FLCS Fluorescence Lifetime Correlation Spectroscopy

Similar documents
LABORATORY OF ELEMENTARY BIOPHYSICS

Dual-Focus Fluorescence Correlation Spectroscopy

Self-calibrated, line-scan STED-FCS to quantify lipid dynamics in model and cell membranes

Fluorescence Lifetime Correlation Spectroscopy (FLCS): Concepts, Applications and Outlook

Laboratory 3&4: Confocal Microscopy Imaging of Single-Emitter Fluorescence and Hanbury Brown and Twiss setup for Photon Antibunching

FLUORESCENCE MICROSCOPY TECHNIQUES PRACTICAL MANUAL FOR

3. Excitation and Detection of Fluorescence

Ultrafast Dynamics and Single Particle Spectroscopy of Au-CdSe Nanorods

Confocal Microscope Imaging of Single emitter fluorescence and Observing Photon Antibunching Using Hanbury Brown and Twiss setup. Lab.

Measuring Colocalization within Fluorescence Microscopy Images

Laboratory 3: Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown, and Twiss Setup for Photon Antibunching

LAB 3: Confocal Microscope Imaging of single-emitter fluorescence. LAB 4: Hanbury Brown and Twiss setup. Photon antibunching. Roshita Ramkhalawon

Selected measurements with FluoTime 300

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown and Twiss Photon Antibunching Setup

I. Proteomics by Mass Spectrometry 1. What is an internal standard and what does it accomplish analytically?

Using Alba with the FemtoFiber laser by Toptica for 2-photon quantitative imaging

The Photon Counting Histogram: Statistical Analysis of Single Molecule Populations

CHEM Outline (Part 15) - Luminescence 2013

Single Molecule Spectroscopy and Imaging

Diagnostics of Filamentation in Laser Materials with Fluorescent Methods

Lecture 32. Lidar Error and Sensitivity Analysis

Lab 3 and 4: Single Photon Source

Supplementary Figure 1 Comparison of single quantum emitters on two type of substrates:

single-molecule fluorescence resonance energy transfer

This document contains the following supporting information: 1. Wide field scanning electron microscope image

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Γ43 γ. Pump Γ31 Γ32 Γ42 Γ41

Singlet. Fluorescence Spectroscopy * LUMO

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

Calibrating Differential Interference Contrast Microscopy with dual-focus Fluorescence Correlation Spectroscopy

Supplementary Figures Supplementary Figure 1: Estimation of the error of the number and brightness of molecules in a single cluster; Simulation

Quantum Optics and Quantum Information Laboratory

Time-resolved methods in biophysics. 3. Fluorescence lifetime correlation spectroscopy

Fluorescence Intensity and Lifetime Distribution Analysis: Toward Higher Accuracy in Fluorescence Fluctuation Spectroscopy

Application of time resolved area normalized emission spectroscopy to multicomponent systems

Photoluminescence Spectrometer (FLS980)

Supplementary Information: Three-dimensional quantum photonic elements based on single nitrogen vacancy-centres in laser-written microstructures

Electronic Supplementary Information

CHAPTER 3 RESULTS AND DISCUSSION

Enhancement of Exciton Transport in Porphyrin. Aggregate Nanostructures by Controlling. Hierarchical Self-Assembly

Technical Program. 3rd International Workshop on "Single Molecule Detection: Basics and Applic ations in Life Sciences" List of Sponsors.

FROM LOCALIZATION TO INTERACTION

Time-resolved Molecule Counting by Photon Statistics Across the Visible Spectrum

Quantitative fluorescence correlation spectroscopy in three-dimensional systems under stimulated emission depletion conditions: supplementary material

Increasing your confidence Proving that data is single molecule. Chem 184 Lecture David Altman 5/27/08

Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll.

Supporting Information

Chapter 6 Photoluminescence Spectroscopy

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement :

Confocal Microscope Imaging of Single-Emitter Fluorescence and Photon Antibunching

Quantum yield determination by low-intensity Fluorescence Correlation Spectroscopy (lifcs)

LAB 4: Gamma-ray coincidence spectrometry (2018)

Measurement Examples. Excitation and Emission Scans. Steady State Fluorescence Anisotropy. Kinetic Measurements

requency generation spectroscopy Rahul N

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

High-Speed Switching of IR-LEDs (Part I) Background and Datasheet Definition

FSA TM Multi-bit Digital Processors

arxiv: v1 [physics.ins-det] 3 Feb 2011

Quantum and Nano Optics Laboratory. Jacob Begis Lab partners: Josh Rose, Edward Pei

