Revision of CMM performance specifications

Similar documents
ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

A Unified Approach to Uncertainty for Quality Improvement

CMM Uncertainty Budget

Journal of Machine Engineering, 2018, Vol. 18, No. 3, 78 86

Uncertainty evaluation for coordinate metrology by intelligent measurement

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005

Comparison of Results Obtained from Calibration of a Measuring Arm Performed According to the VDI/VDE, ASME and ISO Standards

Annex A Procedure for Determining the Uncertainty of Coordinate Measurement Using Multiple Method

Traceability of on-machine measurements under a wide range of working conditions

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Good Practice Guide No. 130

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

University of Huddersfield Repository

Measurement Uncertainty Knowing the Unknown

Uncertainty and its Impact on the Quality of Measurement

Correct Traceability in Flexible and Complex Dimensional Metrology. Uncertainty in coordinate metrology: - Needs - Standards.

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Coordinate Measuring Machines Coordinate INDEX Measuring Machines Coordinate Measuring Machines New Products

Different Approaches in Uncertainty Evaluation for Measurement of Complex Surfaces Using Coordinate Measuring Machine

INTERNATIONAL STANDARD

Temperature Effects in Dimensional Metrology Influences of Temperature on the Accuracy and how to cope with them

ENG56 DriveTrain. Good practice guide. for surface parameter measurement strategies for form and diameter measurements for large bearings

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Know Your Uncertainty

Dimensional X-ray Computed Tomography Conference 2016

Machine Positioning Uncertainty with Laser Interferometer Feedback

Introduction To Metrology

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Guidelines on the Calibration of Static Torque Measuring Devices

Part 5: Total stations

IMPORTANT FEATURES OF ESTIMATION & INTERPRETATION OF MEASUREMENT UNCERTAINTY. B. Bhattacharya, K. Ramani, P.H. Bhave

Uncertainty of the Measurement of Radial Runout, Axial Runout and Coning using an Industrial Axi-Symmetric Measurement Machine

SCOPE OF ACCREDITATION

A laser metroscope for the calibration of setting rings

This document is a preview generated by EVS

COMMON MISTAKES IN DIMENSIONAL CALIBRATION METHODS

Determination of particle size distribution Single particle light interaction methods. Part 4:

Appendix B1. Reports of SMU

THE IMPACT OF ALIGNMENT CONDITION IN THE MEASURING PLAN BY USING THE SOFTWARE CALYPSO ON THE REPEATABILITY OF MEASURED VALUES

Evaluation of AACMM using the virtual circles method

A hybrid Measurement Systems Analysis and Uncertainty of Measurement Approach for Industrial Measurement in the Light Controlled Factory

Calibration and verification: Two procedures having comparable objectives and results

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Calibration Traceability Guidelines

BHARATHIDASAN ENGINEERING COLLEGE, NATTRAMPALLI. DEPARTMENT OF MECHANICAL ENGINEERING FAQ

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Part 14: Water quality Sampling. Guidance on quality assurance and quality control of environmental water sampling and handling

The Fundamentals of Moisture Calibration

Vocabulary of Metrology

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Assessment of the impact of the articulated probe head angular orientation on the probe head errors distribution

White Paper. Moisture Analyzer Routine Testing

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

This annex is valid from: to Replaces annex dated: Location(s) where activities are performed under accreditation

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

Measurement Uncertainty in Mechanical Testing

ISO INTERNATIONAL STANDARD

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Open Access Repository eprint

Uncertainty due to Finite Resolution Measurements

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Copyright 2017 by Select Calibration Inc. All rights reserved. Temperature Compensation

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Metallic materials Knoop hardness test. Part 1: Test method. Matériaux métalliques Essai de dureté Knoop Partie 1: Méthode d'essai

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

ERROR LIMITS AND MEASUREMENT UNCERTAINTY IN LEGAL METROLOGY

ISO INTERNATIONAL STANDARD. Metallic materials Vickers hardness test Part 3: Calibration of reference blocks

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

MECH-201: Precision Shaft Alignment using dial indicators

Metrological Characterization of Hardness Indenter Calibration System

Precision Shaft Alignment... dial indicators Seminar Synopsis

Achieving traceability of industrial computed tomography

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

High Accuracy CMM Measurements at NIST

Internal CQ. Performance criteria for internal QC how to define a significant deviation from the expected results

AM Metrology at NPL. Stephen Brown. Tuesday 8 th December 2015

Acknowledgments. Abstract

Inter laboratory comparison of industrial CT scanners CIA-CT audit. Final report

Traceable Mass Determination and Uncertainty Calculation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF CALIBRATION

