Basic hands-on gamma calibration for low activity environmental levels

Similar documents
Gamma-Spectrum Generator

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

Determination of the activity of radionuclides

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Calibration of A Small Anode Germanium Well Detector

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Indian J.Sci.Res.2(3) :25-32, 2011

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L.

Compton suppression spectrometry

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

Gamma and X-Ray Detection

arxiv:nucl-ex/ v2 21 Jul 2005

*Corresponding author,

K 40 activity and Detector Efficiency

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Simulation of 4π HPGe Compton-Suppression spectrometer

Characterization and Monte Carlo simulations for a CLYC detector

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

First tests of the big volume ultra low background gamma spectrometer

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

BUREAU INTERNATIONAL DES POIDS ET MESURES

The Best Gamma-Ray Detector

Radiometric Measurements of Environmental Radioactivity

Study well-shaped germanium detectors for lowbackground

Characterization of semiconductor detectors for γ-ray and x-ray spectrometry

NGN PhD Studentship Proposal

K 40 activity and Detector Efficiency

Gamma Spectroscopy. References: Objectives:

The IAEA-CU world wide open proficiency test on the determination of radionuclides in soil, spinach and water

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

Compton Camera. Compton Camera

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

IDENTIFICATION AND QUANTIFICATION OF RADIONUCLIDES IN HISTORICAL WASTE AT ANSTO

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

GAMMA RAY SPECTROSCOPY

RMCC CANBERRA Latest developments

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

THALES Project No. 65/1205

Determination of the shielding power of different materials against gamma radiation

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Ultra-Low Background Counting and Assay Studies At SNOLAB

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

PRACTICAL GAMMA SPECTROSCOPY ASSAY TECHNIQUES FOR LARGE VOLUME LOW-LEVEL WASTE BOXES

PHYS 3650L - Modern Physics Laboratory

The Ones That Got Away: Neutron Leakage Spectra

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Attenuation of Radiation in Matter. Attenuation of gamma particles

DEVELOPMENT OF ADVANCED NEUTRON INDUCED PROMPT GAMMA- RAY ANALYSIS SYSTEM FOR SURVEY OF ANTI-PERSONNEL MINES

Gamma background measurements in the Boulby Underground Laboratory

The Determination of Radioactivity Levels in the Environment using High Resolution Gamma-ray Spectrometry

Coincidence gamma-ray spectrometry

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Radioactivity. Lecture 6 Detectors and Instrumentation

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

Quality Assurance. Purity control. Polycrystalline Ingots

ANGLE 4. Advanced Gamma Spectroscopy Efficiency Calibration Software

A coincidence method of thorium measurement

Slides by: Prof. Abeer Alharbi

Radiometric Measurements of Environmental Radioactivity Beta Counting, Alpha and Gamma Spectrometry

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

The metrology of radioactivity. C. Michotte, BIPM

THE COMPTON EFFECT Last Revised: January 5, 2007

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Fundamentals of Radionuclide Metrology

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

IPNDV Working Group 3: Technical Challenges and Solutions Nuclear Material (3) Technology Data Sheet

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

ISO INTERNATIONAL STANDARD. Measurement of radioactivity in the environment Soil Part 3: Measurement of gamma-emitting radionuclides

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Gamma ray coincidence and angular correlation

The NEXT-100 experiment for Neutrino-less Double Beta decay: Main features, Results from Prototypes and Radiopurity issues.

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

High-energy calibration data for neutron activation analysis

Chem 481 Lecture Material 3/20/09

Radionuclide Imaging MII Detection of Nuclear Emission

F. Cappella INFN - LNGS

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

arxiv:physics/ v1 3 Aug 2006

Low Background Counting At SNOLAB

NUCL 3000/5030 Laboratory 2 Fall 2013

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Practical Gamma-ray Spectrometry

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

Sample Spectroscopy System Hardware

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS

Practical Solutions to Radioactive Waste Characterization

Gamma Ray Spectroscopy

Transcription:

