Comparative study on accuracy of 13 C infrared spectrometer with different types

Similar documents
Levene's Test of Equality of Error Variances a

T-Test QUESTION T-TEST GROUPS = sex(1 2) /MISSING = ANALYSIS /VARIABLES = quiz1 quiz2 quiz3 quiz4 quiz5 final total /CRITERIA = CI(.95).

Extraction Process Validation of Isatis Radix

Mass spectrometry is an important part of multicomponent analytical systems designed for diagnosis of metabolic disorders. About 50 different

HYPOTHESIS TESTING II TESTS ON MEANS. Sorana D. Bolboacă

East West Acupuncture & Wellness Center, Inc. Patient Intake Form

LECTURE 1: ELECTROMAGNETIC RADIATION

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400

Chapter 4: Atoms and Elements

McGill University. Faculty of Science MATH 204 PRINCIPLES OF STATISTICS II. Final Examination

MANOVA is an extension of the univariate ANOVA as it involves more than one Dependent Variable (DV). The following are assumptions for using MANOVA:

BIOL 51A - Biostatistics 1 1. Lecture 1: Intro to Biostatistics. Smoking: hazardous? FEV (l) Smoke

BIOS 6222: Biostatistics II. Outline. Course Presentation. Course Presentation. Review of Basic Concepts. Why Nonparametrics.

Coupling Analysis of ECG and EMG based on Multiscale symbolic transfer entropy

The table shows the average background radiation dose from various sources that a person living in Britain receives in one year.

What do the nuclei of different molybdenum isotopes have in common?

Inference for Proportions

Inference for Proportions

Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data

Childhood Nephrotic Syndrome

Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY

The Future of Healthcare? W H A T D O E S T H E F U T U R E H O L D? The Empowered Consumer

Lab [30 pts] Name A Simulation of Radioactive Decay

MAT 2379, Introduction to Biostatistics, Sample Calculator Questions 1. MAT 2379, Introduction to Biostatistics

EARLY DETECTION OF INVASIVE FUNGAL INFECTION

The inductive effect in nitridosilicates and oxysilicates and its effects on 5d energy levels of Ce 3+

AEPHY: Nuclear Physics Practise Test

HYDROXYPROLINE IN URINE by UV/VIS Code Z01010

EGD (Upper Endoscopy)

Introduction to Crossover Trials

with the usual assumptions about the error term. The two values of X 1 X 2 0 1

Infusor / Intermate Patient Guide

Radioactivity Questions NAT 5

Yen Chiropractic, LLC Personal Injury Patient Intake Form Page 1

Treatment Intake Form

RUSSIAN ACADEMY OF SCIENCES A. M. PROKHOROV GENERAL PHYSICS INSTITUTE

Section 9.5. Testing the Difference Between Two Variances. Bluman, Chapter 9 1

TA: Sheng Zhgang (Th 1:20) / 342 (W 1:20) / 343 (W 2:25) / 344 (W 12:05) Haoyang Fan (W 1:20) / 346 (Th 12:05) FINAL EXAM

Section 6-5 THE CENTRAL LIMIT THEOREM AND THE SAMPLING DISTRIBUTION OF. The Central Limit Theorem. Central Limit Theorem: For all samples of

An Analysis of College Algebra Exam Scores December 14, James D Jones Math Section 01

One-Way ANOVA Source Table J - 1 SS B / J - 1 MS B /MS W. Pairwise Post-Hoc Comparisons of Means

Isotope Production for Nuclear Medicine

Fundamentals to Biostatistics. Prof. Chandan Chakraborty Associate Professor School of Medical Science & Technology IIT Kharagpur

Correlation and simple linear regression S5

P7 Radioactivity. Student Book answers. P7.1 Atoms and radiation. Question Answer Marks Guidance

PRELIMINARY RESULTS ON BRAIN MONITORING OF MENINGITIS USING 16 CHANNELS MAGNETIC IN- DUCTION TOMOGRAPHY MEASUREMENT SYSTEM

The independent-means t-test:

T test for two Independent Samples. Raja, BSc.N, DCHN, RN Nursing Instructor Acknowledgement: Ms. Saima Hirani June 07, 2016

Magnetic Janus Nanorods for Efficient Capture, Separation. and Elimination of Bacteria

Inference for Distributions Inference for the Mean of a Population

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Two-Way ANOVA. Chapter 15

How many protons are there in the nucleus of the atom?... What is the mass number of the atom?... (Total 2 marks)

