Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution

Similar documents
Tips & Tricks GPC/SEC: Inter-Detector Delay

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Tips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation

How to use GPC/SEC for compositional analysis

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

c) fitting of the NMR intensity in dependence of the recycle delays 4

The power of multi-detector SEC in the analysis of antibodies

Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?

Differentiation of polymer branching and composition using the Mark Houwink plot

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

GPC/SEC An essential tool for polymer analysis

An Introductions to Advanced GPC Solutions

GPC / SEC Theory and Understanding

Progress toward reliable NC molecular mass distribution by GPC

Sem /2007. Fisika Polimer Ariadne L. Juwono

Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection

GPC/SEC Analysis of Polymer Degradation

How switching mobile phases can improve your OMNISEC results

Gel Permeation Chromatography Basics and Beyond eseminar March 13, Jean Lane Technical and Applications Support LSCA, Columns and Supplies

Macromolecular Chemistry

Molecular Weights of Copolymers Obtained by Gel Permeation Chromatography Light Scattering

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments

EcoSEC - semi-micro GPC/SEC system for high throughput polymer analysis

Tips & Tricks GPC/SEC: Branching Analysis

GPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

COURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download)

Polymer analysis by GPC-SEC. Technical Note. Introduction

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Gel Permeation Chromatography

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

GPC/SEC Troubleshooting Volume 1

Polymer Analysis by Gel Permeation Chromatography

GPC Streamliner. » Continues on Page 2. Focus: accurate GPC/SEC. Sieve Curves. In This Issue

How Molecular Weight and Branching of Polymers Influences Laser Sintering Techniques

Setting the Standard for Advanced GPC/SEC Systems

Particle Characterization Laboratories, Inc.

Chapter 14. Molar Mass Distribution.

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Gel Permeation Chromatography

Optimizing GPC Separations

Setting the Standard for GPC. Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies. The Right Instrument for Your Application

Advanced GPC. GPC On Tour, Barcelona, 28 th February The use of Advanced Detectors in GPC

Practical Steps in GPC Method Development

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Molecular Weight of Polymers *

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

GPC/SEC Column Selection & Method Development

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Chromatographic Separation

Use of SEC-MALS. (Size Exclusion Chromatography - Multi Angle. Light Scattering) for protein quality and characterization

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

MALVERN OMNISEC RESOLVE, REVEAL, REALIZE MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

MALVERN OMNISEC RESOLVE, REVEAL, REALIZE MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

Analysis of Star Polymers Using the Agilent 1260 Infinity Multi-Detector GPC/SEC System

Vol6 No1. Use of Gel Permeation Chromatography in Studies of Acacia Polyacantha Gum ABSTRACT

Lecture 4 : Gel Permeation or Size Exclusion Chromatography

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

Response to Questions Raised by Tom Phillips Concerning the Enforcement Analyitical Method for Avail and Nutrisphere Products. Verdesain Life Sciences

Malvern Macromolecular Solutions. Innovative Solutions in Material Characterization

Comprehensive Polymer Analysis Strategies

Polymers Reactions and Polymers Production (3 rd cycle)

Maximizing Performance Through GPC Column Selection

Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

The ultimate in desktop particle characterization

GPC/SEC standards. Product guide

CASE STUDY. Degradation of Polyethylene by FTIR and High Temperature GPC

GPC/SEC standards. Product guide

Sample preparation and characterization around SAXS

Supporting Information. for. Hexene. Clark R. Landis*,

This document is a preview generated by EVS

Chromatography Outline

Size-exclusion Chromatography as a Useful Tool for the Assessment of Polymer Quality and Determination of Macromolecular Properties

Branching Revealed: Characterizing Molecular Structure in Synthetic and Natural Polymers by Multi-Angle Light Scattering

PAPER No. 6: PHYSICAL CHEMISTRY-II (Statistical

Application compendium. Authors. Greg Saunders, Ben MacCreath Agilent Technologies, Inc. A guide to multi-detector gel permeation chromatography

GPC-Analytics of (Bio)Polymers with Light scattering detection

Speeding up 2D chromatography (HPLCxSEC) at it s limits. ..fast, faster ups broken...

