Portable type TXRF analyzer: Ourstex 200TX

Similar documents
IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

Why do I chose this topic? By: Medani Sangroula 12/6/2013 1

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

Praktikum zur. Materialanalytik

OBSERVATION OF SURFACE DISTRIBUTION OF PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY DURING D 2 GAS PERMEATION THROUGH PD COMPLEXES

Development of a total reflection X-ray fluorescence spectrometer for ultra-trace element analysis

FEASIBILITY OF IN SITU TXRF

S8 TIGER Series 2 for ASTM D 6443

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

DEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

INFLUENCE OF THE SAMPLE MORPHOLOGY ON TOTAL REFLECTION X-RAY FLUORESCENCE ANALYSIS

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM

Repeatability of Spectral Intensity Using an Auger Electron Spectroscopy Instrument Equipped with a Cylindrical Mirror Analyzer

S2 PICOFOX. Innovation with Integrity. Benchtop TXRF Spectrometer TXRF

Auger Electron Spectroscopy (AES)

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

Overview of X-Ray Fluorescence Analysis

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

Elemental analysis by X-ray f luorescence. High performance EDXRF elemental analyzer

Development of a new MeV gamma-ray camera

Secondary ion mass spectrometry (SIMS)

Application of the GD-Profiler 2 to the PV domain

CHEM-E5225 :Electron Microscopy X-Ray Spectrometry

FINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS

Structural Characterization of Giant Magnetoresistance Multilayers with New Grazing Incidence X-ray Fluorescence

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Development of a Compact XAFS Measurement Chamber under Atmospheric Pressure in the Soft X-ray Region

2D XRD Imaging by Projection-Type X-Ray Microscope

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source

Secondary ion mass spectrometry (SIMS)

X-ray fluorescence analysis - calibration of the X-ray energy detector

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Defining quality standards for the analysis of solid samples

FEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada.

OES - Optical Emission Spectrometer 2000

Development of Laser Thickness Gauge in Steel Plate Shearing Line

Electrochemical Cell for in-situ XAFS Measurements

Chemical State Analysis of SiO 2 /Si by Wavelength-Dispersive X-Ray Fluorescence

Extrel Application Note

Advances in Field-Portable XRF

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

A New Cross-Shaped Graphite Furnace with Ballast Body for Reduction of Interferences in Atomic Absorption Spectrometry

X-Ray Fluorescence and Natural History

Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

PROCESS MONITORING OF PLASMA ELECTROLYTIC OXIDATION J.-W. Liaw, C.-C. Hsiao, Clinton Fong, Y.-L. Tsai, S.-C. Chung, Oleg Demin Materials Research

Xylan Adsorption on Cellulose Model Films & Influence on Bond Strength

ICP-Mass Spectrometer

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

PRINCIPLE OF ICP- AES

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

Ground Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2

1+2 on GHD (5 µl) Volume 1+2 (µl) 1 on GHD 1+2 on GHD

Sample Preparation for Evaluation of Detection Limits in X-ray Fluorescence Spectrometry

TRITIUM PRODUCTION IN PALLADIUM DEUTERIDE/HYDRIDE IN EVACUATED CHAMBER

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Shorten your Time-to-Result

AS 101: Day Lab #2 Summer Spectroscopy

FINE PATTERN ETCHING OF SILICON USING SR-ASSISTED IONIZATION OF CF4 GAS

Pulse Height Analysis System (PHA) designed for W7-X Presented by Monika KUBKOWSKA

Iodine-Mediated Chemical Vapor Deposition Growth of Metastable Transition Metal

Supplementary Information for. Silver Nanoparticles Embedded Anti-microbial Paints Based on Vegetable Oil

A Simple Multi-Turn Time of Flight Mass Spectrometer MULTUM II

A DIVISION OF ULVAC-PHI

Multi-element process analyzer

In-Situ FTIR Spectroscopy and Metrology of a Tungsten CVD Process

Diffractometer. Geometry Optics Detectors

Supporting Information

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

ARC SPECTRUM OF IRON /COPPER / BRASS

photo-mineralization of 2-propanol under visible light irradiation

The scanning microbeam PIXE analysis facility at NIRS

Surface Analysis. Dr. Lynn Fuller Dr. Fuller s Webpage:

