RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

Similar documents
PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

A bilateral comparison of a 1 kg platinum standard

Report on EURAMET.M.M-S9

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

PROTOCOL OF MASS AND VOLUME COMPARISONS BETWEEN SIM NMIs

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

INTERLABORATORY MASS COMPARISON BETWEEN LABORATORIES BELONGING TO SIM SUB-REGIONS COORDINATED BY CENAM (SIM.7.31a & SIM.7.31b)

MASS AND VOLUME COMPARISONS AT MIKES

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

INTERLABORATORY COMPARISON BETWEEN TUBITAK-UME TURKEY AND INM ROMANIA IN THE FIELD OF FORCE MEASUREMENTS

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

ASIA PACIFIC METROLOGY PROGRAME

Measurement Uncertainty in Mechanical Testing

ANNEXE 8. Technical protocols for the interlaboratory comparisons

STATISTIQUE DE PERFORMANCE

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Traceable Mass Determination and Uncertainty Calculation

Precision estimated by series of analysis ISO and Approach Duplicate Approach

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Analysis of interlaboratory comparison when the measurements are not normally distributed

Flowchart for the extraordinary calibrations prior to the redefinition of the kilogram. (BIPM proposal)

Measurement uncertainty and legal limits in analytical measurements

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

Comparison protocol EURAMET project

EURAMET Project no Inter-comparison of a 1000 L proving tank

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

Interlaboratory Comparison & Proficiency Testing Program

SIM SIM.M.D-K3. provide. SIM.M.M-K5 for. A set of. Institute LACOMET LATU INTI. Country Costa Rica Uruguay Argentina Chile México Canada Brazil

Appendix B1. Reports of SMU

Interlaboratory Comparison & Proficiency Testing Program

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Mason Technology Ltd

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

ESTA governance Quality Assurance System for Seed Treatment and Treated Seed

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

Progress in organizing national and international comparisons for nuclear medicine measurements

Analysis of a national low dc resistance inter-laboratories comparison

TEST REPORT IEC and/or EN Photobiological safety of lamps and lamp systems

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

ISO INTERNATIONAL STANDARD. Piston-operated volumetric apparatus Part 7: Non-gravimetric methods for the assessment of equipment performance

SIM FORCE STANDARDS COMPARISON UP TO 10 kn

A basic introduction to reference materials. POPs Strategy

Interlaboratory Comparison & Proficiency Testing Program

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011

EUROMET Project 702 EUROMET.M.D-K4

Mason Technology Ltd

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards

Objective: To enable the students to state and differentiate important concepts and terms that are useful to carry out Experiment 1 meaningfully.

R SANAS Page 1 of 7

Participants in the Proficiency Test THM 02/2016

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

CCQM-K45: Sn in tomato paste key comparison

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

THE UNCERTAINTY OF MEASUREMENT IN CALIBRATION USING A COMPARISON FORCE STANDARD MACHINE

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Calibration of Surface Contamination Monitors. UKAS Accredited Organisations Comparison. Final Report

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

Report on the 2011 Proficiency Test for the determination of T-2 and HT-2 toxins in wheat flour

Document No: TR 12 Issue No: 1

Centre for Metrology and Accreditation, MIKES

TEST REPORT IEC and/or EN Photobiological safety of lamps and lamp systems

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY

Final Report August 2010

SOP 21. Applying air buoyancy corrections. 1. Scope and field of application. 2. Principle. 3. Requirements

Calibration and verification: Two procedures having comparable objectives and results

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

CCM short note on the dissemination process after the proposed redefinition of the kilogram

PROGRESS IN ORGANIZING NATIONAL AND INTERNATIONAL COMPARISONS FOR NUCLEAR MEDICINE MEASUREMENTS

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

Physikalisch- Technische Bundesanstalt

Supplementary Comparison EUROMET.EM-S23 EURAMET PROJECT No 815 FINAL REPORT

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Photobiological safety of lamps and lamp systems

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Comparison of V and 10 V DC Voltage References

ISO INTERNATIONAL STANDARD. Statistical methods for use in proficiency testing by interlaboratory comparisons

Portuguese Institute for Quality. Contents

BILATERAL COMPARISON REPORT LNE KIM-LIPI

Report on CCM Pilot Study CCM.R-kg-P1. Comparison of future realizations of the kilogram. Final Report

