The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS)

Similar documents
- A spark is passed through the Argon in the presence of the RF field of the coil to initiate the plasma

ELEMENT2 High Resolution- ICP-MS INSTRUMENT OVERVIEW

Introduction to the Q Trap LC/MS/MS System

PRINCIPLE OF ICP- AES

Mass Spectrometry in MCAL

INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Suzanne Bell Second Edition

The 30-Minute Guide to ICP-MS

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

TANDEM MASS SPECTROSCOPY

a. An emission line as close as possible to the analyte resonance line

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry

Atomic Emission Spectroscopy

ICPMS Doherty Lecture 1

high temp ( K) Chapter 20: Atomic Spectroscopy

Ionization Techniques Part IV

Analysis of Polar Metabolites using Mass Spectrometry

ICP-OES DETERMINATION OF IRON. Introduction:

2101 Atomic Spectroscopy

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Interface (backside) & Extraction Lens

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Atomization. In Flame Emission

Spectrometric Methods of Analysis. OCN 633 Fall 2013

Extrel Application Note

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

and Environmental Science Centre

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Secondary Ion Mass Spectroscopy (SIMS)

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Harris: Quantitative Chemical Analysis, Eight Edition

hydrideicp Hydride Generation System

Exercise 2-2. Titration of a Strong Acid EXERCISE OBJECTIVES

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Thermo VG PQ EXCELL INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

1. The range of frequencies that a measurement is sensitive to is called the frequency

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Determining the Concentration of a Solution: Beer s Law

CHAPTER 4: ANALYTICAL INSTRUMENTATION

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Atomic Spectroscopy AA/ICP/ICPMS:

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

Emission spectrum of H

Introduction to Elemental Analysis

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Air Quality 2 - Determination of Cd and Pb by Anodic Stripping Voltammetry

Analyst Software. Automatic Optimization Tutorial

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

DISCLAIMER: This method:

Chemistry 311: Topic 3 - Mass Spectrometry

MS Goals and Applications. MS Goals and Applications

Trace Analyses in Metal Matrices Using the ELAN DRC II

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

3 - Atomic Absorption Spectroscopy

Protocol Particle size distribution by centrifugal sedimentation (CPS)

Mass Spectrometry. What is Mass Spectrometry?

M M e M M H M M H. Ion Sources

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Information Dependent Acquisition (IDA) 1

GAS CHROMATOGRAPHY (GC)

Agilent G3212 GC-APCI Source

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

Lecture 7: Atomic Spectroscopy

Kinetics of Crystal Violet Bleaching

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Fluorescence Spectrophotometry

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Varian Galaxie Chromatography Data System for Preparative HPLC

Chemistry Instrumental Analysis Lecture 17. Chem 4631

1. Prepare the MALDI sample plate by spotting an angiotensin standard and the test sample(s).

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

MS Goals and Applications. MS Goals and Applications

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Lecture 16 Instrumentation for ICP AES-VIII-Instruments

Chemistry Instrumental Analysis Lecture 35. Chem 4631

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Transcription:

The Easy Guide to: Inductively Coupled Plasma- Mass Spectrometry (ICP-MS) By Arianne Bazilio & Jacob Weinrich December 2012 Contents Introduction... 2 Sample Introduction... 3 Torch... 4 Interface... 6 Mass Spectrometry... 7 ICP-MS User Guide... 9 Works Cited... 11

Introduction Inductively coupled plasma-mass spectrometry (ICP-MS) is a powerful tool for analyzing trace metals in environmental samples. A large range of elements can be detected using an ICP- MS, which are summarized in Figure 1 below. Figure 1. Elements detectable by ICP-MS analysis (Perkin-Elmer) The ICP-MS system can quantitatively measure the colored elements in Figure 1, and give a measurement of the total amount of the specific element of interest. The benefits of using plasma compared to other ionization methods, such as flame ionization, are that ionization occurs in a chemically inert environment, preventing oxide formation, and ionization is more complete. Also, the temperature profile of the torch is relatively uniform, reducing selfabsorption effects. Linear calibration curves are observed over several orders of magnitude for ionization processes. The mechanisms in which ICP-MS analysis occur will be discussed in further detail. The process can be broken down into four stages; sample introduction, ICP torch, interface, and MS. 2

Sample Introduction The first step in ICP-MS is sample introduction. There are multiple ways in which a sample can be introduced, and the method of introduction depends on the physical characteristics of the sample. The various methods of sample entry are shown in Figure 2. While each method may be fundamentally different, they have the same goal: to sweep the sample of interest into the ICP torch in a gaseous or aerosol form for analysis. Figure 2. Potential sample entry pathways. (Skoog, Holler, and Crounch 2007) Typically, samples requiring ICP-MS analysis are in a liquid form. However, the samples must be introduced to the torch in either a gaseous form or an aerosol form. Therefore, liquid samples require sample nebulization, seen in Figure 3. The liquid sample is pumped from a vial, via a peristaltic pump, into the nebulizer. Liquid droplets are formed on the tip of a needle, where they become nebulized due to argon gas flowing through a second needle perpendicular to the sample needle. A small amount of aerosol created is swept into the torch, but the majority of sample condenses on the walls of the nebulizer and is wasted to the drain. 3

