A study of silicon sensor for ILD ECAL

Similar documents
arxiv: v3 [physics.ins-det] 5 Mar 2018

Digital Calorimetry for Future Linear Colliders. Tony Price University of Birmingham University of Birmingham PPE Seminar 13 th November 2013

Status of the CALICE analog calorimeter technological prototypes

First Stage Analysis of the Energy response and resolution of the Scintillator ECAL in the Beam Test at FNAL, 2008

Status of Simulation Tools for ScECAL of ILD

Development of New MicroStrip Gas Chambers for X-ray Applications

First Stage Analysis of the Energy Response and Resolution of the Scintillator ECAL in the Beam Test at FNAL, 2008

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

arxiv: v2 [physics.ins-det] 8 Feb 2013

arxiv: v1 [physics.ins-det] 30 Mar 2014

Technological Prototypes and Result Highlights of Highly Granular Calorimeters

Particle Flow Algorithms

A gas-filled calorimeter for high intensity beam environments

PERFORMANCE OF THE ATLAS LIQUID ARGON FORWARD CALORIMETER IN BEAM TESTS

arxiv: v1 [hep-ex] 18 Feb 2009

PoS(PD07)020. Timing Properties of MCP-PMT. Kenji Inami. Nagoya university, Nagoya, Japan

FoCal Project in ALICE. Yota Kawamura for the ALICE FoCal collaboration TCHoU workshop 2018/3/15

X-ray induced radiation damage in segmented p + n silicon sensors

CLICdp work plan and foreseen documents in preparation for the next European Strategy Update

arxiv: v1 [physics.ins-det] 23 Oct 2017

Components of a generic collider detector

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014

The Hermes Recoil Silicon Detector

Electromagnetic calorimetry for the ILC

Performance of the MCP-PMT for the Belle II TOP counter

Status Report: Charge Cloud Explosion

Fast Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon PIN Diodes and APDs

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Capacitance of the GLAST Prototype Detectors

First Results and Realization Status of a Proton Computed Radiography Device

Functions Represent SiPM Response Especially Linear Behavior After Saturation

The Silicon-Tungsten Tracker of the DAMPE Mission

CALICE Test Beam Data and Hadronic Shower Models

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

CMS (ECAL) precision timing

Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon APDs and PIN Diodes

The CMS ECAL Laser Monitoring System

Lecture 18. New gas detectors Solid state trackers

Edgeless sensors for full-field X-ray imaging

Characterisation of Silicon Photomultipliers for the T2K Experiment

G. Gaudio, M. Livan The Art of Calorimetry Lecture V. The state of art Towards ILC calorimetry

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

Digital Hadron Calorimetry for the Linear Collider using GEM Technology

CEPC Detector and Physics Studies

PoS(TIPP2014)093. Performance study of the TOP counter with the 2 GeV/c positron beam at LEPS. K. Matsuoka, For the Belle II PID group

Test setup, APDs,, preamps Calibration procedure Gain monitoring with LED Beam test results Future R&D options

Particle ID in ILD. Masakazu Kurata, KEK Calorimeter Workshop IAS program 01/19/2018

What detectors measure

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

THE main physics motivation for building the Compact

Design of the new ATLAS Inner Tracker for the High Luminosity LHC era

7th International Conference on New Developments In Photodetection

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

Calorimeter test-beam results with APDs

Search for Displaced Supersymmetry using the Compact Muon Solenoid Detector

CALICE Si-W EM Calorimeter: Preliminary Results of the Testbeams 2006

Triple GEM Detectors for the Forward Tracker in STAR

Hadronic energy reconstruction in the combined electromagnetic and hadronic calorimeter system of the CALICE Collaboration

Silicon Detectors. Particle Physics

ILD and the UK. This talk: ILD and the UK context Relation to CALICE/LCFI LDC ILD GLD ILD-UK Detector Optimisation Summary BaaQuack

