Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations

Similar documents
Compton suppression spectrometry

Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

Researchers at the University of Missouri-Columbia have designed a triple crystal

Efficiency. Calculations for Selected Scintillators. > Detector Counting. Efficiency. > Photopeak Efficiency of Various Scintillation Materials

Gamma ray coincidence and angular correlation

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

SIMPLE USE OF EGS4 BY EXCEL MACRO FILE

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

MCNP analysis and optimization of a double crystal phoswich detector

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT

Chemical Engineering 412

Simulation of 4π HPGe Compton-Suppression spectrometer

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Radiation Detection and Measurement

New Scintillators for the Border Monitoring Equipment

Gamma-Spectrum Generator

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Gamma and X-Ray Detection

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

Chapter 4 Scintillation Detectors

OPPORTUNITY TO JOIN IEEE AND NPSS

Photoneutron reactions studies at ELI-NP using a direct neutron multiplicity sorting method Dan Filipescu

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Gamma Spectroscopy. References: Objectives:

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Development of a Hard X-Ray Polarimeter for Solar Flares and Gamma-Ray Bursts

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

Figure 1. Decay Scheme for 60Co

1 of :32

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

ADVANCED SYNTHETICALLY ENHANCED DETECTOR RESOLUTION ALGORITHM: A SYSTEM FOR EXTRACTING PHOTOPEAKS FROM A SODIUM IODIDE SCINTILLATION DETECTOR SPECTRUM

Mass of the electron m 0

Slides by: Prof. Abeer Alharbi

Basic hands-on gamma calibration for low activity environmental levels

The Best Gamma-Ray Detector

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

ORTEC AN34 Experiment 10 Compton Scattering

Inorganic Scintillators

A New Look at the Galactic Diffuse GeV Excess

Quality Assurance. Purity control. Polycrystalline Ingots

Energy Resolution of Scintillation Detectors New Observations

RESPONSE FUNCTION STUDY FOR ENERGY TO LIGHT CONVERSION IN ORGANIC LIQUID SCINTILLATORS

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

III. Energy Deposition in the Detector and Spectrum Formation

Non-proportionality of organic scintillators and BGO

ISPA-Tubes with YAP:Ce Active Windows for X and Gamma Ray Imaging.

Sample Spectroscopy System Hardware

Preparatory experiments for cold-neutron induced fission studies at IKI

Scintillation Detector

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Improving Scintillation Response in Xenon and Implementation in GEANT4

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Radioactivity. Lecture 6 Detectors and Instrumentation

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

Scintillation Detectors

Detection of γ-rays from nuclear decay: 0.1 < E γ < 20 MeV

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications

Analysis of γ spectrum

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Pulse Shape Analysis

SCINTILLATION DETECTORS AND PM TUBES

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Indian J.Sci.Res.2(3) :25-32, 2011

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Interaction of charged particles and photons with matter

Scintillation Detectors

COMPTON SCATTERING OF GAMMA RAYS

In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Digital Gamma-ray Spectroscopy & Imaging with Semiconductor detectors Frontiers of gamma-ray spectroscopy

Radiation (Particle) Detection and Measurement

in Cross-Section Data

Introduction Polarimeters at MAMI Analysis Future Conclusion. Polarimetry at MAMI. V. Tyukin, Inst. of Nuclear Physics, Mainz, Germany

DETECTORS. I. Charged Particle Detectors

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

Timing resolution measurement of a 3 Lanthanum Bromide detector

Positron Emission Tomography

2 Improvements of Programs 2.1 L-X ray Energy dependent L subshell cross sections In 1996 version of L-X ray calculation [5], the ratio of L sub

Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector

arxiv: v1 [physics.ins-det] 29 Jun 2011

Transcription:

Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations 271 I. ORION 1,2 1 AND L. WIELOPOLSKI 2, 1 St. Luke=s/Roosevelt Hospital, Columbia University, New York, NY 10025,USA 2 Brookhaven National Laboratory, Department of Applied Science, Upton, NY 11973, USA IN VIVO BODY COMPOSITION STUDIES Annals of the New York Academy of Science (May 2000; 904: 271-5) INTRODUCTION The high efficiency of the BGO detectors makes them a very attractive candidates to replace NaI(Tl) detectors, which are widely used in studies of body composition. In this work, the response functions of the BGO and NaI(Tl) detectors were determined at 0.662, 4.4, and 10.0 MeV using three different Monte Carlo codes; 1, 2, and PHOTON 3. These codes differ in their input files and transport calculations, and were used to verify the internal consistency of the set-up and of the input data. The energy range of 0.662-10 MeV was chosen to cover energies of interest in body-composition studies. The superior efficiency of the BGO detectors has to be weighed against their inferior resolution, and their higher price than that of the NaI detectors. Since the price of the BGO detectors strongly depends on the crystal s size, its optimization is an important component in the design of the entire system. METHOD The intrinsic response functions were calculated for a point source at 10 cm from the base of a cylindrical detector, where the source emission was forced into the solid angle subtended by the detector. The energy deposition in the detector, i.e., the photon spectrum, was calculated using 512 bins 20 kev wide. Special attention was paid to the energy cut-off used in the codes. While this value is fixed at 10 kev in the PHOTON code, to improve accuracy we altered it in the other two codes to 1 kev. The main effect of the cut-off value is in the valley area beside the photopeak (Fig 1a.). The calculated photon spectra were convoluted with an energy-resolution function of the type a+be ½ +ce that was fitted to the experimental data for each detector reported in the literature 4. The convolution consisted of Monte Carlo normal sampling of each energy bin. We found that from about 2 MeV and above, the results from the PHOTON code deviated systematically from the results obtained using the other two codes (Figs. 1a, 1b, and 1c). We attributed this discrepancy to inadequate bremsstrahlung cross-sections used in the PHOTON code (we are verifying this with the authors of the code). The Photon code was not used from further at high energies. 1 Address for corresponding: Itzhak Orion, Ph.D., Brookhaven National Laboratory, Medical Department, Bell Ave. Upton, NY 11973. Voice: 516 344-3653; fax: 516 344-5311. orion@bnl.gov