Multiple Comparisons

Gentian Canine CRP Immunoassay Application Note for Siemens ADVIA * 1800/1650

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC

Wayne State University Doctor of Pharmacy Program. Curriculum and Course Prerequisites Classes of 2017, 2018, and 2019

By Mohsin Al-Saleh Clinical Biochemistry Department (SQUH)

Facile Synthesis and Catalytic Properties of CeO 2 with Tunable Morphologies from Thermal Transformation of Cerium Benzendicarboxylate Complexes

Potassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL

Online supplement. Absolute Value of Lung Function (FEV 1 or FVC) Explains the Sex Difference in. Breathlessness in the General Population

PLSC PRACTICE TEST ONE

Faculty of Health Sciences. Regression models. Counts, Poisson regression, Lene Theil Skovgaard. Dept. of Biostatistics

Radioisotopes and PET

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

Pearson Edexcel GCSE Physics Unit P3: Applications of Physics

FORMALDEHYDE IN URINE by UV Code Z65010

BIOSTATISTICS METHODS FOR TRANSLATIONAL & CLINICAL RESEARCH DAY #4: REGRESSION APPLICATIONS, PART C MULTILE REGRESSION APPLICATIONS

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510

Electronic Supplementary Information (ESI) A Green Miniemulsion-Based Synthesis of Polymeric Aggregation-Induced Emission.

x3,..., Multiple Regression β q α, β 1, β 2, β 3,..., β q in the model can all be estimated by least square estimators

a Sample By:Dr.Hoseyn Falahzadeh 1

DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER

The case for culturable air sampling for opportunistic fungi during healthcare construction

- Particle formation -

Tests for Population Proportion(s)

Nuclear Physics and Astrophysics

Multi-period medical diagnosis method using a single valued. neutrosophic similarity measure based on tangent function

Multiple linear regression S6

Draw one line from each type of radiation to what the radiation consists of.

Micro Volume QuEChERS kit

Practice problems from chapters 2 and 3

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Dependence structures with applications to actuarial science

STA 303H1F: Two-way Analysis of Variance Practice Problems

STA 303H1F: Two-way Analysis of Variance Practice Problems

Section 3: Nuclear Radiation Today

TECHNICAL APPENDIX WITH ADDITIONAL INFORMATION ON METHODS AND APPENDIX EXHIBITS. Ten health risks in this and the previous study were

A Comparison of Collaborative Stage with UICC TNM. Darlene Dale Head, Princess Margaret Cancer Registry June 11, 2013

National Standard of People s Republic of China

PubH 7405: REGRESSION ANALYSIS MLR: BIOMEDICAL APPLICATIONS

MTH U481 : SPRING 2009: PRACTICE PROBLEMS FOR FINAL

sample What happens when we are exposed to radiation? 1.1 Natural radiation Cosmic radiation

ALLOWAY METHOD OUTLINE

2018 TANEY COUNTY HEALTH DEPARTMENT. Communicable Disease. Annual Report

For the rapid, sensitive and accurate measurement of Urea levels in various samples.

Biochemistry THE FUNDAMENTAL CHEMISTRY OF LIFE

A Compact Reactor-Pump-Cell-Injection System for In-Situ / On- Line Spectroscopic Studies

Carcino embryonic Antigen Human ELISA Kit

IMPROVING EDUCATION QUALITY OF THE SPECIALTY OF NUCLEAR CHEMICAL AND FUEL ENGINEERING BY USING SUPERIORITY OF CDUT

Transcription:

Comparative study on accuracy of 13 C infrared spectrometer with different types East China Sanatorium, Wuxi, Jiangsu Province, 214065 Abstract Objective: To compare the consistency on test results between the domestic (HY-IREXB, Guangzhou Richen-Frinse Optical & Electronic Co., LTD) and imported (HeliFAN, Fischer GmbH, Germany) 13 C infrared spectrometer via 13 C breath test. Methods: The breath samples of the same patient are collected at the same time and tested respectively on two instruments. DOB values of test results from two instruments are compared, one-way ANOVA are carried out, and the factors are determined including diagnostic sensitivity and specificity of the diagnosis results, positive predictive value, negative predictive value, as well with Kappa consistency. Conclusions: The diagnosis results on two groups with domestic (HY-IREXB) and imported (HeliFAN) 13 C infrared spectrometer are exactly consistent, without any statistical differences. For the domestic 13 C infrared spectrometer, its sensitivity is 100%, specificity 100%, positive predictive value 100% and negative predictive value 100%, with Kappa value equal to 1. So it can be used as the preferred method for clinical detection of Helicobacter pylori gastric infection. Keywords: 13 C breath test, infrared spectrometer, East China sanatorium, Helicobacter pylori, stomach I. Introduction 13 C urea breath test ( 13 C-urea breath test, 13 C-UBT) is a non-invasive method to detect Helicobacter pylori gastric infection. Since 13 C is a stable and non-radioactive isotope without harm on humans and applicable for the detection on all population even including children and pregnant women, 13 C-UBT has been widely used for the diagnosis of HP infection currently. Imported 13 C infrared spectrometer (HeliFAN) was used in the domestic medical institutions very early, with good detection effects recognized by doctors. Domestic 13 C infrared spectrometer (HY-IREXB) is the first instrument under independent research and development at home, and has been used for some time in China. Our study compares the test results of both instruments, and makes evaluations on their clinical significance. II. Materials and Methods 1. Clinical information There are totally 147 cases with comparative data from 232 cases for physical examination in our hospital. 2. Materials

Kit (Heliforce 75mg, Beijing Boran pharmaceutical company); spectrometer: domestic (HY-IREXB) and imported (HeliFAN) 13 C infrared spectrometer 3. Method for 13 C-UBT 1) Patients shall be fasting before test in the morning; 2) Patients maintain normal breathing, and blow the gas into the air bag until full, and immediately tighten its cover for collecting sample of 0 min. 3) Patients take a bottle of [13C] urea mixed reagent (Heliforce) with 80-100ml cool drinking water and then keep waiting quietly. 4) According to the collection methods above, patients collect the sample of 30min after taking Heliforce for 30 minutes and immediately tighten the air bag cover. 5) Breath samples of 0 minute and 30 minutes will be tested respectively on domestic (HY-IREXB) and imported (HeliFAN) 13 C infrared spectrometer at the same time. 4. Statistical analysis The difference between the two sets of DOB data is analyzed by the method of one-way analysis of variance (one way ANOVA), and the consistency of the test results for both methods is evaluated by Kappa value. As for the instruments, the sensitivity, specificity, positive predictive value, and negative predictive value are also calculated respectively. 5. Results The positive rates of the 13 C-UBT results (see Table 1) from domestic (HY-IREXB) and imported (HeliFAN) 13 C infrared spectrometers have no significant difference statistically (see Table 2); the DOB value and Sig. Value measured by two instruments have homogeneity in variances and no significant difference statistically (see Table 3 and 4); the Kappa value of two methods is 1 (147/147) (see table 5), with high compliance. Taken the test results from imported 13 C infrared spectrometer (HeliFAN) as the gold standard, the values are calculated on domestic 13 C infrared spectrometer (HY-IREXB) as: the sensitivity is 100% (86/86), the specificity is 100% (61/61), the positive predictive value is 100% (86 / 86), and the negative predictive value is 100% (61/61). Table 1 Test Results domestic 13C infrared spectrometer (HY-IREXB) imported 13C infrared spectrometer (HeliFAN) Model No.: 14400490 Series No. Code Test time Test date DOB Positive or Negative DOB Positive or Negative 1 00001 08:34:55 14-03-19 30. 6 + 32. 5 + 2 00002 08:38:06 14-03-19 10. 5 + 13. 7 + 3 00003 08:41:18 14-03-19 0, 3 1. 3 4 00004 08:44:29 14-03-19 1. 0 + 9.8 + 5 00005 08:47:40 14-03-19 7.7 + 11. 1 + 6 00006 08:50:57 14-03-19 0. 5 1

