APPLICATION NOTE. Characterization and Classification of Recycled Polyamides by Means of Identify. Dr. Ekkehard Füglein

Similar documents
Identification of polymers by means of DSC, TG, STA and computer-assisted database search

TAWN tests for quantitatively measuring the resolution and sensitivity of DSCs (version 2.1)

Thermal Methods of Analysis

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

High Pressure DSC Differential Scanning Calorimeter

DSC PT 10. Applications

APPLICATION NOTE. Investigation of Alkali Salts with the STA 449 F5 Jupiter. Dr. Alexander Schindler and Dr. Michael Schöneich.

Analyzing & Testing. Thermogravimetric Analysis TGA. Technique, Instrument, Applications TG 209 F1. Leading Thermal Analysis

Nutshells of Thermal Analysis. Heat it up! Burn it! Thermal Analysis

Common Definition of Thermal Analysis

Analyzing & Testing Business Unit. Tau-R Mode for Advanced DSC Analysis Applications Newsletter 1/ /09, Dr. Stefan Schmölzer

DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS

A novel direct coupling of simultaneous thermal analysis (STA) and Fourier transform-infrared (FT-IR) spectroscopy

Evolved gas analysis by simultaneous thermogravimetric differential thermal analysis-fourier transformation infrared spectroscopy (TG-DTA-FTIR)

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14

DSC AS PROBLEM-SOLVING TOOL: BETTER INTERPRETATION OF Tg USING CYCLIC DSC

Characterization of Solid State Drugs by Calorimetry

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

DETERMINATION OF THE DECOMPOSITION PRODUCTS OF CALCIUM OXALATE USING THERMAL GRAVIMETRY AND INFRARED SPECTROMETRY

SILMALAB is a laboratory of analysis for plastics industry that takes advantage of the experience gained by our engineers in 30 years of work and

INTERNATIONAL STANDARD

Katarzyna Lewandowska

TGA Thermal Gravimetric Analysis

APPLICATIONS OF THERMAL ANALYSIS IN POLYMER AND COMPOSITES CHARACTERIZATION. Wei Xie TA Instruments

Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC)

Simultaneous Thermal Analyzer (DTA/TGA)

PCM specific heat capacity c p (T) measurements

How fast are chemical reactions?

Chapter 31. Thermal Methods

Thermal analysis unlocks the secrets of elastomers

Collected Applications Thermal Analysis EVOLVED GAS ANALYSIS

Calorimetric investigations. in multi-component salt systems

CHARACTERIZATION OF PHARMACEUTICALS BY THERMAL ANALYSIS CARACTERIZAÇÃO DE FÁRMACOS POR ANÁLISE TÉRMICA

TOPEM the new advanced multi-frequency TMDSC technique

Thermokinetic investigation of cellulose pyrolysis impact of initial and final mass on kinetic results

CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter

Beating the Heat - Fast Scanning Melts Silk Beta Sheet Crystals. Tufts University, Medford MA, 02155, USA

Kinetic evaluation of decabromodiphenil oxide as a ame retardant for unsaturated polyester

Modulated DSC Paper #8 Use Of Quasi-isothermal Mode for Improved Understanding of Structure Change

Analyzing & Testing. Simultaneous Thermal Analysis. Method, Technique, Applications STA 449 F3. Leading Thermal Analysis

Analyzing & Testing. Photocalorimetry Photo-DSC. Method, Technique, Applications. Photo-DSC 204 F1. Leading Thermal Analysis

Thermal Analysis Premium

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS

A COMPARISON OF DSC AND RADEX FOR THE INVESTIGATION OF SAFETY PARAMETERS FOR INHOMOGENEOUS SYSTEMS

17.2 Thermochemical Equations

APPLICATION OF THERMAL METHODS IN THE CHEMISTRY OF CEMENT: KINETIC ANALYSIS OF PORTLANDITE FROM NON- ISOTHERMAL THERMOGRAVIMETRC DATA

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FTIR

Supplementary Information

Analysis of Polymers and Plastics. Innovation with Integrity. Quality Control & Failure Analysis FT-IR

1) INTRODUCTION 2) ROLE IN PREFORMULATION 3) CLASSIFICATION OF THERMAL ANALYSIS. 4) DIFFERENT METHODS OF THERMAL ANALYSIS.

