Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Similar documents
Activity measurement of 55 Fe within the scope of the BIPM comparison 2006

BUREAU INTERNATIONAL DES POIDS ET MESURES

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

Fundamentals of Radionuclide Metrology

Standardization of 18 F by Digital β(ls)-γ Coincidence Counting

Čerenkov counting and liquid scintillation counting of 36 Cl

Standardization of Tritium by CIEMAT/NIST Method and TDCR Method

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY

Disproof of solar influence on the decay rates of 90 Sr/ 90 Y

APPLICATION OF A FREE PARAMETER MODEL TO PLASTIC SCINTILLATION SAMPLES

Adaptation of PTB s analytical modelling for TDCR-Cherenkov activity measurements at LNE-LNHB. LSC 2017, Copenhagen, 1-5 May 2017

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

TDCR and CIEMAT/NIST Liquid Scintillation Methods applied to the Radionuclide Metrology

The metrology of radioactivity. C. Michotte, BIPM

Primary Standardisation of I-125

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

ɣ-radiochromatography

ARTICLE IN PRESS. Applied Radiation and Isotopes

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods

Applied Radiation and Isotopes

Radiation Detection and Measurement

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

K 40 activity and Detector Efficiency

Slides by: Prof. Abeer Alharbi

Radionuclide metrology using LSC. - Current status and limitations -

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON

Liquid Scintillation at NPL: Practical Applications

Quality Assurance. Purity control. Polycrystalline Ingots

Basic hands-on gamma calibration for low activity environmental levels

Detection and measurement of gamma-radiation by gammaspectroscopy

RADIOPHARMACEUTICAL 11 C ACTIVITY MEASUREMENTS BY MEANS OF THE TDCR-CERENKOV METHOD BASED ON A GEANT4 STOCHASTIC MODELING

Scintillation Detector

Time-Resolved Liquid Scintillation Counting

Setup for an in-situ measurement of the total light extinction of Liquid Argon in GERDA

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

Figure 1. Decay Scheme for 60Co

Researchers at the University of Missouri-Columbia have designed a triple crystal

GUIDE TO LABORATORY SURVEYS. Introduction

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

Radioactivity standardization in South Africa

Basics of a/b Discrimination for Liquid Scintillation Counting

arxiv: v1 [physics.ins-det] 3 Feb 2011

Neutron activation analysis. Contents. Introduction

The 46g BGO bolometer

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

Gamma ray coincidence and angular correlation

Journal of Chemical and Pharmaceutical Research, 2012, 4(9): Research Article

Calibration of A Small Anode Germanium Well Detector

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Single Channel Beta-Gamma Coincidence Detection of Radioactive Xenon Using Digital Pulse Shape Analysis of Phoswich Detector Signals

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

2100TR Liquid Scintillation Counter

Unit 2. Instrumentation. Experts Teaching from Practical Experience

ß + RADIATION. Choose your application of. chromatography controlled. chemical purity TLC PET GABI. PET GABI 2x2 Nal (Tl) 0 1V output.

Received October 7, 1963

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

arxiv:physics/ v1 3 Aug 2006

Experiment Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Gamma Spectroscopy. References: Objectives:

Overview about the Work with Liquid Scintillators

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

PRODUCTS FOR EDUCATION AND TRAINING

New Neutron-Induced Cross-Section Measurements for Weak s-process Studies

MEASUREMENT AND DETECTION OF RADIATION

Development of a Radioxenon Detection System Using CdZnTe, Array of SiPMs, and a Plastic Scintillator

Precise Measurement of αt for the keV E3 transition in 103 Rh A Further Test of Internal Conversion Theory. Vivian Sabla N. Nica, J.C.

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L.

Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring

LS6500 Scintillation Counting Systems

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Gamma-Spectrum Generator

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

BALANCED WINDOW METHOD IN 14 C LIQUID SCINTILLATION COUNTING. P TheodÛrsson 1 S Ingvarsdottir G I Gudjonsson

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

The measurement of tritium in environmental water samples

QuantumMCA QuantumNaI QuantumGe QuantumGold

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Radionuclide Imaging MII Positron Emission Tomography (PET)

Calibration of the IXPE Instrument

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Determination of the activity of radionuclides

Monte Carlo Simulations for Future Geoneutrino Detectors

*Corresponding author,

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS

Compton suppression spectrometry

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Gross Alpha-Gross Beta Analysis in Water by Liquid Scintillation Counting (LSC)

