December 2010 No 036. The electronic Eppendorf Xplorer pipette Versatile adjustment

Similar documents
Technical Report Exact dispensing of whole blood using the Eppendorf positive displacement system Multipette Xstream /Repeater Xstream

N) justment. Register your instrument! Eppendorf Research plus. Factory Adjustment

ept.i.p.s. Version A

Eppendorf Reference : The pipette.

NEW! Your bacteria s favorite! Eppendorf Eporator

No 022 December Application-Specific Requirements for Consumables in the Laboratory

Applications Note 228 September 2010

Raise the limits. Eppendorf Research plus

Simply better pipetting!

Eppendorf Xplorer Eppendorf Xplorer plus. Ordering information. Register your instrument!

Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced

Technical Report Eppendorf Deepwell Plates and Microplates Investigations of sealing methods

ept.i.p.s. Instructions for use Version B

Simply Better Pipetting

The New Legend. Eppendorf Reference 2 Pipette

Applications Note 164 October 2007

The New Legend. Eppendorf Reference 2 Pipette

Larger Display Easier to Use. Generation E. Repeater E3 and E3x: Dispense without compromising accuracy or precision

PhysioCare Concept improved. The feeling of a new philosophy

PhysioCare Concept improved. The feeling of a new philosophy

Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus

96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing

ept.i.p.s. pipette tips are available with various degrees of purity, with and without filters, and as special tips.

Still Making History. Over 50 years of pipetting

1 Product description. Material. 3 Warnings for intended use. Instructions for use ept.i.p.s. Version B. English (EN)

NEW! ep Dualfilter T.I.P.S. Peak of perfection

The Plus for your lab!

Well within Reach. The Eppendorf Pipette Holder System

Applications Note 215 September 2009

N) manual. Biomaster Operating manual

Spa & Wellness for Pipettes

Get the Eppendorf Advantage!

Applications Note 202 October 2008

Multitalent for transfection and cell fusion. Eppendorf Multiporator

NEW. A class of its own. Centrifuge 5430 / 5430 R

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 0

Spa & Wellness for Pipettes

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Multipette / Repeater M4 allows for fast, precise and sterile liquid transfer in cell culture

Cleaning and Decontamination of the ept.i.p.s. Box

Automation of Routine Microplate Pipetting and Dispensing

PhysioCare Concept. Make your Lab a better place

Eppendorf Plate Deepwell 96 and 384: RecoverMax

stance et Helper Eppendorf Pipet Helper Chemical Resistance

Technology that counts

Superior well-to-well Consistency with Eppendorf Cell Culture Plates

Technology that counts

Technology that counts

Applications Note 224 April 2010

Micropipetting Basics

Automated purification of high quality genomic DNA

NGS Made Easy. Optimize your NGS library preparation with the epmotion Automated liquid handling system

mag nanogram kit Description Kit uses

The New Eppendorf X-Drive - Maximum Performance, Flexibility and Longevity for Demanding Shaker Tasks

Now also in Eppendorf LoBind. Conical Competence. Eppendorf Conical Tubes 15 ml and 50 ml

Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX

Automated Illumina TruSeq Stranded Total RNA library construction with the epmotion 5075t/TMX

Natural Winners. Eppendorf Liquid Handling Instruments: Unrivaled in technology, efficiency and handling for over 50 years

Automated Illumina TruSeq Stranded mrna library construction with the epmotion 5075t/TMX

The first address for Titration. TITRONIC 300 and TitroLine 5000

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Automated Illumina TruSight HLA v2 Sequencing Panel Library Preparation with the epmotion 5075t

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1

RayBio Nickel Magnetic Particles

Natural Winners. Eppendorf Liquid Handling Instruments: Unrivaled in technology, efficiency and handling for over 50 years

user manual blotting Hoefer PR648 Slot blot manifold um PR648-IM/Rev.F0/08-12

Eppendorf Vis Cuvettes. Instructions for use

CHEM 334 Quantitative Analysis Laboratory

Promotion period: September 15 - December 31, Super Sale. Quality at unbeatable prices

Eppendorf Consumables it s Your sample. Peak of Perfection. Eppendorf Liquid Handling Consumables

