Chapter 8 Hyper-Pure Germanium Detector

Similar documents
Chapter 8 Hyper-Pure Germanium Detector

Chem 481 Lecture Material 3/20/09

Gamma and X-Ray Detection

Semiconductor Detectors

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Semiconductor Detectors

Compton suppression spectrometry

Outline. Radiation Detectors. Radiation Detectors 1

ORTEC. Review of the Physics of Semiconductor Detectors. Interaction of Ionizing Radiation with Semiconductor Detectors. Heavy Charged Particles

Chapter 4 Scintillation Detectors

Compton Camera. Compton Camera

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Quality Assurance. Purity control. Polycrystalline Ingots

EEE4106Z Radiation Interactions & Detection

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Germanium Detectors. Germanium, a special material. Detectors, big is beautiful. Operational features. Applications. Iris Abt

DETECTORS. I. Charged Particle Detectors

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure

Advantages / Disadvantages of semiconductor detectors

Semiconductor-Detectors

Chapter Seven (Nuclear Detectors)

Chapter 5 Gamma-ray Spectrometry with Scintillators

III. Energy Deposition in the Detector and Spectrum Formation

ORTEC. Overview of Semiconductor Photon Detectors. GEM (P-type) and GAMMA-X (N-type) Coaxial Detectors. Efficiency as a Function of Energy

Alpha-particle Stopping Powers in Air and Argon

Radiation Detection and Measurement

Efficiency and Attenuation in CdTe Detectors

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

Chap. 11 Semiconductor Diodes

The Best Gamma-Ray Detector

Advanced Pulse-Shape Analysis and Implementation of Gamma-Ray Tracking in a Position-Sensitive Coaxial HPGe Detector

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Chapiter VII: Ionization chamber

Airo International Research Journal October, 2015 Volume VI, ISSN:

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

High Purity Germanium Detector Calibration at ISOLDE

Radiation Dose, Biology & Risk

1 Name: Student number: DEPARTMENT OF PHYSICS AND PHYSICAL OCEANOGRAPHY MEMORIAL UNIVERSITY OF NEWFOUNDLAND. Fall :00-11:00

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Silicon Detectors in High Energy Physics

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Determination of the activity of radionuclides

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Analysis of γ spectrum

Radiation (Particle) Detection and Measurement

Chapter 1 Overview of Semiconductor Materials and Physics

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Characterization of Mechanically Cooled High Purity Germanium (HPGe) Detectors at Elevated Temperatures

Chemical Engineering 412

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Application Note. The Continuous Air Monitoring (CAM) PIPS Detector Properties and Applications

MEASUREMENT AND DETECTION OF RADIATION

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract

Simulation of 4π HPGe Compton-Suppression spectrometer

Radionuclide Imaging MII Detection of Nuclear Emission

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

ORTEC Experiment 7. High-Resolution Gamma-Ray Spectroscopy. Equipment Required. Purpose. Introduction

ORTEC. SLP Series Silicon Lithium-Drifted Planar Low-Energy X Ray Detector Product Configuration Guide

NUCL 3000/5030 Laboratory 2 Fall 2013

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Researchers at the University of Missouri-Columbia have designed a triple crystal

Chapter 3 Gas Filled Detectors

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

ORTEC AN34 Experiment 10 Compton Scattering

Gamma Spectroscopy. References: Objectives:

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Lecture 12. Semiconductor Detectors - Photodetectors

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

A Comparison between Channel Selections in Heavy Ion Reactions

The outline. 1) Detector parameters: efficiency, geometrical acceptance, dead-time, resolution, linearity. 2) gaseous ionization chambers

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

Detection and measurement of gamma-radiation by gammaspectroscopy

3.1 Introduction to Semiconductors. Y. Baghzouz ECE Department UNLV

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Gamma Ray Spectrometry

Performance of a Si PIN photodiode at low temperatures and in high magnetic fields

Solid State Detectors

The Ones That Got Away: Neutron Leakage Spectra

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Note that it is traditional to draw the diagram for semiconductors rotated 90 degrees, i.e. the version on the right above.