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

Metal-organic frameworks (MOFs) as precursors towards TiO x /C. composites for photodegradation of organic dye

Transcription:

Excerpted from Adv. X-Ray. Chem. Anal., Japan: 42, pp. 115-123 (2011) H. Nagai, Y. Nakajima, S. Kunimura, J. Kawai Improvement in Sensitivity and Quantification by Using a Portable Total Reflection X-Ray Fluorescence Spectrometer Equipped with a Silicon Drift Detector We have adapted a silicon drift detector (SDD) to a portable total reflection X-ray fluorescence (TXRF) spectrometer using polychromatic X-rays from an X-ray source of single-digit watt. Analytical performance was greatly improved in PB ratio and the sensitivity for lighter elements when compared with the former spectrometer equipped with a Si-PIN detector. In this article improved sensitivity of the new spectrometer is described, and additionally the evaluation of the current wave-guide, a newly developed sample preparation device for precise quantitative evaluation and an application to the analysis of suspended particles are reported. Fig. 1 Photograph of portable TXRF spectrometer, OURSTEX 200TX Portable type TXRF analyzer: Ourstex 200TX This analyzer is a compact type, only 8kg weight. It is easy to be movable at on-site measurement, requires AC100V power. Fig. 2 shows its structural photo. For the X-ray tube, transparent compact X-ray tube is used with 50kV and 200µA of maximum voltage and tube current, respectively. Tungsten is used as the target of the X-ray tube. Electronic cooling type Silicon Drift Detector (SDD, light reception area; 7mm 2, Be thickness; 8µm) is used as the detector. The SDD is composed of united structure of Si-PIN with FET, and ring-shaped electrode gradient on the surface. It can be effectively achieved to drift the electron to the centered anode. By this reason, decrease of thermal noise, higher improvement of electron collecting effect and decrease of leak current etc. are attainable. This improvement contributes to obtain higher energy resolution ability, and PB ratio becomes bigger. Hence, the measurement of high calculating ratio is achievable without decreasing the resolution ability. In general, the wider the light reception area is, the lower the resolution ability is in case of the Si-PIN detector, while in case of the SDD, it has a great capability that it is almost no lowering of energy resolution ability. On the contrary, it has a certain demerit that higher energy X-ray may be easy to transparent because the crystal thickness is thin as in 450µm, so the detecting efficiency is not too high at the high energy area. X-ray wave-guide is used for achieving paralleled of incident X-ray illuminating from the X-ray tube. X-ray wave-guide is a photo-element that enables to implement paralleled of incident X-ray by using 2 pieces of silicon wafers as shown in Fig. 3 (a). As a part of incident X-ray is totally reflected by using the mirror-face silicon wafer, stronger X-ray is obtained than using usual slit. The interval between the silicon wafers is designed and made as 10mm horizontal width and 10µm vertical width 3). Fig. 3 (b) shows the outlook of the X-ray wave-guide. 1 / 5 Fig. 2 Photograph of portable TXRF system Fig. 3 (a) Schematic drawing of the wave-guide device (b) Photograph of the wave-guide device

