Comparative Gamma Spectroscopy with SrI2(Eu), GYGAG(Ce) and Bi-loaded Plastic Scintillators

Similar documents
Non-proportional response and energy resolution of pure SrI 2 and SrI 2 :5%Eu scintillators

New Scintillators for the Border Monitoring Equipment

Applications of Pulse Shape Analysis to HPGe Gamma-Ray Detectors

The Scintillation properties of Pr 3+ doped and Pr 3+, Ce 3+

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

Plutonium 239 Equivalency Calculations

Capabilities for Testing the Electronic Configuration in Pu

OPPORTUNITY TO JOIN IEEE AND NPSS

Keywords: scintillator detectors, nonproportionality, gamma-ray detectors, Onsager mechanism, charge carriers, excitons

Design of a Lanthanum Bromide Detector for TOF PET

Development of a High Intensity EBIT for Basic and Applied Science

Fission-Fusion Neutron Source

arxiv: v1 [physics.ins-det] 29 Jun 2011

S.A. Kreek J.E. Kammeraad D. Beckedahl G.J. Schmid J.J. Blair B. Payne A. Friensehner

A Germanium Detector with Optimized Compton Veto for High Sensitivity at Low Energy

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Neutron Detector Gamma Insensitivity Criteria

Using the X-FEL to understand X-ray Thomson scattering for partially ionized plasmas

August 3,1999. Stiffness and Strength Properties for Basic Sandwich Material Core Types UCRL-JC B. Kim, R.M. Christensen.

Improved scintillation proportionality and energy resolution of LaBr 3 :Ce at 80K

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

Non-proportional scintillation response of NaI:Tl to low energy X-ray photons and electrons

Flipping Bits in the James Webb Space Telescope s Cameras

Characterization of the Lanthanum Chloride Scintillation Detector

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Two-Screen Method for Determining Electron Beam Energy and Deflection from Laser Wakefield Acceleration

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Three-Dimensional Silicon Photonic Crystals

Direct Comparison of the X-Ray Emission and Absorption of Cerium Oxide

RESEARCH OPPORTUNITIES AT THE CNS (Y-12 AND PANTEX) NUCLEAR DETECTION AND SENSOR TESTING CENTERS (NDSTC)

Optimization of NSLS-II Blade X-ray Beam Position Monitors: from Photoemission type to Diamond Detector. P. Ilinski

12/16/95-3/15/96 PERIOD MULTI-PARAMETER ON-LINE COAL BULK ANALYSIS. 2, 1. Thermal Neutron Flux in Coal: New Coal Container Geometry

Simulation of Double-Null Divertor Plasmas with the UEDGE Code

Analysis of Shane Telescope Aberration and After Collimation

A Distributed Radiator, Heavy Ion Driven Inertial Confinement Fusion Target with Realistic, Multibeam Illumination Geometry

Use of Gamma Rays from the Decay of 13.8-sec "Be to Calibrate a Germanium Gamma Ray Detector for Measurements up to 8 MeV

Modeling Laser and e-beam Generated Plasma-Plume Experiments Using LASNEX

Efficiency. Calculations for Selected Scintillators. > Detector Counting. Efficiency. > Photopeak Efficiency of Various Scintillation Materials

Response to Comment on "The National Ignition Facility Laser Performance Status"

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

AC dipole based optics measurement and correction at RHIC

Improvement of -ray energy resolution of LaBr3:Ce3+ scintillation detectors by Sr2+ and Ca2+ co-doping

GA A25658 PRECISION X-RAY OPTICAL DEPTH MEASUREMENTS IN ICF SHELLS

J. C. Batchelder', K. S. Toth2, D. M. Moltz3, E. F. Zganjarl, T. J. Ognibene3, M. W. Rowe3, C. R. Binghan12.~,J. Powell3, and B. E.

