INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

Similar documents
A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

APMP.T-K3.4: Key comparison of realizations of the ITS-90 over the range C to C

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Traceability of temperature measurements in Estonia

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

CERTIFICATE OF CALIBRATION

REPORT to the CCT on Comparison COOMET. T-K3 (COOMET theme No. 285/RU-а/03) Final Report

INTERCOMPARISON OF WATER TRIPLE POINT CELLS FROM INTIBS AND INRIM 1. INTRODUCTION

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

BUREAU INTERNATIONAL DES POIDS ET MESURES

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

FINAL REPORT OF THE BILATERAL COMPARISON OF THE CALIBRATIONS OF STANDARD WEIGHTS BETWEEN CENAM-MEXICO AND INEN-ECUADOR SIM.M.M-S4 (SIM.7.

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

Ajchara Charoensook, Chaiwat Jassadajin. National Institute of Metrology Thailand. Henry Chen, Brian Ricketts and Leigh Johnson

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

BUREAU INTERNATIONAL DES POIDS ET MESURES

PROTOCOL OF MASS AND VOLUME COMPARISONS BETWEEN SIM NMIs

A Case for Periodic Calibration or Verification of RTDs

Document No: TR 12 Issue No: 1

Final Report for the APMP.T-K4 (Draft B on October 27, 2011)

Comparison protocol EURAMET project

CALIBRATION REPORT FOR THERMOHYDROMETER

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

CALIBRATION AND ACCURACY OF SECONDARY STANDARD PRT'S TO THE ITS-90

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

EUROMET Project 702 EUROMET.M.D-K4

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Euramet EM-S40. Bilateral Comparison KIM-LIPI / LNE. Final Report

European Association of National Metrology Institutes

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

Combined method for establishment and dissemination of the international temperature scale

Certificate of Accreditation to ISO/IEC 17025:2005

ANNEXE 8. Technical protocols for the interlaboratory comparisons

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Calibration and verification: Two procedures having comparable objectives and results

A guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006

INTERLABORATORY MASS COMPARISON BETWEEN LABORATORIES BELONGING TO SIM SUB-REGIONS COORDINATED BY CENAM (SIM.7.31a & SIM.7.31b)

Report on EURAMET.M.M-S9

Final Report of COOMET.T-K7: Regional key comparison of water triple point cells (COOMET theme No 395/BY/07)

Calibration of Temperature Block Calibrators

Certificate of Accreditation

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

SIM FORCE STANDARDS COMPARISON UP TO 10 kn

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

CERTIFICATE OF ACCREDITATION

Final Report. SIM / ANDIMET Supplementary Comparison for Volume of Liquids at 100 ml and 100 µl SIM.M.FF-S7

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

MASS AND VOLUME COMPARISONS AT MIKES

RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

SIM Regional Comparison on. The Calibration of Internal and External Diameter Standards SIM.L-K FINAL REPORT July 2012

Portuguese Institute for Quality. Contents

HOW TO EXPRESS THE RELIABILITY OF MEASUREMENT RESULTS. The role of metrological traceability and measurement uncertainty-

Appendix B1. Reports of SMU

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

SIM SIM.M.D-K3. provide. SIM.M.M-K5 for. A set of. Institute LACOMET LATU INTI. Country Costa Rica Uruguay Argentina Chile México Canada Brazil

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT

Physikalisch- Technische Bundesanstalt

Calibration of temperature sensors within Length Standard Section of NMIJ

R SANAS Page 1 of 7

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

Final report on CCQM-K36.1 with erratum

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

Certificate of Accreditation

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

Comparison of V and 10 V DC Voltage References

CCM short note on the dissemination process after the proposed redefinition of the kilogram

The Journey from Ω Through 19 Orders of Magnitude

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Physikalisch Technische Bundesanstalt

Revision of the Guide to the expression of uncertainty in measurement impact on national metrology institutes and industry

Certificate of Accreditation

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards.