Produced wave-guide has a possibility of variation of the slit width by affecting the manufacturing fabrication accuracy and the edge that was cut the silicon wafer. Therefore stainless steel cutter is set at the outlet of the X-ray wave-guide, and by knife-edge scanning method, the evaluation of the slit width of the X-ray wave-guide has implemented. Evaluation of the slit width was determined by obtained data that were differentiated and the full width half maximum (FWHM) value that was calculated by Gaussian fitting. The result is shown in Fig. 4. Calculated slit width was estimated as 16.5µm, which was close to 10µm that was assumed at the initial designing time. Quartz optical flat was used in the sample plate (accuracy of reflection wave surface: λ/20). Fig. 4 A profile of the measured edge-scan, and the evaluated beam profile of the wave-guide device Incident angle was adjusted as setting at ca. 0.05 that is most suitable for transition metals, by the Goniometer that is installed at under the sample plate 6). PB (Peak-Background) ratio improvement by adopting SDD SDD was used in this analyzer. Energy resolution ability of the SDD is 132eV in the FWHM of Mn-K α ray. (Formerly used Si-PIN detector s resolution ability is 180eV in the FWHM of the same ray) Comparison of spectrum between the Si-PIN detector and the SDD is shown in Fig. 5. Spectrum is standardized by the peak intensity of W-L α ray that is the excitation source. Measuring sample: Sc, Cr, Co, As and Sr are in 0.5ppm aqueous solution respectively. The solution is dropped on the optical flat as 2mL and it is dried (1ng). As a measuring result, energy resolution ability as well as the PB ratio in the SDD was more improved than those in the Si-PIN detector. Fig. 6 illustrates respective PB ratio. In Co, PB ratio in the Si-PIN detector is 0.921, while 1.796 on the SDD is detected, almost 2 times higher improved. In the Si-PIN detector, it was impossible to separate the peaks between Ar-K α and Ar-K β ray, and no detect for very little peaks; The SDD enabled to detect the separation of the peaks between Ar-K α and Ar-K β ray, and detect very little peaks effectively. Lower limit of detection (LLD) values estimated from obtained data for Sc, Cr, Co, As and Sr in 1800 seconds measurement are, 29pg, 20pg, 14pg, 51pg and 192pg, respectively. The following theoretical formula was used for the calculation of the LLD value 7). Fig. 5 Total reflection X-ray fluorescence spectra of a dried droplet containing 1ng of Sc, Cr, Co, As and Sr. The data acquisition time was 1800s, and the detector used were a) Si-PIN and b) SDD. : Peak from collimator material of SDD where W: content (ng), t: measuring time (sec), I NET : Net intensity of the peak (cps), I BG : background intensity (cps) Fig. 6 A comparison of PB ratios for various elements obtained with the Si-PIN detector and with the SDD detector 2 / 5

The LLD value of Co obtained this time was the same value as in previous reported one 6) by Kunimura et al. For this reason, as the reception light surface area is 7mm 2 in the SDD, X-ray intensity was lower than the Si-PIN detector, which has the reception light surface area of 13mm 2. It depends on the droplet size though, significant detection sensitivity may be obtained by using bigger caliber of the SDD. Quantitative improvement by using the droplet supporting device On the analysis of the total reflection X-ray fluorescence spectrometry, sample preparation is the most important factor in evaluating the analysis quantitatively. In this analyzer, sample solution is dropped on the optical flat and its dries residual is measured. It is quite important that the sample solution should be dropped on the center position of the optical flat precisely, in order not to lower the quantitative accuracy by changing the X-ray intensity significantly by the relationship with the taking-un angle of the detector. We fabricated the droplet supporting device such as in Fig. 7. The device has a small hole that enables to insert the point of the micro-pipette tip and it can insert always at a constant depth. Hence, it enables to drop the solution always at the center part of the optical flat, and the quantitative accuracy improves much higher. The material used in the device is acryl resin with consideration of the contamination, and the device is reusable by washing treatment. To verify the effect of using the droplet supporting device, two conditions were compared: Measuring the change of the X-ray intensity of Cr-K α in Cr standard solution prepared as in 10ppm and drops 5mL: 1) with the droplet supporting device 2) without the device The results are illustrated in Table 1. Compared to the case of without the device, the change of the X-ray intensity shows smaller in the case of using the device. Fig. 7 (a) Schematic drawing of the sample preparation support device. (b) Photograph of the sample preparation support device. In order to evaluate quantitative verification of the analyzer, drawing the calibration curves of Co and Cr was executed. In Fig. 8, the calibration curves obtained from the experiment are shown. Sample solution preparation: Standard solutions for atomic absorption spectrometry with Cr and Co were diluted and adjusted as in 500ppb, 100ppb, 50ppb, 25ppb and 10ppb, respectively. Also, constant amount of Sc solution as the internal standard is added and 10µL of the solution is dropped on the optical flat. The sample is dried on it, and is measured. As a result, good linearity relation is obtained both for Cr and Co, and it can be confirmed that sufficient quantitativity is approved even in the lower concentration range 3 / 5 Fig. 8 Calibration curve of chromium and cobalt

