NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT

Similar documents
The activities of micro-force measurement below 10 mn in Center for Measurement Standards

Chapter 8. Micro-Metrology

The BIPM Watt Balance

Vacuum Kelvin Force Probe Research Richard Williams August 1st 2008

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2011 C. Nguyen PROBLEM SET #7. Table 1: Gyroscope Modeling Parameters

Dimension measurement. By Mr.Vuttichai Sittiarttakorn

CHAPTER 5 FIXED GUIDED BEAM ANALYSIS

Review of Low Force Transfer Artefact Technologies

Variable Capacitance Accelerometers: Design and Applications

Nanomechanics Measurements and Standards at NIST

Design and Simulation of Comb Drive Capacitive Accelerometer by Using MEMS Intellisuite Design Tool

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2009 PROBLEM SET #7. Due (at 7 p.m.): Thursday, Dec. 10, 2009, in the EE C245 HW box in 240 Cory.

Experimental Validation of Damping Model for a MEMS Bistable Electrostatic Energy Harvester

DESIGN AND FABRICATION OF THE MICRO- ACCELEROMETER USING PIEZOELECTRIC THIN FILMS

ELECTROMAGNETIC OSCILLATIONS AND ALTERNATING CURRENT

CORM Exploring optical radiation pressure as a convenient measure of high-power. laser emission

WATT BALANCES AND THE FUTURE OF THE KILOGRAM

Coulomb s Law and Coulomb s Constant

GENERAL CONTACT AND HYSTERESIS ANALYSIS OF MULTI-DIELECTRIC MEMS DEVICES UNDER THERMAL AND ELECTROSTATIC ACTUATION

CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS

EE C247B / ME C218 INTRODUCTION TO MEMS DESIGN SPRING 2014 PROBLEM SET #1

Catapult tests for microgravity characterization of the MICROSCOPE accelerometers. Manuel Rodrigues On behalf ONERA & ZARM team

Low Frequency Electrical Metrology Programs at NIST

Piezoelectric Resonators ME 2082

Good Luck! Mlanie LaRoche-Boisvert - Electromagnetism Electromagnetism and Optics - Winter PH. Electromagnetism and Optics - Winter PH

Large Scale Computation of Coupled. Electro-Acoustic Systems using ANSYS and CAPA

DESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE

EE C247B / ME C218 INTRODUCTION TO MEMS DESIGN SPRING 2014 C. Nguyen PROBLEM SET #4

Conventional Paper-I-2011 PART-A

Pre-lab Quiz/PHYS 224 Coulomb s Law and Coulomb Constant. Your name Lab section

SIM Regional Comparison on. The Calibration of Internal and External Diameter Standards SIM.L-K FINAL REPORT July 2012

The secondary winding have equal no. of turns. The secondary windings are placed identically on either side of the primary winding.

Louisiana State University Physics 2102, Exam 2, March 5th, 2009.

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

DESIGN, FABRICATION AND ELECTROMECHANICAL CHARACTERISTICS OF A MEMS BASED MICROMIRROR

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Ch. 23 Electromagnetic Induction, AC Circuits, And Electrical Technologies

2014 F 2014 AI. 1. Why must electrostatic field at the surface of a charged conductor be normal to the surface at every point? Give reason.

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Tunable MEMS Capacitor for RF Applications

Lecture 4 Scanning Probe Microscopy (SPM)

EE C245 - ME C218 Introduction to MEMS Design Fall Today s Lecture

Last Revision: August,

and another with a peak frequency ω 2

Technology Readiness Level:

Microstructure cantilever beam for current measurement

Constrained Rectilinear Motion Dynamics Laboratory, Spring 2008

EE C245 ME C218 Introduction to MEMS Design

Mathematical Modeling and Dynamic Simulation of a Class of Drive Systems with Permanent Magnet Synchronous Motors

MEMS INERTIAL POWER GENERATORS FOR BIOMEDICAL APPLICATIONS

Improving the accuracy of Atomic Force Microscope based nanomechanical measurements. Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA

THE WATT BALANCE ROUTE TOWARDS A NEW DEFINITION OF THE KILOGRAM. A. Eichenberger, H. Baumann, and B. Jeckelmann.