M3003 EDITION 2 JANUARY 2007

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

ISO INTERNATIONAL STANDARD. Metallic materials Brinell hardness test Part 2: Verification and calibration of testing machines

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

TITLE OF THE CD (English): New OIML Document

Uncertainty of Measurement A Concept

AERODROMES PANEL (AP) VISUAL AIDS WORKING GROUP (VAWG) FIFTH MEETING. Montréal, Canada, 25 to 27 June 2008

Measurement uncertainty and legal limits in analytical measurements

IS INTERNATIONAL STANDARD. Acoustics - Determination of sound power levels of noise sources using sound intensity - Part 2:

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Robotic manipulation in dimensional measurement

Transcription:

Educational Articles: Article nr. 3 Revision of CMM performance specifications Premise The working group (WG) Specifications is one of several working groups of ia.cmm, the International Association of Co-ordinate Measuring Machine Manufacturers. This working group is active in the area of CMM specifications. The performance specifications for CMM s are defined by the ISO 10360 series. Based on ISO 10360-1, the Specifications WG has generated a structure for a common data sheet within all associated Companies. The general goals of WG Specifications are: The development of human skills by education and training. Giving users worldwide the assurance that machine performance is verified by CMM Manufacturers in compliance with the regulations, methods and equipment stipulated by international standards. Promotion of and support in the adoption of standards for metrology performance testing. Some of the Members of the Specifications WG also take part on national and international meetings of committees for standardisation of CMM performance specifications. One task of the Specifications WG is to inform the Members of ia.cmm and the Users about the upcoming new revision of ISO 10360 part 2 and of part 5, both dealing with acceptance and reverification tests for CMM s. These new revisions will contain more precise definitions, some new performance specifications and new procedures and new artefacts for application to a larger variety of CMM s. They are scheduled for release in 2007. Furthermore, additional new technical specifications related to CMM technology will be released in 2005. These include ISO/TS 23165 Guide to the evaluation of CMM test uncertainty and ISO/TS 15530-3 Uncertainty assessment using calibrated work pieces as one of several techniques for determining task specific uncertainties of measurements with CMM s. Makoto Abbe (Mitutoyo J) Michele Verdi (Hexagon Metrology SpA I) Josef Wanner (Carl Zeiss Industrielle Meβtechnik GmbH D) 1 Coming soon: ISO10360-2 CMM s used for measuring linear dimensions 1.1 Reason for revision Technologies applicable to coordinate measurement in industry have evolved rapidly in recent years. The current trend requires better clarified CMM specifications and enhancement of the corresponding ISO standards. The revised version of ISO10360-2, the acceptance test and reverification test for CMM, has been discussed by ISO/TC213/WG10. A brief overview of this upcoming revised standard is provided in the following. 1.2 Enhanced and clarified test procedures The new test procedure more clearly reveals the performance of CMM s by verifying their conformance or non-conformance to the specification. The test scheme of using a conventional E-test to verify MPE E in seven different positions and orientations in the CMM volume on five different sizes has largely been retained, although the ia.cmm 1/12

conventional E-test is now called E 0 -test for clarification. On the other hand, in keeping with the concept of modularity, the test for determining the probing error (Ptest), which verifies MPE P, has been moved from ISO10360-2 (Part 2) to ISO10360-5 (Part 5). An acceptance test performed according to the upcoming ISO10360-2 requires a tester to refer not only to Part 2 but also to Part 5 in an appropriate manner. The following topics are also dealt with in the new standards. 1.2.1 Thermal properties of test length Major test uncertainty results from thermal property of the test length used, regardless of whether the CMM is thermally compensated or not. Therefore, a CMM stating the specification conforming to the upcoming ISO10360-2 is required to disclose the necessary thermal properties of the test length used for acceptance, i.e. the upper limit of test length CTE (Coefficient of Thermal Expansion) and the upper limit of uncertainty in test length CTE. Furthermore, the default for the calibrated test length is specified as a normal CTE which is a material with a CTE between 8 x 10-6 /K and 13 x 10-6 /K. A specification statement requiring a test length with a non-normal CTE is obliged to indicate MPE*E 0 instead. Especially in a case where use of a low CTE test length is stated by the CMM specification, an additional E 0 -test is performed on a normal CTE test length. 1.2.2 Improving test sensitivity to ram axis roll A new E L -test has been introduced with test sensitivity to ram axis roll proportional to the offset distance between the centreline of the ram axis and the centre of the stylus tip. Two E L -tests are performed with a 150mm offset distance in the default case. The user can choose the position and orientation for the test from eight possibilities along plane diagonals on planes parallel to the ram axis movement. 1.2.3 Repeatability test Repeatability R 0 is calculated and tested using the results from an E 0 -test comprising three repetitive measurements performed on the CMM. Conformance or nonconformance to the specification is verified. 1.3 Newly introduced concept The application possibilities for the coming standard in industry are expected to expand widely. Major possibilities are mentioned below. 1.3.1 Various artefacts that represent a calibrated test length The upcoming standard allows us to adopt a variety of artefacts with a proper test procedure, as far as bidirectional measurement is performed directly or composed of a combination of size measurement on a short gauge block and a series of unidirectional measurements. Various unidirectional measurement possibilities such as laser interferometer, ball bar, ball plate, and so on, are listed in the annex. The typical difficulty of performing the acceptance test on a large CMM, for example, is ia.cmm 2/12

