20-45 minutes) no metal (peacemaker, valves, etc.) not comfortable for the patient and loud P. Gori BIOMED 20/09/2017 3 / 76

Magnetic Resonance Imaging Figure 1: Magnetic resonance images taken from the same anatomical section of the human brain. Contrast between different tissues can change simply by varying some parameters of the acquisition such as the Repetition Time (TR) and the Echo Time (TE). Image taken from [1]. P. Gori BIOMED 20/09/2017 4 / 76

MRI scanner It is composed of: a magnet which produces a very powerful uniform magnetic field B 0 (1.5T - 11T, note that the earth magnetic field is around 60 microt) shim coils to make B 0 homogeneous gradient coils that make the magnetic field linearly vary across the imaging volume. This determines the plane of imaging Radio Frequency (RF) transmission system P. Gori BIOMED 20/09/2017 5 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 6 / 76

Nuclear Magnetic Resonance (NMR) A sample is placed in a large uniform magnetic field B 0 An oscillating current is applied to the coil for few milliseconds, which produces an oscillating magnetic field B 1 in the sample (Jefimenko s equations). The oscillations of the magnetic field are in the radiofrequency (RF) range (3 khz to 300 GHz) P. Gori BIOMED 20/09/2017 7 / 76

Nuclear Magnetic Resonance (NMR) Certain atomic nuclei, such as the hydrogen, are able to absorb specific radio frequency energy and emit a small portion of that energy. The produced time-varying magnetic field will induce voltage in the coil (Maxwell Faraday equation) This current, oscillating at the same frequency as the RF pulse, is the NMR signal P. Gori BIOMED 20/09/2017 8 / 76

Recap about Physics The atomic nucleus is the small, dense region consisting of protons and neutrons at the center of an atom. Spin is an intrinsic form of angular momentum carried by the atomic nuclei. Its magnitude cannot be changed, only its direction can vary. Only nuclei with an odd number of protons and neutrons have a net spin, such as hydrogen (only one proton). Nuclei with spin can be seen as tiny magnets, whose south-north axis is parallel to the spin axis. They possess a magnetic dipole moment, namely a vector µ which points from the south to north pole of the magnet. It is described by: τ = µ x B 0 = γl x B 0 (1) where τ is the torque acting on the magnet, γ is the gyromagnetic ratio and L is the angular momentum. P. Gori BIOMED 20/09/2017 9 / 76

Recap about Physics Remember that τ = dl dt Thus τ = dl dt = Lsin(θ)dθ dt = γlb 0 sin(θ) (2) dθ dt = ω L = γb 0 f L = γb 0 2π (3) This is the precession angular frequency, namely the Larmor frequency. P. Gori BIOMED 20/09/2017 10 / 76

Nuclear Magnetic Resonance (NMR) This means that when a nucleus with spin in placed into a magnetic field B 0, the magnetic field will exert a torque on the nucleus making it process (rotate) around the field. Resonant frequency of NMR The Larmor frequency f L is the resonant frequency of NMR and it is directly proportional to the magnitude of B 0 (f L = γb 0 2π ). Magnetic resonance absorption will only occur at this frequency! P. Gori BIOMED 20/09/2017 11 / 76

Nuclear Magnetic Resonance (NMR) Nucleus γ (MHz/T) 1 H 42.58 13 C 10.71 This means that for a B 0 = 1.5T, if we put a 1 H nucleus inside, we will obtain ω L = 63, 86MHz. Why 1 H? P. Gori BIOMED 20/09/2017 12 / 76

Nuclear Magnetic Resonance (NMR) Nucleus γ (MHz/T) 1 H 42.58 13 C 10.71 This means that for a B 0 = 1.5T, if we put a 1 H nucleus inside, we will obtain ω L = 63, 86MHz. Why 1 H? Because the human body is composed of 50-65% of water (H 2 0)! P. Gori BIOMED 20/09/2017 12 / 76

Nuclear Magnetic Resonance (NMR) Nucleus γ (MHz/T) 1 H 42.58 13 C 10.71 This means that for a B 0 = 1.5T, if we put a 1 H nucleus inside, we will obtain ω L = 63, 86MHz. Why 1 H? Because the human body is composed of 50-65% of water (H 2 0)! Now look at ω L, what do you notice? P. Gori BIOMED 20/09/2017 12 / 76

