IUPAC/CITAC Guide: Investigating out-of-specification test results of chemical composition based on metrological concepts (IUPAC Technical Report)*

Similar documents
SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

INTERNATIONAL UNION OF PURE AND APPLIED CHEMISTRY ANALYTICAL CHEMISTRY DIVISION* INTERDIVISIONAL WORKING PARTY FOR HARMONIZATION OF

IUPAC Technical Report

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis*

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Measurement Uncertainty, March 2009, F. Cordeiro 1

Precision estimated by series of analysis ISO and Approach Duplicate Approach

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

A basic introduction to reference materials. POPs Strategy

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Measurement uncertainty and legal limits in analytical measurements

Rosemary extract liquid

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Conformity assessment of a high-performance liquid chromatography calibration.

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

Method Validation Characteristics through Statistical Analysis Approaches. Jane Weitzel

What is measurement uncertainty?

Validation and Standardization of (Bio)Analytical Methods

Unit 4. Statistics, Detection Limits and Uncertainty. Experts Teaching from Practical Experience

Quality assurance of a test result: classification, modeling and quantification of human errors

Certified Reference Material - Certificate of Analysis

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Workshop on Understanding and Evaluating Radioanalytical Measurement Uncertainty November 2007

CITAC: An aid to demonstrate traceability in chemical measurements. Laly Samuel CITAC Chair MSL, New Zealand

GUIDELINES ON PERFORMANCE CRITERIA FOR METHODS OF ANALYSIS FOR THE DETERMINATION OF PESTICIDE RESIDUES IN FOOD AND FEED CXG Adopted in 2017.

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp

Measurement & Uncertainty - Basic concept and its application

Document No: TR 12 Issue No: 1

Traceable UV-Vis Reference Materials

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Schedule for a proficiency test

serve the goal of analytical lmethod Its data reveals the quality, reliability and consistency of

Certificate of Analysis

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Uncertainty Propagation for Force Calibration Systems

EFFECT OF THE UNCERTAINTY OF THE STABILITY DATA ON THE SHELF LIFE ESTIMATION OF PHARMACEUTICAL PRODUCTS

The Uncertainty of Reference Standards

Certificate of Analysis

Enquiry. Demonstration of Uniformity of Dosage Units using Large Sample Sizes. Proposal for a new general chapter in the European Pharmacopoeia

Analytical Performance & Method. Validation

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents

Measurement & Uncertainty - Concept and its application

Practical Statistics for the Analytical Scientist Table of Contents

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

การใช REFERENCE STANDARDS ในการควบค มค ณภาพยา

Decision rules applied to conformity assessment

International Atomic Energy Agency

Determination of particle size distribution Single particle light interaction methods. Part 4:

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

Joint Committee for Traceability in Laboratory Medicine Terminology. R. Wielgosz and S. Maniguet

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

R SANAS Page 1 of 7

Limit of Detection and Its Establishment in Analytical Chemistry

INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE

Validation of TNCO, TSNAs, and PAHs

Quantifying Uncertainty in Analytical Measurement

Counterfeit and illegal pesticides: Strategies for addressing the issue in the analytical laboratory

Part IVB Quality Assurance/Validation of Analytical Methods

The Role of Proficiency Tests in the Estimation of Measurement Uncertainty of PCDD/PCDF and PCB Determination by Isotope Dilution Methods

TITLE OF THE CD (English): New OIML Document

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Certificate of Analysis

Certificate of Analysis

Unified Approach For Performing An Analytical Methods Comparability Study IVT s LAB WEEK Presenter: Peter M. Saama, Ph.D. Bayer HealthCare LLC

Certificate of Analysis

Certificate of Analysis

Methods for Identifying Out-of-Trend Data in Analysis of Stability Measurements Part II: By-Time-Point and Multivariate Control Chart

EPA's Revision to the 40 CFR Part 136 Method Detection Limit (MDL) Procedure

Application of Gauge R&R Methods for Validation of Analytical Methods in the Pharmaceutical Industry

14 Validation of Analytical Methods Based on Chromatographic Techniques: An Overview

Signal, Noise, and Detection Limits in Mass Spectrometry

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

METROLOGY IN CHEMISTRY MEETING. AZIZ MUKOTA Chemistry Laboratory Uganda National Bureau of Standards

Certificate of Analysis

Introduction to the evaluation of uncertainty

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Essentials of expressing measurement uncertainty

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

Theory of Sampling. Guide to Quality Sample Processing. Jo Marie Cook NACRW 2015

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

Quality by Design and Analytical Methods

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

Measurement Uncertainty Principles and Implementation in QC

TNI V1M Standard Update Guidance on Detection and Quantitation

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Guide to the Expression of Uncertainty in Measurement (GUM) and its supplemental guides

Certificate of Analysis

Transcription:

Pure Appl. Chem., Vol. 84, No. 9, pp. 1939 1971, 2012. http://dx.doi.org/10.1351/pac-rep-11-10-04 2012 IUPAC, Publication date (Web): 9 July 2012 IUPAC/CITAC Guide: Investigating out-of-specification test results of chemical composition based on metrological concepts (IUPAC Technical Report)* Ilya Kuselman 1,, Francesca Pennecchi 2, Cathy Burns 3, Aleš Fajgelj 4, and Paolo de Zorzi 5 1 National Physical Laboratory of Israel, Givat Ram, 91904 Jerusalem, Israel; 2 Istituto Nazionale di Ricerca Metrologica, 91 Strada delle Cacce, 10135 Turin, Italy; 3 Food and Drug Administration, 6 th Avenue and Kipling Street, DFC-Bldg 20, Denver, CO 80225, USA; 4 International Atomic Energy Agency, Vienna International Centre, P.O. Box 100, A-1400 Vienna, Austria; 5 Istituto Superiore per la Protezione e la Ricerca Ambientale, Via Castel Romano, 100 00128, Roma, Italy Abstract: A metrological background for investigating out-of-specification (OOS) test results of chemical composition is discussed. When an OOS test result is identified, it is important to determine its root causes and to avoid reoccurrence of such results. An investigation of the causes based on metrological concepts is proposed. It includes assessment of validation data of the measurement process and its metrological traceability chains, evaluation of measurement uncertainty, and related producer s and consumer s risks. This approach allows distinguishing between OOS test results that indicate an actual change in chemical composition of an analyzed object, and OOS test results that are metrologically related with a certain confidence probability, i.e., caused by measurement problems, while the analyzed object still meets the specification requirements at the time of testing. Practical examples illustrating applications of the described approach in environ mental and food analysis, as well in drug analysis and stability study of drug products, are described. Acceptance limits, warning and action lines for the test results, and corresponding producer s and consumer s risks are discussed. Keywords: acceptance limits; IUPAC Analytical Chemistry Division; measurement uncertainty; out-of-specification test results; producer s and consumer s risks; warning and action lines. CONTENTS 1. INTRODUCTION 1.1 Scope and field of application 1.2 Terminology 1.3 Symbols and abbreviations *Sponsoring bodies: IUPAC Analytical Chemistry Division; IUPAC Interdivisional Working Party on Harmonization of Quality Assurance; Cooperation on International Traceability in Analytical Chemistry (CITAC): see more details on p. 1969. Corresponding author: E-mail: ilya.kuselman@moital.gov.il 1939