overcome technically. The test result thus becomes much more practical and conforms fully to the upcoming standard. 1.3.2 Application possibilities for CMM s with particular functionalities ISO10360-2 is intended to apply to Cartesian CMM s with three orthogonally combined straight guideways equipped with contacting probing systems. Non Cartesian type CMM or optical CMM are of increasing importance. The scope of the upcoming standard mentions that the standard may be applied to these kinds of CMM by mutual agreement between the supplier and the CMM user. The test procedure for a dual ram CMM constituted of two independent arms and in a single coordinate system is specified as well. 1.4 Summary The benefit of the test according to ISO10360-2 is that the measured result has a direct traceability to the unit length. It therefore gives information on how the CMM will perform in connection with similar length measurements. The importance of this standard is increasing year by year, since ISO 10360-2 was published in 1994 and revised in 2001. The upcoming standard, ISO 10360-2, will surely be able to meet the expectations of industry regarding higher productivity. 2 ISO 10360-5: CMM Probing Performance with Contacting Probing System 2.1 Reasons for revision The revision of ISO 10360-5 provides some new definitions, clarifications and better modularity. The scope of ISO 10360-5 will be limited to contacting probing systems. Additional new parts of ISO 10360 to be released later will cover the performance tests for further probing systems. ISO 10360-7 for CMM s equipped with video probing systems is already being prepared. 2.2 Enhanced test procedures As a major change in the long term revision, the test of probing error P will be taken out of part 2 and will be integrated in part 5 as single stylus probing system test for modularity reasons. All tests related to a contacting probing system will be in part 5. Part 2 will contain the basic background description and will be focused on size tests only. ia.cmm 3/12

The concentration of all contacting probing system tests in part 5 will also eliminate double measurements. Measurements taken for the single stylus probing system test can be used for the multiple stylus configuration test as well. The probing system performance is contaminated by the CMM performance. The tests in ISO 10360-5 are sensitive to many errors depending on both the CMM and the probing system, and are to be performed in addition to the length measuring tests given in ISO 10360-2. It is the CMM probing performance which is specified, due to the impracticality of isolating the performance of the probing system from that of the CMM. 2.3 Expanded scope The definition of specific parameters for additional types of contacting probing systems will allow the application of ISO 10360-5 for: Multiple styli connected to the CMM probe (e.g. a star). Installations using an articulating probing system (motorized or manual) that can be pre qualified in empirical mode at each utilised angular position or at a few angular positions in inferred mode with interpolation for any angular positions. Installations using a repeatable probe-changing system. Installations using a repeatable stylus-changing system and multiple-probe installations. 2.4 Renaming of the symbols For a better overview of the high number of definitions, it was necessary to rename the symbols. The symbols for errors and values are in the form Px_Ty with the indications: P x T y for probing performance F for a form parameter L for a location parameter S for a size parameter for a contacting (touch) probing system E for an articulating system using empirical qualification I for an articulating system using inferred qualification M for a fixed multiple-stylus configuration N for a fixed multiple probe configuration U for a single (unique) stylus configuration Examples: PF_TI probing system form error based on an articulating tactile probing system using inferred qualification. ia.cmm 4/12