Nuclear Magnetic Resonance (NMR) Nucleus γ (MHz/T) 1 H 42.58 13 C 10.71 This means that for a B 0 = 1.5T, if we put a 1 H nucleus inside, we will obtain ω L = 63, 86MHz. Why 1 H? Because the human body is composed of 50-65% of water (H 2 0)! Now look at ω L, what do you notice? ω L is in the RF range! This is why the oscillations of B 1 are in the RF range. P. Gori BIOMED 20/09/2017 12 / 76

Nuclear Magnetic Resonance (NMR) To sum up: Spinning hydrogens align with B 0 and rotate around it at f L We apply oscillations (B 1 ) in the RF range which are absorbed by the hydrogen nuclei The frequency of these oscillations depends on the field strength B 0 and on the type of nucleus (γ) P. Gori BIOMED 20/09/2017 13 / 76

Nuclear Magnetic Resonance (NMR) To sum up: Spinning hydrogens align with B 0 and rotate around it at f L We apply oscillations (B 1 ) in the RF range which are absorbed by the hydrogen nuclei The frequency of these oscillations depends on the field strength B 0 and on the type of nucleus (γ) Absorbed?? P. Gori BIOMED 20/09/2017 13 / 76

Nuclear Magnetic Resonance (NMR) A spin is at its lowest energy when it is aligned with B 0 and at its highest energy when it is opposite to B 0 It seems that a 1 H spin can only have two pure states in the presence of B 0 Quantum Physics!! P. Gori BIOMED 20/09/2017 14 / 76

Quantum Physics The spin of 1 H is equal to 1 2 and it can only have two possible measurable states (parallel or anti-parallel to B 0 ). However, without B 0 the two spin states are not measurable! They are in a weighted superposition of both states Zeeman effect: E = γ h 2π B 0 where h is the Planck s constant and E is the energy gap between two nuclear spin states Planck Einstein relation: E = hf The transfer of energy is quantized! P. Gori BIOMED 20/09/2017 15 / 76

Quantum Physics E = γ h 2π B 0 = hf f L = γb 0 2π (4) This is another way to obtain the Larmor equation! So, the difference in energy between the two states is quantized and it depends on f L This means that when a spin goes from its higher energy level (anti-parallel) to its lower energy level (parallel) there is an emission of a photon, whose energy is hf L And viceversa, an increase in energy results from absorption of a photon, whose energy must always be hf L P. Gori BIOMED 20/09/2017 16 / 76

Quantum Physics - Boltzmann distribution p + = N + = exp( (E + E ) ) (5) p N kt where p + = N + N and p = N N are the probabilities of state parallel and anti-parallel, N + and N are the number of particles in state + and, N = N + + N, k is the Boltzmann constant and T is the absolute temperature in Kelvin. At room/body temperature (300 K, red curve), due to thermal collisions and spin-spin interactions, the distribution of low and high energy spins is almost the same: = 1, 0000007!! (Parallel though...) N + N P. Gori BIOMED 20/09/2017 17 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 18 / 76

Net magnetization The net magnetization vector M is the sum of the magnetic moments (µ) of all spins in the sample. We can look at the sample as a whole, instead than looking at each spin! The maximum value of M is: M 0 γ2 h 2 B 0 4kT P D (6) Thus, M 0 is directly proportional to P D, the spin (proton) density, and to the magnetic field strength (B 0 ). P. Gori BIOMED 20/09/2017 19 / 76

Net magnetization When we put a sample within B 0, without any oscillation (i.e. no scanning), M is aligned with B 0. When we apply a precise RF pulse at f L (called B 1 and orthogonal to B 0 ), all the spins of 1 H will start to precess, and so will do M! P. Gori BIOMED 20/09/2017 20 / 76

Net magnetization The direction of B 0 is commonly designated as the z-axis in a (stationary) Cartesian coordinate system placed on the top of the scanner. M has thus a longitudinal component (M z ) and a transverse component M xy, which are time-dependent P. Gori BIOMED 20/09/2017 21 / 76

Net magnetization Remember that as long as a static magnetic field B 0 is present (even the one of earth), spins are always precessing!! This is true even after a RF-pulse B 1. However, when M is aligned with B 0, it does not precess!! P. Gori BIOMED 20/09/2017 22 / 76