NEW: 5.0 ml screw cap tube. The Missing Link. Eppendorf Tubes 5.0 ml Discover a new sample handling system

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Instruction Manual HI Refractometer for Sodium Chloride Measurements

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC

New Promotions July 1 December 31, Take Good Care of Your Cells. More great offers for your cell biology processes

ab Maleimide Quantification Kit - Colorimetric

2. MASS AND VOLUME MEASUREMENTS Mass measurement Analytical and standard laboratory balances Pre-lab Exercises

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Density of Aqueous Sodium Chloride Solutions

Eppendorf UVette. Instructions for use

HPLC Sample Preparation

The Pure Truth. Eppendorf Forensic DNA Grade according to ISO 18385

Electrical safety testing Checklist

L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader

Application Note. Automated pipetting and analysis of polymer suspensions for process control in chemical production

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

For the rapid, sensitive and accurate measurement of Formate levels in various samples.

Eppendorf SmartBlock. Instructions for use

Ammonia Assay Kit Modified Berthelot (Colorimetric)

Automated pipetting made easy new ep Motion 5073 family members for perfect PCR set-up and nucleic acids purifi cation or general liquid handling

Inserting grids into capsule Removing capsules from grid box Immunolabeling using template guide Preparing for TEM insertion

Evaluating the functionality of Eppendorf BioPhotometer and Eppendorf BioSpectrometer using a Secondary UV-VIS Filter Set

Instruction Manual HI Refractometer for Invert Sugar Measurements

Highest Precision. DASGIP PHPO, GA and OD4 for monitoring and control of bioprocesses

Bio 120 Lab 5: Quantitative Analysis

RayBio Rat TNF-alpha ELISA Kit (For Lysates)

Transcription:

Userguide December 2010 No 036 The electronic Eppendorf Xplorer pipette Versatile adjustment Kornelia Ewald, Eppendorf AG, Hamburg, Germany Abstract Piston-stroke pipettes with an air cushion are subject to the effects of temperature, air pressure and air humidity. Such effects are kept to a minimum by constructive measures which are designed to ensure that dispensing accuracy is not significantly affected. Liquids whose physical properties differ significantly from those of water, or temperature differences between the pipette, the pipette tip and the dispensed liquid may also lead to incorrect dispensing volumes. These factors make it necessary to be able to readjust the devices for different physical conditions. Thanks to its numerous adjustment possibilities the new electronic pipette Xplorer from Eppendorf allows for easy compensation of such physical effects. Introduction Since piston-stroke pipettes are mainly used for pipetting of aqueous solutions, they are adjusted with distilled water as a test medium. Depending on the density of the particular liquid the air volume over the liquid expands in different ways. Consequently, other liquid volumes may, under certain circumstances, be aspirated into the tip when dispensing non-aqueous solutions. Liquids with a high vapor pressure, such as organic solvents, can also not be dispensed with the level of accuracy specified for distilled water. The air pressure, which is dependent on the height of the location above sea level, is a further factor to be considered during dispensing operations with air cushion pipettes.