PF_TU PS_TU PL_TM single stylus probing system form error based on a tactile probing system (equal to probing error P as defined in ISO 10360-2:2001). single stylus probing system size error based on a tactile probing system. probing system location value based on a tactile fixed multiple-stylus configuration. The single stylus probing system size error PS_TU is a new definition. For PS_TU it is not required a specification but it is a signed error. It is needed to get comparable results between bidirectional 2-point-distance measurements and other methods for size measurements according ISO 10360-2. 2.5 Clarified test procedures The new ISO 10360-5 will contain definitions which supersede similar definitions in ISO 10360-1:2001, because some symbols used have been revised and expanded for clarification. Specification and test according ISO 10360-5 is necessary for one extension length only. The effective stylus diameters are not identical within a stylus system used for the test and even different nominal diameters could be specified. The test sphere (10-50 mm) needs to be calibrated for form and for size. Form errors of styli and test sphere have to be smaller than 20 % of the specified MPE. They must be covered by the specification and can not be subtracted from the test result. The user is free to choose the location of the test sphere within the measuring volume if not otherwise specified by the manufacturer. 3 ISO/TS 23165 Guide to evaluation of CMM test uncertainty 3.1 Purpose, motivation and status ISO/TS 23165 aims to define the proper criteria for evaluating the CMM test uncertainty. ISO 14253-1 (1) regulates conformance/non conformance decisions concerning acceptance of a work piece or a measuring equipment. In the case of a work piece, the specification is given by the tolerance: in the case of a measuring device (i.e. a CMM) the specification is given by MPE (Maximum Permissible Error), as prescribed ia.cmm 5/12

by ISO 10360 standards (2). All ISO 10360 chapters dealing with CMM compliance to specifications prescribe that the CMM performance is to be verified if the error does not exceed the related MPE, taking into account the uncertainty of measurement as specified by ISO 14253-1. However, no guidance was provided to CMM users and CMM manufacturers for establishing which factors are the main uncertainty contributors and the related expanded uncertainty U by which the specification zone must be either reduced or expanded. Now, with ISO/TS 23165, this gap has been eliminated, and ISO 14253-1 can be uniformly applied in the CMM performance verification which is so important for regulating the related contractual aspects. The application field of ISO/TS 23165 is CMM performance verification, and not the general measurements made by a CMM. In this last case, the ISO 15530 series applies. The TS (Technical Specification) status of the document means that the document is officially released and therefore valid even though it does not yet has the full status of an international standard. 3.2 Key concepts The ISO/TS 23165 guide is based on some key concepts: #1 the concept of test uncertainty. #2 the concept of tester and tester counterpart. The test uncertainty is the expanded uncertainty U associated with the testing equipment and its use in the test, i.e. with the test conditions. A quality index can be in fact associated to both test conditions and CMM performance: test uncertainty U is the quality index associated with the test conditions: o conventional errors like the error of indication, E, or probing error P, are the quality index of CMM performance. ISO/TS 23165 defines the criteria of interaction between test conditions and CMM performance in the spirit of ISO 14253-1. Test conditions are determined by choices of the tester. When a poor material standard of size is used instead of a better one, this will immediately affect the test uncertainty. The tester is the main player in the test uncertainty scenario, and ISO/TS 23165 consequently identifies in the tester the main responsible for the test uncertainty. When the tester is required to provide relevant test inputs (i.e., the uncertainty of CTE of material standard of size), the related contribution to the test uncertainty budget must be considered. 3.3 Test uncertainty contributors ISO/TS 23165 provides the list of the test uncertainty contributors referring to: Error of indication on size E. Probing error P. as stated in ISO 10360-2. ia.cmm 6/12

Concerning probing error P, the main contributors to test uncertainty are: Form error F as reported in the calibration certificate. Standard uncertainty u(f) as reported in the calibration certificate. Concerning the error of indication for size, the main contributors are: Uncertainty due to calibration of the material standard of size, u(ε cal ): this uncertainty contributor quantifies the tester s ability to choose a proper material standard of size for the test. Uncertainty due to the CTE (Coefficient of Thermal Expansion) of the material standard of size u(ε α ): this uncertainty contributor quantifies the tester s ability to choose a proper material standard of size and proper thermal test conditions. It has to be included in the uncertainty budget only when the CMM expects the tester to input a CTE value. In case of thermally uncompensated CMM s, the related inaccuracies are taken into account in the MPE specification. Uncertainty due to the input temperature of the material standard of size u(ε t ): this uncertainty contributor quantifies the tester ability to provide good temperature measurements used for compensating the measuring results. It has to be included in the uncertainty budget only if the temperature measurement depends on the tester s own equipment. In cases involving temperature sensors embedded inside the CMM and thermally uncompensated CMM s, the related inaccuracies are taken into account in the MPE specification. Uncertainty due to misalignment of material standard of size: this uncertainty contributor quantifies the tester s ability to provide a good alignment of the material standard of size. It can be considered as negligible under specific conditions (i.e. if differences between alignment procedures in the calibration and testing phases, the effects of CMM tunnelling errors and improper sampling strategy are properly minimized) Uncertainty of fixturing the material standard of size: this uncertainty contributor quantifies the tester s ability to minimize deformations in the material standard of size. Statistical analysis or expert judgement can be used in determining it. (1) ISO 14253-1 Inspection by measurement of work pieces and measuring equipment Part 1: Decision rules for proving conformance or nonconformance with specifications. (2) ISO 10360-2 CMM s used for measuring sizes. ISO 10360-3 CMM s with the axis of rotary table as fourth axis. ISO 10360-4 CMM s used in scanning measuring mode. ISO 10360-5 CMM s using multiple stylus probing systems. 4 Techniques for evaluation of the uncertainty of measurement 4.1 ISO 15530-1 Overview and metrological characteristics ia.cmm 7/12