Solid Phase Microextraction Analysis of B83 SLTs and Core B Compatibility Test Units

Start-up Noise in 3-D Self-AmpMed

The Radiological Hazard of Plutonium Isotopes and Specific Plutonium Mixtures

Characterization of the new scintillator Cs 2 LiYCl 6 :Ce 3+ *

Journal of Asian Scientific Research, 1(3),PP

national security, nonproliferation, defense, border security, and homeland security applications.

Light yield non-proportionality and intrinsic energy resolution of doped CsI scintillators

BaF 2 :Ce POLYMER COMPOSITE GAMMA-RAY SCINTILLATORS M.B. Barta, J.H. Nadler, Z. Kang, B.K. Wagner

sample-specific X-ray speckle contrast variation at absorption edges $ & ~ 0

Partial Discharge Characterization of High-Voltage Cables and Components

Novel Nanoparticles for Ultrasensitive Detection and Spectroscopy

THE COMPTON EFFECT Last Revised: January 5, 2007

GA A27235 EULERIAN SIMULATIONS OF NEOCLASSICAL FLOWS AND TRANSPORT IN THE TOKAMAK PLASMA EDGE AND OUTER CORE

Alex Dombos Michigan State University Nuclear and Particle Physics

AEC Research and Development REPORT

Safety Considerations for Laser Power on Metals in Contact with High Explosives-Experimental and Calculational Results

: Y/LB-16,056 OAK RIDGE Y-12 PLANT

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

On the HalfiLife of LABORATORY ERNEST ORLANDO LAWRENCE BERKELEYNATIONAL. and Particle Astrophysics Nuclear Science and Physics Divisions

Light yield non-proportionality and energy resolution of Lu 1.95 Y 0.05 SiO 5 :Ce and Lu 2 SiO 5 :Ce scintillation crystals

Scintillation Products Technical Note

PROJECT PROGRESS REPORT (03/lfi?lfibr-~/15/1998):

GA A26474 SYNERGY IN TWO-FREQUENCY FAST WAVE CYCLOTRON HARMONIC ABSORPTION IN DIII-D

The Characterization of new Eu 2+ doped TlSr2I5 Scintillator Crystals

A TER. Observation of Xe with the PNNL ARSA System PNNL UC-7 13

ions which are abundant in waste water. Since zeolites are

Multi-Scale Chemical Process Modeling with Bayesian Nonparametric Regression

Parametric Instabilities in Laser/Matter Interaction: From Noise Levels to Relativistic Regimes

Modified Kriging: Evaluation, Modification, Recommendations

Impurity Transport Studies of Intrinsic MO and Injected Ge in High Temperature ECRH Heated FTU Tokamak Plasmas

Measurement and simulation of proton induced activation of LaBr 3 :Ce

CQNl_" RESPONSE TO 100% INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES TO LOW ENERGY ELECTRONS AND IONS

Europium-doped barium halide scintillators for x-ray and -ray detections

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

GA A23736 EFFECTS OF CROSS-SECTION SHAPE ON L MODE AND H MODE ENERGY TRANSPORT

Chemical Engineering 412

Enhanced α-γ Discrimination in Co-doped LaBr 3 :Ce

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE

Plasma Response Control Using Advanced Feedback Techniques

Direct Determination of the Stacking Order in Gd 2 O 3 Epi-Layers on GaAs

GAMMA RAY SPECTROSCOPY

Final Technical Report. Department of Energy. for

Controlling Backstreaming Ions from X-ray Converter Targets with Time Varying Final Focusing Solenoidal Lens and Beam Energy Variation

Magnetic Measurements of the Elliptical Multipole Wiggler Prototype

The Gas Flow from the Gas Attenuator to the Beam Line

GA A26057 DEMONSTRATION OF ITER OPERATIONAL SCENARIOS ON DIII-D

IMAGING OF HETEROGENEOUS MATERIALS BY. P. Staples, T. H. Prettyman, and J. Lestone

The photoneutron yield predictions by PICA and comparison with the measurements

ust/ aphysics Division, Argonne National Laboratory, Argonne, Illinois 60439, USA

Pathway and Kinetic Analysis on the Propyl Radical + O2 Reaction System

STUDY OF CYCLOTRON RESONANCE AND MAGNETO-PHOTOLUMINESCENCE OF N-TYPE MODULATION DOPED InGaAs QUANTUM WELL LAYERS AND THEIR CHARACTERIZATIONS

A New Computational Method for non-lte, the Linear Response Matrix

~ _- IDOCLRXKT DMSP SATELLITE DETECTIONS OF GAMMA-RAY BURSTS. J. Terrell, NIS-2 P. Lee, NIS-2 R W.