Chapter 15 Molecular Luminescence Spectrometry

Lab 3-4 : Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury-Brown and Twiss Set Up, Photon Antibunching

Chem 681: Student Seminar Series. Two-Photon Induced Fluorescence

File Name: Supplementary Information Description: Supplementary Figures and Supplementary Note. File Name: Peer Review File Description:

Self-assembled Nanoscale DNA-porphyrin Complex for. Artificial Light-harvesting

Implementation and evaluation of data analysis strategies for time-resolved optical spectroscopy

D \ Basis for Detection. Rapid DNA SequencinglSingle-Molecule Spectroscopy. Steven A. Soper, Lloyd A.!?.Davis, and E. Brooks Shera

Parametric down-conversion

Supplementary Figures

Anti-Bunching from a Quantum Dot

MAPPING OF ATOMIC NITROGEN IN SINGLE FILAMENTS OF A BARRIER DISCHARGE MEASURED BY TWO PHOTON FLUORESCENCE SPECTROSCOPY (TALIF)

LASCAD Tutorial No. 4: Dynamic analysis of multimode competition and Q-Switched operation

QuantumMCA QuantumNaI QuantumGe QuantumGold

Modular scanning FCS quantifies receptor-ligand interactions in living multicellular organisms

Supplementary Material for In situ frequency gating and beam splitting of vacuum- and extreme-ultraviolet pulses

Correlation Spectroscopy in Polymer Physics Methodenseminar im Wahlpflichtfach Basics diffusion and brownian motion correlations functions

Solution set for EXAM IN TFY4265/FY8906 Biophysical microtechniques

Radiation-matter interaction.

Determining the orientation of the emissive dipole moment associated with dye molecules in microcavity structures

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Imagent for fnirs and EROS measurements

Rotational Brownian motion; Fluorescence correlation spectroscpy; Photobleaching and FRET. David A. Case Rutgers, Spring 2009

Time-Resolved Liquid Scintillation Counting

Labs 3-4: Single-photon Source

Supporting Information for. Near infrared-to-blue photon upconversion by exploiting direct. S-T absorption of a molecular sensitizer

How DLS Works: Interference of Light

Direct, Reliable, and Correct Oxygen Measurement

Supplementary Materials

P. Lambrev October 10, 2018

PoS(PhotoDet 2012)010

Effects of Temperature and Concentration on the Rate of Photo-bleaching of Erythrosine in Water

ORTEC. Counters/Ratemeters/ Multichannel Scalers. Choosing the Right Counting Solution. The Basic Functions of Counting Systems

PHYS 219 General Physics: Electricity, Light and Modern Physics

Chapter 2. Time-Resolved Raman Spectroscopy. Technical aspects and instrumentation

Fluorescence Depolarization. Ben Berger and Ben Berger

SCINTILLATION DETECTORS AND PM TUBES

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

BMB Class 17, November 30, Single Molecule Biophysics (II)

Secondary Ion Mass Spectrometry (SIMS)

Matthias Lütgens, Frank Friedriszik, and Stefan Lochbrunner* 1 Concentration dependent CARS and Raman spectra of acetic acid in carbon tetrachloride

Fitting Function for Experimental Energy Ordered Spectra in Nuclear Continuum Studies

Transcription:

Application Note FLCS Fluorescence Lifetime Correlation Spectroscopy Peter Kapusta, Michael Wahl, PicoQuant GmbH Aleš Benda, Martin Hof, J. Heyrovsky Institute of Physical Chemistry, Prague, CZ Jörg Enderlein, Forschungszentrum Jülich, Germany Introduction The fusion of Time-Correlated Single Photon Counting and Fluorescence Correlation Spectroscopy, called Fluorescence Lifetime Correlation Spectroscopy (FLCS), is a method that uses pico-second time-resolved fluorescence detection for separating different FCS-contributions. The main emphasis is on the word separating. FLCS does not involve fitting of a complicated model equation to a complicated autocorrelation function. Instead, a separated autocorrelation function is calculated for each expected intensity component, emitted for example by various species in the sample. The on-ly assumption is that various emissions have diffe-rent TCSPC histograms (i.e. different fluorescence lifetimes), which is practically always the case. The core of the method is a statistical separation of dif-ferent intensity contributions, performed on a single photon level. An essential requirement for FLCS is a sub-nano-second pulsed excitation instead of conventional continuous wave (CW) illumination. The second keystone is the ability to simultaneously measure the fluorescence photon arrival time on two diffe-rent time scales: relative to the excitation pulse with picosecond resolution and relative to the start of the experiment with nanosecond precision. The experimental and analysis technique outlined here seriously pushes the limits of FCS. The idea is already four years old [1], but received only little attention until now, though two recent articles [2, 3] present exciting new developments made possible by utilizing this method. This poster explains the basic principles of FLCS. The principle FLCS is best understood in comparison to classical FCS with CW excitation and single channel detec-tion. The autocorrelation function (ACF) is calcu-lated from the various arrival times of detec-ted fluorescence photons, either by a hardware autocorrelator or by software processing of the record of arrival times. If the sample contains, for example, two kinds of diffusing fluorophores, both contribute to the detected intensity fluctuations and the resulting ACF is a complicated superposition, not a linear combination. Taking into account the ubiquitous triplet-populations, scattered excitation light, impurities, etc., the analysis of such an ACF is complex owing to the number of adjustable parameters in the model equation used for fitting. Resolving an ACF is difficult. It is well known, for example, that the diffusion times of two species must differ at least by a factor of 1.6 in order to distinguish them [4]. In FLCS data acquisition, the excitation is pulsed and two independent timings are performed and recorded for every detected photon. 1) The macroscopic arrival time, measured relative to the beginning of the experiment, contains the information related to, e.g., diffusion and is used just like in a classical FCS. 2) The microscopic delay time (in TCSPC traditionally called channel number) measured relative to the onset of the excitation pulse, re-presents a new quality which is essential for determining the contribution of the photon to the ACF. These two timings are evaluated simultaneously when calculating the ACF. However, the calculation itself is not straightforward. One has to separate the various contributions to the detected intensity trace first. As an example, let us consider a sample with two fluorescent components, A and B, which have different decay behaviour. Let us assume that, say, 60% of all photons captured during the FCS ex- -periment were emitted by compound A and the rest by B. In case of CW excitation (see Fig.1), when a single photon is detected there is always a 60% probability that this was emitted by A. This probability of origin is constant in time, because there is no defined moment of excitation. FLCS is based on a simple fact that with pulsed excitation the probability of origin of a detected photon is changing in time, because A and B are decaying differently (see Fig.2). The measured microscopic delay time of a photon (expressed as a channel number, i) allows for a statistical evaluation of its origin. Roughly speaking, if the microscopic delay of a photon is long, then it was most probably emitted by the slower decaying fluorophore, though there is a finite probability that it came from the faster decaying one. Note that the TCSPC histogram of the mixture of A and B, Decay(i), is readily available in FLCS. PicoQuant GmbH, 2006 Application Note: FLCS Seite 1

