Determination of the timing properties of a pulsed positron lifetime beam by the application of an electron gun and a fast microchannel plate

Similar documents
The intense, pulsed positron source EPOS at the Research Centre Dresden-Rossendorf

DETECTORS. I. Charged Particle Detectors

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

positron source EPOS - general concept - timing system - digital lifetime measurement

(Positron Annihilation Lifetime Spectroscopy; PALS) 1) 4) PALS

arxiv:physics/ v1 3 Aug 2006

Positron Beam Lifetime Spectroscopy at

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

EEE4106Z Radiation Interactions & Detection

Advances in the Micro-Hole & Strip Plate gaseous detector

Development status of the positron lifetime beam MePS and the first lifetime measurements of porous ultra-low-k dielectrics with MePS

Capacitive Pick-Up Type DB 040

Assembly and test runs of decay detector for ISGMR study. J. Button, R. Polis, C. Canahui, Krishichayan, Y. -W. Lui, and D. H.

SRF GUN CHARACTERIZATION - PHASE SPACE AND DARK CURRENT MEASUREMENTS AT ELBE*

Chapter 4 Scintillation Detectors

New Concept of EPOS Progress of the Mono-energetic Positron Beam (MePS) Gamma-induced Positron Spectroscopy (GiPS)

Beam Diagnostics Lecture 3. Measuring Complex Accelerator Parameters Uli Raich CERN AB-BI

PoS(PD07)020. Timing Properties of MCP-PMT. Kenji Inami. Nagoya university, Nagoya, Japan

A gas-filled calorimeter for high intensity beam environments

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

STATUS OF THE NOVOSIBIRSK ENERGY RECOVERY LINAC

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

Neutron Transport Calculations Using Monte-Carlo Methods. Sean Lourette Fairport High School Advisor: Christian Stoeckl

Recent Status of Polarized Electron Sources at Nagoya University

Theory English (Official)

Basic structure of SEM

Photoelectric Effect Experiment

The intense Positron Source EPOS at ELBE Radiation Source of Research Center Rossendorf

First Beam Profile Measurements Based on Light Radiation of Atoms Excited by the Particle Beam

CCD readout of GEM-based neutron detectors

v = E B FXA 2008 UNIT G485 Module Magnetic Fields BQv = EQ THE MASS SPECTROMETER

Radioactivity. Lecture 6 Detectors and Instrumentation

Cherenkov Detector. Cosmic Rays Cherenkov Detector. Lodovico Lappetito. CherenkovDetector_ENG - 28/04/2016 Pag. 1

Chopping High-Intensity Ion Beams at FRANZ

Homework 2: Forces on Charged Particles

PoS(TIPP2014)093. Performance study of the TOP counter with the 2 GeV/c positron beam at LEPS. K. Matsuoka, For the Belle II PID group

A Multi-beamlet Injector for Heavy Ion Fusion: Experiments and Modeling

Slow-Positron-Beam Techniques

Influence of gas conditions on electron temperature inside a pinch column of plasma-focus discharge

Weak-Beam Dark-Field Technique

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Lecture 22 Ion Beam Techniques

Beam Shape and Halo Monitor Study

EXPERIMENT 2-6. e/m OF THE ELECTRON GENERAL DISCUSSION

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Outlook: Application of Positron Annihilation for defects investigations in thin films. Introduction to Positron Annihilation Methods

Ion Implanter Cyclotron Apparatus System

Modeling Electron Emission From Diamond-Amplified Cathodes

pp physics, RWTH, WS 2003/04, T.Hebbeker

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

Juliane Rönsch Hamburg University. Investigations of the longitudinal phase space at a photo injector for the X-FEL

Positron-Electron Annihilation

Research Center Dresden Rossendorf

(Total for Question = 5 marks) PhysicsAndMathsTutor.com

MONTE-CARLO SIMULATIONS OF TIME- RESOLVED, OPTICAL READOUT DETECTOR for PULSED, FAST-NEUTRON TRANSMISSION SPECTROSCOPY (PFNTS)

Multiparameter optimization of an ERL. injector

Update on Development of High Current Bunched Electron Beam from Magnetized DC Photogun

Novel Plastic Microchannel-Based Direct Fast Neutron Detection

Nuclear Instruments and Methods in Physics Research A

!"#$%$!&'()$"('*+,-')'+-$#..+/+,0)&,$%.1&&/$ LONGITUDINAL BEAM DYNAMICS

A Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II

Beam Diagnostics. Measuring Complex Accelerator Parameters Uli Raich CERN BE-BI

EPOS an intense positron beam project at the Research Center Rossendorf

Simulation Studies for a Polarimeter at the International Linear Collider (ILC)