International interlaboratory study of forensic ethanol standards

CERTIFICATE OF CALIBRATION

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

International Atomic Energy Agency

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Transcription:

The Fifth International Proficiency Testing Conference Timisoara, Romania () th 8 th September, 0 RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS Adriana Vâlcu, Sterică Baicu National Institute of Metrology, Bucharest, Romania adriana.valcu@inm.ro; sterica.baicu@inm.ro Abstract ISO/IEC 70, [], requires that laboratories have quality control procedures for monitoring the validity of tests and calibrations undertaken. This monitoring may include the participation in interlaboratory comparisons (ILC) or proficiency testing (PT) programs. By interlaboratory comparisons it is determined and continuously monitored the individual performances of laboratories that perform calibrations, thus identifying potential problems for which corrective actions are initiated. In Romania, at an interval of two years, the Romanian Bureau of legal Metrology (BRML) organizes interlaboratory comparisons in different areas. For mass field, an ILC of mass standards was part of interlaboratory comparison program for 03-0. Six standard weights (nominal values: 00 g, 0 g, 0 g, g, 00 mg and 0 mg) used as travelling standards, were circulated in two separate loops between twenty two BRML laboratories, mass Laboratory of National Institute of Metrology (INM) acting as reference and pilot laboratory. The paper presents in tabular and graphical form: laboratory s mass results, uncertainty declared by participants for each weight, the normalized errors obtained for each laboratory, with respect to the INM. Analysis of the performed results belonging to participating laboratories shows that about % of the total reported results are discrepant. Key words Interlaboratory comparison (ILC), normalized error (En), measurement uncertainty. INTRODUCTION In Romania, at an interval of two years, the Romanian Bureau of legal Metrology (BRML) organizes interlaboratory comparisons in different areas.

For mass field, an ILC of mass standards was part of interlaboratory comparison program for 03-0. Six standard weights (nominal values: 00 g, 0 g, 0 g, g, 00 mg and 0 mg) used as travelling standards, were circulated in two separate loops between twenty two BRML laboratories, mass Laboratory of National Institute of Metrology acting as reference and pilot laboratory. The Weights were treated as F accuracy class, even if, for weights of 00 mg and 0 mg the condition.3. from [] was not met. A preparatory document - technical protocol - (containing measurement instructions, transport and use) was drawn up by INM to inform the participating laboratories with requirements of interlaboratory comparison. CIRCULATION SCHEME, TIME SCHEDULE Interlaboratory comparison was organized using a circular scheme, with two petals, Figure. Including the RL, the first petal covered eleven laboratories whereas the second one covered twelve. Fig. Circulation scheme Initial planning of the period for performing measurements was from June 03 to May 0. Subsequently, in the second petal, laboratories no. and requested a delay for completing the measurements. For this reason the last calibration performed by the reference laboratory was at the end of July 0. Each participant had approximately weeks to carry out the measurements. 3 REFERENCE VALUES The weights used as itinerant standards were selected by Mass Laboratory and belong to measuring range between 00 g and 0,0 g as follows: 00 g, 0 g, 0 g, g, 00 mg and 0 mg. Specifications for the artifacts, in accordance with [], are shown in Table.

Table. Specifications for artifacts Parameter Nominal Value Weight 00 g 0 g 0 g g 0, g 0,0g Accuracy class F F F F F F Material brass brass brass brass nickel silver nickel silver Density (kg/m 3 ) 800 800 800 800 800 800 u ρ (kg/m 3 ) 8 8 8 8 8 8 The reference values x ref and U ref for itinerant standards were determined by reference laboratory at three different times: at the beginning (June 03), at the middle (November 03) and at the end of interlaboratory comparison (July 0). For each itinerant standard was calculated the average of the three values obtained. MEASUREMENT INSTRUCTIONS Before starting the intercomparison, the participants received information concerning the nominal value of the weights as well their estimated density. Each participating laboratory determined the conventional mass of the transfer standards and their associated uncertainty. All weighing were performed in air. Measurements were carried out after adequate acclimatization at the mass laboratory, as described in [] for weights of accuracy class F. In addition, instructions were given how to handle, store and transport the weights. For confidentiality purposes, for each laboratory it was assigned a code number ( to 3), the reference laboratory having the code. No other detailed calibration instructions were given to the laboratories. Regarding the method of calibration, laboratories have opted for method of substitution (Borda), using one or more standards. In order to not influence the mass value of the weights during the intercomparison, the participants were informed to not wash the weights, the only accepted method of cleaning being the use of a brush that accompanied the travelling standards. REPORTING RESULTS BY PARTICIPANTS Each participating laboratory determined the conventional mass and associated uncertainty for each itinerant standard. The measurement results were sent to the reference/pilot laboratory in a report where a list of the equipment used for comparison, method used, laboratory environmental conditions, condition of the transfer standard surface and uncertainty budget were included. Also, details on the standards used were required, in order to prove traceability of laboratory s measurement results. The participants were requested to: - specify the uncertainty budget in sufficient detail including: weighing process, reference weight, air buoyancy correction and mass comparator; - report the final results in a calibration certificate. RESULTS OF INTERCOMPARISON. Generalities 3