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

anti-compton BGO detector

EECS130 Integrated Circuit Devices

Queen s University PHYS 352

Fundamentals of Radionuclide Metrology

Modern Physics Laboratory (Physics 6180/7180)

Radioactivity. Lecture 6 Detectors and Instrumentation

Transcription:

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-1 Chapter 8 Hyper-Pure Germanium Detector 8.1. Introduction Silicon semiconductor detectors described in the previous chapter have depletion depths less than 1 mm, which are sufficient for charged particle spectroscopy or soft X-ray detection. For photon spectroscopy in the energy region of hundred kev or several MeV, much thicker semiconductor detectors are required. Fig. 8.1 shows the mean free path of a photon in silicon and germanium. It is obvious that a depletion depth of at least several cm is required. Due to its higher atomic number, Ge has a much lager linear attenuation coefficient, which leads to a shorter mean free path. Thus, Ge is preferred for hard X-ray or gamma-ray detection to achieve higher detection efficiency. 10 1 Mean free path [cm] 10 0 10-1 10 - Si Ge 10-3 10 1 10 10 3 10 4 Photon energy [kev] Fig. 8.1. Photon mean free paths in Si and Ge. For a semiconductor detector, the depletion depth is given by ε V d ( en 0 ) 1/ where, V 0 is the reverse bias voltage and N represents the net impurity concentration in the initial semiconductor material. If silicon or germanium of normal semiconductor purity is employed, the maximum achievable depletion depth is a few mm even at bias voltages close to the breakdown level. Thus, the impurity concentration should be much reduced down to 10 10 atoms/cm 3 in order to realize intended depletion depths of cm order. At this impurity concentration, a reverse bias voltage of 1 kv can produce a depletion depth of 1 cm. The required impurity concentration corresponds to levels less than 1 part in 10 1, which is quite challenging. One way to further improve the net impurity concentration is to compensate the residual impurities with an opposite type impurity material. In both Si and Ge, the material with highest available purity tends to be p-type, which requires addition of donor atoms for compensation. The alkali metals like Li, Na tend to form interstitial donors in Si and Ge. When one of these donor materials (practically Li) is introduced into Si or Ge, the donor atoms are easily ionized and are mobile enough to drift at elevated temperatures under the

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8- influence of a strong electric field. This process is so-called lithium ion drifting and has been applied in both Si and Ge. For convenience, Ge(Li) represents lithium drifted Ge and Si(Li) means lithium drifted Si. The lithium mobility is much greater in germanium and remains high enough at room temperature, which leads to an undesirable redistribution of the lithium. Therefore, the lithium profile of Ge(Li) detectors must always be preserved at LN temperature while Si(Li) detectors can be stored at room temperature for a short period due to low mobility of the Li ion in silicon. The other way to improve the net impurity concentration is to add additional refining processes so that the intended purity of the crystal can be met. Techniques have been developed to achieve high purity Ge crystals in this approach and HyperPure Ge (HPGe) detectors are commercially available. However, no equivalent technique is available for Si yet. Ge(Li) detectors were popular in the past and served as the common type of large-volume germanium detector for a long time. Although there is little difference in detector performance between HPGe and Ge(Li), the maintenance of Ge(Li) is pretty inconvenient, which has made manufacturers stop producing them. 8.. Configurations A. Hyper-Pure Germanium (HPGe) detector fabrication [4] HPGe crystal were first developed in the mid 1970s. The starting material is bulk germanium intended for the semiconductor industry. Although already of very high purity, the material is further purified with the zone refining technique. The germanium is melted in a crucible using radio-frequency (RF) heating coils. The underlying principle is that impurities concentrate in the liquid phase leaving the solid purer than the original melt as a liquid freezes and solid appears. Fig. 8.. Three coil zone refine. Fig. 8.3. Growing Ge crystal.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-3 As shown in Fig. 8., each zone refiner coil melts a small section of the germanium in the crucible. As the RF coils are slowly moved along the length of the crucible, the molten zone moves with them. Thus, the germanium melts as the coil approaches and freezes as the coil moves away. The impurities tend to remain in the molten section, which leads to a higher concentration of impurities in the liquid than the solid. In this way, the impurities are swept to one end. This sweeping operation is repeated many times, until the impurities are concentrated at one end of the ingot. This end is then removed, leaving the remaining portion much purer than the original starting material. The improvement or reduction in impurity concentration actually realized is about a factor of 100 or more at the completion of this process. Large single crystals of germanium are grown using the Czochralski technique. A precisely cut seed crystal is dipped into the molten germanium and then withdrawn slowly, while maintaining the temperature of the melt just above the freezing point as shown in Fig. 8.3. The rate of crystal withdrawal and temperature of the melt are adjusted to control the growth of the crystal. If the remaining extremely low-level impurities are acceptors, the property of the crystal is mildly p-type while high purity n-type is the result in the other case. B. Planar configuration An example of a planar HPGe detector using a p-type crystal is shown in Fig. 8.4. In this configuration, the electric contacts are provided on the two flat surfaces of a germanium crystal. 0.5 mm thick Be Photons p + (0.3 µm B) e p-type h p-n junction n + contact (0.7 mm Li) Fig. 8.4. Planar HPGe detector (p-type). A lithium evaporation and diffusion to form the n + contact is performed over one of the surface. This lithium-diffused layer is about several hundred µm thick. The depletion region is formed by reverse biasing n + -p junction. The surface of the opposite side is modified to the p + layer by ion implantation of boron acceptor atoms to increase the conductivity near the surface. Since both materials are p-type, no semiconductor junction