Physico-chemical characterization and comparison of fluorinated commercial Ski-Waxes.

UNIT 11 DIFFERENTIAL THERMAL ANALYSIS, SCANNING CALORIMETRY AND THERMOMETRIC TITRATIONS

THERMO- GRAVIMETRIC ANALYZER TGA 1000

TECHNICAL UPDATE. Ricon Resins Peroxide Curing Data and Use as a Reactive Plasticizer in Polyphenylene Ether Based CCL and PWB

Total analysis with DSC, TMA and TGA-EGA

HYPHENATED TECHNOLOGY GUIDE

Heat Transfer in Polymers

THERMAL ANALYSIS OF POLYMERIC MATERIALS

HYPHENATED TECHNOLOGY GUIDE

THERMAL BEHAVIOR OF NAFION MEMBRANES

Fabrication and thermal stability studies of polyamide 66 containing triaryl phosphine oxide

Thermal methods. P A Sathe Ramnarain Ruia College, Mumbai

Aspen Polymers. Conceptual design and optimization of polymerization processes

Calorimetry. Chapter 2. Differential Scanning heat flux calorimetry

LFA for Thermal Diffusivity and Conductivity of Metals, Ceramics and Polymers

New incremental isoconversional method for kinetic analysis of solid thermal decomposition

A Thermal Analysis Technique That Combines Differential Scanning Calorimetry and Light Microscopy

Title: Cesa-extend a User Friendly Technology to Enhance Reprocessing and Recycling of Condensation Plastics

Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis

7. Thermal Properties

Thermal Behaviour of Some Azo Dyes Containing Sterically Hindered and Water-Soluble Groups

Single-ion BAB triblock copolymers as highly efficient electrolytes for lithium-metal batteries

Thermal stability of azurite and malachite in relation to the formation of mediaeval glass and glazes

Analyzing & Testing. Comprehensive Material Characterization. Leading Thermal Analysis

Apparent Melting: A New Approach to Detecting Drug-Excipient Incompatibility

Moldflow Material Testing Report MAT2216 ISOFIL H 40 C2 F NAT

Sorption-Desorption Behavior and Characterization of Cyclophosphamide

MORPHOLOGY AND THERMAL STUDIES OF NICKEL CARBONATE NANOPARTICLES

Integrated Electricity Demand and Price Forecasting

ON THE THERMAL STABILITY OF FLAX FABRICS GRAFTED WITH MONOCHLOROTRIAZINYL-β-CYCLODEXTRIN AND TREATED WITH CINNAMIC DERIVATIVES

Melting and solidi cation of Pb nanoparticles embedded in an Al matrix as studied by temperature-modulated di erential scanning calorimetry

11B, 11E Temperature and heat are related but not identical.

Differential Scanning Calorimetry

A DSC STUDY OF THE THERMAL DECOMPOSITION OF 2 METHOXYAMINO 3, 5 DINITRO PYRIDINE

Non-Isothermal Crystallization and Thermal Degradation Kinetics of Biodegradable Poly(butylene adipate-co-terephthalate)/starch Blends

Chem Page IX - 1 LAB MANUAL Differential Scanning Calorimetry 09_dsc131.docx EXPERIMENT IX

Thermal Analysis measurements

VAPOUR PRESSURE OF DSC CALIBRATION SUBSTANCES

Thermochimica Acta 531 (2012) Contents lists available at SciVerse ScienceDirect. Thermochimica Acta

POLYAMIDE-6,9 WITH CARBAZOLE

Polymer Chemistry Research Experience. Support: NSF Polymer Program NSF (PI: Chang Ryu)/RPI Polymer Center

STK-Greifensee Nov. 6 th 2002 Thermal hazards & Storage of self-reactive chemicals

Determination of Nitroglycerine Content in Double Base Propellants by Isothermal Thermogravimetry