Net magnetization To sum up: Apply an oscillating RF pulse B 1 orthogonal to B 0 at f L M starts to rotate around B 0 (the z-axis) tracing a spiral M is tipped away from the z-axis of a flip angle α, which mainly depends on the strength of B 1 and on the duration of the RF-pulse P. Gori BIOMED 20/09/2017 23 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 24 / 76

Relaxation Relaxation Relaxation is the process by which any excited M relaxes back to its equilibrium state, namely parallel to B 0. It can be divided into two separate processes: 1 Longitudinal relaxation (T 1 ) 2 Transverse relaxation (T 2 ) P. Gori BIOMED 20/09/2017 25 / 76

Bloch equations - stationary reference frame Starting from the Larmor precession equation dl dt = γ(l(t) x B 0(t)) (7) In 1946 Felix Bloch found this set of equation about the NMR signal: dm x (t) dt dm y (t) dt dm z (t) dt = γ(m(t) x B 0 (t)) x M x(t) T 2 = γ(m(t) x B 0 (t)) y M y(t) T 2 = γ(m(t) x B 0 (t)) z M z(t) M 0 T 1 (8) This set of equations is in a stationary reference frame positioned on the top of the scanner. P. Gori BIOMED 20/09/2017 26 / 76

Bloch equations In the presence of a constant B 0 along the z-axis, namely B 0 (t) = (0, 0, B 0 ), M z is constant. Calling M xy = M x + im y, it results M xy (t) = M xy (0) exp( iγb 0 t) = M xy (0)[cos(ωt) i sin(ωt)] The M xy component thus rotates around the z-axis (B 0 ) with an angular velocity equal to ω = γb 0 P. Gori BIOMED 20/09/2017 27 / 76

Bloch equations In the presence of a constant B 0 along the z-axis, namely B 0 (t) = (0, 0, B 0 ), M z is constant. Calling M xy = M x + im y, it results M xy (t) = M xy (0) exp( iγb 0 t) = M xy (0)[cos(ωt) i sin(ωt)] The M xy component thus rotates around the z-axis (B 0 ) with an angular velocity equal to ω = γb 0 What if we use a rotating frame of reference around the z-axis with the same angular velocity ω? P. Gori BIOMED 20/09/2017 27 / 76

Bloch equations In the presence of a constant B 0 along the z-axis, namely B 0 (t) = (0, 0, B 0 ), M z is constant. Calling M xy = M x + im y, it results M xy (t) = M xy (0) exp( iγb 0 t) = M xy (0)[cos(ωt) i sin(ωt)] The M xy component thus rotates around the z-axis (B 0 ) with an angular velocity equal to ω = γb 0 What if we use a rotating frame of reference around the z-axis with the same angular velocity ω? M xy would appear stationary with respect to this new reference frame! M z(t) = M z (t) M xy = exp(iωt)m xy (9) P. Gori BIOMED 20/09/2017 27 / 76

Longitudinal relaxation (T 1 ) In this new rotating reference frame, after applying an oscillating RF pulse B 1 at f L and with α = 90 o, the Bloch equation for M z becomes: If α = 90 o then M z(0) = 0: M z(t) = M 0 (M 0 M z(0)) exp( t T 1 ) (10) P. Gori BIOMED 20/09/2017 28 / 76

Longitudinal relaxation (T 1 ) Longitudinal relaxation is also called spin-lattice relaxation because the spins return the energy they have absorbed by the RF pulse to the surrounding lattice (i.e. external environment) This energy is very small compared to normal molecular kinetic energy almost unnoticed at room temperature T 1 depends on γ and on the mobility of the lattice, namely size and motion of the molecules of the sample different tissues have different T 1 values P. Gori BIOMED 20/09/2017 29 / 76

Transverse relaxation (T 2 ) In the new rotating reference frame, after applying an oscillating RF pulse B 1 at f L and with α = 90 o, the Bloch equation for M xy becomes: if M xy(0) = M 0 M xy(t) = M xy(0) exp( t T 2 ) (11) P. Gori BIOMED 20/09/2017 30 / 76

Transverse relaxation (T 2 ) In reality only the spins that are initially aligned with B 0 are in transverse phase coherence with one another after the 90 -pulse There is a (tiny) higher distribution of spins in the transverse plane parallel to M xy P. Gori BIOMED 20/09/2017 31 / 76

Relaxation Click for video P. Gori BIOMED 20/09/2017 32 / 76

Relaxation P. Gori BIOMED 20/09/2017 33 / 76

Relaxation T 1 is always longer than T 2 Liquids have very long T 1 and T 2 values Dense solids (e.g. ice) have very short T 2 values P. Gori BIOMED 20/09/2017 33 / 76

