Dynamic light scattering studies of some modes with wavevectors perpendicular to the twist axis in cholest erics

Similar documents
Polarized Light. Nikki Truss. Abstract:

Probing Atomic Crystals: Bragg Diffraction

Fixturlaser Alignment System

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

Measurements of Liquid Scintillator Light Yield for Future Neutrino Experiments

Modeling Airplane Wings

Apparatus to measure high-temperature thermal conductivity and thermoelectric power of small specimens

Transient Electro-Optic Properties of Liquid Crystal Gels

Determination of Cauchy s Contants

arxiv: v1 [physics.ins-det] 3 Feb 2011

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Tutorials. 1. Autocollimator. Angle Dekkor. General

Polarization of Light and Birefringence of Materials

XRD RAPID SCREENING SYSTEM FOR COMBINATORIAL CHEMISTRY

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

DEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT

Lab #13: Polarization

Learn how reflection at interfaces with different indices of refraction works and how interfaces can change the polarization states of light.

ECE185 LIQUID CRYSTAL DISPLAYS

An accessory to the polarizing microscope for the optical examination of crystals.

Lab E3: The Wheatstone Bridge

Optimizing the Nematic Liquid Crystal Relaxation Speed by Magnetic Field

PH2200 Practice Final Exam Summer 2003

ELECTRIC FIELD EFFECT ON OPTICAL DICHROISM OF SOME AZODYES DISSOLVED IN COMPENSATED CHOLESTERIC LIQUID CRYSTAL

Lab 3-4 : Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury-Brown and Twiss Set Up, Photon Antibunching

Analysis on the birefringence property of lyotropic liquid crystals below Krafft temperature

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

PERFORMANCE OF THE ATLAS LIQUID ARGON FORWARD CALORIMETER IN BEAM TESTS

Atomic and nuclear physics

Confocal Microscope Imaging of Single emitter fluorescence and Observing Photon Antibunching Using Hanbury Brown and Twiss setup. Lab.

DELHI PUBLIC SCHOOL, BAHADURGARH Sample Paper 1 PHYSICS CLASS-XII Date- Duration:3hr

Light Scattering Study of Poly (dimethyl siloxane) in Liquid and Supercritical CO 2.

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

Atomic and nuclear physics

How DLS Works: Interference of Light

This document consists of 10 printed pages.

COVENANT UNIVERSITY NIGERIA TUTORIAL KIT OMEGA SEMESTER PROGRAMME: MECHANICAL ENGINEERING

Polarization control and sensing with two-dimensional coupled photonic crystal microcavity arrays. Hatice Altug * and Jelena Vučković

Direct measurement of electric-field-induced birefringence in a polymer-stabilized blue-phase liquid crystal composite

Tilt Angle and Birefringence of Smectic Liquid Crystal Materials

Electric Field PHYS 296

Supporting Information for Induced smectic phase in binary mixture of twist-bend nematogens

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Cherenkov Detector. Cosmic Rays Cherenkov Detector. Lodovico Lappetito. CherenkovDetector_ENG - 28/04/2016 Pag. 1

THE DIFFRACTION GRATING SPECTROMETER

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Aries. Kapton QFP socket Cycling Test DC Measurement Results. prepared by. Gert Hohenwarter

DETECTORS. I. Charged Particle Detectors

Model 556X User s Manual. Optical Isolator

PC1142 Physics II. The Ideal Gas Law

PHYSICS 2005 (Delhi) Q3. The power factor of an A.C. circuit is 0.5. What will be the phase difference between voltage and current in this circuit?

CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD

LABORATORY OF ELEMENTARY BIOPHYSICS

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

PhysicsAndMathsTutor.com

Velocity and temperature measurements in a large-scale Rayleigh-Bénard experiment using LDA and micro thermistors

SAXS COURSE Richard Koschuch. HECUS M.Braun Graz X-Ray Systems. and. Institute of Biophysics and X-Ray Structure Research,

Pre-lab Quiz/PHYS 224 Coulomb s Law and Coulomb Constant. Your name Lab section

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

PHYSICS 5054/21. Published

Experiment 8. Fresnel Coefficients. 8.1 Introduction. References

Measuring Cosmic Ray Muon Decay Constant and Flux

KERR EFFECT. Monochromatic, vertically polarized light impinges on a PLZT element (leadlanthanum-zirconium-titanium

ELECTRIC FIELD. 2. If you have an equipotential surface that means that the potential difference is zero, along that surface. a. true b.