Pick-up Calibration in CESR Beam Position Monitors

Design of a MEMS Capacitive Comb-drive Accelerometer

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

LASER TRAPPING MICRO-PROBE FOR NANO-CMM

Physics Project. Name:- Dabhi Mihir Roll No:- 3 Class:- 12 th (Science-PCB) Topic:- Moving coil galvanometer

Electrostatic Microgenerators

Lesson 3. Electric Potential. Capacitors Current Electricity

Technical Note Development of a New, High Sensitivity 2000 kg Mechanical Balance

Micro- and Nanoforce Metrology at PTB

Silicon Capacitive Accelerometers. Ulf Meriheinä M.Sc. (Eng.) Business Development Manager VTI TECHNOLOGIES

MEMS Tuning-Fork Gyroscope Mid-Term Report Amanda Bristow Travis Barton Stephen Nary

DEVELOPMENT OF A MEMS REPULSIVE ACTUATOR FOR LARGE OUT-OF-PLANE FORCE

Ion Optics Simulations

Chapter 4. Electrostatic Fields in Matter

E05 Resonator Design

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

RECENT ADVANCES IN SIMULATION OF MAGNETRONS AND CROSSED-FIELD AMPLIFIERS. Abstract

Thickness Optimization of a Piezoelectric Converter for Energy Harvesting

A Vertical Electrostatic Actuator with Extended Digital Range via Tailored Topology

2 Each satellite will have two test masses, each being the end mirror for an interferometer.

THE NEW WATT BALANCE EXPERIMENT AT METAS

PH213 Chapter 24 Solutions

EE C247B ME C218 Introduction to MEMS Design Spring 2016

Experiment 4 Oscillations

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

Design of Electrostatic Actuators for MOEMS Applications

The Pull-In of Symmetrically and Asymmetrically Driven Microstructures and the Use in DC Voltage References

APPLICATIONS OF VIBRATION TRANSDUCERS

The Kilogram Redefined:

Fabrication and performance of d 33 -mode lead-zirconate-titanate (PZT) MEMS accelerometers

2. Determine the surface tension of water with the capillary-rise method.

INTERNAL THALES ALENIA SPACE

Evaluation of a surface acoustic wave motor with a multi-contact-point slider

Nanometrology and its role in the development of nanotechnology

Physics 2B Winter 2012 Final Exam Practice

Hard and Soft X-ray Mirror System Design Status

Automated volume measurement for weihts using acoustic volumeter

Structural Dynamics. Spring mass system. The spring force is given by and F(t) is the driving force. Start by applying Newton s second law (F=ma).

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

EE C245 ME C218 Introduction to MEMS Design

Stepping Motors. Chapter 11 L E L F L D

EE 5344 Introduction to MEMS CHAPTER 6 Mechanical Sensors. 1. Position Displacement x, θ 2. Velocity, speed Kinematic

DESIGN AND EXPERIMENTAL EVALUATION OF AN ELECTROSTATIC MICRO-GRIPPING SYSTEM

A Stacked-type Electrostatic Actuator and Measurement of its Energy Efficiency

Development of Measuring System for the Non-Repetitive Run-Out(NRRO) of Ball Bearing

DEVELOPMENT OF MEASUREMENT STANDARD FOR DYNAMIC PRESSURE AT MIKES

INF5490 RF MEMS. LN03: Modeling, design and analysis. Spring 2008, Oddvar Søråsen Department of Informatics, UoO

Transcription:

NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT John A. Kramar, David B. Newell, and Jon R. Pratt National Institute of Standards and Technology, Gaithersburg, MD, USA We have designed and built a prototype electrostatic force balance for realizing forces in the micronewton range. The active electrodes are concentric cylinders, the outer serving as the reference and the inner suspended and guided by a rectilinear flexure mechanism. The geometry has been designed such that a near-linear capacitance gradient of 1 pf/mm is achieved at a working overlap of 5 mm. We have used this balance in a null-displacement mode to compare an electric realization of force with the force generated by calibrated deadweights of nominal mass 1 mg, 10 mg, and 20 mg. The preliminary measurements reported here agree to within a few parts in 10 4 after including all known correction factors. Introduction The trend toward cost-saving and performance enhancement through miniaturization and scale-down in many areas of manufacturing, but particularly in microelectronics, data storage, and micro-electromechanical systems (MEMS), is well known. With this trend comes the increasing need for calibrated force measurements in the micronewton range and below. Many commercial instruments including nanoindenters and atomic force microscopes have force resolutions that extend into the nanonewton regime, and these levels of measurement are becoming increasingly necessary for the control of manufacturing processes. However, no methods for establishing force measurement traceability at these levels are currently available. Our primary goal is to realize forces in the micronewton to piconewton range. Conceptually, the most straightforward approach would be to use a calibrated mass as a deadweight. However, it is not possible to maintain high levels of precision with this approach as the masses are subdivided. The smallest available calibrated mass is 1 mg (10 µn) having a relative uncertainty at the level of a few parts in 10 4. In principle smaller masses could be calibrated, but they would be difficult to handle, and the achievable relative uncertainty would increase inversely proportionately with the decrease in mass. 1 Another practical means for realizing forces in this range is through the electrical units defined by the International System of Units (SI). This can be done using electromagnetic forces (e.g., the NIST Watt Balance Experiment 2 ) or using electrostatic forces. 3 We have chosen the latter because it is somewhat simpler to execute the required metrology, and the forces generated, although generally less than those feasible electromagnetically, are adequate for the force range of interest. As a validation of our electrostatic force realization, we desire to crosscheck with deadweight forces, at least in our higher force range where the uncertainty achievable mechanically is still competitive. For this reason we have designed our force generator to operate along the vertical axis. Eventually, the plan is also to design transfer artifacts, i.e., calibrated load cells or force generators, through which we can disseminate this realized force to users in industry or academia.

Experimental Design We have built a prototype electrostatic force balance that realizes force through the SI electrical units and has the capability of comparing this force with the deadweight force from SI traceable masses. 4 For this first prototype, the design features a spring suspended platen whose deflection can be measured by an interferometer (see Fig. 1). The spring is a beryllium-copper flexure assembly that constrains the platen to rectilinear vertical motion with a spring constant of 13 N/m. The cross-axis stiffness has been measured to be at least 20 times larger. The spring can be deflected either by adding deadweight or by applying an electrostatic force. For comparison between the two, we typically do a null motion experiment: The spring is first held in a deflected position by the application of an electrostatic force. Then a mass is added and the electrostatic force is decreased such that the final platen position is unchanged as measured by the interferometer. The difference in electrostatic forces is then taken to be equal to the deadweight force. The test masses used for this experiment were calibrated against the NIST working standards. The electrostatic force is generated by applying a voltage across a concentric-cylinder capacitor. The inner cylinder is suspended from the flexure spring. The mechanical work required to change the separation between two electrodes of a capacitor while maintaining constant voltage is dw = F dz = ½ V 2 dc (1) where dw is the change in energy, F is the force, dz is the change in separation, V is the electric potential across the capacitor, and dc is the change in capacitance. Thus force can be realized from electrical units by measuring V and the capacitance gradient, dc/dz: F = ½ (dc/dz) V 2. (2) In our design, the outer diameter of the inner cylinder is 15 mm and the inner diameter of the outer cylinder is 15.8 mm. With this geometry dc/dz is about 1 pf per millimeter of cylinder overlap. We operate at an overlap of 5 mm. spring suspension anchor points glass rod Z a 1 a 2 wire deadweight lifting fork spring suspension inner electrode guard outer electrode reference beams measurement beams interferometer Fig. 1. Schematic block diagram of the NIST electrostatic force balance design.

