Analysis of Geometrical Aspects of a Kelvin Probe

Similar documents
Trek electrostatic voltmeters Setup, environment, working conditions

Magnetic and Thermal Analysis of Current Transformer in Normal and Abnormal Conditions

Virtual Prototyping of Electrodynamic Loudspeakers by Utilizing a Finite Element Method

Electromagnetics in COMSOL Multiphysics is extended by add-on Modules

Microstructure cantilever beam for current measurement

Operation of an Electromagnetic Trigger with a Short-circuit Ring

Comparison of 2D and 3D FEM Models of Eddy Current Pressure Tube to Calandria Tube Gap Measurement

Active elastomer components based on dielectric elastomers

Study and Characterization of the Limiting Thermal Phenomena in Low-Speed Permanent Magnet Synchronous Generators for Wind Energy

Wire Based Flexible Piezoelectric Sensor for Structural Health Monitoring Applications

Prediction of Transformer Core Noise

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2009 PROBLEM SET #7. Due (at 7 p.m.): Thursday, Dec. 10, 2009, in the EE C245 HW box in 240 Cory.

PHYSICS 2B FINAL EXAM ANSWERS WINTER QUARTER 2010 PROF. HIRSCH MARCH 18, 2010 Problems 1, 2 P 1 P 2

Modeling of Ultrasonic Near-Filed Acoustic Levitation: Resolving Viscous and Acoustic Effects

Measurement of the electric field at the near field radiating by electrostatic discharges

Modelling of Different MEMS Pressure Sensors using COMSOL Multiphysics

SIMULATION OF HEAT TRANSFER AND ELECTROMAGNETIC FIELDS OF PROTECTED MICROCOMPUTERS

Large Scale Computation of Coupled. Electro-Acoustic Systems using ANSYS and CAPA

Simulation and verification of

A Modeling Study of Electrical Characteristics of Anisotropic Conductive Film Adhesives

Printed Acceleration Sensor

Modeling Electric Fields in High Voltage Submersible Changeover Switch

o Two-wire transmission line (end view is shown, the radius of the conductors = a, the distance between the centers of the two conductors = d)

Transduction Based on Changes in the Energy Stored in an Electrical Field. Lecture 6-5. Department of Mechanical Engineering

Transducers. Today: Electrostatic Capacitive. EEL5225: Principles of MEMS Transducers (Fall 2003) Instructor: Dr. Hui-Kai Xie

The Influence of Core Shape and Material Nonlinearities to Corner Losses of Inductive Element

E05 Resonator Design

Transduction Based on Changes in the Energy Stored in an Electrical Field

Applicability of Self-Powered Synchronized Electric Charge Extraction (SECE) Circuit for Piezoelectric Energy Harvesting

Filters and Tuned Amplifiers

Modeling and Simulation of Artificial Core-Shell Based Nanodielectrics for Electrostatic Capacitors Applications

Time-Harmonic Modeling of Squirrel-Cage Induction Motors: A Circuit-Field Coupled Approach

REVIEW EXERCISES. 2. What is the resulting action if switch (S) is opened after the capacitor (C) is fully charged? Se figure 4.27.

Biosensors and Instrumentation: Tutorial 2

EXPLORING SCANNING PROBE MICROSCOPY WITH MATHEMATICA

A Fixed Surface Potential Probe with the Swing Capacitive Electrometer Compared to the Vibrating Kelvin Probe

Exercise 1: Capacitors

INF5490 RF MEMS. LN03: Modeling, design and analysis. Spring 2008, Oddvar Søråsen Department of Informatics, UoO

Simulation and Experimental Characterizations of a Thin Touch Mode Capacitive Pressure Sensor

Electrostatic Microgenerators

Sensors & Transducers 2016 by IFSA Publishing, S. L.

Final Exam: 11 May 2001, Friday, 10 AM

MAS.836 PROBLEM SET THREE

Possibilities of Using COMSOL Software in Physics

EE C245 / ME C218 INTRODUCTION TO MEMS DESIGN FALL 2011 C. Nguyen PROBLEM SET #7. Table 1: Gyroscope Modeling Parameters

Robust shaft design to compensate deformation in the hub press fitting and disk clamping process of 2.5 HDDs

not to scale Show that the potential difference between the plates increases to about 80 V. Calculate the energy that is now stored by the capacitor.