Let us now define two functions, f a (i) and f b (i) with the following behaviour: Relative photon detection probability Fig. 1: CW excitation: Note that all photons are equal. On this time scale, the probability of their origin is constant. Relative photon detection probability 0.6 0.4 Because there is no well defined moment of excitation, the detected photons are uncorrelated on the ps/ns timescale. There is always a 60% probability that the detected photon was emitted by compound A. There A is always a 40% probability that the detected photon was emitted by compound B. B A B Photons with channel numbers in this range are more likely emitted by A, rather than B. However, there is a considerable probability of origin from B. A(i), decay component contributed by fluorophore A. (60% of all photons.) Decay(i), decay curve of a mixture of A and B. This histogram is always readily available in FLCS. B(i), decay component contributed by fluorophore B. (40% of all photons.) Photons with these channel numbers are most likely emitted by B. As the channel number increases, there is less and less probability that such a photon was emitted by A, but this probability never decreases to zero. Fig. Note that 2: Pulsed the vast majority excitation: of photons Note has that a very the ambiguous vast majority origin. However, of photons the relative has probability a very of ambiguous origin changes origin. during the However, decay and can the berelative probability determined for of each origin single changes photon, looking during up its the channel decay number. and can be determined for each single photon, looking up its channel number. Now it remains to explain how to use the probability of origin of a photon for the calculation of the ACF of a selected component. The shape of A(i) and B(i) histograms can be obtained from independent TCSPC measurements, or from the analysis (e.g. fitting) of Decay(i). Let us normalize A(i) and B(i) to a unit area, a(i)=a(i)/σa(i), b(i)=b(i)/σb(i), that is Σa(i)=Σb(i)=1. We will refer to these areanormalized curves as to decay patterns. It is obvious that the decay curve of a mixture, Decay(i), can be expressed as a linear combination, Decay(i)=w a a(i)+w b b(i), where w a and w b are the amplitudes of the photon count contribution (in number of photons) of compounds A and B, respectively. Σ f a(i) Decay(i) = w a Σ f b(i) Decay(i) = w b The brackets... denote averaging over infinite number of measurements. Further requirement for f a (i) and f b (i) functions is that they must simultaneously minimize the relative errors: (Σ f a(i) Decay(i) - w a ) 2 (Σ f b(i) Decay(i) - w b ) 2 One can see from the functional form of the above expressions, that f a (i) and f b (i) act like statistical filters, or weighting schemes, applied to the content of each TCSPC histogram channel. These func-tions can be numerically calculated from Decay(i) and from the decay patterns. The relevant matrix manipulation is described in reference [1]. The key point is that using f a (i) and f b (i) makes possible to statistically separate the intensity contributions of A and B, photon by photon. The sign and magnitude of a single photon contribution to an intensity trace can be determined from the photon's channel number through the corresponding filter function. Having obtained f a (i) and f b (i), one can start a software correlation, for example as described in reference [5]. In classical FCS, every photon con-tributes equally and the calculation involves only zeros and ones. In FLCS, there is one more step involved: One has to look up the filter value corresponding to the photon's channel number. It is valuable to have a look at the shape of f a (i) and f b (i) calculated for the above hypothetical example (See Fig. 3). - - - f a (i) f b (i) Fig. 3: Example of statistical filter functions Note that the filter value entering the calculation is not an integer and can be even negative! However, the sum of the two filter values is always exactly 1. Applying the f a (i) filter during software correlation of all photon records one obtains the ACF of compound A. The same holds for f b (i) and compound B. PicoQuant GmbH, 2006 Application Note: FLCS Seite 2

Remarkable features of FLCS The method represents a quasi-multichannel detection scheme. FLCS is a lifetime analogy of a multicolor FCS measurement, with several in-herent advantages (see below). FLCS goes far beyond the capabilities of simple time-gating [6]. It separates the ACFs of diffe-rent signal components (e.g. emitters) quantita-tively, because it uses all photons. It is possible to separate even more contributions, provided that their decay patterns are sufficiently diffe-rent. In general, no assumptions on the functio-nal form of the decay patterns are involved. Owing to the separation principle based on TCSPC decay behaviour, distinct ACFs of two compounds can be obtained even if they have equal diffusion times. FLCS works equally well for emitters with completely overlapping fluores-cence spectra. No assumptions are made on the resulting ACFs. One can continue the analysis with the u-sual fitting of standard FCS models. Once the intensity contributions are separated, cross-correlation is straightforward. [1,2] Note that the signal containing the various contribu-tions is recorded by a single detector, hence the separately calculated ACFs correspond to the same detection volume, which can be inter-preted as two (or more) perfectly overlapped excitation/detection volumes. FLCS is easy to imple-ment. It is not necessary to change the optical hardware of a contem-porary FCS setup whatsoever, because it contains almost all the required components. Affordable diode lasers (e.g., the PicoQuant PDL 800-B, PDL 808 "Sepia" or PDL 828 Sepia II driven laser heads) are well suited for pulsed excitation. Of course, existing picosecond pulsed excitation systems can be used as well. There are also commercially available solutions for the second keystone, the simultaneous mea-surement of the photon arrival time on two dif-ferent time scales. (For example PicoQuant's TimeHarp200 PC board or the Pico- Harp300 TCSPC system.) Measurement examples Quasi-multichannel detection The ability of FLCS to separate the ACFs of two components (for example for simultaneous tracking of concentrations) was demonstrated in references [1] and [2]. When appropriate for the sample, a logical step further is the crosscor-relation analysis of the separated intensities, as described in [2]. Furthermore, FLCS is a very versatile method to purge the ACF from various unavoidable experimental artefacts.[3] 10 5 Decay(i) from FLCS 10 4 Filter function for Atto655 emission Counts 10 3 A(i) Atto655 10 2 B(i) = Decay(i)-A(i) Scattered light, parasitic contributions, most of the backround 10 1 0 50 100 150 200 250 300 - Filter function for the remaining contributions - 0 50 100 150 200 250 300 40 35 30 FLCS with a statistical filter for Atto655 emission yields the correct particle number of N=30 ACF 25 20 15 10 Classical FCS yields an apparent particle number of N=94 5 0 10-7 10-6 10-5 10-4 10-3 10-2 10-1 10 0 time /s Fig. 4.: FLCS can be used to remove the influence of scattered light. The example shows results from measurements with Atto655. PicoQuant GmbH, 2006 Application Note: FLCS Seite 3