Application Handbook. Thermal Analysis in Practice Collected Applications

News & Trends for Thermal Analysis

UPGRADING OF PETROLEUM RESIDUE BY NITROGEN DOPING FOR CO 2 ADSORPTION

Introduction 1. DSC scan 5-bromo-2-aminopyridine..3. DSC scan 5-bromo-2-nitropyridine...4

A NEW MEASUREMENT AND EVALUATION METHOD FOR DSC OF PCM SAMPLES

ph-sensitive Nanocapsules with Barrier Properties: Fragrance Encapsulation and

Transcription:

APPLICATION NOTE Characterization and Classification of Recycled Polyamides by Dr. Ekkehard Füglein Introduction In the field of thermal analysis up to now, researchers had to compare own data with printed collections of measurement results such as Atlas of Thermoanalytical Curves [1] and others [2, 3, 4]. Recently, the first software-based database in thermal analysis, Identify, has been introduced [5]. This database allows for the first time to compare measured thermoanalytical data with library data stored in the database using a software. As a result, the user obtains a list of similarity values, the figure of merit for this comparison, which is given in percent. In the present work, Identify is used in different manners. Various polyamides are investigated employing differential scanning calorimetry (DSC). Using the information about polyamides stored in the Identify database, it will be demonstrated that even small differences in the thermal behavior of the polyamide types are sufficient to significantly distinguish between them. By means of evaluated values such as the glass transition temperature, specific heat capacity, melting temperature or melting enthalpy, a series of recycled polyamide samples was investigated and then classified by comparison with results of a virgin material stored in the database. Thus, the use of Identify as a tool for classification of recycled polyamides will be demonstrated. Materials and Methods The recycled polyamide samples were measured as received. They were designated Pentamid B GV30 batch 001 to 009. The samples used as a reference were PA6 GF30 (durethan, natur), PA6.6 GF30 (ultramid, natur), PA6.10, PA6.12 (grilamid). The melting behavior of the polyamide samples was studied using the DSC 214 Polyma. Aluminum pans (NETZSCH Concavus) with pierced lids were used to heat, to cool, and to reheat the samples at a rate of 20 K/min. The two heating segments were each run to 280 C. The second heating for each polyamide sample was used to evaluate the enthalpy of melting. All samples were prepared with masses of 4.955 (± 0.05) mg. The thermogravimetric measurements were carried out using the thermo-microbalance TG 209 F3 Tarsus. The samples with masses of 11.45 (± 0.35) mg were transferred into aluminum oxide crucibles and heated at a rate of 20 K/min to 800 C in nitrogen. For the consecutive heating to 1000 C the atmosphere was switched to synthetic air (nitrogen:oxygen = 90:10) at 800 C. The total gas flow rate of the inert and reactive gas was 40 ml/min. Results and Discussion In order to prove the capability of the Identify database in terms of identification of materials, virgin polymers here taken as reference materials were tested. The granules were prepared in aluminum pans as described above transferred into the DSC apparatus and heated in a nitrogen atmosphere to temperatures above melting. The second heating was evaluated and the obtained results were compared to results stored in the Identify database. 1 5