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-4 exists at this side. Instead, the p + layer provides the electric contact to collect the charge carriers created by the radiation. The thickness of the implanted boron layer is so thin (a few tenths of µm) that it is a suitable entrance window for low energy photons. To make a reverse bias, a positive high voltage is applied to the n + contact with respect to the p + surface. The depletion region is formed at the region close to the n + contact and then expanded deeply into the p-side as the bias voltage is raised. Once the detector is fully depleted, further increase of the bias (overvoltage) does not make any effect on the sensitive volume, however, it makes the electric field stronger, which subsequently shortens carrier collection times and reduces the risk of carrier losses (recombination and trapping). Planar HPGe detectors can also be fabricated starting with high-purity n-type crystals. Similarly, n + and p + contacts are provided on both sides. Reverse biasing also requires the application of a positive voltage to the n + side. In this case, the depletion region is expanded from the p + contact. When fully depleted and operated with a sufficient overvoltage, the electric field inside the crystal is almost uniform and charge carriers drift under a constant electric field. C. Coaxial configuration From the photon mean free path in germanium, a detector thickness of the order of 5 cm is required for efficient detection of MeV photons. The maximum depletion depth for the planar detectors is limited to less than 1 or cm. To produce a detector with a thicker depletion depth, a different electrode configuration must be employed. HPGe detectors dedicated for MeV photons are constructed in coaxial geometry as shown Fig. 8.5. In this configuration, one electrode is fabricated at the outer surface of a cylindrical crystal and the other electrode is located at the inner surface of the central hole. In this way, much larger active volumes can be achieved. True coaxial Closed-ended coaxial Fig. 8.5. Large volume coaxial HPGe detectors. A closed-ended configuration is one in which only part of the central core is removed and the outer electrode is extended over one flat side surface. This configuration is the most common type of commercial HPGe detector.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-5 depleted p-n junction p-n junction Fig. 8.6. Extension of the depletion region for a coaxial detector. In principle, the p-n junction of a coaxial detector can be located either at the inner or outer surface of the crystal as shown in Fig. 8.6. If the junction is on the outer surface, the depletion region extends inward as the high voltage bias is increased until the crystal is fully depleted. In the opposite case, the depletion region grows outward. The p-n junction at the outer surface requires a much lower full depletion bias and therefore, is preferred. The practical electrode configurations of coaxial Ge detectors are shown in Fig. 8.7. To make the p-n junction at the outer surface, the n + contact is performed over the outer surface for a p-type detector while the p + contact is applied in case of an n-type crystal. The n-type coaxial detectors are often called reverse electrode detectors. The reverse bias requires a positive outside potential for a p-type and a negative potential for an n- type relative to the central electric potential. p-type n-type p + n + n + contact p + contact Fig. 8.7. Electrode configurations for coaxial detectors. Regardless of the crystal type, a coaxial detector can be used for general gamma-ray spectroscopy purposes. For low energy gamma-rays below 100 kev, a significant portion of the incident photons is attenuated for p-type detectors due to the relatively thick surface dead layer, i.e. lithium n + contact. Thus, p-type coaxial detectors are normally limited down to 50 kev. Since the dead layer thickness of n-type detectors is much