LASER INTERFEROMETER GRAVITATIONAL WAVE OBSERVATORY - LIGO - CALIFORNIA INSTITUTE OF TECHNOLOGY MASSACHUSETTS INSTITUTE OF TECHNOLOGY

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

EXPERIMENTAL COMPETITION

Experimental test of an expression for the decay of an autocorrelation function

Photoelectric Effect Experiment

5 Quantum Wells. 1. Use a Multimeter to test the resistance of your laser; Record the resistance for both polarities.

Thin Calorimetry for Cosmic-Ray Studies Outside the Earth s Atmosphere. 1 Introduction

Dr. Radhakrishnan A N Assistant Professor of Physics GPTC, Vechoochira

Coulomb s Law and Coulomb s Constant

Physics 313: Laboratory 8 - Polarization of Light Electric Fields

Density Field Measurement by Digital Laser Speckle Photography

Higher Physics. Particles and Waves

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education

Thermionic Emission. A. Goode Student and A. Finicky Professor. Abstract. φ 2 kt

MAGNET INSTALLATION AND ALIGNMENT FOR THE FUJI TEST BEAM LINE AT KEKB

X-ray fluorescence analysis - calibration of the X-ray energy detector

Structure Report for J. Reibenspies

: Imaging Systems Laboratory II. Laboratory 6: The Polarization of Light April 16 & 18, 2002

CBSE PHYSICS QUESTION PAPER (2005)

Deviations from Malus Law

Surface plasmon resonance based refractive index sensor for liquids

SYNCHROTRON X-RAY MICROBEAM CHARACTERIZATION OF SMECTIC A LIQUID CRYSTALS UNDER ELECTRIC FIELD

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES

Assembly and test runs of decay detector for ISGMR study. J. Button, R. Polis, C. Canahui, Krishichayan, Y. -W. Lui, and D. H.

Lecture 7: Extrinsic semiconductors - Fermi level

I. Pre-Lab Introduction

Evolution of Disclinations in Cholesteric Liquid Crystals

Near field radiative heat transfer between a sphere and a substrate

A Photonic Crystal Laser from Solution Based. Organo-Lead Iodide Perovskite Thin Films

School. Team Number. Optics

The FEA Code of LASCAD

LECTURE 11 ELECTROMAGNETIC WAVES & POLARIZATION. Instructor: Kazumi Tolich

CHARGED PARTICLE INTERACTIONS

Transcription:

Chapter 4 Dynamic light scattering studies of some modes with wavevectors perpendicular to the twist axis in cholest erics 4.1 Introduction In the previous chapter, we described our studies on a well isolated viscoelastic mode with wavevector parallel to the twist axis in a cholesteric. Those experiments were performed in the back scattering arrangement. In this chapter, we study fluctuations with wavevectors perpendicular to the twist axis. Here, the associated modes are very different. The experiments were performed in the transmission arrangement, where the light scattered in the forward direction is analyzed. The interesting result we obtain is that at high scattering wavevectors (equivalently scattering angles), the intensity autocorrelation functions have a non-exponential decay, which implies coupled modes. At low scattering wavevectors they turn out to be single exponential suggesting an isolated relaxation mode [I].. 4.2 The forward scattering arrangement In this arrangement, light is incident in a direction normal to the sample cell. The incident light is linearly polarized with its electric vector perpendicular to the scattering plane. In our experiment, the cholesteric sample is aligned with its twist axis parallel to the cell normal. In forward scattering, light is incident in a direction

Twist axis 9 7 1 0 'sl Polar Figure 4.1: Ray diagram showing the interfaces at which the scattered light experiences refraction.in the figure, 0 shows that the incident electric vector is perpendicular to the scattering plane while t, indicates that the scattered light is analyzed with its electric vector parallel to the scattering plane, parallel to the twist axis of the cholesteric and the light scattered in the forward direction is analyzed. The scattered light is passed through an analyser (who's axis is parallel to the scattering plane) before it enters the detector. This is known as the VH arrangement. V (vertical) stands for the polarization of the incident light and H (horizontal) for the direction in which the scattered light is analysed. The directions H and V are with reference to the scattering plane. In these experiments we used a sample holder kept directly in the oven. In the absence of a refractive index matching medium, it is necessary to apply a correction to the observed angle (Bob,) of scattering. This is shown in figure(4.1). In our calculations of the scattering wavevectors, we have used such a corrected scattering angle (O,,).