Jean Claude BRIENT Laboratoire Leprince Ringuet CNRS IN2P3 / Ecole polytechnique. Pavia CALOR08 J. C.Brient (LLR CNRS/Ecole polytechnique) 1

ATLAS Hadronic Calorimeters 101

PoS(BORMIO2016)013. PANDA Forward Spectrometer Calorimeter. P.A. Semenov IHEP NRC KI S.I. Bukreeva IHEP NRC KI ITEP NRC KI

Development of Radiation Hard Si Detectors

Aspects of radiation hardness for silicon microstrip detectors

Module of Silicon Photomultipliers as a single photon detector of Cherenkov photons

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

Search for Weakly Interacting Undiscovered Particles using Sub-THz Gyrotron

Particles and Universe: Particle detectors

Experimental study of nonlinear laser-beam Thomson scattering

PERFORMANCE GOALS AND DESIGN CONSIDERATIONS FOR A LINEAR COLLIDER CALORIMETER*

ATLAS EXPERIMENT : HOW THE DATA FLOWS. (Trigger, Computing, and Data Analysis)

Challenges for Silicon Pixel Sensors at the XFEL. Table of Content

Top Quark Precision Physics at Linear Colliders

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor

Monika Mittal Shanghai Jiao Tong University, China

Construction of the CALICE High Granular Scintillator Based Electromagnetic Calorimeter Prototype and Its Response to Electrons

pp physics, RWTH, WS 2003/04, T.Hebbeker

Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode

Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)

Neutron irradiation test of Hamamatsu, SensL and AdvanSiD UV-extended SiPMs

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM

Adam Para, Fermilab CALOR2010, IHEP, Beijing May 14, 2010 TOTAL ABSORPTION HOMOGENEOUS CALORIMETER WITH DUAL READOUT

Development of a new MeV gamma-ray camera

The ATLAS Silicon Microstrip Tracker

Calorimeter for detection of the high-energy photons

arxiv:physics/ v2 [physics.ins-det] 18 Jul 2000

arxiv: v1 [physics.ins-det] 18 Feb 2009

arxiv: v2 [nucl-ex] 6 Oct 2008

Reconstruction in Collider Experiments (Part IV)

Characterization of Irradiated Doping Profiles. Wolfgang Treberspurg, Thomas Bergauer, Marko Dragicevic, Manfred Krammer, Manfred Valentan

8 Calibration and Monitoring

M. Diemoz INFN Sezione di Roma, P.le A. Moro 2, Roma, Italy

Dual readout with tiles for calorimetry.

Particles and Universe: Particle detectors

Development of semiconductor imaging detectors for a Si/CdTe Compton camera

Digital Imaging Calorimetry for Precision Electromagnetic and Hadronic Interaction Measurements

REVISED HIGHER PHYSICS REVISION BOOKLET ELECTRONS AND ENERGY

Non-collision Background Monitoring Using the Semi-Conductor Tracker of ATLAS at LHC

Geant4 simulations of the lead fluoride calorimeter

Transcription:

A study of silicon sensor for ILD ECAL arxiv:1403.7953v1 [physics.ins-det] 31 Mar 2014 Tatsuhiko Tomita Shion Chen, Daniel Jeans, Yoshio Kamiya, Kiyotomo Kawagoe, Sachio Komamiya, Chihiro Kozakai, Yohei Miyazaki, Taikan Suehara, Yuji Sudo, Hiraku Ueno, Tamaki Yoshioka Abstract The International Large Detector (ILD) is a proposed detector for the International Linear Collider (ILC) [1, 2]. It has been designed to achieve an excellent jet energy resolution by using Particle Flow Algorithms (PFA) [3], which rely on the ability to separate nearby particles within jets. PFA requires calorimeters with high granularity. The ILD Electromagnetic Calorimeter (ECAL) is a sampling calorimeter with thirty tungsten absorber layers. The total thickness of this ECAL is about 24 X 0, and it has between 10 and 100 million channels to make high granularity. Silicon sensors are a candidate technology for the sensitive layers of this ECAL. Present prototypes of these sensors have 256 5.5 5.5 mm 2 pixels in an area of 9 9cm 2.We have measured various properties of these prototype sensors: the leakage current, capacitance, and full depletion voltage. We have also examined the response to an infrared laser to understand the sensor s response at its edge and between pixel readout pads, as well the effect of different guard ring designs. In this paper, we show results from these measurements and discuss future works. tomita@epp.phys.kyushu-u.ac.jp Kyushu University The University of Tokyo International Center for Elementary Particle Physics, The University of Tokyo Talk presented at the International Workshop on Future Linear Colliders (LCWS13), Tokyo, Japan, 11-15 November 2013.