7 00007 08:58:08 14-03-19 2. 5-1. 1 8 00008 09:01:20 14-03-19 44. 1 + 47. 1 + 9 00009 09:04:31 14-03-19 0. 8 1. 6 10 00010 09:07:42 14-03-19 35. 1 + 36. 8 + 11 00011 09:12:03 14-03-19 0. 4 0. 4 12 00012 09:15:15 14-03-19 25. 7 + 20. 2 + 13 00013 09:18:27 14-03-19 0. 8 0. 3 14 00014 09:21:38 14-03-19 60. 9 + 63. 5 + 15 00015 09:24:50 14-03-19 0. 8 0. 9 16 00016 09:28:56 14-03-19 0. 5 1. 1 17 00017 09:32:08 14-03-19 5. 8 + 7.4 + 18 00018 09:35:20 14-03-19 24. 8 29. 1 + 19 00019 09:38:32 14-03-19 0. 4 1. 9 20 00020 09:41:44 14-03-19 0. 3 0. 9 21 00021 09:46:36 14-03-19 6. 4 + 6. 7 + 22 00022 09:49:48 14-03-19 19. 6 + 20. 6 + 23 00023 09:53:00 14-03-19 30. 4 + 33. 5 + 24 00024 09:56:12 14-03-19 0. 8 0. 7 25 00025 13:57:46 14-03-19 17. 2 + 18.4 + 26 00026 14:04:29 14-03-19 13. 3 + 15. 1 + 27 00027 14:12:53 14-03-19 12. 3 + 12.4 + 28 00028 14:16:05 14-03-19 9. 3 + 8. 4 + 29 00029 14:19:17 14-03-19 0. 6 0. 5 30 00030 14:22:29 14-03-19 0. 1 1. 2 31 00031 14:27:08 14-03-19 19. 0 + 18. 6 + 32 00032 14:30:20 14-03-19 0. 3 0. 4 33 00033 14:33:32 14-03-19 0. 1-0. 1 34 00034 14:36:44 14-03-19 0. 6 0. 9 35 00035 14:39:56 14-03-19 2. 3 0. 5 36 00036 14:43:35 14-03-19 5. 8 + 1. 2 + 37 00037 14:46:47 14-03-19 0. 3 0. 5 38 00038 14:49:59 14-03-19 18. 5 + 18. 8 + 39 00039 14:53:11 14-03-19 0. 2 1. 3 40 00040 14:56:23 14-03-19 18. 1 + 18. 8 + 41 00041 15:01:09 14-03-19 1. 8 + 9.6 + 42 00042 15:04:21 14-03-19 8. 1 + 8. 1 + 43 00043 15:07:33 14-03-19 23. 5 + 21. 7 + 44 00044 15:10:45 14-03-19 26. 1 + 26. 7 + 45 00045 15:13:57 14-03-19 32. 3 + 34.4 + 46 00046 15:17:14 14-03-19 5. 1 + 6. 3 + 47 00047 15:26:51 14-03-19 0. 2 0. 2 48 00048 07:39:47 14-03-20 78.4 + 80. 2 + 49 00049 07:42:58 14-03-20 16. 1 + 17. 1 +

50 00050 07:46:09 14-03-20 16. 0 + 15. 8 + 51 00051 08:15:20 14-03-20 24.4 + 21. 5 + 52 00052 08:18:32 14-03-20 0. 1 0. 4 53 00053 08:21:43 14-03-20 16. 5 + 15. 3 + 54 00054 08:32:11 14-03-20 0. 8 1. 1 55 00055 08:48:21 14-03-20 22. 0 + 25 + 56 00056 08:51:33 14-03-20 25. 0 + 30. 3 + 57 00057 08:58:05 14-03-20 10. 5 + 10.4 + 58 00058 09:02:13 14-03-20 9. 3 + 10. 8 + 59 00059 09:18:22 14-03-20 0. 5 0. 5 60 00060 09:21:34 14-03-20 0. 6 0. 2 61 00061 09:37:41 14-03-20 19.4 + 20. 8 + 62 00062 09:53:46 14-03-20 17. 2 + 18 + 63 00063 10:03:26 14-03-20 11. 3 + 8 + 64 00064 10:06:37 14-03-20 17. 1 + 17. 5 + 65 00065 10:13:02 14-03-20 10. 8 + 12. 5 + 66 00066 09:28:46 14-03-21 0. 5 0. 7 67 00067 09:31:57 14-03-21 3. 7 1. 5 68 00068 09:35:08 14-03-21 0. 8 1 69 00069 09:38:19 14-03-21 32. 2 + 36. 3 + 70 00070 09:41:31 14-03-21 67.4 + 69. 4 + 71 00071 09:45:48 14-03-21 12.4 + 14. 2 + 72 00072 09:48:59 14-03-21 0. 8 1. 6 73 00073 10:04:57 14-03-21 0. 9 0 74 00074 10:08:08 14-03-21 0. 7 0. 3 75 00075 10:11:20 14-03-21 16. 1 + 16. 4 + 76 00076 10:14:31 14-03-21 14. 0 + 14. 8 + 77 00077 10:17:43 14-03-21 0. 4 0. 9 78 00078 10:27:57 14-03-21 38. 6 + 40. 1 + 79 00079 10:31:09 14-03-21 19. 3 + 24. 8 + 80 00080 11:03:26 14-03-21 33. 9 + 34. 8 +