Shock compression of precompressed deuterium

Scaling of divertor heat flux profile widths in DIII-D

Transcription:

LLNL-PROC-462914 Comparative Gamma Spectroscopy with SrI2(Eu), and Bi-loaded Plastic Scintillators N. J. Cherepy November 22, 21 IEEE Nuclear Science Symposium Knoxville, TN, United States October 3, 21 through November 6, 21

Disclaimer This document was prepared as an account of work sponsored by an agency of the United States government. Neither the United States government nor Lawrence Livermore National Security, LLC, nor any of their employees makes any warranty, expressed or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States government or Lawrence Livermore National Security, LLC. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States government or Lawrence Livermore National Security, LLC, and shall not be used for advertising or product endorsement purposes.

Submitted to IEEE Transactions on Nuclear Science, IEEE/NSS Proceedings 21 1 Comparative Gamma Spectroscopy with, and Bi-loaded Plastic Scintillators N.J. Cherepy, Member, IEEE, S.A. Payne, Member, IEEE, B.W. Sturm, Member, IEEE, J.D. Kuntz, Z.M. Seeley, B.L. Rupert, R.D. Sanner, O.B. Drury, T.A. Hurst, S.E. Fisher M. Groza, L. Matei, A. Burger, Member, IEEE, R. Hawrami, Member, IEEE, K.S. Shah, Member, IEEE, and L.A. Boatner Member, IEEE Abstract We are developing new scintillator materials that offer potential for high resolution gamma ray spectroscopy at low cost. Single crystal offers ~3% resolution at 662 kev, in sizes of ~1 in 3. We have developed ceramics processing technology allowing us to achieve cubic inch scale transparent ceramic scintillators offering gamma spectroscopy performance superior to. We fabricated a bismuth-loaded plastic scintillator that demonstrates energy resolution of ~8% at 662 kev in small sizes. Index Terms Scintillators, strontium iodide, ceramic scintillators, gamma ray spectrometers, transparent ceramics, plastic scintillators G I. INTRODUCTION amma ray spectroscopy can be used to identify the presence of weak radioactive sources within natural background. The ability to discriminate close-lying spectral lines is strongly dependent upon the energy resolution of the detector. In addition to excellent energy resolution, large volume detectors are needed to acquire sufficient events, for example, to identify a radioactive anomaly moving past a detector. We have employed a directed search methodology for identifying potential scintillator materials candidates [1], resulting in the discovery of Europium-doped Strontium Iodide, [2-4], Cerium-doped Gadolinium Garnet [5,6],, and Bismuth-loaded Polymers. These scintillators possess very low self-radioactivity, offer energy resolution of 3-8% at 662 kev, and have potential to be grown cost-effectively to sizes similar to the most widely deployed gamma spectroscopy scintillator, Thallium-doped Sodium Iodide,. Manuscript received November 21, 21. This work was supported by the Domestic Nuclear Detection Office in the Department of Homeland Security and by the National Nuclear Security Administration, Office of Nonproliferation Research and Development (NA-22) of the U.S.DOE under Contract No. DE-AC3-76SF98, and was performed under the auspices of the U.S. DOE by Lawrence Livermore National Laboratory under Contract DE-AC52-7NA27344. N.J. Cherepy, S.A. Payne, B.W. Sturm, J.D. Kuntz, Z.M. Seeley, B.L. Rupert, R.D.Sanner, O.B. Drury, T.A. Hurst, S.E. Fisher are with Lawrence Livermore National Laboratory, Livermore, CA 9455 (phone: +1-925-424-3492; e-mail: cherepy1@llnl.gov). K.S. Shah is with Radiation Monitoring Devices (RMD), Watertown, MA M. Groza, L. Matei, and A. Burger are with Fisk Univ., Nashville, TN L.A. Boatner is with Oak Ridge National Laboratory, Oak Ridge, TN Fig. 1. New scintillators for gamma ray spectroscopy developed for DHS and DOE. (upper left) SrI 2(Eu) single crystal under UV excitation, (upper right) ceramic, (bottom) two Bi-loaded polymers under UV excitation. In this study, gamma ray spectra of a variety of sources, were obtained employing,, Bi-loaded polymers,, and. The effects of detector size, energy resolution, and background radioactivity (including self-radioactivity) on the ability to distinguish weak sources is quantified, based on a simple model, and qualitatively compared to laboratory data. A. Strontium Iodide II. SAMPLES SrI 2 (5 mol%eu) crystals are being grown at RMD, Oak Ridge National Laboratory and Fisk University using the vertical Bridgman method. The crystals are grown in silica ampoules using anhydrous SrI 2 beads (Aldrich, 99.99%) and EuI 2 powder (Aldrich, 99.9%), as the starting materials [2-4]. Crystals from all three growers have offered similar performance. The crystal characterized in this report was grown at RMD, then packaged at LLNL. The crystal was cut into a tapered cylinder, polished and encapsulated in an aluminum can [7]. B. Gadolinium Garnet Transparent Ceramics Transparent ceramic garnets were formed at LLNL using stoichiometric mixed metal oxide particles synthesized via flame spray pyrolysis (FSP), a nanoparticle production method developed by Pratsinis and co-workers [8] and Laine and co-workers [9]. Synthesis of FSP nanoparticles was carried out at LLNL, as well as by Nanocerox, Inc. The FSP nanoparticles are formed into a green body, vacuum sintered,

