Infrared Calibration Development at Fluke Corporation Hart Scientific Division

Similar documents
RADIOMETRIC COMPARISON BETWEEN A NATIONAL LABORATORY AND AN INDUSTRIAL LABORATORY

Traceability and Quality Control in a Radiation Thermometry Laboratory

Traceability and Quality Control in a Radiation Thermometry Laboratory

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

Traceability and Quality Control in a Radiation Thermometry Laboratory

A Study on the Effects of Bandwidth of IR Thermometry Measurements

DETERMINING SIZE OF SOURCE FOR HANDHELD INFRARED THERMOMETERS THEORY AND PRACTICE

Standards for Radiation Thermometry

Infrared Thermometer Calibration A Complete Guide

Optimization of Thermal Radiation Source for High Temperature Infrared Thermometer Calibration

IR spectral characterization of customer blackbody sources: first calibration results

Infrared Temperature Calibration 101 Using the right tool means better work and more productivity

Validation of NIST s Low Temperature Infrared Spectral Radiance Scale

NIST Radiance Temperature and Infrared Spectral Radiance Scales at Near-Ambient Temperatures

Abstract. 1. Introduction

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

Emissivity modeling of thermal radiation sources with concentric grooves

Calibrating the Thermal Camera

Radiation properties of IR calibrators with V-grooved surfaces

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

DIRECT RADIOMETRIC TECHNIQUES

Introduction to Blackbody Sources

Sensors - April A Certified-Emissivity Blackbody for Calibrating Infrared Thermometers

On the effects of temperature dependence of spectral emissivity in industrial radiation thermometry

TEMPERATURE COEFFICIENTS OF TOPCON RADIATION THERMOMETERS

Guide to the Realization of the ITS-90

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

Status Update of the NIST Infrared Optical Properties of Materials Program

Total radiation measurements of thermodynamic temperature

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

Characterization of the VIIRS Blackbody Emittance

Potential for improved radiation thermometry measurement uncertainty through implementing a primary scale in an industrial laboratory

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

Calibration of Temperature Block Calibrators

Designed to go where you work

2005 VIRIAL, INC. Prepared by Alexander V. Prokhorov, Ph. D. and Vladimir B. Khromchenko, Ph. D.

Emissivity: Understanding the difference between apparent and actual infrared temperatures

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

Introduction to metrology

Uncertainty of temperature measurements by infrared thermography for metal cutting applications

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

Temperature and Synchrotron Radiation

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

Absolute Radiance Re-Calibration of FIRST

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Reading Problems , 15-33, 15-49, 15-50, 15-77, 15-79, 15-86, ,

On-orbit Absolute Blackbody Emissivity Determination Using the Heated Halo Method

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration

A Case for Periodic Calibration or Verification of RTDs

Operator's Manual. optris BR 400. Calibration source

CALIBRATION REPORT FOR THERMOHYDROMETER

High Accuracy High Speed in Your Process

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Examination of metallic surfaces for IR gray body sources

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Toward Reliable Industrial Radiation Thermometry

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Satish Chandra. Blackbody. Unit IV, BLACK BODY RADIATION. Radiation in a Hollow Enclosure. Pure Temperature Dependence

Calibration and validation of thermal imagers

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Measurement method for the proficiency testing program

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

Infrared thermography

The in-flight calibration system for the airborne imager GLORIA

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

InfraRed Thermometer with Laser Pointer

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

Towards SI traceable humidity measurements with radiosondes

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

The Fundamentals of Moisture Calibration

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Fig 1. Power Tower during Operation

Influence of sun radiation on results of non-contact temperature measurements in far infrared range

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

Routine Cal/Val Activities for GRUAN

Div. 1 Div. 2 Div. 3 Div.4 8:30 am 9:30 pm 12:30 pm 3:30 pm Han Xu Ruan Pan

1 THE CONCEPT OF TEMPERATURE

Thermal and structural design constraints for radiometers operating in geostationary orbits. G. E. Zurmehly, R. A. Hookman

UNCERTAINTY IN THE TEMPERATURE OF SILICON WAFERS MEASURED BY RADIATION THERMOMETRY BASED UPON A POLARIZATION TECHNIQUE

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

International Temperature Scale of 1990 From Wikipedia, the free encyclopedia

Maximum Performance From Dry-Well Thermometer Calibrators. Mark Finch Product Manager EMEA Fluke Calibration

