EXPERIMENT NO.5 Measurement of Screw Thread using Floating Carriage Micrometer

Similar documents
Unit III Introduction sine bar Sine bar Working principle of sine bar

Read the following BEFORE getting started:

Trimos LABCONCEPT Universal Length Measuring Instrument

SIR C.R.REDDY COLLEGE OF ENGINEERING ELURU

Seminar report Autocollimator Submitted in partial fulfillment of the requirement for the award of degree Of Mechanical

External Micrometers. C-1

Tutorials. 1. Autocollimator. Angle Dekkor. General

External Micrometers. C-1

2 - Machining Fundamentals Measurement. Manufacturing Processes - 2, IE-352 Ahmed M El-Sherbeeny, PhD Spring-2015

DEPARTMENT OF MECHANICAL ENGINEERING QUESTION BANK. Sub Code/Name: ME 1304/Engineering Metrology & Measurements Year/Sem :III/ V UNIT-I

Uncertainty Analysis of Experimental Data and Dimensional Measurements

The Program. Wear measurement Page 5. Inspection of parts Page 6. Online certificate Page 7. DKD-calibration Page 8. Deviation analysis Page 11

FAST CHECK THICKNESS GAGE

E.G.S. PILLAY ENGINEERING COLLEGE Nagapattinam DEPARTMENT OF MECHANICAL ENGINEERING V SEMESTER REGULATION 2013 CHENNAI

Course: Technology II Training course topic: Metrology

E x t e r n a l M i c r o m e t e r s

External Micrometers. C-1

ITTC Recommended Procedures and Guidelines

Metrology Prof. Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere

DISCONTINUED PRECISION MEASURING FOWLER CALIPERS 1 - VERNIER CALIPERS 4 - ELECTRONIC CALIPERS

Dhanalakshmi College of Engineering

UNIT-II LINEAR AND ANGULAR MEASUREMENTS

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF DIMENSIONAL METROLOGY

Dhanalakshmi College of Engineering

Wenzhou Tripod Instrument Manufacturing Co., Ltd.

Height Master Page 343. Check Master Page 347. Calibration Tools Page 352

Physical Measurements

PAPER 2 THEORY QUESTIONS

Inclined plane with protractor and pulley, roller, weight box, spring balance, spirit level, pan and thread.

USS ProCoil Company. WELCOME To Micrometer Training. Introduction and Lecture Phase I 6/18/2010

Dial Test Indicators Comparison measuring instruments which ensure high quality, high accuracy and reliability.

Introduction. Earlier. To Scales. We talked about. L Dimensions

LINEAR AND ANGULAR MEASUREMENTS

TESA MICROMASTER Electronic Micrometers with Digital Display

Measurement of basic constants: length, weight and time

3B SCIENTIFIC PHYSICS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

The force on a straight current-carrying conductor in a magnetic field is given by,

MAHARASHTRA STATE BOARD OF TECHNICAL EDUCATION (Autonomous) (ISO/IEC Certified)

Instruction for use Plunger Probes M FB and M78-Plus

Al-Saudia Virtual Academy Pakistan Online Tuition Online Tutor Pakistan

UNIVERSITY OF SURREY DEPARTMENT OF PHYSICS. Level 1: Experiment 2E THE CURRENT BALANCE

Auto collimator. Introduction. Objectives: Apparatus: Theory:

Black Granite Surface Plates

Determination of Cauchy s Contants

Can you divide a millimeter into 20 equal parts? Essentially, a vernier does that exactly.

MC200E PARTS LIST

THE SCREW GAUGE. AIM: To learn to use a Screw Gauge and hence use it to find the dimensions of various regular materials given.

OSCILLATIONS OF A SPRING-MASS SYSTEM AND A TORSIONAL PENDULUM

Mass Mass is the amount of matter in an object. SI unit: kilogram (kg) 1 kg = 1000 g 1 g = 1000 mg. An adult brain: 1.3 ~1.4 kg

JTC TOOL. 6Bore Gauge & Micro Hole Test. Dial Indicator (Option)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

UNIT 2 COULOMB S LAW. Objectives. to understand Coulomb s Law qualitatively and quantitatively

GP-B Telescope Determine Metering Tube Length P0364 Rev -

stone A ruler and a measuring cylinder containing some water are available.