Transcription:

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting Justyna Marganiec-Galazka Ole J. Nähle Karsten Kossert Division 6 Ionizing Radiation Department 6.1 Radioactivity Working Group 6.11 Unit of activity

Outline Introduction Coincidence method 4πβ(LS)-γ coincidence method Detection systems (M1 and M29) Activity measurements Summary

Introduction 88 Y electron capture nuclide (followed by a cascade of two γ-rays). 88 Y used as a radioactive source for the calibration of γ-ray spectrometers. A precise measurement of activity is not trivial due to low energy of X-ray and Auger electrons. Measurements were made by means of 4πβ-γ- coincidence counting with LS and PC detectors in β channels.

Coincidence method Basic principle disintegrations with α-, β- or x-ray and Auger electrons emission plus γ-rays; β-, γ- and coincidence-channels in detection system; the activity follows from the counting rates; a priori knowledge of detection efficiency is not needed. A primary method!

Coincidence method The efficiency extrapolation method The β-channel detector sensitive to the gamma-transition. The efficiency extrapolation method developed to compensate for: the γ-sensitivity in the β-channel, problems induced by conversion electrons, additional coincidences from Compton scattering in the β-channel. The variation of the β-efficiency the activity obtained from the extrapolation.

4πβ(LS)-γ coincidence method the standard coincidence method a proportional counter the new coincidence technique a liquid scintillation detector D, DG T, TG G,F, F, S β-double coincidence β-triple coincidence β-γ-coincidence D=AB or BC or AC T=ABC DG = (AB or BC or AC)G TG = ABCG

Detection systems 4πβ(PC)-γ coincidence system M1 M1 samples consist of: VYNS (polyvinylchloride-polyvinylacetate co-polymer) foils; M1 system: a proportional counter (PC) in the β-channel; two scintillation detectors (NaI) in the γ-channel; samples are changed automatically. coated with a thin Au-Pd layer; mounted on stainless steel rings; pre-treated by electrospraying with a colloidal SiO 2 suspension; the radioactive solution is deposited on the center of the sources and then slowly dried; β-efficiency variation by absorption foils.

Detection systems 4πβ(LS)-γ coincidence system M29 M29 system: a liquid scintillation (LS) counter in the β-channel; an automated sample changer; a scintillation detector (NaI) in the γ-channel; β-γ coincidence and TDCR measurements are possible simultaneously. M29 samples: mixing the radioactive solution directly into the scintillation cocktail (Ultima Gold TM ); low-potassium borosilicate glass or polyethylene vials; β-efficiency variation by chemical quenching (nitromethane-ch 3 NO 2 ) or/and by the neutral density greyfoil filters.

The activity measurement 4πβ(LS)-γ coincidence counting Two sets of samples: set (1) 3 samples, the efficiency variation was made by chemical quenching (nitromethane from 0 μl to 60 μl) combined with ND filters, with light transmission from 69.3% to 6.6%, foils were placed around and under the vial; set (2) 8 samples, the efficiency variation was made by pure chemical quenching (nitromethane from 0 μl to 210 μl). The γ energy window was adjusted to use three different settings: 1) a lower threshold to cover the whole energy spectrum, 2) a window to cover the full-energy peak of one of the most intensive line 898.042 kev 3) a window to cover the full-energy peak of the 898.042 kev line together with the Compton continuum. The coincidence resolving times: 77.5 ns and 76.5 ns for the β channel and for the β-γ coincidences, respectively. The common extendable dead time 30 μs. A delay time of 2.212 μs was used for the β signals to compensate for the slower signal processing of the γ channel.

The activity measurement 4πβ(LS)-γ coincidence counting Set (1) of samples: a = (100.69 ± 0.69) kbq/g Set (2) of samples a = (100.79 ± 0.58) kbq/g Final result a = (100.75 ± 0.58) kbq/g The counting rates of β, γ and β-γ coincidence events, were used to plot the ratio N β N γ /N c as a function of x = (1-N c /N γ )/(N c /N γ ). The counting rate determination includes: the background subtraction, the decay correction to the reference date, and the correction for decay during the measurement. 2 nd deg. polynomial functions were fitted to the experimental data and extrapolated to ε β = 100%. Results were determined for double and triple coincidences in the β channel and for three different γ settings. The activity concentrations were determined separately for each set of samples.