Laboratories presented the results in accordance with the requirements of the technical protocol that accompanied itinerant standards, with the following comments: - some of the participants (,, 8 and ) did not present the final results in a calibration certificate; - laboratory no. has not reported the result for 0 mg weight in the calibration certificate; - laboratories no. and have not sent the primary data.. Presentation of results The measurements results obtained in the interlaboratory comparison are summarized in Table. The results are presented exactly in the form that was sent by the participants. In this table, the first three rows represent the results obtained by the reference laboratory at all the three different times: at the beginning, at the middle and at the end of interlaboratory comparison. The fourth row represents the average of these three results. Table. Deviation (D) from nominal mass and expanded uncertainty (U) for each weight reported by the participants The criterion used for evaluating the performance of laboratories was the normalized error E n, which takes into account both the result and its uncertainty. The normalized error E n is given as [3]: E n U x lab lab x ref U ref Where:

E n x lab x ref U ref U lab = normalized error; = result of measurement carried out by a participant laboratory; = the comparison reference value; = measurement uncertainty reported by reference laboratory; = measurement uncertainty reported by participant laboratory. In the calculation of E n values, the reference values were obtained from the arithmetic mean of the measurements performed at three different times, according to Table. The E n data for each laboratory is presented in the Table 3 and provides supplemental information concerning the measurement capability of the participating laboratories []. Table 3: Normalized deviations E n from reference values Nominal Value 00 g 0 g 0 g g 00 mg 0 mg X ref,033 0,77 0,00 0,003-0,0-0,03 U ref 0,300 0,00 0,00 0,030 0,0 0,008 Laboratory s code Normalized deviation E n -0,7-0, -0,33 0,3-0, -0, 3-0,3 0,0-0,0 0, 0,83 0,0-0, 0,07-0, -0,,0,0-0,7 0,03 0,3 -,0 0,,3-0,3-0, 0,77 0, -0,3-7,0 7-0, -0, -0,0 0, 0,3-0,8 8-0, -0,3-0, 0, 0, 0,7 -,8 -, -0,77-0,03 0,7-0,3 0-0,0 0,0 0,0-0,3 0,3 0,3 0,8-0,3-0,30-0,0-0, 0,0-0,3 0,07 0, 0,0 0,3 0, 3-0, 0, 0,0-0,07-0,0 0,7-0,03-0, 0,00 0, 0, 0,0-0,03-0, -0,0 0,, 0,0 7 0,0 0,07-0,0-0,03 0, 0, 8 0,00-0, 0,0-0,8 0,7 0, -,8 -, -0, 0,7 3,7 3,8 0-0,3 0,07 0, 0,0 0,3 0, 0,0 0,07 0,00 0,0 0,3 0,8 0,33-0, 0,00,8-0,0 -,8 3-0,03 0,7 0, 0, 0, -0,0 E n > E n < - In the Figures to 7 are presented in graphical form (for each weight) the results of measurements reported by participating laboratories.