4.1.1 Purpose, motivation and status ISO 15530-1 aims to introduce the techniques for determining the uncertainty of measurement for a CMM, providing the proper terminology and a reference list of metrological characteristics, i.e. a list of factors that can potentially affect the measurements produced by a CMM. The document is still in a draft version. 4.1.2 Key concepts ISO 15530-1 is based on the following key concepts: #1 Concept of task specific measurement uncertainty. #2 Concept of metrological characteristic. Task specific measurement uncertainty is the expanded uncertainty U including the uncertainty contributors related to the overall process concerning a work piece measured by a CMM. The metrological characteristic of a measuring device is defined as the characteristic of a measuring device which may influence the results of measurements and which can determine an immediate uncertainty contribution. 4.1.3 List of ISO 15530 series standards General title: Geometric Product Specification (GPS) Coordinate Measuring Machines (CMM): techniques for determining the uncertainty of measurement Part 1: Overview and metrological characteristics. Part 2: Use of multiple strategies in calibration of artefacts. Part 3: Use of calibrated work pieces or standards. Part 4: Use of computer simulation. Part 5: Use of expert judgement, sensitivity analysis and error budgeting. 4.2 ISO 15530-2 Use of multiple strategies in calibration of artefacts 4.2.1 Purpose, motivation and status ISO 15530-2 aims to introduce the technique of multiple measurement strategies, for determining the uncertainty of measurement related to a work piece measured by a CMM. The multiple measurement strategy aims to combine several orientations of the work piece and point distributions on its surface in order to get a better estimation for the conventional true value of the measurand. By applying a simple procedure, work pieces can be calibrated by a CMM. The document is still in a draft version. ia.cmm 8/12

4.2.2 Key concepts ISO 15530-2 is based on the following key concept: #1 Concept of multiple strategies. Multiple strategies is a technique that, based on measurement of the work piece in different orientations, repeated measurements varying the point distribution on the work piece, and proper measurements of the material standard of size, permits the generation of work piece calibration values and the related uncertainty by means of a CMM. 4.2.3 Procedure The work piece must be measured in at least three (default four) different orientations corresponding to positions capable of guaranteeing good measurement conditions. In each orientation, the work piece must be measured with at least five different point distributions. In case distance/size measurements are required on the work piece, subsidiary measurements on a material standard of size similar in its length to the above distance/size, must be performed along the three CMM coordinate axes and repeated three times each. The calibration value, and related calibration uncertainty are determined by proper calculation based on the database generated by all of the measuring results obtained. 4.3 ISO 15530-3 Use of calibrated work pieces or standards 4.3.1 Purpose, motivation and status ISO 15530-3 aims to provide a technique for a simple uncertainty evaluation of measurements performed by a CMM. This technique applies to specific measuring tasks and to CMM results obtained from both uncorrected and corrected measurements. The standard was published on 2004-03-01. 4.3.2 Key concepts #1 Concept of non substitution measurement. #2 Concept of substitution measurement. #3 Concept of similarity condition. Non substitution measurements are results in which the CMM indication is not corrected by systematic errors: in substitution measurement, the CMM indication is corrected by systematic errors, where both the work piece and a proper material standard of size are measured. ia.cmm 9/12