Humidity Calibration Solutions

ROUGHNESS ASPECTS IN THERMAL ANALYSIS OF MACHINE TOOLS

ASSET INTEGRITY INTELLIGENCE. Featured Article. ACHIEVING A COMPREHENSIVE FIRED HEATER HEALTH MONITORING PROGRAM By Tim Hill, Quest Integrity Group

Calculation of the Effective Emissivities of Speculardiffuse Cavities by the Monte Carlo Method

Calibration of temperature sensors within Length Standard Section of NMIJ

Transcription:

Infrared Calibration Development at Fluke Corporation Hart Scientific Division Frank E. Liebmann, a Marco A. Cabrera Carrasco b a Fluke Corporation 799 Utah Valley Dr., American Fork, Utah, Estados Unidos de América. Frank.Liebmann@hartscientific.com b INyMET Instrumentación, S. A. de C. V. Salvatierra 32-3, Col. San Bartolo Atepehuacan, 07730, Distrito Federal, México. ABSTRACT A flat plate calibrator is one instrument used in calibrating infrared (IR) thermometers, primarily in the 8 µm to 14 µm band. One such family of flat plate calibrators is the 418X Precision IR Calibrator from Fluke Corporation Hart Scientific Division. This product is calibrated with a radiometric calibration. To support this radiometric calibration and its traceability, a number of developments have been made at Hart Scientific. These developments include the construction of a new IR calibration laboratory with radiometric traceability. This presentation discusses the research done to establish IR calibration capabilities. Among the topics discussed are the need for radiometric traceability for flat plate calibrators, the traceability chain to national laboratories included in radiometric calibrations at Hart Scientific, the development of blackbody cavity baths in Hart s IR calibration laboratory, and Hart Scientific s IR uncertainty budgets. 1. INTRODUCTION In the world of infrared (IR) thermometry, there has been much concern about the accuracy of IR thermometers. Contributing to this concern is a general misunderstanding of their use and operation. The two main aspects of this misunderstanding are a lack of knowledge of surface emissivity and a lack of knowledge of size of source or spot size. This is true as it applies to their use, and it is also true as it applies to their calibration. To address these two issues, Fluke - Hart Scientific has developed two new flat plate IR calibrators. These products are calibrated with a radiometric (non-contact) calibration. To properly support this calibration, Hart has developed an IR calibration laboratory with traceable blackbody cavities. The establishment of this metrology has been the result of much research and experimentation involving IR temperature measurement. It has resulted in NVLAP accreditation for the 418X calibration as well as the calibration of the radiometric transfer standard. 2. 418X PRECISION IR CALIBRATOR The 418X products are flat plate IR calibrators. The major application for these IR calibrators is the calibration of handheld IR thermometers in the 8 14 µm band. This range includes a bulk of the handheld IR thermometers sold today. A near blackbody cavity is the ideal calibrator for an IR or radiation thermometer. However, due to the large spot size or size of source effect of many handheld IR thermometers, the cavity s use is impractical for calibrating devices with larger spot sizes. This necessitates the use of a flat plate for IR thermometer calibrations. The 418X products consist of 2 models, the 4180 and 4181. The two products have a combined temperature range of 15 C to 500 C. They have a number of features that make them an improvement on previous flat plate offerings. Among these features are their radiometric calibration and the metrology and traceability behind that calibration. Along with the radiometric calibration comes a robust uncertainty budget calculated to account for numerous factors and calibration support for the radiometric calibration. Both will be discussed later in this paper. 2.1. Radiometric Calibration The 418X is calibrated with a radiometric calibration. The display temperature on the unit is based on this radiometric calibration. This display temperature shows the user what an IR thermometer with a given emissivity setting should read. This is called apparent temperature. In other words, the display temperature shows what temperature the target appears to be to the IR thermometer. The radiometric calibration is done with a calibrated Heitronics KT19.82II, referred to as a KT19 in this paper. This KT19 is an 8 µm 14 µm radiometer SM2008-M236-1197-1