domestic 13C infrared spectrometer (HY-IREXB) imported 13C infrared spectrometer (HeliFAN) Model No.: 14400479 Series No. Code Test time Test date DOB Positive or Negative DOB Positive or Negative 1 00001 08:43:33 14-03-19 18. 8 + 15. 9 + 2 00002 08:46:44 14-03-19 11.8 + 11. 2 + 3 00003 08:49:55 14-03-19 31. 3 + 35. 6 + 4 00004 08:53:06 14-03-19 42. 0 + 39. 3 + 5 00005 08:58:34 14-03-19 0. 6 1. 9 6 00006 09:02:21 14-03-19 18. 9 + 20. 1 + 7 00007 09:05:32 14-03-19 0. 3 1. 2 8 00008 09:08:43 14-03-19 0. 3 0. 9 9 00009 09:11:54 14-03-19 3. 5 2. 5 10 00010 09:15:06 14-03-19 0. 5 1. 1 11 00011 09:18:37 14-03-19 12. 8 + 11.5 + 12 00012 09:21:48 14-03-19 40.0 + 43. 1 + 13 00013 09:25:00 14-03-19 28. 1 + 25. 4 + 14 00014 09:28:12 14-03-19 4. 0 + 7. 7 + 15 00015 09:31:24 14-03-19 0. 0 1. 3 16 00016 09:34:41 14-03-19 0. 7 3. 8 17 00017 09:37:53 14-03-19 0. 5-0. 5 18 00018 09:41:05 14-03-19 0. 3 0. 7 19 00019 09:44:17 14-03-19 27. 3 + 30. 1 + 20 00020 09:47:29 14-03-19 0. 0 1. 7-21 00021 14:03:05 14-03-19 12. 1 + 16. 8 + 22 00022 14:09:13 14-03-19 31. 8 + 35. 9 + 23 00023 14:12:25 14-03-19 42. 9 + 45. 5 + 24 00024 14:15:37 14-03-19 42. 2 + 42. 4 + 25 00025 14:18:48 14-03-19 7. 9 + 8. 9 + 26 00026 14:22:00 14-03-19 2. 2 1. 9 27 00027 14:25:17 14-03-19 0. 0 0. 8 28 00028 14:28:29 14-03-19 0. 0 1. 1 29 00029 14:31:41 14-03-19 49. 8 + 48 + 30 00030 14:34:53 14-03-19 0. 8 0. 6 31 00031 14:38:05 14-03-19 0. 4 0. 1 32 00032 14:41:22 14-03-19 0. 7 0. 2 33 00033 14:44:34 14-03-19 26. 0 + 25. 7 + 34 00034 14:47:46 14-03-19 12. 6 + 12. 5 + 35 00035 14:50:58 14-03-19 15. 1 + 20 +