Relaxation T1 and T2 as a function of molecular size and tumbling (movement) rate. The minimum value of T1 and dip in the T2 curve occurs when motion is at the Larmor frequency, f 0 = f L. P. Gori BIOMED 20/09/2017 34 / 76

Nuclear Magnetic Resonance (NMR) Certain atomic nuclei, such as the hydrogen, are able to absorb specific radio frequency energy and emit a small portion of that energy. The produced time-varying magnetic field will induce voltage in the coil (Maxwell Faraday equation) This current, oscillating at the same frequency as the RF pulse, is the NMR signal P. Gori BIOMED 20/09/2017 35 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 36 / 76

Free induction decay (FID) The NMR signal is a small electrical current induced in the receiver coil by the precession of M xy during resonance Faraday s Law of Induction (a changing magnetic field induces a current in the coil) The resulting signal is called free induction decay (FID), which is a damped sine wave oscillating at the Larmor frequency (ω L ): sin(ω L t) exp( t T 2 ) P. Gori BIOMED 20/09/2017 37 / 76

T 2 In a real NMR experiment M xy decays much faster than expected. This rate is called T2 where T 2 T 2 T2 results principally from inhomogeneities in the main magnetic field. These inhomogeneities may be the result of intrinsic defects in the magnet itself or from susceptibility-induced field distortions produced by the tissue or other materials placed within the field P. Gori BIOMED 20/09/2017 38 / 76

Repetition time (T R) T R: span of time between consecutive RF pulses (milliseconds) Every RF pulse generates a FID signal If the recovery of the first FID in not complete (T R < T 1 ), the next FID signal will be reduced P. Gori BIOMED 20/09/2017 39 / 76

Spin echo - Echo Time (T E) A spin echo is produced by two successive RF-pulses (usually 90 o and 180 o ) that create a detectable signal called the echo Due to inhomogeneities in the field, spins at different locations may precess at different rates After the 180 o pulse (t) the slower spins are ahead and the faster ones are behind! At T E = 2t we have a complete refocusing that creates an echo. T 2 can be correctly measured P. Gori BIOMED 20/09/2017 40 / 76

Multi-echo pulse sequence Only the peak of the echo falls in the true T 2 decay curve Need a multi-echo pulse sequence using a series of 180 o pulses P. Gori BIOMED 20/09/2017 41 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 42 / 76

Image contrast Image contrast: signal difference between different tissues Tissues have different P D, T 1 and T 2. Signal intensity depends on these parameters Typically one aims to maximize the contrast between tissues focusing on (weighting) one of these parameters Different T E and T R!! P. Gori BIOMED 20/09/2017 43 / 76

T R and T 1 Different tissues have different T 1 If we use a long T R we have a poor contrast since M z recovers completely for both tissues Need to use a short T R to have a good contrast P. Gori BIOMED 20/09/2017 44 / 76

T E and T 2 Different tissues have different T 2 If we use a short T E we have a poor contrast but a high recorded NMR signal If we use a long T E we have a good contrast but a low recorded NMR signal P. Gori BIOMED 20/09/2017 45 / 76

Image contrast P D, T 1 and T 2 are positively correlated maximizing the sensitivity to all of them (short T R and long T E) leads to conflicting (opposite) contrast effects Contrast between gray and white matter is different! Look at the ventricles P. Gori BIOMED 20/09/2017 46 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 47 / 76

Magnetic gradients How do we localize the NMR signal? 3 gradient coils Each gradient coil produces a magnetic field that varies linearly along an axis The three gradient coils produce magnetic gradients along 3 orthogonal directions it can be produced a gradient along any direction! Gradients are added to B 0 and are much weaker P. Gori BIOMED 20/09/2017 48 / 76

Slice selection We select a thin slice (1-2 mm thick) of the body We define a new coordinate system where the z-axis is perpendicular to the slice We produce a gradient field along the new z-axis (B(z) = B 0 + zg z ) together with a tailored RF-pulse Its frequencies match the frequencies of the desired slice only the protons within the chosen slice are excited! ω(z) = ω L + γzg z P. Gori BIOMED 20/09/2017 49 / 76