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Centre for Metrology and Accreditation, MIKES

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

A bilateral comparison of a 1 kg platinum standard

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm

Final Report August 2010

Transcription:

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR M. Araya 1, D. Almeida 2 1 National Laboratory of Temperature of Chile, LCPNT, Santiago, Chile 2 Ecuadorian Standardization Institute, INEN, Quito, Ecuador E-mail: maraya@cesmec.cl dalmeida@inen.gob.ec Abstract. An International Bilateral Comparison of Platinum Resistance Thermometers was performed between the National Metrology Institutes of Chile (LCPNT) and the Ecuadorian Standardization Institute of Ecuador (INEN), both represented by their National Laboratories of Temperature. This comparison was carried out in the range from -39 C up to 232 C. The final results obtained by each laboratory showed to be equivalent (EN < 1) in the measured temperature range. Keywords: Temperature, comparison, LCPNT, Chile, Ecuador, INEN, SIM. INTRODUCTION This comparison was developed under the technical and quality cooperation project signed between the government of Ecuador and the European Union, executed between 2007 and 2008, in which the national temperature laboratory of INEN participated. Two steps were agreed and the whole process was completed between 2007 and 2008. The pilot laboratory was the National Temperature Laboratory of Chile (LCPNT), which is accredited by the German Accreditation Body DAkkS (earlier DKD) since 2001. COMPARISON General guidelines A measurement protocol was given to each participant. It stated the following relevant technical aspects: The first step was performed to compare the calibration of two Industrial Platinum Resistance Thermometer IPRT of 100 Ohm, covering the temperature range from -39 C up to 232 C. These artifacts are calibrated using the reference thermometers of each participant. The working instructions of each participant comply with the calibration by comparison Techniques for approximating the International Temperature Scale of 1990 [1] used the ITS-90 deviation equations and the calculation procedure cross of least squares. The deviation equations used for the range (-40 ºC to +30 ºC): was W(T 90 ) Wr (T 90 ) = A ( W(T 90 ) 1) + B ( W(T 90 ) 1) 2 and for the range (0 ºC to +232 ºC): was W(T 90 ) Wr (T 90 ) = A ( W(T 90 ) 1) + B ( W(T 90 ) 1) 2. Calibrations of each participant were performed by direct comparison in stirred liquid baths. Results were analyzed using the difference in temperature obtained for each artifact when using the calibration coefficients determined for the participants and the Normalized Error (E N ). Only one measurement at -39 C showed an E N > 1 and corrective actions were agreed. A final report for this first step was emitted and signed by both participants. The second step consisted in a comparison of an Standard Platinum Resistance Thermometer of 25.5 Ohm, covering the temperature range from -39 C up to 232 C. The same first step protocol to complete the measurements and analyze the results was followed. All temperatures show an E N < 1. A final report for this second step was emitted and signed by both participants. Uncertainties were evaluated according to the ISO Guide to the Expression of Uncertainty in Measurement [2]. Protocol for Bilateral Comparison for the SPRT Calibration by Comparison Over the Range from - 39 C to 232 C Participant information is as follows: LCPNT Chile