The operating range of the balance is bracketed by the maximum electrostatic force and the noise floor. The maximum force is limited by the dielectric strength of air to roughly 400 µn. The noise floor is the precision with which null can be held. This is a combination of the noise on the interferometer, the performance of the servo controller, and the vertical flexure stiffness. The interferometer noise and controller performance depend among other factors on the amount of averaging. For our experimental conditions, we measured the force noise floor to be less than 10 nn. Procedure The basic weighing procedure is to alternate and bracket weighing sets with measurements of dc/dz. This is done under automated software control over periods of time extending to days. Measurements of dc/dz are typically done by moving the stage that holds the base of the flexure spring over a range of 400 µm, centered about the weighing null point. The stage is scanned up and down ten times while recording the capacitance and the position at 40 µm intervals. Each up and down cycle is averaged to minimize drift effects (e.g., thermal), and fit by a line to give the slope. These individual gradients are in turn averaged and their standard deviation (type A uncertainty 5 ) is calculated to provide an estimate of dc/dz at the weighing null point. For each weighing set, typically 40 mass on/off transitions are performed over a period of 45 min. To minimize any drift effects, the individual weighings are considered to be the difference between the mass on (off) electrostatic force and the mean of the two neighboring mass off (on) electrostatic forces. A rather slow loop time of 125 ms is chosen for the null servo controller to allow time for averaging the interferometer readings and the voltage readings to minimize noise. Furthermore, each mass on (off) voltage is the average of 800 of these servo iterations (100 s). A ring magnet surrounds the inner cylinder to provide eddy current damping of its 12 Hz natural resonance, damping it to a Q of 4.2. Force Comparison and Correction Factors We compared the deadweight force of the test masses with the electrostatic force. We chose the flexure spring axis along which the inner cylinder motion is constrained as a convenient frame of reference for these force comparisons. The force due to the mass along this axis is given by F mf = m g cosθ gf B (3) where m is the mass of the test mass, g is the acceleration of gravity, θ gf is the angle of the flexure axis with respect to gravity, and B = (1 ρ air /ρ mass ) is the buoyancy correction, since the electrostatic force balance operates in air. The uncorrected electrostatic force that balances F mf is given by F efu = ½ (dc/dz f ) (V off 2 V on 2 ) (4) where dc/dz f is the capacitance gradient along the flexure-constrained motion axis and V on and V off is the electrostatic potential between the capacitor electrodes with the mass on and off, respectively. In addition, there are small correction forces due to null-point shift and hysteresis of the flexure spring. These corrections are generally proportional to the applied force. The reference mirror for the interferometer that measures the null position is on the outer

electrode. When the test mass is removed from the platen and the electrostatic force pulling the inner electrode down is increased to hold null position, the outer electrode is also pulled up by the same force. This causes a small shift in the null position equal to F efu /k o, where k o is the stiffness of the outer cylinder with respect to this force. This results in less deflection of the inner cylinder and consequently a decrease in the electrostatic force necessary to hold null by an amount equal to F efu (k f /k o ), where k f is the flexure spring compliance. The other correction force we consider is due to spring hysteresis. As the mass is placed on or taken off the platen, there is a transient deflection of the inner-electrode flexure spring due to the slow response of the servo controller. The resulting spring-hysteresis force is an additional force that must be overcome by the electrostatic force. It is dependent on the magnitude and duration of the spring deflection and so is roughly proportional to F efu by an experimentally determined factor h. Taken together, we arrive at a corrected electrically derived force measurement of F ef = F efu [1 + (k f /k o ) h]. (5) Since the capacitance gradient, dc/dz f, has a dependence on the centering of the electrodes and other factors including the dielectric constant of air, it is best determined in situ, bracketing the sets of mass exchanges. This is done by scanning in z the stage that holds the anchor points of the inner cylinder flexures, as previously discussed. This stage axis, z s, has previously been aligned to the flexure axis to within a few milliradians. During the scanning for the determination of dc/dz s, the z motion is measured by the interferometer, which operates along axis z i. Consequently dc/dz s is overestimated by (1/cosθ si ) if the interferometer reading is taken as the measure of z s. We have also done separate experiments to compare dc/dz f with dc/dz s. For this, we alternate the usual dc/dz s measurements with measurements of dc/dz f, which we obtained by deflecting the flexure axis 160 µm using a 220 mg mass. From this we determine that the capacitance gradients differ by a factor of δ sf, typically about 2.5 x 10-4. Hence, dc/dz f = (dc/dz s ) i (1 + δ sf ) cosθ si (6) where (dc/dz s ) I is dc/dz s but with z s measured along z i. Combining factors and taking the small angle and small error approximations, we arrive at a mechanically derived force F mf = m g [1 (θ gf 2 /2) ρ air /ρ mass ] (7) to be compared with an electrically derived force F ef = ½ (dc/dz s ) i (V off 2 V on 2 ) [1 + (k f /k o ) h + δ sf (θ si 2 /2) ]. (8) Typical numerical values for the correction factors are listed in Table 1. Results We have weighed three different test masses multiple times using the electrostatic force balance. The test masses are bent wires with nominal masses 1 mg, 10 mg, and 20 mg, made of Pt, Au and Au, respectively. The uncorrected electrostatic force data, F efu, from a typical data run are shown in Fig. 2. For each set of weighings, the average dc/dz from the two bracketing dc/dz measurements is used. The value, 103 275.8 nn ± 3.3 nn, is the mean and the standard deviation of the individual weighing set averages. So at this force, our fractional type A uncertainty contribution is roughly 30 x 10-6.