1940 I. KUSELMAN et al. 2. METROLOGICAL APPROACH 2.1 Assessment of validation data 2.2 Evaluation of measurement uncertainty contributions 2.3 Assessment of metrological traceability chains 2.4 Metrologically related OOS test results and acceptance limits 3. HYPOTHESES ON A PRODUCT QUALITY AND OOS TEST RESULTS 3.1 Modeling a distribution 3.2 Probability of OOS test results 3.3 Global producer s and consumer s risks 4. LIMITATIONS ANNEX A. CALCULATION OF GLOBAL RISKS ANNEX B. EXAMPLES MEMBERSHIP OF SPONSORING BODIES ACKNOWLEDGMENTS REFERENCES 1. INTRODUCTION Out-of-specification (OOS) test results of chemical composition are results that fall outside the specifications of acceptance criteria established in pharmaceutical industry [1], or do not comply with regulatory, legislation, or specification limits in other industries and such fields as environmental and food analysis. The problem of OOS test results has been known for analysts working in quality control laboratories since the 1920s [2]. However, this problem attracted special attention in 1993, when Barr Laboratories (a generic-drug manufacturer) was sued by the U.S. government regarding a set of issues influencing the product quality, including the way the company dealt with OOS test results. Judge Wolin s ruling (the Barr Decision) was that following an OOS test result, an investigation must be initiated before any retesting can be done [3]. Identifying OOS test results is described in U.S. Food and Drug Administration (FDA) Guidance [1] as the laboratory (Phase 1) investigation. It includes responsibility of the analyst and his/her supervisor, conditions of the testing in the laboratory, etc. After identification of an OOS test result, it is important to determine its root causes with the purpose to avoid any reoccurrence of the conditions when appearance of a next OOS test result is possible or even inevitable. The FDA Guidance formulates recommendations for such incidences including production process review, additional laboratory testing using a predefined procedure, reporting testing results, and concluding the investigation with identification of the root causes. Thus, this document establishes an empirical organizational approach to the full-scale (Phase 2) investigation and decisions that can be accepted at the different stages of this investigation. Currently, the majority of analysts realize that the measurement uncertainty concept is very important because of the necessity to balance the cost of measurements vs. the product quality risk [4,5]. For example, to assess compliance of a test result within legislation limits in food and feed in Europe, the analyst should report not only an analyte concentration, but also the associated measurement uncertainty [6]. When the compliance assessment is made on the basis of a measurement result accompanied by information on the uncertainty associated with the result, the rules developed in the EURACHEM/CITAC guide [7] can be used. Similar rules are included in the ILAC guidelines [8]. The JCGM guide [9] on the role of measurement uncertainty in conformity assessment has been recently developed. An approach [10] based on measurement uncertainty and other metrological concepts, amplifying recommendations of the FDA guidance for the full-scale investigation of OOS test results, is detailed in the present guide.

IUPAC/CITAC Guide: Investigating OOS test results 1941 1.1 Scope and field of application This Guide is developed for implementation of metrological concepts for investigation of OOS test results of chemical composition. This includes assessment of validation data of the measurement process (of the test method) and its metrological traceability chains, evaluation of measurement uncertainty, and related producer s and consumer s risks. The document is intended for quality control (chemical analytical) laboratories, for accreditation bodies, laboratory customers, regulators, quality managers, metrologists, and analysts. 1.2 Terminology Terminology used in this Guide corresponds to ISO Guide 99 [11], ISO 3534 [12], and ISO 17000 [13]. 1.3 Symbols and abbreviations a year (annus in Latin) ANOVA analysis of variance AOAC Association of Official Analytical Chemists International c amount-of-substance concentration of an analyte in a product or environmental object c a.l action line for c test c l.a.l lower acceptance limit for c test c l.s.l concentration lower specification limit c r parameter expressing c test in parts of corresponding MRL c test concentration test result c true concentration true value c u.a.l upper acceptance limit for c test c u.s.l concentration upper specification limit c w.l warning line for c test CITAC Cooperation on International Traceability in Analytical Chemistry Codex international organization that develops food standards, guidelines, and related Alimentarius documents, named Food Book (Codex Alimentarius in Latin) Commission D OOS deviation of OOS test result from a specification limit EP European Pharmacopoeia EPA U.S. Environmental Protection Agency EURACHEM network of organizations providing a focus for analytical chemistry and quality related issues in Europe F observed frequency f(c true ) pdf of c true distribution f(c test ) pdf of c test distribution f(c test c true ) pdf of c test distribution at a certain c true (likelihood function) f 0 fraction of tested samples in which no pesticide residues were found FDA U.S. Food and Drug Administration GC gas chromatography H 0 null hypothesis H 1 alternative hypothesis HPLC high-performance liquid chromatography i index number ICH International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use

1942 I. KUSELMAN et al. ILAC International Laboratory Accreditation Cooperation INPL National Physical Laboratory of Israel ISO International Organization for Standardization ISRAC Israel Laboratory Accreditation Authority JCGM Joint Committee for Guides in Metrology k coverage factor LC liquid chromatography m mass MRL maximum residue limit MS mass spectrometry mˆ maximum likelihood estimate of a Weibull distribution shape parameter n number of OOS test results N total number of test results NIST National Institute of Standards and Technology OOS out of specification P probability; level of confidence pdf probability density function Q av average rate of air drawn into a sampler during sampling r correlation coefficient R c consumer s risk R p producer s risk s, s c standard deviation; index c is for confidence limit to regression line SI International System of Units t normalized variable having t distribution (Student s distribution) t(0.95, ν) quantile of Student s distribution for level of confidence 0.95, one-sided interval, and ν degrees of freedom TSP total suspended particulates u standard measurement uncertainty U expanded measurement uncertainty USP U.S. Pharmacopoeia UV ultraviolet V volume α probability of Type I error β probability of Type II error βˆ maximum likelihood estimate of a Weibull scale parameter δ an contribution to c test caused by analytical error(s) δ samp contribution to c test caused by sampling error(s) μ population mean ν number of degrees of freedom σ population standard deviation τ time τ 0 shelf life or retest period of a drug product τ l.s.l time of a product storage when c test achieves c l.s.l τ OOS time of a product storage when OOS test results appear τ u.s.l time of a product storage when c test achieves c u.s.l Φ cumulative distribution function

IUPAC/CITAC Guide: Investigating OOS test results 1943 2. METROLOGICAL APPROACH Any OOS test result can indicate a product failure, or be caused by measurement (analytical), i.e., metrological problems. When a result of testing is quantitative and equal to the measurement result, the metrological approach requires, first of all, defining the measurand, i.e., the quantity intended to be measured. In an analytical quality control laboratory, it is amount-of-substance concentration c of an analyte in a product batch or an environmental object. The concentration true value is c true. A model of the concentration test result c test includes c true and contributions caused by error(s) in sampling δ samp and analysis δ an as two stages of testing: c test = c true + δ samp + δ an (1) Distribution functions associated with these contributions can be very different. In particular, distribution of c true values depends on changes of conditions of the production process from batch to batch or changes of the tested environmental object depending on place, day, etc. (global distribution). However, for a well-studied and widely used measurement method including sampling and analysis, distributions of δ samp, δ an, and corresponding c test distribution at one and the same c true value (measurement distribution), are normal more often than not or can be transformed into normal. This c test distribution caused by the measurement uncertainty can be characterized by a probability density function (pdf), general for any c true value in the range under investigation (the likelihood function). Taking into account this model, the full-scale investigation of OOS test results based on the metrological concepts includes the following: 1) assessment of validation data for sampling and chemical analysis; 2) evaluation of contributions to the measurement uncertainty from different stages of the test; 3) assessment of the metrological traceability chains important for the measurement parameters and environmental conditions influencing the test results; and 4) evaluation of consumer s and producer s risks in interpretation of OOS test results. Such an investigation should answer the question, whether the OOS test result is caused by unsatisfactory product (environment) quality, or this result is metrologically related. In other words, are the root causes of an OOS test result deviation from the specification limit found in the measurement/analytical process? 2.1 Assessment of validation data Validation is verification, where the specified requirements are adequate for an intended use [11]. This is a widely used procedure in the pharmaceutical industry. There are the FDA guidance for industry process validation including validation of sampling procedures [14], the ICH guideline for validation of analytical procedures [15], recommendations for analytical method and measuring equipment validation [16], etc. In other industries and analytical fields, validation is regulated by the EURACHEM guide [17], AOAC validation programs, and other national and international documents [18]. The most common validation parameters are repeatability, reproducibility, trueness and bias, limit of detection, selectivity and sensitivity, as well as linearity and limit of quantification [15], robustnesss, and ruggedness [19]. Investigating OOS test results, one should verify where the specified requirements and the validation data are adequate for the intended use. Absence of the adequacy can be a root cause of the OOS test results. Another question to be checked is whether the validation data are complete enough to evaluate contributions of the associated measurement uncertainty.