counts Counts Counts Counts Counts counts Submitted to IEEE Transactions on Nuclear Science, IEEE/NSS Proceedings 21 2 6 5 4 3 2 1 3 2 1 2 x-rays 2 12.4% 4 4 9.8% Am-241 6 7.7% 6 Cs-137 8 Fig. 2. Gamma spectra acquired with various sources using several scintillator materials. crystal is 2.54 cm x 3.81 cm, LaBr 3(Ce) crystal is 1.8 cm x 2 cm, ceramic is 2.77 cm x.62 cm and the SrI 2(Eu) crystal is an encapsulated tapered cylinder with 2.54 cm x 2.7 cm. 1 4 1 3 1 2 1 1 25 2 15 1 5 9 1 1 2 6.4% 3.5% 11 12 3 13 Ba-133 6.1% 3.3% 2.% 1.9% 4 Co-6 14 then hot-isostatic pressed into optically transparent parts [1]. The relative ease with which the ceramic is formed with high transparency is due to deliberate use of intersubstitutional cations, whereby the three different cation sites of the garnet unit cell are each occupied by more than one of the four cation species. This results in flexibility in stoichiometry, and ultimately, a high yield industrial manufacturing process due to excellent phase stability over a range of compositions. C. Bismuth-loaded Polymers Polymer scintillators with 4wt% Bismuth were formed at LLNL by mixing a Bismuth metallo-organic with monomer and 3% fluor, heating to melt, and subsequent polymerization of the homogeneous mixture. Two different fluors were investigated, an Iridium complex, emitting in the green, and diphenylanthracene, emitting in the blue. III. GAMMA SPECTROSCOPY Pulse height spectra were acquired with various radioactive sources, with scintillators coupled to a Hamamatsu R6231-1 PMT, a Tennelec TC243 amplifier (shaping times 8-12 microseconds), and recorded with an Amptek 8A MCA. A comparison of the energy resolution obtained with the new scintillator materials and commercial and crystals, using the Am-241, Ba-133, Cs-137 and Co-6 sources is shown in Figure 2. To explore the ease of detection of a weak source with different scintillators, we measured a Cs-137 source at a 3 foot standoff with, and, all ~1 in 3 in size, with equivalent acquisition times. Figure 3 shows that the 662 kev peak is readily observed in the spectrum, while the poorer energy resolution of degrades its ability to detect the source, and the intrinsic background in (La-138 gammas at 789 and 1436 kev, betas in - 255 kev range, the Ba-138 32-37 kev x-rays, and sums of these) complicates its spectrum [11]. 1 1 1 1 1 1 2 4 6 8 1, 1" x 1" Cs-137, 3 ft ), 1" x 1" Cs-137, 3 ft, 1" x 1" Cs-137, 3 ft 12 14 16 Fig. 3. Pulse height spectra of, LaBr 3(Ce) and SrI 2(Eu), acquired under same conditions, for same scintillator size and acquisition time, using a low count-rate Cs-137 source. Pulse height spectra were acquired with Am-241, Co-57 and Cs-137 sources of 1.5 cm x.3 cm Bismuth-loaded polymer samples. Figure 4 shows that the energy resolution obtained for the sample containing the Iridium fluor is somewhat better than that of the diphenylanthracene activated polymer. This is likely due to the higher light yield with the Ir-complex, which collect both singlet and triplet excitons due to spin-orbit coupling [12]. Since the samples are so small and have only modest stopping power, escape peaks similar in intensity to the photopeak, are observed, from 29 Bi K x-rays at 74.8 and 77.1 kev. These initial results suggest that it may be possible