Similarity conditions define the constraints binding work piece and material standard of size in the uncertainty assessment (i.e. dimensions and materials) the procedure used for measuring them, the environmental conditions, the CMM stylus configuration. 4.3.3 Procedure In case of a nonsubstitution strategy, calibrated artefacts are measured on a CMM in at least ten measurement cycles, each composed of the handling of each artefact and its consequent measurement. A total of at least 20 measurement repetitions must be achieved in different conditions simulating the real measuring process, including position and orientation of the work piece. This means that, in case only one artefact is involved in the assessment, the above cycle must be repeated 20 times. The uncertainty contributors include the following: The calibration uncertainty stated in the artefact certificate. The standard uncertainty assessed by the above procedure. The standard uncertainty resulting from the variations of form errors, roughness, CTE, and other relevant parameters in different corresponding workpieces. In case of a substitution strategy, the same procedure as stated above applies, i.e. the measuring cycle comprises the handling of the calibrated artefact, its consequent measurement, the handling of the work piece, and its consequent measurement. 4.4 ISO 15530-4 Estimating task-specific measurement uncertainty using simulation 4.4.1 Purpose, motivation and status ISO 15530-4 Estimating task-specific measurement uncertainty using simulation aims to define criteria for simulation techniques applied to task-specific uncertainty evaluations. The target of these techniques is to provide measurement results combined with the related measuring uncertainty. Their application will enable the user to take immediate decisions about the consistency of the measuring process and the conformance or non-conformance of the work piece to the specification as required by ISO 14253-1. ISO 15530-4 is the result of pluriennial research activities focused on simulation techniques applied to coordinate metrology (Virtual CMM). In Virtual CMM (VCMM), the VCMM model is assessed on the basis of a set of input parameters by means of proper CMM measurements of sphere/hole plates and of environmental parameters: using VCMM, it is possible to generate calibrated artefacts, i.e. to associate to the measurement result with the related uncertainty. ia.cmm 10/12

ISO 15530 is still in the DTR (Draft Technical Report) phase, namely it is an unpublished document in the first phase of the TR status which does not have the full status of an international standard. 4.4.2 Key concepts ISO 15530-4 is based on the following key concepts: #1 Concept of UES (Uncertainty Evaluating Software). #2 Concept of UES model. #3 Concept of UES validation. The UES is a software used to provide uncertainty evaluation by simulating the overall CMM measuring process on a work piece: it can be resident inside the software equipping the CMM, or in another system working in conjunction with it. UES model can be based on mathematical and/or numerical procedures able to handle input quantities (like the temperature of the air) and to determine the related uncertainty contribution. UES can consider only a part of the metrological characteristics and related uncertainty contributions in its evaluating process: in this case, the uncertainty evaluation must be based on a combination of uncertainty evaluation techniques, some of which are derived from pure simulation, and others of which are based on an experimental approach (as described in ISO 15530-3). The UES manufacturer must provide: A list of CMM metrological characteristics managed by the UES. Proper documentation of the techniques used for the uncertainty evaluation. UES can be validated by following two different approaches: Experimental approach consisting of proper testing with a CMM on a calibrated artefact and/or on a proper calibrated material standard of size. Iterative testing, combined with proper analytical tests focused on some specific factors like scale and probing errors, can consolidate the basis of the uncertainty analysis. Computer-aided techniques (CVE, Computer aided Verification and Evaluation). The process can be based on error vector field analysis, comprising the vector applied to the expected ( true ) measured point and its direction versus the determined position of the measured one. By proper perturbation of a reference true condition, based on the input of different input quantities it is possible to achieve an uncertainty evaluation related to the metrological characteristic under examination. Uncertainties caused by perturbation of contact points derived from a set of reference input in the UES under examination can thus be compared with ones related to contact points given by a testing body by means of coherently simulated instances. Historical data referred to proper predefined tasks can be considered too. ia.cmm 11/12

4.5 ISO 15530-5 Use of expert judgement, sensitivity analysis and error budgeting 4.5.1 Purpose, motivation and status ISO 15530-5 aims to define the criteria of the technique for uncertainty evaluation based on expert judgement. The use of expert judgement can be particularly important when type B standard uncertainty is considered in the uncertainty budget. Personnel aiming to provide expert judgements must be properly qualified. ISO 15530-5 is still in draft status. 4.5.2 Key concepts #1 concept of qualified personnel Qualified personnel, is here defined as personnel possessing the academic qualifications, experience and educational background in the field of accreditation and is required to provide an expert judgement. It is expected that any signatory representing a calibration laboratory accredited according to ISO/IEC 17025 by a national accreditation body is able to provide proper judgements. ia.cmm 12/12