SIM FORCE STANDARDS COMPARISON UP TO 10 kn

Similar documents
FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

Comparison between the CENAM (Mexico) 150 kn and the INRiM (Italy) 1 MN force standard machines

FORCE NATIONAL STANDARDS COMPARISON BETWEEN CENAM/MEXICO AND NIM/CHINA

SIM AUV.A-K1.prev Microphone Intercomparison

PROTOCOL OF MASS AND VOLUME COMPARISONS BETWEEN SIM NMIs

FINAL REPORT OF THE BILATERAL COMPARISON OF THE CALIBRATIONS OF STANDARD WEIGHTS BETWEEN CENAM-MEXICO AND INEN-ECUADOR SIM.M.M-S4 (SIM.7.

INTERLABORATORY MASS COMPARISON BETWEEN LABORATORIES BELONGING TO SIM SUB-REGIONS COORDINATED BY CENAM (SIM.7.31a & SIM.7.31b)

SIM SIM.M.D-K3. provide. SIM.M.M-K5 for. A set of. Institute LACOMET LATU INTI. Country Costa Rica Uruguay Argentina Chile México Canada Brazil

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland

XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

SIM Regional Comparison on. The Calibration of Internal and External Diameter Standards SIM.L-K FINAL REPORT July 2012

Force Key Comparison CCM.F-K4.a and CCM.F-K4.b for 4 MN and 2 MN Forces. Final Report. T.W. Bartel NIST Mass and Force Group U.S.A.

LINKING SIM MASS COMPARISONS TO THE KCRV ON 1 kg. Luis O. Becerra CENAM Querétaro, Qro., Mexico,

Supplementary Comparison

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force

Final Report on the Torque Key Comparison CCM.T-K2 Measurand Torque: 0 kn m, 10 kn m, 20 kn m Dirk Röske 1), Koji Ogushi 2)

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Final Report. SIM / ANDIMET Supplementary Comparison for Volume of Liquids at 100 ml and 100 µl SIM.M.FF-S7

Force Key Comparison APMP.M.F-K2.a and APMP.M.F-K2.b (50 kn and 100 kn) Final Report 6 August Pilot: KRISS, Republic of Korea Yon-Kyu Park

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

INTERLABORATORY COMPARISON BETWEEN TUBITAK-UME TURKEY AND INM ROMANIA IN THE FIELD OF FORCE MEASUREMENTS

Final report on CCQM-K36.1 with erratum

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report on. Regional Key Comparison for Volume of liquids at 20 L and 100 ml Conducted January 2007 to December 2008 SIM.M.

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c)

Uncertainty Propagation for Force Calibration Systems

Study for the requalification of Inmetro's Primary Hardness Standardization Machine for Vickers HV3 scale

ANNEXE 8. Technical protocols for the interlaboratory comparisons

Calibration of Analog Scales Based on. a Metrological Comparison Between. the SIM Guide and OIML R 76-1

THE UNCERTAINTY OF MEASUREMENT IN CALIBRATION USING A COMPARISON FORCE STANDARD MACHINE

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

Comparison of V and 10 V DC Voltage References

TC-Chair Annual Report 2008/2009 TC-M

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

ЕВРО-АЗИАТСКОЕ СОТРУДНИЧЕСТВО ГОСУДАРСТВЕННЫХ МЕТРОЛОГИЧЕСКИХ УЧРЕЖДЕНИЙ (KOOMET)

Key Comparison CCAUV.A-K4

Metrological Characterization of a Primary Vickers Hardness Standard Machine - NIS Egypt

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

INVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE

Final report Comparison Identifier: APMP.SIM.M.P-K1c Other designation: NIST-NPLI/Pressure/2

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Report on EURAMET.M.M-S9

Estimate of the inertial torque in rotating shafts - a metrological approach to signal processing

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

analysis in capillary preparation. In despite of, to 100 mg 20 g in reliability of the the the drop first and weighing readings used in third drop

Uncertainty of Calibration. Results in Force Measurements

Calibration Traceability Guidelines

UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

CERTIFICATE OF ACCREDITATION

CCL-K5. CMM 1D: Step Gauge and Ball Bars. Final Report

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

Loadcell Calibration - Evaluation of Uncertainties

INVESTIGATION OF IMPACT HAMMER CALIBRATIONS

The Journey from Ω Through 19 Orders of Magnitude

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

Final Report On RMO VICKERS KEY COMPARISON COOMET M.H-K1

Comparison protocol EURAMET project

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Proposal for Relevant Comparisons of Natural and Related Reference Materials Lisa Karam, NIST 29 March 2005