FREE FALL. To measure the acceleration of a freely falling object.

Candidate Number. General Certificate of Education Advanced Level Examination June 2010

Week 14 The Simple Pendulum

台灣黑馬牌校正器類 CALIBRATOR METROLOGY

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

When a rigid body is in equilibrium, both the resultant force and the resultant couple must be zero.

RYOBI 10 in. TABLE SAW MODEL NUMBER BTS21

Q11: WHAT IS A MEANT GOOD ELECTRICAL CONNECTION?

1 Measure various dimensions of component using Vernier caliper and dial type vernier caliper. d 3 d 1

3B SCIENTIFIC PHYSICS

Supplementary Notes to. IIT JEE Physics. Topic-wise Complete Solutions. Vernier Calipers Made Easy for Physics Students and Teachers

Measurement devices. Kalevi Aaltonen, Aalto University

3B SCIENTIFIC PHYSICS

Inspection Instrument for Indicators Comparison measuring instruments which ensure high quality, high accuracy and reliability.

Electric Field Mapping Lab 2. Precautions

SAVE UP TO 30% 2016 PRODUCT PROMOTION FULLY CALIBRATED 2 YEAR WARRANTY DISTRIBUTED BY:

Sub. Code:

Lever-type Dial Test Indicators

Determination of Pitch Diameter of Parallel Thread Gauges by Mechanical Probing

Force and Motion. Thought Experiment

Lever-type Dial Test Indicators

1.105 Solid Mechanics Laboratory Fall 2003 Experiment 3 The Tension Test

Coulomb Balance. Figure 1: The Coulomb Balance apparatus.

5. Is it OK to change the spacing of the tick marks on your axes as you go across the page? a. Yes b. No - that screws up the analysis of the data.

Fixturlaser Alignment System

Answer all questions. For each question there are four different answers, A, B, C and D. Choose the one you consider correct.

Metrology Prof. Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere. Lecture 26 Measurement of Gear Elements

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

mag 6 & mag 7 magnifiers


Electric Field Mapping

Educational Objectives Determine which variable affects the frequency of a simple pendulum.

Rotational Motion. Figure 1: Torsional harmonic oscillator. The locations of the rotor and fiber are indicated.

Current Balance. Introduction

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

Electronic Indicators / Dial Gauges / Dial Test Indicators (Lever-Type)

Lab M4: The Torsional Pendulum and Moment of Inertia

Electric Field Mapping

TUTORIAL SHEET 1. magnitude of P and the values of ø and θ. Ans: ø =74 0 and θ= 53 0

MECHANICS LAB AM 317 EXP 4 TORSION OF CIRCULAR RODS

22 Which of the following correctly defines the terms stress, strain and Young modulus? stress strain Young modulus

Increasing of the Stern Tube Bushes Precision by On-Line Adaptive Control of the Cutting Process

PHYSICS 30 ELECTRIC FIELDS ASSIGNMENT 1 55 MARKS

Standard Test Methods for Thickness of Solid Electrical Insulation 1

Force and Motion 20 N. Force: Net Force on 2 kg mass = N. Net Force on 3 kg mass = = N. Motion: Mass Accel. of 2 kg mass = = kg m/s 2.

Transcription:

EXPERIMENT NO.5 Measurement of Screw Thread using AIM:To measure the Major, Minor and Effective diameter of external parallel screw threads using. APPARATUS: EDMM-100CL, Prism (A to D), Wire (1 to 4mm), Specimen, Master, Hooks etc. THEORY:In order to ensure the manufacture of screw threads to the specified limits laid down in the appropriate standard it is essential to provide some means of inspecting the final product. For measurement of internal threads thread plug gauge is used and to check these plug gauges is used for measuring Major, Minor and Effective diameter. FLOAING CARRIAGE MEASURING MACHINE: Fig: EDMM-100CL SPECIFICATIONS: 1. Weight of the machine : Approx. 25kg. 2. L x W x H : Approx. 350mm x 150mmx 140mm 3. List count of micrometer : 0.001 mm 4. Standard micrometer or electronic type. Electronic Micrometer has digital display and Std. 5. Dial type Fiducial Indicator with 0.01mm standard dial. 6. Admit between centre 200 mm 7. Max Diameter capacity 100mm Measuring machine shown in the figure has Base with two small and one big adjustable support knobs provided for leveling the assembled unit. Base hastwo parallel integral V grooves one short and other long. Long groove is for guide pegs located at the bottom of Intermediate Piece or Carriage (B) and smaller for a ball. One more V pair is in the Centre of the base, which is provided for accommodating Centers (E) to hold work -1-