TABLE 1. Measurement Sequence Measurement Sequence Contact: Mauricio Araya Santiago CHILE INEN Ecuador Contac: Diego Almeida Quito - Ecuador The instructions and procedures given below must be followed by the participants. The laboratories agree to follow the general instructions and technical protocol written in this document, the MRA Appendix F document Guidelines for CIPM Key Comparisons, and the JCRB document A Note on Supplementary Comparison by T.J. Quinn. The MRA Appendix F and JCRB documents are found at: 1) www.bipm.fr\pdf\guidelines.pdf, 2) and, www.bipm.org/cc/documents/jcrb/supplementary_ Compns.doc, respectively. Description 4) The SPRTs is returned to the LCPNT and is calibrated a second time to close the loop. E-mail: maraya@cesmec.cl 5) Results after each measurement set (steps 1, 2, Contact: and 3) are Ing. sent Diego to INEN. Almeida (Table 1). Measurement Sequence) Technical Instructions Time Period 1 LCPNT complete measurements of SPRT (pre-submission to INEN) 2007, November 2 LCPNT deliver SPRT to INEN 2008, June 3 INEN complete measurements 2008, July 4 LCPNT complete measurements of SPRT (post-submission to INEN) 2008, September 5 Draft A Report submitted to participants 2008, September LCPNT measurements (Pre-submission to INEN) The 25.5 Ω SPRT chosen by the LCPNT must meet the ITS-90 purity requirements and may have either a glass or metal sheath. The triple point of (TPW) was measured after every annealing cycle. After annealing and stabilizing the SPRT, the measurement sequence to be followed is the TPW, 232 C, 200 C, 157 C, TPW, 80 C, 30 C, 15 C, 0 C (Ice Point optional), TPW, -20 C, -39 C and TPW. All results must be reported for 1 ma. This comparison is designed as a bilateral comparison. The LCPNT supply one SPRT 25 Ohm. Comparison will be performed by a calibration by comparison covering the range of temperature from -39 C to 232 C (points; -39 C, -20 C, 0 C, 15 C, 30 C, 80 C, 157 C, 200 C, 232 C). Participants will use their own procedures to perform their calibrations. Results of the comparison will be calculated by the LCPNT personnel. Transfer artifacts The LCPNT will supply one SPRT for the comparison. The measurement sequence of the SPRT is as follows: 1) The SPRT is calibrated by the LCPNT by 2) comparison prior to sending the artifacts to INEN. 3) The SPRT is calibrated by the INEN by comparison. Calibration certificate is sent to LCPNT. INEN measurements Upon receipt of the SPRT, INEN will inspect the artifacts for damage. If there is damage, the LCPNT and the INEN will discuss and agree on how to proceed. If no damage has been sustained, the SPRT will be measured at the TPW upon receipt. The as received [R(273.16 K) Ω, 0 ma] values will be reported to the LCPNT prior to annealing the SPRT. Prior to the comparison measurements, the SPRT will be stabilized by annealing at a temperature of 425 C. The stability criteria value for an SPRT is calculated from the difference between the R(273.16 K) Ω, 0 ma values measured before and after an anneal, the results are described in the tables 6 and 8. After annealing and stabilizing the SPRT, the measurement sequence to be followed is the TPW, 232 C, 200 C, 157 C, TPW, 80 C, 30 C, 15 C, 0 C

(or Ice Point), TPW, -20 C, -39 C and TPW. All results must be reported for 1 ma. On completion of the set of measurements, INEN return the artifacts to LCPNT. Transport may be in the form of hand carrying or shipping. Hand carrying is the preferred method. LCPNT measurements (Post-submission to INEN) In order to close the measurement loop, the LCPNT will repeat steps outlined above in the LCPNT measurements (Pre-submission to INEN) section of this protocol. Reporting of Data The participating laboratories must submit the following: 1. Record the artifact Rt(1 ma) results, reference t90 (defined by the participating laboratory) and ITS-90 deviation function coefficients. 2. A calibration certificate will be submitted. Reporting of Uncertainties The individual uncertainty components should be listed along with the total combined uncertainty assigned to each of the calibration points. All expanded uncertainties should be expressed as k=2. In an effort to harmonize the calibration point uncertainty budgets used by the participants. The uncertainties declared in the calibration certificates by the participants are the reported like Calibration and Measurements Capabilities. The calibration point measurement reported in each calibration report and its associated uncertainties for the SPRT transfer artifact will be combined to generate only one difference and associated uncertainty for calibration point. If the SPRT transfer artifact fails or is found to be unstable during the comparison (e.g. SPRT changes at the TPW during the measurements by more than the laboratory s allowable limit), the participating laboratories will discuss and agree upon a course of action. If the SPRT is stable at LCPNT but unstable (e.g. SPRT changes at the TPW during the measurements by more than the laboratory s allowable limit) for the second set of data, then only the first set of LCPNT data will be used. The two outcome results to be reported are: 1. bilateral differences with associated uncertainties at each measured calibration point among the participants, 2. the Normalized Error (EN). Will be used the uncertainties (k=2) reported in the calibration certificates of the participants. Artifacts The LCPNT provided the artifacts to be compared in both steps. Artifacts features are shown in table 1.1. Standards and Equipment Features about standards and equipment are shown in table 2. Complementary equipment is shown in table 3. TABLE 1.1. Comparison artifacts feature. Manufacturer Model Series Range of Measurement R(0 C) [] Nominal Alpha Coefficient [// C] Dimension / Min. Immersion [cm] Date of Measurement Isotech T100-NTS587 K0812A -40 a 232 100 0,00392 0,635 x 46 / 30 2007 ASL T100-450-3 B595354-40 a 232 100 0,00392 0,635 x 45 / 30 2007 Hart Scientific 5628 61084-40 a 232 25 0,00392 0,7 x 51 / 30 2008