Figure 3. Schematic of liquid sample introduction to torch. (Skoog, Holler, and Crouch 2007) If a solid sample requires analysis, the most likely method of introduction would be via electrothermal vaporization. An electrothermal vaporizer uses electric current to rapidly vaporize a solid sample, which can then be swept into the ICP torch via argon gas flow. One fundamental aspect of ICP-MS analysis, which can be viewed as both a strength and weakness, is that elements are detected and quantified in total. In other words, no separation of compounds containing the element of interest occurs. However, prior separation can occur via a chromatographic technique, such as gas chromatography, supercritical fluid chromatography, or liquid chromatography. Because gas and supercritical fluid chromatography separates samples in the gas phase, the eluents produced after column separation can be directly swept into the ICP- MS torch. However, liquid chromatography elutes, being in a liquid phase, require nebulization similar to that of standard liquid sample introduction. Torch The ICP torch consists of a copper induction coil wrapped around a concentric quartz structure, a schematic of which can be seen in Figure 4. 4

Figure 4. ICP torch schematic. (Skoog, Holler, and Crouch 2007) Argon gas is continuously flowing throughout the quartz torch, and a radio-frequency (RF) generator provides power to the RF coil at oscillating frequencies. Plasma (an electrical conducting gaseous mixture) generation occurs when the argon gas is seeded with a spark from a Tesla unit. The spark ionizes some of the argon, and the cations and electrons produced from that accelerate towards the RF coil. The cations and electrons collide with other argon molecules during this acceleration, creating high temperatures. With ample argon supplied, the plasma will reach equilibrium and remain at a constant temperature of about 6,000 C for the duration of analysis. The aerosol produced via nebulization enters this high temperature plasma, where it is first dried to a solid, and then heated to a gas, referred to as atomization. These atoms will continue to travel through the plasma, absorbing energy until they release an electron, becoming ionized, referred to as ionization. These newly formed ions then travel out of the torch and come to the interface. 5

Interface Generally speaking, the interface can be described as the point at which sample from the ICP portion of the instrument is introduced to the mass spectrometry (MS) portion of the instrument. The interface portion of the instrument serves to allow the ICP and MS portions to be coupled. A general schematic of this interface can be seen in Figure 5. Figure 5. General schematic of interface. (Skoog, Holler, and Crouch 2007) The first component the sample matrix confronts after ionization in the ICP torch is the sampler cone. This is a water cooled cone with a small orifice, allowing for the hot plasma gas to enter a depressurizing chamber. In this chamber, rapid cooling, and thus rapid expansion, of the gas occurs. A fraction of this gas then passes through a skimmer cone, and into a chamber that is maintained at a vacuum of that of the MS. This two-step pressure reduction allows the ionic gas to enter the MS at proper temperature and pressure. 6

Mass Spectrometry After passing through the sample and skimmer cones, the ion stream is focused into the quadrupole region by single ion lenses. Ions generated in plasma are nearly all positively charged and have a tendency to repel each other. The ions pass through a charged metallic cylinder which keeps the ion beam from diverging. The Elan DRC-e ICP-MS is equipped with a dynamic reaction cell (DRC). The DRC is located in the vacuum chamber between the lens and the quadruple. Chemical modification of the ion beam to eliminate interferences occurs in the DRC when operating in DRC mode. The type of reaction gas and pressure is set by the user in the computer software. Interference is prevented by interrupting the sequence of reactions that would otherwise create interference. When DRC mode is off, the DRC is a multipole device which transfers the ions to the MS analyzer chamber. Mass spectrometry is used to provide information about the elemental composition of samples of matter; the structures of inorganic, organic, and biological molecules; the qualitative and quantitative composition of complex mixtures; the structure and composition of solid surfaces; and isotopic ratios of atoms in samples. Energetic electrons collide with analyte molecules and impart enough energy to leave the molecules in an excited state. Relaxation then often occurs by fragmentation of part of the molecular ions to produce ions of lower masses. Ions are dispersed in the mass analyzer based on their mass-to-charge ratio (m/z). Mass spectrometers require low pressures in all of the instrument components except the signal processor and readout. The capability of a mass spectrometer to differentiate between masses is usually stated in terms of its resolution which is defined by Equation 1: Equation 1 In the equation, is the mass difference between two adjacent peaks that are just resolved, and m is the nominal mass of the first peak. 7