piece. Carriage has two parallel V grooves, one to accommodate two balls and other to accommodate one ball. Underneath the floating top (C) there is one V groove on one side and flat portion on other side. Digital Micrometer is in one bracket (with less width) and dial type fiducial on the other side lever is provided to tighten it. Fig: Different views of -2-

DEFINITIONS RELATED TO PARALLEL SCREW THREADS Major Diameter: The diameter of an imaginary cylinder (termed the major cylinder) which just embraces the crests of the external thread or the roots of an internal thread. Minor Diameter: The diameter of an imaginary cylinder (termed the minorcylinder) which just embraces the roots of an external thread or the crests of an internal thread. Simple effective(or Pitch) Diameter: The diameter of an imaginary cylinder(termed the pitch cylinder ) which intersects the surface of the thread in such manner that the intercept on an generator of the cylinder between the points where it meets the opposite flanks of the thread groove is equal to one half the basic of the thread. The two methods for thread measurement as follows Two Wire Method: The effective diameter of a screw thread may be ascertained by placing two wires orrods of identical diameter between the flanks of the thread, andmeasuring the distance over the outside of these wires. The effective diameter is thencalculated as E=T+P Where T= Dimension under the wires =M 2d M=dimension over the wires, d= diameter of each wire -3-

Fig (a) Fig (b) The wires used are made of hardened steel to sustain the wear and tear in use. Theseare given a high degree of accuracy and finish by lapping to suit different pitches.dimension T can also be determined by placing wires over a standard cylinder of diametergreater than the diameter under the wires and noting the reading R1 and then taking readingwith over the gauge, say R2. Then T=S (R1 R2). P=It is a value which depends upon the diameter of wire and pitch of the thread. If P= pitch of the thread, then P= 0.9605p 1.1657d (for Whitworth thread). P= 0.866p d (for metric thread). to give the effective diameter. The expression for the value of P in terms of p (pitch), d(diameter of wire) and x (thread angle) can be derived as follows: In Fig. since BC lies on the effective diameter line BC= ½ pitch=½ p OP= (d cosec x/2) 2 PA=d (cosecx 2 1) 2 PQ=QC cot x 2=p 4 cot x 2 AQ=PQ AP=(p cot x 2) 4 d (cosec x 2 1) 2 AQ is half the value of P.. P value=2aq =p 2 cot x 2 d (cosecx 2 1) Two wire method can be carried out only on the diameter measuring machinedescribed for measuring the minor diameter, because alignment is not possible by two wires andcan be provided only by the floating carriage machine. In the case of three wire method, two wire, on one side help in aligning the micrometer square to the thread while the third placedon the other side permits taking of readings. -4-

Three Wire Method: This method of measuring the effective diameter is an accurate method. In this threewires or rods of known diameter are used; one on one side and two on the other side [Fig. (a) and (b)]. This method ensures the alignment of micrometer anvil faces parallel tothe thread axis. The wires may be either held in hand or hung from a stand so as to ensurefreedom to the wires to adjust themselves under micrometer pressure. M=distance over wires E=effective diameter r=radius of the wires d=diameter of wires h =height of the centre or the wire or rod from the effective x=angle of thread. Fig (a) From fig.(b), AD = AB cosec x 2 = r cosec x 2 H = DE cot x 2 = p 2 cot x 2 CD = ½H = p 4 cot x 2 H = AD CD r = cosec x 2 p 4 cot x 2 Distance over wires=m = E+2h+2r = E+2(r cosec x 2 p 4 cot x 2)+2r = E+2r (l+cosecx 2) p 2 cot x 2 or M = E+d (1+cosec x 2) p 2 cot x 2 (since 2r = 0 ) Fig (b) (i) In case of Whitworth thread: X = 55, depth of thread = 0.64 p, so that E= D 0.64 p and cosec x 2 = 2.1657 Cot x 2 = 1.921 M = E+d(1l+cosec x 2) p 2 cot x 2 = D 0.64p+d(1+2.1657) p 2(1.921) = D+3.1657d 1.6005p M = D+3.1657d 1.6p -5-