No 036 Page 2 Adjustment Solutions whose physical data is very different from water with regard to their density, viscosity, surface tension and/ or vapor pressure may lead to incorrect dispensing volumes. This makes it necessary to readjust piston-stroke pipettes with an air cushion to account for this. If the density of an aqueous solution changes by approximately 10 %, for example, because of the salt concentration, the volume will change by approximately 0.2 %. This statement does only apply if other relevant properties of the liquid do not change at the same time. Another reason for changing the factory settings can be, for example, the altitude of the location at which the pipette is used. If the pipette is used at an extremely high altitude, it must be adjusted to the ambient air pressure. At 1000 meters above sea level, the volume error of a 100 µl pipette is about -0.3 %. In addition, when using pipette tips that significantly differ from standard tips in their geometry, changing the adjustment can also improve dispensing accuracy. In contrast to performing calibration which involves determining the measured random (precision) and systematic (accuracy) errors from the nominal value and which does not require any alterations that will permanently change the dispensing system, performing an adjustment will change the device for all subsequent dispensing operations. Changes made to the adjustment do not affect dispensing precision. Precision can only be improved by exchanging parts. Precision is also considerably affected by handling errors. The actual volume of a pipette can be checked by weighing as follows: If the set volume corresponds to the actual volume, no correction is necessary. If there is a difference between the actual volume and the set volume, make sure to check the following factors before readjusting a pipette: Is there any liquid dripping from the tip? Is the pipette tip fitted leak-proof? Is the tip cone undamaged? Are the piston and the cylinder leak-proof? Does the temperature of the pipetted liquid correspond to the temperature of the device and the ambient air? Is the weighing location free from drafts? Does the work method and pipetting speed permit complete aspiration and dispensing of the liquid? Has the correct value for Density liquids at weighing temperature been used for the calculation of the actual volume? Is the pipette volume setting correct? Is the balance sufficiently sensitive (balance resolution 0.001 mg) for very small volumes (< 10 µl)? Were original ept.i.p.s. pipette tips used with the correct volume (see technical data) as test tips? The factory setting of the Eppendorf Xplorer may only be changed after all of the above listed points have been thoroughly checked and the pipette volume setting is still different from the measured volume. The Eppendorf Xplorer offers several individual adjustment options which can be used as an alternative to adjusting the pipette, depending on the application. The following adjustment options are available for selection: ACTUAL volume = Mean value of the weighings Density of liquids at weighing temperature

No 036 Page 3 Adjustment Liquid type ethanol 75 % or glycerol 50 % When this adjustment option is selected, the factory setting is changed by an internal factor which considers the density of the ethanol or the glycerol. This means that the substance can be dispensed with greater accuracy (smaller systematic error) with the Xplorer pipette. ept.i.p.s. long This enables the pipette to be adjusted to the different lengths of a pipette tip, such as, ept.i.p.s. 1200 µl elongated or ept.i.p.s. 10 ml L. When this adjustment option is selected, the tip geometry of the longer tip is considered in the internal volume calculation. This means that the technical data of the original tip (test tip) is reached. This function can also be used when the Xplorer pipette is not used with its test tip, e.g., when using the 300 µl ept.i.p.s. for the pipette size 5 100 μl. When using the following tips dispensing accuracy can be increased with this function: Eppendorf Xplorer 10 µl (medium gray): ept.i.p.s. long 20 µl L (light gray tray) Eppendorf Xplorer 100 µl (yellow): ept.i.p.s. 300 µl (orange tray) Eppendorf Xplorer 1000 µl (blue): ept.i.p.s. long 1250 µl L (dark green tray) Eppendorf Xplorer 1200 µl (green, 8-channels): ept.i.p.s. long 1250 µl L (dark green tray) Eppendorf Xplorer 10 ml (turquoise): ept.i.p.s. long 10 ml L Geographic altitude The mean air pressure of a location depends on its height above sea level. When testing a piston-stroke pipette it is thus important to take fluctuations in pressure into account. At higher altitudes with a drop in air pressure the aspiration volume of a piston-stroke pipette is reduced. The Xplorer pipette s stroke is corrected taking into account the air pressure at the respective altitude. This can be easily accomplished using the Geographic altitude adjustment function that allows the altitude to be adjusted in increments of 250 m (820 ft). The maximum altitude that can be selected is 5000 m. The adjustment options Liquid Type, ept.i.p.s. long and Geographic Altitude can be combined with each other.