1 Introduction The International Linear Collider (ILC) is a future high energy lepton collider [1]. It can make precision measurements of the Higgs boson, top and electroweak physics. ILC has a capability of exploring new physics as well. To realize these measurements, we should prepare a detector with excellent precision [2]. In the case of ILC, many events will include jets in their final state. For precise measurements, these jets should be reconstructed as well as possible. We plan to use Particle Flow Algorithm (PFA) [3] to achieve Jet Energy Resolution (JER) 3-4%. To make full use of PFA, we should separate each shower made from individual particles in jets. Thus, we need high granular calorimeter especially in ECAL. In this paper, we show basic properties of silicon sensors which should be measured for quality control, and we also report the response to an infrared laser. 2 Overview of silicon sensors for ILD ECAL The silicon sensor chips are produced by Hamamatsu (HPK). The sensors are a kind of silicon PIN diode, and they have 256 5.5 5.5mm 2 pixels in an area of 9 9cm 2 (Figure 1). The thickness of sensors is 320 µm. The guard ring can collect surface current of the chips, but it also limit sensitive region of chips. Furthermore, from a former test beam result, 1 guard ring structure makes fake signals along it when particles come into near guard ring. In this way, guard ring has advantage and disadvantage, thus we have to decide which design should be used for ILD ECAL. Figure 1: Silicon sensor: each chip has 256 channels. Figure 2: Guard ring structures: the guard ring width is 80 µm (1 guard ring), or 20 µm (2, 4 guard rings).

There are four types of guard ring structure. Currently, we have only 1 guard ring structure in this size. The other structures are provided by smaller chips 3 3 pixels (4 guard rings), or 4 4 pixels (no guard ring, 2 guard rings). We measured only 1 guard ring type chip, but in the near future we will measure the other chip and compare the properties of each structure. 3 Basic properties: leakage current and capacitance We made two setups for the measurements of basic properties. One of the them is for the measurement of leakage current, the other is for capacitance measurement (Figure 3). Figure 3: Setup: the left one is for leakage current, the other one is for capacitance. Figure 4: Setup box: copper plate combine 256 channels to 1 channel. Figure 5: Setup for the measurements: source/lcr meter is controlled from a PC.

We can control temperature and humidity of dark box, thus we are able to assume these two dependence are negligible in both measurements. In our measurements, we set temperature at 27.5 C (or 20.0 C for capacitance measurement) and humidity at 50%. During both measurements, we used copper plate to combine signals from all sensor cells channel at one readout channel (see Figure 4). Between sensor and read out circuit are connected by using springs. The operation and data taking were controlled from a PC (see Figure 5). For leakage current measurement, we took data from 0 V to 800 V in 10 V step. Each step has 1 second, and if the chip reaches to breakdown voltage we stop the measurement to prevent the chip from getting damaged. As shown in Figure 6, the leakage current is about 100 na with applied bias voltage of 100-150 V, and its voltage dependence is very small in this region. We repeated this procedure 10-20 times on each chip, and found that the breakdown voltage becomes higher than the previous scan at first 3-4 scans. This phenomenon needs to be understood. However, we can conclude the leakage current is quite small and stable at the operation voltage of 100-150 V. If a chip has bad channel of high leakage current, we expect that the chip reaches breakdown on much smaller voltage. Therefore, this measurement is also important to test the chips before assembly, especially in mass production phase. Figure 6: Leakage current of the chip (0-800 V) under the situation of 27.5 C. Horizontal axis shows applied voltage, and vertical axis shows the total leakage current of silicon sensor. 47 scans of the same chip are overlaid. Figure 7: Capacitance curve: the horizontal axis shows applied voltage, and vertical axis shows 1/capacitance 2. This measurement did in the situation of 20 C and 50 % of humidity.