Transient Finite Element Analysis of a Spice-Coupled Transformer with COMSOL-Multiphysics

COUPLED MAGNETO-MECHANICAL FINITE ELEMENT ANALYSIS OF A POWER TRANSFORMER IN SHORT CIRCUIT CONDITIONS

Physics 102 Spring 2006: Final Exam Multiple-Choice Questions

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

COURSE OUTLINE. Introduction Signals and Noise Filtering Sensors: Piezoelectric Force Sensors. Sensors, Signals and Noise 1

NAME: PHYSICS 6B SPRING 2011 FINAL EXAM ( VERSION A )

Capacitor Action. 3. Capacitor Action Theory Support. Electronics - AC Circuits

Problem Solving 9: Displacement Current, Poynting Vector and Energy Flow

Optimizing micromechanical force detectors for measuring. magnetization at high magnetic fields

Physics Investigation 10 Teacher Manual

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

Exam 2 Solutions. Note that there are several variations of some problems, indicated by choices in parentheses.

Analysis of the thermal heating of poly-si and a-si photovoltaic cell by means of Fem

Multidomain Design and Optimization based on Comsol Multiphysics: Applications for Mechatronic Devices

Modeling of the Material/Electrolyte Interface and the Electrical Current Generated during the Pulse Electrochemical Machining of Grey Cast Iron

PHYA4/2. General Certificate of Education Advanced Level Examination January Unit 4 Fields and Further Mechanics Section B

Calculation of Cable Parameters for Different Cable Shapes

ECE 341 Test 2, Spring 2019 Your Name Tue 4/16/2019

PIEZOELECTRIC TECHNOLOGY PRIMER

Study of DC Cylindrical Magnetron by Langmuir Probe

time/s force/n Polesworth High School 1

SIMULATION OF A TIME DEPENDENT 2D GENERATOR MODEL USING COMSOL MULTIPHYSICS

Modelling a Resonant Near Field Scanning Microwave Microscope (RNSMM) Probe

Resistance and Conductance

ACTIVE VIBRATION CONTROL PROTOTYPING IN ANSYS: A VERIFICATION EXPERIMENT

Exam 4 (Final) Solutions

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

fiziks Institute for NET/JRF, GATE, IIT-JAM, JEST, TIFR and GRE in PHYSICAL SCIENCES

CIRCUIT ELEMENT: CAPACITOR

University of California, Berkeley Physics H7B Spring 1999 (Strovink) SOLUTION TO PROBLEM SET 11 Solutions by P. Pebler

Finite Element Modeling of Ultrasonic Transducers for Polymer Characterization

PURPOSE: See suggested breadboard configuration on following page!

Piezoelectric Resonators ME 2082

Physics 24 Exam 2 March 18, 2014

Physics 6B Summer 2007 Final

Books by Dieter Stotz

Scanning capacitance spectroscopy of an Al x Ga 1Àx NÕGaN heterostructure field-effect transistor structure: Analysis of probe tip effects

Towards Rotordynamic Analysis with COMSOL Multiphysics

ANALOGY BETWEEN ELECTROSTATIC FIELD AND HEAT TRANSFER SIMPLE MODELS

Characteristics and Definitions Used for Film Capacitors

Electrical Indicating Instruments

PHYS 1444 Section 004 Lecture #22

Creep characteristics of piezoelectric actuators

De La Salle University Manila Physics Fundamentals for Engineering 2 Quiz No. 3 Reviewer

AN019. A Better Approach of Dealing with Ripple Noise of LDO. Introduction. The influence of inductor effect over LDO

Pretest ELEA1831 Module 11 Units 1& 2 Inductance & Capacitance

Simulation and Experimental Characterizations of a Thin Touch Mode Capacitive Pressure Sensor

2. What are the 4 steps of the Scientific Method as described by Mr. Martin?

Final Exam: 11 May 2001, Friday

Induction_P1. 1. [1 mark]

PROPERTY STUDY ON EMATS WITH VISUALIZATION OF ULTRASONIC PROPAGATION

Simulation of Piezoelectric Induced Lamb Waves in Plates

An Optimised High Current Impulse Source

Transcription:

Analysis of Geometrical Aspects of a Kelvin Probe Stefan Ciba 1, Alexander Frey 2 and Ingo Kuehne* 1 1 Heilbronn University, Institute for Fast Mechatronic Systems (ISM), Kuenzelsau, Germany 2 University of Applied Science, Faculty of Electrical Engineering, Augsburg, Germany *Corresponding author: ingo.kuehne@hs-heilbronn.de Abstract: The presented analysis investigates the capacitance characteristic of a kelvin probe regarding the geometrical transition from a movable electrode plate to a narrow tip. Moreover, predictions can be done concerning optimum geometry, sensitivity and suitable electrical measurement circuitry. A further aim of this study is to provide optimal tip geometries for different sized kelvin probes. This includes pitching the probes impedance and maximizing the current yield out of the contact potential difference between two metals. Keywords: Kelvin probe, contact potential difference, work function, non-contact measurement technique, FEM 1. Introduction A kelvin probe is a non-contact, nondestructive measurement technique to investigate the work function difference of material combinations [1, 2]. It is based on a time varying capacitor where the movable electrode consists of a material A and a fixed electrode of a material B with different work functions [3]. For electrically connected electrodes (e.g. external wire bond) the Fermi levels are equilibrated. Electrons originating from the material with the lower work function flow to the material with the higher work function. Consequently, a contact potential difference establishes and the capacitor is charged. The vibration of the movable electrode results in a change of the stored electrical energy within the capacitor forcing an electrical current to flow. The work function difference can be determined by measuring this current. Alternatively, a compensated operation can be established by applying an external bias voltage so that the current becomes zero. As a result the bias voltage is equivalent to the work function difference of the electrode material pair [4]. On the one hand, the movable electrode can be a plate in order to form a parallel plate capacitor. This guarantees a large capacitance change which results in a current that can be directly measured without a sophisticated electrical compensation scheme. On the other hand, the movable electrode can be formed as a narrow tip what gives the ability to get a lateral resolution by scanning over some area [5]. Here, the capacitance change and so the resulting current are quite low what calls for an electrical compensation circuit. 2. Use of COMSOL Multiphysics The model is set up with the Electromechanics Interface. For numeric efficiency, a 2D axial symmetry is established in order to model a vibrating capacitor with circular plate/tip geometries. The movable electrode is configured to a prescribed harmonic displacement and a defined contact potential difference. The counter-electrode is mechanically fixed and set to electrical ground. The electromechanical field equations used are the electrostatic force equation and the electric flux density equation: ρ δ2 u δt σ=fv 2 D=ρ v The transient behavior is observed and the lateral dimension of the movable electrode is varied by a parametric sweep (cf. figure 1). Furthermore a parametric sweep of probe length is done and an analysis of different tip geometries. Figure 1. 2D axial symmetric model setup of the kelvin probe.

3. Transient behavior of kelvin probe Since the motion of the movable electrode is set up as a sine wave, the capacitance gets slightly distorted, like predictable by the well known analytical plate capacitors formula for the capacitance C: C (t )=ε 0 ε r π r me g 0 +m(t ) m(t)= a sin (2π f t) in which m is the motion with an amplitude a=30μm and a frequency f=1hz (this value has been chosen for calculation purposes note that a real kelvin probe needs higher frequencies for optimal impedance matching), g 0 =100μm is the nominal gap of the movable electrode. These parameters are used in every model for comparability. Figure 2 shows the characteristic transient behavior for a single sine input cycle. 2 This phenomenon makes numerical calculation necessary which also is comfortable. COMSOL AC/DC module uses the energy method to calculate the capacitance out of the electric energy density and makes an easy evaluation for every single time step data possible [6]. The average difference of the analytical results amount at least 8% at r me =5mm and increases dramatically by downscaling geometry parameters. There is not only an offset in nominal capacitance, rather than an increasing difference of the capacitance change as well. By matching figures 3 and 4 it becomes obvious that a kelvin probe design requires progressive numerical calculation techniques if aimed for high lateral resolution, so that expensive series of experiments can become as short as possible. Figure 3. Nominal capacitance as a function of radius. Figure 2. Characteristic transient capacitance behavior of the kelvin probe. Non applicability of analytical model For further reviewing the capacitance behavior in downscaling, the radius r me of the movable electrode was swept. By downscaling the plate capacitors radius into a sub millimeter domain, the well known analytical plate capacitor model becomes more incorrect. That is because the electrical stray capacitance gets a higher influence to the capacitive behavior each step in downscaling. These capacitances are caused by the upper side and the outer sides of the plates geometry, which also contain electrical charges. Thus the capacitance C(t)=2 W e (t) / V 2 has to be calculated out of the electrical energy W e all over the domain and the contact potential difference V. Figure 4. Capacitance change as a function of radius.