No 036 Page 4 Individual adjustment Another possibility of readjusting the Xplorer is to use the Individual adjustment function. With this option, the gradient or axis intercept is changed taking into account the exact density of the solution to be dispensed. 1-point, 2-point or 3-point adjustment are available for selection. To create the weighing results, the use of a fine balance with a high resolution is required. Dispensing volumes below 10 µl require a balance with a resolution of 0.001 mg. The arithmetic operations required for performing the correction are automatically performed by the Xplorer pipette during the 1-3-point adjustment, so that the user does not have to carry out any complicated calculations. Piston stroke of manual pipette 1-point adjustment After you input the density, the selected volume and the corresponding weighing result, the Xplorer determines a correction factor. The factor is only valid for the selected volume and the selected work technique (speed, prewetting, wall dispensing method, etc.). Piston stroke of Xplorer min. Digital volume display of manual pipette Manual pipette; change by a fixed amount for all volumes Fig. 2: Example for the change of the piston stroke of a mechanical pipette 2-point adjustment After you input the density, two different volumes and the corresponding weighing results, the Xplorer determines a correction factor. The factor applies to the volume range between the tested volumes and only for the selected work technique. However, the factor is also used here for the entire volume range, that is, also below and above the two measuring points (Fig. 3). Piston stroke of Xplorer min. 25 % 50 % 75 % Volume display of Xplorer 1-point adjustment Fig. 1: Example for the piston-stroke correction of a 1-point adjustment In this example the factory piston-stroke setting is increased by a factor (Fig.1). Strictly speaking, the correction only applies to the testing volume, but it is used for the entire volume range. The correction is different to that for a mechanical pipette. A correction with a factor results in a smaller stroke correction for a small volume than for a large volume. min. 25 % 50 % 75 % 2-point adjustment at 25 % and 50 % Fig. 3: Example for a 2-point adjustment Volume display of Xplorer In the case of a mechanical pipette the stroke can only be changed by a fixed volume (Fig. 2). This volume change applies to the entire measurement range of the mechanical pipette. The existing adjustment is changed in parallel by a fixed amount.

No 036 Page 5 Individual adjustment 3-point adjustment After you input the density, three different volumes and the corresponding weighing results, the Xplorer determines two correction factors. The factors are correct from measuring point to measuring point in the selected volume regions and only for the selected work technique. However, the respective factor is also used below and above the first or third measuring point (Fig.4). Piston stroke of Xplorer In contrast to 1-point and 2-point adjustment, 3-point adjustment is more accurate. If three significantly different volumes are used for calculating the two correction factors, the corrected volume range is considerably larger and thus also more precise. However, 3-point adjustment is more time-consuming. It is necessary to carry out a gravimetric test for each Xplorer pipette whose factory setting has been changed by one of the above-mentioned adjustment options. This is the only way of ensuring that the selected adjustment meets the required measurement errors. To ensure that other users are informed of the changed adjustment, each Xplorer pipette that has been changed by the adjustment must be additionally marked by a clearly visible label indicating the type of change made. The precise work technique relating to the individual adjustment options is described in the adjustment instructions of the Eppendorf Xplorer (see Xplorer Adjustment Instructions). min. 25 % 50 % 75 % Volume display of Xplorer 3-point adjustment at 25 %, 50 % and 75 % Fig. 4: Example for a 3-point adjustment Factory settings A pipette that has been readjusted can be reset to the original settings using the Reset to Factory Settings adjustment option at any time. Outlook Today, it is expected that pipettes should not only offer ease of use and precision. Other features, such as simple volume adjustment, have become standard requirements for such devices. The electronic pipette Eppendorf Xplorer meets such requirements in every respect, making it ideally suited as an instrument for daily use in the lab.