Development status of non-destructive assay of nuclear material by using laser Compton scattered gamma-rays

IPBI-TN June 30, 2004

Differential Cross Section Measurements in Ion-molecule Collisions

Revision Guide. Chapter 7 Quantum Behaviour

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

DESIGN OF A WIEN FILTER AND MEASUREMENT OF LONGITUDINAL POLARIZATION OF BETA PARTICLES BY S. S. ABHYANKAR AND M. R. BHIDAY

Study of DC Cylindrical Magnetron by Langmuir Probe

Gaetano L Episcopo. Scanning Electron Microscopy Focus Ion Beam and. Pulsed Plasma Deposition

Instruction Manual for EP-20 e/m of the Electron Apparatus

Scintillation Detectors

Relativistic Electrons

Scintillation Detectors

1 Introduction. KOPIO charged-particle vetos. K - RARE Meeting (Frascati) May Purpose of CPV: veto Kl

CHARGED PARTICLES IN FIELDS

The CeB6 Electron Gun for the Soft-X-ray FEL Project at SPring-8

Search for an electric dipole moment with polarized beams in storage rings

Dense plasma formation on the surface of a ferroelectric cathode

Study of the hyperfine structure of antiprotonic helium

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Longitudinal Dynamics

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

High quantum efficiency S-20 photocathodes for photon counting applications

Experiment 5 Deflection of Electrons

A measurement of the air fluorescence yield

Longitudinal Measurements at the SLAC Gun Test Facility*

Response function measurements of an NE102A organic scintillator using an 241 Am-Be source

Stability and loss in an ion-trap resonator

Radionuclide Imaging MII Detection of Nuclear Emission

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

Single bunch longitudinal measurements at the Cornell Electron-Positron Storage Ring

PoS(PhotoDet 2012)010

Time of Flight Technique

Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI)

A Comparison between Channel Selections in Heavy Ion Reactions

Transcription:

Applied Surface Science 252 (2006) 3148 3153 www.elsevier.com/locate/apsusc Determination of the timing properties of a pulsed positron lifetime beam by the application of an electron gun and a fast microchannel plate A. Laakso *, A. Pelli, K. Rytsölä, K. Saarinen, P. Hautojärvi Laboratory of Physics, Helsinki University of Technology, P.O. Box 1100,FI-02015 Hut, Finland Available online 3 October 2005 Abstract We show that the timing properties of a pulsed low-energy positron lifetime beam can be conveniently tested by an electron beam. We apply this method to study the time resolution of the beam and electron scattering in flat and sawtooth shaped choppers. The results show that (i) time resolution of 160 ps is obtained, (ii) the scattering of the electrons and the secondary electron yield of the flat chopper make the time resolution worse and background poor, and (iii) both these problems can be solved by using a sawtooth shaped chopper. We also compare these results to beam simulations. # 2005 Elsevier B.V. All rights reserved. Keywords: Positron spectroscopy; Pulsed positron beam; Electron scattering 1. Introduction Positron lifetime spectroscopy is a powerful technique to study vacancies and other open volume type defects in solid materials, for example in semiconductors. Conventional positron lifetime spectroscopy based on using b + -active isotopes (e.g. 22 Na) is limited to thick samples and is not applicable to thin * Corresponding author. Tel.: +358 9 4515801; fax: +358 9 4513116. E-mail address: antti.laakso@hut.fi (A. Laakso). samples or thin layers. To expand the applicability of lifetime spectroscopy, the positron lifetime beam has been developed [1 3]. The timing properties of the beam can be determined with positrons and a suitable target, e.g. stainless steel, using a scintillator and a photomultiplier tube (PMT) as a detector. When the system is tested it is advantageous if the measurement times are short and the resolution good. This can be accomplished if the positron source is replaced with an electron gun and a fast microchannel plate (MCP) is used as a detector. 0169-4332/$ see front matter # 2005 Elsevier B.V. All rights reserved. doi:10.1016/j.apsusc.2005.08.057