The activity measurement 4πβ(PC)-γ coincidence counting To validate the result from the new PTB custom-built 4πβ(LS)-γ coincidence counting system. Three samples were prepared 10 mg, 15 mg and 22 mg. The β counting efficiency was varied by successive addition of VYNS absorption foils. Both the β and γ channels included dead-time units with the same non-extendable dead time of about 8 μs. To compensate for the delay between the γ and β events, a delay unit was used. The coincidence resolving time: about 1 μs. The γ energy window was adjusted with a lower threshold to cover the whole energy spectrum.

The activity measurement 4πβ(PC)-γ coincidence counting With G1 detector: a = (100.73 ± 0.29) kbq/g With G2 detector a = (100.75 ± 0.27) kbq/g Final result a = (100.74 ± 0.29) kbq/g The counting rates of β, γ and β-γ coincidence events, were used to plot the ratio N β N γ /N c as a function of x = (1-N c /N γ )/(N c /N γ ). The counting rates were corrected for background, decay to the reference date, decay during the measurement and analyzed based on the expression given in Smith D. (Nucl. Instrum. Meth. 152:505-519, 1978) for the coincidence counting rate. 1 st deg. polynomial functions were fitted to the experimental data and extrapolated to ε β = 100%.

Comparison 4πβ(LS)-γ coincidence counting a = (100.75 ± 0.58) kbq/g 4πβ(PC)-γ coincidence counting a = (100.74 ± 0.29) kbq/g 0.01 % relative deviation The good agreement between results from both systems confirms the correct operation of the 4πβ(LS)-γ coincidence counting system.

Uncertainties Component u(a)/a in % 4 (PC)- CC 4 (LS)- CC Counting statistics 0.18 * 0.02 * Weighing 0.08 0.02 Dead time negligible 0.20 Background 0.04 0.03 Resolving time negligible 0.20 Counting time negligible negligible Adsorption 0.05 0.05 Decay correction <0.01 <0.01 Extrapolation of efficiency curve 0.05 0.04 Impurities (no radioactive impurity <0.03 <0.03 detected) Fitting uncertainty 0.19 0.49 Square root of the sum of quadratic components 0.29 0.57

Uncertainties Component u(a)/a in % 4 (PC)- CC 4 (LS)- CC Counting statistics 0.18 * 0.02 * Weighing 0.08 0.02 Dead time negligible 0.20 Background 0.04 0.03 Resolving time negligible 0.20 Counting time negligible negligible Adsorption 0.05 0.05 Decay correction <0.01 <0.01 Extrapolation of efficiency curve 0.05 0.04 Impurities (no radioactive impurity <0.03 <0.03 detected) Fitting uncertainty 0.19 0.49 Square root of the sum of quadratic components 0.29 0.57 This uncertainty might be reduced by using a Monte-Carlo-simulation to provide a proper fitting function.

Uncertainties Component u(a)/a in % 4 (PC)- CC 4 (LS)- CC Counting statistics 0.18 * 0.02 * Weighing 0.08 0.02 Dead time negligible 0.20 Background 0.04 0.03 Resolving time negligible 0.20 Counting time negligible negligible Adsorption 0.05 0.05 Decay correction <0.01 <0.01 Extrapolation of efficiency curve 0.05 0.04 Impurities (no radioactive impurity <0.03 <0.03 detected) Fitting uncertainty 0.19 0.49 Square root of the sum of quadratic components 0.29 0.57 In the 4KAM module, the determination of coincidences for the β channels (double and triple) is realized by a delay line with increments of 2 ns, while the determination of β-γ coincidences is set by the FPGA clock resulting in a jitter and increments of about 5.882 ns. Insufficient knowledge of these differences was the reason to estimate rather large uncertainties for the uncertainty components due to dead time and resolving time. A significant decrease of these uncertainties could be achieved by a more comprehensive analysis of the 4KAM module.

Summary The 4πβ(LS)-γ coincidence counting is a good primary activity standardization technique, and sample preparation is much easier than in classical coincidence counting. The activity measurement of 88 Y is important for the calibration of γ-ray spectrometers. The good agreement between results from the 4πβ(LS)-γ coincidence counting and the 4πβ(PC)-γ coincidence counting confirms the correct operation of the 4πβ(LS)-γ coincidence counting system.

Thank you for your attention