1944 I. KUSELMAN et al. 2.2 Evaluation of measurement uncertainty contributions Measurement uncertainty is a non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used [11]. Evaluation of measurement uncertainty can be done using repeatability, reproducibility, and trueness estimates from the validation data [20]. A number of examples of uncertainty calculation in the field of environmental analysis are available in the handbook [21]. Other methods for quantifying uncertainty in analytical measurement are described in the EURACHEM/CITAC guide [22]. Methods and approaches for evaluating measurement uncertainty arising from sampling are discussed in the EURACHEM/CITAC guide [23]. There are two important measurement uncertainty aspects and questions in the full-scale investigation of OOS test results: 1) is the measurement uncertainty adequate for the intended use? and 2) are the contributions to the measurement uncertainty the values of the same order of magnitude? Any negative answer for one or both of these questions can indicate a cause of the OOS problem. If a dominant contribution is detected while answering the second question, this contribution should be studied thoroughly. 2.3 Assessment of metrological traceability chains Metrological traceability is a property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty. Traceability chain is a sequence of measurement standards and calibrations that is used to relate a measurement result to a reference [11]. There are the EURACHEM/CITAC guide [24] on this topic, the IUPAC technical report on metrological traceability of measurement results in chemistry [25], and many other documents and publications, e.g., [26]. Assessment of metrological traceability chains is important for measurement parameters and environmental conditions influencing the test results. For example, traceability chains of measurement results to SI units of mass (kilogram), of amount of substance (mole), and of thermodynamic temperature (kelvin) should be realized for practically every chemical test. The reason is that a test portion is quantified by mass, measuring instruments are calibrated by certified reference materials, and temperature is to be under control. In particular, the pharmaceutical industry s practice of using a one-point calibration raises questions regarding the traceability chain of the measurement result to mole. This calibration consists of comparison of responses of the measuring system obtained for the test portion and a working standard. The working standard is certified by comparison with a USP s or other reference standard. Commutability [11] between the reference and the working standards, and adequacy of the working standard to the substance or drug product under analysis should be a point for investigation as well, first of all for impurities and degradation products. Any broken metrological traceability chain can lead to OOS test results. 2.4 Metrologically related OOS test results and acceptance limits When root causes of metrologically related OOS test results are found, corresponding corrective actions of the measurement/analytical process will be helpful. However, even though a problem is not found in the investigation, and an OOS test result c test differs from specification limit in the range of expanded measurement uncertainty U(c test ), it can be considered also as a metrologically related OOS test result because of the uncertainty. For example, when concentration upper specification limit c u.s.l should be taken into account, the difference D OOS = c test c u.s.l U(c test ) may be metrologically related with certain probability. When D OOS > U(c test ), this difference is probably not caused by metrological problems and indicates violation of the product or environmental quality.

IUPAC/CITAC Guide: Investigating OOS test results 1945 For interpretation of such results, acceptance limits for c test set by a testing laboratory, manufacturer, or regulator, different from specification limits by measurement uncertainty, are applied according to the EURACHEM/CITAC guide [7]. Upper acceptance limit c u.a.l for test results smaller than upper specification limit c u.s.l by expanded measurement uncertainty can be used as a warning line c w.l = c u.a.l = c u.s.l U(c test ), which are seen used in quality control charts [27]. When a test result exceeds the warning line, i.e., c test > c w.l, the sampling and measuring systems should be checked and a decision to repeat the test may be made. At the same time, any decision about the product quality is still based on comparison of the test results with the upper specification limit c u.s.l. Upper acceptance limit c u.a.l larger than c u.s.l by expanded measurement uncertainty can be used as an action line c a.l = c u.a.l = c u.s.l + U(c test ), separating the metrologically related and violating OOS test results. The c a.l values resemble action lines that are used in quality control charts. Similar acceptance limits c l.a.l for test results different from concentration lower specification limit c l.s.l by expanded measurement uncertainty, as well as warning and action lines to lower specification limit c w.l = c l.a.l = c l.s.l + U(c test ) and c a.l = c l.a.l = c l.s.l U(c test ), respectively, are applicable also. Examples from the environmental field, food quality control, and pharmaceutical analysis, including a stability study of drug products, are provided in Annex B. 3. HYPOTHESES ON A PRODUCT QUALITY AND OOS TEST RESULTS Any decision on a product quality and its conformity assessment is based on comparison of null hypothesis H 0 that the quality is satisfactory and an alternative hypothesis H 1 about unsatisfactory product quality [27]. For example, when an upper specification limit c u.s.l is discussed, there are H 0 : c true c u.s.l against H 1 : c true > c u.s.l. True value c true is unknown, and decisions are made using test results c test. The distribution of c true values in different batches of a product (the global distribution) and the measurement distributions of c test values for two of these batches under testing according to the model in eq. 1 are illustrated in Fig. 1 as pdfs, i.e., f(c true ) and f(c test ), respectively, truncated normal for simplicity. Centers (means) of the c test distributions are shown by vertical dotted pointers. These dotted pointers reach the true values c true of the analyte concentration in the particular batches of the product under testing, i.e., considered coinciding with them. The upper specification limit c u.s.l is represented by a dotted vertical line. The range of c true > c u.s.l values corresponding to the product failure is shown by the horizontal dotted pointer. The shaded area under the f(c true ) curve to the right side of c u.s.l in Fig. 1a equals the probability P of the product failure. The range of OOS test results c test > c u.s.l is shown by horizontal dotted pointers in Figs. 1b and 1c. The shaded area under the f(c test ) curve in Fig. 1b is the probability α of Type I error in the decision on the product quality. This error, named also false positive, appears when c true c u.s.l, while an OOS test result c test > c u.s.l is obtained, hypothesis H 0 is rejected and hypothesis H 1 is not rejected (accepted). The probability α of Type I error is the producer s risk. Type II error in the decision on the product quality, named also false negative, is possible when product failure is analyzed, c true > c u.s.l, while c test c u.s.l and hypothesis H 0 is not rejected. This situation is illustrated in Fig. 1c. The shaded area under the f(c test ) curve to the left side of c u.s.l is probability β of Type II error. It is the consumer s risk. 3.1 Modeling a distribution When a measurement distribution of c test results (like in Fig. 1b and/or in Fig. 1c) is known and a number of tested batches are statistically significant, the global distribution of c true values shown in Fig. 1a can be approximated by the empirical distribution of test results accumulated from batch to batch or from day to day of environmental monitoring, etc. The empirical distribution is fitted by a theoretical distribution (a model) with unknown, as a rule, parameters. Goodness-of-fit techniques for a control of

1946 I. KUSELMAN et al. Fig. 1 OOS test results, producer s risk α and consumer s risk β. Functions f(c true ) in Fig. 1a and f(c test ) in Figs. 1b and 1c are pdf of c true and c test, respectively; vertical dotted pointers are means of c test distributions equal to certain c true ; P is the probability of the product failure. Reproduced from ref. [10] by permission of Springer. the model adequacy are described, e.g., in the textbook [28]. A model can be also chosen based on knowledge about the production process and properties of the product or about the environmental object under testing. Examples of lognormal, Weibull, Student s, and normal models of global c true distributions are discussed in Annex B, Examples 1 4, respectively. 3.2 Probability of OOS test results When a global c true distribution is approximated by a corresponding model, adequate to the c test data, probability P of the product failure in Fig. 1a is transformed in the probability P OOS of OOS test results. Therefore, P OOS can be evaluated by the following equations at upper, lower, and both specification limits, respectively:

IUPAC/CITAC Guide: Investigating OOS test results 1947 P OOS = 1 Φ(c u.s.l ), P OOS = Φ(c l.s.l ) and P OOS = 1 [Φ(c u.s.l ) Φ(c l.s.l )] (2) where Φ is the cumulative distribution function modeling the global c true distribution, i.e., the integral of the modeling pdf. The integrals Φ(c u.s.l ) and Φ(c l.s.l ) in eqs. 2 have the left integration limit equal to zero, since concentration of an analyte in a product or environmental object is a non-negative property. Since calculation of the probability by eqs. 2 is per se integration of the distribution tails, results of such calculation can be larger than observed frequency values of OOS test results, when OOS test results appear mostly close to the specification limit(s), i.e., far from zero and infinity. Examples of such calculations are provided in Annex B. 3.3 Global producer s and consumer s risks While α and β are producer s and consumer s risks, respectively, for one and the same batch of a product, such risks evaluated in general for a statistically significant number of batches, forming a global c true distribution, are named global producer s risk R p and global consumer s risk R c. Equations for their calculation are presented in Annex A, examples of the calculations are in Annex B. 4. LIMITATIONS The metrological approach does not spread in the present guide to cases of semi-quantitative and qualitative (e.g., organoleptic) testing and human errors. If investigation of an OOS test result indicates a product failure caused by technological problems, this approach cannot be directly useful. However, when a significant contribution to measurement uncertainty arising from sampling is identified as a cause of OOS test results, an optimization of the technological parameters may be required to increase the product homogeneity. The global producer s risk R p and consumer s risk R c do not take into account possible economical, health, and social consequences of false decisions on quality of a material or environment under testing. ANNEX A. CALCULATION OF GLOBAL RISKS There are three scenarios for calculation of the global risks according to the JCGM guide [9]: 1) around an upper specification limit, 2) around a lower specification limit, and 3) when both specification limits should be taken into account. The global producer s risk R p and consumer s risk R c around an upper specification limit c u.s.l can be evaluated by the following equations: where f(c test c true ) is the measurement pdf of c test distribution at a certain c true value (the likelihood function). The risks around a lower specification limit c l.s.l are: (3) (4) (5) (6)