36 00036 14:54:10 14-03-19 0. 2 1. 9 37 00037 14:57:39 14-03-19 11. 9 + 15. 4 + 38 00038 15:00:51 14-03-19 2. 0 1. 3 39 00039 15:04:03 14-03-19 0. 2 1. 6 40 00040 15:07:15 14-03-19 9. 0 + 13. 1 + 41 00041 15:10:27 14-03-19 30. 0 + 34. 2 + 42 00042 15:14:23 14-03-19 48. 0 + 55. 5 + 43 00043 15:17:35 14-03-19 0. 3 1. 4 44 00044 15:20:47 14-03-19 50. 6 + 56. 4 + 45 00045 15:23:59 14-03-19 57. 3 + 57. 1 + 46 00046 08:03:17 14-03-20 13. 7 + 8. 5 + 47 00047 08:52:29 14-03-20 0. 9 0. 5 48 00048 08:55:41 14-03-20 33. 7 + 37. 6 + 49 00049 08:59:57 14-03-20 0. 9 0. 1 50 00050 09:06:21 14-03-20 44. 7 + 44. 7 + 51 00051 09:16:00 14-03-20 64. 2 + 67. 2 + 52 00052 09:25:36 14-03-20 1. 0 1. 2 53 00053 09:35:18 14-03-20 0. 5 0. 4 54 00054 09:48:11 14-03-20 22. 1 + 19. 8 + 55 00055 10:04:18 14-03-20 20. 9 + 22. 6 + 56 00056 10:07:30 14-03-20 14. 2 + 18.8 + 57 00057 09:33:09 14-03-21 0. 5 3.3 58 00058 09:36:20 14-03-21 38. 6 + 40 + 59 00059 09:39:31 14-03-21 14. 7 + 10.8 + 60 00060 09:42:42 14-03-21 0. 8 2.3 61 00061 09:45:54 14-03-21 0. 4 0.4 62 00062 09:52:47 14-03-21 0. 1 0.2 63 00063 09:25:19 14-03-21 0. 4 2.2 64 00064 09:28:30 14-03-21 15. 5 + 16.5 + 65 00065 09:31:42 14-03-21 18. 5 22.1 + 66 00066 09:34:53 14-03-21 0. 1 2 67 00067 09:38:05 14-03-21 24. 6 + 23.5 +

Table 2 Comparison on tests results of domestic and imported 13 C infrared spectrometers Method spectrometer + positive result ratio 13C UBT HY-IREXB 86 61 58. 80% HeliFAN 86 61 58. 80% Table 3 Descriptives DOB 95% Confidence Interval N Mean Std. Std. for Mean Minim Deviation Error Lower Upper um Maximum Bound Bound HY-IREX 147 14. 301 16. 5879 1. 3681 11. 597 17. 005 0 78. 4 B HeliFAN 147 15. 215 17. 2403 1. 422 12. 405 18. 025-1. 1 80. 2 Total 294 14. 758 16. 8946 0. 9853 12. 819 16. 697-1. 1 80. 2 Note: P=0.634 >0.05 Table 4 Test of Homogeneity of Variances (DOB values) Levene Statistic dfl df2 Sig. 0. 185 1 292 0. 667 Note: P=0.634 >0.05 Table 5 Distribution of tests results from domestic and imported 13C infrared spectrometers HY-IRXB HeliFAN + - total + 86 0 86-0 61 61 total 86 61 147

III. Discussion Currently, HP infection has been recognized as a major cause for chronic active gastritis, purulent ulcer, and has relationships with a variety of stomach cancer [1]. In 1994, the World Health Organization included the HP level as one biological carcinogen factor at first class. Successful eradication of HP can cure the related diseases, and therefore, the detection of HP is very important. According to statistical data, there are much difference on the stomach HP detection rates with 30-80% in different Chinese regions and different ethnic groups [1]. In terms of HP inspection, the clinics are constantly looking for more simple, more secure, and more convenient diagnostic method for patients. It can be classified as invasive and non-invasive two types according to the need for endoscopy or not. Since 13 C is a stable and non-radioactive isotope without harm on humans and applicable for the detection on all population even including children and pregnant women, 13 C-UBT has been widely used for the diagnosis of HP infection currently, and this method is particularly suitable for large-scale health institutions. Imported 13 C infrared spectrometer (HeliFAN) was used in the domestic medical institutions very early, with good detection effects recognized by doctors. Domestic 13 C infrared spectrometer (HY-IREXB) is the first instrument under independent research and development at home, and is applied later than the Germany instrument for some time. Our study compares the test results of both instruments, and makes evaluations on their clinical significance. And it can be concluded that the DOB value and Sig. value measured by two instruments have homogeneity in variances, and no any statistical differences, with the exact same clinical diagnosis results. Compared to the Germany instrument HeliFAN, the domestic 13 C infrared spectrometer (HY-IREXB) has its sensitivity, specificity, positive predictive value and negative predictive value all equal to 100%. In this paper, the data were collected on 232 cases in three days, with high detection frequency at 80 cases per day. By using advanced MEMS technology, domestic 13 C infrared spectrometer (HY-IREXB) has high precision, high stability, rapid detection with 10 channels. Compared to the 4-channel Germany instrument HeliFAN, it has more advantage and shall be the first choice for large medical institutions to detect H. pylori. Reference [1] Wandai Zhang, Shudong Xiao, Fulian Hu, Common opinions on the relevant issues of H. pylori (2003, China), [J] National Medical Journal of China, 2004, 84(6): 522