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-6 thinner, n-type detectors are still effective for low energy photons (down to several kev) when coupled to a thin Be entrance window. Therefore, n-type detectors cover quite a wide spectral region compared to p-type coaxial and planer detectors. D. Electric field and capacitance The electric field in HPGe detectors governs the charge carrier collection process. The signal pulse shape and the timing behavior are consequently dependent on the variation of the electric field strength. As discussed in the general properties of semiconductor detectors, the governing equation is Poisson s equation: ρ V = ε The charge density ρ is given as ρ = -en A for p-type crystals and as ρ = en D for n-type crystals. For convenience, p-type crystals are assumed in the following derivation. Planar detectors From the solution derived in the semiconductor detector chapter, the depletion depth corresponding to a reverse bias V 0 is ε V0 1/ ε V0 1/ d ( ) = ( ) en en A where, N represents the impurity concentration in the initial semiconductor material. If the crystal thickness is T, the high voltage bias corresponding to the depletion depth T is T en A Vd = ε In this case, the detector is fully depleted and V d is defined as the depletion voltage. In one-dimensional slab geometry as shown in Fig. 8.8, Poisson s equation becomes T d V ( x) ρ en A = = (0 x d) n+ dx ε ε p-type p+ For a partially depleted voltage, the equation is valid for the depleted region. The boundary condition of the electric potential becomes V ( x = 0) V ( x = d) = V 0 where x = 0 represents the n + contact position. The electric field is obtained by integrating the Poisson s equation: x=0 x=d Fig. 8.8. Partial depletion of a planar detector.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-7 V0 en A d E( x) = + ( x) d ε (0 x d) For a given bias voltage (if V 0 < V d ), this solution is limited in the depleted region and the electric field outside the depleted region is zero. The variation of the electric field 1500 strength for a planar HPGe detector is N A =10 10 cm -3, T=1 cm shown in Fig. 8.9. The depletion voltage for N A = 10 10 cm -3 and T = 1 cm is 565 V. The capacitance per unit area is ε ερ 1/ C = ( ) d V 0 for partial depletion conditions. Once the bias V 0 exceeds V d, the detector capacitance becomes constant and independent of the bias voltage. Coaxial detectors For the true coaxial geometry, Poisson s equation becomes E(x) [V/cm] 1000 500 400 V 100 V 565 V 00 V 1000 V 0 0.0 0. 0.4 0.6 0.8 1.0 X [cm] Fig. 8.9. Electric field variation of a planar detector. d V + 1 dr r dv dr ρ en A = = ε ε For partial depletion conditions, solving for the depletion depth is a little complicated, so that the bias voltage is assumed to be larger than the depletion voltage. Then the boundary condition can be set as E(x) [V/cm] 4000 3000 000 1000 000 V 3000 V V d =1556 V V ( r = V ) V ( r1 ) 0 where r 1 and r represent inner and outer radii. Fig. 8.10. Electric field variation of a true coaxial detector. r The electric field can be set as E ( r) = re ˆ ( r) and is solved as en A V0 ( en A / 4ε )( r r1 ) en A V0 ( en A / 4ε )( r r1 ) E( r) = r E( r) = r + ε r ln( r / r ) ε r ln( r / r ) 1 The depletion voltage V d can be calculated from the condition N A =10 10 cm -3, r 1 =4 mm, r =.5 cm 0 0.0 0.5 1.0 1.5.0.5 r [cm] 1