1948 I. KUSELMAN et al. are: In the case when both specification limits should be taken into account simultaneously, the risks (7) (8) The integrals are calculated numerically. Examples of such calculations are presented in Annex B: 1) around an upper specification limit, Examples 1 3; 2) around a lower specification limit, Example 3; and 3) for the case of both specification limits, Example 4. ANNEX B. EXAMPLES CONTENTS EXAMPLE 1. INVESTIGATING OOS TEST RESULTS OF TOTAL SUSPENDED PARTICULATE MATTER CONCENTRATION IN AIR B-1-1 Introduction B-1-2 Experimental B-1-3 Global distribution B-1-4 Causes and probability of OOS test results B-1-5 Risks of stone producer and inhabitant EXAMPLE 2. MULTICOMPONENT OOS TEST RESULTS: PESTICIDE RESIDUES IN TOMA- TOES B-2-1 Introduction B-2-2 Experimental B-2-3 Global distribution B-2-4 Causes and probability of OOS test results B-2-5 Risks of tomato producer and consumer EXAMPLE 3. OOS TEST RESULTS IN LONG-TERM STABILITY STUDY OF DRUG PROD- UCTS B-3-1 Introduction B-3-2 Experimental B-3-3 Regression analysis and shelf life of the products EXAMPLE 4. B-3-3-1 Shelf life of sodium chloride injection B-3-3-2 Shelf life of epinephrine injection B-3-4 Risks of setting a shelf life B-3-4-1 When a measured attribute increases with time B-3-4-2 When a measured attribute decreases with time OOS RESULTS OF CETIRIZINE DIHYDROCHLORIDE ASSAY B-4-1 Introduction B-4-2 Experimental B-4-3 Global distribution B-4-4 Causes and probability of OOS test results B-4-5 Risks of producer and consumer

IUPAC/CITAC Guide: Investigating OOS test results 1949 EXAMPLE 1. INVESTIGATING OOS TEST RESULTS OF TOTAL SUSPENDED PARTICULATE MATTER CONCENTRATION IN AIR B-1-1 Introduction The objective of this example was an application of the metrological approach in the environmental field for investigating OOS test results of total suspended particulate (TSP) matter concentration in ambient air of some industrial zones [29]. EPA method IO-2.1 [30] for characterizing TSP uses a high-volume sampler for collection of particles with aerodynamic diameters of 100 μm or less. The sampler design causes air to be drawn into the sampler by means of a blower and through a glass or quartz fiber filter located downstream of the sampler inlet in order that the airborne particulate matter can be collected uniformly on the filter surface. A large volume, V, of (1600 to 2400) m 3 of air is typically sampled at an average rate Q av of (1.13 1.70) m 3 min 1 during sampling. Thus, V = Q av τ, whereτ is the total elapsed sampling time in min. In order to determine a metrologically traceable value of the air volume, the flow rate measurement device should be calibrated and the total volume of sampled air corrected to V st at the EPA s standard temperature and pressure by EPA method IO-2.4 [31]. The mass m of the matter in the sampled air volume is measured as the difference between the results of weighing the filter before and after sampling, in mg. The filter media characteristics and performance, as well as its conditions before and after sampling, are prescribed in EPA method IO-3.1 [32]. The TSP concentration in ambient air is the measurand in this method, while practically measured is the average value of the TSP concentration over the sampling/test period. Therefore, the test result is c test = m/v st. There are national regulations of air quality including upper specification limits c u.s.l for TSP concentration depending on the period of averaging. For example, in Israel c u.s.l = 0.200 mg m 3 for 24 h. OOS test results of TSP concentration in ambient air of stone quarries located in Israel were investigated during a year as a case study. B-1-2 Experimental High-volume samplers and glass fiber filters were used. Three quarries were monitored by the National Physical Laboratory of Israel (INPL) according to EPA method IO-2.1 at all four points in the compass approximately (1 3) km from each quarry, four to five times per month, i.e., once per week or more frequently during the quarries work. Each test lasted for 24 h. A total of 496 test results were obtained. The results were sorted with analysis of variance (ANOVA). The main found factor of variation was a quarry. Some details of the data distribution for every quarry, including number N of the test results, mean μ and standard deviation σ of the result natural logarithms, number n of OOS test results and their observed frequency F = n/n, are presented in Table 1. A total of 20 out of the 496 were OOS test results. The distributions are shown in Fig. 2 as histograms. Table 1 Observed frequency F and probability P OOS of OOS test results. Quarry N μ σ n F P OOS 1 220 2.326 0.434 7 0.032 0.049 2 176 2.031 0.280 11 0.063 0.066 3 100 2.338 0.403 2 0.020 0.035

1950 I. KUSELMAN et al. Fig. 2 Histograms and pdf of lognormal distributions of test results c test /mg m 3. Data for quarry 1 are shown in Fig. 2a, quarry 2 in Fig. 2b, and quarry 3 in Fig. 2c. The upper specification limit c u.s.l is indicated by the dotted vertical line, the range of OOS test results by the dotted pointer. Reproduced from ref. [29] by permission of Springer. The expanded relative measurement uncertainty was evaluated as U(c test )/c test = 0.14 to 0.21 or (14 to 21) % for normal distribution and the range of the levels of confidence P = 0.95 to 0.99, with the coverage factor k = 2 to 3. B-1-3 Global distribution Lognormal distributions were used for modeling pdf of the global c true distributions for every quarry: (9) where μ and σ values for a quarry are from Table 1. These pdfs are shown in Fig. 2 for quarry 1 (Fig. 2a), quarry 2 (Fig. 2b), and quarry 3 (Fig. 2c) by solid lines smoothing the empirical c test distributions, while the empirical distributions are presented here by the histograms. The upper specification limit c u.s.l = 0.200 mg m 3 is shown by a dotted vertical line, common for all parts of Fig. 2. The range of OOS test results is indicated by dotted pointer.

IUPAC/CITAC Guide: Investigating OOS test results 1951 Using the lognormal models of the c true distributions by eq. 9, one can estimate probabilities P OOS of OOS test results for every quarry by eq. 2 for an upper specification limit. The cumulative lognormal distribution functions Φ in eq. 2 were calculated for c test > 0 and upper specification limit c u.s.l = 0.200 mg m 3. Results of calculations are presented in Table 1. Comparison of the frequency values with probabilities of the OOS results shows that their annual numbers n may be larger than the observed ones. B-1-4 Causes and probability of OOS test results All observed OOS test results, their deviations D OOS from the upper specification limit, and U(c test ) values for P = 0.95 to 0.99 are listed in Table 2. In this table i = 1, 2,, n is the number of an OOS test result for a quarry. Answers to the question is the OOS test result metrologically related? are presented in the last column of the table. Only 2 out of 20 OOS test results indicate decidedly that the TSP concentration in ambient air violates the national regulations. The other 18 OOS test results may be caused by metrological problems. Table 2 Deviations D OOS of OOS test results c test from the upper specifications limit c u.s.l in comparison with the expanded measurement uncertainty U(c test ). Quarry OOS test results D OOS /mg m 3 U(c test )/mg m 3 Metrologically i c test /mg m 3 P = 0.95 P = 0.99 related? 1 1 0.210 0.010 0.029 0.044 Maybe 2 0.210 0.010 0.029 0.044 Maybe 3 0.204 0.004 0.029 0.043 Maybe 4 0.231 0.031 0.032 0.049 Maybe 5 0.210 0.010 0.029 0.044 Maybe 6 0.224 0.024 0.031 0.047 Maybe 7 0.223 0.023 0.031 0.047 Maybe 2 1 0.223 0.023 0.031 0.047 Maybe 2 0.288 0.088 0.040 0.060 No 3 0.211 0.011 0.030 0.044 Maybe 4 0.204 0.004 0.029 0.043 Maybe 5 0.255 0.055 0.036 0.054 No 6 0.215 0.015 0.030 0.045 Maybe 7 0.211 0.011 0.030 0.044 Maybe 8 0.216 0.016 0.030 0.045 Maybe 9 0.226 0.026 0.032 0.047 Maybe 10 0.225 0.025 0.032 0.047 Maybe 11 0.232 0.032 0.032 0.049 Maybe 3 1 0.206 0.006 0.029 0.043 Maybe 2 0.218 0.018 0.031 0.046 Maybe

