Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Similar documents
Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Trace Analyses in Metal Matrices Using the ELAN DRC II

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Automated Determination of PPQ Levels of Thorium in High Purity Copper

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

The 30-Minute Guide to ICP-MS

Analysis Repeatability of Trace and Major Elements in a Water Sample

OES - Optical Emission Spectrometer 2000

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

ICP-OES Application Note Number 35

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Fergus Keenan Shona McSheehy Julian Wills

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Sample Analysis Design PART II

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

Contamination Monitoring of Semiconductor Processes by VPD HR-ICPMS

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Samples compliant with the WATER

PRINCIPLE OF ICP- AES

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

Florida State University Libraries

Outsmart Interferences with Triple Quadrupole ICP-MS. The world leader in serving science

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

ICP-MS Analysis of Bulk & Process Chemicals for Semiconductor Processes. Ann O Connell Chemical Analysis Eng Intel Corporation

Sample Analysis Design PART III

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Characterization of Catalysts and Surfaces. Elemental Analysis (ICP, AAS etc.) Fall Semester 2016 Bodo Hattendorf HCI G105

Analysis of noble metals at low levels in geological reference materials and ores

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Current State of Extraction Don t Be Deceived! Sharon F. Webb, Ph.D. Director of Quality Program

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Concentration Measurements by Single Particle ICP-MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Sources of Errors in Trace Element and Speciation Analysis

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

materials and their properties

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

ANY MATRIX ANY PARTICLE SIZE ANY INTERFERENCE. NexION 2000 ICP Mass Spectrometer

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

ICP/MS Multi-Element Standards

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Transcription:

APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity levels in silicon-based semiconductor devices is critical because even ultratrace amounts of impurities, including alkali and alkali-earth elements and transition metals, can cause defects, such as voltage breakdown or high dark current. For quality control purposes, there are two types of silicon that are routinely analyzed: bulk silicon and the surface of silicon wafers. Bulk silicon analysis can be performed by totally digesting the silicon using a very aggressive acid, such as hydrofluoric acid (HF). Vapor phase decomposition is the most common method used for the surface analysis of silicon wafers and involves collecting impurities on the wafer surface using a very small amount of acid (typically HF) deposited on the surface as a droplet. This results in a typical sample volume of around 200 μl. For bulk silicon analysis, sample volume is not an issue; however, small sample volumes are desirable in order to minimize time-consuming sample preparation. As such, both types of silicon analyses require the ability to handle small sample volumes and high silicon matrices, as well as an HF-resistant sample introduction system. Since a typical analysis may take 2-3 minutes per sample, low-flow nebulizers with sample uptake rates from 20-100 μl/min are routinely used.