Slice selection ω = γg z z : every slice has a finite width z which contains a range of frequencies ( ω) centered around ω L ω is usually fixed (1-2 khz), only variable is G z (limit of the scan) z = ω γg z strong gradients leads to small thickness We choose ω L, ω and the RF-pulse (sinc pulse or Shinnar-Le Roux (SLR) algorithm) P. Gori BIOMED 20/09/2017 50 / 76

NMR signal is the sum of all signals generated across the slice how to localize in the x-y plane? P. Gori BIOMED 20/09/2017 50 / 76 Slice selection ω = γg z z : every slice has a finite width z which contains a range of frequencies ( ω) centered around ω L ω is usually fixed (1-2 khz), only variable is G z (limit of the scan) z = ω γg z strong gradients leads to small thickness We choose ω L, ω and the RF-pulse (sinc pulse or Shinnar-Le Roux (SLR) algorithm)

Frequency encoding Once selected the slide, we can use a frequency-encoding gradient (G f ) to specify one direction (i.e. x) within the slide B(x) = B 0 + xg f ω(x) = ω L + γxg f Pixels A,B,C,D,E,F in a static B 0 would resonate at the same ω Thanks to G f, A,B,C resonate at the same ω which is lower than the one of D,E,F We can discriminate between the two columns! P. Gori BIOMED 20/09/2017 51 / 76

Frequency encoding Once selected the slide, we can use a frequency-encoding gradient (G f ) to specify one direction (i.e. x) within the slide B(x) = B 0 + xg f ω(x) = ω L + γxg f Pixels A,B,C,D,E,F in a static B 0 would resonate at the same ω Thanks to G f, A,B,C resonate at the same ω which is lower than the one of D,E,F We can discriminate between the two columns! What about the other direction (i.e. y)? P. Gori BIOMED 20/09/2017 51 / 76

Phase encoding We use another gradient G p for the last direction (i.e. y) We measure the sum of protons which all have the same frequency we use several measures with different phase shifts! If we focus on 2 pixels. At step 0: S 0 (t) = (I A + I B )e iωt At step 1, B has a 180 o phase wrt A: S 1 (t) = (I A I B )e iωt Algebra: I A = 1 2 S 0 + S 1, B = 1 2 S 0 S 1 P. Gori BIOMED 20/09/2017 52 / 76

Spatial localization A RF pulse is first applied at a precise ω L together with a slice-selection gradient (z-axis) Two opposite G f gradients are applied to produce an echo signal Each x is characterized by a different f Before the echo a G p is kept for a time τ, protons will have a phase shift along the y-direction φ(y) = γyg p τ Given a G f we need several different G p (there are hundreds of pixels) P. Gori BIOMED 20/09/2017 53 / 76

Spatial localization - Math We read the NMR signal while the x-gradient G f is active. This means that at time t the protons will have a phase shift equal to φ(t) = γg f xt + γyg p τ where γyg p τ is the phase shift already acquired due to G p The NMR signal sums the contributions from all the locations, namely: S(t) = x y I(x, y) exp[i(ω Lt + γg f xt + γyg p τ)] x y I(x, y) exp[i(γg f xt + γyg p τ)] Calling k x = (γg f t) and k y = (γg p τ) S(t) = kx x y I(x, y) exp[ i2π( 2π x + ky 2π y)] This is a spatial 2D Fourier transform! It goes from the k-space of spatial frequency amplitudes (S(k x, k y )) to the real space of intensities (I(x, y)) P. Gori BIOMED 20/09/2017 54 / 76

K-space P. Gori BIOMED 20/09/2017 55 / 76

K-space Field of view (FOV): it can be seen as the size of the object of interest FOV and pixel width w (spatial resolution) are related to the sampling in the k-space ( k) and to the dimension of the k-space k depends on the direction. One can increase the frequency encoding steps (x-axis, no time penalty) and the number of phase encoding steps (y-axis, time penalty) to reduce k. Usual k-space (matrix) size are 128x256 or 192x256. k also depends on the strength and duration of the gradient P. Gori BIOMED 20/09/2017 56 / 76

K-space (Aliasing and Low resolution) k has doubled (more space between samples in the k-space) Same w (spatial resolution) but half FOV Aliasing k is the same Only certain spatial frequencies have been considered (reduced k F OV ) w has augmented, spatial resolution has decreased P. Gori BIOMED 20/09/2017 57 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 58 / 76