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-8 en A r r1 E ( r 1 ) = 0 Vd = r1 ln( r / r1 ) ε Fig. 8.10 shows the variation of the electric field for a true coaxial HPGe detector. The capacitance per unit length of a fully depleted true coaxial detector is πε C = ln( r / r 1 ) To make the capacitance as low as possible, the radius of the center core r 1 is kept to a minimum. For the closed-ended coaxial detectors, the Poisson s equation is hard to solve analytically and therefore a numerical method like the Finite Difference Method must be employed. An example of the potential variation and the electric field direction is shown in Fig. 8.11. Fig. 8.11. Potential and electric field distribution of a closed-ended coaxial detector [5]. 8.3. Operational characteristics A. Cryostat and dewar Although silicon detectors can be operated at room temperature in principle, operating germanium detectors at room temperature is impossible due to the small bandgap energy, which generates an intolerable leakage current. Therefore, germanium crystals must be cooled to reduce the leakage current. HPGe detectors are allowed to warm to room temperature when they are not in operation, however, they are always maintained at LN temperature practically.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-9 Fig. 8.1. Modern arrangement of HPGe detector and preamp within the cryostat housing. The Ge crystal is encased in a vacuum tight container (cryostat) to reduce thermal conduction between the crystal and the air. The basic function of a cryostat is to cool the germanium detector to LN temperature. The germanium detector preamp is normally included as part of the cryostat package. Since the preamp should be located as close as possible so that the overall capacitance can be minimized, the preamp is installed together. The input stages of the preamp are also cooled. The entire cold assembly is maintained by the cryostat under high vacuum for both thermal insulation and protection of the internal components from contamination. All of the cryostat materials around the detector should be as low Z as possible to reduce photon scatter. Hence, aluminum, magnesium, beryllium, Teflon, and Mylar are used whenever possible.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-10 There are various types of dewars so that the detection geometry can be selected depending on the application. For general radioactive or neutron-activated samples, the vertical type is widely used. For in-beam spectroscopic purposes, either the horizontal or side-looking type is more convenient. Portable dewars are also available for outside field measurements. Vertical Horizontal Side-looking B. Energy resolution Fig. 8.13. Various configurations of LN dewars and cryostats. The most important characteristic of HPGe detectors is their excellent energy resolution. In Fig. 8.14, two spectra from an HPGe and a NaI(Tl) are compared for a 60 Co source. The great superiority of the HPGe detector allows the separation of many closely spaced gamma lines, which is very beneficial for measuring multi-gamma emitting radioactive sources. 5x10 4x10 NaI (75 kev FWHM) HPGe (.35 kev FWHM) 1173 kev Counts/channel 3x10 x10 1x10 133 kev 0 0 500 1000 1500 000 Photon energy [kev] Fig. 8.14. Comparison of NaI(Tl) and HPGe spectra for 60 Co.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-11 The energy resolution of the HPGe photon spectroscopy system is governed by the variation in the number of charge carriers, variation in the charge carrier collection, and the contribution of electronic noise. Hence, the FWHM of a peak in the spectrum can be represented as T D X W = W + W + W E where W values on the right-hand side are peak widths expected due to carrier statistics, carrier collection and electronic noise. The variation in the number of the charge carriers is given by W FδN F E D =.355 =. 355 γ W D = (.355) F = (.355) FwEγ Eγ N N N where E γ is the photon energy, N is the number of e-h pairs, F is the Fano factor, and w is the energy required for creating a electron-hole pair. If the Fano factor is assumed to be 0.1, the equation predicts the W D value of 1.3 kev at 1.33 MeV when the w value of 3 ev is used. The contribution of the second term is due to incomplete charge collection and becomes important for large volume or low electric field detectors. The last term represents the broadening effects of all electronics following the detector. This component can be measured with a precision pulser as discussed in the pulse processing chapter. The pulser output is fed to the test input of the preamp and it undergoes exactly the same path as that of the actual signal from the detector. The figure shows the contribution of each component as a function of the photon energy. At low energies, electronic noise and charge collection are relatively important while the additional broadening due to carrier statistics becomes significant at higher energies. Fig. 8.15. Dependence of resolution on photon energy. C. Pulse shape and timing properties The detailed shape of the signal pulse is important for determining the shaping time of the pulse height spectroscopy and for applications requiring timing information. The