Suppressing of scattered light and various para sitic contributions At very low concentrations, typical for FCS, a considerable portion of the detected intensity is Rayleigh and Raman scattered excitation light. In this example the sample was a 10 picomolar solution of Atto655 in ethylene glycol. Analysis of the TCSPC histogram reveals that only 60% of the photons are emitted by Atto655 molecules. The decay pattern of pure Atto655 was obtained in a separate measurement of a nanomolar solution, where the amount of parasitic con-tributions can be neglected. Using the appro-priately scaled Atto655 pattern, first the "decay curve" of the remaining contribution was deter-mined, and then appropriate statistical filters were calculated (See Fig. 4) Note the difference of ACF amplitudes. FLCS yields the correct particle number (i.e. concentration). The effect of the presence of scattered photons is strongly non-linear. Their contribution to the total detected intensity is only 40%, but the apparent particle number is increased by a factor of more than 3. Detector afterpulsing is a very common and often overlooked instrumental artefact that leads to a dis-tortion of the ACF at short lag times. A fast initial decay of an ACF can be easily misinterpreted as a triplet contribution. At high excitation repetition ra-tes (in this example: 80 MHz), the spurious after-pulsing events are evenly distributed among the histogram channels, therefore appear as an increa-sed background. [3] This simple temporal be-haviour makes it easy to obtain the two patterns and the corresponding statistical filters (See Fig. 5). Of course, only the filter corresponding to the "pure" decay is suitable for correlation analysis. The ACF obtained with afterpulsing suppression by means of FLCS can be fitted with a simple diffusion model, unlike the conventionally calculated ACF. The sample was a 100 picomolar water solution of Atto655. 10 5 10 4 Decay(i), from FLCS Filter function for "pure" decay Counts 10 3 10 2 B(i) Background level 10 1 Decay(i)-B(i) Background subtracted decay curve 10 0 0 100 200 300 400 500 600 - - Filter function for the background - 0 100 200 300 400 500 600 ACF 3.5 3.0 FLCS: The effect of afterpulsing is suppressed. The ACF can be fitted with the model of a single diffusing particle. Classical FCS: The detector afterpulsing causes a fast initial decay of the ACF. 10-7 10-6 10-5 10-4 10-3 10-2 10-1 time / s Fig. 5: Example of FLCS used to remove influence of afterpulsing from detectors PicoQuant GmbH, 2006 Application Note: FLCS Seite 4

References [1] Böhmer M., Wahl M., Rahn H.-J., Erdmann R., Enderlein J., Time-resolved fluorescence correlation spectroscopy, Chemical Physics Letters, Vol.353, p.439-445 (2002) [2] Benda A., Hof. M., Wahl M., Patting M., Erdmann R., Kapusta P., TCSPC upgrade of a confocal FCS microscope, Review of Scientific Instruments, Vol.76, 033106 (2005) [3] Enderlein J., Gregor I., Using fluorescence lifetime for discriminating detector afterpulsing in fluorescence-correlation spectroscopy, Review of Scientific Instruments, Vol.76, 033102 (2005) [4] Meseth U., Wohland Th., Rigler R., Vogel H., Resolution of fluorescence correlation measurements, Biophysical Journal, Vol.76, p.1619-1631 (1999) [5] Wahl M., Gregor I., Patting M., Enderlein J., Fast calculation of fluorescence correlation data with asynchronous time-correlated single-photon counting, Optics Express, Vol.11, p.3583-391 (2003) [6] Lamb D. C., Schenk A., Röcker C., Scalfi-Happ C., Nienhaus G. U., Sensitivity enhancement in fluorescence correlation spectroscopy of multiple species using time-gated detection, Biophysical Journal, Vol.79, p.1129-1138 (2000) Copyright of this document belongs to PicoQuant GmbH. No parts of it may be reproduced, translated or transferred to third parties without written permission of PicoQuant GmbH. All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or ommisions. Specifications and external appearences are subject to change without notice. PicoQuant GmbH Rudower Chaussee 29 (IGZ) D-12489 Berlin Germany Phone +49-(0)30-6392-6560 Fax +49-(0)30-6392-6561 Email info@picoquant.com WWW http://www.picoquant.com