4. Geometry scaling of kelvin probe The geometry scaling was done stepwise to isolate individual effects, by scaling length and radius of the probe separately. While working on the models it became apparent that the individual effects increase dramatically by downscaling both radius and length at the same time (cf. figure 7). 4.1 Impact of scaling lateral dimension The radius r me of the movable electrode was scaled, while the length remains constant at 0.05mm just to force the impact of scaling the lateral dimension. The simulation results (cf. figure 3 and 4) show, the bigger the movable electrode, the higher the nominal capacitance as well as the capacitance change. The good news is, that surface effects can provide higher capacitances in a real measurement setup. But it also leads to the conclusion, that a higher lateral resolution end at the point where the measurement equipment can not provide reliable values, if just downscaling the lateral dimension. Therefore the kelvin probe should have a minimum length. This effect becomes more critical, if scaling down the radius at the same time (cf. figure 7). Bear in mind that figures 5 and 6 show the characteristics at a constant radius of 0.5mm. The logarithmic scale in figure 7 adumbrate the tremendous gain in these characteristics. Further postprocessing showed, that there is no fixed proportion between length and radius for analytical simplifications. Each single geometry must be reviewed separately. Figure 5. Nominal Capacitance as a function of length. 4.2 Impact of scaling the length The length of the probe is scaled, while the radius remains constant at 0.5mm. The simulation results (cf. figure 5 and 6) show, that the nominal capacitance scales in a different way as the capacitance change. After underrunning a length of approximately 2mm the capacitance change decreases significantly. Figure 6. Capacitance change as a function of length. Figure 7. electrode. Left: The nominal capacitance, right: The capacitance change as functions of length and radius of the movable

5. Tip geometries of kelvin probe Different tip geometries provide different capacitance behaviors. That is because of different electric energy densities. To study the consequence of different tip geometries three relativeley simple geometries are chosen, in which the flat tip acts as reference (cf. figure 8). The simulation results show that a flat tip has the highest nominal capacitance (cf. figure 10) and capacitance change (cf. figure 11). Figure 10. Nominal capacitance of different tips. Figure 8. Tips: flat tip, round tip, spikey tip (45 ). To maintain comparability, each geometry length was set to 2mm while the radius was swept. The biggest challenge though is the meshing in COMSOL. Different airspace geometries were set up to provide exact calculations while enabling the sine movement and the parametric sweeps of the different geometries. For instance the meshing which was done for the round tip is shown in figure 9. Figure 11. Capacitance change of different tips. Figure 9. Additional airspace meshing of the round tip model. The smaller the lateral tip dimensions are, the closer the capacitance values of the different tip geometries. In case of narrow tips (r me < 100 μm) the tip geometry (flat, round, spikey) looses importance. Figure 12 shows the electric potential of the different tip geometries. The electric potential is comparable in the close gap. The vertical boundaries account for the stray capacitances, which have minor influence on the kelvin probe. Figure 12. Electric potential of three different tips: flat tip, round tip, spikey tip (45 ).

6. Equivalent electrical circuit of kelvin probe In order to find a suitable equivalent electrical measurement circuit, the model was exemplarily extended by a simple load resistor (cf. figure 13). The Spice model is based on the analytical equations of a plate capacitor (see section 3.) and shows a good correlation. 7. Conclusions A COMSOL model of a kelvin probe was successfully established and different probe geometries were investigated. Moreover, the model was expanded by a simple electrical measurement circuitry to demonstrate a complete kelvin probe setup. Future investigations should lead to a highly sensitive electrical circuit (e.g. compensated scheme or differential amplifier scheme) in order to optimize performance. Figure 13. Equivalent electrical circuit of kelvin probe with simple voltage measurement application. For comparison purposes the model was set up in COMSOL (Electrical Circuit) and in Spice. Figure 14 shows the capacitance characteristic at a frequency of 100Hz for two periods. Figure 14. Characteristic transient capacitance behavior of the kelvin probe including electrical circuit. Figure 15 presents the voltage drop over the load resistor (R L =10MΩ). 8. References [1] Sir William Thomson (Lord Kelvin): Electrostatics and Magnetism, MacMillan & Co., London, 1872. [2] J. Erskine-Murray: On Contact Electricity of Metals, Proceedings of Royal Society of London Vol. 63, pp. 113-146, London, January 1898. [3] J. C. Riviere: Contact Potential Difference Measurements by the Kelvin Method, Proceedings of the Physical Society of London - Section B, Vol. 70, pp. 676-686, London, 1957. [4] Hans Lueth: Solid Surfaces, Interfaces and Thin Films, Springer Verlag, pp. 564-570, Berlin, 2010. [5] Laurent Nony: Principles of Kelvin Probe Force Microscopy and applications, 1st German- French Summer School on noncontact - AFM Porquerolles, pp. 80, France, 2013. [6] COMSOL: AC/DC Module User's guide, Chapter 3: Modeling with the AC/DC Module, pp. 68, 2013. 9. Acknowledgements Figure 15. Characteristic transient voltage drop over load resistor. Supported by the foundation for the promotion of the Reinhold Würth University of Heilbronn University in Künzelsau.