Determination of Polymer Modifier in Asphalt

Evaluation of round robin test results for a size exclusion chromatography method for the analysis of nitrocellulose

Characterize NOM in typical china s water using HPLC-SEC with continuous wavelength UV absorbance detector

Size exclusion chromatography of branched polymers: Star and comb polymers

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

HPLC Background Chem 250 F 2008 Page 1 of 24

[VIM = 4 R3 gx ( 3)

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Puzzled? Look it up in our GPC Glossary

High Performance Liquid Chromatography

No.106. Aqueous SEC Columns for Analysis of Cationic Polymers: TSKgel PWXL-CP Series. Contents. Page

The ph-responsive behaviour of aqueous solutions of poly(acrylic acid) is dependent on molar mass

Calibration and Normalization of MALS Detectors

PHASE 1: IUPAC SEC/GPC Round Robin Project Report. Title: Repeatability and Reproducibility of Sample Preparation and Analysis in High-Temperature SEC

Effects of Secondary Interactions in Size Exclusion Chromatography

High Performance Liquid Chromatography

Automated and accurate component detection using reference mass spectra

Transcription:

Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution Peter Kilz and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany. Molar masses cannot be measured directly by gel permeation/size-exclusion chromatography (GPC/SEC). The direct results of a GPC/SEC run are chromatograms or elugrams, which show the detector signal intensity versus the elution volume. These elugrams can be easily compared to each other on the same column set. However, without additional information (such as that from a calibration curve) the molar mass averages and polydispersity are not available. These properties can be determined if the relationship between elution volume and molar mass is known. This article explains more. Photo Credit: liangpv/getty Images Many macroscopic properties of macromolecules can be derived from their molar mass distribution (MMD). In contrast to molar mass averages such as M n or M w, which provide reduced information, the MMD describes the complete sample characteristics. Two samples can have the same molar mass averages but have very different molar mass distributions and therefore macroscopic behaviour. The difference between an MMD and a chromatogram can be easily understood using the following example: Two laboratories inject the same sample on different instruments. They have a different number of columns with different lengths and inner diameter. This results in two different chromatograms, which is the primary information. Without additional information it is not possible to decide if these chromatograms result from the same sample or not. It is not even possible to tell from the chromatogram if two peaks in the sample correspond to a species with a narrow or a broad molar mass distribution. A broader-looking peak can have a narrower molar mass distribution than a smaller peak, if the broad peak elutes in a column region with high resolution. However, inter-laboratory comparison and distribution information is easy to achieve if MMDs are compared. This process eliminates the experimental conditions. Ultimately only correctly calculated molar mass distributions allow the direct inter-laboratory and long-term comparison of samples and sample properties. GPC/SEC Data Processing The primary information provided by gel permeation/size-exclusion chromatography (GPC/SEC) measurements is the apparent concentration distribution 4

(chromatogram, h[v]). This is a convolution of sample-related parameters and experimental conditions. The molar mass distribution can be calculated from the signal heights in the elugram by the slice method: Therefore the eluted peak is separated into equidistant time, or volume slices. 1, For the transformation the retention axis (x-axis, volume or time) will first be changed into a molar mass axis. This is achieved using the information from the molar mass versus elution volume relation. In the second step the y-axis is converted into mass fractions in one molar mass interval, w(log M). When determining the correct molar mass distribution, the normalized signal height (h i ) must be divided by the slope of the calibration curve at every elution volume. This correction can only be neglected in the case of strictly linear calibration curves over the complete separation range. However, this is a feature that even most commercial linear mixed bed columns usually do not exhibit. As soon as a typical GPC/SEC fit function (for example, cubic fit, polynomial 3, or polynomial 5) is used to achieve higher result accuracy, the slope correction is necessary because the data recording occurs linearly, while the molar mass change is not linear. In practical terms this means that for the same measured height (h i ) the number of polymer chains on the high molar mass fraction of the elugram is much smaller than on the low molar mass fraction. The differential distribution, w(m), of the molar mass (M) is defined as w(m) = dm/ dm, the mass fraction (m) of the molecule in a dm interval (molar mass). By simple transformation w(m) can be expressed by measured quantities with h(v) detector signal and σ(v) slope of the calibration curve. w(m) h(v) M(V) σ(v) ; σ (V) d log M dv The molecular weight averages can be calculated as: Number average molecular weight: Mn w(m) w(m) M Weight average molecular weight: Mw = w(m) M w(m) [1] [] [3] Please note that the GPC software modules of many high performance liquid chromatography (HPLC) software programmes do not perform the correction with the slope of the calibration curve. This results in incorrect molar mass 4 7

Figure 1: Chromatogram versus molar mass distribution. While a chromatogram shows the detector signal height (y-axis) versus the elution volume (x-axis), a MMD displays w(log M) on the y-axis versus log M on the x-axis. Peak shape and breadth can change during transformation depending on the slope of the non-linear calibration curve (displayed in red). distributions for all setups with typical non-linear GPC/SEC calibration behaviour. The errors caused by this neglect will increase with the width of the sample and decrease with the data recording frequency. This is dangerous when submitting GPC/SEC results to regulatory organizations such as the US Food and Drug Administration (FDA) or the European Medicines Agency (EMEA) or for Registration, Evaluation, Authorization & restriction of CHemicals (REACH) registration. Elution Volume to Molar Mass There are several options to determine a molar mass/elution volume relation. From a practical point of view, the methods can be distinguished between those that use reference materials and those which use static light scattering detectors to measure the molar mass on-line for every sample. 4 8