No 036 Page 6 Eppendorf Xplorer Technical Specifications and Ordering Information Single-channel Eppendorf Xplorer, (incl. charger) Volume range Volume Systematic deviation Random deviation Order no. Medium gray multi-function rocker for 20 µl ept.i.p.s. pipette tips 0.5 10 µl 1 µl ±2.5 % ±0.025 µl ±1.8 % ±0.018 µl 4861 000.015 5 µl ±1.5 % ±0.075 µl ±0.8 % ±0.04 µl 10 µl ±1.0 % ±0.1 µl ±0.4 % ±0.04 µl Yellow multi-function rocker for 200 µl ept.i.p.s. pipette tips 5 100 µl 10 µl ±2.0 % ±0.2 µl ±1.0 % ±0.1 µl 4861 000.023 50 µl ±1.0 % ±0.5 µl ±0.3 % ±0.15 µl 100 µl ±0.8 % ±0.8 µl ±0.2 % ±0.2 µl Orange multi-function rocker for 300 µl ept.i.p.s. pipette tips 15 300 µl 30 µl ±2.5 % ±0.75 µl ±0.7 % ±0.21 µl 4861 000.031 150 µl ±1.0 % ±1.5 µl ±0.3 % ±0.45 µl 300 µl ±0.6 % ±1.8 µl ±0.2 % ±0.6 µl Blue multi-function rocker for 1,000 µl ept.i.p.s. pipette tips 50 1,000 µl 100 µl ±3.0 % ±3 µl ±0.6 % ±0.6 µl 4861 000.040 500 µl ±1.0 % ±5 µl ±0.2 % ±1 µl 1,000 µl ±0.6 % ±6 µl ±0.2 % ±2 µl Violet multi-function rocker for 5 ml ept.i.p.s. pipette tips 0.25 5 ml 500 µl ±3.0 % ±15 µl ±0.6 % ±3 µl 4861 000.058 2,500 µl ±1.2 % ±30 µl ±0.3 % ±6.25 µl 5,000 µl ±0.6 % ±30 µl ±0.15 % ±7.5 µl Turquoise multi-function rocker for 10 ml ept.i.p.s. pipette tips 0.5 10 ml 1,000 µl ±3.0 % ±30 µl ±0.6 % ±6 µl 4861 000.066 Multi-channel Eppendorf Xplorer, (incl. charger) 5,000 µl ±0.8 % ±40 µl ±0.2 % ±10 µl 10,000 µl ±0.6 % ±60 µl ±0.15 % ±15 µl Volume range Volume Systematic deviation Random deviation Order no. 8-channel Order no. 12-channel Medium gray multi-function rocker for 20 µl ept.i.p.s. pipette tips 0.5 10 µl 1 µl ±5.0 % ±0.05 µl ±3.0 % ±0.03 µl 4861 000.104 4861 000.112 5 µl ±3.0 % ±0.15 µl ±1.5 % ±0.075 µl 10 µl ±2.0 % ±0.2 µl ±0.8 % ±0.08 µl Yellow multi-function rocker for 200 µl ept.i.p.s. pipette tips 5 100 µl 10 µl ±2.0 % ±0.2 µl ±2.0 % ±0.2 µl 4861 000.120 4861 000.139 50 µl ±1.0 % ±0.5 µl ±0.8 % ±0.4 µl 100 µl ±0.8 % ±0.8 µl ±0.25 % ±0.25 µl Orange multi-function rocker for 300 µl ept.i.p.s. pipette tips 15 300 µl 30 µl ±2.5 % ±0.75 µl ±1.0 % ±0.3 µl 4861 000.147 4861 000.155 150 µl ±1.0 % ±1.5 µl ±0.5 % ±0.75 µl 300 µl ±0.6 % ±1.8 µl ±0.25 % ±0.75 µl Green multi-function rocker for 300 µl ept.i.p.s. pipette tips 50 1,200 µl 120 µl ±6.0 % ±7.2 µl ±0.9 % ±1.08 µl 4861 000.163 600 µl ±2.7 % ±16.2 µl ±0.4 % ±2.4 µl 1,200 µl ±1.2 % ±14.4 µl ±0.3 % ±3.6 µl The data for systematic and random deviations only applies when using Eppendorf ept.i.p.s. pipette tips. Charging stand Order no. Order no. for 1 pipette 4880 000.000 for 4 pipettes (incl. charging adapter) 4880 000.026 Your local distributor: www.eppendorf.com/worldwide Eppendorf AG 22331 Hamburg Germany Tel: +49 40 53801-0 Fax: +49 40 538 01-556 E-mail: eppendorf@eppendorf.com Eppendorf North America, Inc. 102 Motor Parkway Hauppauge, N.Y. 11788-5178 USA Tel: +1 516 334 7500 Toll free phone: +1 800-645-3050 Fax: +1 516 334 7506 E-mail: info@eppendorf.com Application Support Europe: Tel: +49 1803 666 789 (Preis je nach Tarif im Ausland; 9 ct/min aus dem dt. Festnetz; Mobilfunkhöchstpreis 42 ct/min) E-mail: support@eppendorf.com North America: Tel: +1 800 645 3050 E-mail: techserv@eppendorf.com Asia Pacific: Tel: +60 3 8023 6869 E-mail: support_asiapacific@eppendorf.com eppendorf, Eppendorf Xplorer and ept.i.p.s. are registered trademarks of Eppendorf AG. All rights reserved, including graphics and images. Copyright 2010 by Eppendorf AG. Order No. AU036 WW 020/GB2/WEB/1210/NEUH