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-1 ultimate time resolution obtainable from HPGe detectors depends on both the average rise time and the variation in the pulse shape from event to event. Two factors limit the ultimate time resolution of HPGe detectors. The first factor is the slow process of the charge carrier collection. If the electric field is strong enough, the drift velocities of electron and hole are saturated. The electron saturation drift velocity is ~ 10 5 m/s, which makes a travel time of 100 ns/cm. Typical pulse rise times will be of this order and are much longer than those of organic scintillation detectors. The second factor is the large variation in the shape of the output pulses. The leading edge of the pulse changes substantially from event to event depending on the position of electronhole creation. The pulse shape of planar detectors can be described from the energy balance as follows. The infinitesimal induced charge is given by dw = V0dQ where, dw is the work done on the charge carriers, V 0 is the applied voltage and dq is the induced charge. The work done on the charge carriers with charge q 0 is dw = q0dv ( x) dw V = q E x q 0 0 ( ) 0 dx T where, E(x) is the electric field and T is the detector thickness. The energy absorbed by the motion from x 0 to x is x q0v x 0 q0v0 W = dw = dx = ( x x0 T T ) x0 x0 which induces charge Q as W x x0 vt Q = dq = = q0 = q0 V0 T T The corresponding signal voltage V R can be represented as Q q0 vt V R = = C C T There are two components of the induced charge: one from the motion of electrons, the other from holes. Let s suppose charge carriers were produced at a distance x from the n + layer after a photon interaction as shown in Fig. 8.16. In the initial stage of the charge carrier collection, both electrons and holes are drifting, thus, ve vh Q( t) = q0 ( t + t) T T h If electrons have been collected, but holes are still drifting, l vh Q( t) = q0 ( e + t) T T In the opposite case (holes have been collected), ve T le Q( t) = q0 ( t + ) T T e x

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-13 After both holes and electrons are collected Q ( t) = q 0 Fig. 8.16 shows typical patterns of pulse shapes depending on the production positions of e-h pairs. Fig. 8.16. Shape of the leading edge of the output pulse Q(t) for a planar HPGe detector. 8.4. Photon spectroscopy with HPGe detectors A. Detector response [6] Fig. 8.17. Pulse height spectrum of 54 Mn single gamma emitter.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-14 The response of a HPGe detector for a single gamma emitter is shown in Fig. 8.17. Interaction events contributing the full-energy peak are more likely to consist of multiple interactions, such as Compton scattering followed by photoelectric absorption of the scattered photon. Absorption of the full photon energy in a single photoelectric effect is pretty rare. The table below summarizes a Monte Carlo simulation result. Number of interactions contributing to the 133 kev peak for a coaxial HPGe detector 1 3 4 5 6 4. % 17.8 % 7.0 % 5.1 % 15.4 % 7.0 % As the photon energy increases above m e c, the pair production interaction becomes dominant and the escape peaks are quite important. The spectra in Fig. 8.18 show different patterns of the full-energy, single escape, double escape peaks depending on the photon energy. Fig. 8.18. The relative intensity of the full-energy, single escape and double escape peaks for a coaxial detector.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-15 Fig. 8.19. Pulse height spectrum of a HPGe detector for 56 Mn multi-gamma emitter. Fig. 8.19 is an example of the response for a multi-gamma emitter. Thanks to the excellent resolution, each peak is well identified, which is pretty important for analyzing a complicated spectrum. B. Compton suppression system Ge Ge Fig. 8.0. Guard detector arrangements for Compton suppression. The Compton continuum in photon spectra from HPGe detectors is generated primarily by photons that undergo one or more scatterings in the detector followed by escape of the scattered photon from the active volume. If those events contributing to the continuum are rejected, the peak-to-continuum ratio is greatly improved and accordingly, the minimum detectable radioactivity can be reduced significantly. To realize a rejection process, the HPGe detector is surrounded with a Guard detector (inorganic scintillation type) as shown in Fig. 8.0. Then the pulse processing system can reject the escape events by the anticoincidence technique when proper timing electronics are implemented.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-16 In other words, the pulse processing system accepts each detection event from the HPGe only when there is no detection event from the guard detector in coincidence. Figs. 8.1 and 8. show a standard pulse processing system for Compton suppression and an example of the suppression effect. Fig. 8.1. Standard pulse processing system for Compton suppression. Fig. 8.. 137 Cs spectrum with and without the Compton suppression. C. Efficiency calibration Conventionally, the relative efficiency is conveniently used for quoting the peak efficiency of a HPGe detector. The relative efficiency is defined as Rel. eff. = HPGe peak effi/nai(tl) (3 3 ) peak effi at 133 kev from 60 Co. The source-detector distance of 5 cm is uniquely used for this definition. The absolute peak efficiency of a 3 3 NaI(Tl) is 1. 10-3 in this geometry.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-17 A typical variation of the full-energy peak efficiencies for coaxial detectors is shown in Fig. 8.3. Due to the relatively thick dead layer, the p-type efficiency starts dropping at ~ 100 kev while the n-type efficiency is constant down to ~ 5 kev. Fig. 8.3. Absolute efficiencies of n-type and p-type coaxial detectors. A practical example of the efficiency calibration for a wide dynamic range is shown in Fig. 8.4. The experimental efficiency data points were obtained from standard calibration sources up to ~ 1.5 MeV and extended up to 1 MeV with both 14 N(n,γ) and 35 Cl(n,γ) prompt gamma-rays. The measured data points are converted into the log E versus the log of the efficiency and then fitted with a polynomial so that a wide energy region is covered. Fig. 8.4. Fitting of a HPGe detector efficiency in a wide energy region.