Portuguese Institute for Quality. Contents

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

BILATERAL KEY COMPARISON SIM.T-K6.2 ON HUMIDITY STANDARDS IN THE DEW/FROST-POINT TEMPERATURE RANGE FROM 20 C TO 20 C

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

Evaluation of Force Standard Machines by using Build-up System

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Analysis of drift correction in different simulated weighing schemes

STUDY OF THE BEST MEASUREMENT CAPABILITY IN ROCKWELL SCALE AT THE BRAZILIAN NMI INMETRO S PRIMARY HARDNESS STANDARD MACHINE

DESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE

EVALUATION OF STATIC AND DYNAMIC PARASSITIC COMPONENTS ON THE INRIM 1 MN PRIMARY FORCE STANDARD MACHINE BY MEANS THE 500 kn SIX-COMPONENT DYNAMOMETER

Establishment of a Traceability Chain in Metrology for Materials

CCM short note on the dissemination process after the proposed redefinition of the kilogram

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

Key Comparison. CCQM K Electrolytic Conductivity at 0.5 S m -1 and 5 ms m -1. Final Report

The National Unitary Institution "Belarusian State Institute of Metrology" (BelGIM)

The calibration system of force measurement devices - conceptions and principles

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

FINAL REPORT ON THE KEY COMPARISON EUROMET.AUV.A-K3

ISO defines the Rockwell superficial hardness value, HRN, as:

ISO 376 Calibration Uncertainty C. Ferrero

Final Report August 2010

Document No: TR 12 Issue No: 1

Transcription:

SIM FORCE STANDARDS COMPARISON UP TO 10 kn Jorge C. Torres-Guzmán, Daniel A. Ramírez-Ahedo Centro Nacional de Metrología, CENAM.. ABSTRACT A comparison in the quantity of force was carried out among the SIM national laboratories in order to estimate the level of agreement for the realization of the quantity and the uncertainty associated to its measurement. This comparison was carried out up to 10 kn. The equipment used consisted on two force transducers (load cells); both with the same measuring range (10 kn). With the purpose of obtaining maximum accuracy on the transducers, the comparison range was selected from 4 kn up to 10 kn. This comparison provides a link to CIPM Key Comparisons. The results obtained, as well as the reference values selected for the comparison are included in this document. Two different methods were used to analyze the level of agreement and to state the conformity declaration. 1. INTRODUCTION This force comparison was performed among national laboratories within the Interamerican Metrology System (SIM) region. Each laboratory used its national standard for the established measuring range. There were six participating NMIs from three different areas of SIM, Noramet, Andimet and Suramet. The reason the two other SIM areas did not participate in this comparison was that they did not count with standard with the required range. The Centro Nacional de Metrología (CENAM),, had the role of coordinator and pilot laboratory. The comparison started in August 00 and finished in September 004. In order to link this comparison with the CIPM Force Key Comparison CCM.F-K1.a and CCM.F-K1.b, the force points of 5 kn and 10 kn were measured in this comparison. This paper presents the results of the laboratories which used primary standards (dead weight machines).. SCOPE The objective of the comparison was to estimate the level of agreement for the realization of the quantity and the uncertainty associated to its measurement up to 10 kn within the SIM region. Two transducers (load cells) were used as the comparison standard, to obtain its maximum accuracy, the comparison range was selected from 4 kn to 10 kn (starting at 40% of the full load cells range)..1 Participating Laboratories Table 1. Participating laboratories. SIM area Laboratory Person in charge Country Suramet Centro de Física, Instituto Nacional de Tecnología Industrial Luis Giobergia (INTI-CEFIS) Instituto Nacional de Metrologia, Normalização e Qualidade Jorge Cruz Industrial (INMETRO) Instituto de Investigaciones y Control del Ejército (IDIC) Christian Villarroel Chile Laboratorio Tecnológico del Uruguay (LATU) Claudia Santo Uruguay Andimet Centro de Control de Calidad y Metrología, Superintendencia Arisitides C. Dajer Colombia Industria y Comercio (SIC) Noramet Centro Nacional de Metrología (CENAM) Jorge C. Torres