Table 1: Typical correction factors and standard uncertainties, u c. Table 2: Typical standard uncertainties, u c, for standard transfers. Correction Factor Value (x10-6 ) u c (x10-6 ) Standard Transfer, Q u c = dq/q (x10-6 ) Uncertainty Type θ gf 2 /2 4 4 g 1 ρ air /ρ mass 59 5 k f /k o 30 5 C 10 V 6 B h 70 70 z 1 δ sf 250 250 1 mg 380 θ si 2 /2 6 4 m 10 mg 38 20 mg 19 A and B The force comparisons with their corresponding uncertainties are listed in Table 3. The data include all the known corrections and uncertainties from Tables 1 and 2. For the mechanical force, the uncertainty is dominated by the uncertainty in the mass, since the uncertainties in the correction factors only contribute a few parts in 10 6. For the electrically derived force, the uncertainties in the correction factors dominate all other sources including the type A contribution from the weighings. All uncertainties are reported as standard uncertainties (coverage factor, k = 1). We define the fractional difference as (F ef F mf )/F mf. We find that in all cases, the electrically derived force is less than the mechanical force by a few parts in 10 4. For the 10 mg and 20 mg nominal masses, the fractional difference is about 6 x 10-4 whereas the combined uncertainty is a factor of two smaller at 3 x 10-4. This points to a yet undiscovered systematic error at this level. One possibility we are still exploring, for example, is that dc/dz could be a function of electrostatic potential. Electrostatic Force (nn) -103 220-103 240-103 260-103 280-103 300 Mean Force = -103 275.81 nn ± 3.3 nn Number of Weighings = 7 x 38-103 320 0 2 4 6 8 10 12 14 Time (hour) Fig. 2. A typical data run showing the individual uncorrected electrostatic force readings and the mean and uncertainty for each set of weighings.

Table 3: Force comparisons Nominal mass (mg) Mechanical Force (µn) 1 9.1421(35) 10 103.3057(38) Electrical Force (µn) Number of Weighing Sets Fractional Difference (x 10-5 ) Fractional Uncertainty (x 10-5 ) 9.1395(37) 13 28 56 9.1385(33) 4 39 52 103.235(30) 12 68 30 103.248(24) 7 56 24 20 205.0025(39) 204.898(60) 10 51 29 Conclusion We have successfully built an electrostatic force balance that operates in the micronewton range, and have used it to compare electrically and mechanically derived forces. We have demonstrated an agreement between the two at the level of a few parts in 10 4. Work is in progress to discover the source of the remaining systematic differences. At the same time, we are designing a second prototype balance 6 that will increase our sensitivity, primarily by lowering the vertical spring constant by a factor of 800. It is based on an equal-arm balance design. The smaller spring stiffness will also allow us to measure dc/dz f more directly by driving the guided inner-cylinder motion using electrostatic forces. This will eliminate the largest correction factor, δ sf, thereby also eliminating the largest source of uncertainty. References 1 Jabbour, Z L., and Yaniv, S. L., The Kilogram and the measurements of mass and force, Journal of Research of the National Institute of Standards and Technology, 106, 2001, pp. 25-46. 2 Williams, E. R., Steiner, R. L., Newell, D. B., and Olsen, P. T., Accurate measurement of the Planck constant, Phys. Rev. Lett., 81, 1998, pp. 2404-2407. 3 Funck, T. and Sienknecht, V, Determination of the Volt with the Improved PTB Voltage Balance, IEEE Trans. on Inst. and Meas., 40(2), 1991, pp. 158-161. 4 Newell, D. B., Pratt, J. R., Kramar, J. A., Smith, D. T., Feeney, L. A., and Williams, E. R., SI traceability of force at the nanonewton level, 2001 NCSL International Workshop and Symposium Proceedings, Washington D.C., July 29 August 2, 2001. 5 Taylor, B. N. and Kuyatt, C. E., Guidelines for evaluating and expressing the uncertainty of NIST measurement results, NIST Technical Note 1297, 1994 6 Pratt, J. R., Newell D. B., and Kramar, J. A., A flexure balance with adjustable restoring torque for nanonewton force measurement, (these Proceedings), 2002.