1952 I. KUSELMAN et al. B-1-5 Risks of stone producer and inhabitant Global producer s risk R p is here the probability that the satisfactory quality of air (c true c u.s.l ) will be determined falsely as violating the national regulations since c test > c u.s.l. Corresponding global consumer s/inhabitant s risk R c is the probability that air quality violating the national regulations, when c true > c u.s.l, will be accepted falsely as conforming, since c test c u.s.l. These risks were estimated by eqs. 3 and 4, where the f(c true ) was modeled by lognormal pdf by eq. 9 with parameters μ and σ shown in Fig. 1 and Table 1, whereas the likelihood function f(c test c true ) was approximated by a normal pdf with μ = c true and corresponding σ = u(c test ) for every c true : (10) The warning lines c w.l to the upper specification limit were c w.l = c u.s.l U(c test ) = c u.s.l 0.14c test = 0.200/(1 + 0.14) = 0.175 mg m 3 or c w.l = c u.s.l 0.21c test = 0.200/(1 + 0.21) = 0.165 mg m 3 for P = 0.95 and 0.99, respectively. When a test result exceeds the warning lines, i.e., c test > c w.l, the sampling and measuring systems should be checked and a decision to repeat the test may be made. The action lines c a.l to the upper specification limit were c a.l = c u.s.l + U(c test ) = c u.s.l + 0.14c test = 0.200/(1 0.14) = 0.233 mg m 3 or c a.l = c u.s.l + 0.21c test = 0.200/(1 0.21) = 0.253 mg m 3 for P = 0.95 and 0.99, respectively. When a test result exceeds the action lines, i.e., c test > c a.l, the air quality is violated. Results of R p estimation for different c u.a.l are displayed in Fig. 3 by solid line 1, while R c estimation results are shown by solid line 2. The upper specification limit is presented by a dotted line. The risks for quarry 1 are shown in Fig. 3a, quarry 2 in Fig. 3b, and quarry 3 in Fig. 3c. Acceptance limits for the range of the levels of confidence P = 0.95 to 0.99 are indicated by grey bars. The left one is the warning lines c w.l, while the right one is the action lines c a.l. When acceptance limits are not in use and c u.a.l = c u.s.l, R p and R c are equal to 0.008 and 0.006, respectively, for quarry 1; 0.016 and 0.011 for quarry 2; and 0.006 and 0.005 for quarry 3. This means for quarry 1, for example, that the producer may be punished mistakenly in 8 cases of the ambient air testing from 1000, while violation of the national regulations may be not determined in 6 cases of the testing from 1000. From Fig. 3, one can see as acceptance limits influence the global inhabitant s and producer s risks. For example, for the same quarry 1, when the level of confidence P = 0.95 was chosen and upper acceptance limit c u.a.l was equal to the warning line c w.l = 0.175 mg m 3, risks R p = 0.043 and R c = 0.0003, while when acceptance limit c u.a.l was equal to the action line c a.l = 0.233 mg m 3, R p and R c were already equal to 0.0001 and 0.026, respectively.

IUPAC/CITAC Guide: Investigating OOS test results 1953 Fig. 3 Global producer s risk R p and inhabitant s risk R c in dependence on upper acceptance limit c u.a.l /mg m 3. The risks for quarry 1 are shown in Fig. 3a, quarry 2 in Fig. 3b, and quarry 3 in Fig. 3c. R p is displayed by solid line 1, and R c by line 2. The upper specification limit is shown by the dotted line. Acceptance limits in the range of the levels of confidence P = 0.95 to 0.99 are indicated by grey bars. The left one is warning lines, while the right one is action lines. Reproduced from ref. [29] by permission of Springer. EXAMPLE 2. MULTICOMPONENT OOS TEST RESULTS: PESTICIDE RESIDUES IN TOMATOES B-2-1 Introduction The objective of this example was an application of the metrological approach in the field of food analysis for investigating multicomponent OOS test results of pesticide residuals in tomatoes [33]. In this field, OOS test results are c test > MRL, where MRL is a national legal maximum residue limit expressed in mg of the residue in kg of tomatoes (mg kg 1 ) [34], known also in the United States as a tolerance [35]. Investigated data were obtained during a year by the Israel Laboratory for Pesticide Residue Analysis. The laboratory has participated successfully in 22 proficiency testing rounds of the Food Analysis Performance Assessment Scheme for pesticide residues and has been accredited by the national laboratory accreditation authority (ISRAC). Periodically, ISRAC assesses the test method validation data, metrological traceability chains of the measurement/test process, and measurement uncertainty. Nevertheless, when an OOS test result is obtained, the question about causes of the result arises: is it because of a laboratory s metrological problem or a problem of a farmer/producer of the commodity violating the national legal limits? The European guidance [34] requires, in the case of official food control by regulatory authorities, to check compliance with MRL assuming the lower limit of the uncertainty interval c test U(c test )

1954 I. KUSELMAN et al. with U(c test )/c test = 0.50 or 50 %, if a laboratory proves its own calculated uncertainty to be less than 50 %. In other words, this requirement sets an acceptance limit c u.a.l for test results, meaning that the concentration of pesticide residues in a sample does not violate the national legal limits when c test c u.a.l = 2MRL (mg/kg) at the 0.95 level of confidence. What are the global risks to the farmer/producer and buyer/consumer with such an acceptance limit? How can the risks be changed at the same measurement uncertainty? Answers to these questions are discussed further using the case study of tomatoes. B-2-2 Experimental Sampling was conducted by certified inspectors according to the Codex Sampling Guidelines [36] directly from the field, packing houses, and logistic centers before sending the product to the market. Laboratory samples of tomatoes were in the amount of 1 kg. Sample preparation for gas chromatography (GC) was performed by the known method based on extraction of analytes with acetone from a test portion of 15 g. For liquid chromatography (LC), sample preparation was performed by the method employing acetonitrile extraction from 10 g test portions. The test portions were taken from the homogenized (blended) laboratory samples. The extracts were analyzed by GC methods with flame photometric and halogen selective detectors, as well as with mass spectrometry (MS). Electron ionization was applied in the MS in full scan. LC of the extracts was performed by LC/tandem MS method with a triple quadrupole instrument and electrospray ionization. A total of 217 reference standards (reference materials) for calibration of the chromatographs and for quality control purposes (125 for GC, 45 for LC, and 47 for both GC and LC methods) were used for simultaneous determination of the pesticides, as well as some of their metabolites and degradation products in the samples. The analytical methods used were validated by the validation technique of the European guidance [34]. Relative expanded uncertainty U(c test )/c test, including sample preparation and measurement/analytical components, was evaluated from the ongoing validation data. When averaged for all analytes, it was about 39 % with the coverage factor 2 at the 0.95 level of confidence. Therefore, the intralaboratory value of U(c test )/c test = 39 % was replaced for 50 % according to the European guidance [34] and used in the following discussion when the product is not yet marketed. B-2-3 Global distribution In 46 out of 169 tested samples, i.e., in f 0 = 46/169 = 0.272 or 27.2 % of them, no pesticide residues were found. A total of 39 analytes from 130 pesticides, authorized for use in Israel for tomatoes cultivation, were determined in 1 f 0 = 0.728 or 72.8 % of samples (123 out of 169). The analyte names, their occurrence (numbers of samples in % of 169), test results, and MRL values by the national regulations are listed in Table 3. Five (n = 5) out of N = 169 were OOS test results, and three of them violated the national legal limits by the 2MRL criterion of the European guidance [34].

IUPAC/CITAC Guide: Investigating OOS test results 1955 Table 3 Analytes, their occurrence, test results c test and the national MRL values in tomatoes. Analyte Occurrence/% c test /mg kg 1 MRL/mg kg 1 Azoxystrobin 2.4 0.04 0.11 0.5 Bifenazate 3.0 0.02 0.04 0.05 Boscalid 3.6 0.01 0.10 0.2 Carbendazim 0.6 0.41 0.1 Carbosulfan 0.6 0.01 0.1 Chlorothalonil 18.3 0.01 1.33 5 Chlorpyrifos 1.8 0.01 0.39 0.5 Clofentezine 0.6 0.13 1 Cymoxanil 0.6 0.02 0.05 Cypermethrin 0.6 0.08 0.5 Cyprodinil 2.4 0.02 0.31 0.5 Diafenthiuron 0.6 0.05 0.05 Diazinon 0.6 0.03 0.5 Diethofencarb 1.2 0.01 0.04 0.1 Difenoconazole 0.6 0.09 0.1 Dimethoate 0.6 0.01 1 Endosulfan 0.6 0.08 0.5 Fenazaquin 1.2 0.02 0.33 0.1 Fenhexamid 0.6 0.03 0.5 Fludioxonil 1.8 0.01 0.11 0.3 Folpet 0.6 0.20 0.5 Iprodione 3.0 0.16 0.76 5 Iprovalicarb 0.6 0.04 0.05 Lufenuron 0.6 0.02 0.05 Mepanipyrim 0.6 0.11 0.1 Metalaxyl 3.0 0.01 0.09 0.5 Metominostrobin 1.2 0.01 0.14 0.2 Myclobutanil 1.2 0.06 0.08 0.3 Novaluron 0.6 0.02 0.2 Penconazole 3.0 0.03 0.11 0.2 Propargite 0.6 0.10 2 Pyrimethanil 0.6 0.03 0.05 Spiromesifen 3.6 0.01 0.28 1 Tebuconazole 1.8 0.03 0.11 0.2 Tebufenpyrad 0.6 0.03 0.05 Tetradifon 1.2 0.01 0.07 1 Thiamethoxam 0.6 0.03 0.02 Triadimenol 5.9 0.01 0.19 0.5 Trifloxystrobin 2.4 0.03 0.75 0.2 In order to enable analysis of multiresidue data as a statistical sample from the same population for different pesticide residues, the test results c test were expressed in parts of corresponding MRL using a new dimensionless parameter c r : c r = c test /MRL (11) Such a transformation led to universal characterization of a concentration of any analyte in a sample from the point of view of the concentration adjacent to MRL. When c test = 0, c r = 0 also, and when c test = MRL, c r = 1. Any c r > 1 indicates an OOS test result, and c r > 2 denotes an OOS test result violating regulations by the European guidance [34], etc. The histogram of the experimental data in Fig. 4