Basic hands-on gamma calibration for low activity environmental levels Iolanda Osvath Presented by Mats Eriksson Environment Laboratories Marine Environment Laboratories, Monaco Radiometrics Laboratory International Atomic Energy Agency

Presentation outline Detector characteristics Commonly used types of detectors Calibration of detectors Energy Resolution Efficiency Efficiency calibration of HPGe detectors Experimental calibration Reference data used for calibration Commonly used calibration standards and approaches Preparation of volume standards Calibration curves Commonly encountered problems Geometry, self-attenuation and coincidence summing corrections Combined experimental and numerical calibration Demonstration of GammaVision calibration routine

Basic characteristics of gamma-ray spectrometric systems Environmental gamma-ray spectrometry nowadays relies on low-background high resolution HPGe detectors, with performance characterized by - efficiency - resolution - peak/compton ratio Important characteristic: the system s background, which can be decreased using - low activity detector components - passive shield (lead + copper, tin, plexiglass) - active shield (anti-cosmic, anti-compton)

but more information comes with a detector on delivery Specifications Detector model & s/n Cryostat model PA model Nominal volume Resolution @ 122 & 1332 kev Peak/Compton ratio Endcap dimensions Physical characteristics of crystal Electrical characteristics Resolution and relative efficiency

and even more is available upon request from the supplier* The data provided by the supplier are necessary, but not sufficient, for modeling the detector. Ge crystal details Detector chamber details (materials and dimensions) Exact plans are confidential, certain parameters are even not known exactly to the producer, but by combining experiment and modeling, a set of effective detector parameters can be defined. *ask for it when you buy the detector, otherwise you may have to pay for it

and finally the off-the-shelf solution Have your detector characterized by the producer (for an additional cost, but may finally save money) Laboratory Sourceless Calibration Software (LabSOCS CANBERRA) Eliminates the cost of purchasing, tracking, and disposing of radioactive standards No radioactive sources needed for accurate efficiency calibrations in the laboratory Calibrations valid from 10 kev to 7000 kev Not yet developed for well detectors!

Commonly used types of Ge detectors (CANBERRA)

Common types of semiconductor detectors used for environmental applications (CANBERRA)

How to chose your detector? Most environmental gamma spectrometrists aim primarily to minimize the MDA (minimum detectable activity) Detection efficiency Background Resolution Other issues to consider: Energy range of interest Efficiency vs. energy Typical sample size Some common-sense tips: Large ultra-low background well detectors will be good for low singleline radionuclides (e.g. Cs-137), but require the user to master coincidence corrections for other applications Broad energy Ge detectors are excellent for normal environmental levels Important: choice of adequate shielding!!!

Why do we need calibration? To interpret a spectrum in terms of - energy (identify radionuclides) energy calibration - activity (quantify radionuclides) efficiency calibration To carry out automated spectrum analysis peak width calibration Sediment spectrum, 200ks 200% well, Pb+Cu+anti-cosmics shielding (GammaVision)

Energy calibration Energy = A + B * CH + C * CH 2 Energy = A + B * CH sufficient in general with nowadays electronics for environmental applications

Peak width calibration FWHM = a * E 1/2 + b * E + c FWHM = 2.35 * σ (CANBERRA) Other functions may apply depending on the energy range and taking into account the peak shape

Efficiency calibration: which efficiency is used in common applications? Relative efficiency a general performance measure relating the efficiency of detection of the 1332 kev Co-60 gamma ray for a point source placed centrally 25cm above the detector to that of the standard 3 x 3 sodium iodide scintillation detector Absolute full energy peak efficiency (FEPE) the ratio of the number of events when the complete energy E was deposited in the detector to the number of photons of energy E emitted from the source Absolute total efficiency relates the number of gamma-rays emitted by the source to the number of counts detected anywhere in the spectrum. This takes into account the fullenergy peak and all incomplete absorptions represented by the Compton continuum Intrinsic efficiency (full energy peak or total) relates the counts in the spectrum to the number of gamma-rays incident on the detector. This efficiency is a basic parameter of the detector

Efficiency calibration The detection efficiency depends on the energy of the incident gamma ray the detector crystal (intrinsic) the materials surrounding the crystal the source/sample - detector geometry attenuation in the source matrix Can be determined experimentally measuring calibration standards with known activity.