Sensitivity improvement for light elements In order to confirm analytical sensitivity of the analyzer in the light elements field, P analysis in the solution was implemented. With the formerly used Si-PIN detector, it has lower energy resolution ability that the peak of P-K α ray is hidden in the tailing part of the peak of Si-K α ray that is due to the optical flat. So trace amount of P analysis was not achievable. Now the experiment for the analyzer with the SDD is implemented whether trace amount of P analysis is achievable or not. Standard P solution is diluted as in 1ppm, 0.5ppm and 0.1ppm, respectively. As the measuring sample, 5mL of it is dropped on the optical flat, and after drying measurement is executed. Absolute amount of sample is 5ng, 2.5ng and 0.5ng, respectively. Measured P spectra are shown in Fig. 9. It can be confirmed that even in 0.5ng concentration range, sufficient detection is achievable. The LLD value estimated by 0.5ng peak intensity and the background intensity with the formula (1) becomes 286pg in 1800 second measurement. It can be confirmed that high sensitive measurement is attainable even in the light element field in air atmospheric environment. Fig. 9 Total reflection X-ray fluorescence spectra of phosphorus This achievement could be concluded as: 1) Energy resolution ability improves much higher by the SDD adoption 2) On the structure of total reflection X-ray fluorescence spectrometer, the distance between the sample and detector is extremely short; the detector is located on upper of the sample; and it is no affection of the absorption by the polymer film, etc. that is used as sample holding in usual fluorescence X-ray spectrometer. Application field As the examples of application fields measured by a portable total reflection X-ray fluorescence spectrometer, there are bunch of reports as in measured soil eluding water 8), leachate from tea or toys, etc 9). In this report, the possibility for element analysis in air particulate matter is examined. Optical flat is on the Petri dish and it is left at the sampling place 1 hour remains. Accumulating particles on the optical flat is directly measured. Sampling place is along national route in Neyagawa City, Osaka Pref., that is relatively high traffic-laden street. Measured result spectra are shown in Fig. 10. Compared to the blank, the elements of Al, S, K, Ca, Ti, Cr, Mn, Fe, Zn etc. were detected. Most likely, as the sampling place is high traffic-laden street, elements from automobile exhaust gas and frictional metal by the break pad etc. may be detected presumably. Fig. 10 Total reflection X-ray fluorescence spectra of suspended particulate matter. (Collected place of sample: Neyagawa City, Osaka, Japan) By improving the sampling method, this analyzer might be utilized at the application of environmental monitoring of airborne particulate matter. 4 / 5

Conclusion By loading the SDD as the detector of a portable total reflection X-ray fluorescence spectrometer utilizing non-color X-ray generating from small X-ray tube in a few watt, energy resolution ability is improved, and compared to the former used Si-PIN detector, PB ratio improves about 2 times higher. Moreover, analysis sensitivity of P in light element field is improved; it is obtained 286ng as the lower limit detection. Stable sample preparation is achievable by utilizing the droplet sample supporting device; and it enabled to improve quantitative accuracy drastically. This analyzer is compact and light-weighted, so it may be possible to implement measurements in case of having difficulty of sample storage and/or transferring the sample, such as the air particulate matter at site (ex. at Antarctic), as well as by measuring the hair of the cattle at site, cattle s health condition may be determined as routinely. References 1) Y. Yoneda, T. Horiuchi: Rev. Sci. Instrum., 42, 1069 (1971) 2) A. Iida, Y. Gohshi: Jpn. J. Appl. Phys., 23, 1543 (1984) 3) S. Kunimura, J. Kawai: Anal. Chem., 79, 2593 (2007) 4) S. Kunimura, J. Kawai: Chemistry and Industry, 61, 1050 (2008) 5) ditto: Analytical Chemistry, 58, 1041 (2009) 6) ditto: Adv. X-Ray. Chem. Anal., Japan, 41, 29 (2010) 7) I. Nakai: Fluorescence X-ray analysis (2005), Asakura Publishing Co. 8) S. Kunimura, et al.: Adv. X-Ray. Chem. Anal., Japan, 38, 367 (2007) 9) S. Kunimura, et al.: Analytical Chemistry, 57, 135 (2008) 5 / 5