For capacitance measurement, we applied bias voltage from 0.1 V to 130 V by step of 0.1 V. Figure 7 shows good linearity in 5 V to 45 V region. From 0 V to 5 V, the capacitance curve is slightly winding because of its Metal-Oxide Semiconductor structure [4]. Shown in Figure 7, the capacitance saturates at around 60 V, which is consistent of the full depletion voltage of 65 V in the specification. We also estimated that the chip thickness is 318µm. This value is also consistent with the specification. 4 Response to the laser To understand the difference between different guard ring structures, we prepared a laser system. We use an infrared laser (class-3b) whose wavelength is 1064 nm. Photons of this wavelength can pass through the surface of silicon with high transmission to produce a pair of electron-hole in the active area. Since infrared photons cannot go through electrode, we shoot the laser into the gap (between a pixel and a pixel, or a pixel and a guard ring). At first, we measured the response from a specific pixel to check our system. We used an oscilloscope to see waveform, and we also check the fluctuation of peak value by using the waveform. The waveform taken at 100 V bias voltage is shown in Figure 8. The decay tail was caused by the time constant of the pre-amplifier. We obtained the peak fluctuation value of 3%, shown in Figure 9, which corresponds to the stability of our measurement. Figure 8: Wave form: the horizontal axis shows time (µs), vertical axis shows output value (mv) time constant of pre-amplifier is about 150 µs. This measurement ran under applied voltage of 100 V. Figure 9: Peak fluctuation: the horizontal axis shows peak value of response, and vertical axis shows the number of counts. The measurement did in 100 V applied voltage.

Figure 10: The voltage dependence of peak value: The horizontal axis shows applied voltage, and vertical axis shows peak value. The error bar comes from the fluctuation of peak value (shown in Figure 9). We also investigated the response for various bias voltages. The result is shown in Figure 10. In this figure, the response saturates at 80 V, which indicates that the chip reached full depletion. Since this saturation is not consistent to the full depletion voltage of 65 V (obtained from previous measurements), we are now investigating the reason of the difference. 5 Summary We established a silicon sensor test system. We can measure basic properties of silicon sensors such as leakage current and capacitance. Our first result from these measurements meet the specification. We also investigated the response to an infrared laser. We have a small disagreement for the full depletion voltage between the capacitance measurement and the laser measurement. Now we are trying to understand this issue. Our final goal is to make a decision on the chip design including guard ring structures, thus we will compare these properties on prototypes of various structures in the near future. References [1] T,Behnke, James E. Brau, B. Foster, J. Fuster, M. Harrison, et al, The International Linear Collider Technical Design Report - Volume 1: Executive Summary. 2013. [2] T,Behnke, James E. Brau, Philip N. Burrows, J. Fuster, M. Peskin, et al. The International Linear Collider Technical Design Report - Volume 4: Detectors. 2013.

[3] M.A. Thomson. Particle Flow Calorimetry and the PandoraPFA Algorithm. Nucl. Instrum. and Meth. A611 (2009) 25-40. [4] E.Fretwurst, H.Herdan, G.Lindstroem, U.Pein, M.Rollwagen, H.Schatz, P.Thomsen, R.Wunstorf, Nucl. Instr. and Meth. A288 (1990) 1-5.