1956 I. KUSELMAN et al. Fig. 4 Histogram of c r values and pdf of the Weibull distribution. The pdf of c r calculated by eq. 12 is shown by the solid line; MRL by the dotted vertical line; the range of OOS test results by the dotted pointer. Reproduced from ref. [33] by permission of Springer. was plotted for c r > 0, i.e., for 123 samples in which one or more pesticide residues were detected, identified, and quantified. The pdf of the Weibull distribution used for modeling the global distribution of true c r values c true, shown in Fig. 4 by solid line, was (12) where mˆ = 0.652 and βˆ = 0.204 are the maximum likelihood estimates of the shape and scale parameters, respectively. The MRLs are displayed in Fig. 4 by a dotted vertical line at c r = 1, common for all analytes. The range of OOS test results is indicated by a dotted pointer. B-2-4 Causes and probability of OOS test results Analytes and other details of the observed OOS test results are presented in Table 4. Note, occurrences of these analytes (Table 3) were minimal. Any OOS test result can indicate a pesticide concentration in the tomato sample violating the national legal limits, or be caused by measurement (metrological) problems, i.e., be metrologically related, especially when a specific analyte (a kind of measurement) is rare relatively to others. Dividing both terms of the condition of a metrologically related OOS test result c test MRL U(c test ) by the MRL leds to the following requirement: c r 1/[1 U(c test )/c test ]. Therefore, if U(c test )/c test = 0.50 in compliance with the European guidance [34], OOS test results can be classified as metrologically related at the level of confidence P = 0.95, when c r 2. For the same relative standard uncertainty u(c test )/c test = 0.50/2 = 0.25 or 25 %, the expanded uncertainty U(c test )/c test achieves 0.25 3 = 0.75 (75 %) at the coverage factor 3 corresponding to the

IUPAC/CITAC Guide: Investigating OOS test results 1957 level of confidence P = 0.99. In such a case, a metrologically related OOS test result may appear up to c r 4. When c r > 1/[1 U(c test )/c test ], the OOS test result was not caused by metrological problems. Answers to the question is the OOS test result metrologically related? are presented in the last column of Table 4. Only carbendazim residue in the sample can be classified as definitely (with more than 0.99 confidence) caused by a farmer s/producer s problem violating the national legal limit. The other OOS test results may be metrologically related with different probabilities. Table 4 Comparison of the OOS test results c test with MRL (c r ) and the expanded measurement uncertainty at different levels of confidence P. OOS test results Metrologically related?* Analyte c test /mg kg 1 c r P = 0.95 P = 0.99 Carbendazim 0.41 4.1 No No Fenazaquin 0.33 3.3 No Maybe Mepanipyrim 0.11 1.1 Maybe Maybe Thiamethoxam 0.03 1.5 Maybe Maybe Trifloxystrobin 0.75 3.8 No Maybe * No is for P = 0.95 when c r > 2, and for P = 0.99 when c r > 4. Using the Weibull distribution modeling the global empirical c r distribution in Fig. 4, one can calculate probability P OOS of OOS test results by eq. 2 for an upper specification limit, where Φ is the cumulative Weibull function and c u.s.l is equivalent to c r = 1. When the distribution parameters are mˆ = 0.652 and βˆ = 0.204, the probability is P OOS = 0.06. Therefore, the annual number of OOS test results may be larger than the observed frequency F = n/n = 5/169 = 0.03. B-2-5 Risks of tomato producer and consumer Since pesticide residues were not detected in every sample, the global risk R p of tomato producer/farmer and the global buyer s/consumer s risk R c were calculated here after the following modification of eqs. 3 and 4: (13) where 1 f 0 = 0.728 is the frequency/probability of a pesticide residue detection in a tomato sample (c test > 0). The global pdf f(c true ) was modeled by the Weibull pdf with eq. 12, whereas the likelihood function f(c test c true ) was approximated by a normal pdf as in eq. 10 having the mean μ = c true /MRL = c r and the standard deviation σ = u(c r ) = u(c test )/c test for every c true. Simultaneously, MRL in the role of integration limits in eqs. 13 and 14 was replaced for c r = 1, and acceptance limits c u.a.l were expressed also in c r values. The values of the upper acceptance limit c u.a.l for test results lower than MRL under the measurement expanded uncertainty were used as warning lines: c w.l = c test = MRL U(c test ) = MRL 0.5c test = MRL/1.5 = 0.67MRL mg kg 1 or c r = 0.67 for P = 0.95, and c w.l = c test = MRL 0.75c test = MRL/1.75 = 0.57MRL mg kg 1 or c r = 0.57 for P = 0.99. When a test result exceeds the warning lines, i.e., c test > c w.l, the sample preparation and measurement/analytical systems should be checked and a decision to repeat the test may be made. (14)

1958 I. KUSELMAN et al. Acceptance limit values larger than MRL by the measurement expanded uncertainty were used as action lines: c a.l = c test = MRL + U(c test ) = MRL + 0.5c test = 2MRL mg kg 1 or c r = 2 for P = 0.95, as required in the European guidance [34], and c a.l = c test = MRL + 0.75c test = 4MRL mg kg 1 or c r = 4 for P = 0.99. When a test result exceeds the action lines, i.e., c test > c a.l, the quality of tomatoes is violated with corresponding probabilities 0.95 or 0.99. Results of R p estimation for different c u.a.l are displayed in Fig. 5 by solid line 1, while R c estimation results are shown by solid line 2. The c u.a.l values were expressed here in parts of MRL, i.e., as c r. Acceptance limits in the range of the levels of confidence P = 0.95 to 0.99 were indicated by grey bars. The left one is warning lines c w.l, while the right one is action lines c a.l. Fig. 5 Global producer s risk R p and consumer s risk R c in dependence on upper acceptance limit c u.a.l. R p is displayed by solid line 1, and R c by solid line 2. The c u.a.l values are expressed in parts of MRL (as c r ). Acceptance limits in the range of the levels of confidence P = 0.95 to 0.99 are demonstrated by grey bars. The left one is warning lines, while the right one is action lines. Reproduced from ref. [33] by permission of Springer. From Fig. 5, one can see as acceptance limits influence the global producer s and consumer s risks. For example, when the level of confidence P = 0.95 is chosen and acceptance limit c u.a.l is equal to the warning line c w.l = 0.67MRL mg kg 1 or c r = 0.67, the risks are R p = 0.040 and R c = 0.001, while when acceptance limit c u.a.l is equal to the action line c a.l = 2MRL mg kg 1 or c r = 2, R p and R c are already 1 10 7 and 0.034, respectively. When acceptance limits are not in use, c u.a.l = MRL and c r = 1 (dotted line in Fig. 5), e.g., according to the system of tolerances of EPA [35], R p and R c are equal to 0.008 and 0.007, respectively. That means the farmer/producer may be punished mistakenly in 8 cases of the tomatoes testing from 1000, while violation of the national regulations may be not determined in 7 cases of the testing from 1000.