The relation between the observed and the corrected angles is given by Here, p,c is the refractive index of air and psample is the average refractive index of the sample. The average refractive index of the sample is nearly 1.55. The details of the scattering arrangement are shown in figure(4.2). A simple geo- metric analysis yields the following expressions for the components of the scattering wavevect or q. and Here, the parallel and perpendicular directions are defined with respect to the twist axis of the cholesteric. 4.3 Experimental 4.3.1 Sample preparation The cholesteric used in the experiment was a three component mixture consisting of cholesteryl chloride (CCL), cholesteryl nonanoate (CNN) and cholesteryl oleyl carbonate (COC). In the particular mixture used in the experiment, the weight percentages of the compounds were the same as that mentioned in chapter 3. This mixture is a room temperature cholesteric. The cholesteric samples are aligned with the twist axis perpendicular to the glass plate. The details of the alignment procedure and sample preparation have already been described in chapter 3. 4.3.2 Design of the sample cell temperature control unit To do experiments in the forward scattering arrangement, we constructed a different temperature control unit consisting of two parts, the sample oven and a sample holder. The cholesteric sample cell is mounted on the holder which is housed inside an oven. 75

Sample Figure 4.2: The forward scattering arrangement. Here, the sample is aligned such that the twist axis is parallel to the cell normal. The incident wavevector is parallel to the twist axis of the cholesteric and the scattering wavevector is predominantly perpendicular to the twist axis. The VH arrangement is indicated in the figure by the symbols 0 and t, respectively.

PPS DMM L I I Figure 4.3: A schematic diagram of the oven and sample holder used for the forward scattering experiments. A cross-sectional view from above is shown. DMM is a digital multimeter that measures the RTD resistance. PPS is a programmable power supply that provides power to the heating elements. PC is a personal computer that controls the multimeter and the power supply via a GPIB interface. The oven used in this experiment was designed for good temperature stability and hence had a large thermal mass. A schematic diagram of the sample oven and holder is shown in figure(4.3). The oven consists of two blocks of copper firmly bolted together with the heating elements (Kapton type, Minco Inc.) sandwiched between them. Each of the two blocks had a central cylindrical tunnel in it to accommodate the sample holder and allow the passage of incident and scattered light. The heating elements were annular in shape. The central annulus was exactly of the same diameter as the tunnels in the copper blocks. The heating elements were mounted such that they encircle the tunnel. The advantage of using annular heating elements is that the sample would

be uniformly heated from all sides. This would prevent any temperature gradients across the sample. A small amount of heat sink material was smeared on to the two sides of the heating elements to ensure good thermal contact with the copper blocks. The holder was cylindrical in shape and was also made of copper. The diameter of the holder was almost exactly the same as that of the tunnel. The holder fits snugly in the tunnel to ensure efficient conduction of heat from the oven to the holder. The holder itself has a hole running through its entire length to allow for the passage of the incident light. The sample cell wm mounted on the holder with the aid of two metallic screws and was held firmly in contact with the body of the holder with teflon washers. A resistance temperature device (RTD) was positioned such that after mounting the sample, the RTD could be made to slide and come in direct contact with the sample cell. The entire oven was enclosed in a 3mm thick teflon jacket to reduce radiative heat loss. The PI temperature controller described in chapter 2 was used in conjunction with this heater. The experiments described in this chapter were carried out for samples maintained at 32f 0.05"C. The body of the oven was designed with flat outer surfaces so that it could be removed from the goniometer stand and placed on the stage of a polarizing microscope directly. This feature of the oven enabled us to examine the sample alignment between different runs in the light scattering experiments without having to remove the sample cell from the oven. 4.3.3 Apparatus and measurements Light from a 488 nm Argon ion laser (Spectra Physies Model 163) is incident in a direction normal to the sample cell. The sample oven is positioned such that the scattering volume lies exactly at the intersection of the goniometer axis and the incident beam. Light scattered from the sample enters a photomultiplier tube (PMT) after passing through an analyser. The PMT is mounted on the movable arm of the goniometer. The signals from the PMT are passed on to a digital autocorrelator (Malvern 4700c) via a pre-pmplifier discriminator system fabricated by us described in chapter 2. We selected different delay times depending on the angle at which the