Figure : Chromatogram versus molar mass distribution. A MMD displays w(log M) on the y-axis versus log M on the x-axis. Graphs that display the signal height or similar on the y-axis for example probably do not show true molar mass distributions. The most common technique when reference materials are used is to calibrate the system with polymer reference materials with a narrow molar distribution. 3 Here different molar masses are measured and the M p values of the reference materials are plotted versus their measured elution volume. A fit function is then chosen that describes the relation between log M and the elution volume. Because GPC/SEC separates according to hydrodynamic volume and not to molar mass, only apparent molar masses (related to the calibration standards) are obtained if the calibration standards and the samples are chemically or structurally different. The deviation for the molar mass averages and the molar mass distribution can often not be predicted. Still the results for different samples can be compared to each other. Since the method is robust and easy-to-use, many laboratories apply such procedures for quality control and sample comparison as well as for applications where absolute molar masses are not required. Calibration methods to overcome this limitation use matching reference materials or any of the following techniques: t Universal calibration using Mark-Houwink coefficients; t Broad standard calibration; t Cumulative match calibration 4; or t Calibration using an on-line viscometer. 5 Another popular approach is the use of on-line static light scattering detectors, such as multi-angle light scattering (MALS), right-angle light scattering (RALS), and low-angle light scattering (LALS). 6 The major advantage here is that a combination of a concentration detector with a light scattering detector measures the molar mass at every elution volume on-line with each sample. It is therefore not necessary to establish a calibration curve with molar mass standards. However, for accurate molar mass determination several sample and instrument parameters need to be known independently on the instrument: t refractive index increment (dn/dc) t slice concentration (c i ) (either calculated from the injected mass or by intensity calibration of the concentration detector) t inter-detector delay between light scattering and concentration detector (RI, UV) t refractive index of the eluent, n Eluent t detector constant of the light scattering detector (instrument calibration with toluene or by a molar mass standard) t normalization coefficients for detector angles (MALS detectors only) If these parameters are known it is then possible to measure the molar mass at every elution volume and to use this to tranfer the chromatogram into a molar mass distribution. Summary t Inter-laboratory comparison of GPC/SEC data requires the use of molar mass distribution data, as the primary result of GPC/SEC runs (the chromatogram) is a convolution of sample related parameters and experimental conditions. t The transfer of the chromatogram into the MMD is a two-step process. First, the x-axis is transferred from elution volume to molar mass, then the y-axis is transferred from detector signal height 4 9

to the mass fraction of the molecule in a molar mass interval the slope of the calibration curve must be taken into account. t For both steps it should be known which molar mass elutes at the elution volume slice. This information can be taken from a calibration curve created using several techniques or using an on-line static light scattering detector to measure the molar mass on-line with the sample. References 1. A. Striegel, W.W. Yau, J.J. Kirkland, and D.D. Bly, Modern Size-Exclusion Liquid Chromatography: Practice of Gel Permeation and Gel Filtration Chromatography (John Wiley & Sons, New York, USA, 009).. E. Schröder, G. Müller, and K.-F. Arndt, Polymer Characterization (Hanser, Munich, Germany, 1998). 3. D. Held, The Column 4(6), 18 1 (008). 4. D. Held and P. Kilz, The Column 9(14), 11 15 (013). 5. D. Held and P. Kilz, The Column 8(), 1 16 (01). 6. D. Held and P. Kilz, The Column 5(4), 8 3 (009). Daniela Held studied polymer chemistry in Mainz (Germany) and works in the PSS software and instrument department. She is also responsible for education and customer training. Peter Kilz studied polymer chemistry in Mainz (Germany) and Liverpool (UK). He is one of the founders of PSS and head of the software and instrument department. He is also involved in customer support and training. E-mail: Website: Dheld@pss-polymer.com www.pss-polymer.com OPTIMIZED BIOPHYSICAL CHARACTERIZATION FROM A SINGLE EXPERIMENT Unleash the full potential of your SEC system! Conjugate composition Aggregate content and size Molecular weight, size and structure Material relationships By adding Malvern s advanced detectors to your SEC you can: Characterize and understand protein behaviour in solution Assess protein oligomeric state Measure aggregate content and size Quantify conjugation in PEGylated and membrane proteins r Measure molecular size and broad conformational changes CLICK HERE TO FIND OUT MORE AT www.malvern.com/fullpotential 4 30