IUPAC/CITAC Guide: Investigating OOS test results 1959 EXAMPLE 3. OOS TEST RESULTS IN LONG-TERM STABILITY STUDY OF DRUG PRODUCTS B-3-1 Introduction The objective of this example was an application of the metrological approach in pharmaceutical field for investigating OOS test results in a stability study of drug products [37]. When stability of a stored drug product is studied, it is important to establish a retest period or shelf life of the product, during which its properties are not influenced as yet and the drug can be used according to a physician prescription. International harmonized guideline ICH Q1E [38] recommends the establishment of a retest period or shelf life for a drug product using regression analysis of stability data (e.g., assay results vs. time) accumulated during long-term storage of the product. For a measured attribute (property) of the product known to increase with time, the regression one-sided upper 0.95 confidence limit should be compared to the acceptance criterion. The retest period or shelf life is estimated as the earliest time at which the confidence limit intersects the criterion. A similar rule is recommended for a measured property of the product known to decrease with time. The regression onesided lower 0.95 confidence limit should be compared in such a case to the acceptance criterion. The acceptance criterion may be formulated as a requirement to an amount-of-substance analyte concentration in a product not to exceed the upper specification limit c u.s.l, or not to be less than the lower specification limit c l.s.l. However, true values of the concentration c true are unknown, and test results c test are affected by the measurement uncertainty. Therefore, OOS test results c test > c u.s.l or c test < c l.s.l in a stability study can indicate an actual change (e.g., degradation) of the product or be metrologically related with a certain confidence probability P, i.e., be caused by the measurement problems, though the product still meets the quality requirements at the time of testing. As examples, the test results of sodium chloride injection in plastic containers and epinephrine injection in ampoules, accumulated in the Research & Quality Control Laboratory of the Medical Corps, the Israel Defense Forces, are discussed. The sodium chloride assay specification limits are c l.s.l = 95.0 % and c u.s.l = 105.0 % of the labeled amount [39], whereas the labeled amount is, for example, 0.9 % weight per volume, i.e., 0.9 g of sodium chloride in 100 ml of the solution. During long-term storage, an amount of water permeates from inside the container into the over-wrap space due to evaporation through the plastic. The water loss increases the sodium chloride concentration with time. Therefore, the test results c test of sodium chloride concentration in the stored product (relative also to the labeled amount) were compared with the upper specification limit c u.s.l = 105.0 %. The assay specification limits for epinephrine injection in ampoules are c l.s.l = 90.0 % and c u.s.l = 115.0 % of the labeled amount of L-adrenaline, i.e., (L)-4-[1-hydroxy-2-(methyl- amino)ethyl]benzene-1,2-diol [40], whereas the labeled amount is, for example, 1 mg ml 1. L-adrenaline in the solution is subject to degradation during long-term storage, caused by oxidation, sulfonation, and racemization. Products of these reactions, including D-adrenaline and adrenaline sulfonate, do not have pharmacological activity comparable to L-adrenaline [41]. Therefore, the test results c test of L-adrenaline concentration in the stored product (relative to the labeled amount) were compared with the lower specification limit c l.s.l = 90.0 %. Besides the assay results, concentration of some impurities and other properties of these products should be under control. Thus, the examples with sodium chloride and L-adrenaline assay were used here as a model only for discussion of OOS test results in both situations of the measured product property changes (increasing and decreasing) specified in guideline ICH Q1E [38]. B-3-2 Experimental Samples of 18 batches of sodium chloride injection in 500 ml plastic containers (labeled as 0.9 %) were manufactured by B. Braun Melsungen AG, Germany, and Teva Medical Ltd., Israel. Samples of 93

1960 I. KUSELMAN et al. batches of epinephrine injection in 1 ml ampoules (labeled as 1 mg ml 1 ) were manufactured by Teva Pharmaceutical Industries Ltd., Israel. The samples were stored under controlled conditions recommended by their manufacturers. Choice of these samples does not mean any preference or criticism. Sodium chloride assay was performed by titration with silver nitrate of test portions sampled from a bag. The titration end-point was determined potentiometrically [42] with an automated titrator. L-adrenaline chiral HPLC assay with UV vis detection was performed as described in the paper [41]. The expanded relative measurement uncertainty associated with a routine sodium chloride assay result was evaluated as U(c test )/c test = 0.008 to 0.012 or 0.8 to 1.2 % for normal distribution and the range of the levels of confidence P = 0.95 to 0.99, with the coverage factor k = 2 to 3, respectively. The measurement uncertainty for L-adrenaline assay results evaluated for the same conditions was U(c test )/c test = 0.060 to 0.090 or 6.0 to 9.0 %. B-3-3 Regression analysis and shelf life of the products Results of regression analysis of the accumulated data c test vs. time τ of the product storage are demonstrated in Fig. 6 for the sodium chloride injection and in Fig. 7 for the epinephrine injection. Fig. 6 Specification limits and shelf life of the sodium chloride injection. The ordinate is c test /% axis; c u.s.l /% is displayed by the dotted line 1; c l.s.l /% by line 2 coincided with the abscissa, i.e., storage time τ/a (years); the regression is shown by solid line 3; the one-sided upper 0.95 confidence limit to this line is indicated by thin line 4. The product shelf life is shown by solid pointer 5. Grey bar 6 illustrates the corridor of test results c test for the levels of confidence P = 0.95 to 0.99. Dotted pointers 7 and 8 indicate corresponding τ values. Grey bar 9 demonstrates the corridor of OOS test results at P = 0.95 to 0.99. Dotted pointers 10 and 11 show the storage time values corresponding to the corridor borders. Reproduced from ref. [37] by permission of Springer.

IUPAC/CITAC Guide: Investigating OOS test results 1961 Fig. 7 Specification limits and shelf life of the epinephrine injection. c u.s.l and c l.s.l are displayed by dotted lines 1 and 2, respectively. Other symbols and signs are the same as in Fig. 6. Reproduced from ref. [37] by permission of Springer. One-sided 0.95 confidence limit to the linear regression lines ĉ test (τ) was calculated by the known formulas: c c.l = ĉ test ± t(0.95, ν)s c (15) t(0.95, ν) is the quantile of t distribution (Student s distribution) for the level of confidence 0.95, onesided interval, and the number of degrees of freedom ν = N 2, N is the number of observed test results c test used in the regression analysis, s c is the standard deviation of the predicted ĉ test, and τ is the mean of the τ range (the mean storage time). The optimal range of storage time τ values for the study was estimated from eq. 15 as the range providing the minimal s c when τ = τ 0 = τ, where τ 0 is the shelf life of the product. Such a range at any symmetrical distribution of τ is from 0 to 2τ 0. According to the manufacturer recommendations (known a priori ), the shelf life for sodium chloride injection was 3 a and the optimal range of the storage time was 2 3 = 6 a. In practice, there were not any test results during 4.5 to 6 a of storage, and the studied range was limited by 4.5 a as shown in Fig. 6. The manufacturer recommendation concerning the shelf life of the epinephrine injection was 1.5 a, and the optimal range of the studied storage time was 2 1.5 = 3 a as in Fig. 7. The problem also is, eq. 15 does not have any direct solution τ(c) for calculation of the actual shelf life τ 0 in a general form, as the earliest time at which the confidence limit intersects the critical c test value. Therefore, the confidence limit was approximated by parabola ĉ c.l = b 2 τ 2 + b 1 τ + b 0. This approximation allowed calculation of the shelf life using the parabola roots: The sign + was used for the square root in the case of upper specification limit, while the sign was necessary in the case of lower specification limit. (16)