TABLE 2. Standards and equipment Laboratory Model Manufacturer Model Serial Number Measure /Calibration Range Calibration Method Traceability SPRT 25 Ω Rosemount 162 CG 4592 SPRT 25 Ω Rosemount 162 CG 4593 LCPNT Resistor 100 Ω Tinsley 5685 A 274560 100 Ohm Comparison CENAM Bridge ASL F18 4135-001-163 0 1,2999999 -- -- Triple point of NPL Type 32 1064 0,01 C Comparison NPL SPRT 25 Ω Hart Scientific 5681 1503 SPRT 25 Ω Hart Scientific 5681 1514 INEN Bridge Fluke- Hart Fluke- Hart 1590 A63325 1Ω 10KΩ Comparison Scientific Scientific Triple point of Hart Hart Scientific Type B D-G 1096 0,01 C Comparison Scientific TABLE 3. Complementary equipment Laboratory Description Manufacturer Model Medium LCPNT Stirred liquid bath Hart Scientific 7037 Alcohol / distilled Stirred liquid bath Hart Scientific 6022 Silicon oil INEN Stirred liquid bath Hart Scientific 7312 Alcohol / distilled Stirred liquid bath Isotech 915 Alcohol Stirred liquid bath Hart Scientific 6022 Silicon oil Comparison round. The comparison sequence for both steps was performed as follows: LCPNT performs the initial measure, LCPNT transports the artifacts by hand to INEN, INEN performs its measure, LCPNT transports the artifacts by hand (return), LCPNT checks the artifact to complete its measure. RESULTS participant, the Rt 90 [Ohm] reported in the calibration certificate by the INEN is used to fix a temperature point to be compared. The temperature t 90 calculated by each participant for each artifact was obtained using the coefficients informed in the individual calibration certificate. The expanded uncertainties (k=2) U used to evaluate de normalized error E N correspond to the uncertainties informed by each participant in the individual calibration certificate for each artifact. The normalized error E N is defined by the following equation; Each participant documents their results in a traceable calibration certificate. To evaluate the temperature difference t(inen - LCPNT) between each