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-18 8.5. Medical applications A. In vivo X-ray spectrometry [7,8] In our department, planar HPGe and Si(Li) detectors have been used for in vivo X-ray spectrometry to analyze both trace (Pb, Hg, As etc.) and major (Sr) elements in the body quantitatively. The targeted elements are excited either with radiation from a radionuclide or an X-ray generator and then the characteristic X-rays are emitted and detected. Fig. 8.5 shows a system developed for Hg and an example phantom spectrum. Fig. 8.5. In vivo X-ray spectrometry system for measuring mercury and a phantom spectrum [8]. B. Collimator-less Compton imaging (a) (b) Fig. 8.6. Concept of collimator-less Compton imaging (a) and three projected cones intersecting at one point (b) [9,10]. Segmented (or position sensitive) semiconductor detector technology has been developing rapidly. For medical applications, this class of semiconductor detector makes possible collimator-less gamma-ray imaging. In principle, Compton imaging is effective even with a single segmented detector if the scattering process inside the detector can be tracked (gamma-ray tracking [11]) with sufficient spatial resolution. With a highly

Med Phys 4R06/6R03 Radioisotopes and Radiation Methodology 8-19 segmented structure of Ge or Si, accurate information on both interaction position and deposited energy can be analyzed. From this information, the interaction sequence of the photon inside the detector can be extracted with a proper tracking algorithm using the Compton scattering formula studied in the chapter 5. Then, the direction cone of the incident photon is obtained. Since the segmentation of the electrode is physically limited, the pulse shape analysis technique is employed to further improve the spatial resolution. Finally, the source position is constructed as shown in Fig. 8.6. Both planar and coaxial configurations have been developed. Fig. 8.7. Two-dimensionally segmented closed-ended coaxial (40-fold) and planar detectors ( 39-fold). References 1. G.F. Knoll, Radiation Detection and Measurement - 3 rd edition (Chapters 1 and 13), John Wiley & Sons, 1999.. G. Gilmore, J.D. Hemingway, Practical Gamma-Ray Spectrometry, John Wiley & Sons, 1995. 3. K. Debertin, R.G. Helmer, Gamma- and X-ray Spectrometry with Semiconductor Detectors, North-Holland, 1988. 4. HPGe Detector Manufacturing, ORTEC, URL: http://www.ortec-online.com/. 5. Th. Kröll et al., Nucl. Intstr. and Meth. A 371 (1996) 489. 6. Online Ge(Li)-Si(Li) Gamma Spectrum Catalog, Gamma-ray Spectrometry Center, Idaho National Laboratory, Published 3-9-1999, URL: http://www.inl.gov/gammaray/catalogs/. 7. D.R. Chettle, X-Ray Spectrom. 34 (005) 446. 8. J. Grinyer, M. Popovic, D.R. Chettle, X-Ray Spectrom. 36 (007) 99. 9. L. Mihailescu et al., Nucl. Intstr. and Meth. A 570 (007) 89. 10. K. Vetter, M. Burks, L. Mihailescu, Nucl. Intstr. and Meth. A 55 (004) 3. 11. I.Y. Lee, Nucl. Intstr. and Meth. A 4 (1999) 195.