which serves as a radiometric transfer standard. The KT19 is calibrated using blackbody cavities at Hart that will be discussed later in this paper. The purpose for the radiometric calibration is to account for factors that cannot be accounted for with a contact calibration. The two main factors are the difference between contact temperature and surface temperature, and the difference between UUT emissivity and the target s surface emissivity. The first problem with relying on a contact calibration is a lack of knowledge of the surface temperature. This is due to the fact that a contact calibration would not take place on the calibrator s surface, but below it, between the calibrator and the heat source. Heat flow creates a temperature drop. Tests at Hart Scientific have shown that this error can be up to 1,1 C at 500 C [1]. The second problem with relying on a contact calibration is the lack of knowledge of emissivity. Emissivity is especially troublesome because it can be both wavelength and temperature dependent. This makes assuming emissivity to be an arbitrary value questionable. Such an assumption can cause large temperature uncertainties. A surface s emissive dependence on wavelength and temperature can be verified by Fourier Transform Infrared (FTIR) tests [2, 3]. Results from two such tests are shown in Fig. 1. These graphs show how non-gray a material can be. Gray is defined as a material having the same emissivity regardless of wavelength [4]. 2.2. Use of the 8 µm 14 µm Band The 418X is meant for use in the 8 µm 14 µm band. This is because this band is common to most handheld IR thermometers. 2.3. Larger Target Size The 418X products provide the user with a 152 mm diameter flat plate. This gives the user a larger temperature controlled surface to calibrate IR thermometers. The target size needed to calibrate a given IR thermometer is dependent on the IR thermometer s optical system. Many IR thermometers come with a spot size diagram. These diagrams can be misleading because not all of the energy the IR thermometer detects is within the given spot diagram. To determine an IR thermometer s size of source effect, ASTM provides a method [5]. Also, the IR thermometer manufacture may be able to provide you with information on the target size needed for calibration. The distance between the IR thermometer and the IR calibrator is also important for an IR thermometer calibration. From testing done at Hart Scientific on a variety of IR thermometers [6], the spot-size diagram contained anywhere from 81 % to 98 % of the energy received by the IR thermometer. ASTM suggests that target size for a radiation thermometer be determined by the diameter where 99 % of the energy is received by the IR thermometer [5]. Emissivity Emissivity (%) Emissivity - Paint 1 100 98 96 94 92 90 88 86 84 82 80 0 2 4 6 8 10 12 14 16 18 20 Wavelength (um) Emissivity - Paint 2 100 98 96 94 92 90 88 86 84 82 80 0 2 4 6 8 10 12 14 16 18 20 Wavelength (um) 3. RADIOMETRIC CALIBRATION The 418X calibration is done with a radiometric transfer standard (KT19), which is calibrated with Hart Scientific s blackbody cavity baths. The cavity s radiometric temperature is determined by contact thermometry. To verify the quality of the data, a number of controls and tests have been implemented. 3.1. 418X Calibration Hart s radiometric calibration for the 418X is done with a KT19 that is calibrated with the cavity baths which are described in detail later in this paper. The 418X is set to a number of set-points. Its surface is measured by the KT19 with an emissivity setting of 0,95. This data is fed back to the 418X controller which makes the proper compensation, so that the display temperature will read within a given tolerance of the KT19 s calibrated temperature. Fig 1. Results of FTIR Testing. SM2008-M236-1197-2