Where D=outside dia. (ii) In case of metric threads: Depth of thread=0.6495p So, E = D-0.6495p. x = 60, cosec x 2 = 2;cot x 2 = 1.732 M = D 0.6495 p+d(l+2) p 2 (1.732) = D+3d (0.6495+0.866)p = D+3d 1.5155p. PROCEDURE: 1. The diameter of setting master or a cylinder should be nearly same as the diameter of the thread gauge to reduce error. 2. First meet the anvil of micrometer to dial gauge and rotate micrometer in forward to rotate needle in dial by 10mm with setting its scale zero at needle positionand make digital micrometer to zero by pressing button. 3. Put master in between centers and make dial needle to zero by moving micrometer. Take the reading on micrometer as R S.. Similarly replace master with a threaded work piece again second reading is taken as R 4. Hang the required set of prisms (from Appendix-1 according to pitch of work piece) on the hooks provided on both sides of standard. Take the reading on micrometer and zero indicator in such a way that portion of prism touches master and flat portion of prism is on micrometer & zero indicator. 5. Take reading for master as R P and work piece as R q on micrometer as needle on dial indicating zero. 6. Now place two small wires or cylinders in place of prism. The wires should be choosen so that, when placed between the threads, they should contact about halfway down theflanks. Suitable sizes of these cylinders for various threads are given atappendix-1 7. Take the readings for master as R W and work piece as R OW on micrometer as needle on dial indicating zero. OBSERVATION: D = Diameter of Setting master = Dimension of work piece = 14M*2-6-

Observation Table: MAJOR DIAMET (M) MINOR DIAMET (C) R S M = D ± (Difference between R and R S ) = R R p C = D ± (Difference between R p and R q ) R q = EFFECTIVE DIAMETER (E) R W T = D ± (R W - R OW ) = R OW E = T + P = CALCULATIONS: Major diameter: R S = Micrometer reading over setting master. R = micrometer reading of threaded work piece. M = D ± (Difference between R and R S ) (Note: The + or is determined by relative size of the master and the work piece.use sign as master is of greater size than work piece and vice-versa for + sign.) Minor diameter: -7-

C = core or minor diameter of the work piece. R p = reading over master with prisms. R q = Reading over work piece with prisms. C = D ± (Difference between R p and R q ) Effective diameter: E = effective or pitch diameter R OW = Reading measured over work piece with wires. R W = Reading measured over setting master with wires. T = measured dimension under cylinders. T = D ± (R W - R OW ) E = T + P P = Constant depending on the pitch and angle of the thread and the mean diameter of wires used. (Reading P values for different threads are given along wires with wires to be used in Appendix -1) -8-

RESULT : Major Diameter = Minor Diameter = Effective Diameter = CONCLUSION: Using the diameter of imaginary cylinder i.e. Effective or pitch diameter can be evaluated. SAMPLE CALCULATIONS OBSERVATION: D = Diameter of Setting master = 16.004mm Dimension of work piece = 14M*2 Observation Table: MAJOR DIAMETERR S 16.05 Major Diameter = D ± (Difference between R an R 14.05 R S ) =14.004mm MINOR DIAMETERR p 26.37 C = D ± (Difference between R p and R q ) R q 21.491 =11.125mm EFFECTIVE DIAMETER R W 18.699 R OW 15.009 T = D ± (R W - R OW ) =12.314mm E = T + P =12.696mm Major Diameter = D ± (Difference between R and R S ) =16.004 (16.05-14.05) = 14.004mm Minor diameter C = D ± (Difference between R p and R q ) = 16.004- (26.37-21.491) =11.125mm Effective Diameter T = D ± (R W - R OW ) = 16.004-(18.699-15.009) =12.314mm E = T + P = 12.314 +0.382 = 12.696mm -9-