APPLIED. Measuring Proof and Color in Whiskey. Application Summary. Analytes: ethanol concentration, product color. Detector: OMA-300 Process Analyzer

Similar documents
S reacts with the catalyst s metallic surface to substitute sulfur atoms for oxygen atoms.

Monitoring Nitrogen Trichloride in Chlorine Manufacture. solvent. O NaOH + 1 /2H 2

S is referred to as sour. The sulfide loading lends an unpleasant odor and makes the water corrosive. Additionally, the presence of H 2

A Powerful New Diagnostic Tool for Transit Time Ultrasonic Meters page 1. A Powerful New Diagnostic Tool for Transit Time Ultrasonic Meters

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Fast, Effective XPS Point Analysis of Metal Components

DG Filters 2500 Series Refilling Instructions

Instrumental Limitations for High Absorbance Measurements in Noise

AN2970 Application note

XR Analog Clock - Manual Setting Model Troubleshooting Guide

Human C-reactive protein ELISA using Sword Peroxidase Reagents

December 2010 No 036. The electronic Eppendorf Xplorer pipette Versatile adjustment

Chromatography & instrumentation in Organic Chemistry

for XPS surface analysis

ncounter PlexSet Data Analysis Guidelines

J.T.Baker Brand HYDRA-POINT KARL FISCHER REAGENTS. Europe Middle East Africa

Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

NOTE: Performing this procedure invalidates the factory calibration!!

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

STTH60R04. Ultrafast recovery diode. Main product characteristics. Features and benefits. Description. Order codes

AN2044 APPLICATION NOTE

Open the Beer s Law Lab simulation on your laptop or tablet:

How to use GPC/SEC for compositional analysis

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Applications Note 202 October 2008

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

STTH61R04TV. Ultrafast recovery diode. Main product characteristics. Features and benefits. Order codes. Description A1 K2 ISOTOP ISOTOP STTH61R04TV1

STTH6006W. Turbo 2 ultrafast - high voltage rectifier. Features and benefits. Description. Main product characteristics

STTH6004W. Ultrafast high voltage rectifier. Table 1: Main product characteristics I F(AV) 60 A V RRM. 400 V T j (max) 175 C V F (typ)

Positive and Nondestructive Identification of Acrylic-Based Coatings

1 Measurement Uncertainties

Dräger simultaneous test sets Dräger-Tubes

Rosemount 3051S MultiVariable Configuration Data Sheet

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

LD29300xx. 3 A, very low drop voltage regulators. Features. Description

1 Complementary Access Tools

Obsolete Product(s) - Obsolete Product(s)

AN922 Application note

Low-Frequency Raman Spectra of Carbon Nanotubes Measured with an Astigmatism-Free Schmidt-Czerny-Turner Spectrograph

of mass spectrometry

Route and retain the clinometer cable so that it does not interfere with moving parts where the unit is mounted.

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

HYPHENATED TECHNOLOGY GUIDE

AMBERLITE HPR1300 H Ion Exchange Resin

Linear Systems and Matrices. Copyright Cengage Learning. All rights reserved.

Spectrophotometric Determination of an Equilibrium Constant 1

L78LxxAB L78LxxAC - L78LxxC

KNOWLEDGE

Model TS318-11C55 Thermopile Sensor

AN-1301 APPLICATION NOTE

Energy Chemicals Product Guide

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

STTH16R04CT. Ultrafast recovery diode. Main product characteristics. Features and benefits. Description. Order codes TO-220AB TO-220FPAB

HYPHENATED TECHNOLOGY GUIDE

Order codes Part numbers DPAK (tape and reel) PPAK (tape and reel)

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Subminiature Transmissive Optical Sensor with Phototransistor Output, RoHS Compliant, Released for Lead (Pb)-free Solder Process

Aquaflow nonionic synthetic associative thickeners selection guide

Kinetics of Crystal Violet Fading AP* Chemistry Big Idea 4, Investigation 11 An Advanced Inquiry Lab

Augmented Reality Market Research Report- Global Forecast till 2023

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

STTH6110TV. Ultrafast recovery - high voltage diode. Main product characteristics. Features and benefits. Order codes. Description ISOTOP ISOTOP

89BSD CALCULATION METHOD APPLICATION NOTE

XBC300 Gen2. Fully-automated debonder and Cleaner

Varian Galaxie Chromatography Data System for Preparative HPLC

Model TS105-10L5.5mm Thermopile Sensor

Electronic Gas Flow Computer Smith Meter microflow.net Gas. Calculations. The Most Trusted Name In Measurement

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

DO-15 DO-201AD STTH3R04Q STTH3R04

Spectroscopy of Various Light Sources: The Interactions between Light and Matter ASTR 170B1, Spring 2017, Lab #2. 1 Introduction.

Recovering MC56F8300 Family Devices from Extreme RFI

AN1755 APPLICATION NOTE

Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM. Application Note

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

RamanStation 400: a Versatile Platform for SERS Analysis

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces


Determining the Rate Law and Activation Energy for the Methyl Blue Reaction:

Analysis of Cocoa Butter Using the SpectraStar 2400 NIR Spectrometer

Optimization of temperature-sensitive assays using the Spark Te-Cool

INFORMATION NOTIFICATION

Solar Radiant Heating and Angle of Incidence NS 696 V: Weather and Climate for Educators

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

The Plus for your lab!

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

Analyst Software. Peptide and Protein Quantitation Tutorial

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Eppendorf Reference : The pipette.