. Comparison Standard The comparison standard was conformed of transducers with the same range, but with different metrological characteristics. Each laboratory used its own electronic amplifier (DMP 40). A resistance calibrator was sent to allow corrections due to each amplifier possible deviations. Table. Comparison Standard Data. Transducer Type: Load cell Load cell Resistance calibrator Range: 1 kn to 10 kn 1 kn to 10 kn -,0 to,0 mv/v Make: HBM HBM HBM Model: C3H C3H3 K3608 Serial number: F 44 067 G 51 316 03 50 014 Uncertainty (k = ): ±0,010% of the reading ±0,031% of the reading ±0,005% of full scale.3 General Guidelines and Procedure The relevant aspects of the measurement protocol are summarized here, but they were carried out widely in [1] and []. Just a few relevant points are mentioned here: a) The transducers measurements were made in mv/v. The reference temperature was C ± 1 C. b) The reading without a load was referred as the zero reading. The measurements on the comparison standard were performed strictly in ascending order up to the measuring force load. c) The DMP40 resistance was verified before and after de measurements with the K3806 calibrator provided, in: 0,0 mv/v; 0, mv/v; 0,8 mv/v; 1,0 mv/v; 1, mv/v; 1,8 mv/v;,0 mv/v. d) For both transducers, five measurements positions were used 0, 90, 180, 70 y 360. e) The load application cycles for each position are shown in Figure 1. f) After finishing the corresponding readings, each participating laboratory sent to the pilot laboratory, the complete data file report of the measurements, including the associated uncertainty g) Each participating laboratory technical staff performed the measurements and it was their sole responsibility to fulfill the requirements of the agreed regulating documents of this comparison. 100% 100% 80% 50% 40% 0% 90s 90s Load cycle 1 (0 ) 10 0% 100% 40% 50% 80% 40% 50% 80% Load cycles (90 ), 3 (180 ) and 4 (70 ) Load cycle 5 (360 ) Figure 1. Comparison Load Cycles.

3. PARTICIPATING LABORATORIES STANDARDS Four of the participating laboratories used Dead Weights Machines (DWM) and used Force Transfer Standard Machine (FTSM). In Table 3, general information of the standards from each laboratory is listed. The uncertainties declared are those included in the BIPM CMCs data base. Table 3. Participating Laboratories Standards General Information. Laboratory Machine Type Range Declared Uncertainty INTI-CEFIS, DWM 1 kn 110 kn 300 10-6 INMETRO, DWM 1 kn 110 kn 0 10-6 IDIC, Chile FTSM 500 N 50 kn 500 10-6 LATU, Uruguay FTSM 1 kn 10 kn 600 10-6 SIC, Colombia DWM 100 N 10 kn 50 10-6 CENAM, DWM 500 N 50 kn 0 10-6 4. RESULTS A spreadsheet in Excel was provided to register the measurements of each transducer and the readings obtained from the DMP40 with the K3806 [3]. The uncertainties calculated by each laboratory were based mainly on four contributing elements: the standard used by the laboratory, repeatability, reproducibility and resolution of the comparison standard (instrument). Each laboratory applied all necessary corrections to the measured force. CENAM carried out five calibrations on the force transducers and the maximum difference among the various calibrations is presented in Graph 1 and Graph, for each of the force transducers. Nominal Force Applied, kn Graph 1. Maximum deviations for the force transducer G51316. Nominal Force Applied. kn Graph. Maximum deviations for the force transducer F44076.

As it is shown in graphs 1 and, the stability during the complete period of the comparison is less than 180 10-6 (relative to the applied force) for one transducer and less than 130 10-6 for the other. For the worst case at 5 kn force point (force transducer G51316), the estimated standard uncertainty due to the transducer stability is, u stability = 0,9 N / (3) 1/ = 0,6 N, which corresponds to 180 10-6 in relative terms, as presented in Graph 1. For each force measurement point the transducer stability uncertainty can be estimated from: u stability of transf.std. r r 3 CENAM 5 CENAM 1 = (1) For the purposes of the comparison, the values presented for CENAM s measurements are the mean values of the different calibrations performed. Since INMETRO and CENAM are primary laboratories and declared the smallest uncertainties, the force reference values used for the comparison are the average measurements obtained by them. V + V INMETRO CENAM V = () ref The following two graphs, present the results for the measurements made to the two comparison transducers as described in Table. The SIC (Colombia) was unable to make measurements in the force transducer F44076. - Force Reference Value, in kn Graph 3. Relative deviation to the reference value of force with the force transducer G51316. Colombia - Force Reference Value, in kn Graph 4. Relative deviation to the reference value of force with the force transducer F44076.