Total delay time (ps) Figure 4.4: A typical normalized correlogram G2(7), obtained in the experiment. The corrected scattering angle was 29.62 degrees. The sample temperature was maintained at 32f 0.05"C. The delay time was 200 0s and the total integration time was 200s. The continuous line is obtained from a two exponent relaxation model described in the text. correlogram data was being acquired. Data was acquired for observed scattering angles between 35 and 50 degrees which corresponded to a range of corrected scattering angles between 29.62 degrees and 21.72 degrees. The total integration time for which each datum was collected was 200 seconds. We observe that the signal to noise ratio of the experiments reduces uniformly as we approach lower scattering angles. At these angles, the signal to noise ratio could be improved by reducing the effective area of the detector by using smaller pinholes. The delay time used is such that the decaying part of the correlation function is recorded in approximately the first 150 channels. The rest of the channels record the tail of the autocorrelation function. 79

Total delay time T (p) Figure 4.5: The natural logarithm of the intensity correlation function shows a change in slope. There is an initial fast decay and then a slower one. The two slopes correspond to the two time scales present in the fluctuations in the system. The solid lines are drawn as cues to the eye.

Figure 4.6: The variation of the inverse relaxation times of the two modes obtained in -the experiment as a function of the square of the scattering wavevector. The circles indicate the fast mode and the triangles indicate the slow mode. For the slow mode, the error bars are smaller than the size of the markers and hence are not indicated. These experiments were performed at 32k0.05"C. The solid lines are drawn as cues to the eye.

4.4 Analysis of the intensity autocorrelat ion functions A typical correlogram obtained in these experiments is shown in figure(4.4). It can be seen that it is not a single exponential decay. We plotted the natural logarithm of the correlation function with the delay time. We could discern two linear portions of different slopes separated by a cross over region. This is shown in figure(4.5). It indicates that there are two relaxation times in the system. Based on this trend, we decided to fit a two exponential decay to our data. The model we have used is, Here, cl is a base line parameter, c2 and c4 are strength parameters and cs and cs are the relaxation times. We define the normalized autocorrelation function as: Where, and g2(oo) is the value of the base line. This value is automatically calculated by the correlator by considering very long delay times. A fit of our data, the normalized autocorrelation function G2(7), to the two exponential model is shown as a continuous line in figure(4.4). One can notice that the fit is quite good. The error estimates on the relaxation times are calculated from the standard errors obtained by taking five data sets at every angle [2]. 4.5 Results and discussions In the forward scattering arrangement, the scattering wavevector has a large component in a direction perpendicular to the twist axis. This perpendicular component of

the wavevector senses fluctuations within the cholesteric planes. We have detected fluctuations with two distinct time scales. The inverse relaxation times of these fluctuations as a function of the square of the scattering wavevector are shown in figure(4.6). It can be seen that the two modes behave quite differently. The relaxation time of the faster mode decays rapidly with wavevector and seems to saturate at lower wavevectors 131. On the other hand, the slower mode doesn't appear to have such a marked dependence on wavevector. We speculate that' we are observing some new coupled modes. These modes might be arising due to the coupling of the twist and umbrella fluctuations. The analysis of these modes is quite complex in this forward scattering arrangment since it is not possible to isolate them. It is also possible that these modes are due to layer undulations. According to an argument by de Gennes [4] we can assume that at some level of coarse graining, cholesterics and smectics become indistinguishable from each other. This is called the 'jelly roll' model. In this model there can be layer undulations that cause scattering in a direction perpendicular to the twist axis. More experiments on &(T) as a function of temperature, pitch, sample thickness and other parameters are necessary to investigate the origin of these coupled modes.

Bibliography [l] M. S. Veshchunov. Sou. Phys. JEPT, 49, 769 (1979). [2] S. K. Muthu. Probability and Errors for the Physical Sciences. Orient Longman Limited, New Delhi (1982). [3] T. Harada and P. P. Crooker. Phys. Rev. Lett., 34, 1259 (1975). [4] P. G. de Gennes and J. Prost. The Physics of Liquid Crystals, 2nd ed., Clarendon press, Oxford, (1993).