1 Comparison of DSC results of the 2 nd heating runs for polymer samples used as references, PA6 GF30 (red), PA6.10 (green), PA6.12 (blue and PA6.6 GF30 (black) Figure 1 compares the second heating run for each of the four different polyamides, PA6 GF30 (1), PA6.10 (2), PA6.12 (3) and PA6.6 GF30 (4). Whereas the melting temperature of PA6.6 GF30 was detected at a significantly higher temperature, the main endothermal melting effect for PA6 GF30, PA6.10 and PA6.12 was in the same temperature range. However, the database is able to distinguish and identify these samples. Tables 1a to 1d show the similarity values delivered by the database, when comparing the measured data (Figure 1) to already existing data of the library. If, for instance, the results for PA6 GF30 (curve 1 in Figure 1) are requested for comparison with database data, the similarity with polyamide 6 data stored in the database is 97%. The results stored in the database, serving for this comparison, are of course not taken from the identical measurement, but from a different measurement of a similar, but Tab. 1a Results of the database search for reference sample PA6 to be identified PA6 PA6.12 PA6.10 PVA PBT PA6 97 87 85 76 70 Tab. 1b Results of the database search for reference sample PA6.10 to be identified PA6.10 PA6.12 PA6 PBT PVA PA6.10 98 85 86 81 56 not identical sample. This is why the similarity is not exactly 100%, but therefore the proof to be able to also identify unknown samples via this procedure is much more reliable. Other polyamides, melting in the same temperature range such as PA6.10 and PA6.12, were found to have significantly lower similarities, namely 87% and 84%, respectively. The same applies if PA6.10 or PA6.12 is the sample to be identified and compared to library data. The results are summarized in tables 1a, 1b and 1c. Since polyamide 6.6 melts at an about 40 K higher temperature compared to the above-mentioned polyamides, the further data in the library are no polyamides but ETFE, PET, PPS and FEP. This procedure confirms, along with recently published data [6] [7], the capability of the Identify database to distinguish between samples with a similar thermal behavior. Tab. 1c Results of the database search for reference sample PA6.12 to be identified PA6.12 PA6.10 PA6 PBT PVF PA6.12 96 87 77 64 46 Tab. 1d Results of the database search for reference sample PA6.6 to be identified PA6.6 ETFE PET PPS FEP PA6.6 96 87 60 51 47 2 5 Phone.: +49 9287/881-0 Fax: +49 9287/881505

As a consecutive step, a series of recycled polyamide 6 samples was studied. All samples were of the same origin, but taken from different batches, namely having the same composition PA6 GF30. In order to proof the composition and to get a clear picture about the margin of deviation regarding the samples composition as well as about the repeatability of sampling, thermogravimetric measurements were carried out. Tab. 2 summarizes the samples compositions with respect to the content of volatiles, the polymer content, the carbon black content and the residual mass. As long as the samples do not contain any other chemically inert ingredients, the latter should be equivalent to the amount of added glass fibers. Samples PA6.10 and PA6.12 do not show any residuals inside the crucible after the measurements. All other samples show a white to light yellow molten residual. Tab. 2 Comparison of the thermogravimetric results (mass loss in %) for all recycled polyamides and reference materials tested PA6 samples / Volatiles Polymer Carbon black reference samples 25 to 250 C 250 to 800 C 800 to 1000 C Residual Mass 001 1.16 66.66 1.49 30.69 002 1.10 67.01 1.45 30.45 003 1.25 66.77 1.74 30.24 004 1.11 67.05 1.44 30.40 005 1.23 68.41 1.04 29.31 006 1.15 67.54 1.45 29.86 007 1.14 67.72* 1.23 29.90 008 1.12 67.87 1.70 29.31 009 1.19 66.74 1.66 30.41 PA6 GF30 0.71 69.73 0.29 29.27 PA6.10 0.09 98.66 0.10 1.15 PA6.12 0.45 98.73 0.25 0.60 PA6.6 GF30 0.41 68.02 1.10 30.48 * Within the temperature range of 250 to 800 C, this sample shows an additional mass-loss step of 1.54% which is most probably due to the release of carbon dioxide originating from the decomposition of chalk. This would refer to a chalk content of 3.5%. According to the detected glass fiber contents of 30.0% (±0.7), the TGA results could confirm the expected amount within an uncertainty of 2.5%. The thermal behavior of all batches of polymaide 6 samples (001 to 009) was then studied using differential scanning calorimetry (DSC). The second heating run each was compared to the database and also to the PA6 GF30 samples. 3 5 Phone.: +49 9287/881-0 Fax: +49 9287/881505