There are a number of steps that have been taken to lessen uncertainties in the 418X calibration. First, the background temperature is controlled at near room temperature. Background temperature is defined as the temperature of any object facing the surface being measured [4]. This radiation can cause the apparent temperature of a surface to change. This is especially true at lower temperatures. Second, for both the KT19 calibration and the 418X calibration, scatter is controlled by using a cooled aperture that is controlled at a constant temperature. This temperature is close to room temperature. Testing has been done to verify the effect of KT19 scatter on measurements with Hart s cavities [6]. This test follows an ASTM guideline [5] for testing size of source. Third, the lower temperature range of the 4180 is 15 C. There are 2 calibration points below ambient, 15 C and 0 C. Any radiometric calibration done between 15 C and the dew point has the risk of causing dew or ice to form on the calibrator surface. To solve the problem with humidity at lower temperatures during calibration of the 4180, a purge system has been developed. This purge system involves enclosing everything between the KT19 and the 4180 s surface. This area is purged with a dry gas, so that a positive pressure is maintained within this system. The frost point of the air inside the enclosure is monitored to ensure it is well below the calibration temperature. 3.2. KT19 Transfer Standard As mentioned above, the KT19 is calibrated using Hart s cavity baths. The KT19 calibration uses the same calibration geometry that is used in the radiometric calibration of the 418X. This means that the calibration uses the same distance from KT19 to aperture, the same aperture size and the same controlled aperture temperatures. All of these values have tolerances and are accounted for in the KT19 and 418X uncertainty budgets. The KT19 is calibrated with the cavities at a number of set points. The temperature of the bath fluid during this calibration is monitored by a platinum resistance thermometer (PRT). 3.3. Self Consistency Checks To show the quality of Hart Scientific s radiometric measurement data, a number of tests have been performed to show that Hart s data is self consistent. First, there are a limited number of points within Hart s cavity bath temperature range that are covered by more than one bath. At those points, radiometric temperature has been compared between cavities. A summary of one such comparison done with a TRT is listed in Table 1. The 2 nd column lists the difference in radiometric temperature between the two cavities. The 3 rd column lists Hart s radiometric uncertainty for the cavity. Table 1. Comparison of radiometric temperature. Temperature Temperature Difference ( C) UHART ( C) 100 C 0.0173 0.1000 200 C 0.0814 0.1600 Second, a Heitronics TRTII (TRT) [7] has been used to measure Hart s cavities over their entire temperature range. This TRT is a dual band radiation thermometer. The data is curve-fit to the Sakuma-Hattori Equation [8]. This curve-fit is compared to Hart s uncertainties. A summary of this data is shown in Fig. 2. The error bars in these graphs represent the uncertainty with a coverage factor of 2 (k=2). Similarly, the transfer standard s calibration has been curve-fit to the Sakuma-Hattori Equation. It has been found that the KT19 calibration can be better curve-fit into a polyfunction. The Sakuma-Hattori Equation contains 3 parameters while the polyfunction uses 5. The calibration of the KT19 uses 7 points making the curve fit over-determined. Finally, the 418X calibration uses an overdetermined curve fit to calculate apparent temperature from the control sensor resistance. After the calibration is performed, the adjustments are verified by an as-left calibration. Additional testing has been done at Hart to evaluate temperature set-points between the calibration points. A summary of one such check is shown in Fig. 3. The 418X calibration uses 5 points (-15 C, 0 C, 50 C, 100 C and 120 C for the 4180 and 35 C, 100 C, 200 C, 350 C and 500 C for the 4181). The additional points from these tests are taken between the standard calibration points for these units. Note that no additional error is observed on the points between the calibration points. The SM2008-M236-1197-3