A. Laakso et al. / Applied Surface Science 252 (2006) 3148 3153 3149 2. Setup and methods 2.1. The pulsed positron lifetime beam The general structure of the magnetically guided pulsed positron beam in HUT Laboratory of Physics is described elsewhere [1,4] and only the main details are repeated here. The source, velocity selector and the beam bunching and chopping stages float at high voltage whereas the sample is at ground potential. The source area comprises a 22 Na source followed by a W- moderator. Slow positrons are separated from fast ones with a 908 bend in the beam. The first pulsing stage, the prebuncher, is an unmatched, nonresonant, two frequency (33.3 and 66.6 MHz) double gap buncher, followed by a drift tube and a pre-accelerator where positrons reach an energy of 1 kev. After this acceleration the beam is chopped in a deflection type chopper shown in Fig. 1, where a deflecting field drives the positrons into the walls of the chopper inbetween the pulses. The chopper is described in more detail in Section 2.3. The chopped beam is then bunched in the main buncher at a frequency of 166.7 MHz. The final acceleration will be done with an accelerator decelerator stage. 2.2. Electron test setup The testing of the system requires that the timing spectra can be obtained fast and accurately. With conventional isotope sources and moderators and a BaF 2 scintillator detectors the count rate is limited to hundreds of counts per second resulting in relatively long measurement times. In our case it is also difficult to provide the necessary magnetic shielding for the photomultiplier tube of the detector in the test configuration of the beam. To overcome these limitations we developed a new method for the testing of timing properties of a pulsed positron beam. We used a small, low-temperature electron source to obtain a narrow electron beam with high brightness. This enables to obtain easily high, controllable count rates. As a detector we used a microchannel plate, which is insensitive to magnetic fields. Other additional advantages of this setup are that (i) the energy distribution of beam is narrower, (ii) the beam diameter is much smaller compared to that generated from a positron source and (iii) the time resolution of the MCP detector is better than the resolution of a scintillator detector. The electron source was a self-made electron gun where a BaO-disc (Kimball Physics ES-015 BaO cathode) was used as the electron emitter. An adjustable slit was placed 250 mm away from the electron gun to reduce the electron current to sub falevel during timing tests. The e-gun was also used for beam alignment, where a phosphor screen and a picoammeter were used as beam monitors. The timing detector, a microchannel plate (Hamamatsu F4655-12), was mounted next to the buncher and floating at 2.6 kv voltage. The accelerator decelerator structure was not employed. It was observed that the MCP was very sensitive to the residual gas ions and electrons created in the ion pumps. Therefore, the system was pumped only with turbopumps during the tests. The stop signal from the MCP was connected to the timing electronics through a insulated BNC-feedthrough and double DC-block (both signal and ground connection are isolated) made of high-voltage high-q ceramic capacitors in a coaxial arrangement. The start-signal was obtained from the pulsing electronics via a fiber optic link. 2.3. The chopper Fig. 1. The sawtooth grooves in chopper reduce the number of scattered electrons passing through to detector. The flat portions between the slopes and ridges are necessitated by the machining. According to simulations, these do not change the performance of the chopper. The lines illustrate simulated trajectories of positrons. The chopper is made of two grounded sandwiched copper blocks with a thin double sided printed circuit board in between them, see Fig. 1. The beam passes through the chopper via a slit machined to both copper blocks. An isolated area in the thin central copper plate acts as an electrode at the entrance end of the chopper. This electrode is part of a 16.67 MHz resonator circuit