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

Stabilus gas springs. Linear Motion

Transcription:

Application Summary Analytes: ethanol concentration, product color Detector: OMA-300 Process Analyzer Process Stream: whiskey Zeroing Fluid: air Calibration Fluid: standard (known proof) whiskey samples Path Length: 1 cm Response Time: every 1-5 seconds Introduction For the purposes of quality control, whiskey manufacturers desire a method to monitor alcohol content (proof) and color across the bottled yield. The absorbance spectrum of a whiskey sample is an excellent, non-invasive tool for determining these specifications in real time. Ethanol and water both have unique structural features in their absorbance spectra which will have stronger or weaker prominence depending on their concentrations in the measured whiskey sample. Similarly, the absorbance spectrum of a whiskey sample can easily be correlated to a host of color scales (both standard and arbitrary), such that product color can be verified using absorbance spectroscopy. The OMA is a single analyzer providing extremely fast response on both ethanol concentration and product color. Measuring a full, high-resolution absorbance spectrum in the continuously drawn whiskey sample, the instrument is able to lock in the background absorbance (i.e. the total absorbance structure of all the non-ethanol compounds in whiskey) and isolate the absorbance of ethanol, which correlates directly to its real-time concentration. APPLIED TM A N A L Y T I C S [1]

Test Results of Ethanol and Water Concentration Analysis The case study system was calibrated using eight standard samples of fine, high-proof Bourbon whiskey furnished by Brown- Forman of Louisville, Kentucky. Below, the characteristic absorbance peaks of both ethanol and water are displayed. The height of the ethanol peak correlates directly to its concentration in the sample. The eight samples display a variety of proofs based on absorbances at this peak wavelength. The OMA measures the real-time concentration of both ethanol and water in order to calculate proof. Figure 01: Absorbance curves of ethanol and water in a whiskey background matrix. When tested against lab readings on a whiskey stream with unknown proof, the OMA demonstrated exceptional accuracy and background correction: 80 80 70 Concentration (%) 60 Concentration (%) 40 60 50 40 Water (%) Water (%) Water (%) OMA Water (%) OMA Alcohol (%) Alcohol (%) Alcohol (%) OMA Alcohol (%) OMA 30 20 20 1 2 3 4 5 6 7 8 Sample # (3 trials per) Figure 02: Trend graph comparing OMA accuracy with lab results across 8 samples. [2]

The OMA software can simultaneously produce a 1st derivative visualization for the absorbance spectrograph: Figure 03: 1st derivative of water and alcohol absorbance curves. Running the OMA on our whiskey samples and comparing the continuous measurement log against the labels on the samples produced the results in the table below. Any error in the sample prepration would have to be considered as a factor for any discrepancy between the bottle label and the OMA reading. Figure 04: Table of Results [3]

Test Results of Whiskey Color Analysis Product color is known to correlate to the steepness of the curve within the 400-600nm wavelength range of the whiskey absorbance spectrum. In the experimental spectra (Figure 3), the colors of the samples can easily be differentiated by the separation in the overlapping curves. In the graph, the samples range from darkest to lightest as follows: (DARKEST) S1 = S5 > S3 = S7 > S4 = S8 > S2 = S6 (LIGHTEST) ` 2.0 Absorbance (AU) 1.0 0 400 600 800 1000 Wavelength (nm) Figure 05: Color curves of the whiskey absorbance spectra. Steepness of curve indicates relative darkness. These absorbance curves in the spectra above can easily be translated to any sort of arbitrary color scale or standard scales (e.g. Degrees Lovibond, APHA, and more). To demonstrate how we can normalize the color analysis to an arbitrary scale, we subtracted the absorbance at the 590 nm wavelength from the absorbance at the 460 nm wavelength to produce the graph and table below: Figure 06: Spectrograph of whiskey color range. The vertical lines indicate measurement and reference wavelengths. [4]

Using the two indicated wavelengths above, we produced the table of values below. We used 3-wavelength ranges for the measurement wavelength and the reference wavelength to allow for some statistical averaging of multiple data points. This signifcantly reduces the effect of signal noise at any single photodiode since the correlated color value from each of the 3 absorbance data points is averaged. Figure 07: Color to Absorbance Correlation Results Conclusion As demonstrated herein, the OMA is highly suited to monitoring whiskey product quality, defined in this particular application as consistent proof (ethanol level) and color. The OMA offers an elegant solution as a one-stop device for both parameters. Ethanol concentration and whiskey color are both monitored in real-time in a continuously drawn sample. The accuracy of the system aligns closely with lab measurement of these parameters. [5]

Revised 10 April 2017 Further Reading Subject OMA-300 Process Analyzer Advantage of Collateral Data Multi-Component Analysis Location http://www.a-a-inc.com/documents/aa_ds001a_oma300.pdf http://www.a-a-inc.com/documents/aa_tn-202_collateraldata.pdf http://www.a-a-inc.com/documents/aa_tn-203_multicomponentanalysis.pdf is a registered trademark of Applied Analytics, Inc. www.aai.solutions Headquarters Applied Analytics, Inc. Burlington, MA sales@aai.solutions North America Sales Applied Analytics North America, Ltd. Houston, TX sales@appliedanalytics.us Asia Pacific Sales Applied Analytics Asia Pte. Ltd. Singapore sales@appliedanalytics.com.sg Middle East Sales Applied Analytics Oil & Gas Operations, L.L.C. Abu Dhabi, UAE sales@appliedanalytics.ae India Sales Applied Analytics (India) Pte. Ltd. Mumbai, India sales@appliedanalytics.in Europe Sales Applied Analytics Europe, AG Genève, Switzerland sales@appliedanalytics.eu Brazil Sales Applied Analytics do Brasil Rio de Janeiro, Brazil vendas@aadbl.com.br 2017 Applied Analytics, Inc. Products or references stated may be trademarks or registered trademarks of their respective owners. All rights reserved. We reserve the right to make technical changes or modify this document without prior notice. Regarding purchase orders, agreed-upon details shall prevail. [6]