error bars indicate the 418X calibration uncertainty with a coverage factor of 2 (k=2). concerns of drift during transit. This necessitated constructing blackbody cavities for use on site. Residual ( C) 0.25 0.2 0.15 0.1 0.05 0-0.05-0.1 Sakuma-Hattori Residuals TRT 8-14µm B d Residuals 4.1. NIST Cone One of Hart s earliest designs of a cavity came from work done with NIST [9]. The NIST cone was a blackbody cavity with a spectral surface. The cone had an angle of 36,4 and was constructed of copper. The cone s temperature was maintained by a Hart Scientific model 6024 bath. It had a temperature range of 20 C to 200 C. Residual ( C) -0.15-0.2-0.25 0.5 0.4 0.3 0.2 0.1 0-0.1-0.2-50 0 50 100 150 200 250 SP ( C) Sakuma-Hattori Residuals TRT 3.9µm B d Residuals 4.2. IR Calibration Facilities at HS To facilitate the KT19 calibration, Hart Scientific has developed 3 cavity baths for use in Hart s new Infrared Calibration Laboratory. The cavity inside these baths is shown in Fig. 4. They are based on existing Hart models and use a cylinder-cone cavity that is 304 mm deep and 51 mm in diameter with a conical angle of 120. The cavities have emissivity greater than 0,999. This number was verified by modeling with STEEP3 [10-12] which is discussed later in this paper. -0.3-0.4-0.5 0 100 200 300 400 500 600 SP ( C) Bath Fluid Purge Tube Fig. 2. Residuals from Sakuma-Hattori Curve-Fit of TRT Data. Blackbody Cavity 4181 Calibration Error 1.000 Cal Error (C) 0.800 0.600 0.400 0.200 0.000-0.200-0.400-0.600-0.800-1.000 0 100 200 300 400 500 Set Point (C) Error Aperture (Water Cooled) Fig 4. Blackbody cavity in bath. Hart s cavity bath fluid temperature is monitored by Hart Scientific Model 5626 PRTs that are calibrated in Hart s Primary Calibration Laboratory. Their calibration is traceable to NIST. An illustration of this traceability is shown in Fig. 5. Fig 3. Error at Points between Calibration Points 4. DEVELOPMENT OF CAVITY BATHS Part of Hart Scientific s radiometric calibration of its flat plates necessitated a calibrated transfer standard, the KT19. It was determined to be more favorable to have the transfer standard calibrated on site rather than rely on an outside laboratory to calibrate this instrument. This was mainly due to 4.3. STEEP3 Modeling To gain better temperature uniformity on the cavity walls, the end of the cavity is purged with dried shop air that is heated to the bath temperature by a heat exchanger in the bath fluid. Hart has done experimentation to measure the temperature on the cavity walls by contact thermometry. The results from these tests are used in the STEEP3 model. This model reveals the results in Table 2. This table shows the effective emissivity at various temperatures plus the isothermal emissivity. SM2008-M236-1197-4

NIST FP CELLS HART CL1 KT19 5626 HART IR CAVITIES 418X KEY: CONTACT TRACEABILITY RADIOMETRIC TRACEABILITY Fig 5. Radiometric traceability at hart. Table 2. Effective Emissivty of Hart s Cavity Bath s Bath Bath Temperature Effective Emissivity LT & Isothermal 0.9996 MT LT 15 C 0.9998 LT 0 C 0.9997 LT 50 C 0.9995 LT 100 C 0.9994 HT Isothermal 0.9996 HT 200 C 0.9993 HT 350 C 0.9992 HT 500 C 0.9991 To verify Hart s cavity modeling technique, STEEP3 was used to model cavities in a number of published papers [13-15]. 5. HART S IR UNCERTAINTY BUDGETS Hart s IR uncertainty budgets are the result of much research into the uncertainties related to IR thermometry. The determination and calculation of these uncertainties involves a complex process of calculation and experimentation. 5.1. Determination of Uncertainties The determination of Hart s uncertainties is an involved process. Where possible, experimentation has been performed to provide the uncertainty budgets with type A uncertainties. Where experimentation is not possible, modeling has been performed to provide knowledge of the uncertainty. Two examples of this are the determination of cavity effects and determination of the effects of aperture temperature. Cavity effects are the effects of the cavity not acting as a perfect blackbody. These effects are evaluated in STEEP3. The effects of aperture temperature were evaluated by experimentation. The aperture temperature was varied, and the change in radiometric temperature was noted at several temperatures. The knowledge of the change in radiometric temperature versus the change in aperture temperature was used to calculate the effect of aperture temperature uncertainty in the KT19 uncertainty budget. 5.2. Evaluation of Uncertainties As suggested by uncertainty budget guidelines [16], uncertainties are evaluated using the system s measurement equation. This measurement equation is based on Planck s Law evaluated over the 8 µm 14 µm band [6]. This implementation of the measurement equation takes into account the effects IR thermometer s spectral response and considers the calibrator s emissivity, aperture geometry and background temperature. This implementation of Planck s Law is difficult to solve. Previous publications suggest the Sakuma-Hattori Equation or a narrow band form of Planck s Law should be used [8, 17]. In the case of the 418X, these were not used because of the wide band effects of emissivity and spectral response. 6. CONCLUSION Developments behind the 418X project have resulted in a new measurement capability for Fluke - Hart Scientific. A radiometric calibration has been established as the standard calibration for these units. As a result, an infrastructure to support this calibration has been developed. This infrastructure includes radiometric transfer standards and blackbody cavities to support the 418X calibration. These developments plus education of the user will help to bring more accuracy and metrology to lower temperature IR thermometry. ACKNOWLEDGMENTS The author greatly appreciates the help of many people in developing this technique in calculating IR thermometry uncertainties. Among the people who have helped are Tom Wiandt, Mike Hirst and Rick Walker of Fluke Corporation Hart Scientific Division; Ingrid Gooding of Fluke; Patrick Parker, Alex Nikittin, Steve King, Medwin Schreher, Thomas Heinke and Reno Gärtner of Raytek; Leonard Hanssen, Benjamin Tsai, Sergey Mekhontsev and Howard SM2008-M236-1197-5