At the first step, results showed to be equivalent for both IPRTs Isotech (see table 4) and ASL (see table 5) in the measured temperature range. Only at the nominal temperature equal to -39 C, the E N > 1 was obtained for both artifacts. Investigation showed that the stirred liquid bath used by the INEN had a mechanical problem in the system of agitation, and the rheostat was repaired. Corrective actions were agreed for the second step. IPRTs instability measured at the triple point of R 0.01 [Ohm] for both artifact at the complete comparison round at the first step showed to be < 4 mk (see table 6). In the case of the second step, results showed to be equivalent for the Hart Scientific SPRT (see table 7) in the measured temperature range. All measured temperatures showed E N < 1. Corrective actions agreed in the first step showed to be appropriate. SPRTs instability measured at the triple point of R 0.01 [Ohm] at the complete comparison round at the second step showed to be < 2.5 mk (see table 8). TABLE 4. Results for Isotech IPRT, s/n K0812A Rt90 [] Nominal Value t90 / INEN 84.457 88-39 -38.768 30-38.809 15 40.5 1.21 (1) All uncertainties in table 4 are expressed as expanded uncertainties (k=2). TABLE 5. Results for ASL IPRT, s/n B535954 84.471 65-39 -38.768 30-38.808 15 40.5 1.21 (1) All uncertainties in table 5 are expressed as expanded uncertainties (k=2). TABLE 6. IPRTs instability at the first step U (1) / INEN t90 / LCPNT U (1) / LCPNT t(inen - LCPNT) 189.051 86 232 231.505 30 231.493 15 11.8 0.35 177.080 36 200 199.393 30 199.382 15 10.6 0.32 160.950 66 157 156.639 30 156.622 15 17.5 0.52 131.611 89 80 80.275 20 80.269 12 6.1 0.26 111.973 75 30 30.151 20 30.147 12 3.6 0.15 106.143 11 15 15.417 20 15.413 12 3.8 0.16 100.014 02 0 0.001 10-0.004 10 5.0 0.35 92.078 92-20 -19.850 30-19.856 15 6.2 0.19 Rt90 [] Nominal Value t90 / INEN U (1) / INEN t90 / LCPNT U (1) / LCPNT t(inen - LCPNT) 189.147 43 232 231.506 30 231.501 15 4.8 0.14 177.167 37 200 199.395 30 199.391 15 3.9 0.12 160.023 38 157 156.631 30 156.624 15 6.8 0.20 131.661 90 80 80.276 20 80.270 12 5.8 0.25 112.00845 30 30.151 20 30.148 12 2.9 0.13 106.173 23 15 15.417 20 15.413 12 3.6 0.15 100.039 59 0 0.001 10-0.004 10 4.1 0.29 92.097 74-20 -19.850 30-19.857 15 7.2 0.21 IPRT Laboratory R(0.01 C) [] T(max min) * Isotech ASL LCPNT (initial) 100.019 60 INEN 100.018 07 LCPNT (final) 100.018 60 LCPNT (initial) 100.044 80 INEN 100.043 99 LCPNT (final) 100.044 20 3.8 2.0 EN EN

* The T(max min) absolute value is calculated with de difference between the R(0.01 C) maximum R(0.01 C) minimum measured for each artifact by each participant. TABLE 7. Results for Hart Scientific SPRT, s/n 61084 Rt90 [] Nominal Value t90 / INEN 20.980 237-39 -39.220 20-39.220 15 0,5 0.02 (1) All uncertainties in table 7 are expressed as expanded uncertainties (k=2). TABLE 8. SPRTs instability at the second step U (1) / INEN t90 / LCPNT U (1) / LCPNT t(inen - LCPNT) 47.143 986 232 232.111 25 232.109 15 2,3 0.08 44.167 191 200 200.066 25 200.064 15 1,7 0.06 40.126 042 157 157.070 25 157.068 15 1,2 0.04 32.789 211 80 80.445 15 80.444 12 1,0 0.05 27.868 946 30 30.052 15 30.051 12 1,6 0.08 26.393 262 15 15.089 15 15.087 12 1,9 0.10 24.899 335 0 0,010 4 0.008 9 2,3 0.24 22.902 602-20 -20.037 20-20.038 15 1,4 0.06 SPRT Laboratory R(0.01 C) [] T(max min) * Hart Scientific LCPNT (initial) 24.899 58 INEN 24.899 34 LCPNT (final) 24.899 46 * The T(max min) absolute value is calculated with de difference between the R(0.01 C) maximum R(0.01 C) minimum measured for each artifact by each participant. 2.4 EN CONCLUSION Concluded the bilateral comparison, final results in the second step (see table 7) showed an improvement respect the results in the first step (see table 5 and 6). Improvement proposed by the LCPNT-CESMEC to INEN in its measurement system and its working instructions showed to be effective. Calibration performed by comparison in the range -40 C to 232 C by the participant showed to be equivalent in accordance with the claimed CMCs. ACKNOWLEDGMENTS This comparison was performed with partial support of Physikalisch-Technischen Bundesanstalt and the European Union. REFERENCES [1] R.E. Bedford, T.J. Quinn; Techniques for approximating the International Temperature Scale of 1990, Metrología, 1997. [2] Guide To The Expression Of Uncertainty In Measurement; ISO TAG 4 WG 3. BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML; 1995.