Energy Resolution (%) Submitted to IEEE Transactions on Nuclear Science, IEEE/NSS Proceedings 21 3 1 4 1 3 1 2 27% Am-241, 59.5 kev Co-57, 122 kev, 46 kev (escape) Cs-137, 33 kev and 662 kev, 586 kev (escape) 24% 22% 17% Escape 8.5% Photopeak 8.1% 2 1 Bi-loaded polymer 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 1 1 1 3 1 2 36% 32% 24% 2 4 Fig. 4. Pulse height spectra of two 1.5 cm x.3 cm Bismuth-loaded polymer samples, acquired with Am-241, Co-57 and Cs-137 sources. to produce polymer-based gamma spectrometers, with useful energy resolution for radioisotope identification, however larger samples must be fabricated and tested. Figure 5 shows energy resolution obtained as a function of gamma ray energy for the scintillators described herein. The data may be fit to equation 1, comprised of the photon statistics term and a B factor that accounts for all nonidealities in the measurement, including intrinsic light yield non-proportionality, material response non-uniformity and light collection non-uniformity. R = {[(A / E).5 ] 2 + B 2 ]}.5 (1) Am-241 Co-57 Cs-137 9.5% 6 In equation 1, A is a constant, E is the gamma energy, and B is required when there is deviation from pure photon statistics. For the data shown in Figure 5, the values for B obtained from the fits are shown in Table I. The large value of B for is thought to be due to its intrinsic non-proportionality. The much smaller value for likely results from light trapping, which may be addressed by careful optical design [13,7]. The origin of B value for is under investigation. Material response non-uniformity is likely the 9% Fig. 5. Energy resolution as a function of gamma energy for several scintillators; data is fit to Eq. 1. origin of the large B value for the Bi-loaded polymer scintillator, possibly due to its amorphous structure and phase segregation of Bi on the nanoscale, observable by electron microscopy. From the analysis of Nelson, et al. [14], we have deduced that in the energy resolution range applicable to scintillators (4-15% at 2 kev) a false alarm rate (FAR) for identification of spectral anomalies can be expressed: FAR ~ R 3.4 / (S PP / B TOT ).54 ( 2) where R is the resolution of the instrument, S PP is the photopeak efficiency (product of the stopping power and the photofraction), and B TOT is the sum of the natural background (B NAT ) and the self-radioactivity (B SELF ) of the scintillator material. The self-radioactivity for is approximately equivalent to the natural background rate. In terms of this performance metric, and appear promising. For resolution in the 2-keV region, offers an overall reduction in false alarms of a factor of 3 compared to and a factor of 14 better than. These estimates may be qualitatively compared with the data shown in Fig. 3, where in low count spectra, the weak isolated gamma from Cs-137 is most readily identifiable with. For, the advantage is less pronounced, but offers more than 2x better performance than in the 2 kev region, and in the 12 kev region, as seen in the Co-6 spectrum in Figure 2, the advantage of becomes more pronounced. The ease of manufacture and handling of transparent ceramic oxides, could result in an economic advantage for over. Finally, the intrinsic energy resolution of both Scintillator TABLE II COMPARISON OF PROPERTIES AND PERFORMANCE OF SEVERAL SCINTILLATORS FOR GAMMA RAY SPECTROSCOPY Light Yield (Photons/MeV) Energy Resolution (662 kev) NP Energy Resolution a (662 kev) B (Eq. 1) Energy Resolution (2 kev) S PP, %/cm (2 kev) Radiation Background Relative False Alarm Rate, d 1.5 x 1.5 cm 4, 7% 5.% 5.3 9%.3 B NAT 1 LaBr 3(Ce) 63, 3% 2.2%.3 5.4%.34 B NAT+ B SELF.24 SrI 2(Eu) 9, 3% 2.2% 1.1 4.3%.36 B NAT.7 (Gd,Y) 3(Al,Ga) 5O 12(Ce) 5, 4.8% 1.9% 1.3 7.3%.38 B NAT.43 Standard PVT 15, 8% b 3.6% --- --- c.6 B NAT --- c Bi-loaded polymer 1,-3, 8% 3.5% 6.4 15%.8 B NAT 12 a Non-proportionality-limited energy resolution determined by fits to data from the Scintillator Light Yield Non-Proportionality Characterization Instrument, see Reference 14. b Obtained from fitting the Compton Edge of the pulse height spectrum of Eljen EJ-2 plastic scintillator. c Not readily obtained due to extremely weak photopeak for this material. d False Alarm Rate (FAR) is described by Equation 2, for identifying a spectral anomaly in the vicinity of 2 kev.