Table 1. Instrumental parameters and sample introduction components for the NexION 300S ICP-MS. Spray Chamber: PFA Scott Nebulizer: 20 µl/min PFA-ST Torch: High Efficiency Quartz Plasma Gas: 18 L/min Torch Injector: PFA-Platinum Auxiliary Gas: 1.1 L/min Sampler Cone: Platinum RF Power: 1600 W Skimmer Cone: Platinum Integration Time: 1 sec/mass Adding to the complexity of the analysis for silicon impurities is the fact that many of the critical analytes are difficult to analyze by inductively coupled plasma mass spectrometry (ICP-MS) because they suffer from plasmabased molecular and isobaric interferences, such as ArO +, ArH +, and Ar +. By providing a low flow of the proper reaction gas into the Universal Cell and using the unique Dynamic Bandpass Tuning (DBT) feature, both part of PerkinElmer s NexION 300 ICP-MS, interferences can be chemically removed from the ion beam before they enter the analyzer quadrupole of the mass spectrometer. Another advantage of the NexION 300 ICP-MS is that it always operates under robust, hot plasma conditions, effectively decomposing the sample matrix. The NexION 300 ICP-MS also has the ability to combine elements run in Reaction mode (with reaction gas) with elements run in Standard mode (without a reaction gas) in a single analytical run, eliminating the need for running the sample twice or under two different plasma conditions. The results obtained by both modes (Reaction and Standard) are combined by the instrument s software and printed out in a single report. This application note demonstrates the ability of the NexION 300S ICP-MS to analyze impurites in low-volume bulk silicon samples using a low-flow nebulizer. Experimental Conditions Sample Preparation: A bulk silicon sample was digested with a small amount of concentrated high-purity HF and HNO 3 acids. Typically, 0.1 g of crushed silica wafer is weighed into a clean, dry PTFE bottle, followed by the addition of 1 ml of HF and 0.5 ml of HNO 3. The addition of HNO 3 should be done with small, successive aliquots because of the high reactivity involved. No heating is necessary; the addition of HF and HNO 3 will dissolve the silicon. Several aliquots of the digested sample were then diluted to obtain samples containing 100, 500, 1000, 2000, and 5000 ppm of Si. The final acid concentration in each sample was adjusted to contain 2% HF and 1.5% HNO 3. Two samples were prepared for each Si concentration, and one of each pair was spiked with a standard solution (PerkinElmer Pure, PerkinElmer, Shelton, CT, USA) in order to perform spike recovery tests. Calibration solutions in 2% HF and 1.5% HNO 3 (Tama Chemicals, Tokyo, Japan) were prepared from multi-element solutions (PerkinElmer Pure, PerkinElmer, Shelton, CT, USA). Instrumental Conditions: The instrument used was the PerkinElmer NexON 300S ICP-MS. The instrument operating conditions and sample introduction system used for this work are shown in Table 1. The elements determined and instrument modes used are shown in Table 2. Table 2. Analytes and the measurement modes. Analytes Measured Mode B, Na, Mg, Cu, Mo and Cd Standard mode (without reaction gas) Al, K, Ca, V, Cr, Mn, Fe, Ni, Co and Zn Reaction mode (with reaction gas) 2

Results Since silicon is a refractory element, it tends to form oxides in the plasma, particularly when cool plasma conditions are used. These silicon oxides deposit on the surface of the interface cones, causing significant signal drift. The ability of the NexION 300S ICP-MS to use hot plasma conditions for all analyses greatly reduces this signal drift. In order to show the benefit of using the more robust hot plasma conditions, a 2000-ppm silicon sample spiked with a 100-ppt multielement standard solution was continuously introduced into the NexION 300S without washing for two hours, while taking readings every 5 minutes. As shown in Figures 1 and 2, the signal for elements in Standard and Reaction modes (in a single method) were very stable even during constant nebulization of the sample solution. Figure 3. Ca calibration, with NH3 cell gas flow of 1 ml/min. For quantitative measurements, additions calibrations with 2000 ppm of silica in 2% HF and 1.5% HNO3 solution were used to determine the concentration of each analyte in the various silicon samples. Calibration curves for Ca, Fe and Co are shown in Figures 3 5, respectively. Figure 4. Fe calibration, with NH3 cell gas flow of 0.6 ml/min. Figure 1. Continuous analysis of 100 ppt multielements spike in 2000-ppm silica solution, first group. Figure 5. Co calibration, with NH3 cell gas flow of 0.3 ml/min. Figure 2. Continuous analysis of 100 ppt multielements spike in 2000-ppm silica solution, second group. 3