5. DISCUSSION To compare (in a better way) the measurement results from the participating laboratories, the normalized error was calculated using a modified equation (such as in [4]) of the one described in NORAMET s document 8 [5] and EAL P7 [6]. Equation 3 is used to compare all laboratories with the reference value in one graph. The reference uncertainty is INMETRO s and CENAM s combined uncertainties and combining, as uncertainty, the transfer standard stability as indicated in equation (4). e V V lab ref = (3) n u + u lab ref Where, e n - normalized error calculated at each calibration force V lab - laboratory s force value V ref - CENAM and INMETRO force measurements average u lab - laboratory s standard uncertainty u ref - CENAM and INMETRO combined standard uncertainty (see equation 4) ref [( u + u )/ 4 u ] 1 / u + = (4) INMETRO CENAM stability of transf. std. Graph 5 shows the normalized error equation graph for the force transducer G51316. 1.00 Normalized Error 0.75 0.50 0.5 0.00-0.5 Colombia Force reference value, in kn -0.50-0.75-1.00 Graph 5. Normalized error (using standard uncertainty, k=1) for the force transducer G51316. Graph 6 presents the normalized error equation graph results for the force transducer F44076. As previously said, SIC was unable to make measurements in this transducer. 1.00 0.75 0.50 Normalized Error 0.5 0.00-0.5-0.50 Force reference value, in kn -0.75-1.00 Graph 6. Normalized error (using standard uncertainty, k=1) for the force transducer F44076.

Since two force transducers were used, the Youden Plots [7] can be plotted to visualize the results of the comparison. Unfortunately, SIC measurements could not be included due to the fact that they were unable to make measurements to the force transducer F44076. Graphs 7, 8, 9 and 10 show the comparison results for each force measuring point. The standard reference uncertainty is included at the center. -6.5E-04-4.5E-04 -.5E-04 - Reference - - - 3.5E-04 Graph 7. Youden Plot for 4 kn. -6.0E-04-5.0E-04 - Reference - - - 3.0E-04 Graph 8. Youden Plot for 5 kn. R e la tive D evia tion -5.0E-04 - Reference 3.0E-04 4.0E-04 - - - -5.0E-04 - Reference 3.0E-04 - - - -.5E-04 Graph 9. Youden Plot for 8 kn. Graph 10. Youden Plot for 10 kn. 6. CONCLUSIONS Six national laboratories (INTI-CEFIS, IDIC, INMETRO, LATU, SIC and CENAM) compared their force standards by means of two force transducers without performing preliminary measurements prior to the reported data. Two of the national laboratories did not have primary standards and their declared uncertainties were higher. To keep adequate detail on the study their results will be presented separately. In general, the results demonstrated agreement among the four primary national laboratories with negligible differences observed. It is important to notice the fact that the ISO standard uncertainties were used for the normalized error technique and for the Youden Plot results. This made the analysis of comparability among the laboratories more severe. Observing the results of comparability, the participating laboratories showed agreement within their declared uncertainties. The normalized error equation that was used has been proposed as means of assessing comparability between the four laboratories, the Youden Plot was useful for the three laboratories which calibrated the two force transducers; the use of both techniques facilitated the visualization of compatibility of force measurement.

The values obtained by means of the normalized error equation were, for almost all cases, below 0,75; only one laboratory had two points on 0,8 (4 kn and 5 kn, graph 5). For many points was below 0,5. The Youden Plot for the 3 compared laboratories showed full agreement between INMETRO and CENAM; when including the declared uncertainty INTI-CEFIS shows a good agreement with the reference values. ACKNOWLEDGEMENTS The authors wish to express their appreciation for the work performed by the laboratories and specially for the work of the persons in charge mentioned in Table 1. Additionally, the authors wish to acknowledge the collaboration of CENAM, SIM and PTB for the financial support given to meet the expenses involved in such continental exercise. REFERENCES [1] Torres Guzmán J. C., Ramirez Ahedo D., General guidelines for the force national standards comparison within the Interamerican Metrology System (SIM), range up to 10 kn. [] Torres Guzmán J. C., Ramirez Ahedo D., Procedure for the force national standards comparison within the Interamerican Metrology System (SIM), range up to 10 kn. [3] Torres-Guzmán J. C., Ramirez Ahedo D., Data sheet file for: the force national standards comparison within the Interamerican Metrology System (SIM), range up to 10 kn. [4] Torres-Guzmán J., Jõger Jochen, Soriano B., Pressure Standards Comparison between Germany and (Primary and Secondary Laboratories). NCSL International, Proceedings of the 00 Workshop and Symposium (Pressure Metrology). San Diego, USA. 00. [5] Document No. 8. Noramet. 1998. [6] EAL P7, EAL Interlaboratory Comparisons. 1996. [7] Youden W. J., Graphical Diagrams of Interlaboratory Test Results, Journal of Industrial Quality Control, Vol. 15, No. 11, May, 1959. Authors Addresses: J. C. Torres-Guzmán, D. A. Ramírez-Ahedo, Force and Pressure Division, Centro Nacional de Metrologia. km 4,5 carretera a Los Cués, El Marqués, Queretaro, Telephone [+5] (44) 11 057, fax [+5] (44) 11 0578 jorge.torres@cenam.mx dramirez@cenam.mx