Submitted to IEEE Transactions on Nuclear Science, IEEE/NSS Proceedings 21 4 and, as estimated from fitting data from the Scintillator Light Yield Non-Proportionality Characterization Instrument, is better than the current experimental values, shown in Table 1, and therefore we expect that with improved material quality and device design, their energy resolution may improve further [15,16]. IV. SUMMARY We have identified several new scintillator materials through a directed search methodology. Among them, offers the highest energy resolution. Transparent ceramic offers high light yield, and gamma spectroscopy with better resolution than. We have synthesized and characterized Bi-loaded polymer scintillators that can be used for gamma spectroscopy, and larger size samples are under development. [14] K E Nelson, T B Gosnell, D A. Knapp, The effect of gamma ray detector energy resolution on the ability to identify unknown radioactive sources, internal LLNL document, 2/25/21. [15] Payne, S.A.; Cherepy, N.J.; Hull, G.; Valentine, J.D.; Moses, W.W.; Choong, W.-S., Nonproportionality of Scintillator Detectors: Theory and Experiment, IEEE Trans. Nucl. Sci., vol. 56, 256-2512 (29). [16] P. Dorenbos. J.T.M. de Haas, and C.W.E. van Eijk, Non- Proportionality in the Scintillation Response and the Energy Resolution Obtainable with Scintillation Crystals, IEEE Trans. Nucl. Sci., vol. 42, p. 219, 1995. REFERENCES [1] N.J. Cherepy, S.A. Payne, S.J. Asztalos, G. Hull, J.D. Kuntz, T. Niedermayr, S. Pimputkar, J.J. Roberts, R.D. Sanner, T.M. Tillotson, E. van Loef, C.M. Wilson, K.S. Shah, U.N. Roy, R. Hawrami, A. Burger, L.A. Boatner, W.-S. Choong, Scintillators with Potential to Supersede Lanthanum Bromide, IEEE Trans. Nucl. Sci., vol. 56, 873-88 (29). [2] N.J. Cherepy, G. Hull, A. Drobshoff, S.A. Payne, E. van Loef, C. Wilson, K. Shah, U.N. Roy, A. Burger, L.A. Boatner, W-S Choong, W.W. Moses Strontium and Barium Iodide High Light Yield Scintillators, Appl. Phys. Lett. 92, 8358 (28). [3] N. J. Cherepy, B. W. Sturm, O. B. Drury; T. A. Hurst. S. A. Sheets, L. E. Ahle, C. K. Saw, M. A. Pearson, S. A. Payne, A. Burger, L. A. Boatner, J. O. Ramey, E. V. van