00 g weight,0,70,0 0,70 3 7 8 0 3 7 8 0 3 0,0-0,30-0,80 -,30 Fig. The difference between participant s results and reference value and uncertainty for 00 g weight 0 g weight 0,0 3 0,0 3 0,0 3 7 8 0 7 8 0 0,00-0,0-0,0 Fig.3 The difference between participant s results and reference value and uncertainty for 0 g weight

0 g weight 0,0 0,0 0,30 0,0 0,0 0,00 3 7 8 0 3 7 8 0 3-0,0-0,0-0,30 Fig. The difference between participant s results and reference value and uncertainty for 0 g weight g weight 0,30 0, 0,0 0, 0,0 0,0 0,00 3 7 8 0 3 7 0 3-0,0-0,0 8-0, Fig. The difference between participant s results and reference value and uncertainty for g weight 7

00 mg weight 0,00 0,00 0,000-0,00-0,00-0,0-0,00 3 7 8 0 3 7 8 0 3-0,0-0,300 Fig. The difference between participant s results and reference value and uncertainty for 00 mg weight 0 mg weight 0,00 0,00 0,000-0,00-0,00 3 7 8 0 3 7 8 0 3-0,0-0,00-0,0-0,300 Legend: o Fig.7 The difference between participant s results and reference value and uncertainty for 0 mg weight Reference mass value; Measurement uncertainty U (k=) reported by reference laboratory; Result of measurement carried out by a participating laboratory; Measurement uncertainty U (k=) reported by a participating laboratory. Figure 7 presents a complete image, a centralization of normalized errors obtained from participants results, for all the weights. 8

Normalized errors 3 7 8 0 3 7 8 0 3 00 g 0 g 0 g g 00 mg 0 mg - +.3 Analysis of results Fig. 7 Centralization of normalized errors As general observations it can be mentioned the following: - Lab : in the handwritten tables are some components of uncertainty that are not in the printed tables; - Lab : the characteristics of the balances from the annex of the technical protocol are not in accordance with the primary data; - In Table it can be seen that laboratories 8,, and have some results where the number of decimal places of conventional mass is not consistent with that of calibration uncertainty; - Laboratories codified with numbers,,, 8, 0, 3, 8, and presented some results with an excessive number of decimal places; - Laboratories,,, 3, 8, and reported an uncertainty less than /3 of the maximum permissible error, which led (for some weights) to a non compliance with index value within the [-, ]. From the Table 3 it can be seen as following: -for values of the normalized deviations in the range [-...+], laboratory i is compatible with reference laboratory and is considered able to perform calibration of class F standard weights in the analyzed domain with corresponding measurement uncertainties []; - participating laboratories (, 3, 7, 8, 0,,, 3,, 7, 8, 0,, 3) have obtained all results consistent with those of pilot laboratory for F accuracy class. Table presents a centralization of uncertainty components reported by each participant laboratory. Thus, can be also viewed the deficiencies of each participating laboratory in reporting the expanded uncertainty.

Table : The measurement uncertainty components of each participating laboratory and deficiencies observed by the reference/pilot in reporting of results * the two components were listed in the same column because the participants have differently noted them. o ) for some balances this component should be reported; o ) it was not properly calculated; o 3) even is used the same balance, this component it was not introduced in the uncertainty budget for all the weights; not applicable; v these components should be introduced in the budget table even if their effect was negligible. 7 CONCLUSIONS The results of interlaboratory comparison show that approximately % from the total results are discrepant. Fourteen participating laboratories (, 3, 7, 8, 0,,, 3,, 7, 8, 0,, 3) have obtained all results compatible with those of the reference laboratory for class F. Two laboratories ( and ) have only one incompatible result, four laboratories (,, and ) have two and laboratory no. has four inappropriate results. Table presents the weights for which these laboratories have obtained inconsistent results. Given the participants' comments, as their proposed corrective actions, the reference laboratory / pilot made the following proposals: - validation of the calculation file used; - for ensuring the quality results, it is recommended replicated calibrations using the same or different methods; - a better checking of input data (environmental conditions, measuring instruments used, etc.); - reducing the time between calibrations of measuring instruments (where applicable); 0

- it is also recommended that the number of performed measuring cycles not be in excess (see section. C..3 of []). By performing this ILC it was also checked the correctness of reporting measurement results, the method of calculating the mass and associated uncertainty and also in establishing the degree of compatibility between the participants results and results of measurements obtained by the reference laboratory. REFERENCES [] ISO/IEC 70:00(E): General requirements for the competence of testing and calibration laboratories. [] OIML R - Weights of classes E, E, F, F, M, M-, M, M-3 and M3, 00. [3] SR EN ISO/CEI 703, Conformity assessment General requirements for proficiency testing. [] Adriana Vâlcu. Interlaboratory Comparison of Five Standard Weights of Class F in Several Romanian Laboratories. SPRINGER (MAPAN Journal), 7, -.