Figure 6. A quadrupole mass spectrometer (Skoog, Holler, and Crouch, 2007) The quadrupole mass spectrometer (Figure 6) is the most common type of mass spectrometer used in atomic mass spectroscopy. Four parallel cylindrical rods serve as electrodes; opposite rods are connected electrically with one pair connected to the positive side of a variable direct current (dc) source, and the other to the negative terminal. Variable radiofrequency alternating current (ac) voltages 180 out of phase are applied to each pair of rods. Ions are accelerated into the space between the rods by a potential of 5 to 10 V while the ac and dc voltage son the rods are increased simultaneously but with a constant ratio. The quadrupole is a mass filter which only allows ions having a limited range of m/z values reach the transducer. Ions that are not of the correct m/z collide with the rods or exit the path between the rods and are pumped out of the system. Quadrupole instruments easily resolve ions that differ in mass by one unit. The MS in the Elan DRC-e scans the mass range from 5 to 270 amu in milliseconds. Ions that exit the mass analyzer enter the detector. The Elan DRC-e is equipped with a dual mode detector. The ions exiting the quadrupole produce an amplified signal that is processed by the detection electronics and sent to the computer for data processing. The detector is comprised of 26 dynodes. The ions exiting the MS strike the active surface of the first angled dynode. This allows the ions to follow a curved path, separating the signal ions from the background noise (neutral and radiation from the ion source). The dynode releases an electron each time an ion strikes it. The electrons released from the first dynode strike the second dynode and more electrons are released. Cascading of electrons continues until measurable pulse is created. A counting pulse system counts ions that first hit the dynode. If the pulse count exceeds about 2 x 10 6 counts, an isolated dynode is switched to ground potential, disabling the pulse section of the detector. The analog portion of the detector is still available to measure the current produced by the ions hitting the first 23 dynodes. (ELAN DRC-e Hardware guide) 8

ICP MS User Guide Instrument Startup and Daily Performance Check 1. Turn on gas (Argon). 2. Check gas pressure in tank (Argon). The delivery pressure should be at least 60psi. The instrument is always on. 3. Place sample line in beaker with Mili Q water (disconnect from auto sampler if necessary). Place tubing on peristaltic pump. 4. On the computer, open the ELAN icon. 5. File Open Workspace Daily Performance Instrument tab Start (Plasma). The vacuum pressure should be less than 10 5 torr. Check that water is being drawn up the tube. When the ignition sequence is complete, the diagram of the instrument in the window should be green; if not green (e.g. if gas is not in line after changing tank), repeat. 6. The Ready and Plasma lights on the instrument (by the beakers) should be green. 7. Wait 10 15 minutes for the plasma to warm up before performing Daily Performance Check. 8. Fill the beaker labeled daily performance (a quarter to a half) with Daily Performance Check solution (ELAN 6100 Setup Solution). 9. Switch the feed tube from the beaker of MiliQ water to the beaker with Daily Performance Solution. Analyze Sample (Start). 10. When the run is complete, a report will print. To view on screen select Rpt View. In this report: a. Mg must be > 50,000 b. In must be > 250,000 c. Ur must be > 200,000 d. Net Intensity mean of CeO should be 3% (0.03) e. Bkgd Intens. Mean should be 1. 11. If (d) and (e) are not satisfied, wait 5 15 minutes and run again. 12. Fill in values in the Log book (Populate all fields, including your name and the date). Place the printed report in the Log and Daily Performance binder located next to the computer. Preparing Standards and Samples 1. Standards should be prepared for a range of concentrations encompassing expected sample concentrations. 2. A minimum of 5mL of standard or sample should be used for each run. 3. Filter samples with a 0.45 µm filter. 4. 2% by volume of nitric acid (HNO 3 ) should be added to samples and standards. 5. If testing for Bromide or Iodide, 0.1% ammonium hydroxide (NH 4 OH) should be added. 9

Creating a Method 1. Select Method Tab. File New Quantitative Analysis. 2. To select the analytes you wish to quantify: right click in the analyte column Select element from periodic table. You can select multiple isotopes of an element (Ctrl + Select). N.B. If measuring multiple Isotopes, change the names, e.g. Br, Br1 3. The other fields will automatically populate after elements are selected. 4. Save the method. Use your initials and the date; may include elements analyzed. Making a Sample List 1. In the Sample Tab, File Open or Create New. Select Batch Mode (not Manual). 2. Start from Sample location 11. Location 11 is the 1 st position in the auto sampler tray. Locations 1 to 10 run horizontally on the autosampler. 3. Input sample names in the Sample I.D column. Sample list can be created in Excel and copied. 4. Select Measurement Action Run Sample 5. Click Analyze Batch. In the pop up window, Method Load Existing. Running Samples 1. Check pressure in gas tank, rinse solution level and waste collection container. 2. Sign Log book. 3. Run Daily Performance Check. 4. Connect sample tube to auto sampler tube. 5. Place samples in auto sampler tray and create sample list. 6. Select samples to run in the sample list (highlight rows). 7. Click Summary Summarize by Analyte Output to File Browse. Name the file to be created Save OK. Build Run List Autostop. 8. Check that Sample info is correct. Select Analyze Batch. 9. If prompted Are you sure you want to use Auto Stop? Select Yes. 10. Start sample run. 11. Remove your samples from the tray after run. Disconnect tubing from the peristaltic pump and leave lying flat (to prevent wear of the tubes). 12. To access data: C:\Elandata\ReportOutput 10

Works Cited ELAN DRC-e Hardware guide. (n.d.). PerkinElmer SCIEX. Skoog, D. A., Holler, F. J., & Crouch, S. R. (2007). Principles of Instumental Analysis. Thomson Higher Education. 11