Reference decay data (NUCLEONICA)

Efficiency calibration: standardized solutions www.nuclitec.de

Efficiency calibration: standardized solutions Attention to the uncertainties specified by the supplier! These will enter into the uncertainty budget of the results. Choose solutions with <2% uncertainty, slightly higher (3-4%) acceptable for some radionuclides (e.g. Pb- 210) Not all solutions can be mixed! Because of the different chemical forms precipitates can form and you will have an inhomogeneous secondary standard! E.g. NG4 (Pb-210) should not be mixed into any of the other solutions. Do not use Reference Materials for calibration. Their use is limited to internal quality control and proficiency tests. www.nuclitec.de

Geometry reference sources commercially made to order (QSA www.nuclitec.de) (QSA)

Geometry reference sources commercially made to order (QSA)

Efficiency calibration: experimental A ( ) i n K i Y( E) t E 1 ε(e) FEPE for a gamma ray of energy E Ai n net area under the full energy peak of radionuclide i Λ i reference activity of radionuclide emitting gammaray of energy E Y(E) emission probability of radionuclide at energy E t spectrum collection time (live time) K 1 decay-correction factor reference date counting date

Efficiency calibration (FEPE) Various functions and combinations of functions can be used to describe the energy dependence of efficiency. Example ln ( E) 2 a0 a1 ln ( E) a2 ln ( E) where ε - efficiency, E - energy

Efficiency calibration: experimental Ideally: calibration standard identical in all aspects (geometry, matrix, radionuclide composition) to sample. In this case a direct calibration can be made for the radionuclides of interest. Many different radionuclides can be present in the sample, therefore efficiencies may be required at energies other than those of the gamma rays emitted by the calibration standard. Calibration curves have then to be determined based on the experimental efficiencies calculated for the gamma-rays emitted by the calibration standard. All commercial software include several options for the type of fit used to describe the efficiency-energy dependency (linear, quadratic, polynomial). You can also use your own routines to fit the experimentally obtained efficiencies.

Commonly encountered problems Sample and calibration standards not identical (matrix, density) The numerous counting geometries and matrices usually encountered in an environmental lab require many calibration standards: costly procurement, storage and disposal Radionuclides with coincident gamma-ray emissions in standard and sample

Combined experimental and numerical approach Corrections Geometry: different geometry sample standard Self-attenuation: important for low energy gamma rays when the sample and the standard have different composition (matrix effect) or density (differently compacted) Coincidence summing: important for radionuclide with cascade lines in close counting geometries, particularly in well detectors

efficiency Geometry: to be respected! 8.E-01 7.E-01 6.E-01 5.E-01 4.E-01 3.E-01 2.E-01 h (cm) 0.1 0.5 1 1.5 2 2.3 3 4.7 1.E-01 0.E+00 28 60 128 275 591 1271 2729 E (kev) Well detector, 150% relative efficiency Geometry: cylindrical tube r = 1cm, h = 0.1 4.7cm Water (Monte-Carlo calculations, GESPECOR)

Self-attenuation: important at low energies 1.00E+00 1.00E-01 phosphogypsum water 1.00E-02 1 10 100 1000 10000 Well-type detector, 200% relative efficiency Geometry: cylindrical tube r = 1cm, h = 4.5cm Phosphogypsum density 1.26g cm -3 (Monte-Carlo calculations, GESPECOR)

Coincidence-summing corrections: can be very important

Uncertainties The uncertainty associated with calibration contributes to the final uncertainty budget of the analysis result for the sample Uncertainty from calibration Standardized solution Preparation of calibration standard Counting of calibration standard Sample treatment and preparation for counting Sample counting (ISO 11929) GUM: Guide to the Expression of Uncertainty in Measurement

How to obtain a calibration curve? Demo: GammaVision