1962 I. KUSELMAN et al. B-3-3-1 Shelf life of sodium chloride injection The upper specification limit c u.s.l is displayed in Fig. 6 by dotted line 1. The lower specification limit c l.s.l (line 2) coincides with the abscissa. The linear regression for N = 18 observations was ĉ test = 2.2837τ + 97.982 with the squared correlation coefficient r 2 = 0.7192 (shown by solid line 3). The onesided upper 0.95 confidence limit c c.l to the regression line 3, shown as thin line 4 in Fig. 6, was approximated by parabola ĉ c.l = 0.1717τ 2 + 1.5319τ + 99.587, r 2 = 0.9999. The τ 0 value was calculated by eq. 16 where the parabola coefficients were b 2 = 0.1717 % a 2, b 1 = 1.5319 % a 1, and b 0 = 99.587 %, while ĉ c.l = c test = c u.s.l = 105.0 %. Thus, the actual product retest period or shelf life for the sodium chloride injection in the storage conditions of the Defense Forces was τ 0 = 2.7 a. It is shown in Fig. 6 by solid pointer 5. The calculated τ 0 value is close to the 3 a recommended by both the injection manufacturers. However, c test may be less than c u.s.l because of the measurement uncertainty. For example, the grey bar 6 in Fig. 6 illustrates the corridor of values c test = 105.0 U(105.0) from 104.2 to 103.7 % at the levels of confidence P = 0.95 to 0.99, respectively. Dotted pointers 7 and 8 indicate the storage time values of 2.4 and 2.2 a corresponding to these levels of confidence (the bar borders). In other words, there is a chance of a significant change of the injection quality after 2.2 a of the storage. Metrologically related OOS test results c test > c u.s.l appear up to c test = 105.0 + U(105.0). The grey bar 9 in Fig. 6 demonstrates the corridor of such OOS test results from 105.8 to 106.3 % at the levels of confidence P = 0.95 to 0.99, respectively. Dotted pointers 10 and 11 show the storage time values of 3.0 and 3.2 a corresponding to the corridor/bar borders. Thereby, even after 3.2 a of the storage, the injection quality may be satisfactory. OOS test results greater than 106.3 % would unlikely be classified as metrologically related but are, as a rule, evidence of the product change. B-3-3-2 Shelf life of epinephrine injection The assay upper and lower specification limits are shown in Fig. 7 by dotted lines 1 and 2, respectively. The linear regression for N = 93 observations displayed as solid line 3 in Fig. 7 was ĉ test = 10.157τ + 106.18 with r 2 = 0.8632. The parabolic approximation of the lower one-sided 0.95 confidence limit (line 4) was ĉ c.l = 0.2950τ 2 9.1646τ + 104.76, r 2 = 0.99999. Substituting the coefficients of this parabola in eq. 16 for ĉ c.l = c test = c l.s.l = 90.0 %, one can calculate the actual time value τ 0 indicated in Fig. 7 by solid pointer 5. The product retest period or shelf life calculated for the epinephrine injection ampoules was τ 0 = 1.5 a for the Defense Forces storage conditions. This is exactly what was recommended by the manufacturer. As in the situation with the sodium chloride injection, c test may exceed here c l.s.l because of the measurement uncertainty. For example, the grey bar 6 in Fig. 7 demonstrates the corridor of such test results c test = 90.0 + U(90.0) from 95.4 to 98.1 % at the levels of confidence P = 0.95 to 0.99, respectively. Dotted pointers 7 and 8 show the storage time values of 1.0 and 0.7 a corresponding to the corridor borders. Therefore, there is the necessity in increasing quality control measures after 0.7 a of the product storage. Metrologically related OOS test results appear for epinephrine injection when c test < c l.s.l. The grey bar 9 in Fig. 7 illustrates the corridor of values c test = 90.0 U(90.0) from 84.6 to 81.9 % at the levels of confidence P = 0.95 to 0.99, respectively. Dotted pointers 10 and 11 show the storage time values of 2.1 and 2.3 a corresponding to these levels of confidence (the bar borders). Thus, even after 2.3 a of the storage, the injection quality may be satisfactory. OOS test results smaller than 81.9 % would unlikely be classified as metrologically related but provide an indication of the product degradation. B-3-4 Risks of setting a shelf life B-3-4-1 When a measured attribute increases with time When a decision is to be made about a shelf life of a product in which the measured attribute increases with time, such as the assay of the sodium chloride injection, there are two risks. One of them is a prob-

IUPAC/CITAC Guide: Investigating OOS test results 1963 ability that the product still meeting the quality requirements (c true c u.s.l ) will be falsely determined as violating the upper specification limit, since c test > c u.s.l. This is the global manufacturer s/producer s risk R p, and a product owner (e.g., the Defense Forces) is in the role of producer. There is also a probability that the product quality violating the upper specification limit (c true > c u.s.l ) will be falsely accepted as conforming, since c test c u.s.l. This is the global consumer s risk R c, and a patient is in the role of consumer. As in the previous examples, the global producer s and the consumer s risks were estimated by eqs. 3 and 4. However, the global c true pdf was modeled in this case by the Student s pdf with location parameter ĉ test and scale parameter s c calculated from the regression data ĉ test (τ) at corresponding time τ: (17) where Γ is the Gamma function, t = (c test c true )/s c is the normalized variable c test, and c true = ĉ test. The likelihood function is here a normal pdf of measurement/test results c test for a product with the true value of the measured property c true, mean value of the test results equal to each c true which can be assumed by the c true pdf (in practice to each c true = ĉ test ) and corresponding standard deviation s = u(c test ): R c and R p values vs. c u.a.l for sodium chloride injection assay and the c true Student s pdf with ν = 16 degrees of freedom, referring to time τ 0 = 2.7 a, are displayed in Fig. 8 by solid lines 1 and 2, respectively. The upper specification limit is presented by a vertical dotted line. The range of OOS test results is shown by a horizontal dotted pointer. Acceptance limits for the levels of confidence P = 0.95 to 0.99 are indicated by grey bars 3 and 4, similar to bars 6 and 9 in Fig. 6, respectively. These limits can be interpreted and used respectively as warning and action lines in quality control charts. When c u.a.l = c u.s.l, R c and R p are equal to 0.015 and 0.067, respectively. This means that the violation of the upper specification limit may be not determined in 15 cases of the testing from 1000, while the sodium chloride injection may be mistakenly found as not acceptable for use in 67 cases of the product testing from 1000. When the level of confidence P = 0.95 is chosen, for example, and the acceptance limit is equal to the warning line c w.l = 104.2 %, the risks are R c = 0.0004 and R p = 0.460. Acceptance limit equal to the action line c a.l = 105.8 % leads to R c and R p of 0.043 and 0.001, respectively. Different risk values R c and R p correspond also to every product storage timeτ, since c true and c test change with τ, as discussed above. The dependences of R c and R p on τ for the sodium chloride injection are displayed in Fig. 9 by solid lines 1 and 2, respectively, where the R c and R p values were calculated with respect to an acceptance limit equal, for each τ, to the relevant one-sided upper 0.95 confidence limit to the regression line. Time corresponding to the upper specification limit, τ u.s.l = τ 0 = 2.7 a, is shown by vertical dotted line, while the range of time values τ OOS leading to OOS test results is indicated by horizontal dotted pointer. Grey bars 3 and 4 present the time acceptance limits for the levels of confidence P = 0.95 to 0.99 (time warning and action lines, respectively) corresponding to bars 6 and 9 in Fig. 6. It is clear that setting the shelf life τ 0 > 2.7 a increases significantly the consumer s risk R c. (18)

1964 I. KUSELMAN et al. Fig. 8 Global risks of consumer R c and of producer R p of the sodium chloride injection vs. upper acceptance limit c u.a.l /% for test results.r c and R p are displayed by solid lines 1 and 2, respectively; c u.s.l is presented by the vertical dotted line. The range of OOS test results is shown by the horizontal dotted pointer. Acceptance limits for the levels of confidence P = 0.95 to 0.99 are indicated by grey bars 3 (warning lines) and 4 (action lines). Reproduced from ref. [37] by permission of Springer. Fig. 9 Global risks of consumer R c and producer R p of the sodium chloride injection vs. storage timeτ/a. R c and R p are displayed by solid lines 1 and 2, respectively. Time τ u.s.l corresponding to c u.s.l is shown by the vertical dotted line, while the range of time values τ OOS led to OOS test results indicated by the horizontal dotted pointer. Grey bars 3 (warning lines) and 4 (action lines) present the time acceptance limits for the levels of confidence P = 0.95 to 0.99. Reproduced from ref. [37] by permission of Springer. B-3-4-2 When a measured attribute decreases with time When a decision is to be made about the shelf life of a product whose measured attribute decreases with time, like the epinephrine injection assay, the risks can be estimated by eqs. 5 and 6. Results of calculation of R c and R p vs. lower acceptance limit c l.a.l for epinephrine injection assay and the global c true Student s pdf with ν = 91 degrees of freedom, relevant to time τ 0 = 1.5 a, are shown in Fig. 10. The

IUPAC/CITAC Guide: Investigating OOS test results 1965 Fig. 10 Global risks of consumer R c and producer R p of the epinephrine injection vs. lower acceptance limit c l.a.l /% for test results. Symbols and signs are the same as in Fig. 8. Reproduced from ref. [37] by permission of Springer. pointer of OOS test results in Fig. 10 is directed in the opposite verse to the corresponding in Fig. 8: for the epinephrine injection, the direction is c test < c l.s.l, whereas for the sodium chloride injection it is c test > c u.s.l. Corresponding consumer s risk (solid line 1) decreases for the epinephrine injection in Fig. 10 when the acceptance limit increases, in contrast to the situation with the sodium chloride injection in Fig. 8. A similar difference in the behavior of the producer s risks (solid lines 2) is also observed. The risk values vs. τ are demonstrated in Fig. 11 with respect to an acceptance limit equal for each τ to the corresponding value of the one-sided lower 0.95 confidence limit to the regression line. Fig. 11 Global risks of consumer R c and producer R p of the epinephrine injection vs. storage time τ/a. Symbols and signs are the same as in Fig. 9. Reproduced from ref. [37] by permission of Springer.