272 10 6 a) Source Energy: 0.662 MeV 10 5 NaI(Tl) 6"x6" PHOTON 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 E (MeV ) 10 6 b) Source Energy : 4.4 M ev 10 5 N a I( Tl) 6" x6" PH OTON 0 1 2 3 4 5 c) Source E nergy: 10 Me V NaI(Tl) 6"x6" Yield (per 10 5 histories) PH OTON 10 0 0 1 2 3 4 5 6 7 8 9 1 FIGURE 1. Comparison between the results of 4B2,, and PHOTON Monte Carlo codes: (a) at a source energy of 0.662 MeV; (b) at 4.4 MeV; and (c) at 10 MeV.

273 BGO 5"x3" FWHM = 5.4 % BGO 3"x3" FWHM = 4.3 % NaI(Tl) 6"x6" F WHM = 3.8 % 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 FIGURE 2. Simulated 4.4 MeV pulse-height distribution with the specific energy-resolution function for BGO 5 x 3, BGO 3 x 3, and NaI(Tl) 6 x 6 using the Monte Carlo code. BGO 5"x3" FWHM = 3.6 % BGO 3"x3" FWHM = 3.4 % NaI(Tl) 6"x6" FWHM = 3.3 % Yield (per 10 5 histories) 0 1 2 3 4 5 6 7 8 9 1 FIGURE 3. Simulated 10 MeV pulse-height distribution with the specific energy-resolution function for BGO 5 x 3, BGO 3 x 3 and NaI(Tl) 6 x 6 using the Monte Carlo code. DISCUSSION AND CONCLUSIONS At high energies, the higher energy resolution of the BGO detectors had little effect on the distribution of pulse height compared to that of the NaI detectors (Figs. 2 and 3). However, the high efficiency of the BGO detectors led to a higher peak-to-background ratio (Fig. 4). At higher

274 1.0 0.9 0.8 0.7 Peak / Total Ratio 0.6 BGO 6" x 6" 0.5 6" 3" 5" x 3" 0.4 4" 4" 3" x 6" 3" x 3" 0.3 6" x 2/3" 2" x 6" 0.2 NaI(Tl) 6" x 6" 0.1 3" x 3" 0.0 0 1 2 3 4 5 6 7 8 9 1 Incident Photon Energy (MeV) by size. FIGURE 4. Calculated peak-to-total ratios versus incident photon energy for each detector by type and TABLE 1. The yield per history of the single-escape (SE) and double-escape (DE) peaks and the photopeakto-escape-peak ratios at two source energies. E = 4.4 MeV E = 10 MeV Detector P/N SE/N DE/N P/SE P/DE P/N * SE/N * DE/N * P/SE P/DE NaI(Tl) 3 x3 0.082 0.051 0.019 1.6 4.2 0.031 0.045 0.016 0.7 1.9 6 x6 0.218 0.066 0.009 3.3 25.7 0.124 0.080 0.009 1.7 13.4 BGO 3 x3 0.354 0.074 0.006 4.8 62.3 0.240 0.078 0.007 3.1 35.1 5 x3 0.427 0.064 0.004 6.7 112.4 0.323 0.070 0.004 4.6 72.6 ABBREVIATIONS: N = 10 6, Number of histories; N * = 10 5, Number of histories; P = photopeak counts; SE = single escape peak net counts; DE = double escape peak net counts.

275 0.9 0.8 0.7 2"x6" 4"x4" 3"x3" 3"x6" 6"x 2 / 3 " 5"x3" 6"x3" 0.662 MeV 6"x6" E f (per incident photon) 0.6 0.5 0.4 0.3 4.4 MeV 10 MeV 0.2 0.1 Fit to E f =(A1-A2)/(1-V 0.3 3 )+A2 A1 A2-0.53±0.17 0.91±0.02-2.63±0.31 0.74±0.04-2.68±0.33 0.64±0.04 0.0 0 500 1000 1500 2000 2500 3000 V (cm 3 ) FIGURE 5. The BGO detector s intrinsic efficiency (E f ) dependence on volume fitted to the logistical equation (the parameters are listed for three incident-photon energies). energies the shape of the Compton region is almost independent of the type and the size of the detector. For the BGO detector, the photopeak intrinsic efficiency versus detector volume is effectively represented by a single logistical equation of the form A1 A 2 E f = + A 2, 0.33 1 V as shown in Fig. 5. The main differences in the response function for the detectors of different size and type were in the escape peaks, (see TABLE 1). In summary, the BGO detectors, despite their worse energy resolution, appear to be superior to NaI(Tl) detectors for high-energy gamma detection. REFERENCES 1. Nelson, W. R., H. Hirayama, and D. Rogers. 1985. The Code. SLAC-265. 2. Briesmeister, J.F. Ed. 1997. The Transport Code, Version 4B. LA-12625-M. 3. Puzovic, J. M. and I. V. Anicin. 1998. User-friendly Monte Carlo program for the generation of gamma-ray spectral responses in complex source-detector arraignments. NIM A 414: 279 282. 4. Heat, R. L. 1964. Scintillation Spectrometry Gamma-Ray Spectrum Catalogue. Vol. 2, 2 nd edition. U.S. Atomic Energy Commission IDO- 16880-2.