3150 A. Laakso et al. / Applied Surface Science 252 (2006) 3148 3153 which creates a sinusoidal electric field perpendicular to the beam. The phase of the field is adjusted so that the positron bunch from the prebuncher passes the electrode at the zero crossings (zero field) letting the bunch straight through the chopper. Between the bunches, the electric field increases the transversal momentum of the positron, which in turn increases the radius of larmor precession resulting the positrons to collide with the chopper walls. The chopper time acceptance window can be adjusted by varying the slew rate at the zero crossing. This is simply done by adjusting the amplitude of the chopping RF-signal. When low energy electrons are used to test the system care must be taken to avoid disturbance caused by the electron scattering. This is especially important in the chopper that is based on the beam colliding with the chopper walls by deflection. In this collision there is a relatively large probability for scattering and secondary electron production when the energy is in the 1 kev range [5,6]. These electrons, which have a lower energy than those in the original beam, are detrimental to the time resolution of the beam. To reduce the number of these electrons we designed a chopper which is opaque to the scattered and secondary electrons. Two geometrically different choppers were designed and studied experimentally and the results were compared to simulations [7]. The first chopper was called as the flat chopper and the second the sawtooth chopper. Both choppers consist of a deflection plate as described earlier, followed by a long slit machined to the copper body of the chopper. In the flat chopper the slit is smooth, while in the sawtooth chopper the longitudinal cross-section of the slit resembles a sawtooth (see Fig. 1). This structure consists of tilted slopes and six ridges perpendicular to the beam minimizing the probability for deflected electrons to pass through the chopper. The angle of the tilted surface and the height of the ridge were chosen so that the mirror-like reflection does not occur in the forward direction more than once. In addition to the reflected electrons there are always secondary electrons emitted with low velocities to nearly random directions so that they can pass over the ridge and scatter again. However, the probability is small for an electron to pass through all six ridges and exit the chopper. This kind of sawtooth surface has been used in large proton and positron accelerators to reduce photoelectron emission from accelerator chamber surfaces [8]. 3. Results 3.1. Results from the measurements 3.1.1. The flat chopper and the chopping power The electron scattering effects in the flat chopper are illustrated by the timing spectra is shown in Fig. 2. In this figure the timing spectra for a beam that was only chopped is shown. With a low amplitude of the field in the chopper (dashed curve) the peak is broad and the background is high. When the field amplitude is increased the peak gets narrower and the background becomes asymmetric (solid curve) showing a larger number of electrons arriving earlier (right hand side). This asymmetry can be understood if the spectrum is thought to be composed of two spectra: the chopped spectrum and the background spectrum due to scattered electrons. The electrons in the chopped spectrum have passed through the chopper with unmodified energy when the chopping field is negligible in the chopping window and no scattering Fig. 2. Increasing the chopping power from 5 to 50 W improves pulse width but changes the pulse shape due to scattering and secondary electron effects.

occurs. The background spectrum is from those electrons that arrive outside chopping window and have been deflected and subsequently scattered. In the inelastic scattering the electrons lose energy and the flight time to the target increases. This causes the background spectrum to shift to later times i.e., to the left and the hole in the background spectrum (due to chopped spectra) appears on the left hand side of the main peak. This explains the shoulder on the right hand side of the main peak, which is due to electrons scattered during the previous cycle of the deflecting field. This effect is not visible in the low amplitude spectrum since the chopping window is large and the chopping amplitude too low so that some electrons can pass through the chopper all the time and the number of inelastically scattered electrons is low. Thus, the background due to scattering is constant. From these results, it is evident that the flat chopper is not suitable for lifetime beam testing if electrons are used as test particles. A. Laakso et al. / Applied Surface Science 252 (2006) 3148 3153 3151 3.1.2. The sawtooth chopper The results from the measurements with the sawtooth chopper and using all pulsing components are shown in Fig. 3. The amplitudes and phases of different pulsing components were adjusted to yield as narrow final bunches as possible. The topmost curve shows how the prebuncher compresses electrons into the chopping time window increasing the efficiency of the beam (curve a). The full width at half maximum (FWHM) is 3.1 ns and the asymmetry of the peak can be explained by the phase difference between f and 2f components in the prebuncher waveform. The spectrum shown by the curve b is collected when only the chopper is running. It shows that the chopper is working: the FWHM is 2.3 ns, the background is flat and low and the peak-to-background ratio is approximately 100:1. When the prebuncher and the chopper are operating simultaneously, the peak-to-background ratio improves to 300:1 and FWHM of the peak is 1.8 ns, as seen from the curve c. The spectrum recorded with only the main buncher is shown by curve d in Fig. 3. The periodic structure of five similar peaks is due to the 6 ns main bunching period in a 30 ns base time window (166.67 MHz bunching rate versus 33.33 MHz base frequency). The width of an individual peak is good, 160 ps, but the background is high (P/B ratio of 30:1). Fig. 3. The spectra measured from bunchers and the chopper and their combinations show that a time resolution of 160 ps, is obtainable with the sawtooth chopper: (a) prebuncher, (b) chopper, (c) prebuncher and chopper, (d) buncher and (e) all together. The FWHM for each component and combinations of them is given in the figure. When all pulsing stages are operating, the spectrum of Fig. 3 curve e is obtained. The signal phases of different pulsing components have been adjusted so that the middlemost bunching peak matches the prebuncher and chopper time window. The FWHM of the final peak is 160 ps. The peak-tobackground ratio is good, better than 1000:1. The background is not completely flat, due to the five main bunching periods during one base operating cycle. This uneven background should not be a problem for data analysis, since its shape is not dependent on the acceleration energy and thus it can be subtracted from the data. If the peak positions in the spectra in Fig. 3 are compared, a small peak shift is observed when the prebuncher is on. This is due to a small additional energy introduced by RF-acceleration in the prebuncher because of nonoptimal initial energy of the beam [9].