Tab. 3 Results of the database search for nine different recycled PA6 GF30 samples compared to a virgin PA6 GF30 Samples Similarity in % PA6 GF 30 100 008 98 003 87 001 84 006 81 009 77 005 76 002 75 007 74 004 63 The similarity values are summarized in table 3. Figure 2 depicts the visual comparison of these results together with the PA6 GF30 sample used as a reference (dashed line). The curves are displayed according to the similarity values given in table 3 with decreasing similarity values from bottom to top. Besides variation of the peak temperature, the melting enthalpy, the change in specific heat capacity and the glass transition temperature, it is obvious that the recycled materials also show additional effects which were not expected and not detectable for the virgin sample. Additional 2 Comparison of the DSC results of the second heating runs for all recycled polyamide samples compared with a virgin polyamide sample used as references (dashed line). The curves are displayed according to the similarity values given in Tab. 3 with decreasing similarity values from bottom to top. endothermic effects were detected for some samples in the temperature range around 22, 105 and 245 C. They are most probably due to impurities or caused by foreign substances, mainly additives or other polymers. Those additional - unexpected effects of course reduce the similarity values since they are not typical for the virgin materials and therefore not part of the data stored in the library database. This on the other hand means the database comparison takes into account if expected effects are missing or if additional effects are detected which are not stored in the database for this kind of material. 4 5

A comparison of the results obtained for the material serving as a reference (PA6 GF30, dashed black line, middle) with the most similar sample (blue) and the most different sample (green) according to the similarity results obtained from the database is shown in figure 3. The sample showing the weakest similarity does not only show additional endothermic effects at around 22 and 105 C, respectively, but also the evaluated values for the melting temperature and the glass transition temperature are more shifted to lower values compared to the reference material than for the sample showing the most similar thermal behavior. 3 Comparison of evaluated values from DSC second heating runs for sample 004 (green) having the lowest similarity value with the polyamide reference material (dashed black line, middle) and with sample 008 (blue) showing the highest similarity value Conclusion The recently introduced Identify database is the first thermoanalytical software offering a software-based comparison of measured DSC data with DSC measurement or literature values stored in the library. A series of recycled polyamide samples were measured using the differential scanning calorimeter (DSC 214 Polyma). The evaluated values for the glass transition and melting were used as identification criteria. The Identfiy database offers the capability to not only distinguish between different types of polyamides such as PA6, PA6.6, PA6.10 and PA6.12 but it furthermore allows to detect and quantify difference in temperature or enthalpy for the above-mentioned caloric effects. Depending on the desired quality or processing demands, the similarity values can serve for classification of the material and might be used as a quality control tool. References [1] G. Liptay, editor. Atlas of thermoanalytical curves (TG-, DTG-, DTA-curves measured simultaneously). Akadémiai Kiadó: Budapest; 1976. [2] Kaisersberger E, Möhler H. DSC on polymeric materials. NETZSCH annual for science and industry, volume 1. Selb:, D 8672; 1991. [3] Kaisersberger E, Knappe S, Möhler H. TA for polymer engineering. NETZSCH annual for science and industry, volume 2. Selb:, D 8672; 1993. [4] Kaisersberger E, Knappe S, Möhler H, Rahner S. TA for polymer engineering. NETZSCH annual for science and industry, volume 2. Selb:, D 8672; 1993. [5] (a) Schindler A. Automatic evaluation and identification of DSC Curves, Plastics engineering. 2014. www.plasticsengineering.org/ ProductFocus/productfocus.aspx?ItemNumber=20498 (b) A. Schindler, NETZSCH Application Note 059, Analysis of 200 Unknown DSC Curves by Using its Polymer Libraries, 2014 (c) A. Schindler, C. Strasser, Application Note 060, Stability of Identify Database Search Results with Regard to Sample Mass and Heating Rate, 2014 (d) A. Schindler, NETZSCH Application Note 061, The Identify Database as an Archive for NETZSCH and User Data, 2014 [6] (a) Fueglein E, Kaisersberger E. About the development of databases in thermal analysis. J. Therm. Anal. Calorim. DOI: 10.1007/ s10973-014-4381-3 (b) Füglein E., Kaisersberger E., Identification of Polymer Mixtures (PE/PP) by, NETZSCH Application Note 078, 2015 (c) Füglein E., Kaisersberger E., Identification of PC-PBT Mixtures by, NETZSCH Application Note 079, 2015 [7] Fueglein E., About the use of IDENTIFY - a thermoanalytical database - for characterization and classification of recycled polyamides J. Therm. Anal. Calorim. DOI: 10.1007/s10973-015-4583-3 5 5