It can be deduced from the calibration plots that the background level of the NexION 300S system is not affected by the plasma temperature, and as a result, the Ca, Fe and Co background equivalent concentration (BEC) levels were only a few parts per trillion (ppt), with a sample uptake rate at 20 μl/min. The benefit of using hot plasma conditions for this particular sample matrix is the accelerated decomposition of polyatomic ions such as CaF 2, which are formed in the HF matrix. This species cannot be decomposed under cool plasma conditions, typically causing low sensitivity for Ca. However, using hot plasma conditions in the NexION 300S, this species is effectively decomposed, and the Ca sensitivity is maintained. Furthermore, the use of Reaction mode in the Universal Cell eliminated the ArF + polyatomic interference on Co, allowing low-levels to be measured. A spike recovery test was carried out to determine the level of silicon in a sample that could be analyzed without significant matrix suppression. The recovery results for the samples containing different concentrations of Si with a 100 ppt multielement spike are summarized in Figure 6. This test shows that samples containing up to 5000 ppm Si can be analyzed against an additions calibration curve containing 2000 ppm of Si. Also evident from this test is that a signal suppression of less than 20% occurs, which is excellent for this matrix type. Figure 6. Spike recovery tests of 100 ppt multielement standard in silica matrix ranging from 100 ppm to 5000 ppm. A separate spike recovery test of a 50-ppt multielement spike into 2000-ppm silicon was carried out. The results are shown in Table 3 and indicate that all of the elements show greater than 90% spike recovery, which is excellent for this matrix. Table 3 also shows detection limits, which were calculated using three times the standard deviation of the 2000-ppm Si in 2% HF/1.5% HNO 3 blank solution. Since matrix-matched calibration curves are not required to analyze Si matrix samples, the detection limits achieved on the NexION 300S ICP-MS are independent of the concentration of Si in the solution. Table 3. Detection limits (DLs) and 50 ng/l spike recoveries for all analytes in 2000-ppm Si solution (units in ng/l). Cell Gas Flow* DL 50 ppt Analyte Mass (ml/min) RPq (ppt) Recovery Li 7 0 0.25 2.6 110% Be 9 0 0.25 5 101% B 11 0 0.25 5.8 97% Na 23 0 0.25 2 106% Mg 24 0 0.25 1.9 111% Al 27 0.6 0.5 4 107% K 39 0.6 0.5 2 107% Ca 40 1 0.5 1.7 109% V 51 0.3 0.5 0.9 93% Cr 52 0.3 0.5 3 97% Mn 55 0.6 0.7 0.7 105% Fe 56 0.6 0.5 2.8 103% Co 59 0.3 0.5 0.8 99% Ni 60 0.3 0.7 3 96% Cu 63 0.3 0.5 3.5 98% Zn 64 0.3 0.65 4.2 105% Ga 71 0.6 0.5 3 97% Ge 74 0.3 0.65 4 110% As 75 0 0.25 10 96% Sr 88 0.6 0.5 1.5 103% Zr 90 0 0.25 2.8 101% Nb 93 0 0.25 1.8 103% Mo 98 0 0.25 2 102% Ag 107 0 0.25 1 115% Cd 114 0 0.25 1 110% In 115 0 0.25 1.6 109% Sn 120 0 0.25 2 110% Sb 121 0 0.25 2 110% Ba 138 0 0.25 1.6 104% Ta 181 0 0.25 2 99% W 184 0 0.25 2 99% Au 197 0 0.25 3 98% Tl 205 0 0.25 1.8 100% Pb 208 0 0.25 1.7 103% Bi 209 0 0.25 2 102% U 238 0 0.25 2.8 100% *Cell gas used is NH 3. 4

Conclusion The data presented in this application note show that the NexION 300S ICP-MS can effectively eliminate the 40 Ar + interference on 40 Ca +, the 40 Ar 19 F + interference on 59 Co +, and the 40 Ar 16 O + interference on 56 Fe +, as well as other common, troublesome interferences in ICP-MS using ammonia as the reaction gas. By adjusting the dynamic bandpass tuning parameters to eliminate unwanted reaction byproducts and combining Reaction mode and Standard mode elements in the same analytical method, multiple elements can be determined in the same sample analysis, increasing laboratory productivity. In addition, spike recovery tests demonstrate the ability of the NexION 300S ICP-MS to perform these analyses utilizing robust hot plasma conditions, significantly reducing matrix suppression effects with silica matrix up to 2000 ppm. This means that the NexION 300S ICP-MS can perform silicon analysis with minimal matrix suppression using a simple external calibration curve. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2012-2014, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010281A_01