3152 A. Laakso et al. / Applied Surface Science 252 (2006) 3148 3153 Fig. 4. The time spectrum measured with the flat (left) and sawtooth (right) chopper when all pulsing components are on. The FWHM 160 ps is good in both cases, but the background and peak-to-background ratio is relatively poor for flat chopper. 3.1.3. Comparison between the sawtooth and the flat chopper The measured timing spectra using a flat chopper and sawtooth chopper are compared in Fig. 4 (the right hand side is the same as curve e of Fig. 3, but now with linear scale). The superiority of the sawtooth chopper over the flat one is evident: the flat chopper spectrum shows side peaks, the peak-tobackground ratio 1:10 is low and there is a significant nonlinear background. None of these problems are present in the spectra obtained with the sawtooth chopper. 3.2. Results from simulations Computer simulations were done to compare the properties of the two chopper designs [7]. The main challenge with simulation is the treatment of the scattering in a proper way. Theoretical and experimental data of scattering probabilities and secondary electron yields were found from the literature [5,6]. Also, detailed studies using a model from [7] were done. As a result it was found that a simple model where 33% of electrons colliding with the wall scatter elastically and 67% inelastically is sufficient to Fig. 5. Both the measured and the simulated spectra for the flat and the sawtooth chopper show that the introduction of the grooves improves time distribution significantly.

A. Laakso et al. / Applied Surface Science 252 (2006) 3148 3153 3153 describe the scattering. The elastically scattered electrons were assumed to reflect from the surface, while the inelastic ones lose on average 40% of their energy in the scattering [7]. The direction of the inelastic scattering is randomized but weighted in the calculation so that it points to the direction of forward scattering. Measured and simulated spectra for the two chopper designs are compared in Fig. 5 (the measured and calculated time distributions here are for a beam that is only chopped with a low power RF-amplifier). It is clearly seen from the measured data (left in Fig. 5) that the sawtooth chopper gives significantly better results compared to the flat chopper: the background is negligible and the peak is narrow. This result is verified by simulations seen in the right hand side of Fig. 5, which replicates the measured spectra relatively well. These results suggest that the electron scattering and secondary electron generation plays an important role in the shape and the width of the final timing spectra. 4. Conclusions and discussion We have shown that it is possible to test pulsing properties of a pulsed positron lifetime beam with electrons. For successful electron tests the beam chopper should be designed properly; in this case a sawtooth design was used to prevent electron scattering through the chopper. This design improves the system performance also with positrons by reducing positron scattering. We also show that it is possible to simulate electron scattering in the beam realistically and estimate its effect on the final time distribution. The electron gun and an MCP proved to be an efficient way to study timing properties of a pulsed beam and to obtain accurate results rapidly. Tests showed that our system is working in the expected way with electrons and we can obtain a FWHM of 160 ps for the peak and a peak-to-background ratio better than 1000:1 in its test configuration. Acknowledgements The authors thank Dr. Mikko Hakala for valuable discussions about electron and positron scattering. The financial support from the Magnus Ehrnrooth Foundation is acknowledged. References [1] A. Laakso, K. Saarinen, P. Hautojärvi, Physica B 308 310 (2001) 1157. [2] D. Schödlbauer, P. Sperr, G. Kögel, W. Triftshauser, Nucl. Instr. Meth. B 34 (1988) 258. [3] R. Suzuki, Y. Kobayashi, T. Mikado, H. Ohgaki, M. Chiwaki, T. Yamazaki, T. Tomimasu, Jpn. J. Appl. Phys. 30 (1991) 532. [4] K. Fallström, T. Laine, Appl. Surf. Sci. 149 (1999) 44. [5] S.A. Schwarz, J. Appl. Phys. 68 (1990) 2382. [6] W.S.M. Werner, I.S. Tilinin, M. Hayek, Phys. Rev. B 50 (1994) 4819. [7] A. Laakso, M.O. Hakala, A. Pelli, K. Rytsölä, K. Saarinen, P. Hautojärvi, Mater. Sci. Forum 445 446 (2004) 489. [8] Y. Suetsugu, M. Tsuchiya, T. Nishidono, N. Kato, N. Satoh, S. Endo, T. Yokoyama, J. Vac. Sci. Technol. A 21 (2003) 186. [9] P. Sperr, G. Kögel, Mater. Sci. Forum 255 257 (1997) 109.