Loef, J. Glodo, R. Hawrami, W. M. Higgins, K. S. Shah, W. W. Moses, SrI 2 scintillator for gamma ray spectroscopy, Proc. SPIE, 7449, 7449- (29). [4] N.J. Cherepy, S.A. Payne, R. Hawrami, A. Burger, L.A. Boatner, E. van Loef, K.S. Shah Prospects for High Energy Resolution Gamma Ray Spectroscopy with Europium-Doped Strontium Iodide, Mater. Res. Soc. Symp. Proc. Vol. 1164, 1164-L11-4 (29). [5] N.J. Cherepy, J.D. Kuntz, J.J Roberts, T.A. Hurst, O.B. Drury, R.D. Sanner, T.M. Tillotson, S.A. Payne, Transparent Ceramic Scintillator Fabrication, Properties and Applications, Proc. SPIE., 779, 779x 28. [6] N. J. Cherepy, J. D. Kuntz, Z. M. Seeley, S. E. Fisher, O. B. Drury, B. W. Sturm, T. A. Hurst, R. D. Sanner, J. J. Roberts, and S. A. Payne, Transparent ceramic scintillators for gamma spectroscopy and radiography, Proc. SPIE, 785, 785I, 21. [7] B.W. Sturm, N.J. Cherepy, O.B. Drury, P.A. Thelin, S.E. Fisher, S.A. Payne, A. Burger, L.A. Boatner, J.O. Ramey, K.S. Shah, R. Hawrami, Effects of Packaging SrI 2(Eu) Scintillator Crystals, Nucl. Instr. Meth. A, in press, 21. [8] L. Mädler, H.K. Kammler, R. Mueller, S.E. Pratsinis, Controlled synthesis of nanostructured particles by flame spray pyrolysis, Aerosol Sci. 33, 369, 22. [9] C.R. Bickmore, K.F. Waldner, D.R. Treadwell, R.M. Laine, "Ultrafine Spinel Powders by Flame Spray Pyrolysis of a Magnesium Aluminum Double Alkoxide," J. Am. Ceram. Soc., 79, 1419-23, 1996. [1] J.D. Kuntz, J.J. Roberts, M. Hough, N.J. Cherepy Multiple Synthesis Routes to Transparent Ceramic Lutetium Aluminum Garnet, Scripta Mat., vol. 57, p. 96, 27. [11] F. Quarati, A.J.J. Bos, S. Brandenburg, C. Dathy, P. Dorenbos, S. Kraft, R.W. Ostendorf, V. Ouspenski, A. Owens, X-ray and gamma-ray response of a 2 x 2 LaBr 3:Ce scintillation detector, Nucl. Instr. Meth. A, vol.574, 115-12, 27. [12] H. Campbell and B. K. Crone, Efficient Plastic Scintillators Utilizing Phosphorescent Dopants App. Phys. Lett. 9, 12117, 27. [13] J. Glodo, E. V. van Loef, N. J. Cherepy, S. A. Payne, and K. S. Shah, "Concentration Effects in Eu Doped SrI 2," Nuclear Science, IEEE Transactions on, vol. 57, pp. 1228-1232, 21.