Yoon of NIST and Kendall Johnson of the Space Dynamics Laboratory at Utah State University. REFERENCES [1] F. Liebmann, Infrared Thermometer Calibrator Development at Fluke Corporation Hart Scientific Division, Proceedings of NCSLI 2007. [2] L. M. Hanssen, A. V. Prokhorov, V. B. Khromchenko, S. N. Mekhontsev, Comparison of Direct and Indirect Methods of Spectral Infrared Emittance Measurement, Proceedings of TEMPMEKO 2004, pp. 539-544, 2004. [3] L. M. Hanssen, V. B. Khromchenko, S. N. Mekhontsev, Infrared Spectral Emissivity Characterization Facility at NIST, Proceedings of SPIE, 5405, pp. 1-12, 2004. [4] D. P. DeWitt G. D. Nutter, Theory and Practice of Radiation Thermometry, Wiley Interscience, New York, 1988, pp. 41-43, 72-73, 194. [5] 1256 95, Annual Book of ASTM Standards Vol. 14.03, ASTM International, West Conshohocken, PA, 2005, pp. 490-491. [6] F. Liebmann, Determining Size Of Source For Handheld Infrared Thermometers Theory And Practice, Proceedings of MSC 2008. [7] O. Struß, Transfer Radiation Thermometer Covering the Temperature Range from -50 C to 1000 C, AIP Conference Proceedings, Vol. 684, pp. 565-570, 2003. [8] F. Sakuma, M. Kobayashi, Interpolation Equations of Scales of Radiation Thermometers, Proceedings of TEMPMEKO 96, pp. 305-310. [9] J. B. Fowler, A Third Generation Water Bath Based Blackbody Source, Journal of Research of the National Institute of Standards and Technology, Vol. 100, Num. 5, pp. 591-599, 1995. [10] A. V. Prokhorov, L. M. Hannsen, Effective Emissivity of a Cylindrical Cavity with an Inclined Bottom, Metrologia, Vol. 41, pp. 421-431, 2004. [11] V. I. Sapritsky, A. V. Prokhorov, Spectral Effective Emissivities of Nonisothermal Cavities Calculated by the Monte Carlo Method, Applied Optics, Vol. 34, No. 25, pp. 5645-5652, 1995. [12] A. V. Prokhorov, Monte Carlo Method in Optical Radiometry, Metrologia, Vol. 35, pp. 465-471, 1998. [13] J. Ishii, M. Kobayashi, F. Sakuma, Effective Emissivities of Black-body Cavities with Grooved Cylinders, Metrologia, Vol. 35, pp. 175-180, 1998. [14] F. Sakamura, L. Ma, Cavity Emissivity of Fixed- Point Blackbody, SICE 2003 Annual Conference, Vol. 2, pp. 2187-2190, 2003. [15] C. K. Ma, Method for the Measurement of the Effective Emissivity of a Cavity, Proceedings of TEMPMEKO 2004, pp. 575-580, 2004. [16] U. S. Guide to the Expression of Uncertainty in Measurement, John A. Wehrmeyer, Chair (National Conference of Standards Laboratories, Boulder, CO, 1997), pp. 4-8. [17] Saunders, Fischer, Sadli, Battuello, Park, Yuan, Yoon, Li, van der Ham, Sakuma, Ishii, Ballico, Machin, Fox, Hollandt, Matveyev, Bloembergen, Ugur, Uncertainty Budgets For Calibration Of Radiation Thermometers Below The Silver Point, Paper presented at TEMPMEKO 2007, not yet published SM2008-M236-1197-6