Magnetic Resonance imaging We usually select and acquire one slice at a time or multi-slice using a composite RF-pulse to stimulate multiple slices. In both cases, we need to wait for full relaxation between RF-pulses only long TR (i.e. T2-weighted) To accelerate: using symmetry of Fourier plane, full volume with 2 phase encoding, use smaller flip angles, etc To increase SNR: greater B 0, greater voxel volume and slice thickness (more spins), acquisition time (average of repetitions of pulse sequence), increase FOV, decrease k-space size, decrease T E, increase T R P. Gori BIOMED 20/09/2017 59 / 76

Intrinsic parameters MRI T 1 : longitudinal relaxation time T 2 : transverse relaxation time P D : proton density field heterogeneity physiological motion P. Gori BIOMED 20/09/2017 60 / 76

Extrinsic parameters MRI External magnetic fields Sequence parameters: T R: repetition time T E: echo time flip angle number of slices FOV slice thickness slice orientation gradient parameters type of coils P. Gori BIOMED 20/09/2017 61 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 62 / 76

MRI artifacts Partial volume effect Several tissue are averaged together in a single pixel. Need for smaller pixels (higher resolution). P. Gori BIOMED 20/09/2017 63 / 76

MRI artifacts Motion artifact P. Gori BIOMED 20/09/2017 64 / 76

MRI artifacts Magnetic susceptibility artifacts Due to metal implants, characterized by geometric distortion, very dark and very bright areas P. Gori BIOMED 20/09/2017 65 / 76

Summary 1 Introduction 2 Nuclear Magnetic Resonance (NMR) 3 Net magnetization 4 Relaxation 5 Free induction decay (FID) 6 Image contrast 7 Spatial localization 8 Recap 9 MRI artifacts 10 MR modalities P. Gori BIOMED 20/09/2017 66 / 76

Anatomical (structural) MRI - Brain It provides information to qualitatively and quantitatively describe the shape, size, and integrity of gray and white matter structures T1-w: good contrast between gray matter (GM) and white matter (WM). No for CSF T2-w: good contrast between CSF and brain tissue. Sensitive to subtle white matter alterations P. Gori BIOMED 20/09/2017 67 / 76

Diffusion weighted MRI DWI detects water diffusion Brownian, unconstrained movement in large spaces (i.e. ventricles) produces isotropic diffusion. Constrained motion along a direction (i.e. white matter axons) produces anisotropic diffusion P. Gori BIOMED 20/09/2017 68 / 76

Diffusion weighted MRI Several diffusion gradients are applied (usually along 50-60 different directions) to detect water diffusion For each direction we use two opposite gradients that change linearly in the space. If there is no movement, the two signals cancel out, otherwise their difference is proportional to the movement of the water molecules along the direction S b S 0 = exp( b ADC), b is related to the strength of the gradient, ADC is the apparent diffusion coefficient since water is not free in biological tissues P. Gori BIOMED 20/09/2017 69 / 76

Diffusion weighted MRI P. Gori BIOMED 20/09/2017 70 / 76

Functional MRI - Brain First fmri: blood-oxygen-level dependent (BOLD) contrast It measures brain activity by detecting hemodynamic responses the more the brain is active in an area, the more the blood flows Hemodynamic response: blood releases oxygen to active neurons at a greater rate than to inactive neurons. Change in oxygenated and deoxygenated blood. Detectable with MRI since different magnetic susceptibility. P. Gori BIOMED 20/09/2017 71 / 76

Functional MRI - Brain It is usually used for research purpose Detect signal changes in response to different stimuli P. Gori BIOMED 20/09/2017 72 / 76

MRI contrast agent Contrast media used to improve the visibility of healthy or pathological tissues (i.e. tumor) The most common are the gadolinium-based P. Gori BIOMED 20/09/2017 73 / 76

MR angiography Technique used to image blood vessels (e.g. arteries and veins) Usually based on contrast agents and short-tr sequences P. Gori BIOMED 20/09/2017 74 / 76

Cardiac MRI Non-invasive assessment of the function and structure of the cardiovascular system Coupled with diffusion imaging V IDEO P. Gori BIOMED 20/09/2017 75 / 76

References 1 Richard B. Buxton (2002). Introduction to Functional Magnetic Resonance Imaging: Principles and Techniques. Cambridge University Press 2 F. Bloch. Nuclear Induction. Phys. Rev. 70-1946 3 http://mriquestions.com/index.html P. Gori BIOMED 20/09/2017 76 / 76